Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Similar documents
CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

IMRT in clinical practice at the UMC-Utrecht. Utrecht

SnapShot IMRT with compensators and FFF beams

Clinical Implementation of Volumetric Modulated Arc Therapy

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

Math Review. Overview

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

THE development of more advanced techniques in radiotherapy,

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Constancy checks of well-type ionization chambers with external-beam radiation units

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Quality control for Intensity-modulated radiation therapy

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Monitor Unit Verification for Small Fields

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

Quality Management Determined from Risk Assessment

SIGMA MASCOT. Product Hand Assembly made easy. PCB Hand Assembly made easy

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Automatic treatment planning an MCO perspective

Review of fundamental photon dosimetry quantities

Proton Therapy QA Tools

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Torrild - WindSIM Case study

Surface-based Respiratory Motion Classification and Verification

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Surface buildup dose dependence on photon field delivery technique for IMRT

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

Dynamic wedges dosimetry and quality control

Real-Time Smoothness Measurements on Concrete Pavements During Construction

A beam-matching concept for medical linear accelerators

Technical Note. Determining the surface tension of liquids by measurements on pendant drops

INTRODUCTION TO PATTERN RECOGNITION

RESPIRATORY MOTION IN RADIATION THERAPY

Phase I-II of the Minnesota Highway Safety Manual Calibration. 1. Scope of Calibration

Introduction. Materials and Methods

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Scientific Journal of Nuclear Medicine & Radiation Therapy

ALMA Status Update. October ALMA Doc 1.17, ver. 1.1 October 02, 2013

Accuracy and Precision of High-Speed Field Measurements of Pavement Surface Rutting and Cracking

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Pendant Drop Measurements

MIKE NET AND RELNET: WHICH APPROACH TO RELIABILITY ANALYSIS IS BETTER?

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Impact of Lead Free on Automated X-Ray Inspection

Dosimetric Calculations. Lonny Trestrail

CORRELATION BETWEEN SONAR ECHOES AND SEA BOTTOM TOPOGRAPHY

Evaluation of six TPS algorithms in computing entrance and exit doses

Planning Assignment (3 field rectum)

BASKETBALL PREDICTION ANALYSIS OF MARCH MADNESS GAMES CHRIS TSENG YIBO WANG

Modeling of Hydraulic Hose Paths

WindProspector TM Lockheed Martin Corporation

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Iranian Journal of Medical Physics

technocurling delivery stick

If a fair coin is tossed 10 times, what will we see? 24.61% 20.51% 20.51% 11.72% 11.72% 4.39% 4.39% 0.98% 0.98% 0.098% 0.098%

Traffic Parameter Methods for Surrogate Safety Comparative Study of Three Non-Intrusive Sensor Technologies

Commissioning of a new total body irradiation protocol

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

955730_1 4/17/18. FlowSense Operator s Guide For Gen2 20/20 SeedSense Displays

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

How to Design Medical Accelerator Systems for Reliability: IBA PT System

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

The ESS neutron bunker design, materials and timeline.

Software Engineering. M Umair.

Anthropogenic Noise and the Marine Environment

PROJECT CYCLOPS: THE WAY FORWARD IN POWER CURVE MEASUREMENTS?

Recommendations for the Risk Assessment of Buffer Stops and End Impact Walls

Junction design rules for urban traffic networks

Bayesian Optimized Random Forest for Movement Classification with Smartphones

Best Practice RBI Technology Process by SVT-PP SIMTECH

Open Research Online The Open University s repository of research publications and other research outputs

Zoubir Ouhib Lynn Cancer Institute

ENVIRONMENTALLY ADAPTIVE SONAR

Building an NFL performance metric

New Technology used in sports. By:- ABKASH AGARWAL REGD NO BRANCH CSE(A)

Relative Dosimetry. Photons

SEMMME3-073 Checking that completed assemblies comply with specification

LaserSnake Underwater Laser Cutting Phase 1 TWI

Decision Trees. Nicholas Ruozzi University of Texas at Dallas. Based on the slides of Vibhav Gogate and David Sontag

Gripping rotary modules

SEMDAC MSC support and surveillance in Chile s EEZ, Desventuradas Islands and Easter Island. Laura Fontan Bouzas Fisheries Analyst, OceanMind

Transcription:

Treatment plan complexity metrics for predicting IMRT pretreatment quality assurance results Scott Crowe, QUT

Pre-treatment QA Verification of fluence delivery to ensure that dose can be delivered accurately To identify: if leaf sequence doesn t produce expected fluence if plan isn t being correctly transferred to linac if treatment isn t being delivered correctly What s popular for contemporary treatments? comparing 2D/3D measurements against predictions

Genesis QLD? Current procedure involves evaluation of predicted portal dose against EPID images Therapists responsible for portal image acquisition (after RO approval of plan) Physicists responsible for analysis of results, where: pass for IMRT = mean GAI > 90% and no beam GAI < 85% pass for RapidArc = no beam GAI < 90% both calculated using 3%, 3mm done within the Epiqa software GAI = gamma agreement index, % of (sig.) points in portal image that agree with predicted dose Past procedures included MapCheck system

Deliverability What are some possible limitations could prevent the linac from delivering the calculated dose? limitations in the machine leaf movement, dose rate, gantry movement, etc. limitations in the software fluence optimisation, dose calculation, etc. Deliverability is related to complexity/modulation Cost functions are used to control complexity in inverse planning algorithms

Potential for time saving Since pre-treatment QA results correlate with complexity... Identifying overly complex plans before QA could potentially save time. Why bother with the QA if you expect the plan will be sent back for adjusting anyway?

Established measures Modulation index deviations between adjacent bixels against standard deviation of fluence map Fluence map complexity deviations between adjacent bixels against sum of each bixel Modulation complexity score indication of planned mobility of leaves in the beam (aperture area variations between control points and differences between adjacent leaf positions)

Novel measures Novel metrics added to TADA code: Mean field area segment weighted by MU Mean aperture displacement mean of difference between centre of beam and centre of leaf apertures Cross axis score % of leaf pairs blocking central axis Closed leaf score % of leaf pairs in field closed could indicate problem with transmission modelling Small aperture score percentage of leaf pair apertures below a threshold (i.e. 1mm, 5mm) could indicate limits of dose calculation accuracy and others

MapCheck results Analysis of 122 prostate beams, planned in BrainLab Modulation index and small aperture score significantly correlated to QA pass rates SAS had useful threshold (beams with SAS(5mm)<0.27 all passed)

Epiqa results Analysis of IMRT and Rapidarc Plans over multiple treatment sites. Modulation index distinguished between treatments for different sites more complex: anus/rectum, endometrium less complex: prostate and brain

Epiqa results Strong correlation for many metrics for IMRT treatments Weak correlation for some metrics for RapidArc treatments, even when split up into sub-arcs. CAS / MAD: interleaf leakage not modelled in Eclipse, which has a greater effect for IMRT, as collimator angles chosen for RapidArc to minimise effect

Other observations arcs Representative sample of arcs re-qa d after beams split into 6 60 sub arcs: sub-arc GAI values significantly lower

Other observations GAI 2%,2mm & 3%,3mm were both used in Epiqa GAI(2%,2mm) 1.08 GAI(3%,3mm)

References Crowe et al. 2014. Treatment plan complexity metrics for predicting IMRT pre-treatment quality assurance results. Australas Phys Eng Sci Med (in press) Kairn et al. 2014. Predicting the likelihood of QA failure using treatment plan accuracy metrics. J Phys Conf Ser 489(1), 012051 Kairn et al. 2014. Experienced-based management of IMRT quality assurance. Upcoming CSM poster!