OPERATOR S MANUAL. Pediatric/Infant cap-one Mask YG-232T/YG-242T A. General. Safety Information WARNING WARNING WARNING. Components WARNING

Similar documents
English. ivolve Full Face Mask. User Manual

ATI Skills Modules Checklist for Oxygen Therapy

EasyFit SilkGel Full Face Mask

Micro Dial-Flowmeter. Instructions for Use

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

VENTI-O 2 plus. O 2 switching valve for: VENTImotion from serial number 4000 VENTIlogic from serial number 4000 VENTImotion 2

Dialflow Regulator. Instructions for Use

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

Instructions for Use. OptiLife Nasal Lab Mask

Oxygen Dialflow Meter. Instructions for Use

Med Aire Alternating Pressure Pump and Pad System

Mirage Vista Nasal Mask

AUTOVENT 4000 VENTILATOR

Oxygen Dialflow Meter. Instructions for Use

REF CH-302B IDENTIFICATION OF PARTS

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

Oxygen Delivery Systems:

O 2 Star - Oxygen Therapy Accessories Consumables and Accessories

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Rev. B. Operating Instructions

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

Oxygen Therapy Devices

English Español Português Italiano Deutch Français INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR MODEL CH-656C. - Eng 1 -

Mirage FX. User Guide. English NASAL MASK

SC505 Sof Care Inflator. Operating Instructions and Service Manual

Hamilton accessories CONTENT

ISO INTERNATIONAL STANDARD. Sleep apnoea breathing therapy Part 2: Masks and application accessories

INSTRUCTIONS. Sterilization lot number

PULSOX -2 INSTRUCTION MANUAL

NOTES BEFORE USE BE SURE TO READ THIS MANUAL BEFORE USING THE HELMET. KEEP THE MANUAL IN A SAFE PLACE FOR FUTURE REFERENCE.

Stroller. Sprint S T R O L L E R S P R I N T

Oxygen Demand Valve. Instructions for Use

Instruction Manual for Configura Cushionair Portable Pump

.$2;,)*3/6$ USER MANUAL

U S E R M A N U A L. (CGA 870 Connection)

Patient Sleep Test Hookup

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

MODEL CH-432B INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR. - Eng 1 - English Español Português Deutch. Italiano Français CH-432B

HELiOSTM Plus & Marathon

ASAHI Neurovascular Guide Wire

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

MACS Mask CPAP System

PRODUCT NUMBER: Rd40 Aluminium Canister K2

Mirage Micro. User Guide. English. nasal mask

Oxygen Saturation Monitor PULSOX -2. Instruction Manual

USER MANUAL. AIR-OXYGEN BLENDER (DISS and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS

U S E R M A N U A L CAUTION. Models: 19MFA Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Anesthesia Breathing Circuit

Paradise Pump Series Operator s Manual

O-Two Self-Study Guide. Equinox Relieve

REF CH-602B IDENTIFICATION OF PARTS

Complex Care Hub Manual: Oxygen Therapy

Model UB-328. Wrist Digital Blood Pressure Monitor. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Great products available through ROI!

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

C-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8

Swift LT NASAL PILLOWS SYSTEM. User Guide English

Pressure Regulators. Instructions for Use

Hi Flow Stroller H I F L O W S T R O L L E R

First and Only Disposable Infant T-Piece Resuscitator. Because they are not miniature adults...

INTRODUCING CARDIOSAVE Hybrid MORE THAN A PUMP. A REVOLUTION. SPECIFICATIONS

TITLE: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION DATE: 10/18/12 EMPLOYEE NAME. DATE INITIAL RE-EVALUATION ANNUAL (if required) PRN

Alternating Pressure / Low Air Loss Mattress System

UsER manual for Watersens ph -REDOX

Oxygen Concentrator Instruction

Domiciliary Oxygen Switch. Instructions for Use

Servo UNIVERSAL ADVANCED VENTILATION PLATFORM ALL PATIENT CATEGORIES

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

e.do Gripper Safety requirements, technical features and assembly and integration instructions

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

EasyLife INSTRUCTIONS FOR USE. Respironics Inc Murry Ridge Lane Murrysville, PA USA JDH 3/25/10

Invacare Twilight II Nasal Mask

Home Oxygen Therapy Patient Handbook

Pressure Regulator. Instructions for Use

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Checklist for care of Respirators

Use this key to determine which sections of this Product Manual apply to your job.

M5400 Operators Manual

2010 Koninklijke Philips Electronics N.V. All rights reserved. INSTRUCTIONS FOR USE

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

Resusci Anne Simulator

User Manual. Hopkins IMPACT Digital Wrist BP RELATED ITEMS. Item #526014

TJF-Q180V Cleaning and Disinfection Checklist

Air Plus User Manual

User Guide Antidecubitus systems EURO BASIC

TJF-Q180V Cleaning and Disinfection Checklist

Corporate Overview and Product Summary

Welcome to Maxstar Medical

Welcome to the Specialized Medical Services Respiratory training webinar series!

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

English USER GUIDE User guide

Instruction Manual Mode d emploi Gebrauchsanweisung Manuale di istruzioni Manual de instrucciones Gebruiksaanwijzing РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

DM-WU (English) Dealer's Manual. Wireless Unit SM-EWW01

Model 405 &407 Small Animal Ventilator. User Manual v3.5

Operating Manual. 3S-R AirElite Full Face Mask Advantage AirElite Full Face Mask Advantage 4000 AirElite Full Face Mask. AirElite - Full Face Masks

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

OC4000 USER'S GUIDE I O OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY. by Supera Anesthesia Innovations GLOSSARY OF SYMBOLS. : ON (power switched on)

CPAP Devices - Masks - Accessories for respiratory assistance in CPAP

Transcription:

OPERATOR S MANUAL 0614-905493A First Edition: 12 Dec 2011 Second Edition: 12 Feb 2015 Printed: Pediatric/Infant cap-one Mask YG-232T/YG-242T Symbol Description Attention, consult operator s manual Date of manufacture YG-232T Pediatric cap-one Mask CO 2 sensor attachment Oxygen connector Do not reuse Expiration date Lot number The CE mark is a protected conformity mark of the European Community. Products marked with this symbol comply with the requirements of the Medical Device Directive 93/42/EEC. Transparent film Strap YG-242T Infant cap-one Mask Oxygen cannula tube General The YG-232T pediatric cap-one mask and YG-242T infant cap-one mask are for suppling air or oxygen to the patient. The cap-one mask is used with the TG-970P CO 2 sensor kit to measure the partial pressure of the expired CO 2. Refer to the operator s manual of the CO 2 sensor kit together with this operator s manual. Model YG-232T YG-242T Applicable Patients For children 20 to 40 kg For infants 7 to 20 kg Dead Space 3.5 ml 2.5 ml Qty 10 10 Supply Code V933 V935 The applicable patient weight is for reference only. The cap-one mask might not be appropriate for some ventilation conditions. Components Mask: PVC Oxygen cannula tube: PVC Safety Information Use the cap-one mask only for supplying oxygen and attaching the CO 2 sensor kit to measure the partial pressure of the expired CO 2. Never use the cap-one mask in a flammable environment, such as close to fire or static electricity. It may cause skin burn to the patient and damage the mask and CO 2 sensor. Select the cap-one mask taking into consideration the patient weight and ventilation volume. If an inappropriate cap-one mask is used, desired oxygen concentration or correct measurement value cannot be obtained. Before and while using the cap-one mask, check that the oxygen connector is not loose or disconnected from the oxygen supply unit and that the oxygen cannula tube is not blocked. Always check the conditions of the patient and oxygen supply circuit. 1

If arterial oxygen partial pressure does not increase, immediately stop using the cap-one mask and select another way to supply oxygen. This cap-one mask is for infants and children. Do not use this cap-one mask for neonates. Do not use a damaged or deformed cap-one mask. It causes incorrect measurement or insufficient oxygen supply. Do not use a cap-one mask which is past the expiration date. The cap-one mask is non-sterilized and disposable. Use only for a single patient and single use. Failure to follow this instruction causes cross infection. United States law restricts this product to sale by or on the order of a physician. When using the cap-one mask, follow the physician s instruction. Only transport and store the cap-one mask in the environment conditions which are described on the package. Do not leave the cap-one mask in a place with high temperature such as on a dashboard of a car. This may deform the mask and cause inaccurate measurement. Attaching the cap-one Mask 1. Open the package of the cap-one mask just before using it to avoid dust gets on the transparent film of the mask. 2. Firmly connect the oxygen connector of the cap-one mask to a socket of the oxygen supply unit. When using a humidifier unit with the oxygen supply unit, check their compatibility and safety thoroughly. When using the cap-one mask on a patient with bleeding disorder, poor general medical condition or malnutrition, observe the patient condition all the time. The cap-one mask touches the nose and mouth and may cause pressure sores. Failure to follow the instructions below degrades the anti-fogging ability of the transparent film and results in incorrect measurement. Replace the cap-one mask with a new one every 72 hours. Replace the cap-one mask with a new one if blood, sputum or mucus adheres to the transparent film. Do not damage the transparent film. Do not let dust or detergent contact the transparent film. Do not touch, wipe or clean the transparent film with fingers or cleaners. Do not reuse the cap-one mask. It is disposable. Connect the cap-one mask only to a socket of an oxygen supply unit which can be firmly connected with the oxygen connector of the cap-one mask. Check the socket connection of the oxygen supply unit and oxygen connector occasionally. If the socket connection is loose, replace the cap-one mask with a new one. 3. Check that the oxygen cannula tube is not blocked. If block or leakage is found on the cap-one mask, replace it with a new one. Make sure that oxygen comes out from the mask. Do not bend or twist the cap-one mask. This may cause damage or block on the cap-one mask. 4. Connect the CO 2 adapter connector of the TG-970P CO 2 sensor kit to the input socket of the monitor. Wait for about 10 seconds until the CO 2 sensor kit warms up. 2

5. Attach the CO 2 sensor to the CO 2 sensor attachment of the cap-one mask until it clicks. The CO 2 sensor can be attached in either direction. 8. Attach the cap-one mask firmly to the patient using one of the following three ways. The gastric tube can be inserted through the cutout on the sides of the mask. CO 2 sensor attachment CO 2 sensor Fasten the cap-one mask to the patient s head with the strap. Pull the ends of the strap to adjust the strap length. Strap 6. Perform zero calibration. For details about zero calibration, refer to the operator s manual of the connected monitor. Calibrating with air: Expose the CO 2 sensor kit to the air and calibrate it with the connected cap-one mask. Adjust it assuming the partial pressure of CO 2 in the air is about 0.5 mmhg. Hook the oxygen cannula tubes around the patient s ears and adjust the tightness with the adjuster. Adjuster Perform calibration when: the cap-one mask is replaced with a new one. the operating temperature changes. the measurement room is changed. 7. Position the CO 2 catcher (the small cup inside the mask) under the patient s nose and check that the waveform of the expired CO 2 appears on the monitor. Adjust the position as necessary. Gastric tube Oxygen cannula tube Tighten the oxygen cannula tubes on the patient s head and adjust the tightness with the adjuster at the top of the head. Keep the distance between the CO 2 catcher and the top of the mouth and nose within 1 cm. CO 2 cannot be measured correctly if the CO 2 catcher is far from the nose and mouth. Gastric tube CO 2 catcher Oxygen cannula tube Adjuster Do not tighten the strap or oxygen cannula tubes too tight. It may cause blood congestion. Observe the patient condition periodically. Oral care is possible when the mask is attached. However pay attention not to get the CO 2 sensor attachment dirty. 9. Tape the tubes and CO 2 sensor cable on the cheek with surgical tape. 3

10. Supply the oxygen to the patient. Oxygen concentration changes depending on the respiratory volume of the patient. Adjust the oxygen flow rate to obtain the desired oxygen concentration. Refer to the following table as a guide for the oxygen concentration. Oxygen Concentration Oxygen Flow Rate YG-232T YG-242T (Children) (Infants) 2 L/min 40% 60% 4 L/min 50% 70% 6 L/min 60% 80% Removing the cap-one Mask 1. Remove the surgical tape, then the cap-one mask from the patient. 2. Hold the CO 2 sensor attachment firmly with one hand and the CO 2 sensor with the other hand as shown. Pull the CO 2 sensor straight up with three fingers to remove the sensor from the mask. Follow the physician s instruction when controlling the oxygen flow rate. Periodically check the oxygen concentration of the patient with a pulse oximeter or arterial blood gas analyzer. Use the cap-one mask with 2 L/min or more of oxygen flow rate to reduce rebreathing. Otherwise the patient may rebreathe the expiratory gas in the mask. Bad examples If the patient s respiratory volume is low and the oxygen flow rate is high, CO 2 data becomes inaccurate due to the cap-one mask s dead space. Do not pull the oxygen cannula tube excessively while supplying the oxygen to the patient. The tube may disconnect from the cap-one mask and oxygen may not be supplied to the patient. 11. Check that the partial pressure of the expired CO 2 is measured correctly on the monitor. Refer to the operator s manuals of the monitor and CO 2 sensor kit for details. If the measured partial pressure of the inspired CO 2 is high, do the following. i) Remove the CO 2 sensor from the cap-one mask. ii) Reattach it to the mask. iii) Perform calibration again. Refer to step 6. Do not let the patient bite the mask, tubes or CO 2 sensor cable. Arrange the tubes and CO 2 sensor cable so they do not coil around the patient. When removing the CO 2 sensor, do not hold the cable or one side of the CO 2 sensor. The cable may break or excessive force may damage the CO 2 sensor. 3. Disconnect the oxygen connector from the oxygen supply unit. Hold the oxygen connector, not the oxygen cannula tube. Check Before Use To use the cap-one mask properly and safely, check the following items before attaching the cap-one mask to the patient. No loose or disconnection on the oxygen connector. No block on the oxygen cannula tube. No damage, deformation or dirt on the cap-one mask. Cleaning, Disinfection and Sterilization The cap-one mask is disposable. You cannot clean, disinfect or sterilize it. Immediately replace it with a new one when it becomes dirty. 4

Expiration Date The cap-one mask is disposable. The expiration date is 36 months including the month of manufacture. Disposing For detailed information about disposal, contact your Nihon Kohden representative. Dispose of Nihon Kohden products according to your local laws and your facility s guidelines for waste disposal. Otherwise, it may affect the environment. If there is a possibility that the product may have been contaminated with infection, dispose of it as medical waste according to your local laws and your facility s guidelines for medical waste. Otherwise, it may cause infection. Specifications Mechanical Strength Resistance to shock: MIL STD 810F 516.5 4.5.5:2000 (when the product is in the package) Resistance to kinking: EN 13544-2:2002 5.1.5 EN 13544-2 Amendment 1:2009 5.1.5 (except for the thinner oxygen cannula tubes) Airtightness: EN 13544-2:2002 5.1.4 EN 13544-2 Amendment 1:2009 5.1.4 Strength of connection: Oxygen cannula tube and oxygen connector: EN 13544-2:2002 5.1.3 EN 13544-2 Amendment 1:2009 5.1.3 Oxygen connector and oxygen supply unit: EN 13544-2:2002 5.1.4 EN 13544-2 Amendment 1:2009 5.1.4 Environment Store the cap-one mask away from direct sunlight, high temperature and humidity. Operating environment Temperature: 10 to 40 C (50 to 104 F) Humidity: 30 to 85% (noncondensing) Atmospheric pressure: 700 to 1060 hpa Transport and storage environment Temperature: 20 to +65 C ( 4 to +149 F) Humidity: 10 to 95% Atmospheric pressure: 700 to 1060 hpa Safety Standard Safety standard: IEC 60601-1:1988 IEC 60601-1 Amendment 1:1991 IEC 60601-1 Amendment 2:1995 ISO 10993-1:2009 Options Refer to the operator s manual of the TG-970P CO 2 sensor kit for details. Description Model Qty Supply Code CO 2 sensor kit TG-970P 1 P909 CO 2 sensor TG-221T 1 P924 CO 2 adapter JG-970P 1 K987 Surgical tape #1527 1 Y242 Copyright Notice The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. Manufacturer NIHON KOHDEN CORPORATION 1-31-4 Nishiochiai, Shinjuku-ku, Tokyo 161-8560, Japan Phone +81 3-5996-8036 Fax +81 3-5996-8100 http://www.nihonkohden.com/ NIHON KOHDEN AMERICA, INC. 15353 Barranca Parkway, Irvine, CA 92618, U.S.A. Toll-free +1-800-325-0283 Phone +1 949-580-1555 Fax +1 949-580-1550 5 European Representative NIHON KOHDEN EUROPE GmbH Raiffeisenstrasse 10, D-61191 Rosbach, Germany Phone +49 6003-827-0 Fax +49 6003-827-599 SHANGHAI KOHDEN MEDICAL ELECTRONIC INSTRUMENT CORP. 567 Huancheng Bei Road Shanghai Comprehensive Industrial Development Zone Shanghai 201401, China Phone +86 21-5743-6998 Fax +86 21-5743-6939