Operating Instructions for Direct Flow Mass Flow Meter for gases. Model: DMW

Similar documents
Operating instructions for variable area flow meter with needle valve. Model: UVR/UTR

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

DMW. Direct Mass Flow Meter and Controller. for Gases. Measuring ranges from Nl/min air. Accuracy: ±3% of full scale

Operating Instructions for Flow Restrictors. Model: REG

Design Features. General Description

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

MANUAL KPS Pressure Control Valve

Model PDT Dewpoint Transmitter

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N

2/2-Way Solenoid Control Valve

Mass Flow Controller (MFC) for Gases

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

SITRANS P measuring instruments for pressure

Mass Flow Controller (MFC) for Gases

Mass Flow Meter (MFM) for Gases

Mass Flow Meter (MFM) for gases

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Proportional pressure regulators VPPE

Mass Flow Controller (MFC) for Gases

Gas density monitor With integrated transmitter Model GDM-100-TI

Model 601CV. Mass Flow Controller. Product Features:

Digital Mass Flow Meters and Controllers for Gases

OPTIBAR P 1010 C Technical Datasheet

QPL... Pressure Switches. Siemens Building Technologies HVAC Products

KMT. Thermal Mass Flowmeter. for compressed air and gases. Measuring range: Nm³/h Nm³/h

Mass Flow Controller (MFC) for Gases

Bypass Mass Flow Controller (MFC) for gases

Precision pressure sensor Basic version Model CPT6020

Mass Flow Controller (MFC) for gases. Inline MFC for nominal flow rates from 25 I N. /min to I N

Operating Instructions for Manometer. Model: MAN

Pressure Transmitter P41

Best Practice Guide, Servomex 2700

3G Operating instructions. Digital Purge Gas Valve 3G Type: Document no.: D0001 Version: 01. July 2011/Rev.

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

Electronic relative and differential pressure switch type 616

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Mass Flow Controller (MFC) for Gases

High-performance submersible pressure transmitter For level measurement Model LH-10

2/2-Way Proportional Valve (motor-driven)

ATM.ECO - Analog Pressure Transmitter

Mass Flow Controller (MFC) for Gases

C-Flow Coriolis Mass Flow Meters

Operating Instructions DB41

High-performance submersible pressure transmitter For level measurement Model LH-10

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

Instruction Manual Differential Pressure Transmitter

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

x act i Precision Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor accuracy according to IEC 60770: 0.

KTE6000 / KTU6000 Series OEM pressure transmitters for industrial media

Features. Description

P499 Heavy Duty Pressure Transducer

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/16 Siemens FI US Edition

Pressure Measurement Single-range transmitters for general applications

Differenz-Drucktransmitter Differential pressure transmitter Transmetteur de pression différentielle bar

Relative and differential pressure transmitter type 692

Thermal Energy Flowmeter Model KEC. Application Range. Special Features

AMS 2710 PCB pressure sensor module with V output

deltaflowc deltaflowc Venturi or Probe

Solenoid valves VZWF, force pilot operated

1 LS unloading (control valve, orifice, none) 2 LS max pressure relief. 3 Spool type

ATM - Analog Pressure Transmitter

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

Pressure Measurement. Transmitters for basic requirements SITRANS P Z for gauge pressure. 2/4 Siemens FI Overview

ATM/GR. Electronic pressure switches CUSTOMER BENEFITS

Instruction and Maintenance Manual

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

Variable Area Flow Meters

The affordable consumption counter for compressed air and gases R 3 /4 (DN 20)

2/2-way proportional valve

Orifice Plate Flow Meter

Technical Data Sheet. OPTIMASS 7000 Custody Transfer Mass Flowmeter

Installation and Operation Manual

SPCH/ Blowing block

P900. P900, Datenblatt Seite 1 DESCRIPTION

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

Instruction Manual Differential Pressure Transmitter

2/2-Way Solenoid Control Valve

Programmable pressure and temperature transmitters PTM/RS485

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

2/2- and 3/2-way Rocker Solenoid Valve - bistable (impulse) version

Instruction and Maintenance Manual

Pressure Transmitters 36 F-27839

Measurement of Low GAS & Liquid Flows

Air Purge Unit. Series: LSP-compact Type: EJG 2-2. (Controlled with DigiFlow 515) Installation & Operation Manual 11/2014

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

ATM.1ST. Pressure transmitters CUSTOMER BENEFITS

Product information. Capacitive. Level detection in liquid VEGACAP 62 VEGACAP 63 VEGACAP 64 VEGACAP 66 VEGACAP 69. Document ID: 29983

GasBloc Multifunctional gas control Combined regulator and safety shut-off valves Integrated gas-air system GB-GD 057 D01

Level transmitter LT 100

Transcription:

Operating Instructions for Direct Flow Mass Flow Meter for gases Model: DMW

1. Contents 1. Contents... 2 2. Note... 3 3. Instrument Inspection... 3 4. Regulation Use... 3 5. Operating Principle... 4 5.1. Operating Principle DMW-A/-C... 4 5.2. Operating Principle DMW-B/-D... 5 6. Mechanical Connection... 6 7. Electrical Connection... 7 8. Commissioning... 8 8.1. Trouble shooting... 8 9. Technical Information... 9 10. Order Codes... 10 11. Maintenance... 11 12. Dimensions... 12 13. Pressure Loss Diagram... 14 14. Declaration of Conformance... 15 Manufactured and sold by: Kobold Messring GmbH Nordring 22-24 D-65719 Hofheim Tel.: +49(0)6192-2990 Fax: +49(0)6192-23398 E-Mail: info.de@kobold.com Internet: www.kobold.com page 2 DMW K01/0910

2. Note Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The devices are only to be used, maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health & Safety and prevention of accidents. When used in machines, the measuring unit should be used only when the machines fulfil the EWG-machine guidelines. as per PED 97/23/EG In acc. with Article 3 Paragraph (3), "Sound Engineering Practice", of the PED 97/23/EC no CE mark. Diagram 6, Pipe, Group 1 dangerous fluids 3. Instrument Inspection Instruments are inspected before shipping and sent out in perfect condition. Should damage to a device be visible, we recommend a thorough inspection of the delivery packaging. In case of damage, please inform your parcel service / forwarding agent immediately, since they are responsible for damages during transit. Scope of delivery: The standard delivery includes: Mass Flow meter for gases, model: DMW Operating Instructions 4. Regulation Use Any use of the Direct Mass Flow Meter for gases, model: DMW, which exceeds the manufacturer s specifications may invalidate its warranty. Therefore, any resulting damage is not the responsibility of the manufacturer. The user assumes all risk for such usage. DMW K01/0910 page 3

5. Operating Principle The new model DMW-... direct mass has been specially designed for use in gas measuring technology. An inexpensive analogue output is fitted. The model DMW is available in three versions: with or without integrated 3 1/2-digit LCD display or with an 8-digit counter. The direct stream procedure is ideally suited for measurements from 5 LN/min (max.) up to 7.50 Nm³/min air (model: DMW-B*876). In case of small flow rates the measurement is executed via a bypass-system (DMW-A/C). In contrast with most variable area s the mass DMW-... has no moving parts, and no temperature pressure correction is required. The meter may be installed in any position and the pressure loss is negligible. The simple mechanical construction ensures the highest degree of reliability when used with aggressive gases, and as well under rugged industrial conditions. 5.1. Operating Principle DMW-A/-C The flow sensor is operating on a principle of heat transfer by sensing the delta-t along a heated section of a capillary tube. The delta-t is directly proportional to the gas mass flow. The heat transfer function between gas mass flow and temperature difference can be described by the equation: T = K x Cp x Φm T = temperature difference Cp = specific heat K = constant factor Φm = mass flow When using a flow regulator a control valve is integrated directly into the instrument. The central valve is a N/C solenoid valve. The baffle plate of the valve is being actuated from the magnetic field of the solenoid. For adjustment of the k v -value the valve orifice can be changed. Standard seal material is FPM. Please check compatibility of sealing material on appropriate resistant tables. page 4 DMW K01/0910

5.2. Operating Principle DMW-B/-D The flow sensor consists of two stainless steel probes that are positioned in the flow path; a heater probe and a temperature probe. A constant difference in temperature (delta-t) is created between the two and the energy required to maintain the delta-t is dependent of the mass flow rate. This can be described by the following equation: P = P0 + C. Φm n P = total heater power P0 = heater power offset at zero flow C = constant Φm = mass flow n = dimensionless number (typ. 0.5) Using a flow controller a directly operating control valve is integrated. The integrated control valve is a normally closed solenoid valve. The plunger is lifted by the magnetic field of the coil. For adapting the suitable k v -value the orifice of the valve is changeable. The typical sealing material is FPM. The compatibility of the material with your operating media may be seen in suitable resistant tables. DMW K01/0910 page 5

6. Mechanical Connection Ensure that the piping is absolutely clean and free of damage. It is recommended to install an inlet filter, and if backflow can occur, also an outlet filter. During installation of the instrument pay attention to the arrow marking on the body which defines the flow direction. Do not install small diameter piping on high flowrates. Do not use abrupt angles or create other disturbances directly on the inlet of the instrument. Use only metric fittings. NPT-fittings will destroy the thread of the body. On instruments with special design with low pressure loss p an inlet pipe straight will be supplied together with the unit and must be mounted at the inlet. The recommended inlet pipe straight is 10x the nominal pipe diameter. Please leakproof the system before commissioning. If explosive and/or corrosive gases are to be used, purge the system with a dry, inert gas, for at least 30 minutes. This is also required to remove these gases from the system. The preferred installation of our instruments is horizontal. Please consult the factory for recommendations if high flow rate controllers must be installed in alternative positions. page 6 DMW K01/0910

7. Electrical Connection All instruments carry the CE-mark. Therefore they have to comply with the EMC requirements as are valid for these instruments. When connecting the instrument to other devices (e.g. power supply), ensure that the operating voltage corresponds with the one on the name plate and the function of the shielding is impacted. Do not use unshielded wire terminals. Ensure that the electrical supply of the unit is in agreement with its operational data. Make certain that the Power Supply lines are powerless. To prohibit damage or destruction to electronics and sensor, all units are provided with polarity protection. Attention! In case of a polarity-reversal, the power supply circuit gets short-circuited. Connect the sensor according to the connection diagram shown here. Terminal pin assignment PIN DMW-A/B DMW-C/D with Poti DMW-C/D 1 + V S + V S 2 GND (Power) GND (Power) 3 GND (Signal) GND (Signal) 4 + Signal + Signal 5 N/C + setpoint 6 N/C N/C Note: Wiring for units with cable connection and flying leads is printed on the cable. DMW K01/0910 page 7

8. Commissioning After switching on power, allow for at least 30 seconds to warm-up and stabilise. After this time the instrument works with an accuracy of 4% FS. After additional 30 minutes the optimal stability is reached and the instrument now has its optimum accuracy of 3% FS. This may be done with or without gas flow. Be sure that the specified pressure(s) has/have been applied. Pressure surges, which may occur during system pressurisation must be avoided. The control valve (in conjunction with the regulator) is not suitable for complete shut-off. When starting up, always bring the instrument gradually up to the level of operating conditions. 8.1. Trouble shooting Problem: No output signal Oscillation of the controller Flow although input signal is zero Setpoint > actual flow Check: - no power supply - connections of the cable - pressure - for inline stoppage - pressure; valve - seals of the valve - inlet outlet pressure; kvvalue of the valve page 8 DMW K01/0910

9. Technical Information Measuring system: and Bypass Measuring range: 3...100 % Gas type: all gases, depending on the material compatibility Accuracy (with calibration for air): ±3 % f.s. incl. non-linearity (better upon request) Repeatability: ±0.5 % f.s. (better upon request) Time constant (63.2%): τ = 0.7 s (standard, better upon request) Display: 3 1/2-digit LCD (flow rate); 8-digit LCD (counter) Pressure sensitivity: 0.2 % / bar typically (air) Temperature coefficient: ±0.1 %/ C Temperature range: 0...70 C Pressure: 10 bar; higher upon request Leakage rate: < 2 x 10-9 mbar L/s He Preheating time: 30 min for optimal accuracy; 30 s for accuracy ±4 % f. s. Mounting position: any, flow rate in arrow direction Inlet section: 10 x D (DMW-A) without limitation for other types kvs-values controller: 0.066; 0.35; 1.0 (DMW-C/D) Materials Sensor: Stainless steel 316L Casing: Stainless steel 1.4305 or Aluminium (anodised) Filter /rings: Stainless steel /PTFE Seal: FPM, EPDM or Kalrez Power supply: 24 V DC ±10 % Max. current values DMW-A/C: DMW-B/D: With valve control: Signal input (with controller only): 75 ma max. current input 250 ma max. without flow 75 ma max. 100 % flow rate 175 ma max. + 250 ma max. Poti, 0...5 V DC or 4...20 ma DMW K01/0910 page 9

Signal output: Connector: Protection: Special design (upon request): real gas calibration (upon request): 0...5 V DC or 4...20 ma active round pin plug/ opposite plug, 6-pole DIN 45322 IP 40; IP 65 (option)»low pressure loss«version, dry and oil-free; fast response version; digital design with Bus connection for example: H2, Helium, CO2, Methane, Propane, Butane, Argon (No oxygen) 10. Order Codes Order Details mass (Example: DMW-A71 12 G2 F 3 2) Measuring system Bypass Measuring tube Aluminium Sensor housing Stainless steel small DMW-A71... DMW-A21... 4 mm DMW-B70... DMW-B20... 8 mm DMW-B71... DMW-B21... 16 mm DMW-B72... DMW-B22... 32 mm DMW-B73... DMW-B23... 56 mm DMW-B75... DMW-B25... 84 mm DMW-B78... DMW-B28... *L N = Norm liter at 1013 mbar and 0 C L S = Standard liter at 1013 mbar and 20 C (others upon request) Measuring ranges also available in m³ N /min, m³ N /h, L N /h or others. Measuring range [LN/min air]*..12.. = 5.0...100.0 ml N /min..22.. = 10.0...200.0 ml N /min..52.. = 0.025...0.500..13.. = 0.050...1.000..23.. = 0.100...2.000..53.. = 0.25...5.00..14.. = 0.50...10.00..14.. = 0.50...10.00..24.. = 1.00...20.00..24.. = 1.00...20.00..54.. = 2.5...50.0..15.. = 5.0...100.0..15.. = 5.0...100.0..25.. = 10.0...200.0..45.. = 20...400..16.. = 50...1000..26.. = 100...2000..26.. = 0.10...2.00 m³ N /min..46.. = 0.20...4.00 m³ N /min..56.. = 0.25...5.00 m³ N /min..56.. = 0.25...5.00 m³ N /min..66.. = 0.30...6.00 m³ N /min..76.. = 0.30...7.50 m³ N /min Connection..G2.. = G 1/4 IG..G2.. = G 1/4 IG..G2.. = G 1/4 IG..G4.. = G 1/2 IG..G4.. = G 1/2 IG..G6.. = G 1 IG..G6.. = G 1 IG Seal Indication/ protection Output..F.. = FPM..E.. = EPDM..P.. = Kalrez..3.. = without indication, IP 40..6.. = without indication, I P 65..D.. = flow rate indication, IP 40..Z.. = counter, IP 40..2 = 0-5 V DC..4 = 4-20 ma page 10 DMW K01/0910

Order Details mass flow controller (Example: DMW-C71 12 G2 F 3 2) Measuring system Bypass Bypass Measuring tube Sensor housing Stainless Aluminium steel small DMW-C71... DMW-C21... medium DMW-C72... DMW-C22... Measuring [LN/min air]..12.. = 5.0...100.0 ml N /min..22.. =10.0...200.0 ml N /min..52.. = 0.025..0.500..13.. = 0.050...1.000..23.. = 0.100...2.000..53.. = 0.25...5.00..14.. = 0.50...10.00..14.. = 0.50...10.00..24.. = 1.00...20.00..54.. = 2.50...50.0 Connection..G2.. = G 1/4 IG..G4.. = G 1/2 IG 4 mm DMW-D70... DMW-D20.....14.. = 0.50...10.00..24.. = 1.00...20.00..G2.. = G 1/4 IG 8 mm DMW-D71... DMW-D21... 16 mm DMW-D72... DMW-D22.....24.. = 1.00...20.00..54.. = 2.5...50.0..15.. = 5.0...100.0..15.. = 5.0...100.0..25.. = 10.0...200.0..45.. = 20...400..G2.. = G 1/4 IG..G4.. = G 1/2 IG 32 mm DMW-D73... DMW-D23.....45.. = 20...400..16.. = 50...1000..G4.. = G 1/2 IG Plug-in power supply Seal Indication/ Protection Input/ Output Input: 90-264 V AC Output: 24 VDC 500 ma Model ZUB-SNT035L..F.. = FPM..E.. = EPDM..P.. = FFKM..3.. = without indication, IP 40..6.. = without indication, IP 65..D.. = flow rate indication, IP 40..Z.. = counter, IP 40..2 = 0-5 VDC..4 = 4-20 ma..6 = Poti/ 0-5 VDC (to 100 L/min)..8 = Poti/ 4-20 ma (to 100 L/min) 11. Maintenance No routine maintenance is required to be performed on the instruments. Inlet filters should periodically be cleaned ultrasonically. In such a case, the unit must be cleaned well, and a new calibration becomes inevitable; send this unit under consideration of safety instructions and cleaning at the latest every 24 months to your supplier. DMW K01/0910 page 11

12. Dimensions DMW-A..., DMW-B*0..., DMW-B*1..., DMW-B*2... Model B C F G DMW-A... G ¼ G ¼ 95 96 DMW-B*0... G ¼ G ¼ 95 96 DMW-B*1... G ¼ G ¼ 95 96 DMW-B*2... G ½ G ½ 99 100 DMW-B*3..., DMW-B*5..., DMW-B*8... Model A B C D E F G H DMW-B*3 116 G ½ G ½ 50 50 123 124 35 DMW-B*5 130 G 1 G 1 70 70 143 144 35 DMW-B*8 160 G 1 G 1 100 100 172 173 35 page 12 DMW K01/0910

DMW-C*1..., DMW-C*2..., DMW-D*0..., DMW-D*1... Model A B C D F G H DMW-C*1... 95 G ¼ G ¼ 30 95 96 35 DMW-C*2... 95 G ½ G ½ 30 97 98 35 DMW-D*0... 95 G ¼ G ¼ 30 95 96 35 DMW-D*1... 95 G ¼ G ¼ 30 95 96 35 DMW-D*2..., DMW-D*3... Model A B C D F G H I DMW-D*2... 145 G ½ G ½ 50 131 45 95 44 DMW-D*3... - G ½ G ½ Dimensions upon request DMW K01/0910 page 13

13. Pressure Loss Diagram For model DMW-C71... and DMW-C-21... all <50 mbar page 14 DMW K01/0910

14. Declaration of Conformance We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Flow Meter Model: DMW which relates to this certificate, conforms to the standards listed below: EN 50081-1 EN 55022 EN 50082-2 EN 61000-4-2 ENV 50140 EN 61000-4-4 EN 50141 Class B 8 kv/ad4 kv C 10 V/m 2 kv/5 khz 10 V also fulfils the following EWG guidelines: 89/336/EEC EMC Directive 2004/108/EC EMC Directive Hofheim, 16. Jan. 2007 H. Peters M. Wenzel General Manager Proxy Holder DMW K01/0910 page 15