THE UNIVENTOR 400 ANAESTHESIA UNIT

Similar documents
The Univentor 1250 Anaesthesia Unit

In Vivo Scientific, LLC INSTRUCTION MANUAL

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

MAGNA-CHEM AUTOMATIC CHLORINATOR - MODEL: CSA-2P MAGNA ORP Chlorine Pump

92831 TEL: (714) FAX:

MODEL 100 NITROGEN INFLATION CART

Budget Range Operators Handbook

User s Guide. CO2 Controller for miniature incubators

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Operating Instructions Part No

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

User s Guide. CO2 & O2 Controller for miniature incubators

92831 TEL: (714) FAX:

DPC-30 DPC-100. Reference Manual

Portable oxygen concentrator

User s Guide. Vacuum Controller for liquid delivery systems

UNIVERSAL MASK TEST SYSTEM

PM105 PRESSURE METER USER S MANUAL

CSA Sample Draw Aspirator Adapter Operator s Manual

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

RAM 4021 Operation Manual

INSTRUCTION MANUAL. Includes: OPERATION SETUP USE & CARE SERVICE. SPORTS ATTACK LLC

Digital Melting Point Apparatus

1. Connect the attached anesthesia mask to your anesthesia system using the included tubing and connectors.

NovelBright LED Illuminator

Advanced Test Equipment Rentals ATEC (2832)

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

! Warning, refer to accompanying documents.

R E D I C O N T R O L S

Issue OWNER S MANUAL. GAMMA RACQUET SPORTS GAMMA Sports All Rights Reserved

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

MODEL Maverick Heavy Duty Aerosol Compressor Operator s Manual

RK LEL Sample Draw Aspirator Adapter Operator s Manual

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

MODEL NUMBER: AS18-2 USER GUIDE

SPECIFICATIONS APCEPH1

AA GWR Water Retention Meter for Accurate Measurement of Water Retention

RK-IR Sample Draw Aspirator Adapter Operator s Manual

USER MANUAL SAVE THESE INSTRUCTIONS. 300 Held Drive Tel: (+001) Northampton, PA USA Fax: (+001) ISO Certified

User Manual MFCS TM -EZ

Instructions for Use

Cocoa Patio Pond with Lit Spillway

U S E R M A N U A L. Model: PM4300 Series SAVE THESE INSTRUCTIONS

PERSONAL AIR BREATHING UNIT

MODEL CALIBRATION GAS DELIVERY SYSTEM

DIGITAL AUTOMATIC TYRE INFLATOR

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

Operating Instructions Part No

ACV-10 Automatic Control Valve

COMBAT SWIM BOARD. RJE International, Inc. RJE International Feb 1, TAC-300 Revision 6

Model No.: SFC-ED Torque Tool Serial No.: SFC Ref No.:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

KJ4000 Operating Instructions & Parts Manual NSN

COMBAT SWIM BOARD. RJE International, Inc. RJE International January 16, TAC-300 Revision 4

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

700 SERIES BMS II BUBBLER MONITORING SYSTEM

NOVALYNX CORPORATION MODEL WEATHER STATION INSTRUCTION MANUAL

444C DUAL PERFORMANCE VALUE PACK

200 PSI HIGH-FLOW AIR SOURCE KIT

SPECIFICATIONS ATTENTION

M5400 Operators Manual

Ever. EverFlo. Home oxygen system. Oxygen simplified

Installation, Operation & Maintenance Instructions

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

400C & 450C DUAL PERFORMANCE VALUE PACKS

Corning Step-R. Repeating Pipettor. Instruction Manual

K-SYSTEMS Instruction Manual. CO ² Bench Top Incubator G85

O2 Portable. User Manual

Instruction and Maintenance Manual

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

DIVER NAVIGATION BOARD. RJE International, Inc. RJE International Feb 1, TAC-100D Revision 10

Installation, operating and maintenance Instructions for Seemag bypass level indicator

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Qi Litre Ultrasonic Cleaner

Hallowell EMC Anesthesia WorkStation Operating Manual

200 PSI FAST-FILL AIR SOURCE KIT

ANALOX O2 PORTABLE OPERATORS MANUAL

GasSense NDIR User Manual

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

PH01 Perfusion Cannula Manual

CUT OFF TOOL MODEL: CAT113

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09

12V Mini Air Compressor

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

Digital Vacuum Regulator

OPERATING AND MAINTENANCE MANUAL GOLD LINE EXTRACTION CLEANERS M14-M26

Instruction and Maintenance Manual

Model 130M Pneumatic Controller

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

INFUSION SYSTEM (110V) (220V) USER S MANUAL

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Measurement of Low GAS & Liquid Flows

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

Transcription:

THE UNIVENTOR 400 ANAESTHESIA UNIT

TABLE OF CONTENTS EDITION 3 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2 - INTRODUCTION 2.1. INTRODUCTION 3 Section 3 - GENERAL 3.1. UNPACKING & INSTALLATION 4 3.2. POWER SUPPLY 4 3.3. CONTROL PANEL 4 3.4. DISPLAY 5 3.5. SAFETY FEATURES 5 Section 4 - OPERATION 4.1. INSTRUCTIONS 5 Section 5 - ROUTINE MAINTENANCE 5.1. CLEANING 7 5.2. LEAKAGES 7 5.3. REPLACING PARTS 7 5.4. STORAGE 7 5.5. TROUBLE SHOOTING 7 Section 6 - SPECIFICATIONS 9 Section 7 - ACCESSORIES & REPLACEMENT PARTS 9 INDEX 10 1

1.1. WARRANTY Section 1 - WARRANTY & SERVICE Univentor Ltd guarantees all components of the 400 Anaesthesia Unit to be free from defects of material and workmanship for a period of two years after initial purchase. Univentor will repair or replace, at its discretion, all defective components during the aforementioned warranty period. For warranty service or repair, all Univentor s products must be returned to Univentor or to an authorised Univentor representative. The client is responsible for shipping charges to Univentor. The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the client, unauthorised modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. For any product expressly covered under this warranty, Univentor is liable only to the extent of replacement or repair of the defective items. Univentor shall not be liable for any personal injury, property damage, or consequential damages of any kind whatsoever. The foregoing warranty is in lieu of all other warranties of merchant ability and fitness for a particular purpose. 1.2. DAMAGED SHIPMENTS Damage to any part of this instrument during shipping should be reported immediately to Univentor or an authorised representative. You must retain the original packing box and contents for inspection by the freight handler. Univentor will replace any new instrument damaged in shipping with an identical product as soon as possible after the claim filing date. Claims not filed within 30 days after the shipping date will be invalid. Do not return damaged goods to Univentor without first contacting Customer Service for a Return Authorisation Number (RA#). When a defective part is returned to Univentor, the RA# immediately identifies you as the sender, and describes the item being returned. Univentor refuses all unauthorised return shipments. 1.3. SERVICE Univentor has a skilled service staff available to solve any technical problem. For further details contact Univentor or Univentor s representative. Following discussion of your specific difficulties, an appropriate course of action will be described and the problem resolved accordingly. Do not return any products for service until a RETURN AUTHORISATION NUMBER (RA#) has been obtained. The RA# identifies you as the sender and describes in full detail the problems you have.turnaround time for service can be quoted to you at the time your RA# is issued, although we can not determine the actual amount of service required until we have received your unit and diagnosed the problem. All correspondence and shipments should be sent to Univentor Ltd. or your Univentor representative. 2

2.1. INTRODUCTION Section 2 - INTRODUCTION The Univentor 400 Anaesthesia Unit is designed to control the mixture of liquid anaesthetic and air with the precision required to successfully operate on animals weighing from 20-500 grams. The Univentor 8323001 is designed to operate with compressed air reduced to not more than 1 bar whereas the Univentor 8323002 has been adapted to work with a pulse-free airpump. Air and anaesthetic is mixed in a vaporiser according to settings and may be delivered into an anaesthetised box, through a mask or to a ventilator. Your unit is calibrated for Isoflurane and must at all times be operated with a 10 ml/stroke 60 mm gas-tight glass syringe. Should you wish to use other liquid anaesthetics you will have to take the different properties into consideration and adapt your percentage accordingly. Whilst reliability and safety features have been incorporated we recommend continuous monitoring of the anaesthetised animal at all times. We urge you to read this manual and to follow the instructions! CONTROL PANEL FLOW METER KNOB AIR INLET CONNECTOR LIQUID INLET CONNECTING TUBE LIMIT SWITCH DIAL TILT FOOT VAPORISER POINTER SYRINGE HOLDER PUSHER GAS OUTLET FILTER LIQUID ANAESTHETIC FIGURE 2.1. - The Univentor 400 Anaesthesia Unit 3

Section 3 - GENERAL 3.1. UNPACKING AND INSTALLATION Remove the instrument from the shipping packaging and inspect both the instrument and the packaging for any signs of damage. If any damage is noted, contact the freight handler and your Univentor representative immediately, see section 1.2. Missing Items? Your 400 Anaesthesia Unit is delivered with: 1. 1 10mL gas-tight glass syringe 2. 1m PVC tubing 3. 1 syringe and Liquid Inlet connecting tube 4. Power supply 5. Mains lead 3.2. POWER SUPPLY MAINS - Use an earthed wall plug and the power supply supplied with the 400 Anaesthesia Unit. The power supply can automatically handle input voltage in the range from 100V AC to 240V AC, 50 to 60 Hz. BATTERY - 12V. Consumption: 400 ma. 3.3. CONTROL PANEL The following functions, together with brief instructions, are located on the panel of the Univentor 400 Anaesthesia Unit. (See Figure 3.3.) TEXT FUNCTION Fast Feed Set Start/Stop forward/reverse fast feed to position pusher. increasing/decreasing liquid flow rate. starting/stopping liquid delivery. Lights green when delivering and red in case of alarm. ref:3.5. Fast Feed Set Start/Stop Air Inlet Air Flow UNIVENTOR 400 Anaethesia Pump AIR FLOW 200 ml/min ISOFLUI 1.4 % Liquid Inlet Gas Outlet Use 10ml/stroke 60 mm gastight glass syringe only. 1. Set air pressure from source to 1 bar. 2. Connect air source to the air inlet. 3. Set air flow by turning the air inlet knob. Air flow rate is displayed in ml/min. 4. Fit the filled syringe and connect with tubing to Liquid inlet. Use to position the pusher. 5. Connect the mask tubing to the gas outlet. 6. Use to set liquid flow rate in ml/hour. 7. Press Start/Stop to start delivery. 8. When running display shows concentration. FIGURE 3.3. - The control panel of the Univentor 400 Anaesthesia Unit. 4

3.4. DISPLAY 1. AIR-FLOW: current air-flow in ml/min. 2. LIQUID: a) when switched on set liquid flow rate in ml/hour is displayed. b) when switched on and running the current anaesthetic concentration is displayed in % as calculated. 3.5. SAFETY FEATURES An audible alarm will be triggered should any one of the following occur: 1. The pusher has reached set stopping position. 2. The air flow is blocked. 3. The air flow is below 50mL/min. 4. The air flow is above 999 ml/min. 5. The liquid delivery is stopped. Refer to SECTION 5 - ROUTINE MAINTENANCE & TROUBLE SHOOTING. 4.1. INSTRUCTIONS Section 4 - OPERATION HANDLING OF ISOFLURANE AND/OR ANY OTHER CHEMICAL SHOULD AT ALL TIMES FOLLOW THE MANUFACTURER S INSTRUCTIONS AS WELL AS REGULATIONS APPLICABLE TO YOUR COUNTRY. ALWAYS WORK ON A VENTILATION BENCH. 1. Connect the instrument to the power supply and earthed mains. 2. Set air pressure from source to 1 bar in the case of 8323001 or connect air pump, making sure not to exceed 0.5 bar, in case of 8323002. 3. Remove protective cover and connect air source to the Air Inlet connector. NOTE: make sure that your air source is adequately filtered to prevent liquids entering the system. 4. Set the air flow, minimum 50 ml/min and maximum 999 ml/min, by turning the Flow Meter knob. Air Flow is displayed in ml/min. Mice - approximately 150-250 ml/min Rats - approximately 400-500 ml/min Should you use a chamber you will need to increase the air flow, and anaesthetic, until the animal is asleep: Mice - approximately 400-500 ml/min Rats - approximately 700-800 ml/min Once asleep turn the stopcock and route the gas to the mask and decrease the airflow. NOTE: the unit will NOT operate without air! 5

5. Press Start/Stop to start the instrument and set the concentration percentage by pressing Set arrows and then press Start/Stop again to stop the instrument during which time the liquid flow rate, ml/ hour, is displayed. Concentration - approximately 1.7-2.5% The following setting would give a concentration of 1.7%: Air flow rate - 200 ml/min Liquid flow rate - 1 ml/hour Should you use a chamber the induction dose should be approximately 4% NOTE: To increase and/or decrease the % concentration during operation, you may find it easier to adjust the air flow. 6. Following the manufacturer s handling instructions fill the 10mL gas-tight syringe (stroke 60 mm) with anaesthetic and make sure that there are no air bubbles. 7. Using Fast Feed reverse/forward the pusher as necessary and clamp syringe. Connect the syringe and Liquid Inlet with the supplied tubing and move the pusher forward until the liquid in the connecting tube is visibly just about to enter the vaporiser. NOTE: The fast feed forward will not work without air and the fast forward be pressed continuously. button must 8. Protect the syringe by turning the limit switch dial until the pointer is in the required stopping position. 9. Remove protective cover and connect one end of the supplied PVC tubing to the Gas Outlet and the other to the anaesthetised box, mask or ventilator. Keep tubing as short as possible. 10. Calculate the operating time available by dividing the liquid volume with the flow rate and make sure that the time is adequate. Flow rate is displayed when the instrument is in stop mode for this purpose. eg: with 10 ml Isoflurane in the syringe and set flow rate at 1.5 ml/hour the available time will be 6.666 hours (6 hours 40 minutes). 11. Press Start/Stop to start delivery. 12. Monitor the animal at all times and adjust the concentration percentage as necessary by pressing Set arrows or by increasing/decreasing the air flow. 13. When ready press Start/Stop to stop the liquid delivery and allow the air flow to continue for a period of 15 minutes to clean the vaporiser. DO NOT STORE ANAESTHETIC IN SYRINGE WHEN NOT IN USE NOTES 1. All values are approximate due to variations in weight and procedures. 2. The instrument can be tilted to maintain visual contact with display even while seated. 3. Delivery can be instantly terminated by pressing Start/Stop to stop the instrument. However, in case of emergency remove the mask due to gas in outlet tubing. When started again the previous settings will be displayed. 6

Section 5 - ROUTINE MAINTENANCE 5.1. CLEANING THE INSTRUMENT 1. After every use stop the liquid delivery and allow the air flow to continue for a period of 15 minutes to clean the vaporiser. 2. Keep your Anaesthesia Unit clean. 3. DO NOT STORE ANAESTHETIC IN SYRINGE WHEN NOT IN USE 5.2. LEAKAGES We recommend that the unit is regularly checked for gas leakages. 1. Switch the unit on. 2. Set air flow on 500mL/min. N.B. If air supply pressure exceeds 1 bar the unit will be damaged. 3. Block the gas outlet. 4. Reading on display should rapidly decrease to less than 10mL/min. Should unit leak (i.e. more than 10mL/min on display) contact your Univentor representative. 5.3. REPLACING PARTS We recommend that the gas outlet filter part number 2423017 is changed at least yearly or as necessary according to use. 5.4. STORAGE If the Univentor 400 Anaesthesia Unit is not to be used for a significant length of time, it is recommended to clean the instrument, fit protective covers on air inlet and outlet and store it safely in the shipping carton. 5.5. TROUBLE SHOOTING 1. Unit does not start delivery: a) The air flow is below 50 ml/min or above 999 ml/min check settings ref. Section 4/OPERATIONS/Step 4. check source and tubing. 2. The liquid delivery is stopped. a) Pusher has reached set stopping position check if syringe is empty. check position of limit switch and reset if necessary. b) Pusher is blocked - make sure stroke is clear. 7

3. Display shows SLIP ERROR STOP. Occurs when force exceeds 100 N due to: a) Pusher is blocked - make sure clear. b) Syringe and/or tubing blocked - check and replace as necessary. c) Gearbox is slipping - contact your Univentor representative. d) Cannula to vaporiser is blocked - to test fill syringe with air, connect to luer lock and try pushing plunger. If air passes easily it is not blocked. If very tough to push and plunger bounces back when released it is blocked. To unblock remove gas outlet adaptor and the tubing from the luer lock to vaporiser and push a thin wire through the cannula to clean any dirt that may be blocking the liquid. 4. Animal not being anaesthetised: a) Faulty air sensor - contact your Univentor representative. b) Syringe no longer gas tight - replace. c) Cannula to vaporiser is blocked - to test fill syringe with air, connect to luer lock and try pushing plunger. If air passes easily it is not blocked. If very tough to push and plunger bounces back when released it is blocked. To unblock remove gas outlet adaptor and the tubing from the luer lock to vaporiser and push a thin wire through the cannula to clean any dirt that may be blocking the liquid. d) Stroke of 10mL syringe not 60mm - replace 5. Animal over anaesthetised: a) Faulty sensor - contact your Univentor representative b) Air leakage check air connections check internal tubing - contact your Univentor representative c) Stroke of 10mL syringe not 60mm - replace 8

Section 6 - SPECIFICATIONS Power POWER SUPPLY: 100-240V AC 50-60 Hz. BATTERY: 12V 400 ma. Dimensions 120 mm (w) x 285 mm (l) x 95 mm (h). Weight 1.8 kg. Shipping weight 3 kg. Drive motor Pulse free DC motor with variable speed setting. Fast feed Pusher movement of 45 mm/min. Syringe 1 glass and gas-tight 10 ml syringe with 60 mm stroke. Min. liquid flow rate 0.4 ml/hr. Max. liquid flow rate 10 ml/hr. Min. air flow rate 50 ml/min. Max. air flow rate 999 ml/min. Min. air pressure/8323001 0.5 bar. Max. air pressure/8323001 1.0 bar. Min. air pressure/8323002 0.3 bar. Max. air pressure/8323002 0.5 bar. Max. pusher force 100 N. Pusher movement tolerance +/- 0.01 mm or +/- 1% of total distance. Concentration tolerance +/- 0.15 units of displayed % value. Display 2x16 Characters. Safety features Audible alarm and red LED. Section 7 ACCESSORIES & REPLACEMENT PARTS ORDERING INFORMATION CAT. No. DESCRIPTION 8323001 Univentor 400 Anaesthesia Unit. 8323002 Univentor 400 Anaesthesia Unit adapted for airpump. 2423017 Gas Outlet Filter. 8433005 Gas Routing Switch 8329001 Induction Chamber 0.8l. 8329002 Induction Chamber 1.4l. 9

INDEX ANAESTHETIC...3/5 AIR FLOW RATE...5/7 ALARM...5 ACCESSORIES...9 BATTERY...4 BODY WEIGHT...3 CLEANING THE INSTRUMENT...7 CONCENTRATION - ANAESTHETIC...6 CONTROL PANEL...4 DAMAGED SHIPMENTS...2 DISPLAY...5 LEAKAGES...6 LIMIT SWITCH...6 LIQUID FLOW RATE...5/6 MAINTENANCE AND STORAGE...7 OPERATING TIME...6 OPERATING INSTRUCTIONS...5-6 PACKING LIST...4 POWER SUPPLY...4 REPLACING PARTS...7 SERVICE...2 SPECIFICATIONS...9 SYRINGE...3/5 TOLERANCES...9 TROUBLE SHOOTING...7-8 UNPACKING AND INSTALLATION...4 WARRANTY...2 10

BLB029, Bulebel Industrial Estate, Zejtun ZTN 3000, Malta. Tel: (+356) 21 895824. Fax: (+356) 21 895835. E-mail: info@univentor.com Homepage: http://www.univentor.com