The Puritan Bennett 980 Neonatal Ventilator System. Helping to Protect Our Most Vulnerable NEWBORNS

Similar documents
HELPING PROTECT YOUR MOST VULNERABLE NEWBORNS. Puritan Bennett 980 Neonatal Ventilator System

THE PURITAN BENNETT 980 VENTILATOR CLINICAL APPLICATIONS LESSON PLAN

Puritan Bennett 840 Universal Ventilator

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

PATIENT SETUP QUICK REFERENCE GUIDE

Inspiration 7i Ventilator

Intelligent Ventilation solution from ICU to MRI

D Beyond essentials DRÄGER SAVINA 300 SELECT

Mechanical Ventilation

fabian HFO: Modular, Powerful, Versatile.

Selecting the Ventilator and the Mode. Chapter 6

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Supporting you in saving lives!

Using Common Ventilator Graphics to Provide Optimal Ventilation

evolution 3e Ventilators

Astral in AirView: Improving patient care through connectivity. ResMed.com

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

HAMILTON-T1 HAMILTON-T1. Intelligent transport ventilation

Neonatal tidal volume targeted ventilation

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Minimum size for maximum performance

Mechanical Ventilation

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

D Beyond essentials DRÄGER SAVINA 300 SELECT

Mechanical Ventilation

Astral in AirView: Improving patient care through connectivity. ResMed.com

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Flight Medical presents the F60

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

VENTILATION SERVO-i INSPIRATION IN EVERY BREATH

GE Healthcare. Aestiva/5. More than superior ventilation

HELP THEM BREATHE MORE NATURALLY.

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

SLE5000 Infant Ventilator with HFO

Operating Instructions for Microprocessor Controlled Ventilators

HAMILTON-C6. The next generation of intelligent ICU ventilators

Panther 5 Acute Care Ventilator

HAMILTON-G5. The modular high-end ventilation solution

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Focused on essentials DRÄGER SAVINA 300

MT Excellent performance and reliability EVITA 4 EDITION

Test lungs from IMT Analytics simulate a range of realistic ventilation situations during training.

Introduction to Conventional Ventilation

Blease900 Ventilator PATIENT CENTERED VENTILATION

Oxygen injection sites: Important factors that affect the fraction of inspired oxygen

GE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations

Initiation and Management of Airway Pressure Release Ventilation (APRV)

Managing Patient-Ventilator Interaction in Pediatrics

Test your PICU/ NICU/ CHSU Knowledge

Technical Data and Specifications

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

THE PURITAN BENNETT TM 980 VENTILATOR EDUCATION PRESENTATION

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE

A Bench Study of the Effects of Leak on Ventilator Performance During Noninvasive Ventilation

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Corporate Overview and Product Summary

Operator s Manual Addendum

A bench study comparison of volume delivery in the presence of leak

CHAPTER 02 CH02_FIGURE 15A A. Schematic of a simple oxygen mask. B. Oxygen mask in place on a

INTENSIVE CARE VENTILATORS

Health Professional Info

Principles of Mechanical Ventilation: A Graphics-Based Approach

Safe and Intuitive Neonatal Ventilation

MT The secret of harmonious breathing BABYLOG 8000 PLUS

VENTILATORS PURPOSE OBJECTIVES

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients

NAVA Neurally Adjusted Ventilatory Assist In Neonates

WILAmed INTENSA. Intensive Care Ventilator System

HAMILTON-C3. The compact high-end ventilator

V8800 Ventilator System. Your healthcare, we care.

Intelligent Ventilation. Hamilton Medical. Intelligent Ventilation since 1983

Made to meet your needs DRÄGER SAVINA 300

Software enhancements

Airox Supportair Ventilator

Meet the changing needs of your patients. Trilogy Series, with advanced ventilator technology

A few seconds after turning on, the screen changes to Set Up Screen Before adjusting the pressure, calibrate the O 2 sensors

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Focused on essentials DRÄGER SAVINA 300

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Giving neonates a future

D Designed for non-invasive ventilation DRÄGER CARINA

Conflict of Interest Disclosure Robert M Kacmarek

VENTIlogic LS VENTIlogic plus. 100 % Mobility and Reliability in IV and NIV

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB

Trust Guideline for Neonatal Volume Guarantee Ventilation (VGV)

Mechanical Ventilation. Flow-Triggering. Flow-Triggering. Advanced Concepts. Advanced Concepts in Mechanical Ventilation

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Oxygen Injection Site Affects F IO2 During Noninvasive Ventilation

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Patient Setup Information

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

TESTCHEST RESPIRATORY FLIGHT SIMULATOR SIMULATION CENTER MAINZ

Oxylog 2000 plus. Step up your performance. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

ReVel ventilator po r table critica l ca re ven t i l a t i on

Safe ventilation at rest and on the move

Safe ventilation at rest and on the move

Transcription:

The Puritan Bennett 980 Neonatal Ventilator System Helping to Protect Our Most Vulnerable NEWBORNS

2

Helping Provide Comfortable Care When newborns first weeks or months of life are spent in the NICU, it is emotionally challenging for the baby as well as the family. Everyone strives to keep these babies as comfortable as possible. However, routine care required for NICU patients can be a source of discomfort and stress. 1-3 3

The Challenge of Ventilatory Support One cause of patient discomfort can be the need for ventilatory support. The often weak respiratory drive and vulnerable lungs of babies can make effective ventilation difficult, and can lead to a potentially uncomfortable patient-ventilator interaction. 4, 5 To confirm that sufficient gas exchange is taking place, ventilated newborns require significantly more procedures. 2 Mechanically ventilated NICU patients are also unable to receive continual soothing parental contact, which can have a considerable impact on not only the baby s outcome, but also the parents satisfaction and well-being. 1 REDUCING THE NEED FOR INVASIVE VENTILATOR SUPPORT AND IMPROVING MECHANICAL VENTILATION WHEN POSSIBLE CAN HELP LIMIT EXPOSURE TO UNCOMFORTABLE PROCEDURES AND ENABLE NEWBORNS TO SPEND MORE TIME IN THE ARMS OF THEIR LOVED ONES. 1, 2 4

Managing Ventilatory Support Significant air leaks are common in newborns intubated with uncuffed tubes, and vary substantially over time because of changing patient conditions, sedation levels or muscle relaxation, patient position and wakefulness. 6 Year Author Key results 2010 Mahmoud 7 An air leak of 40% indicated the displayed Vt was underestimated by 1.2 ml/kg, thus by about 24% of target Vt (generally 5 ml/kg). 2001 Main 6 Leaks larger than 20% resulted in inconsistent tidal volume delivery and gross overestimation of compliance (C) and resistance (R) regardless of ventilator mode. 5

Noninvasive Ventilation Noninvasive ventilation is increasingly preferred to avoid the complications associated with intubation. 8 A clinician s ability to continue providing noninvasive support and avoid intubation is related in part to how well the newborn is tolerating noninvasive ventilation. 8 One factor contributing to a patient s tolerance of noninvasive ventilation is synchrony between the patient s spontaneous breathing efforts and the ventilator settings. 8 Large airway leaks are common with noninvasive interfaces, for example nasal masks, and can result in a variety of trigger and cycle asynchronies. Trigger asynchrony can lead to an increase in work of breathing for the baby. 9 6

What Leaks Can Mean for NICU Patients Leaks place ventilated patients at a higher risk of auto-triggering, and can cause the ventilator to provide unwanted breaths. 5, 8 To mitigate auto-triggering, clinicians could increase sensitivity settings on the ventilator, placing babies at risk of ineffective triggering, making it difficult for the baby to get a breath despite making an effort. 5, 8 Both ineffective triggering and auto-triggering can cause an increased 5, 7-9 work of breathing, distress and fatigue for the baby. FOR EXAMPLE Trigger sensitivity is set by the clinician at 0.5 liters per minute (Lpm). The baby has a leak of 0.2 Lpm. The effective trigger sensitivity is 0.3 Lpm. The baby s condition changes over time. As a result the leak changes to 0.6 Lpm. The ventilator will autocycle because the leak is greater than the set trigger sensitivity. Trigger sensitivity is changed by the clinician to 0.8 Lpm to compensate for the leak. The baby s leak decreases to 0.2 Lpm. Baby s inspiratory work of breathing increases significantly. 7

Engineered to Help Protect Your Most Vulnerable Patients from Leak-Related Asynchrony The all new Puritan Bennett 980 neonatal ventilator is designed to address the challenges of safe and effective neonatal ventilation. The Puritan Bennett Leak Sync software automatically detects and compensates for fluctuating leak sizes, resulting in a decreased risk of breath trigger and cycle asynchronies, helping clinicians keep babies at a more desirable work of breathing. It also results in more reliable displayed volume. 8

Leak Sync software Providing Safe and Responsive Care The Leak Sync software can compensate for leaks up to 15 Lpm in neonate applications, helping to ensure newborns receive the flow and volume selected by the clinician. This software works with multiple airway interfaces, helping provide peace of mind during both invasive and noninvasive neonatal ventilation. 10 Puritan Bennett Leak Compensation Performance During Invasive Ventilation 10 Invasive ventilation The Puritan Bennett 840 ventilator outperformed all other ventilators in this study by requiring fewer breaths to achieve synchronization during increasing and decreasing leaks. 10 Puritan Bennett Leak Compensation Performance During Noninvasive Ventilation 10 Noninvasive mechanical ventilation The Puritan Bennett 840 is one of the only two ventilators in this study that adapted well to increasing or decreasing leaks. 10 By effectively managing gas leaks during both invasive and noninvasive ventilation, the Puritan Bennett 980 ventilator can help clinicians provide the right ventilation, supporting the clinician s effort to keep babies safe and comfortable. 9

Advancing the Legacy The Puritan Bennett 980 ventilator was built on the reliability and sophisticated breath delivery technology clinicians have come to expect from Puritan Bennett ventilation. SOFTWARE YOU DEPEND ON Proximal Flow Sensor Measures lower flows, pressures and tidal volumes right at the patient wye for neonatal applications. C20/C Helps ensure patients don t suffer from over distention by monitoring compliance of the last 20% of each inspiration compared to the entire inspiration. Configurable FiO 2 elevation Elevates FiO 2 for two minutes according to the configurable % set between 21-100%. Trending Captures up to 72 hours and more than 57 parameters of patient data to help review the effectiveness of treatment over time. Security for You and Your Patients Our new ventilator assurance program includes: VENTILATOR ASSURANCE FEATURE The ventilator will continue to deliver ventilatory support in the event of certain system failures. STATUS DISPLAY There is an additional screen located on the breath delivery unit (BDU) with data display even if the graphic user interface (GUI) is unavailable. STANDBY MODE This feature pauses ventilation while the patient is disconnected from the ventilator and preserves settings; auto-detects patient upon reconnection and resumes ventilation. 10

Service You Can Trust THE COVIDIEN SERVICE DIFFERENCE Quality The Covidien Service team worked hand-in-hand with Engineering during the development of the Puritan Bennett 980 ventilator to ensure they are able to service the ventilator with the high level of quality that our customers have come to expect. Consistency The Covidien Service team operates on a solid foundation of experience and expertise, with more than 50 years as a qualified provider of service for Puritan Bennett ventilators. Responsiveness Covidien Service has more than 40 Customer Support Engineers across the country. Our fully integrated Sales, Service and Clinical Support team enables us to respond quickly to your service and account needs. Integrity Strict compliance with industry standards for Quality Management Systems and with our manufacturerrecommended service maintenance schedule is a priority for our Service team. Satisfying your needs with steadfast integrity enables us to build our relationship with you as a valued customer. 11

References 1. De Lima J, Carmo KB. Practical pain management in the neonate. Best Pract Res Clin Anaesthesiol. 2010;24(3):291-307. 2. Simons SH, van Dijk M, Anand KS, Roofthooft D, van Lingen RA, Tibboel D. Do we still hurt newborn babies? A prospective study of procedural pain and analgesia in neonates. Arch Pediatr Adolesc Med. 2003;157(11):1058-1064. 3. Bellù R, de Waal KA, Zanini R. Opioids for neonates receiving mechanical ventilation. Cochrane Database Syst Rev. 2008;23;(1):CD004212. 4. Bougatef A. Neonatal mechanical ventilation. In: Gullo A. Anaesthesia Pain Intensive Care and Emergency Medicine. Milan, Italy: Springer; 2005: 73-81. 5. Bernstein G, Knodel E, Heldt GP. Airway leak size in neonates and autocycling of three flowtriggered ventilators. Crit Care Med. 1995;23(10):1739-1744. 6. Main E, Castle R, Stocks J, James I, Hatch D. The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med. 2001;27(11):1788-1797. 7. Mahmoud RA, Proquitté H, Fawzy N, Bührer C, Schmalisch G. Tracheal tube airleak in clinical practice and impact on tidal volume measurement in ventilated neonates. Pediatr Crit Care Med. 2011;12(2):197-202. 8. Vignaux L, Vargas F, Roeseler J, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med. 2009;35(5):840-846. 9. Wilkins RL, Stoller JK, Scanlan CL. Egan s Fundamentals of Respiratory Care. 8th ed. Louis, MO: Mosby; 2003. 10. Oto J, Chenelle CT, Marchese AD, Kacmarek RM. A comparison of leak compensation in acute care ventilators during non-invasive and invasive ventilation; a lung model study. Respir Care. 2013; Available at http://rc.rcjournal.com/content/early/2013/05/21/respcare.02466.full. pdf+html. Accessed January 1, 2014. COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. 2013, 14 Covidien. 13-VE-0031(1) 6135 Gunbarrel Avenue Boulder, CO 80301 800-635-5267 covidien.com/rms