TYPICAL OPERATION PROCEDURES

Similar documents
Scanning Electron Microscope JEOL JSM F

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

SEM LEO 1550 MANUAL. I. Introduction. Generality

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

[needs to be discussed with faculty if any new material to be done]

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Thermo K-Alpha XPS Standard Operating Procedure

IVIS Spectrum Whole Animal Imager

David Cockayne Centre for Electron Microscopy.

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

1)! DO NOT PROCEED BEYOND THIS MARK

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date:

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

r β tr t t s t r s t s tr tt r t r t r r r tt r e s tt e tt r e s tt r e r r e r 2 t ö t t r s s t t r r r ts r rr t

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

OPERATION OF THE DIMPLER

Oerlikon Sputtering Evaporator SOP

MIMS PROCEDURES. Updated 10 Sept 2008

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Angstrom Dielectric Sputterer Operation Manual

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Standard Operating Procedure. For. PVD E-Beam

Arizona State University NanoFab XACTIX ETCHER. Rev A

Operating Procedures for Metal Evaporator I

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Standard Operating Procedure I PHI 5000 VersaProbe II XPS/UPS

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

CircuFlow Quick Setup Guide

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

5890II GC Standard Operating Procedure 9/2/2005

Armfield Distillation Column Operation Guidelines

Boyle s Law: Pressure-Volume Relationship in Gases

FLUORESCENCE DETERMINATION OF OXYGEN

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

Basic ICP Operating Procedures

Standard Operating Manual

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Approved by Principal Investigator Date: Approved by Super User: Date:

JASCO 810 CD SPECTROPLOARIMETER STANDARD OPERATING PROCEDURE

Approved by Principal Investigator Date: Approved by Super User: Date:

The MRL Furnaces USED FOR THIS MANUAL COVERS

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

SPUTTER STATION STANDARD OPERATING PROCEDURE

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

CONSUMER MODEL INSTALLATION GUIDE

Table of Content IMPORTANT NOTE: Before using this guide, please make sure you have already set up your settings in

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM

UNIVERSITY OF WATERLOO

[MYLAPS INTEGRATION]

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

RM-80 respiration monitor

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel

Lab 4: Pressure Gradients over a Wing

University of Minnesota, MN Nano Center Standard Operating Procedure

Free Golf Scorer. Installation: Page: 1 / 9 28/06/10

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Xactix XeF2 OPERATION MANUAL

MP15 Jockey Pump Controller

Using the Akta Prime plus October 22, 2012

Approach. S3 Owner s Manual. May _0B Printed in Taiwan

WATER CONTROL SYSTEM QUICK START

ECL Comfort 110, application 131 (valid as of software version 2.00)

Approach. owner s manual

Page 1 GM-FAQ Club Profile FAQs. Page

Approved by Principal Investigator Date: Approved by Super User: Date:

OAI Model 200 Tabletop Mask Aligner Portland State University

BVIS Beach Volleyball Information System

Experiment C-6 Gas Solubility

Cryo-Evaporator Operation

Drift-Chamber Gas System Controls Development for the CEBAF Large Acceptance Spectrometer

Wire Buddy Manual IOpath Software

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

COBILT CA-800 Mask Aligner Equipment Operation

[USING THE VITROBOT April 9, Using the Vitrobot

Mass Spec will not Autotune

Introduction. The Shearwater Petrel is an advanced technical diving computer for open and closed circuit divers.

LAMINAR FLAME SPEED ANALYSIS USING PIV (PARTICLE IMAGE VELOCIMETRY)

Cambridge NanoTech: Savannah S100. Table of Contents

Software for electronic scorekeeping of volleyball matches, developed and distributed by:

ACCOUNT MANAGER REFERENCE GUIDE

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Initiated Chemical Vapor Deposition

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

Xactix Xenon Difluoride Etcher

Standard Operating Manual

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Transcription:

JSM-7500F TYPICAL OPERATION PROCEDURES Reminders! Remember to sign in on the sign up sheet. WEAR GLOVES: always wear gloves when handling your samples, the sample holders and the specimen exchange rod. Ensure that samples are dry and free of excessive out gassing always discuss a new sample with the MCF staff before you use the SEM to appropriately prepare your sample. MAGNETIC SAMPLES: please discuss with Dr. Bisrat first they require different procedure than conventional non-magnetic samples. Login with your user name and password (Do not share your password with any body! The SEM computer keeps a record of who logged in.) Make sure you transfer your files into your own jump drive the same day. MCF will not back up your files and the USERS folder will be cleaned frequently without notice. Please acknowledge the use of the MCF when you publish your paper and e-mail me a copy for our records. 1. LOADING SAMPLE INTO CHAMBER (a) Sign up the login sheet: Please write your FULL NAME, your ADVISOR S FULL NAME and a complete ACCOUNT NUMBER legibly. (b) Login to windows with your NetId (NetId@tamu.edu) and NetId pass word. Then select your account and submit. (c) Start the SEM software by clicking on PC_SEM icon on the desktop and login. (d) Turn on the specimen IR chamber camera, IR Camera icon on located on the desk top. If it does not turn on, the procedure for setting up the camera can be found at the end of the SEM procedure manual on the SEM desk. (e) Ordinarily, the airlock chamber is under high vacuum (9.6x10-5 Pa). Before loading, the lights in the airlock button should be: VENT OFF; EVAC ON; HLDR OFF (f) Check that the stage is in the exchange position - EXCH POSN light on the side of the airlock should be ON and the stage parameters x,y, z, R & T should read: If not: x=0; y=0; z=38; R=0; T=0 LEFT click the grey box of any of the stage parameters (x, y, z, R, T) and type in the values manually and select move. (g) Put on your GLOVES and mount your sample on the sample holder. (h) Press and hold the VENT button until it starts to blink; the isolation valve closes and N2 gas vents into the airlock. (The VENT button can also be clicked from the SEM monitor window) (i) When the VENT light stops blinking, release the latch on the airlock to open the airlock door. (j) Load your sample, close the airlock and press the EVAC button until it starts to blink. 1

(k) Wait until the EVAC light stops blinking. Now the lights in the airlock buttons should be VENT OFF; EVAC ON; EXCH POSN ON; HLDR OFF (l) Fully insert the specimen exchange rod, keeping the holder horizontal until it comes to a firm stop and make sure the HLDR light is ON. (m) Retract the exchange rod FULLY and tilt up; now the lights on the airlock should be VENT OFF; EVAC ON; EXCH POSN ON; HLDR ON Remove your GLOVES now. (n) When the Select Specimen holder window pops up select the holder type and insert the specimen offset. Function Buttons Image information area Vacuum pressure Figure 1. PC- SEM window 2

2. FINDING THE DESIRED AREA ON THE SAMPLE (a) Wait until the Vacuum pressure reads 3.6x10-4 Pa or better. (b) Select/Input the value for Acceleration voltage (acc. vol.). Start with 5kV if you need to increase or decrease the acc. vol at some point during your work, make sure you do it in steps of 2kV to avoid damaging the gun. (c) Turn on the Acceleration voltage (circled in yellow on Fig.1): [ON] (d) Watch the Emission Current readout increase. Wait for proper Emission Current usually 10µA or 20A. (e) Check the Probe Current setting, use the following guideline FUNCTION BUTTONS (Fig.1.): [Observation] Probe Current (generally 8 is a good choice). (f) Use Low Magnification Mode (LM) to find area of interest on sample and have it centered. (KNOBSET on the operational panel : LOW MAG knob) (g) Auto adjust contrast and Brightness (KNOBSET on the operational panel : ACB button) (h) Focus just enough to identify site of interest on your sample (KNOBSET on the operational panel : FOCUS knob) (i) Turn off LM and select SEM Mode (LOW MAG button light off). 3. SETTING THE APPROPRIATE WORKING DISTANCE (WD) Be mindful of the actual distance between the top of the sample and the lens! Make sure the IR camera is on whenever you are making any Z movement or inserting and retracting the LABE detector. (a) Select the desired WD IMAGE INFO AREA: left click [WD] and select (WD= 15mm is recommended for a start) (image info area is the black bottom of the image viewing part of the screen as shown in Fig.1) (b) Make sure you are at the lowest magnification possible for the selected WD in SEM mode. (c) Move stage from the exchange position close to the selected WD as follows STAGE INFO AREA: Left click Z, input a value close to the desired value (WD + offset+ 3mm) (d) Adjust contrast and brightness using auto contrast/brightness, ACB. (e) If the contrast of the image is too high (if the image looks too bright or too dark), change your detector to LEI (circled in blue in Fig.1). If the image looks too dark (not enough electron yield), you will need to increase the probe current. (f) Bring image into focus with Stage Z control (manual control of stage) as follows: Activate [ZFC] (circled in red on Fig.1) use TRACKBALL RING until the sample surface comes into focus. (g) Turn off Stage focusing [ZFC] 3

4. ALIGNING THE BEAM using the Operation Panel (a) Set magnification to ~5000X 50000X KNOBSET: MAGNIFICATION knob (b) Focus the image the best you can KNOBSET: FOCUS knob (c) Turn Wobbler on Operation Panel : WOBB button. (d) If image shifts, adjust to stop image shifting KNOBSET: X, Y knob (e) Turn Wobbler off Operation Panel : WOBB button (f) Increase the magnification and then focus to check for any astigmation Using the FOCUS knob, check for astigmatism by going through, over & under focus while looking for directionality (or streaking) of focus in the image (over and under focus directionality will be at right angle to each other). (g) Correct Astigmatism if necessary see step 5 below 5. CORRECTING THE ASTIGMATION using the Operation Panel (a) Stop focus at center of over and under focus (image may not be sharp but has no directionality of focus). (b) Adjust the X & Y knobs on the operation panel to obtain the sharpest image. (c) Increase magnification and focus the image with the FOCUS knob checking for astigmation. (d) Usually several iterations of steps (a) (c) are required before acquiring a crisp image. Note: Focusing and astigmation are generally necessary for any change in operating parameters except for magnification. In fact, for a good image acquisition it is best to adjust these controls at an exaggerated magnification (at least 50kX more than the desired magnification), then reduce the magnification to the desired level. 6. TAKING PHOTOS OF IMAGES Acquisition of images can be performed using the FREEZE or NORMAL mode. Once acquisition is completed, save it in your folder. Create a folder with your full name at C: images\users\ (a) Using the NORMAL mode (recommended for non-charging specimens) (i) Set exposure visually KNOBSET: ACB button and/or IMAGE CONTRAST, BRIGHTNESS knobs (ii) Press PHOTO on the KNOBSET or NORMAL on the software button for slow acquisition Important! make sure the Export box is checked in the save as window to ensure all the image information is saved as part of the image. 4

(b) Using the FREEZE mode (recommended for charging specimens) (i) Check if the Freeze button set to integration box is checked (setup\ operation settings\image scan\integration) (ii) Press the FREEZE button (it will blink until it finishes acquisition) (iii) Once it stops blinking, go to FILE SAVE AS and save your file in the appropriate folder Important! You can use the LUT (Look Up Table) to adjust the contrast and brightness of the frozen image before saving it. 7. SHUTTING DOWN (a) Exit INCA software first if you have run EDS analysis. (b) Turn the Accelerating Voltage off [Observation]:[OFF] (c) Make sure the IR camera is turned on. (d) Retract the LABE detector if used 5 uncheck the LABE check box (Note: Make sure that the top of your sample is at least 8mm away before you retract the LABE.) (e) Move the stage to the exchange position a. Click the Function button: Specimen button in the main menu b. Click the Specimen Exchange button in the SEM monitor window Make sure the stage parameters read as follows: x=0; y=0; z=38; R=0; T=0 (f) Before unloading, the lights in the airlock button should be: VENT OFF; EVAC ON; EXCH POSN ON; HLDR ON (g) Wear gloves! Fully insert the specimen exchange rod keeping it horizontal until you feel it comes to a firm stop. (h) Fully retract the specimen exchange rod and confirm that the HOLDR light goes out. (i) When the rod is FULLY retracted, tilt up the exchange rod and secure it to the black clip. (j) Press and hold the VENT button momentarily until it starts to blink (the isolation valve closes; N2 gas vents into the airlock). (k) When the VENT button light stops blinking, unlock the airlock chamber by releasing the latch. Open the airlock door. (l) Remove the specimen. (m) Close and lock the airlock chamber and press the EVAC button until the light starts blinking. When the EVAC light stops blinking, the airlock button should be: VENT OFF; EVAC ON; EXCH POSN ON; HLDR OFF (n) Make sure the pump is pumping down well and is in the 10-3 Pa range. (o) Exit the SEM monitor: go to FILE Exit and turn off the IR Camera. (p) Exit the PC_SEM software. (q) Log off windows and remove your user name. (r) Complete the login sheet.

BACKSCATTERED ELECTRON (LABE) IMAGING 1. Make sure you have the camera on. 2. Make sure there is at least 8mm gap between the objective lens and your sample surface. If you have been working at shorter WD, then lower the stage to have 8mm gap using ZFC before you insert the LABE detector. (Once you insert the detector, use ZFC to bring up the stage to your desired WD.) 3. Insert LABE Detector in: in the SEM Monitor Window select the LABE detector check box (circled in green on Fig. 1.), wait till the Gun valve opens. 4. First using the SEI mode, (a) choose the sample location (b) adjust magnification, focus and astigmatism (c) adjust contrast and brightness (ACB) (d) make sure the IR Camera is off (other wise it will blind the LABE detector and you won t be able to see any signal, you will see a white screen) 5. Now set detector to LABE: IMAGE INFO AREA: [SEI] [LABE] 6. Set scan speed to slow: KNOBSET: FINE VIEW button 7. Adjust the Brightness and Contrast 8. If the contrast is too weak, it may be necessary to increase the Probe Current to a large number, at least PC 10, or you may need to increase the Acc. voltage. 9. View and/or acquire BE image using the FREEZE or PHOTO mode. 10. When finished retract the LABE detector: in the SEM Monitor Window uncheck the LABE detector box (Make sure you have at least 8mm gap between the objective lens and your sample surface.) 11. Set detector selector back to SEI: IMAGE INFO AREA: [LABE] [SEI] 6

ADDITIONAL OPERATIONAL STEPS AND TIPS (Jeol Applications Laboratory, p.93) 1. CHOOSING BETWEEN THE TWO SECONDARY ELECTRON DETECTORS (LEI) Vs (SEI) -for best/high resolution: SEI especially at Low KV -for good resolution: SEI, LEI 4.5 6mm WD 8mm WD 2. REDUCING THE EFFECT OF SAMPLE CHARGING IN THE IMAGE - Reduce the accelerating voltage: Make sure you change the acceleration voltage in steps of 2kV and wait for few seconds for it to stabilize between each step. (Otherwise, it will shorten the lifetime of the gun or even damage the gun) - Use LABE detector (since BEs have more energy than SEs your image is less susceptible to Charging - One can also use a BE imaging at low accelerating voltage (at or < 2kV) using GB mode - Reduce the probe current [Observation]:[Condition]:Probe current slider - Use integration to acquire image KNOBSET: FREEZE button - Use the Lower Detector (LEI) For WD longer than 6mm - Change to GB mode (see step 3 below) - Reduce the emission current Emission Current: [select desired current 20, 10, 5, 2] 3. WHAT TO DO IF SIGNAL IS WEAK If your image looks darker or not clear even after adjusting the contrast and brightness, you may have weak signal (not enough electron yield) and so try the following: - Check if the emission current has reduced from what you have started, if so reset by clicking the reset button - increase the probe current - increase emission current to 20A if using different - if working at WD > 6mm, try using LEI instead of SEI 7

MCF Acknowledgement Policy The TAMU MCF provides instrumentation and consulting services that address the needs of the TAMU community involved in teaching and research related to materials and the life sciences. To facilitate interdisciplinary interactions between the TAMU community and the MCF staff, the following policy regarding acknowledgement of MCF contributions is now in effect. In addition, this acknowledgement helps the MCF to demonstrate to the funding agencies - the Division of Research, the College of Engineering (TEES), and the College of Science, the significance of their continued support of the MCF MCF users who have incorporated data obtained from MCF resources into publications and presentations will acknowledge the MCF facilities and personnel who contributed. The MCF recognizes that samples and all research materials involved in these investigations are the property of the Principal Investigator (PI), who serves as the responsible individual representing the University on funded projects. Just as the intellectual property at the heart of the investigation may be owned by the PI, so the ideas for new methods developed by the MCF staff may be the intellectual property of the Technical Staff, to be used with their authorization. In many cases, techniques will be developed as the result of joint efforts. Ideally, these different facets will augment each other and can be published together or separately as appropriate. In any case, the two parties should discuss the best way to disseminate the information and must recognize and respect each other s' contributions either by authorship or acknowledgement. In the case of co-authorship, a contribution to the writing of a paper is both a privilege and requirement. Cases of disagreement or potential conflict arising from prior agreements (or more likely, lack of prior agreements) will be resolved through the Office of the Vice-President for Research in consultation with the Director of the MCF, or an ad hoc committee of their choosing. Research carried out in part or in full using MCF facilities will thus fall into one of three categories, with services and/or contributions requiring the following acknowledgements: o o o Use of facilities only (PI and associates carry out the research) requires acknowledgement of facilities in all resulting publications and presentations. The following acknowledgement should be used: "Use of the TAMU Materials Characterization Facility is acknowledged." Provision of routine procedures and methods by MCF Staff requires acknowledgement of facilities and individuals in the resulting publications. The following acknowledgement should be used: "Use of the TAMU Materials Characterization Facility and (Dr./Mr./Ms.) are acknowledged." Contributions of MCF Staff that include development and implementation of original techniques, experimental design, interpretation, or other significant intellectual input require co-authorship and acknowledgement of facilities. The following acknowledgement should be used: "Use of the TAMU Materials Characterization Facility is acknowledged." 8