VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

Similar documents
Kinetics of Inhaled Anesthetic Agents

Anesthesia Machine Intro

Anes Machine Use and Pre-use Check - CA1 focus

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

New Anesthesia Delivery Systems What Do I Need to Know About My New Machine?

Understanding Rebreathing Kinetics

QED-100 Clinical Brief

Anesthesia Machines. Inhaled Anesthetic Agents

Historically, rebreathing anesthesia systems

Anesthesiology Research and Scholarship. Follow this and additional works at:

Draeger Fabius GS Pre-Use Check

Breathing Circuits. Product training

Errors in Monitoring. BWH Clinical Conference 10/06/04. Copyright 2004, James H Philip, all rights reserved.

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

What our equipment teaches us

Materials and Methods

Checking a Breathing System

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

How Technology Will Secure the Future of Inhalation Anesthesia. How Technology Will Secure the Future of Inhalation Anesthesia. Jan Hendrickx, MD, PhD

Fabius will not compensate for Lung Compliance or Changes in lung compliance

ROUTINE PREOXYGENATION

Unit 15 Manual Resuscitators

Technology Topics 2017 James H Philip MEE, MD, CCE page 1

GE Healthcare. Avance Carestation. Upgrade promotions guide

Flow meter. bellow vaporizer. APL valve. Scavenging system

Selecting and Connecting Breathing Systems

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Breathing Systems. Professor Khalid Bashir

S/5 Aespire. Quick Reference Guide

Anesthesia. Midmark Animal Health Products

Comparison Of The Efficacy Of Two Low Fresh Gas Flow Techniques In Low Flow Anaesthesia

NARKOVET ANAESTHETIC UNIT

Recommendations on Checking Anaesthesia Delivery Systems

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Figure 5. Nasal mask in place

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Effect of Fresh Gas Flow on Isoflurane Concentrations during Low-flow Anaesthesia

Medical Ventilators. Presented by: Edwin Lim

LAB 7 HUMAN RESPIRATORY LAB. Complete the charts on pgs. 67 and 68 and read directions for using BIOPAC

Selecting the Ventilator and the Mode. Chapter 6

Monitoring, Ventilation & Capnography

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Keeping anaesthetic pollution to minimum in theatre environment

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Name: Oasis: Questions EPCP. Professional Development: ETCO2 Monitoring

Aespire 7100 Essential performance Compact design

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

PROBLEM SET 8. SOLUTIONS April 15, 2004

KOALA. Disposable Adult Anaesthetic Circuits for Low Flow Anaesthesia KO.S5.22MM. Adult Anaesthetic Circuits. Medical Pty Ltd

LISTADO DE CIRCUITOS DE PACIENTE

Performance Redefined. Automatic Gas Control (AGC) FLOW-i Anesthesia Delivery System

HYPERCAPNIC HYPERVENTILATION SPEEDS EMERGENCE FROM INHALED ANESTHESIA. Nishant A Gopalakrishnan

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

THE GE AISYS ANAESTHETIC MACHINE

GE Healthcare. Aestiva/5. More than superior ventilation

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

Nitrous oxide Quiz 2017 Answers

Canine Endotracheal Intubation

We Warm Animals. User Manual For. Darvall Heated Smooth-Wall Circuits. Darvall Heated ZDS Qubes

Breathing Process: Inhalation

Disclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14

State-of-the-art anesthesia technology

CARBON DIOXIDE ELIMINATION FROM SEMICLOSED SYSTEMS

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

New A7 Anesthesia Workstation

Your patients, our primary focus. Siesta i Whispa anesthesia machine

Technology Insights for protective ventilation and spontaneous breathing in the OR

Patient Information for the: Humanitarian Device for use in the Control of Air Leaks

McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

ekom Compressor for Medical Air Sonata with Ekom

Hallowell EMC Vacuum / Scavenging Interface

Lung Volumes and Capacities

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

Anesthesia Workstation

Hallowell EMC Anesthesia WorkStation Operating Manual

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

Fiberoptic Intubation Made Easi(er) Know Your Scope. Indications. Christine Whitten MD Department of Anesthesia Kaiser Permanente San Diego

Guidelines on Monitoring in Anaesthesia

Mechanical Ventilation

Your patients, our primary focus

Fabius Tiro TM. Designed for your environment. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Mechanical Ventilation. Flow-Triggering. Flow-Triggering. Advanced Concepts. Advanced Concepts in Mechanical Ventilation

ACUC ۰ ROUTINE PROCEDURE

Respiratory Signs: Tachypnea (RR>30/min), Desaturation, Shallow breathing, Use of accessory muscles Breathing sound: Wheezing, Rhonchi, Crepitation.

Merlin Training for Vets and Nurses

VT900A + VAPOR. Technical data. One Solution. Auto-detection to Ensure Patient Safety. Key benefits and features

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

M5400 Operators Manual

APNOEA AND PRE-OXYGENATION

Isocapnic hyperpnea (IH): fast-filtering foreign gases

FRESH GAS UTILIZATION OF EIGHT CIRCLE SYSTEMS

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

27/10/58. Background and Rationale. Background and Rationale. Background and Rationale. Objective. Background and Rationale

Technical Data and Specifications

Transcription:

VIMA (Volatile Induction and Maintenance Anesthesia) How and Why James H. Philip M.E.(E), M.D. Copyright 1995-2007, James H Philip, all rights reserved

VIMA (Volatile Induction and Maintenance Anesthesia) How and Why James H. Philip, M.E.(E.), M.D. Anesthesiologist and Director of Bioengineering, Brigham and Women's Hospital Associate Professor of Anaesthesia, Harvard Medical School Medical Liaison, Partners Biomedical Engineering

I will use Gas Man Info: gasmanweb.com

Three unique uses for Sevoflurane Almost unique Sevoflurane Sedation Sevoflurane for a difficult airway Sevoflurane Vital Capacity Induction

Sevoflurane Sedation 2 L/min FGF Sevoflurane 2% PO Oral Ibuprofen (Advil, Motrin) $0.50

Why do VCI? It works No need for other drugs IV Induction agent (Pent/Prop) Opioid Muscle relaxant (any NMB) Reversal agents for NMB

No relaxants No Propofol

No relaxants No Propofol No use

No relaxants No Propofol No use No waste

Why do VIMA? Inexpensive Satisfying to Patient Anesthesiologist Surgeon

How to do Volatile Induction

Premedication None, typically Anything, as required Try talking, first

Sedation possibility Propofol 10 mg BWH Pharmacy makes 10 ml syringes 2 ml syringes would be better

Volatile Induction Fast* Volatile agent alone, with or without nitrous oxide Deep breath 51 ± 4 seconds, 1-4 breaths 39 ± 3 seconds, if 1 breath * Philip BK, Lombard L, Calalang I, Philip J. Sevoflurane Vital Capacity Induction Compared with Propofol Intravenous Induction for Adult Ambulatory Anesthesia. Anesth & Analg 87:623-7, 1999.

2 minute Preparation Volatile Prime for anesthesia circuit Oxygen Prime for lungs Mental Prime for patient

Volatile Prime Goal Prime the circuit with 8% Sevoflurane 66-75% N 2 O

Semi-Closed Circuit Volatile Prime 8% Sevoflurane 1 L / min O 2 3 L / min N 2 O Prime for 2 minutes Same time as pre-oxygenation

Volatile Prime Required to provide sudden, large, controlled increase in inspired anesthetic concentration Inspired Control by any means is OK New machines and techniques may make this easier

How to do Volatile Prime

Volatile Prime Empty the circuit FGF = 0 - flows all off Open relief (APL, pop-off) valve Squeeze bag to empty circuit Attach mask Seal mask Glove, stopper, cork, plug, other

One way - seal mask with glove

Another way - seal mask and circuit with tightfiting occluder

Prime the Circuit Empty the circuit first FGF=0 Attach mask Seal mask Open APL valve fully Empty reservoir bag through APL

Prime the Circuit Empty the circuit first FGF=0 Attach mask Seal mask with glove Open APL valve fully Empty reservoir bag through APL Let reservoir bag fill

Normal flow in breathing circuit Flow goes forward In the direction of the valves Emptying and filling reservoir bag and lungs, alternately

Normal flow in the breathing circuit - Inspiration Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Normal flow in the breathing circuit - Expiration Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Priming the Circuit Flow goes backwards Blocked by each valve Filling the reservoir

Back-fill the breathing circuit (#1) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Back-fill the breathing circuit (#1) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Back-fill the breathing circuit (#1) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Empty the reservoir bag (#1) Exhaust Squeeze CO 2 Absorbant Sampled 200 ml/min Fresh

Fill the reservoir bag again (#2) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Empty the reservoir bag (#2) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Fill the reservoir bag again (#3) Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Close the Exhaust Valve Exhaust CO 2 Absorbant Sampled 200 ml/min Fresh

Empty reservoir bag (#3) and fill the inspired limb Closed Exhaust CO 2 Squeeze Absorbant Sampled 200 ml/min Fresh

Empty reservoir bag (#3) and fill the inspired limb Closed Exhaust CO 2 Squeeze Absorbant Sampled 200 ml/min Fresh

Empty reservoir bag (#3) and fill the inspired limb Closed Exhaust CO 2 Squeeze Absorbant Sampled 200 ml/min Fresh

Empty reservoir bag (#3) and fill the inspired limb Closed Exhaust CO 2 Squeeze Absorbant Sampled 200 ml/min Fresh

Empty reservoir bag (#3) and fill the inspired limb Closed Exhaust CO 2 Squeeze Absorbant Sampled 200 ml/min Fresh

Open the Exhaust - Fill the Expired Limb (#3) CO 2 Release Absorbant Sampled 200 ml/min Fresh

Open the Exhaust - Fill the Expired Limb (#3) CO 2 Release Absorbant Sampled 200 ml/min Fresh

Remove Glove, Apply mask to patient (#4) Exhaust CO 2 Release Absorbant Sampled 200 ml/min Fresh Patient

Circuit is ready Gas Monitor shows Circuit is Ready

Circuit is ready Gas Monitor shows Circuit is Ready After third bag-fill

Sample to gas monitor Here is how to test when ckt is ready

Bag # Sample from Reservoir Bag Attachment

2 minute Preparation Volatile Prime for anesthesia circuit Oxygen Prime for lungs Mental Prime for patient

While priming circuit Preoxygenate

Simple Circuit

Simple parts

Simple parts Oxygen source Reservoir bag (with cut end) 15-22mm adapter Sampling (oxygenating) Elbow Mask

Option in some countries: SiBI Connector Facilitates pre-oxygenation during circuit prime We have some at BWH Skip the next 4 slides to skip this section

SiBI(TM) Connector

SiBI Close-up

November 4, 2001 SiBI

South Lake Regional Hosp Ont 01/22/02

Peak Inspired Agent Monitoring Some monitors report Inspired = Peak Inspired This is a problem if Mean and Peak differ Mean is representative of what affects uptake GE-Datex-Ohmeda ADU puts FGF after Insp Valve This makes Mean and Peak differ greatly

Normal Breathing Circuit FGF path CO 2 Absorbant E Most ADS I

GE ADU FGF path CO 2 Absorbant I,E ADU

GE ADU wave data

Exp SiBI -> Ckt Insp Exp Insp

Exp SiBI -> Ckt Mean Inspired = 2%, different from 7.1% peak Insp Exp Insp

Anesthetizes like 2%, not 7%

2 minute Preparation Volatile Prime for anesthesia circuit Oxygen Prime for lungs Mental Prime for patient

Mental Prime Patient Instruct and practice VC VC = Vital Capacity breath Breathe all the way out Take a deep breath in Hold it in Maintain positive verbal contact Create positive experience through words and actions

2 minute Preparation is done Volatile Prime for anesthesia circuit Oxygen Prime for lungs Mental Prime for patient Ready for Volatile Induction

VI - Volatile Induction Say and Do Say: Exhale Do: Switch to Sevoflurane mask Say: Inhale and hold inspiration Do: Maintain positive verbal contact Maintain high (7%) inspired sevoflurane conc.

> One Breath Induction If patient breathes out Instruct patient to take another deep breath and hold it Take another deep breath and hold it Take another deep breath and hold it

Data from BWH First Series 18 patients with perfect data for analysis In first 20 seconds after VCI Sevoflurane %M SD Inspired = 6.2 + 0.8 Expired = 4.2 + 0.8 Nitrous Oxide Mean = 58 + 6 Most lost lash reflex in 1 breath Some required 2 breaths Few required 3 breaths

Anesthetize and Oxygenate together VCI Sevo 8% in Oxygen Slightly slower No Second Gas Effect No additional effect of N2O Always quite adequate Allows same settings from VCI to airway securement (Tracheal Tube, LMA)

Volatile Induction Phase Two Deepen Inhalation Anesthesia After loss of consciousness in 1 min Continue induction with 8% Sevo Assist or control ventilation Prepare to secure airway if desired O 2 + Sevo Secure airway LMA OK in 2 minutes ETT OK in 3-5 minutes (no relaxant)

Produce apnea before removing mask Hyperventilate to apnea Avoid losing anesthetic from lungs blood brain

Or, keep anesthesia light Natural airway or Minor airway interventions Oropharyngeal airway Nasopharyngeal airway Deepen anesthesia before incision

Monitor Carefully Often BP q 1 minute (for a while) Always automatic

Induction is complete Maintenance begins

VM - Volatile Maintenance Anesthesia We have always done this Maintain desired anesthetic depth Decrease fresh gas flows to save cost Increase vaporizer setting Maintain previous expired anesthetic conc. Maintain previous anesthetic depth Increase and decrease depth as required React or anticipate

VM - Volatile Maintenance Anesthesia Depth = Constant + Adjust as required Three anesthetic conditions Constant Increase Decrease

VM - Volatile Maintenance Anesthesia Depth = Constant + Adjust as required Three anesthetic conditions Constant Increase Vaporizer to Max = 8% Decrease Vaporizer off = 0% Increase FGF if required Readjust vaporizer to near prior setting Overpressure Underpressure

-30 Up and down using Overpressure and Underpressure -20 time (minutes) I E -10 10 8 6 4 2 0

For short cases, remember Awakening begins just after induction Don t be afraid to turn the vaporizer off Bring expired concentration down to 1.3 MAC Keep expired at 0.7 MAC near the very end

Prepare for recovery - always Recovery begins with wake up Recovery ends with patient back to normal Normal means normal - Patient is satisfied Anesthesiologist is satisfied Surgeon is satisfied

Volatile Emergence Volatile emergence is smooth Create and maintain zero (0.0) inspired agent concentration Empty the Reservoir Bag, Flush O 2, High FGF Ventilate anesthetic out of lungs Minute ventilation > 4 L/min p ET CO 2 around 36 mmhg

FGF is low Problems occur with wake up if FGF or Ventilation is low Poor circuit Increase FGF wash out; FGF too low

Problems occur Ventilation with wake up if FGF is or Ventilation low is low Poor lung wash out. Ventilation is too low

Better wake up w experience Usually, wake up is better Low inspired Low expired

1.0 A/I.8 Volatile Induction Zerothane Des 0.88 Sev 0.83 0.71.6 Iso 0.56.4 Hal F E F I.2 Data from Yasuda & Eger, 1991 0.0 0 Minutes of administration 30

1.0 0 Minutes of discontinuation 30.8 Volatile Emergence.6 After a long anesthetic Hal F E F E0.4 Iso 0.44.2 A/I 0 0.0 Sev Des Zerothane Alveolar response to inspired = 0 0.29 0.17 0.12

1.0 0 Minutes of discontinuation 30.8 Volatile Emergence.6 After a long anesthetic Hal F E F E0.4 Iso 0.44 0.33 0.29 MAC Awake.2 A/I 0 0.0 Sev Des Zerothane Alveolar response to inspired = 0 0.17 0.12

1.0 0 Minutes of discontinuation 30.8 Volatile Emergence.6 After a long anesthetic Hal F E F E0.4 Iso 0.44 0.33 0.29 MAC Awake 0.20 A/I 0 0.0 Sev Des Zerothane MAC climb to strether (approx) Alveolar response to inspired = 0 0.17 0.12

Clinical Benefits of VIMA No need for IV Induction agent Opioid Muscle relaxant (NMB) Reversal agents for NMB Other drugs Complications Avoid Side Effects Avoid Costs

VIMA provides Deep anesthesia from beginning to end No additional drugs for intubation No additional drugs for intra-op pain Rapid awakening and return to normal Post-op pain requires local anesthesia infiltration anesthesia and post-op analgesic

How to avoid slow wake ups

FGF is low Problems occur with wake up if FGF or Ventilation is low Poor circuit Increase FGF wash out; FGF too low

Problems occur Ventilation with wake up if FGF is or Ventilation low is low Poor lung wash out. Ventilation is too low

Better wake up w experience Usually, wake up is better

Better wake up w experience Usually, wake up is better Low inspired Low expired

Summary VIMA is Volatile Induction and Maintenance Anesthesia. VIMA is an acceptable equivalent alternative to IV plus Volatile combinations. VIMA preparation requires 2 minutes for circuit anesthetic prime, lung oxygen prime, and brain psychological prime. VIMA Sevoflurane induction is faster than IV Propofol (51 ± 4 sec vs. 81 ± 12 sec). VIMA wake up is as fast and uncomplicated as Propofol induction and VM anesthesia.

Thank you Jphilip@partners.org