Design and development of microcontroller based human airway pressure measurement system with real time graphical display for anaesthesia ventilator

Similar documents
Selecting the right pressure sensor for your application

AMS 6916 Board mount pressure sensor with ratiometric analog output

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18

AMS 6915 Board mount pressure sensor with digital output (I²C)

628 Differential Pressure Decay Leak Tester

AMS 2710 PCB pressure sensor module with V output

THE PRESSURE SIGNAL CALIBRATION TECHNOLOGY OF THE COMPREHENSIVE TEST

Pressure Sensor Bridge Configurations

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

GasSense NDIR User Manual

Technical Explanation

Precision level sensing with low-pressure module MS

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

RM-80 respiration monitor

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

HumiSys HF High Flow RH Generator

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Recommendations on Checking Anaesthesia Delivery Systems

The Development of A System for Measuring the Level of LP Gas in the Home

CARDIOVIT AT-104 ergospirometry

Technical Data Sheet MF010-O-LC

Application Note 183. Page 1 of Aero Camino, Goleta, CA Tel (805) Fax (805)

SHORT COMMUNICATION A SIMPLE, SENSITIVE AND VERSATILE SOLID-STATE PRESSURE TRANSDUCER

Mitos Fluika Pressure and Vacuum Pumps Datasheet

BUBBLER CONTROL SYSTEM

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Digital Pressure Gauge Main features: Data Sheet: DFP.418.R4.EN

Crystal Breath. High Level Design. Nicholas Castro, Suong Do, Joe Duffy, Joe Lervi, John Mullaney

MEGAS 2.0. Gas analysis with direct C-Level Calculation and Bus-Connection. Data sheet

Long term stability tests of INO RPC prototypes

A9600 Anesthesia Workstation. Your healthcare, we care.

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

fabian HFO: Modular, Powerful, Versatile.

UNDERSTANDING, ACCELERATED

BUBBLER CONTROL SYSTEM

Final Report. Remote Fencing Scoreboard Gator FenceBox

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

Anesthesia Workstation

Code Basic module and level control complete with optionals code

Technical Note. PendoTECH Single Use Pressure Sensors: Calibration, Accuracy, and Implementation. Introduction

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

P-5215 Differential Pressure Transmitter

Pulmonary Capacity Analyzer

3.0 Pressure Transmitter Selection

Safe Accurate Reliable. CWM-301 Anesthesia System

Panther 5 Acute Care Ventilator

Pressure measurement. Absolute pressure D Gauge pressure C Atmospheric pressure

Pro-V Multivariable Flowmeter Model M22 In-line Vortex

SX150A. Features. Typical Applications. Description

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

Aespire 7100 Essential performance Compact design

BiPAP Vision Ventilator Testing with Various Patient Interfaces Report February 18, 2002

Electronic Automatic Transmission for Bicycle

Implementation of Height Measurement System Based on Pressure Sensor BMP085

Design Features. General Description

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

DATA SHEET. VENTILATION SERVO-s

THE GE AISYS ANAESTHETIC MACHINE

HASTAC High stability Altimeter SysTem for Air data Computers

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

SCIENTIFIC DATA SYSTEMS, INC. Depth Tension Line Speed Panel. DTLS Manual

Milestones of Aeonmed Ventilators

SLE5000 Infant Ventilator with HFO

PC10 Piezoresistive Silicon Pressure Sensor

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

V8800 Ventilator System. Your healthcare, we care.

How Ventilators Work. Chapter 3

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

Load Station for Air Gas, CO and N O. Flow Instruments & Engineering GmbH Heiligenstock 34 c-f D Solingen

Cover Page for Lab Report Group Portion. Pump Performance

MS52XX-C&T SMD Pressure Sensor

TH-1800-T TRX-700-CNG Converter

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

Flow meter. bellow vaporizer. APL valve. Scavenging system

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

Constant Pressure Inlet (CCN) Operator Manual

86BSD Backside Digital Output

e360t and e360t+ Model

Automated Unified Trolley System for LPG Leakage Detection with Safety Measures and Refill Booking

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

Medical Instruments in the Developing World

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

Automatic Isokinetic Sampler. ST5 Evo

Manual of SF6 Comprehensive Tester

Narkomed MRI-2. A perfect fit for the MRI environment. Because you care. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Transcription:

Journal of Scientific & Industrial Research Vol. 65, December 2006, pp. 986-991 Design and development of microcontroller based human airway pressure measurement system with real time graphical display for anaesthesia ventilator R N Sengupta, Jaspreet Kaur* and Jagdish Kumar Central Scientific Instruments Organisation, Sector 30, Chandigarh 160 030 Received 08 December 2005; accepted 21 July 2006 A microcontroller based hardware design and system software associated with the graphical visualization of the airway pressure data acquisition system has been designed and developed, and successfully incorporated in indigenous prototype of anaesthesia ventilator. An inexpensive airway pressure monitoring system based around 80C31 microcontroller with RAM, EPROM, AD574 ADC, peripheral interface devices and graphical LCD etc., is a module of the indigenously developed instrument. What has been envisaged here is to get a real time pressure waveform on LCD. Air pressure changes over a respiratory cycle. This makes anaesthetist to have a feeling of how air pressure in the tract changes with time. An accurate and sensitive piezoresistive type of air pressure sensor has been used outside the mouth and inside the piping. Keywords: Airway pressure, Anaesthesia ventilator, Microcontroller, Piezoresistive pressure, Sensor IPC Code: G12B7/00 Introduction Ventilators 1 of different types in functional characteristics and construction have been used for anaesthesia and intensive care. The device used earlier in a ventilator has been a pressure pick up system 2, 3 from the patient s mouth to the transducer and then to implement at suitable display device for it. A dial type pressure indicator with a pointer at the centre was being used for a long time and still in vogue as this has been a convenient and reliable method of air pressure measurement where the Pascal s law of transmittance of air pressure is used. Measuring and displaying airway pressure (AP) waveform on front panel helps in monitoring functions such as disconnection in system, unrecognised leaks in the ventilator circuit 4. At the same time, a reduction in AP to the level of atmospheric pressure would indicate disconnection of device. By displaying real time airway graph, it is possible to realize the rise and fall of air pressure over a breathing cycle and their trends (Fig. 1). This paper presents design and development of AP measurement system with real time graphical display for anaesthesia ventilator. *Author for correspondence Tel: 0172-2657811, 2651831/Extn 467 E-mail: jaspreet_k123@rediffmail.com Materials and Methods Instruments Anaesthesia ventilator developed is an electropneumatic type of instrument, which controls tidal volume, respiration rate, inspiration flow rate, AP, I: E ratio etc. It has three basic subsystems: 1) Bellow system supported by pneumatic driving components; 2) Microcontroller based control system for ventilator parameter control, monitoring and alarms for safety; and 3) A closed circuit rebreathing circle absorber type patient circuit. Ventilators provide the force that inflates lungs, and with controlled ventilation 5, also determine respiratory time cycle. AP is monitored in new design through a LED bargraph display using piezoresistive Fig. 1 Schematic of the inspiratory and expiratory airway pressure versus time waveform

SENGUPTA et al: MICROCONTROLLER TO MEASURE AIRWAY PRESSURE OF ANAESTHESIA VENTILATOR 987 Fig. 2 Ventilator system diagram showing pressure display miniaturized pressure sensor (Fig. 2). Also, as an additional facility, real time AP waveform is displayed on a separate LCD graphic display unit. By this, it is possible to realize the rise and fall of air pressure over a breathing cycle. Several audio and visual alarms have been added to AP measurement system so as to make the system safe and failure proof to the extent possible. Main alarm conditions are AP high and AP low. System Design Choice of transducer needs to be capable of measuring required pressure (0-1.71 psig or 0-120 cm of H 2 O) and having response time of 10 milliseconds. Transducer has high accuracy, repeatability, longterm stability, high input impedance, linearity, negligible hysterisis, temperature compensation and small size. Further, for amplification of the output signal from transducer, precise low level, low time drift, accurate closed loop gain, high CMRR, high slew rate type amplifier system is designed and used. The range of values of air pressure (0-120 cm H 2 O) can be given either by a dial type pressure gauge of very large diameter or a water manometer with long heights. Both of these are very inconvenient as a part of a bigger system like a ventilator. Also, a low power consuming display device (graphics LCD module) with good contrast and backlight having full dot matrix graphics capability, is used to display graphics as well as characters. Microcontroller based Hardware Design Basic modules of the system contain a pressure sensor, instrumentation amplifier, low pass filter, buffer, bar graph driver, 10 segment LED bar graph, amplifier, 12-bit analog to digital converter and micro controller based signal processing circuitry (Fig. 3). Block diagram can be subdivided in following major parts: i) Micro controller & memories; ii) Transducer based circuitry; iii) Peripherals and interfaces; and iv) Measurement and signal processing. System design is based around 80C31 micro controller 6 connected through address bus, data bus and control bus to the 64Kbytes of EPROM 27C512 for monitor and control program, 8Kbytes of RAM and peripheral I/O device 8255 for interfacing, 12-bit

988 J SCI IND RES VOL 65 DECEMBER 2006 Fig. 3 Microcontroller based hardware design for airway pressure measurement with real time graphical display A/D converter, 30 characters 8 lines alphanumeric / graphic LCD display. AP is seen in the form of bar graph and as well as on LCD screen in graphical form. Transducer based Circuitry Transducer-based system consists of pressure sensor, instrumentation amplifier, low pass filter, buffer, bargraph driver and 10 segment LED bargraph (Fig. 4). Sensor consists of a sensing element (0.254 cm 2 silicon chip) with internal sensing diaphragm and 4 piezoresistors. Sensing resistors are connected as 4 active elements constituting a bridge. LM 3914 bargraph display driver is a monolithic IC that senses the analog voltage levels and drives 10 LEDs. Each successive LED of bargraph will glow for every 125 mv increase in input voltage. Data and control pins of the module are controlled through software to bring out display of values and real time bargraph display of AP. Fig. 4 Block diagram of transducer based system Signal Processing In AP display system of anaesthesia ventilator, the sensor selected has a pressure range of 0-5 psig. Pressure applied to the diaphragm to flex causes change in the resistance of the resistors. As AP changes, it is sensed by transducer, thus an amount of voltage proportional to AP is obtained. Output signal obtained from transducer is of very low magnitude (0-17 mv). Thus there is a need for some amplification where low noise, low thermal and noise

SENGUPTA et al: MICROCONTROLLER TO MEASURE AIRWAY PRESSURE OF ANAESTHESIA VENTILATOR 989 Fig. 5 Flow diagram for system software

990 J SCI IND RES VOL 65 DECEMBER 2006 drifts, high input impedance and accurate closed loop gain are required. Output of filter is amplified and then passed to a buffer known as voltage follower (Table 1). Use of unity gain buffer amplifier reduces loading effect in measurement system. After calculating analog value of AP, it is converted in digital form by using 12-bit A/D converter. Microcontroller processes digital data and displays on LCD in graphical form. Measurement and design requirements are as follows: pressure range, 0-120 cm of H 2 O; response time of sensor, 10 millisec; resolution, 7 cm of H 2 O; time period (X-axis), 1-15 sec; maximum frequency of signal, 1 Hz; output signal of pressure transducer, 0-17 mv; output signal from instrumentation amplifier, 0-1.25 V; and gain required in instrumentation amplifier, 73. System Software On the basis of flow diagram (Fig. 5), system software has been developed using C cross compiler 7,8 for Intel 8031 micro controller in modular form. System software is very much in optimised form and also maintains integrity of real time data and provides a very efficient way to represent in graphical form and also provides FREEZE button to store the graph for quick analysis of real time data, which is very useful for an anaesthetic analyst. The main program calls the sub functions and executes them accordingly. System software is stored in the system EPROM. Results and Discussion The system, fabricated and thoroughly tested for AP, has been found working satisfactorily giving accurate results. Its noise immunity has been observed to be excellent as it works in harsh environments. LCD module helps to monitor AP both alphanumeric as well as graphics without going through complex sequential operations, thus making the system ideal for ventilators. Total system is operated by air pressure and calibration is done by using Fluke Pressure Meter. For calibration of AP measurement system of anaesthesia ventilator, air is supplied with Fluke Pressure pump and this pressure goes simultaneously to the input to pressure sensor and to standard Fluke pressure calibrator. Output of AP measurement system is then calibrated with pressure calibrator. Graph between AP measurement system output and the pressure calibrator output is a straight line and thus shows satisfactory results (Fig. 6). It is Table 1 Instrumentation amplifier input and buffer amplifier output S No. Instrumentation amplifier input Buffer amplifier output S No. Instrumentation amplifier input Buffer amplifier output mv V mv V 1 1.4 0.13 16 10.1 0.73 2 2.1 0.18 17 10.3 0.75 3 2.9 0.23 18 11.2 0.81 4 3.5 0.28 19 11.7 0.84 5 4.4 0.35 20 12.1 0.87 6 4.6 0.36 21 12.5 0.90 7 5.5 0.42 22 13.0 0.93 8 5.9 0.45 23 13.3 0.96 9 6.5 0.48 24 13.8 0.99 10 7.0 0.52 25 14.2 1.01 11 7.5 0.55 26 14.6 1.04 12 8.1 0.58 27 15.0 1.11 13 8.5 0.63 28 15.5 1.14 14 9.1 0.66 29 16.3 1.22 15 9.6 0.70 30 17.1 1.27 Fig. 6 Graph between airway pressure measurement system output and pressure calibrator output reading observed that variation in the voltage required to glow an LED is less than 5% of the true value, which is quite acceptable. The resolution of bar graph LED is 7 cm of H 2 O. Conclusions AP measurement is useful for anaesthetist to manually control AP to reach maximum level. In the present work, provision is made to display only air pressure but display capacity can be increased up to 8 ventilator parameters by altering software a little and incorporating a small additional hardware.

SENGUPTA et al: MICROCONTROLLER TO MEASURE AIRWAY PRESSURE OF ANAESTHESIA VENTILATOR 991 Acknowledgements Authors thank Department of Science and Technology, New Delhi for sponsorship of this work and Director, CSIO, Chandigarh. References 1 Khandpur R S, Handbook of Biomedical Instrumentation, 2 nd edn (Tata McGraw Hill Publishing Co. Ltd., New Delhi) 2003, 841-852. 2 Mushin W W, Automatic Ventilation of the Lungs, 2 nd edn (Blackwell Scientific Publications, Oxford & Edinburgh) 1969, 1-18. 3 Carr J J & Brown J M, Introduction to Biomedical Equipment Technology, 3 rd edn, (Prentice Hall, New Jersey) 1998, 243-253. 4 Webster J G, Encyclopaedia of Medical devices and instrumentation (Wiley-Interscience Publication, New York) 1988, 2847-2857. 5 Cromwell L, Arditti M, Weibell F J, Pfeiffer E A, Steele B & Labok J, Medical Instrumentation for Healthcare (Prentice Hall, New Jersey) 1976, 274-282. 6 Thomas W S, C and the 8051 Hardware, Modular programming and multitasking, 2 nd edn (Prentice Hall, New Jersey) 1995, 97-113. 7 8051 C Compiler Programming Guide for Intel s 8051 Microcontroller Family, 8 th edn (IAR Systems, Uppsala) 1996, 9-45. 8 WINIDEA TM Version 9.0, Integrated Development Environment (isystem GmbH, Neuhimmelreich) 2000, 5-17.