Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume

Similar documents
Pipetting Small Volumes

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Gel Loading Practice MiniLab

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

Real World Metric System

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

MOLEBIO LAB #1: Microquantity Measurement

Pipetting and Determining Protein Concentration

Inquiry Module 1: Checking the calibration of a micropipette

Experiment 1 Use of Micropipettes: measuring volume accurately

Adjust the Volume Display

CHM250 Calibration and Measurement Lab. Balance Calibration

Chapter 2. Pipetting techniques. Guide to Pipetting

Pipettor. User Manual

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

User s Manual. Autoclavable & UV r e s i s t a n t. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

User s Manual. Autoclavable & UV resistant. Multi-channel digital micro pipette. In Vitro Medical Diagnostic Devices (98/79/EC)

RayPette Plus Autoclavable Pipette. User Manual

MULTICHANNEL. Labnet International, Inc.

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

Improving Pipetting Techniques

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

The pipettes cover a volume range from 0.1µl to 10ml.

Varipette Operating manual

PIPETMAN F ENGLISH. Other languages available on our website:

How to Use: 1. Hold the pipette like you are going to shake hands with it.

Other languages available on our website:

CALIBRATION AND ADJUSTMENT

Technical Procedure for General Laboratory Equipment

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

MICROMAN ENGLISH. Other languages can be consulted and downloaded on our website,

User s Manual. Autoclavable & UV resistant Nichipet EX II. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

Nichipet EX Plus II. User s Manual. Autoclavable & UV resistant. Digital micro pipette for liquid handling

Transferpette. Lightweight precise rugged reliable!

Transferpette. Lightweight precise rugged reliable!

SOP-111 Pipetting Small Volumes

Operating Instructions. Rainin Classic Manual Pipette

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Determination of Sodium using Atomic Emission

APPENDIX A CORRECT USE OF DIGITAL PIPETTES

How to Measure R7.1. Reference. I. Linear dimensions

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

BASIC LABORATORY TECHNIQUES (Revised )

Transferpette. Single and Multichannel Pipettes

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

A N F G J I C D D L A B C D E F

User's Manual. Autoclavable Nichipet EX Plus Digital micro pipette for liquid handling CERTIFIED ISO9001

BASIC LABORATORY TECHNIQUES (Revised )

Before you start sampling, be sure to read

VOLUMETRIC TECHNIQUES

Operating manual for the pressure transducer unit CCS 1000

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

DECIMALS. Decimal Place Value and Rounding Decimals

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Transferpette -8/-12

Recording Significant Figures

NCERT solution Decimals-2

Laboratory Safety Training. Environmental Health and Safety

Rounding Up or Down? Name: Date:

Lab Equipment ANALYTICAL BALANCE

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

VWR HIGH PERFORMANCE PIPETTOR

The pipette of Choice for Problem Liquids!

A N B P F G J I C D D L A B C D E F

Userts Manual. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC) Annex III self-declared ISO 8655 STANDARD

Ensuring pipette & weighing device performance: Verification & Calibration

The use of the analytical balance, and the buret.

Corning Step-R. Repeating Pipettor. Instruction Manual

Module 1- Mid Module Review Part 1- Classwork

CHM Introductory Laboratory Experiment (r17sd) 1/13

User Manual for the Mars Calibration Bench

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

Purpose. Introduction

Measuring Mass and Volume

Ultra-Low Copper Concentrations Determined by Rotating Disk Electrode Stripping Voltammetry

Ergonomic Pipetting for Problem Liquids!

Standard Operating Procedure. Air Displacement Pipette Calibration

[USING THE VITROBOT April 9, Using the Vitrobot

Version BA D

Loopamp Bovine Embryo Sexing Kit Instruction Manual

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

OXYTEMP Instructions 1

Place Value

Chemistry and Measurements (Lecture PPTs)

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Any laboratory is equipped with specific tools, equipment,

WITOPET PREMIUM PIPETTE USER S GUIDE

Intermolecular Forces

Pipette Operation Guide

VITLAB micropipette. Standard Operating Procedure (SOP)

NEO. User s Guide. Do not lubricate this pipette. The use of lubricant cancels the warranty of this pipette. NOTICE

Transcription:

Using a micropipette to transfer exact liquid measurements in scientific experiments By: Nicole Hume Introduction A micropipette is a scientific tool used to transfer small, exact volumes of liquid between 1-1000ul. Proper use of a micropipette is essential to successfully carry out a scientific experiment that requires exact volumes of a reagent during the sample preparation. Micropipettes are most often used in general or organic chemistry, biochemistry, and molecular biology experiments. Each micropipette has a control button, eject button, volume knob, number windows, and a tip. v The control button is used when drawing up and expelling liquid. Each control button has two stopping points. o The first stopping point is used to draw up the liquid. o The second stopping point is used when the liquid is expelled from the plastic tip attached to the tip of the micropipette. v The eject button is used when discarding the plastic tip. v The volume knob is used to adjust the volume for each micropipette. v The number windows display the set volume based on the adjustments made on the volume knob v The tip of the micropipette is the part of the micropipette that will attach to the plastic tips v The plastic tips hold the volume of the reagent being used in the experiment Anatomy of a micropipette However, each micropipette can be used only in a specified range of liquids. Once the correct micropipette is chosen, the volume within the range of the micropipette must be specified. Next, the proper tip must be obtained and inserted onto the micropipette. The volume of liquid is drawn up and then expelled using the control button using the two stopping points on the button. Then, the tip is ejected. The whole process, when mastered, can be carried out in less than one minute. CAUTION When using a micropipette, it is critical to follow these safety measures: v Never attempt to pipet a volume outside of the specified volume range for each micropipette v Always use gloves when handling biological fluids, strong acids, strong bases, or toxic materials v Always dispose of plastic pipette tips in a biohazard bag when handling biological fluids v Always work in a fume hood when using volatile (readily evaporated) liquids

Choosing the Correct Micropipette Choosing the correct micropipette is essential in preventing calibration errors and transferring an accurate volume of liquid. Determine the volume of liquid you need to transfer. This will be stated in the steps to complete the experiment. The volume must be between.2-1000ul. For volumes greater than 1000ul, a serological pipette would be more useful. Look at the top of the control button to see the pipette size v If you need to pipet between 200-1000ul, you should use the blue P1000 pipette v If you need to pipet between 20-200ul, you should use the yellow P200 pipette v If you need to pipet between 2-20ul, you should use the yellow P20 pipette v If you need to pipet between.2-2ul, you should use the gray P2 pipette If the volume you wish to pipet falls within the range of more than one pipette (such as 2ul, 20ul, or 200ul), the pipettes manufactured for smaller volumes should be used. Accuracy is decreased when trying to pipet smaller volumes on larger volume pipettes. Micropipette control buttons

Adjusting to your Needed Volume Adjusting the volume of the pipette is the trickiest part of the micropipetting process. A simple mistake could lead to a large error in the preparation of your experiment. You will need to refer to the volume knob and number windows to accurately input the volume you are going to pipet. To adjust the volume, first refer to the number window. If the volume is lower than the volume you need to pipet, turn the volume knob clockwise to increase the volume until the desired volume is read in the number window. To decrease the volume, the volume knob should be turned counterclockwise. The following table describes how to read the number window for each pipette: P1000 Pipettes If you are using a P1000 pipette, the top, red number is the thousands digit. The middle number is the hundreds digit, and the lower number is the tens digit. In the example on the left, 350ul of liquid would be pipetted. The numbers should range from 020 and 100 for this pipette. P200 Pipettes If you are using the P200 pipette, the top number is the hundreds digit, the middle number is the tens digit, and the bottom number is the ones digit. In the example on the left, 95ul of liquid would be pipetted. The set numbers should range between 020 and 200 for this pipette. P20 Pipettes If you are using the P20 pipette, the top number represents the tens digit, the middle number represents the ones digit, and the bottom, red number represents the tenths digit. There is a decimal between the middle and bottom number. In the example on the left, 2.5ul of liquid would be pipetted. The numbers should range from 020 to 200 with this pipette. P2 Pipettes If you are using the P2 pipette, the top number represents the ones digit, the second, red number represents the tenths digit, and the bottom, red number represents the hundredths digit. In the example on the left,.5ul of liquid would be pipetted. The numbers should range from 020 to 200 with this pipette. How to read the number window on a P1000, P200, P20, and P2 micropipette

Pipetting the Liquid 1. Obtain the micropipette tips that match the micropipette you will need to use. v P1000 pipettes use the large, blue tips v P200 and P20 pipettes use the yellow tips v P2 pipettes use the clear tips 2. Insert the tip of the micropipette firmly into a plastic tip. The plastic tip should be attached securely to the pipette. 3. Press the control button down to the first stopping point. Do not push the control button all the way down. 4. Insert the plastic pipette tip into the liquid you wish to pipet. Once the tip is suspended in the liquid, slowly release the control button to its starting point. Then, remove the plastic tip from the liquid. 1 11 2 3 Visual depiction of steps 3-4 detailing how to draw up liquid before pipetting it into a new container. (1) The control button is pressed down to the first stopping point, (2) the tip is inserted into the liquid, and (3) the control button is released while the tip is still in the liquid. 5. Insert the plastic pipette tip into the container with the liquid you are transferring the pipetted liquid to. Push the control button down all the way. The button should pass the first stopping point and reach the second stopping point where you cannot push the control button down any further. The liquid should now be pipetted into the new liquid container. Do not release the control button. 6. Remove the plastic tip from the container. Once the tip is removed from the container, release the control button to its starting point. A visual depiction of this process can be seen on the following page. 7. Place the micropipette, which is still connected to the plastic pipette tip, over a biohazard bag. Press down on the eject button to release the plastic tip into the bag.

1 2 Visual depiction of steps 5-6 detailing how to transfer liquid to a new container. (1) The control button is pressed down to the second stopping point and the liquid is transferred to the new container. (2) The control button is then released to the starting point. Tips v Avoid touching plastic pipette tips to hands, counter, or other objects that could cause contamination v Dispose of plastic pipette tips after each individual use to avoid cross-contamination v Use smooth, controlled movements when using the control button v Check the numbers on the number window periodically to confirm volume