Operating Instructions. Flowfit W CCA250. Flow assembly for chlorine sensors. BA062C/07/en/

Similar documents
Operating Instructions Flowfit CUA252

Operating Instructions. Cleanfit W CPA450. Retractable Assembly for 12 mm Sensors for DO/pH/ORP Measurement BA183C/07/EN/13.

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Retractable Process Assembly

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Indumax CLS54D. Operating Instructions

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Mounting and Operating Instructions EB 3017 EN. Self-operated Regulators

Technical Information Conducal CLY421

Operating Instructions CCS120. Sensor for total chlorine. BA388C/07/en/

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

Type BBS-03, BBS-05, BBS-06, BBS-25

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

Siphon vessel / Priming aid

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Oxymax H COS21D. Technical Information

GI Series Diaphragm Pressure Gauge Guard

Technical Information Cleanfit CUA451

Oxymax COS61/COS61D. Technical Information

Operating Instructions Turbimax CUS52D

book : t95575.fm Seite 1 Mittwoch, September 20, :10 AM. Avensys. Exposed Single Lever Mixer

OUS31. Technical Information

OPERATING INSTRUCTIONS MANUAL FOR QACV PNEUMATIC DOSING PUMP

Pressure relief valve

Pressure relief valve

Flowmeter. Original operating manual DFM

Preliminary filter Precision measurement chamber Flow meter Instruction manual

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

Operating Manual VSA100A. tina29e1 ( )

Operating Instructions. Oxymax W COS41. Dissolved oxygen sensor. BA284C/07/en/

Pneumatic proportional controller Types M Types FM Function tested for use of the float in Ex-zone 0

Pressure relief valve DHV 716 set range: 0,5-10,0 bar

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

Modular valve block. Operating manual Series MVB 100/200. Version BA EN Print-No TR MA DE Rev001

Type /2-Way Globe Valve 3/2-Wege-Geradsitzventil Vanne à siège droit 3/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Type Type Differential Pressure Regulators (opening)

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Mounting and Operating Instructions EB 2555 EN. Self-operated Pressure Regulators. Pressure Reducing Valves Type 50 ES Type 50 EM.

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

Type 2000 INOX. Quickstart. English Deutsch Français. 2/2-way angle seat valve 2/2-Wege Schrägsitzventil Vanne à siège incliné 2/2 voies

Sensor for Measuring Dissolved Oxygen

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

Mounting and Operating Instructions EB 8546 EN. Supply Pressure Regulator Type Fig. 1 Supply pressure regulators

DIAPHRAGM PRESSURE CONTROL VALVE FDV 30 / 31, FDV 1.30 / 1.30 FDV 300 / 301, FDV / 1.301

Mounting and Operating Instructions EB EN. Type Supply Pressure Regulator. with increased air capacity

Premium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw.

Operating instructions Diaphragm pump MP-520. Article No Original operating instructions. T-Dok-397-EN-Rev.0web

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Pressure relief valve DHV 712 DN 65-80: 0,5-10 bar, DN : 0,3-4 bar, DN 100: 0,5-6 bar

Pneumatic switch module 3/2-way valve (On/Off) Types P Types FP Function tested for use of the float in Ex-zone 0

Declaration of Conformity as per Directive 97/23/EC

Mounting and operating instructions EB 2530 EN. Self-operated Pressure Regulator. Pressure Reducing Valve Type M 44-2

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

Instruction Manual Contact Pressure Vacuum Gauge

Installation and operating instructions KEMPER Backflow Preventer BA Figure 361

Pressure Reducing Valve DMV 750 set range: bar

Flexi-pro Aerosol Delivery System (FADS) Manual

Mounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

Mounting and Operating Instructions EB 8546 EN. Type 4708 Supply Pressure Regulators. Translation of original instructions. Type Type

Inverted Bucket Steam Trap IB 16A-7

Float Operated Level Controllers

553 Series.

SRV66 Sanitary Pressure Reducing Valve

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

THE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691

SPECIFICATIONS ATTENTION

Pressure Relief Valve DHV 718

Installation, commissioning and servicing instructions

Declaration of Conformity as per Directive 97/23/EC

Type C-CUT. Quickstart. English Deutsch Français

Armatures for analytical probes

GM Series Dual-Block Multi-Function Gas Control Valves

Declaration of Conformity as per Directive 97/23/EC

Instruction Manual. Alfa Laval SB Pressure Exhaust Valve ESE02965-EN Original manual

Pressure Reducing Valve DMV 750

GILMONT ACCUCAL FLOWMETERS

INSTRUCTION MANUAL Pressure Relief Device LPT

LRS(H)4 Pressure-Reducing Regulator User Manual

Type 2080 FALTENBALGVENTIL BELLOWS VALVE VANNE A SOUFFLET. Operating Instructions. Bedienungsanleitung Manuel d utilisation

Instruction Manual. Standard Surge Eliminator. Standard X. Flushable X - X

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

Installation and Operating Instructions Bypass Flow Meter DST

TECHNICAL DATA. Trimpac 251a. Spetember 16, 2013

Type 0404 / /2-way solenoid valve 2/2-Wege-Magnetventil Électrovanne 2/2 voies Operating Instructions Bedienungsanleitung Manuel d utilisation

THE HF-210 SERIES. Operating and Service Manual. Series includes all variants of HF-210/211. Issue A October 2014 *Section 6.

Pressure maintenance valve SDV-P / SDV-P-E

Modular Analytical Panel System with Sample preparation for Steam and Power Applications

PRESSURISED PAINT CONTAINER

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Series 42 Self-operated Regulators Differential Pressure Regulators with Type 2424/Type 2428 Actuator (closing) Type Type 42-28

TECHNICAL DATA. TRIMPAC Model B-5 & B-5B

Pressure relief valve DHV 725 set range: 0,2-10,0 bar

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

Neotecha SAPRO - Aseptic sampling valve SV Installation, operation and maintenance instructions

Transcription:

Operating Instructions Flowfit W CCA250 Flow assembly for chlorine sensors BA062C/07/en/07.09 71099066

Brief overview Fig. 1: CCA250 with sensors 1 ph sensor (accessories, not in scope of delivery) 2 PML connection 3 Vent screw 4 Chlorine sensor (accessories, not in scope of delivery) 5 Outlet (adapter depending on version) 6 Needle valve for flow <120 l/h (32 gal/h) 7 Locking cap with calibration vessel 8 Inlet (adapter depending on version) 9 Inductive limit switch for flow measurement (optional) 10 Installation position for a second ph/orp sensor a0007322

Table of contents 1 Safety instructions................ 4 1.1 Designated use............................ 4 1.2 Installation, commissioning and operation........ 4 1.3 Operational safety.......................... 4 1.4 Return................................... 4 1.5 Notes on safety icons and symbols.............. 5 2 Identification.................... 6 2.1 Nameplate............................... 6 2.2 Product structure.......................... 6 2.3 Scope of delivery........................... 6 3 Installation...................... 7 3.1 Incoming acceptance, transport, storage.......... 7 3.2 Installation conditions....................... 7 3.3 Installation instructions...................... 8 3.4 Post-installation check...................... 10 4 Commissioning.................. 10 5 Maintenance.................... 11 5.1 Cleaning the assembly...................... 11 5.2 Cleaning the sensor........................ 11 5.3 Cleaning agents........................... 11 6 Accessories..................... 13 6.1 Accessories kits........................... 13 6.2 Sensors................................. 13 7 Trouble-shooting................ 15 7.1 Replacing damaged parts.................... 15 7.2 Return.................................. 15 7.3 Disposal................................ 15 8 Technical data.................. 16 8.1 Environment............................. 16 8.2 Process................................. 16 8.3 Mechanical construction.................... 17 Index......................... 18 3

Safety instructions 1 Safety instructions 1.1 Designated use The flow assembly has been designed for the installation of the membrane covered chlorine sensors CCS120, CCS140, CCS141, CCS240, CCS241 and CCS142D. Additionally, there are two more installation positions for the installation of sensors with Pg 13.5 and installation length 120 mm (4.72 inch), e.g. ph resp. ORP sensors. Its mechanical design permits its use in pressurised systems (see "Technical data"). Any other use than the one described here compromises the safety of persons and the entire measuring system and is not permitted. The manufacturer is not liable for damage caused by improper or non-designated use. 1.2 Installation, commissioning and operation Please note the following items: Installation, commissioning, operation and maintenance of the measuring system must only be carried out by trained technical personnel. Trained personnel must be authorized for the specified activities by the system operator. Electrical connection must only be carried out by a certified electrician. Technical personnel must have read and understood these Operating Instructions and must adhere to them. Before commissioning the entire measuring point, check all the connections. Ensure that electrical cables and hose connections are not damaged. Do not operate damaged products and secure them against unintentional commissioning. Mark the damaged product as being defective. Measuring point faults may only be rectified by authorized and specially trained personnel. If faults can not be rectified, the products must be taken out of service and secured against unintentional commissioning. Repairs not described in these Operating Instructions may only be carried out at the manufacturer s or by the service organization. 1.3 Operational safety The assembly has been designed and tested in accordance with the latest industry standards and left the factory in perfect functioning order. Relevant regulations and standards have been met. As the user, you are responsible for complying with the following safety conditions: Installation instructions Local prevailing standards and regulations. 1.4 Return If the assembly has to be repaired, please return it cleaned to the appropriate sales center. Please use the original packaging, if possible. Please enclose the completed "Declaration of contamination" (copy the second to last page of these Operating Instructions) with the packaging and the transportation documents. No repair without completed "Declaration of contamination"! 4

Safety instructions 1.5 Notes on safety icons and symbols # " Warning! This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if ignored. Caution! This symbol alerts you to possible faults which could arise from incorrect operation. They could cause damage to the instrument if ignored.! Note! This symbol indicates important items of information. 5

Identification 2 Identification 2.1 Nameplate You can read the following information from the nameplate on the assembly: Product name Order code Serial number or week code Permissible process pressure and maximum process temperature You can find possible assembly versions and the resulting order codes in the product structure. 2.2 Product structure Process connection, adapter A G1/2, none B G1/2, 2x NV 1/2" C G1/2, 2x D 6/12 D G1/2, D 6/12 + D 16 M NPT 1/2", none Q NPT 1/4", none Proximity switch 0 none 1 with inductive proximity switch CCA250- complete order code 2.3 Scope of delivery The scope of delivery comprises: Flowfit assembly (ordered version) a PML adapter Operating Instructions (English) If you have any questions, please contact your supplier or your local sales center. 6

Installation 3 Installation 3.1 Incoming acceptance, transport, storage Make sure the packaging is undamaged! Inform the supplier about any damage to the packaging. Keep the damaged packaging until the matter has been settled. Make sure the contents are undamaged! Inform the supplier about damage to the contents. Keep the damaged products until the matter has been settled. Check that the order is complete and agrees with your shipping documents. The packaging material used to store or to transport the product must provide shock protection and humidity protection. The original packaging offers the best protection. Also, keep to the approved ambient conditions (see "Technical data"). If you have any questions, please contact your supplier or your local sales center. 3.2 Installation conditions 3.2.1 Dimensions Fig. 2: Dimensions a depending on version: G1/2, NPT 1/2" or NPT 1/4" a0007328 7

Installation 3.2.2 Notes on installation Parts description used at... two shut-off valves one shut-off valve aperture in the main conduit particle filter (500 μm or finer) pressure-relief valve bypass version version with open outlet bypass version if the process water contains large dirt particles if the process water pressure is above the maximum value (see chapter " technical data") 3.3 Installation instructions 3.3.1 Measuring system A complete measuring system comprises: Flowfit CCA250 Chlorine sensor, e.g. CCS142D Measuring cable, e.g. CYK10 Transmitter, e.g. Liquiline M Optional: up to two ph sensors, e.g. Orbisint CPS11D Junction box for cable extension, e.g. RM junction box Fig. 3: Measuring system 1 Transmitter Liquiline M CM44 with weather protection cover 2 Supply line of the transmitter 3 Chlorine sensor CCS142D 4 Flowfit CCA250 5 Assembly inlet (outlet at backside, not to be seen in figure) 6 ph-sensor 7 Measuring cable CYK10 a0007341 8

Installation 3.3.2 Installing the assembly into the process To get a flow through the bypass, pressure p1 has to be higher than pressure p2. Therefore, you have to install an aperture in the main conduit ( å 4, pos. 5). Fig. 4: Installation example with bypass and aperture in the main conduit 1 Stop valve (to be provided by customer) 2 Dirt trap (filter), d = 500 μm (to be provided by customer). Installation of dirt trap is mandatory! 3 Pressure reducer (with p > 4 bar (58 psi)) 4 Stop valve (to be provided by customer) 5 Aperture in the main conduit a0007337 In case of an open outlet installation, no pressure increasing procedure is needed ( å 5). 1 2 3 4 Fig. 5: Installation example with open outlet Legend å 4 a0007338! Note! The flow assembly must be installed upright. Connect the process water with commercial fittings. You can use the normal sealing (e.g. teflon tape) or an O-ring (e.g. EPDM). Installation in the by-pass is preferable to installation in the process pipe as the by-pass pipe can be blocked off without process interruption (you have to install shut-off valves upstream and downstream from the flow assembly). This permits maintenance of the sensors without interrupting the process. A stop valve and a dirt trap (e.g.,a strainer) with a mesh size of 500 μm must be located upstream from the flow assembly. Pressure reducers usually contain a dirt trap, i.e. an additional dirt trap is not required when a pressure reducer is installed. 9

Commissioning " Caution! For the version with an open outlet, you have to install a shut-off valve upstream. Install a DN 5-8 sampling cock downstream from the flow assembly to permit reference measurements according to the DPD method. The medium pressure must not exceed the maximum permissible pressure of the flow assembly or of the sensors. If the medium pressure exceeds 4 bar (58 psi), you have to install a pressure-relief valve.! Note! 3.3.3 Sensor installation 1. Installation of proximity switch: Screw in all the way and lock with hex nut. 2. Loosen and remove the clamping screw. 3. Insert the chlorine sensor in the mounting position and firmly tighten with the clamping screw. 4. Remove the two dummy plugs, place the ph/orp sensors in the openings with the Pg13.5 threads and screw in. 5. Connect the sensors via the corresponding measuring cable to the transmitter. Vent the assembly by unscrewing and screwing the vent screw at the time of the first commissioning. 3.4 Post-installation check After installation, check that all connections are firmly in position and leak-tight. Check all hoses for damage. # Warning! 4 Commissioning Before the first commissioning, make sure of the following items: all seals are correctly seated (on the assembly and process connection) the sensor is correctly installed and connected Danger of squirting medium. Before applying the process pressure to the assembly, make sure the connections are correctly fitted. 10

Maintenance # Warning! 5 Maintenance Risk of injury! Before starting maintenance work on the assembly, make sure that the process line is depressurised, empty and rinsed.! Note! 5.1 Cleaning the assembly To ensure a reliable measurement, the assembly and the sensor must be cleaned at regular intervals. The frequency and intensity of the cleaning operation depend on the process medium. All parts in contact with the medium, e.g. the sensor and the sensor holder, must be cleaned at regular intervals. Remove the sensor 1). Remove light dirt using suitable cleaning agents (see chapter "Cleaning agents"). Remove severe fouling with a soft brush and a suitable cleaning agent. Remove persistant fouling by soaking in a liquid cleaner and if neccessary by cleaning with a soft brush. A typical cleaning interval for e.g. drinking water is at least half a year.! Note! 5.2 Cleaning the sensor You have to clean the sensor: before every calibration regularly during operation before being returned for repair depending on the local conditions, at least twice a year Clean the sensor outside only. Do not open it! Do not use any abrasive cleaning agents. This can lead to irreparable damage of the sensor. After cleaning the sensor, rinse the rinse chamber of the assembly with copious amounts of water. Otherwise, remaining residues of cleaning agent can corrupt measurement. If required, re-calibrate after cleaning. 5.3 Cleaning agents The selection of the cleaning agent is dependent on the degree and type of contamination. The most common contaminations and the suitable cleaning agents are listed in the following table. Type of contamination Greases and oils Calciferous deposits, metal hydroxide deposits, lyophobic biological deposits Sulphide deposits Protein deposits Fibres, suspended substances Light biological deposits Cleaning agent Substances containing tensides (alkaline) or water-soluble organic solvents (e.g. Ethanol) approx. 3% hydrochloric acid Mixture of 3% hydrochloric acid and thiocarbamide (commercially available) Mixture of 3% hydrochloric acid and pepsin (commercially available) Water under pressure, poss. with surface-active agents Water under pressure 1) in reverse sequence of operations to the installation procedure 11

Maintenance " Caution! Do not use organic solvents containing halogen or acetone. These solvents could destroy plastic components on the assembly or the sensor and it is also partly suspected that they cause cancer (e.g. Chloroform). 12

Accessories! Note! 6 Accessories In the following sections, you find the accessories available at the time of issue of this documentation. For information on accessories that are not listed here, please contact your local service. 6.1 Accessories kits For G1/2 process connection versions only! NV 1/2 2 PVC nipples for connection to PVC pipework for pipes with OD 16 mm (0.63 inch) order no. 50003228 SV 1/2 2 PVC nipples different adaptersfor hose connection with Ø 6/12 mm (0.24/0.47 inch) inlet and Ø 16 mm (0.63 inch) outlet order no. 50003232 identical adapters for hose connection with Ø 6/12 mm (0.24/0.47 inch) inlet and outlet order no. 50003230 6.2 Sensors CCS120 Membrane covered amperometric sensor for total chlorine Measuring range 0.1 to 10 mg/l Ordering acc. to product structure, see Technical Information (TI388C/07/en) CCS140 Membrane covered amperometric sensor for free chlorine Measuring range 0.05 to 20 mg/l Ordering acc. to product structure, see Technical Information (TI058C/07/en) CCS141 Membrane covered amperometric trace sensor for free chlorine Measuring range 0.01 to 5 mg/l Ordering acc. to product structure, see Technical Information (TI058C/07/en) CCS142D Membrane covered amperometric sensor for free chlorine Memosens technology Measuring range 0.01 to 20 mg/l Ordering acc. to product structure, see Technical Information (TI419C/07/en) CCS240 Membrane covered amperometric sensor for chlorine dioxide Measuring range 0.05 to 20 mg/l Ordering acc. to product structure, see Technical Information (TI114C/07/en) CCS241 Membrane covered amperometric trace sensor for chlorine dioxide Measuring range 0.01 to 5 mg/l Ordering acc. to product structure, see Technical Information (TI114C/07/en) Orbisint CPS11/11D ph electrode for process applications with dirt-repellent PTFE diaphragm Optional Memosens technology (CPS11D) Ordering acc. to product structure, see Technical Information (TI028C/07/en) 13

Accessories Ceragel CPS71/CPS71D ph electrode with double junction reference system and integrated bridge electrolyte Optional with Memosens technology (CPS71D) Ordering acc. to product structure, see Technical Information (TI245C/07/en) Ceragel CPS72/CPS72D Redox sensor with double junction reference system and integrated bridge electrolyte Optional with Memosens technology (CPS72D) Ordering acc. to product structure, see Technical Information (TI374C/07/de) 14

Trouble-shooting 7 Trouble-shooting # Warning! 7.1 Replacing damaged parts Damage to the assembly which affects the pressure safety must only be repaired by authorized technical personnel. After every repair and maintenance activity, suitable measures must be taken to test whether the assembly shows any signs of leaking. The assembly must then correspond to the specifications stated in the technical data. Replace all other damaged components immediately. To order accessories and spare parts, please use the "Accessories" and "Spare parts" chapters or contact your local sales center. 7.2 Return If the assembly has to be repaired, please return it cleaned to the appropriate sales center. Please use the original packaging, if possible. Please enclose the completed "Declaration of contamination" (copy the second to last page of these Operating Instructions) with the packaging and the transportation documents. No repair without completed "Declaration of contamination"! 7.3 Disposal Please dispose of the device in accordance with the local regulations. 15

Technical data 8 Technical data 8.1 Environment Ambient temperature range 0 to 50 C (32 to 120 F) Storage temperature 0 to 50 C (32 to 120 F) 8.2 Process Process temperature 0 to 45 C (32 to 110 F), non-freezing Process pressure max. medium pressure: 4 bar (58 psi) at 40 C (104 F) Temperature-pressurediagram Fig. 6: Temperature-Pressure diagram a0007413 Flow optimum 30 l/h (7.9 gal/h) 30 to 120 l/h (7.9 to 31.7 gal/h), adjustable 16

Technical data 8.3 Mechanical construction Design, dimensions see chapter "Installation" Weight 0.5 to 0.8 kg (1.1 to 1.8 lbs), depending on process connection Material In contact with medium: Assembly body: PMMA Mounting parts: PVC, stainless steel 1.4571 (AISI 316 Ti), EPDM Process connection G1/2, NPT 1/2" or NPT 1/4" 17

Index A Accessories.................................... 13 Sensors.................................... 13 Assembly Cleaning................................... 11 C Checking Installation................................. 10 Cleaning Agents.................................... 11 Assembly.................................. 11 Sensor.................................... 11 Cleaning interval............................... 11 Commissioning................................. 4 Scope of delivery................................. 6 Sensor Cleaning................................... 11 Installation................................. 10 Storage........................................ 7 Symbols....................................... 5 T Transport...................................... 7 U Use........................................... 4 D Designated use.................................. 4 Dimensions.................................... 7 Disposal...................................... 15 E Environment.................................. 16 I Icons......................................... 5 Incoming acceptance............................. 7 Installation................................. 4, 7 8 Sensor.................................... 10 M Maintenance.................................. 11 Maintenance interval............................ 11 Measuring system................................ 8 Mechanical construction.......................... 17 N Nameplate..................................... 6 O Operation...................................... 4 Operational safety................................ 4 Ordering information............................. 6 P Parts Replacing.................................. 15 Process....................................... 16 Product structure................................ 6 R Replacing Parts...................................... 15 Return..................................... 4, 15 S Safety icons.................................... 5 18

Declaration of Hazardous Material and De-Contamination Erklärung zur Kontamination und Reinigung RA No. Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility. Bitte geben Sie die von E+H mitgeteilte Rücklieferungsnummer (RA#) auf allen Lieferpapieren an und vermerken Sie diese auch außen auf der Verpackung. Nichtbeachtung dieser Anweisung führt zur Ablehnung ihrer Lieferung. Because of legal regulations and for the safety of our employees and operating equipment, we need the "Declaration of Hazardous Material and De-Contamination", with your signature, before your order can be handled. Please make absolutely sure to attach it to the outside of the packaging. Aufgrund der gesetzlichen Vorschriften und zum Schutz unserer Mitarbeiter und Betriebseinrichtungen, benötigen wir die unterschriebene "Erklärung zur Kontamination und Reinigung", bevor Ihr Auftrag bearbeitet werden kann. Bringen Sie diese unbedingt außen an der Verpackung an. Type of instrument / sensor Geräte-/Sensortyp Serial number Seriennummer Used as SIL device in a Safety Instrumented System / Einsatz als SIL Gerät in Schutzeinrichtungen Process data/prozessdaten Temperature / Temperatur [ F] [ C] Pressure / Druck [psi] [ Pa ] 2 Conductivity / Leitfähigkeit [µs/cm] Viscosity / Viskosität [cp] [mm /s] Medium and warnings Warnhinweise zum Medium harmful/ Medium /concentration Identification flammable toxic corrosive irritant other * harmless Medium /Konzentration CAS No. entzündlich giftig ätzend gesundheitsschädlich/ sonstiges* unbedenklich reizend Process medium Medium im Prozess Medium for process cleaning Medium zur Prozessreinigung Returned part cleaned with Medium zur Endreinigung * explosive; oxidising; dangerous for the environment; biological risk; radioactive * explosiv; brandfördernd; umweltgefährlich; biogefährlich; radioaktiv Please tick should one of the above be applicable, include safety data sheet and, if necessary, special handling instructions. Zutreffendes ankreuzen; trifft einer der Warnhinweise zu, Sicherheitsdatenblatt und ggf. spezielle Handhabungsvorschriften beilegen. Description of failure / Fehlerbeschreibung Company data /Angaben zum Absender Company / Firma Address / Adresse Phone number of contact person / Telefon-Nr. Ansprechpartner: Fax / E-Mail Your order No. / Ihre Auftragsnr. P/SF/Konta XIV We hereby certify that this declaration is filled out truthfully and completely to the best of our knowledge.we further certify that the returned parts have been carefully cleaned. To the best of our knowledge they are free of any residues in dangerous quantities. Wir bestätigen, die vorliegende Erklärung nach unserem besten Wissen wahrheitsgetreu und vollständig ausgefüllt zu haben. Wir bestätigen weiter, dass die zurückgesandten Teile sorgfältig gereinigt wurden und nach unserem besten Wissen frei von Rückständen in gefahrbringender Menge sind. (place, date / Ort, Datum) Name, dept./abt. (please print / bitte Druckschrift) Signature / Unterschrift

www.endress.com/worldwide BA062C/07/en/07.09 FM+SGML 6.0 71099066