Instructions for Use for CAIMAN Instrument (12 mm x 44 cm) Item #: US731SU

Similar documents
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.

Instructions for Use Reprocessed LigaSure Maryland Jaw Sealer/Divider. Reprocessed Device for Single Use. Reprocessed by

Prestige Endoscopic Graspers Instructions for Use

10 Best Practices for Electrosurgical Unit Testing

MyoSure Hysteroscopic Tissue Removal System MyoSure LITE Tissue Removal Device Instructions for Use

VividTrac R. Video Intubation Device. INSTRUCTIONS FOR USE. English

ASAHI Neurovascular Guide Wire

MyoSure Setup Simplified

ALAN. Electrosurgery Re-defined. AN ISO : 2003 COMPANY

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

Instruction Manual. ErgoLAP BIPOLAR

Fingertip Pulse Oximeter

Introduction. J-PLASMA and Laser have many similarities, but also some important differences.

USER MANUAL LTH-255A Hand Piece

HelixAR. Electrosurgical Generator with Argon Beam Coagulation

DSL# Rev. 12/ All Rights Reserved. PR1011-HS. Product Catalog

Safety Powder Spray Systems

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

GATOR Battery-powered Pistol Grip Cable Cutters

AUTOVENT 4000 VENTILATOR

PreludeSYNC. Radial Compression Device INSTRUCTIONS FOR USE

OWNER S MANUAL. Page: 1 of 8

REL-510H WARNING NOTICE 12 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD

Manual Actuated Boiler Blowdown Valves

REL-46 WARNING NOTICE 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD. Compatible with RELIABLE R15 and P Style dies. REL-46 Manual

Aquamantys System Generator Verification

CinchLock SS (Sports Sheath) Knotless Labrum Restoration SURGICAL TECHNIQUE

Leadership in Fluid Management

Frigitronics CRYO-PLUS

Use this key to determine which sections of this Product Manual apply to your job.

Simplified Automated Ventilator Training Slides

Injector system for implantation of 1stQ foldable intraocular lenses

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Paradise Pump Series Operator s Manual

Preparation and Installation of the Sanitary BDI-FLX Sensor and Connection to the BDI-FLX Interface Cable

Indications for Use: Caution: Note:

25+ instruments answer the most common barrier to 25G instrument flexibility.

PDY TON HYDRAULIC CRIMPING TOOL WARNING

Mirage Vista Nasal Mask

TJF-Q180V Cleaning and Disinfection Checklist

Med Aire Alternating Pressure Pump and Pad System

AnastoClip GC Closure System

Hercules 3 Universal Stabilizer Arm. Catalog Number

TJF-Q180V Cleaning and Disinfection Checklist

Domiciliary Oxygen Switch. Instructions for Use

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

Pneumatic Kerrison. Technology that makes life easier. Aesculap Neurosurgery

PS Plasma 375 Plasma Cutter Kit Assembly & Operating Instructions

Confidence comes with knowing you are Code-Ready.

Oxygen Dialflow Meter. Instructions for Use

POWERPRESS UNIT. User Manual. Gradient Pneumatic Sequential Compressor. Neomedic

750 Cable Cutter INSTRUCTION MANUAL

SC505 Sof Care Inflator. Operating Instructions and Service Manual

User Instructions 1789 Parapet Wall Anchor

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

2009 G3 Genuine Guide Gear Rev -

INSTRUCTIONS FOR USE (US)

O-Two Self-Study Guide. Equinox Relieve

Defibtech Automated External Defibrillator

THIS PAGE INTENTIONALLY BLANK

77-I0PN230 Rev. A. Model 1030 User Manual

Hemorrhoid Multiple Band Ligator Set Instructions for Use

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Oxygen Dialflow Meter. Instructions for Use

Instructions for Use. OptiLife Nasal Lab Mask

Regalia Oxygen Concentrator

Composite Pistol-Type Air Needle Scaler OWNER S MANUAL

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

LEEP PRECISION TM Integrated System, 120 VAC Model LP Operating Manual Rev. A 10/13

INSTRUCTIONS FOR USE - US RaplixaSpray (Raplixa Delivery Device)

CS CANNULATION. Step-by-Step Guide. This guide contains recommendations as provided by Dr. Seth Worley, MD POWERED BY

TEM# Air Hammer Combo USER'S MANUAL

AND OPERATION INSTRUCTIONS

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Electrical Safety in Hospitals

USER S INFORMATION MANUAL

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

MEDIANA HEARTON AED A15. Mediana Automated External Defibrillator A15

This section provides general information and recommendations required prior to installing, operating or maintaining your KAT Master Hitch.

trufreeze Spray Kit CO CO Instructions for Use

Rev. B. Operating Instructions

INTENDED USE TECHNICAL SPECIFICATIONS

COMMON PRECAUTIONS Data Sheet - Be Sure To Read Before Handling Page 1 of 6

OPERATION INSTRUCTIONS

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

NWC EMSS Skill Performance Record Application & Use LUCAS CPR DEVICE

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

LEEP System 1000 TABLE OF CONTENTS. I. Professional Use Guide...2. II. System Features III. Electrical Precautions...14

MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS

BIORAPTOR Knotless Suture Anchor Arthroscopic Shoulder Instability Repair. Shoulder Series Technique Guide by: Thomas Youm, MD

Micro Dial-Flowmeter. Instructions for Use

U.S. CONCRETE, INC. SAFETY POLICY and PROCEDURE MANUAL

Arbor Terminator Press Operation Manual For IDT Connectors Order No Engineering No: AM

NOTE: Parts are no longer available for this tool. The manual will continue on the next page.

Instructions for Installation, Use and Maintenance FSI Safety Tank Shower Systems

STS PECVD Instructions

Reflow Oven HHL3000 INSTRUCTION MANUAL POHUA - jedyny autoryzowany przedstawiciel w Polsce

INSTRUCTION MANUAL. (Variante 2 / System 2) Surgery table standard (Variante 1 / System 1)

Recommendations - USA

Transcription:

Page 1 of 8 Instructions for Use for CAIMAN Instrument (12 mm x 44 cm) Item #: US731SU CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. CAREFULLY READ ALL INSTRUCTIONS PRIOR TO USE. OBSERVE ALL WARNINGS AND PRECAUTIONS NOTED THROUGHOUT THESE INSTRUCTIONS. FAILURE TO DO SO MAY RESULT IN COMPLICATIONS. The Lektrafuse Radio Frequency (RF) Generator Instructions for Use (IFU) is supplied as a separate document. The Generator IFU contains detailed information describing the Lektrafuse RF System components, generator setup, guide to operation, maintenance, troubleshooting, and specifications. Please read the IFU prior to operation of the Lektrafuse RF Generator. BRIEF SYSTEM DESCRIPTION The CAIMAN Instrument contained in this package is a sterile, single use component of the Lektrafuse System and is designed to deliver radiofrequency (RF) energy from the RF Generator. The CAIMAN Instrument can only be used in conjunction with the dedicated Lektrafuse RF Generator. The CAIMAN Instrument is only compatible with the Lektrafuse RF Generator, which supplies a maximum voltage of 115Vrms. INDICATION FOR USE The CAIMAN Instrument is a dedicated bipolar electrosurgical instrument intended for use in general surgical and gynecologic laparoscopic procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structures (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue. The indications for use include general laparoscopic procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological laparoscopic procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resections, and oophorectomy etc., or any procedure where vessel ligation (cutting and sealing), tissue grasping, and dissection is performed. The device can be used on vessels up to and including 7 mm and bundles as large as will fit in the jaws of the instrument. The CAIMAN Instrument has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures. CONTRAINDICATIONS The CAIMAN Instrument is not effective for use in tubal sterilization or tubal coagulation for sterilization purposes. Do not use the CAIMAN Instrument for these procedures. WARNINGS Inspect the CAIMAN Instrument for damages. Do not use the instrument if the Tyvek tray seal has been compromised or if there is any evidence of damage, such as broken or missing components. Either contact Aesculap, Inc. for a replacement or dispose of the CAIMAN Instrument abiding all regulation on disposing disposable medical instruments.

Page 2 of 8 The CAIMAN Instrument is intended for use ONLY with the dedicated Lektrafuse RF Generator. Use of the CAIMAN Instrument with any other generators produced by other manufacturers may result in injury to the patient and/or damage to the Instrument. Only properly trained personnel should use the CAIMAN Instrument. The CAIMAN Instrument should not be used when, in the judgment of the treating physician, the RF procedures would be contrary to the best interests of the patient. The CAIMAN Instrument is not effective for use in tubal sterilization or tubal coagulation for sterilization purposes. Do not use the CAIMAN Instrument for these procedures. The CAIMAN Instrument is intended for sealing and dividing tissues, including ligaments and vessels, or any procedure where vessel ligation (cutting and sealing), tissue grasping, and dissection is performed. Use caution during surgical cases in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurismal vessels, etc.). For best results, apply the seal to unaffected vasculature. Do not use the CAIMAN Instrument on vessels in excess of 7 mm in diameter. Do not use hybrid trocars that are comprised of both metal and plastic components. Capacitive coupling of RF current may cause unintended burns. Inspect the Instrument and cable/cords for breaks, crack, nicks, or other damage before use. If damaged, do not use. Failure to observe this caution may result in injury or electrical shock to the patient or surgical team or cause damage to the Instrument. Position CAIMAN Instrument cords to avoid contact with the patient or other leads. Do not wrap cords around metal objects. This may induce currents that could lead to shocks, fires, or injury to the patient or surgical team. Examine all connections to the Lektrafuse RF Generator and all Instruments before using. Improper connection may result in arcs, sparks, accessory malfunction, or unintended surgical effects. Do not use the CAIMAN Instrument in the presence of flammable anesthetic or other flammable liquids, gases, or other flammable materials. Where the CAIMAN Instrument is utilized, prevent the accumulation of oxygen or other oxidizing gases. The risk of igniting flammable gases or other materials is inherent in electrosurgery and cannot be eliminated by device design. Precautions must be taken to avoid contact of the CAIMAN Instrument electrodes and flammable materials and substances, including some forms of anesthetic or skin preparation agents, flammable materials and substances produced by natural processes within body cavities, and flammable materials and substances such as surgical drapes or other materials present in the operative field. During a seal cycle, energy is applied to the area between the instrument jaws. This energy may cause water to be converted into steam. The thermal energy of steam may cause unintended injury in close proximity to the jaws. Care should be taken in surgical procedures occurring in confined spaces in anticipation of this possibility. Studies have shown that smoke generated during electrosurgical procedures may be harmful to surgical personnel. These studies recommend using an appropriate surgical mask and adequate ventilation of the smoke by use of a surgical smoke evacuator or other means.

Page 3 of 8 Use the CAIMAN Instrument with caution in the presence of an internal or external pacemaker. Interference produced by electrosurgical devices can cause devices such as pacemakers to enter an asynchronous mode or can block the pacemaker effect entirely. Consult with the pacemaker manufacturer or hospital biomedical engineering personnel for further information. Use the CAIMAN Instrument with caution in the presence of an internal cardiac defibrillator (ICD). Electrosurgery may cause multiple activations of ICDs. Consult with the ICD manufacturer or hospital biomedical engineering personnel for further information when use of electrosurgical application is planned in patients with ICDs. The CAIMAN Instrument has a marker at the distal tip of the jaws which indicates the furthest extent to which the cutting blade can be advanced. Any tissue distal to this marker will NOT be cut during the dissection process. Contact between any active instrument electrode and any metal object (hemostats, staples, clips, retractors, etc.) may increase current flow and can result in unintended surgical effects such as an effect at an unintended site or insufficient energy deposition. Do not activate the Lektrafuse RF Generator until the Instrument has been applied with the proper jaw pressure. Activating the system before this is done may result in improper sealing and may increase thermal spread to tissue outside the surgical site. Do not activate the Lektrafuse RF Generator in an open circuit condition. Activate the generator only when the instrument is near or in direct contact with the target tissue to reduce the possibility of unintended burns. Always keep the external surface of the instrument jaws away from adjacent tissue while activating the Lektrafuse RF Generator. Do not activate RF energy if the jaw cannot be visualized. Do not activate the CAIMAN Instrument while the instrument jaws are in contact with, or in close proximity to, other conductive instruments including metal cannulas, as localized burns to the patient or physician may occur. PRECAUTIONS Avoid placing fingers in the handle ratchet mechanism (between the handle and the jaw closure lever). Injury to the User may result. Carefully insert and withdraw CAIMAN Instrument from cannulas to avoid possible damage to the devices and/or injury to the patient. The CAIMAN Instrument must be disposed of properly in accordance with the applicable national/local/state environmental laws. The external surfaces of the CAIMAN Instrument jaws may remain hot enough to cause burns after the RF current is deactivated. Inadvertent activation or movement of the RF activated CAIMAN Instrument outside of the field of vision may result in injury to the patient. Do not open and close the jaws during the RF sealing cycle. An incomplete seal may occur. Reseal with RF energy prior to division.

Page 4 of 8 ANIMAL TESTING Aesculap, Inc. conducted chronic animal studies to demonstrate the effectiveness of the Lektrafuse RF System with the CAIMAN Instrument for laparoscopic general surgery and gynecological procedures and the sealing of vessels up to and including 7 mm. Three canines and one porcine were selected to undergo general surgical and gynecological procedures utilizing the CAIMAN Instrument to seal and divide vessels up to and including 7 mm. The vessels sealed ranged from 2 7 mm. A total of 115 visible vessels were sealed. The animals were survived for 21 days to determine if there was any leakage at the vessel occlusion site due to thermal injury and subsequent tissue necrosis and if any injuries to adjacent structures or organs had occurred. The outcomes demonstrated no visible macroscopic vessel leaking and no injury or damage to adjacent structures or organs. Potential adverse events from thermal spread that might cause or contribute to injury or damage to adjacent structures or organs were not observed in this study Clearance of the CAIMAN Instrument was not based on human clinical testing. USE OF THE LEKTRAFUSE GENERATOR SYSTEM DURING SURGERY PATIENT AND OPERATING ROOM SAFETY The safe and effective use of electrosurgical equipment is largely dependent upon factors solely under the control of the user and associated operating room personnel. There is no substitute for a properly trained and vigilant surgical team. It is important that the operating instructions supplied with the Lektrafuse RF Generator Operator s Manual be read, understood, and followed. The CAIMAN Instrument is designed for general laparoscopic procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological laparoscopic procedures where ligation and division of vessels is performed. The device can be used on vessels up to and including 7 mm and bundles as large as will fit in the jaws of the Instrument. OPERATION OF THE LEKTRAFUSE SYSTEM The Lektrafuse RF System is designed to seal and divide tissue with a single application (50 mm represents complete coverage of the CAIMAN Instrument jaws by tissue). After first performing necessary tissue dissection; the user determines length of tissue to be sealed and divided with any individual application of the CAIMAN Instrument (within the range of 10 50 mm), and where within the CAIMAN Instrument jaws the tissue is to be placed. CAUTION: The selected tissue should not be bunched, folded, or layered within the Instrument jaws. Failure to correctly position the tissue within the Instrument jaws may result in inadequate tissue sealing and/or division. TISSUE AND VESSEL SEALING SETTINGS Lektrafuse s Technology has the capability that enables the user to seal and divide 10 50 mm length of variable thickness tissue. Aesculap, Inc. has achieved this through a series of proprietary designs.

Page 5 of 8 The jaw design optimizes the pressure between the two jaws resulting in uniform force distribution along the tissue length. This allows the user to uniformly and consistently seal and divide any length of tissue of varying thickness or compressibility. In addition, the RF Generator software is capable of sensing electrical parameters to determine the level of tissue coverage and the thickness of such tissue to optimize the sealing process. Based on these parameters, the risk of arcing is reduced. Therefore the energy settings allow for uniform and consistent sealing regardless of the degree of vascularity, vessel size, and length of tissue being sealed. Sealing Duration and Power Setting do not require any User adjustment. The sealing parameters are monitored by the software based on: Impedance RF Voltage RF Current The Lektrafuse technology allows the CAIMAN Instrument to be used on vessels up to and including 7 mm in diameter. If sealing conditions are not met, a Regrasp Indicator series of audible tones is triggered by the Lektrafuse RF Generator to inform the user that adequate tissue sealing may not have been achieved following the delivery of RF energy. An additional tissue sealing cycle may then be performed by the user, or the user may remove the CAIMAN Instrument and divide the sealed tissue manually (using surgical scissors). When RF energy is applied, the maximum RF cycle is 17.5 seconds. The energy setting and application duration has been validated by performing bench, animal and software validation. SET UP (Connecting the CAIMAN Instrument to the Lektrafuse RF Generator) Open the CAIMAN Instrument package using standard sterile technique. The CAIMAN Instrument RF cable ends in a connector; directional arrows are indicated on one side. On the left front of the Lektrafuse RF Generator is the connector port with a red dot in the superior (12 O clock) position. Align the arrows on the cable connector with the red dot on the connector port and push the connector into the connector port. If you feel any resistance, stop pushing, realign the arrows and the red dot, and try again. There should be little to no resistance if the connector is correctly aligned within the connector port. Advance the connector fully into the connector port, until no further connector may be advanced. Inadequate or incomplete advancement of the connector into the connector port may result in failure of the System to correctly or adequately operate. If the connector is correctly advanced into the connector port, the green ring encircling the connector port on the Lektrafuse RF Generator will illuminate.

Page 6 of 8 STEP BY STEP USE OF THE CAIMAN INSTRUMENT: Rotating Instrument Shaft and Jaws Turn the rotator (a) on the Instrument until the jaws are in the required position. A black line on the rotator designates when the CAIMAN Instrument jaw is in the midline or horizontal position. Articulating Instrument Jaw Wrist Turn the articulation lever (b) with a left and right movement to the required position and the jaw wrist (c) will move accordingly. 1. The CAIMAN Instrument allows jaw rotation via the rotator (a), and articulation with right and left movement of the jaw articulation (c) via the articulation lever (b). 2. Maneuver the CAIMAN Instrument jaw to the desired tissue by turning the rotator (a) or pulling/pushing right and left on the articulation lever (b). 3. Place the tissue in the CAIMAN Instrument jaws centering it between the black line at the base of the jaws and the black line on the tip of the jaws. 4. With the selected tissue (10 50 mm in length) positioned in the CAIMAN Instrument, squeeze (close) the Jaw Closure Lever (d) (the jaws will close) until the Lever palpably locks into the closed position. The jaws will remain locked. Note: If not satisfied with the selection or placement of tissue within the jaws, gently squeeze the Jaw Closure Lever (d), opening the jaws and reposition the tissue selected in the jaws. 5. To apply RF energy, either press the red RF button (f) on the handle, tap the Foot Pedal, or push the RF Delivery Control Button (on the front of the RF Generator). 6. The termination of audible tones indicates completion of tissue sealing and is accompanied by the loss of illumination of the blue ring encircling the RF Delivery Control Button on the RF Generator. 7. If division (cutting) of the sealed tissue is desired, fully retract the Cutter Lever (e) located on the bottom of the Instrument by pressing towards the handle in a trigger like motion. The Cutter Lever automatically returns to its original position. 8. To open the CAIMAN Instrument jaws (following sealed tissue division, or following tissue sealing if division is not required), gently squeeze the Jaw Closure Lever (d) to unlock as the jaw closure lever returns to its former open position. 9. Upon removal of the Instrument away from the dissected tissue, visually inspect the jaw surfaces. If cauterized tissue and/or blood is readily visible on the CAIMAN Instrument electrodes, gently remove the material using a sterile surgical pad/gauze dampened with sterile water. REGRASP INDICATOR CAUTION: The Lektrafuse RF System is designed to identify electrical characteristics indicating potentially inadequate sealing of the tissue within the instrument jaws. If such potentially inadequate sealing is identified, a unique and distinct audible series of tones will immediately sound from the RF Generator following automatic termination of the audible tones associated with tissue sealing. This unique and distinct tone is a Regrasp tone and is accompanied by the illumination of the amber Regrasp indicator light on the front of the RF Generator and a Regrasp message on the RF Generator Screen.

Page 7 of 8 If a Regrasp occurs, 1. do not divide the tissue with the cutter 2. open the instrument jaws 3. visually examine the sealed tissue 4. manually divide the tissue with surgical scissors OR 1. reapply the jaws to the tissue and lock the jaws in place 2. deliver an additional cycle of RF, which may result in a normal seal process (which can then be followed by routine sealed tissue division, jaw release, and instrument removal) If a Regrasp reoccurs, the sealed tissue should not be divided, the jaws should be released, and the sealed tissue divided manually with surgical scissors and hemostasis achieved where required. A single press and release action on either the Foot Pedal, RF Delivery Control Button (on the front of the RF Generator), or the RF Button (on the CAIMAN Instrument) will (1) automatically terminate the Regrasp tones, amber Regrasp indicator light on the front of the RF Generator, and the Regrasp message on the RF Generator Screen; and (2) setup the System for the next standard tissue sealing and division process. DISCARD USED INSTRUMENT Discard the CAIMAN Instrument after use in accordance with hospital policy for biohazards and sharps. DO NOT RESTERILIZE. Manufactured for: Aesculap Inc. 3773 Corporate Parkway Center Valley, PA 18034 USA 800 258 1946 www.aesculapusa.com SOP AIC 5001019 Rev. 04 07/14

Page 8 of 8 Functional Operation of the CAIMAN Instrument (12 mm x 44 cm) The below figure identifies the operational rotator/levers/button used to operate the CAIMAN Instrument