Ju L ENGINEERING D A T A TRANSMITTAL PW. 1. l.edt BD ( 0 4 / 9 4 ) GEFO97 BD

Similar documents
Spreadsheet Description Document for Saturation Temperature Calculation. UC: NA Charge Code: Total Pages:

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

This test shall be carried out on all vehicles equipped with open type traction batteries.

O3 3E 1 F Gas Sensor Module

Calibration Requirements for Direct Reading Confined Space Gas Detectors

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Phosgene Sensor. Sensoric COCl2 3E 1

id? UL 30 ita5 ENGINEERING DATA TRANSMllTAL SWMFE SWM 1000 C (04194)GEFO~PGMACPATH-GEF Page 1 of 1.EDT 11. Receiver Remarks: BD I

Sensoric 4-20 ma Transmitter Board Operation Manual

Applications Note: Use of "pentane equivalent" calibration gas mixtures

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

MODEL GT820 OXYGEN SENSOR

Ozone. Sensoric O3 3E 1

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Sensepoint XCD. Calibration Guide

Application Note AN-107

Operating Characteristics and Handling Manual for the NAP-55A / NAP-50A Explosive/Flammable Gas Sensor NEMOTO

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

P-5215 Differential Pressure Transmitter

The Discussion of this exercise covers the following points: Range with an elevated or suppressed zero Suppressed-zero range Elevated-zero range

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

General Accreditation Guidance. User checks and maintenance of laboratory balances

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

GasLab SAN-102 Micro Oxygen Monitor User Manual

LCLS Heavy Met Outgassing Tests * K. I. Kishiyama, D. M. Behne Lawrence Livermore National Laboratory

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

SPECIFICATION. date of manufacture. Intermittent: -20 o C to +50 o C. Pressure Range:

USER'S MANUAL HOT-WIRE TYPE GAS SENSOR NAP-100A SERIES ( NET10A(R) ) (For All Combustible Gases)

Drift Characteristics of Paroscientific pressure sensors

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

Transmitter CS 21 Operation Manual

Best Practice Guide, Servomex 2700

@ No. D A Bragg/D B Parkman. 241-C System/Bldg/Fadliv

It s Just Calibration How Hard Can It Be?

Usage Policies Notebook for Thermco Atmospheric Diffusion Furnace system

Do Not Print This Page.

Digiquartz Water-Balanced Pressure Sensors for AUV, ROV, and other Moving Underwater Applications

Specifications and information are subject to change without notice. Up-to-date address information is available on our website.

RGC-IR Remote Gas Calibrator for IR400

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.

Gas Sampling in the DST

Test Plans & Test Results

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

CALIBRATION KIT. Instruction Manual

PG7000 high line differential mode for defining differential pressure at line pressures above atmospheric pressure (PG7000 V2.

Model Description

WEBINARS. Hazard Communication

Fail Operational Controls for an Independent Metering Valve

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

A review of best practices for Selection, Installation, Operation and Maintenance of Gas meters for Flare Applications used for Managing facility

Cover Page for Lab Report Group Portion. Pump Performance

LABORATORY EXERCISE 1 CONTROL VALVE CHARACTERISTICS

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

23 RD INTERNATIONAL SYMPOSIUM ON BALLISTICS TARRAGONA, SPAIN APRIL 2007

CCT-7320/ROC-2313 Reverse Osmosis Controller

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Scope: This plan applies to all personnel, including contractors, who enter or work in confined spaces, or supervise such activities.

Exhibit 4. Determination of Static Pressure Performance of the Healy Clean Air Separator (Executive Orders VR-201-N and VR-202-N)

Engineering Evaluation of Nitrogen Purge Fill Station

International Research Journal of Electronics and Computer Engineering Vol 1(3) Oct-Dec 2015

Navigator 600 Silica analyzers

Standard Operating Procedure for H 2 S and SO 2 Converter

Transmitter CS 21 Operation Manual

PUBLISHED PROJECT REPORT PPR850. Optimisation of water flow depth for SCRIM. S Brittain, P Sanders and H Viner

AMT-Ex Dewpoint Transmitter

GasLab SAN-102 Micro Oxygen Monitor User Manual

AMS 6916 Board mount pressure sensor with ratiometric analog output

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

Title: Standard Operating Procedure for Sabio 2010 and 4010 Dilution Calibrator

EN00 Pre-Instructional Survey

The Discussion of this exercise covers the following points:

BAPI Pressure Line of Products - FAQs

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

Electro Controls. Level WattsIndustries.co.uk

Manual of SF6 Comprehensive Tester

DPC-30 DPC-100. Reference Manual

APPENDIX C. 40 CFR PART 60 App. B SPEC. 1-4, 6

MIL-STD-883G METHOD

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Installation and Operation Manual

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Plasma. Argon Production Helium Liquefaction Truck Loading Pure Gas Bottling

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Digital Vacuum Regulator

OP CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST

UTAS EH&S INSTRUCTION No. 036 June 2012 Revision No. 3. Purging of Flammable Gas Equipment and/or Piping (Explosion Hazard)

Constant Pressure Inlet (CCN) Operator Manual

Transcription:

W Ju L 3 0 1336 PW. 1 ENGINEERING D A T A TRANSMITTAL l.edt NO.' 615278 ~~ BD-7400-172-2 ( 0 4 / 9 4 ) GEFO97 BD-7400172.1

WHC-SD-WM-TRP-249, Rev. 0 Combustible Gas Sensor Interference Test Report B. J. Webb Westinghouse Hanford Company, Richland, WA 99352 U. S. Department o f Energy Contract DE-AC06-87RL10930 EDT/ECN: c 1527Y Org Code: 74840 B&R Code: EW3120072 Key Words: UC: 2030 Charge Code: T o t a l Pages: N1737 ys /6 G3tA4 Flammable Gas; Hydrogen; Ammonia Abstract: Result o f Laboratory I n t e r f e r e n c e t e s t i n g o f t h e S i e r r a Monitor Corporation Gas Sensor, number model 4101-2, are documnented. TRADEMARK DISCLAIMER. Reference herein t o any s p e c i f i c comnercial product, process, o r service by trade name, trademark, manufacturer, o r otherwise, does not necessarily c o n s t i t u t e o r imply i t s endorsement, reconmendation, o r favoring by the u n i t e d States Goverment or any agency thereof o r i t s contractors or subcontractors. P r i n t e d i n the United states of America. TO obtain copies of t h i s d a c m n t, contact: UHC/BCS Document Control Services. P.O. Box 1970, Mailstop Hb-08, Richland UA 99352, Phone (509) 372-2420; Fax (509) 376-4989. f;;:jajl R e l e a d Approval Dbe Approved for Public Release A-6400.073 (10/95) GEF321

COMBUSTIBLE GAS SENSOR INTERFERENCE TEST REPORT Approval Designator: SQ Eric K. Straalsund Characterization Monitoring Development and Brent J. Webb Waste Tank Stabilization Group May, 1996 Prepared for United States Department of Energy Westinghouse Hanford Company P.O. Box 1970 Richland, Washington 99352

WHC.SD.WM.TRP.249, Revision 0 COMBUSTIBLE GAS SENSOR INTERFERENCE TEST REPORT TABLE OF CONTENTS 1.0 2.0 3.0 4.0... BACKGROUND... SUMMARY RESULTS AND OBSERVATIONS... CONCLUSIONS... INTRODUCTION 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9 4.10 4.11 4.12 4.13 4.14 4.15 5.0 1 2 3 5... 5... 5... 5... 6 8... 8... 8... 8... 9... 10... 10... 1 0... 11... 12 D e t a i l e d T e s t D e s c r i p t i o n and D a t a Pretest A c t i v i t i e s T e s t A1 T e s t B1 T e s t C 1... T e s t s I1 and I 2 T e s t s D1 a n d D2 T e s t K2 T e s t K3 T e s t K4 T e s t K5 R e s p o n s e T i m e T e s t s Summary T e s t s E 2 and F1 T e s t M2 T e s t M 3... CONCLUSIONS... 12 13 L I S T O F FIGURES Figure Figure Figure Figure Figure Figure 1 2 3 4 5 6... 6... 7... 7... 9... 10... 11 MC S e n s o r R e s p o n s e Raw vs. Filtered S i g n a l T e s t A 1.SMC L i n e a r i t y T e s t B1.SMC S t a b i l i t y T e s t C 1.SMC W a r m U p T i m e SMC H u m i d i t y T e s t SMC R e s p o n s e T e s t L I S T O F TABLES Table 1 Table 2 T e s t El R e s u l t s T e s t F1 R e s u l t s...... I 12 12

1.0 INTRODUCTION Tests on the SMC combustible gas monitor, Model 4101-2, S/N 95-B-4101, were conducted in the 306E gas sensor laboratory beginning July 14, 1995. These tests were conducted per the "Combustible Gas Sensor Interference Test Plan," WHC-SD-WM-TP- 374. The purpose of the tests was to evaluate sensor range, accuracy, and response time, and the effects of ammonia, nitrous oxide, and humidity on the sensor in the presence of hydrogen gas. I

2.0 BACKGROUND Twenty-five (25) underground storage tanks (USTS) on the Hanford reservation are currently listed on the flammable gas watch list (FGWL). These tanks have been identified as having the potential to contain trapped gases--most notably, trapped quantities of hydrogen, a recognized flammable gas. WHC requires strict work controls and monitoring for all activities conducted in FGWL tanks. Typically, this monitoring for flammable gases on FGWL is accomplished through either the use of a standard hydrogen monitoring system (SHMS) or a hand-held combustible gas analyzer. However, several work activities are planned that require additional process controls such as the ability to automatically shut down or interlock process equipment in the event of a measurable quantity of hydrogen and/or other flammable gas. This test was specifically conducted to assess the applicability of using an SMC combustible gas monitor, Model No. 4101-2, for use in a shutdown interlock system for FGWL tanks. The results of evaluation testing of the sensor are documented herein. 2

3.0 SUMMARY RESULTS AND OBSERVATIONS Testing conducted on the SMC sensor verified that the accuracy and range of the sensor for measuring flammable gas was consistent with the manufacturer's specification of 3% of full scale or f 10% of applied concentration, whichever is greater. The lower detection limit of the SMC sensor was determined to be about 1000 ppm hydrogen or 2.5% LFL. Sensor response time was measured to be nominally 23 seconds to 60% of final reading when subjected to a step response of 7070 ppm hydrogen in air. In all cases tested, mixing hydrogen gas with either anunonia or nitrous oxide produced conservative results. No evidence of short-term sensor poisoning from anunonia, nitrous oxide, or water vapor was observed. Several important observations are worth noting that have direct impact on applications: 1. The sensor tested was not rated for operation in a Class 1, Division 1, Group B (hydrogen) environment. 2. The output signal of the model tested was noisy and needs to be processed to prevent spurious alarms. This can be rectified by using a Newport voltmeter with signal processing. 3. The response of the sensor was significantly affected by the gas flow rate used in testing. This is common for combustible gas (catalytic bead) sensors as the flow rate affects the heat transfer and thus, affects resultant readings from the sensor. Extreme care should be taken to ensure that the gas flow should be well regulated and constant during normal operations as well as during calibration. 4. The response of the sensor was significantly affected by the vertical and angular orientation of the sensor. This is attributed to the fact that orientation affects heat transfer between the active and passive beads. Because of this, sensor orientation during calibration match that of its intended operation. 5. During subsequent testing of the sensor, calibrations were performed that attempted to "boostr' the output and improve low-end resolution by reducing the full scale span of the instrument. In all cases, this "boosted" calibration 3

.... WHC-SD-WM-TRP-249, Revision 0 resulted in poor sensor stability and increased drift. It is recommended that the instrument be calibrated using the primary gas of interest (in this case, hydrogen) and calibrated per factory instructions. 6. The SMC sensor is equipped with automatic power-on delay. If power is interrupted to the sensor, output will be set to zero for approximately four minutes. 4

4.0 CONCLUSIONS If proper care is taken to ensure that calibration conditions match operating conditions, the SMC sensor should provide fast, repeatable detection of flammable gas. For applications involving fixed mounting conditions at known gas flow rates, the sensor should provide conservative results provided that the appropriate calibration gas is used. However, because of NEC classification issues and effects of sensor position, the sensor tested should not be considered for the mobile color camera sensor. Several other combustible gas sensors are on the market which would more adequately match this application. 4.1 Detailed Test Description and Data Tests conducted on the SMC sensor were conducted per approved. test procedure. A copy of the completed test procedure and data sheets is provided as the Attachment. 4.2 Pretest Activities The SMC sensor was powered up, calibrated, and hooked up to the LabView data acquisition system. Because the LabView system is a high-speed data acquisition system, we were able to detect an anomaly with the SMC signal that appeared to be a result of the sensor s circuit design. The noise was measured by an oscilloscope and determined to be 120 cycles/second and 8.3 khz. A Newport voltmeter with output processing was installed into the system-and the noise was effectively filtered, even at the lowest filter setting. Figure 1 shows both the raw and filtered response of the SMC to a series of calibration gas injections. 4.3 Test A1 The SMC sensor was calibrated per manufacturer s recommendations and subjected to 10% step increases from zero to nominal 10,100 ppm hydrogen. The instrument exhibited a nonlinear behavior at the low end and lower detection limit of about 1000 ppm. This test was repeated with similar results as Test A2. This experiment was also repeated after recalibrating the sensor using more gain. Data from this experiment is shown as Figure 2. Examination of Figure 2 shows that with the exception of the high span gas (10,100 ppm hydrogen), all data points are slightly under-reported by the sensor. 5

WHC-SD-WM-TRP-249, Revision 0 4.4 Test B1 The SMC sensor was subjected to a prolonged injection (over 2 hours) of 6705 ppm hydrogen gas to determine stability. Results of this test suggest a stable signal with no appreciable drift over this time frame. Data from Test B1 is shown graphically as Figure 3. Figure 1. SMC Sensor Response R a w v s. Filtered Signal 1431) I1 I PPM Standard - I Km a m. r1 Test Time 6

WC-SD-WM-TRP-249, Revision 0 Figure 2. Test A 1 - SMC Linearity n mn 4m m m lixkxl 12mo PPM Measured Figure 3. Test B1 - SMC Stability / Drift 7wo PPM Standard 4Mo m 1Mo O'b0.00 011:21 02R.IR 0.4312 057.36 112.00 1.2621 1:POIR 1:5512 20936 221:W Elapsed Test Time 7

4.5 Test C1 The SMC sensor was again tested with 6705 ppm hydrogen, this time after the unit was powered off for over 12 hours. Results from this test verified that the SMC WILL NOT measure hydrogen for 4 minutes following power-down conditions. Response after the four minutes was essentially instantaneous. Figure 4 shows a plot of data from Test C1. 4.6 Tests I1 and I2 These tests, not described in the original test procedure, were conducted to assess the effects of humidity on the SMC response. The tests involved passing the 6705 ppm hydrogen through a bubbler prior to the sensor. The first test indicated a reduced response; however, this was determined to be a result of the time required to purge the relatively large bubbler volume. The test was repeated as Test 12, this time allowing the bubbler to purge prior to hooking up gas to the SMC. Results from this test showed no appreciable decrease in sensitivity or response time as a result of the humidity. Data from Test I2 is shown graphically as Figure 5. 4.7 Tests D1 and D2 Tests D1 and D2 were performed to determine the response time of the SMC sensor. Test D1 applied a step response of 4040 ppm hydrogen, and Test D2 applied a step response of 7070 ppm hydrogen to the SMC sensor. Both tests showed similar results with the SMC sensor approaching 90% of final reading in about 50 seconds. Data from both tests indicate a "dead" period of about 14 seconds before the sensor starts to respond. This "dead" period may be a result of sensor performance, diffusion through the sintered metal filter, or purge time for the calibration housing. Additional response time tests were performed as Test Series K to examine the repeatability and effects of special calibrations on response time. 4.8 Test K2 A response time test, with a "step" increase of hydrogen from 0 to 7070 ppm hydrogen, was similar to Test D2. Results similar to Test D2 showed about 3 seconds less of "dead" time in sensor response. I attribute this to a better "live" zero calibration. Again, test data emphasize importance of calibrations. a

4.9 Test K3 A response time test was performed immediately after recalibration. The sensor was subjected to a 0-7070 ppm "step" increase. Response time was consistent with response time from Test K2. Figure 4. Test C1 - SMC Warm Up Time PPM Standard Elapsed Test Time 9

4.10 Test K4 The sensor was recalibrated, and Test K3 was repeated as Test K4. Results were consistent with Test K3. It was noticed that the output of the sensor was less stable for special boosted calibrations in which the full scale span of the sensor was reduced. 4.11 Test K5 A sensor response time test was performed. The sensor was subjected to a step increase of 3030-7070 ppm. The response time was consistent.with previous response time tests. 4.12 Response Time Tests Summary All response time tests showed similar results: the sensor starts to respond in approximately 10 seconds and reaches 60% response in about 20-25 seconds. This calculated response time calculated from this data is considered to be conservative. This is because it is assumed that the sensor was subjected to a step response in gas concentration while, in reality, the small volume in the calibration head requires time to mix, and a true step 10

response cannot be achieved in our current laboratory setup. Test data for all response time tests are shown in Figure 6. 4.13 Tests E2 and F1 Tests E2 and F1 were performed to assess the effects of nitrous oxide or ammonia on the SMC. During conduct of the first nitrous oxide test, the lab gas divider malfunctioned and had to be repaired. The test was repeated and no negative interference or "poison" effects from nitrous oxide were observed. However, an interesting phenomenon was noted in that the sensor signal did not stabilize in the presence of high quantities of nitrous oxide. Ammonia testing again showed no negative or poison effects but rather a small (approximately 25%) increase in signal. Data from these experiments are shown in Table 1 for nitrous-oxide and Table 2 for ammonia. Figure 6. SMC Response Time Tests Elapsed Test Time (sec) 11

Table 1. Test E2 Results H2 Standard g SMC Readinq GC Readinq 7450 0 7085 7517 0 67700 *1190 0 7450 0 7015 7549 3725 34700 3395 3815 3725 0 2609 3789 *Output climbing after 4 hours. Table 2. Test F1 Results H2 Standard SMC Reading GC Reading 10100 0 9790 10179 0 10100 5050 5050 5090 0 2545 0 2110 10083 6020 4690 0 10218 5163 5176 4.14 Test M 2 Approximately 5050 ppm hydrogen at 100 cc/minute was applied to the sensor, and the sensor was rotated about its horizontal axis. Sensor response was significantly affected by orientation, with over 50% reduction in signal at 180 degree rotation. This implies that sensor position in calibration and operation mode MUST be identical. 4.15 Test M3 Approximately 5050 ppm hydrogen was applied to the sensor originally mounted in the horizontal position. Sensor was progressively rotated about the vertical axis in 5 degree increments. Sensor effects were significant but not to the degree noted with rotation about the horizontal axis. Again this test highlights the importance of consistency of position in calibration and operation modes. 12

5.0 CONCLUSIONS Tests of the SMC Model 4101-2 combustible gas monitor were conducted in the 306E gas sensor laboratory. The purpose of the tests was to evaluate sensor range, accuracy, and response time, and the effects of ammonia, nitrous oxide, and humidity on the sensor in the presence of hydrogen gas. Tests conducted verified the accuracy and range of the sensor were consistent with the manufacturer's specification of 3% of full scale or f 10% of applied concentration, whichever is greater. The lower detection limit of the SMC sensor was determined to be about 1000 ppm hydrogen or 2.5% LFL. Tests show the sensor begins to respond in -10 seconds after introduce of flammable gases and is capable of reaching 60% of final reading (7070 ppm hydrogen in air) within 20-25 seconds. Calibration of the sensor should be performed under the same conditions as sensor operation. No evidence of short-term sensor poisoning from ammonia, nitrous oxide, or water vapor was observed. Sensor orientation during calibration should match that of its intended operation. 13

To Distribution Project Title/Work Order From Combustible Gas Sensor Interference Test Report, WHC-SD-TRP-249, Rev. 0 WN m67& Waste Tank Process Engineering Text Name MSlN With All Attach. Text Only Page 1 of 1 Date May 24, 1996 EDT NO. dl5-z 7d ECN No. N/A Attach./ Appendix Only EDTlECN Only D. 8. Engelman M. A. Lane W. H. Meader B. J. Webb D. T. Vladimiroff R1-49 S8-05 S8-05 S5-08 S5-08 A-6000-135 (01193) WEfO67 I I '