SpectraPure PUMPED RO SYSTEMS (PSP) User s Manual for PSP-1500 Systems

Similar documents
Light Commercial Reverse Osmosis System. EE-1000 Manual

Installation & Instruction Manual For Portable Series R.O. units:

REVERSE OSMOSIS INSTALLATION Watermark 4 Stage

Pure Water Power Purification System 4-Stage RODI Operations Manual

ULTRA PURE SYSTEMS 400 GPD RODI INSTALLATION AND OPERATION MANUAL

IMPORTANT PLEASE READ BEFORE COMMENCING INSTALLATION

MN-800-TF REVERSE OSMOSIS TREATMENT SYSTEM OWNERS MANUAL

CART FOR CENTURION 1500+

ULTRA PURE SYSTEMS 625 GPD RODI INSTALLATION AND OPERATION MANUAL

Operation Manual. Carbon Filtration Industrial Drive, Orlinda, TN fax

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL

Ultra-High Efficiency RO/DI (U.H.E RO/DI)

TOMAR TSM-E SERIES (400, 800 & 1200 GPD) REVERSE OSMOSIS TREATMENT SYSTEM OWNER S MANUAL

Operating Instructions Model and Hydrostatic Test Pump

TOMAR TLC-SERIES (400, 800 & 1200 GPD) REVERSE OSMOSIS TREATMENT SYSTEM OWNER S MANUAL

1020 Industrial Drive, Orlinda, TN fax

BLS 1900 Optimum REVERSE OSMOSIS SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

SpectraPure. MAXCAP UHE RO/DI System. Ultra-High Efficiency RO/DI INSTALLATION AND OPERATING MANUAL. Optional Accessories Model Optional Part

INSTALLATION, OPERATION AND MAINTENANCE MANUAL ZERO WASTE REVERSE OSMOSIS SYSTEM

Commercial Coffee Lovers System

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

BWS1500. Advanced Membrane Separation System Installation, Operation & Maintenance Manual Rev. 2.1

Reverse Osmosis System Installation Guide and Operation Manual.

Chloramines Removal Filter CLF20 and CLF35 Models

ULTIMATE REVERSE OSMOSIS SYSTEM

Owners Manual Models: 200-CHS-XXXX-XX. US Water Systems Cobalt Hyper-Safe Reverse Osmosis System. Visit us online at

DRS4-RM Manual. Set Up Instructions for DRS4 Series Single Tank

Sterisil Ac+ Installation & Operating Manual

US WATER DENTAL OFFICE PURIFICATION SYSTEM

Pressure Dump Valve Service Kit for Series 3000 Units

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 3035 INSTRUCTION MANUAL

Owner s Manual AMERICAN. LiquaGen 5 Stage (RO/DI) Aquarium Reef System. YEAR Y 1

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

Installation and Operation Guide MRS-100, MRS-225, MRS-100CC, MRS225CC Reverse Osmosis Mineral Reduction System For Foodservice Applications

Commercial Coffee Lovers System

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

The MAXCAP RO/DI System

This Page Intentionally Left Blank 2

ULTIMATE REVERSE OSMOSIS SYSTEM

Installation, Operation & Maintenance Manual for Flo-Max Coupler Bracket Model FM150

UltRo Dual Flow Reverse Osmosis Water System

TOMAR TSM-1500 REVERSE OSMOSIS TREATMENT SYSTEM OWNER S MANUAL

User Instruction Manual

Pressure Dump Valve Service Kit for Series 2300 Units

Installation and Instruction Manual for 6 stage ROCCS Requires Professional Installation

Hydraulic Punch Drivers

OPERATIONS MANUAL. Ultra Pure Systems HydroTube. Reverse Osmosis/Deionizing water production machines designed for water fed pole window cleaning.

4-Stage Drinking Water Systems (MODEL: MPDW, DWS, DWS-PP)

NGP-250/500 Nitrogen Generator Quick Start Guide

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

BRILLIANT WONDERS LED BUBBLERS INSTALLATION INSTRUCTIONS CMP SERIES. Brilliant Wonders LED BUBBLERS

3M Water Filtration Products

Installation, Operation, and Maintenance Manual

ROES-PH75 INSTALLATION INSTRUCTION & OWNER S MANUAL

WatMan WATER RO SYSTEMS FOR DESALINATION

Drinking Water Kit with Non-Air Gap Faucet for RO and RO/DI Systems (Including -MF Systems)

Installation and Instructions 2. Product Features 3-6. Key Pad Functions 7. Distributor Information Programming Guide 8. Master Programming Guide 9-14

This Page Intentionally Left Blank 2

RG1200 Service and Repair Manual

BRILLIANT WONDERS LED BUBBLERS INSTALLATION INSTRUCTIONS & PRODUCT MANUAL CMP SERIES

Softeners Clack Model V125DTH Medical / Industrial Series Operation and Service Manual

INSTALLATION & START-UP INSTRUCTIONS. Clack WS1 & WS1.25 METER WATER SOFTENER SYSTEMS

ESSENCE REVERSE OSMOSIS SYSTEM INSTALLATION INSTRUCTION

Installation and Maintenance Manual. ECO Filtration Unit with 6-way-Top-Mount-Valve. Art. Nr

This Page Intentionally Left Blank 2

Jacuzzi. J-CQ420 Cartridge Filter Installation and Operating Instructions

SpectraPure. The MAXCAP RO/DI System. Reverse Osmosis/Ion Exchange Water Purification System INSTALLATION AND OPERATING MANUAL

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

BWS350. Advanced Membrane Separation System Installation, Operation & Maintenance Manual Rev. 4

Copyright, 2005 GPM Hydraulic Consulting, Inc.

Instruction Manual Updated 7/26/2011 Ver. 2.2

TECHNICAL INFORMATION

Easy Nest Kits. Installation Suggestions

Halsey Taylor Owners Manual STOP!

OPERATION MANUAL NTF-60 Plus

ULTRA REVERSE OSMOSIS SYSTEM

INSTALLATION INSTRUCTION & OWNER S MANUAL

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

Vessel Installation and Operating Manual

Installation Guide & Owner s Manual. Model V4700. Water Filtration for Dental & Medical Instrument Washing Equipment

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

START UP MANUAL MADE IN USA REV US and Foreign patents and patents pending see

OPERATION MANUAL NTF-15

Code AWC20HP Air Compressor

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

Operating Instructions Models and Hydrostatic Test Pumps

KJ4000 Operating Instructions & Parts Manual NSN

AMP Oil Free Manual AMP 50-8-TC AMP 50-6-D AMP General User and Maintenance Instructions

SPECIFICATIONS TABLE Maximum Air Pressure 115 PSI Air Tank Capacity 20 Gallons

Halsey Taylor Owners Manual 4410 Freeze Resistant Tubular Fountain STOP!

NXT 5 Pure Water System

TABLE OF CONTENTS. Purchase Records... Introduction... Component List... Diagram... Specifications... Quick Connect Guide...

4-Stage Drinking Water Systems (MODEL: MPDW, DWS, DWS-PP)

AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: CHAMP 3 PART NO: LS0115

Customer Support Troublshooting

INSTALLATION INSTRUCTIONS

RIX COMPRESSORS. MICROBOOST Owner s Manual CONFIGURATION A MB-115 MB-230. November 21, 2002 APPLICABILITY [ A]

Transcription:

SpectraPure PUMPED RO SYSTEMS (PSP) User s Manual for PSP-1500 Systems

2

3 PSP-1500 SYSTEM DESCRIPTION Reverse Osmosis RO Reverse Osmosis utilizes the unique properties of a semi-permeable membrane to allow fluid to pass while restricting the flow of dissolved ionic material. While pressure is applied to impure water on one side of such membrane material, pure water will pass though, leaving most of the impurities behind. The rejection of the dissolved ionic material is a function of both molecular weight and ionic charge. Most metallic salts are rejected in excess of 98% PSP-1500 System Components RO (Reverse Osmosis) System High/low level floats 10 ft. float control cable 10 ft. 3/8 inch blue product water tubing 10 ft. 3/8 inch black reject (waste water) tubing 8 ft. power cord PSP-1500 System Specifications (115 / 230 volts 60 Hz / 10.4 / 5.2 amps) MODEL Production Rate * (GPD / LPD) Description Element size (inches) Maximum Recovery (%) Motor HP (SinglePhase) Inlet / Outlet Connections PSP-1500-EF-115 1500/ 5677 Dual Electric Float Level Control PSP-1500-BT-115 1500 / 5677 For use with Bladder Tank 4.0 x 40 33 3/4 3/4 pipe 3/8 tube 4.0 x 40 33 3/4 3/4 pipe 3/8 tube

4 PSP-750 SYSTEM DESCRIPTION (continued) PSP-1500 Systems Technical Operating Data 1. Number of Modules ONE (4.0 x 40 ) 2. Type of Membrane Ultra high Rejection TFC Membrane 3. Operating Pressure 110 psig max 4. Operating Design Temperature 40 F to 86 F (4.5 C to 30 C) 5. Membrane Rejection greater than 99% average 6. RO System Recovery 33% 7. Feed flow (GPM) 3.12 GPM at 150 psi, 77 F 8. Reject flow (GPM) 2.08 GPM at 150 psi, 77 F 9. Product flow (GPM) 1.04 GPM at 150 psi, 77 F 10. System Inlet port 3/4 female pipe 11. Membrane Housing Material PVC 12. High Pressure Lines SS 13. Permeate (Product) port 3/8 14. Retentate (Reject) port 3/8 15. RO (Reverse Osmosis) pump Rotary Vane 16. Pump MODEL No. 3306 PH 17. Motor Horsepower 0.75 hp (1725 RPM) at 60 Hz, 0.75 hp (1432 RPM) at 50 Hz 18. Power Requirements Single phase, 60 Hz., 115 VAC, 10.4 Amps Single phase, 60 Hz., 230 VAC, 5.2 Amps. Single phase, 50 Hz., 230 VAC, 5.2 Amps.

5 RO PRETREATMENT RO Feed Water Requirements For a Pumped Reverse Osmosis System to function effectively & yield long membrane life, the feed water MUST meet the following requirements: ph Range: 3-11 Max. Temperature: 110 F (43 C) Max. Feed Turbidity: 1.0 NTU Max. Silt Density Index: 5.0 (based on 15 min. test time) Maximum Chlorine: < 0.1 ppm Maximum TDS: 2000 ppm Maximum Hardness: < 1 grain per gallon (17 ppm as CaCO3) Maximum Iron: < 0.1 ppm Maximum Manganese: < 0.1 ppm Maximum Hydrogen Sulfide: 0 ppm Langlier Saturation Index (LSI): must be negative Minimum Water Pressure 30 PSI Pretreatment RO pre-filtration should be a combination of sediment,carbon filtration.and water softening The sediment filters bring the feed water NTU value to < 1 and the carbon filter further reduce the sediment & turbidity levels in your filters reducing chlorine, chloramines and VOC levels in the water down to acceptable levels. Because of the high recovery of the pumped RO systems we highly recommend the use of a water softener ahead of the reverse osmosis system. This will greatly increase membrane life and reduce maintenance costs of the RO system. (See Figure 1)

6 RO PRETREATMENT (continued) Premeate to Storage Water Lines Drain Lines Electrical Lines Purity Monitor Brine to drain 7 CONTROL BOX M 1 City Water PG PG 4 5 6 level Control Switches Storage tank 2 3a 3b 1 Back flow preventor assembly (Optional) 2 GAC carbon filter 3a Twin Bed water softener (OR Single Bed Softener) 3b Brine Tank 4 Sampling ball valve 5 Shut-Off Valve 6 Sediment Pre-filtration Assembly 7 RO System Figure 1: Water Treatment System Components Sequence & Hook-UP

7 SYSTEM INSTALLATION & START-UP Securely mount RO system to wall. Carbon Filter and Water Softener Installation Follow Figure 1 for sequence and hook-up of components of water purification system. Important Note: On carbon filter or water softener not equipped with system lock-out controls, never operate RO System during carbon filter backwash or water softener regeneration. System lock out control is not necessary when used with twin bed water softeners. Plumbing Connections 1. Carbon Filter & Water Softener feed lines- 3/4 2. Drain lines - 1/2 minimum 3. Install a sampling port and a shut-off valve ahead of the pre-filter (See Figure 1, # 4, 5). 4. Feed water line to prefilter - 3/4 minimum. 5. Prefilter outlet to RO System inlet - 3/4 pipe. 6. Reject water line to drain 3/8 black tubing. 7. Product water line to storage tank 3/8 blue tubing. Electrical Requirements Single Phase, 115 VAC / 60 Hz / 10.4 Amps Singe Phase 220 VAC / 60 Hz / 5.2 Amps Float Switch Set-Up (PSP-1500-EF System) 1. Determine the position of the High and Low float switches in the storage tank, at the levels you want to RO system to turn off and then turn back on again. 2. Drill 5/8 holes at these positions using a 5/8 wood boring bit. Remove any burrs from the holes. 3. Mount the floats from the inside of the tank. Position the rubber gasket on the inside of the tank. (Continued next page.)

8 SYSTEM INSTALLATION & START-UP (continued) Float Switch Set-Up (PSP-1500-EF System) cont. 4. Orient the floats as shown in Figure 2, and secure with the plastic nut on the outside of the tank. The float switches should be in the Horizontal (closed) position... (See Figure 2). When the water level rises above the high float switch the system shuts Service Position Indicator 24 hr. Gear Permeate From RO System High Level Switch with Female Sockets High Level Switch with Male Pins Low Level Level Switch Switch with Female Male Sockets Pins Permeate (Product Water) from RO System STORAGE TANK Model 2510 Manual Backwash Knob SWITCH OPEN SWITCH CLOSED SWITCH CLOSED MINIMUM 8" INCHES HIGH FLOAT LOW FLOAT Figure 2: Float Switch OFF and when water level falls below the low float switch the RO system turns back on. 5. Connect the float control cable to each float and plug the other end of the cable into the 3 terminal connector found on the control box. Red Time Set Button Skipper Wheel (shows every other day regeneration) Figure 3: Carbon Filter Control Valve Figure 4: Solenoid Valve

9 SYSTEM INSTALLATION & START-UP (continued) Water Purification System Start-Up 1. Close shut-off valve ahead of pre-filter. (# 5, Figure 1). 2. Set the carbon filter by-pass valve to by-pass position. 3. Turn the manual regeneration knob (rotate clockwise on click from the service position) in the carbon filter control valve to backwash position. Make sure carbon filter is plugged into a 120 V outlet. Now slowly turn the red by-pass valve handle located at the back of the carbon filter to the service position to bleed air from tank. Do not rapidly open this valve. Leave valve partially open until the air has been bled from the tank. Then fully open the valve. 4. Backwash the carbon filter for at least 15 minutes after air has been bled from the tank. 5. Return the valve back to service position by rotating the manual backwash valve clockwise to the service position.(figure 3) 6. Set time of day on the control valve on the carbon filter by depressing the red button and aligning the 24 hour gear to the correct time. (See Figure 3). 7. Open the shut-off valve ahead of the pre-filter. 8. With system in off position and all plumbing and electrical connections completed, open feed water valve and let pressure build up to set low pressure switch. 9. Check feed water connections for any leaks. 10. Rotate solenoid valve lever counter-clockwise as viewed from top to manually open the inlet solenoid valve. (See Figure 4). Leave the valve open for 5 minutes to purge any air from the system and then shut the valve by turning lever counter clockwise to close the inlet solenoid valve. 11. Turn the power switch on the control panel to the ON position (UP). This will immediately open the inlet solenoid valve. After 1 minute, the RO pump will turn on. This motor protection circuitry assures a flooded inlet to the pump prior to pump startup; thus eliminating any dry starts which could damage the pump. 12. After a few minutes of operation, adjust the pressure relief valve,located beneath pressure gauge, to the pressure listed on system tag by loosening the nut on the threaded shaft and turning the brass handle. When adjusted to this pressure, lock the nut back in place. 13. Check the product water to waste water ratio. It should be at least 1:2. Note: If not, contact the factory (the flow restrictor may require adjustment). 14. Check for leaks and carefully tighten any compression fittings if necessary.

10 SYSTEM OPERATION & MAINTENANCE Sediment pre-filter cartridge Replacement Maintainence Regime: Pre-filter Output pressure < 30 PSI when system is operating. Materials Needed: 20 1.0 micron gradient density cartridge, filter wrench Procedure: a. Turn the power switch on the control panel to the OFF position (DOWN position). b. Close the shut-off valve, prior to the pre-filter. (# 5, Figure 1). c. Manually open inlet solenoid valve on RO System to relieve pressure in the filter housing and then shut solenoid valve. (See previous page: Water Purification System Start Up # 10.) d. Locate the filter wrench for removing the blue canister. e. Using filter wrench, unscrew blue housing from the housing head. Some water may leak out of the canister in the process. f. Throw out the used cartridge. g. Place the new micron cartridge in the canister. h. Screw the blue canister back on. i. Open the shut-off valve. j. Turn the power switch on the control panel to the ON position (UP position). k. Check for leaks. RO Membrane Replacement Maintainence Regime: As Needed (Refer to troubleshooting chart.) Materials Needed: Replacement membrane, pliers. Figure 5: White Retaining Ring Figure 6: Pulling End Plug

11 SYSTEM OPERATION & MAINTENANCE (continued) Membrane Replacement Procedure: a Turn the power switch on the control panel to the OFF position (DOWN position). b Close the shut-off valve. (# 5, Figure 1). c Remove white retaining ring by holding it with your left hand, grip the pull tab with thumb and forefinger, and pull ring towards center of membrane. Now pull ring straight out. (See Figure 5). Grasp the threaded port with pliers, and pull bottom end plug straight out. (See Figure 6). d Pull the membrane from the housing with pliers. (See Figure 7). Insert new membrane into the housing. Black rubber brine seal must remain at the open end. (See Figure 8) Push the membrane into the housing until it bottoms out. e Replace O-ring (See Figure 9) on end plug and firmly push plug into membrane housing to seat it. f Replace white retaining ring in groove. g Open the shut-off valve. (# 5, Figure 1). h Turn the power switch on the control panel to ON position (UP position). i Divert the product water to a drain and operate the RO system for 1/2 hour at 150 psi and check for leaks. Tighten fittings if necessary. j RO water can now be sent to storage tank for safe use. Figure 7: Pulling Membrane Figure 8: Replacing Membrane Figure 9: Replacing O-ring

12 TROUBLESHOOTING GUIDE This troubleshooting guide is to assist you in identifying the most common operating problems you may experience with your water purification system. PROBLEM PROBABLE CAUSE CORRECTIVE ACTION Low system inlet pressure. Insufficient feed water pressure or flow Clogged pre-filter Check feed pressure, open feedwater valve, check for restrictions Replace pre-filter cartridge. Closed valve ahead of RO System Check for closed valves. Low system operating pressure. Pump not operating correctly Repair or Replace pump. Defective pressure regulator valve Replace or repair valve. Low permeate flow rate Low operating pressure See above possible causes for low operating pressure. Dirty or fouled membrane Replace it. Excessive Permeate flow rate. Operating on cold water Defective product water O-rings. Damaged membrane (possibly due to chlorine damage) Ruptured membrane If possible operate with a feedwater temperature of 72-80 F (22-27 C). Check product water tube O-rings and replace if necessary. Replace membrane. Replace membrane. Water flowing when system is turned off RO System cycles ON and OFF. Inlet solenoid valve not closing or seating properly Insufficient system inlet pressure. Clean valve or replace. See above possible causes for low system inlet pressure.

13 TROUBLESHOOTING GUIDE (continued) PROBLEM PROBABLE CAUSE CORRECTIVE ACTION High product water conductivity (declining rejection) Dirty or fouled membranes Product water tube O-ring seal broken or damaged Membrane put in backwards Change in incoming water quality Replace membrane. Replace O-ring, check sealing surfaces on O-ring groove, interconnectors and end caps. Replace damaged parts. Install membrane correctly. Power switch on, unit not operating Float switch has cut power to machine Defective Float Thermal overload in motor No power to system Motor and/or pump not operating properly Micro-switch on either carbon filter or water softener not functioning properly Carbon filter or water softener control valve not in service position Low system inlet pressure Check position of float in the storage tank. Replace Float Allow motor to cool and re-start system. Check amp draw of motor. Check fuses or circuit breakers, measue voltage. Repair or Replace pump/ motor. Replace micro-switch. Set the control valves to the service position. See previous page for possible causes for low system inlet pressure.