Good Pipetting Practice

Similar documents
Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

DISCOVERY PRO - features

Improving Pipetting Techniques

High-Throughput. Faster, Better 96 & 384-Well Pipetting Ultra Efficient and Extremely Versatile

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

MULTICHANNEL. Labnet International, Inc.

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

Qi Litre Ultrasonic Cleaner

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

Transferpette. Single and Multichannel Pipettes

Chapter 2. Pipetting techniques. Guide to Pipetting

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Pipette Tips. Purity and Performance Rainin BioClean Ultra Tips Deliver

Pipettor. User Manual

Adjust the Volume Display

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Transferpette. Lightweight precise rugged reliable!

RayPette Plus Autoclavable Pipette. User Manual

Transferpette. Lightweight precise rugged reliable!

Operating Instructions. Rainin Classic Manual Pipette

The pipettes cover a volume range from 0.1µl to 10ml.

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

Transferpette -8/-12

A N F G J I C D D L A B C D E F

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

Liquidator 96 Manual benchtop pipetting system

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Proline Plus Mechanical Pipette

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

E-Man Hybrid. Manual pipette with real-time display. Continuously- Adjustable Microliter Pipette with LCD. volume ranges from 0.

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Other languages available on our website:

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

For accurate measurements and to prevent damage to the micropipets, follow these important guidelines:

Ergonomic Assessments of the Melni Splice Connector and the Traditional Cable Splice Method

User s Manual. Autoclavable & UV resistant. Multi-channel digital micro pipette. In Vitro Medical Diagnostic Devices (98/79/EC)

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

IE 366. Chapter 14. Handtools

New Technological Guidelines Required for Pipette Management (Proposal of a Pipette Management Method)

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

A N B P F G J I C D D L A B C D E F

Pipetting Small Volumes

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

User s Manual. Autoclavable & UV r e s i s t a n t. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

How to Measure R7.1. Reference. I. Linear dimensions

Real World Metric System

Technical Procedure for General Laboratory Equipment

A N B P G J I C D L D E

Gel Loading Practice MiniLab

Register your instrument! Eppendorf Research plus. User Adjustment

GUIDELINE for MSD Hazard Control Measures

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

VITLAB micropipette. Standard Operating Procedure (SOP)

MSD Risk Assessment Checklist

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

WorkWell Consulting Ergonomic Risk Identification & Assessment. Job Title: Equipment Operator III (Sweeper) Elgin Ford #2108 Date: September 17, 2003

Thermo Scientific Matrix MultiChannel Equalizer Pipette with Expandable, Equal Tip Spacing. User Manual

SoCal BioMed, LLC 4695 MacArthur Court, 11th Floor Newport Beach, CA Phone: +1 (424) Fax: +1 (323)

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

Pipetting and Determining Protein Concentration

ali-q TM Gravimetric Verification Procedure

Manual Handling Procedure

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Operating Instructions. Liquidator 96 Manual benchtop pipetting system

MOLEBIO LAB #1: Microquantity Measurement

User's Manual. Autoclavable Nichipet EX Plus Digital micro pipette for liquid handling CERTIFIED ISO9001

Bioscience Laboratory Ergonomic Check List

MVS Volume Verification Using Any Microtiter Plate or Small Volume Container, Such as a Tube or Vial

Air Displacement Pipetting Module. Total Integrated Liquid Handling

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

CALIBRATION AND ADJUSTMENT

Chapter 13: Manual Handling

PIPETMAN F ENGLISH. Other languages available on our website:

SOP-111 Pipetting Small Volumes

Nichipet EX Plus II. User s Manual. Autoclavable & UV resistant. Digital micro pipette for liquid handling

CHM250 Calibration and Measurement Lab. Balance Calibration

User s Manual. Autoclavable & UV resistant Nichipet EX II. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

Softball Pitching Mechanics

Standard Operating Procedure. Air Displacement Pipette Calibration

ASSESSING AND CONTROLLING MANUAL HANDLING RISK

INTERNATIONAL STANDARD

Phone: Fax: More information:

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

Corporate Standard Manual Handling HB#

CanNeed-DSST-100 Digital Secure Seal Tester --- To evaluate the secure seal quality of PET bottles, glass bottles and cans

Using PV Diagram Synchronized With the Valve Functioning to Increase the Efficiency on the Reciprocating Hermetic Compressors

ISO INTERNATIONAL STANDARD. Ergonomics Manual handling Part 1: Lifting and carrying

Air screwdrivers CSEZ

Feature Picture Example Description Use Task Examples

List 4 Example Of How Manual Handling Risks Can Be Reduced

Transcription:

Good Pipetting Practice The Ergonomics of Pipetting Forces are Affected by Pipette Selection The process of pipetting is well understood by scientists around the world. Pipettes are ubiquitous, providing a simple mechanism to complete workflow requirements, from the starting liquid sample preparation to introducing the sample into a final analysis system. Using many different pipettes throughout the day is common and each unit may have its own unique challenges high volume, low volume or multichannel. The ergonomic factors associated with a variety of pipettes and applications is an important consideration poor ergonomics can lead to fatigue, pain and the risk of injuries. Several critical areas have been identified as key to degrading the comfort level of the pipette and subsequent development of various painful ailments, generally called Repetitive Stress Injuries (RSIs). This information has been published by ergonomics scientists in an effort to create awareness of the usability issues associated with pipetting. 1, 3, 4, 5, 6, 7 A common theme among all studies is that pipetting is a forceful and repetitive activity, and that there is a strong association between pipetting and developing RSIs. In fact, pipetting just over an hour a day over the course of a year is enough to put you at risk, and the chances increase exponentially with workload and 1, 3, 4, 7 age. A key reference point is the amount of force required to perform an activity, such as depressing the plunger or ejecting tips. Most of these forces are transmitted through the thumb. Generally speaking, the goal of pipette manufacturers has been to lower pipetting forces as much as possible in order to improve their ergonomic impact. Part of the GPP - Good Pipetting Practice Series

Tip loading and tip ejection Loading a tip on the end of a pipette and subsequently ejecting it at the end of the pipetting cycle are very closely linked and constitute some of the heaviest forces associated with pipetting. With most pipetting systems the design of the distal end of the pipette is conical in shape. The process of loading the tip on the pipette requires a certain amount of force to ensure an adequate seal of the system too little force and there is a risk of leakage and volumetric error. User technique for tip loading can vary as there is no clear way to know when enough force had been used to seal the system. Typically a user will adopt a technique that provides reasonable assurance that the tip has sealed and will stay with that technique for most of their working life. Unfortunately, for traditional conical tip systems, there is a strong correlation between the amount of force needed to load the tip and the amount of force required to eject it, so for those who adopt a technique that uses high force to add the tip, a high force will be required to eject it (see Figure 1, red line). Tip Ejection Forces at Increasing Tip Loading Forces Brand X Traditional Tip vs. Rainin LTS Ejection Force (kg) 4.0 3.0 2.0 1.0 0.0 0.0 5.0 10.0 15.0 Insertion Force (kg) Brand X Tip, 250 µl Rainin LTS Tip, 250 µl Figure 1 The relationship between tip loading and ejection forces a traditional conical tip compared with a Rainin LTS tip. In the paper by Lichty et al, 6 a number of different manufacturer s pipettes are reviewed and force measurements compiled for plunger forces and tip ejection forces. It is clear that tip ejection forces are the most significant force associated with the many steps involved in pipetting, but there is great variation in tip ejection forces between pipettes (See Figure 2). 2

Rainin Pipet-Lite (LTS) 2.1 kg - Female 3.O kg - Male Manuf. 3 s Manuf. 2 s Manuf. 4 s 0 1 2 3 4 5 kg Force Data from Work, 39, (2011), 177-185. Lichty et al Figure 2 Comparison of tip ejection forces from four popular pipettes. The maximum recommended thumb exertion force is 2.1 kgf for females and 3.0 kgf for males. 2 Rainin solved this problem with LiteTouch System (LTS) of pipettes and tips (See Figure 3). Here both the distal end of the pipette and the inside of the tip have a closely matched cylindrical surface. LTS tips also have a sealing system close to the entry point and a hard stop end point inside the tip. The end stop prevents the shaft from being pushed further into the tip with increased force. This strictly limits the maximum sealing force, while still guaranteeing that there will be no leakage. A strict limit on sealing force ensures a minimal ejection force (See Figure 1). This unique yet simple design provides a tool that significantly reduces one of the major sources of muscle fatigue and causes of RSI for pipettors. Traditional Large seal Conical Friction fit LTS Small seal Cylindrical Positive Stop Figure 3 Mechanics of Rainin LTS tip sealing system compared to traditional conical tip sealing systems 3

Plunger forces The plunger system of manual pipettes constitutes the second area of ergonomic interest to the user where repetitive tasks can influence the outcome of RSI s. After tip insertion, pressing the plunger down to the first stop against the stroke spring (See Figure 4) prepares the pipette/tip system for aspiration. A clear distinctive first stop is critical in that it defines the distance needed to move the plunger and aspirate the desired volume. Home Position First Stop Second Stop Figure 4 Pressing manual pipette plunger to show home position (start dispense), first stop (full dispense) and second stop (blowout) Mechanically, this first stop is created by the plunger hitting a stiffer pre-loaded blowout spring that has greater resistance than the stroke spring. After the plunger is pressed down to the first stop, the tip is inserted into the solution and the plunger slowly released to aspirate. The stroke spring performs the aspiration by pushing the plunger back up to the home position. The sample is dispensed at the last step pushing the plunger down against the force of the stroke spring, through the first stop position and then against the force of the blow-out spring to remove all of the liquid. With all liquid dispensed, the plunger is released and returns back to the home position. In the paper by Lichty, 6 blow-out forces are the second largest force in the process of pipetting (after tip ejection forces) and stroke spring forces the smallest. 4

An important aspect of the forces in this process is the internal piston seal system (Figure 5). The seal produces a partial vacuum that enables liquid to be drawn into the tip as the piston moves up the shaft. The seal system must be tight and strong enough to hold the vacuum required for the entire operating range of the pipette. If the seal is too tight the force required to push the plunger down and then return it to the home position will be excessive. If the seal is relatively loose, the forces required to move the plunger back and forth will be less, but the potential for leakage increases. Piston Piston spring Seal retainer Seal O-ring Figure 5 Sealing system in manual pipettes For manufacturers of pipettes there is a balance between the seal design and the stroke spring forces that must be considered in order to meet both the technical needs for accurate pipetting and the individual user s need for an ergonomic solution. This challenge is met differently by various manufacturers, resulting in variable plunger forces (See Figure 6). Manuf. 4 s Manuf. 3 s Manuf. 2 s Rainin Pipet-Lite* (LTS) 0 1 2 3 4 5 6 7 kg Force Depress Hold Aspirate Dispense and Blowout Figure 6 Comparison of plunger forces from four popular pipettes. The maximum recommended thumb exertion force is 2.1 kgf for females and 3.0 kgf for males 2 *New Pipet-Lite XLS reduces plunger forces by up to 43% 5

Rainin developed high precision, low friction seals to help reduce stroke spring plunger forces. These seals require minimal force to move while maintaining a vacuum, which allows for a lighter stroke spring. A lighter stroke spring allows for a lighter blow-out spring, because the difference in spring forces defines the first stop. Thus blow-out forces are reduced as well. Ultimately, the best solution for avoiding plunger forces is to use an electronic pipette, where a microprocessor-controlled motor moves the plunger, resulting in near zero plunger forces. This can be particularly helpful with multichannel pipettes, which have the highest plunger forces, and when a researcher needs to perform intensive repetitive mixing of the liquid samples. Volume setting changing the aspiration volume Changing the volume of a manual pipette is often necessary to meet the requirements of the experiment (e.g., changing the setting of a 200 µl pipette down to 50 µl for specific buffer addition, and then back up to 200 µl for the next buffer). The process of changing the volume setting screw normally a small diameter button requires several turns of the hand-fingersthumb combination. Experienced pipettors may even do this with one hand, both holding and adjusting the pipette at the same time, while holding a liquid sample in the other hand. A study by Asundi et al 5 indicates that volume adjustment is an activity that requires high muscle activity. In the four different thumb-related muscles that were tested, volume changes generated significant muscular response, indicating a potential for fatigue. Different pipette designs can either reduce or increase the forces required for changing volume. A further study by Lichty et al 6 found a positive correlation between ease of volume adjustment and worker productivity, probably because significant time is spent adjusting the volume setting during routine operation. Rainin reduced the forces associated with volume adjustment by providing a large easy-grip button and low-friction screw mechanism. An adjustable lock prevents the inadvertent changing of the volume setting between pipetting steps. With this combination an operator can easily unlock, adjust and lock the volume setting mechanism with one hand and with minimal force requirements, (see Figure 7). Figure 7 Setting the pipette volume with one hand 6

Repetition Pipetting once will not affect your hand, but pipetting hundreds of times over the course of several hours certainly will. The chance of getting injured increases greatly with repetitive work. Fortunately there are several straight-forward solutions. The first kind of solution is to use products that do the work for you. Multichannel pipettes and 96-well pipetting platforms reduce repetition by pipetting multiple samples at once. Electronic pipettes can semi-automate certain functions, such as dispensing multiple aliquots from a single aspiration, automatically mixing, or automatically changing the programmed volumes in a pre-determined sequence. The second kind of solution is to modify pipetting habits to decrease the effects of repetitive work. Try to take regular breaks, switch hands, stretch your limbs and maintain good posture (minimize twisting of your shoulders, arms and wrists; manipulate objects at shoulder height or below). Grip Grip comfort has been positively associated with reducing hand and arm discomfort 6. Factors important for grip comfort include the shape of the pipette body and fingerhook, and the distance your thumb needs to reach to the top of the plunger. Try to grasp the pipette with as little force as necessary to maintain control. Conclusion and Recommendations The ergonomics of pipetting are generally well understood. Higher pipetting forces and long hours of pipetting are associated with developing RSIs. The following recommendations will assist researchers in minimizing their chance of developing RSIs. Look for lower plunger forces Check spring forces associated with the plunger. Compare new pipettes with old ones. Use electronic pipettes when possible. Service your pipettes regularly to maintain optimal mechanics. Keeping the piston and seals clean will both lower friction, thus reducing pipetting force, and maintain better accuracy and precision. Look for lower tip ejection forces Compare the tip ejection forces of different pipettes. Consider that a pipette with a high tip loading force will probably have a high tip ejection force. Remember that multichannel pipettes have generally higher forces and can be particularly hard on the thumb. Other considerations Use pipettes with a finger hook that fit comfortably in your hand without gripping tightly. Lower the forces associated with volume change by using a pipette with a volume lock, low-friction volume change mechanism and a large dial with good grip. Reduce repetitive pipetting by using electronic pipettes, multichannels and 96-well pipetting platforms. Take breaks and, if possible, switch hands and stretch arms and hands. 7

References 1 Björksten MG, Almby B, Jansson ES. Hand and shoulder ailments among laboratory technicians using modern plunger-operated pipettes. Appl Ergon. 1994 Apr;25(2):88-94. 2 Kroemer KH. Cumulative trauma disorders: their recognition and ergonomics measures to avoid them. Appl Ergon. 1989 Dec;20(4):274-80. 3 McGlothlin JD, Hales TR. Health Hazard Evaluation Report at Scientific Application International Corporation, Frederick, Maryland. Cincinnati, OH: U.S. Department of Health and Human Services, Public Health Service, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health. HETA Report No. 95-0294- 2594. (August 1996.) 4 Fredriksson K. Laboratory work with automatic pipettes: a study on how pipetting affects the thumb. Ergonomics. 1995 May;38(5):1067-73. 5 Asundi KR, Bach JM, Rempel DM. Thumb force and muscle loads are influenced by the design of a mechanical pipette and by pipetting tasks. Hum Factors. 2005 Spring;47(1):67-76. 6 Lichty MG, Janowitz IL, Rempel DM. Ergonomic evaluation of ten single-channel pipettes. Work. 2011;39(2):177-85. 7 David G, Buckle P. A questionnaire survey of the ergonomic problems associated with pipettes and their usage with specific reference to work-related upper limb disorders. Appl Ergon. 1997 Aug;28(4):257-62. Rainin Instrument, LLC 7500 Edgewater Drive, Oakland, CA 94621 Phone +1 510 564 1600 www.mt.com/rainin For more information a METTLER TOLEDO Company Subject to technical changes 02/2012 Rainin Instrument, LLC Printed in USA WP-504 Rev A 55% recycled fiber, 30% PCW