Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

Similar documents
VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

STANDARD OPERATING PROCEDURES

School of Chemistry SOP For Operation Of Glove Boxes

Schlenk Line. Standard Operating Procedure. Lab: Engineering Sciences Building 155. Department: Materials Science and Engineering. PI: Paul V.

Solvent System Walkthrough

Figure 1. Basic vacuum/inert gas manifold (Adapted from chemistryviews.org)

EHS Laboratory Guidance: Cryogenic Material

ARC Unofficial Manual to the Solvent Dispensing System. Designed by J. C. Meyer

MB-SP Series Solvent Purification System

1.07 Experimental Methods and Techniques: Basic Techniques

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

How to use Solvent Purification System A brief manual

Chemical Class Standard Operating Procedures. Pyrophoric Chemicals (PYR) & Water Reactive Chemicats (WRC) Diethylzinc

Technical Bulletin AL-134

General Safety Rules

Digital Vacuum Regulator

School of Plant Biology - Liquid Nitrogen Induction

Solvent Dispensing System Manual. Designed by J. C. Meyer

Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases.

[USING THE VITROBOT April 9, Using the Vitrobot

Equipment Operating Procedure Glove Box

CRYOGENIC LIQUIDS AND THE SCIENTIFIC GLASSBLOWER. Michael D. Wheeler, Arizona State University Chair, A.S.G.S. Safety and Hazards Committee

DRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35

Each cryogenic liquid has its own specific properties, but most cryogenic liquids can be placed into one of three groups:

Working with Cryogenic Liquids

Laboratory Standard Operating Procedure: Cryogenic Liquids

Instruction Manual Updated 7/26/2011 Ver. 2.2

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

General Precautions for handling compressed gases

GUIDELINES FOR CONDUCTING HYDROGENATION REACTIONS

TEMPERATURE S RELATIONSHIP TO GAS & VAPOR PRESSURE

DUQUESNE UNIVERSITY. Liquid Nitrogen Safety

Dilution Refrigerator manual

Hydrogen Cyanide (HCN)

POWER FLUSH KIT INSTRUCTIONS

CVI Valve Line. Exceeds the industry s highest standards for reliability and performance

THERMAL EVAPORATION UNIT (for Al evaporation)

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

Operating Instructions

Manual Solvent Purification System. Innovative Technology, Inc.

The Safe Use of Pyrophoric Substances

Standard Operating Procedure

Physics Experiment 17 Ideal Gas Law Qualitative Study

Research Laboratory Safety Self-Inspection: Chemical & Physical Safety Checklists

Measuring Carbon Dioxide in Breath

STANDARD OPERATING PROCEDURES

Department of Physics Standard Operating Procedures (SOP) Approved by: Physics Safety Committee

Lab-Specific Standard Operating Procedure (LSOP)- Pyrophoric Chemicals

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Section 3- Repair and Calibration Procedures

LABORATORY SAFETY. A general overview

Instructions for completing the Unit Operations Lab safety inspections using the Safety Inspection Checklist

Inert Air (N2) Systems Manual

ZUERCHER TANK REGULATORS PRINCIPLES OF TANK VENTILATION

Cryogenic Materials Operating Procedure and Safety Requirements

D R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00

University of MN, Minnesota Nano Center Standard Operating Procedure

MMRC PVD Evaporator Instructions

General Chemistry I Percent Yield of Hydrogen Gas From Magnesium and HCl

CRYOGENIC LIQUID HANDLING STANDARD

B. Assemble the Method-25 sample train with a heated probe/filter.

Rejuvenation Instructions

Research Teaching Production

Operating Procedures for Metal Evaporator I

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Rev. 1. Cryogenics Thermoflasks. 2122, 2123, 2124, 2129, and 2130

LIQUID OXYGEN. General. Health Effects. Safety Considerations

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

9.0 Equipment Cleaning and Repair

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

THE UNIVERISTY OF NEWCASTLE - SCHOOL of BIOMEDICAL SCIENCES

Alkylphosphines. Storage and Handling Recommendations

Both physical and chemical effects come into play either separately or in combination.

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Experiment 18 Properties of Gases

Portable Oil Lube Air Compressors

How To Use Getters and Getter Pumps

Tank Blanketing. Application Jordan Valve Product Offering Competition

PURDUE UNIVERSITY TEFLON CORE PCB THRU-HOLE PREPARATION FOR BLACK HOLE APPLICATION RF LAB. Professor in charge: William Chappell MSEE 289

MV5: Automated concentration system

Compressed Gases and Cryogens

Vapor Pressure of Liquids

Standard Operating Procedure (SOP)

Photolithography. Operating Instructions

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

Ammonia Treatment Instructions Page 1 of 5

Figure. Experimental set up of catalyst, reactants and product syringes.

MS.RAJA ELGADY/PRESSURE PAPER 3

Focus on VOC Emissions Reduction Using an Oxygen Based Inerting Control System For Inert Gas Blanketing of Chemical Process Vessels

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

CHAPTER 26 WELDING AND CUTTING

Cell Disruption System

Oxygen Contamination

Safe handling of cryogenic liquids

Norton Group Toepler Pump Operating Instructions. Version 3.1

Laboratory Personal Protective Equipment & Hazard Assessment Tool

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

1. The Safe Use of Pyrophoric Reagents

Welcome to the Specialized Medical Services Respiratory training webinar series!

ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS

Transcription:

Standard Operating Procedure Inert Vacuum Line Ryan Mulvenna, May 2012 Description of Process The inert vacuum lines located in the fume hoods one (Conjugated Polymer Synthesis: CPS) and two (ATRP polymer synthesis: ATRPPS) (Location FRNY 2182). These pieces of equipment are used to inert reaction vessels with nitrogen (N 2 ) or Argon (Ar) via flame dry purge/refilling, freeze-pump-thaw or vacuum distillation techniques. The flame dry purge/refilling technique will inert the atmosphere in the vessel. During evacuation, running the butane flame gun over the glass aids in desorption of oxygen (O 2 ) as well as water (H 2 O) from the interior surface of the glass. The freeze-pump-thaw technique is often used after the flame dry purge/refilling technique to remove oxygen from injected liquids into the vessel prior to reaction. The vacuum distillation technique is used for separating or transferring a liquid or solid from a more volatile liquid or solvent. This technique requires the use of the liquid nitrogen trap (LN 2 ) to condense vacuumed vapors to prevent cavitation in the vacuum pump. Required Training Purdue Department of Chemical Engineering Safety Orientation P.O.W.E.R. Lab safety orientation (including fume hood induction) Engineering and Ventilation Controls The inert vacuum lines are located inside of fume hood one (conjugated polymer synthesis) and two (ATRP polymer synthesis) (Location FRNY 2182). Ensure that while working in the fume hoods (regardless of whether it contains a vacuum line), the sash must be at least at the minimum working height during experiments! This minimizes exposure to the chemicals as well as projectiles from explosion/implosion of vacuum line vessel in case of breach. Personal Protective Equipment (while using fume hood/vacuum line) EYE PROTECTION: Safety glasses or goggles (mandatory!) or face shield, if desired PROTECTIVE CLOTHING: Laboratory coat and gloves (nitrile/neoprene) (mandatory!) Thermal insulated gloves for freeze 1

Use of Vacuum Line (for flame dry/evacuating/refill) 1. Ensure that the vacuum pump is on (switch on back) and the regulator supply pressure to the desired fume hood is 0.5 bar at all times (refer to Appendix 2 & 3). 2. If you wish to use the vacuum, ensure that it is not currently in use and its pressure is 0.3-0.5 torr (refer to Appendix 4). Ensure that during experiment AS WELL AS setup, sash is at or below the minimum safe working height! 3. Ensure that the inert gas line is working properly by attaching a half cut syringe with sheathed needle into a vacant line. Ensure first that ALL other gas/vacuum supply valves on the manifold are closed (close FINGER TIGHT CLOCKWISE). Unsheathe the needle and open the BOTTOM valve on your connected syringe line and to test the inert gas flow. If you can feel gas flow from the needle 2-3cm away from your hand, the gas flow is sufficient. If you cannot feel the gas flow, check again to make all other gas/vacuum supply valves in the manifold are closed. If unsuccessful, adjust the inert gas supply line regulator (in the fume hood) to get flow. If this does not work, CLOSE the open gas supply valve AND the inert gas supply line regulator (in the fume hood) FIRST BEFORE you iteratively adjust the regulator supply pressure to the desired fume hood to 0.5 bar to get sufficient gas flow. 4. Connect the reaction flask using thick walled red rubber tubing to a vacant port. Once again, ensure that during experiment AS WELL AS setup, sash is at or below the minimum safe working height! (refer to Appendix 1 & 4). 5. To use the vacuum, ensure that all connections are on tight (use Keck clamps/rubber septums/vacuum greased stopcocks where required). Ensure first that ALL other gas/vacuum supply valves on the manifold are closed (close FINGER TIGHT CLOCKWISE). Open the TOP valve on your connected port to evacuate your reaction vessel. Note: Vacuum line pressure should steadily decrease to below 0.5 torr in a few minutes (approx. <0.1 torr after 5-10 minutes) depending on vessel size. If this does not happen, check your connections and try again. 6. To use the inert gas, ensure that all connections are on tight (use Keck clamps/rubber septums/vacuum greased stopcocks where required). Ensure first that ALL other gas/vacuum supply valves on the manifold are closed (close FINGER TIGHT CLOCKWISE). Open the BOTTOM valve (COUNTERCLOCKWISE ~90 on your connected port to fill your reaction vessel. If you are refilling a vessel, only open for 2-3 seconds! You MUST have a relief if you are purging the vessel. 2

7. While performing steps 5 of 6 (2-3 times), ensure that the line that you are using is isolated from all other reactions using the vacuum line. 8. To use a hotplate for your reaction, immerse the vessel in the silicone oil bath (operating temperatures <200 C). Temperature control can be achieved in one of two different ways. The first method is without the thermocouple. Upon removal of the temperature thermocouple AND reinsertion of the attached grey plug, temperature control is achieved via adjusting the temperature dial to the desired value via the LED display. The second method is with the thermocouple. Insert the thermocouple into the oil bath/medium you wish to heat. Ensuring that the thermocouple is plugged into the hotplate, adjust the thermocouple LED display to the desired temperature (not on the hotplate!). After the desired value has been set, adjust the temperature dial on the hotplate from 250 C to 500 C. The temperature dial on the hotplate is the proportional heat control for the thermocouple controlled heating. 9. After evacuation/refill procedure is complete, ensure that you write your reaction and reactions conditions with contact details on the sash using a marker. Use of Vacuum Line (Freeze-Pump-Thaw Technique) 1. Ensure that the vacuum pump is on (switch on back) and the regulator supply pressure to the desired fume hood is 0.5 bar at all times (refer to Appendix 2 & 3). 2. Follow the previous procedure outlined in Use of Vacuum Line (for flame dry/evacuating/refill) section to prepare the vessel for solid/liquid insertion. 3. For solid insertion, break the seal under an inert gas blanket flow and add. Reseal and repeat the Use of Vacuum Line (for flame dry/evacuating/refill) outlined above. 4. For liquid insertion, inject using a needle under a positive inert gas pressure using a separate syringe needle for pressure relief. Upon completion, remove the relief needle and close the inert supply stopcock to the vessel 5. Wearing thermal gloves, immerse the vessel in the fluid container and leave for five minutes. During this time, ensure that the LN 2 level does not fall below the liquid that you are to freeze. After 5 minutes or when the liquid is observed to be completely 3

frozen, open the TOP valve (COUNTERCLOCKWISE ~90 on your connected port to evacuate your reaction vessel and the vessel stopcock to vacuum. 6. Upon the LED vacuum gauge displaying less than 0.3 torr, CLOSE the vacuum and isolate your reaction vessel from vacuum. Lift the vessel from the LN 2 bath and immerse in warm water. Do not stir contents during thaw cycle. 7. Upon melting of contents in the vessel and ceasing of bubble formation, repeat steps five though seven two to three more times. Repeating this cycle multiple times will remove oxygen from the reaction vessel. Use of Vacuum Line (Vacuum Distillation Technique) 1. Ensure that the vacuum pump is on (switch on back) and the regulator supply pressure to the desired fume hood is 0.5 bar at all times (refer to Appendix 2 & 3). 2. Wearing thermal gloves, remove and fill the liquid nitrogen vacuum trap to approx. ¾ full of LN 2 (refer to Appendix 4). During this time, ensure that the LN 2 level does not fall below the liquid that you are to freeze. (Each complete fill of the liquid nitrogen vacuum last from approx. 9-12 hours. If you are using the vacuum line where liquids will be entrained for longer periods of time, you will need to organize to have the liquid nitrogen vacuum trap refilled as long as it is in active use). 3. After waiting for approx. ten minutes, prepare the vacuum distill head for vacuum distillation, as well as multiple appropriately sized round bottom flasks (r.b.f.). Connect the liquid r.b.f. to be distilled to the evaporator and the collection flask to the side arm. Ensure connection is secure at all times. 4. Ensure the condenser is off before securing the lines to the side arm of the vacuum head. Ensure to use ties to properly seal both the condenser and vacuum line. Vacuum grease the thermometer and hold into the vacuum head. 5. Quickly open and close the vacuum line and refill from the inert line to ensure distillation occurs under low concentrations of oxygen. 4

6. If under the presence of heat, immerse the liquid material into thermostat controlled oil bath while under vacuum to distil the required components at the required temperatures/pressures. Appendices Minimum safe working height of fume hood sash (~30cm) ATRPPS fume hood CPS vacuum pump ATRPPS vacuum pump CPS fume hood Appendix 1: Fume hoods containing vacuum lines in FRNY 2182. 5

Reserve tank ATRPPS vacuum line N 2/Ar supply tank CPS vacuum line N 2/Ar supply tank N2 purge-dry tank Tank pressure indicator (bar/psi) Regulator supply pressure (bar/psi) Tank open/ close valve Regulator supply on/off needle valve Supply pressure adjustment valve (bar/psi) Appendix 3: N 2 /Ar regulator for vacuum lines. 6

Vacuum supply line pressure (torr) Vacuum line to vacuum pump Inert gas supply line regulator (L/min) Liquid nitrogen vacuum trap Vacuum supply valves (along top to each port) Vacuum supply line (to each port) Inert gas supply line (to each port) Inert gas supply valves (along bottom to each port) Vacuum/inert gas port s (to reaction vessel) Appendix 4: Vacuum line setup. 7