MANUAL GAS MANIFOLDS

Similar documents
AUTOMATIC GAS MANIFOLDS

DUPLEX PRESSURE REDUCING SET

CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

AREA VALVE SERVICE UNITS

PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL

Specialists in HTM Medical Gas Pipeline Equipment AVSU/Module Operating and Maintenance Instructions

Manifolds & pressure reducing sets

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

Tri-Tech Medical Inc.

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Installation & Maintenance Manual. Economy Manifold

Economy Manifold Alert -2. Installation and Maintenance Manual V1.6

PureGAS Manifold Control Panel - HTM02-01

Installation & Maintenance Manual. C2 & C3 Manifolds

SERIES-O REGULATOR. Instructions for use

Pressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions

Pressure Regulators. Instructions for Use

Dialflow Regulator. Instructions for Use

Pressure Regulator. Instructions for Use

Technical Specification

System Pressure Manager Standard & System Pressure Manager Plus

Manual Actuated Boiler Blowdown Valves

Protocol Switchover Station

Budget Range Operators Handbook

MEDICAL EQUIPMENT CATALOG

Peninsula Dental Social Enterprise (PDSE)

SEMEM3-73 Maintaining medical gas pipeline systems and equipment

Medical Gas Control Panel

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES

Oxygen Dialflow Meter. Instructions for Use

LARGE DRY CHEMICAL, TWIN-AGENT, AFFF, HALON 1211, AND AQUASONIC (NITROGEN-OPERATED SYSTEMS) Pressure Regulator Test Manual

MEDICAL GAS CYLINDERS SELECTION AND MAINTENANCE OF SEALS USED ON HIGH PRESSURE CYLINDERS

Fully Automatic Gas Manifold. Patent US 7,013,906. Installation & Operation Manual

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

User manual. ZX trolley.

AUTOMATIC GAS MANIFOLD

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Unimatic II Manifolds for Oxygen, Nitrous Oxide & Nitrogen Service Manual

Installation and Maintenance Manual. Compact Medical Gas Outlets

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Flow meter. bellow vaporizer. APL valve. Scavenging system

EDUCATION DEPARTMENT ACCU-TEST

Zeus Area Valve Service Unit (AVSU) and AVSU Module. Installation, Operation and Maintenance Manual

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

Installation & Operating Instructions for Genesys Series Manifolds

Oxygen Dialflow Meter. Instructions for Use

BOC Healthcare Training

CS150 CAP STAPLER OWNER S MANUAL

632 Series Gas Distribution System

SAFETY MANUAL FOR FLAMMABLE PRODUCT TRANSFER

Installation & Operating Instructions for Genesys Analog Series Manifolds

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units Digital Range

PRODUCT OPERATING MANUAL

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

Installation and Maintenance Manual. Microprocessor Based Digital Manifold NFPA

Operating Instructions Part No

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

User Information Sheet 013

Micro Dial-Flowmeter. Instructions for Use

Inert Air (N2) Systems Manual

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

Operating Instructions Part No

Oxygen Demand Valve. Instructions for Use

BOC Healthcare training

SINGER MODEL 106/206-RPS-L&H

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

CLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1

Installation, Operation, and Maintenance Manual

Training. Testor Training Manual

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Operating Instructions

This written Compressed Gas Plan is kept at the corporate office and in the written Safety & Health Program.

AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: CHAMP 3 PART NO: LS0115

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

MD R a n g e O p e r a t o r s H a n d b o o k

Installation & Operating Instructions for Genesys Analog Series Manifolds


OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Pressure Regulators. Operating Instructions. Instrumentation

Tri-Safem Model I1 Regulator. INSTRUCTIONS and PARTS

Working Environment Safety Assessment

East SP and Zeus SP Terminal Units Installation, Operation and Maintenance Manual

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS. This work covers part of outcome 2 of the Edexcel standard module:

INSTALLATION, OPERATION AND SERVICE MANUAL ABS AIR BAG LIFT

Installation and Operating Instructions For MANUAL MANIFOLDS SD and SDHP SERIES

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

PRESSURE TEST PUMP OPERATING & MAINTENANCE INSTRUCTIONS. Model No: PTP100. Part No: GC01/09

453 Series Steam Heated Vaporizing Regulator

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

TECHNICAL DATA. Q= Cv S

Premium Series. Economical Price & Reliable Performance. High Quality. Rust-Resistant. Long Life. Tamper-Resistant locking Screw.

1 Exam Prep NFPA 99 Health Care Facilities Questions and Answers (Plumbing Contractor)

Installation, Operation, and Maintenance Manual

Analgesic Demand Valve. Instructions for Use

Dri-Line Mk3 Monnier Compressed Air Drain Trap

Air Mortar User Manual

623 Series Dome Load Regulators

100C Air Compressor Kit

200 PSI COMPRESSORS - MODEL NUMBERS

Transcription:

MANUAL GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com U.K., SK3 0SB www.p3-phoenix.com

Contents Page Section Page Product Description General 3 Operation General 3 Pneumatic 3 Manual Control Panel Layout 4 Pressure Settings 4 Safety General 5 Installation General 5 Mechanical 5 Electrical 5 Installation Drawings 6 Electrical Schematic 7 Testing General 7 Testing 7 Commissioning 7 Maintenance General 7 Preventative Maintenance 8 Spare Parts General 8-2 - Manual Gas Manifolds Rev 2

1. Product Description 1.1 General Manual Gas Manifolds comprise a control panel, manifold headers & racks, tailpipes, an isolation valve, a non-return valve and a pressure relief valve. The manifold provides a means of manually changing the gas supply from one bank of cylinders to the other bank without interrupting the continuity of the supply. Manifolds are available for oxygen, nitrous oxide, 50% oxygen/50% nitrous oxide mixture and compressed air. An isolation and pressure relief valve assembly is provided with copper stub pipes for on site copper to copper joints to the hospital distribution pipeline and exhaust pipeline. A pressure switch is provided for connection to an alarm system to warn that the cylinders supplying gas are approaching empty. A non return valve is incorporated into each point where the tailpipe connects into the header to prevent a whole bank of cylinders becoming depleted in the event of a cylinder being disconnected or a tailpipe rupture. Tailpipes are available with pin-index, bull nose or CGA cylinder connections. 2. Operation 2.1 General The Manual Control Panel is not automatic and must be operated manually when required. The manifold should be kept in the following state; (a) All cylinders should be full of gas, open and connected to the tailpipes. (b) The shut-off valve for one bank of cylinders should closed and the shut-off valve for the other cylinder bank open, supplying pressure to the regulator. (c) The isolation valve to the distribution pipeline should be open. When the cylinder bank supplying gas becomes exhausted, the cylinder valve on the other full cylinder bank should be opened and the shut-off valve on the empty cylinder bank closed. The empty cylinders should then be disconnected from the tailpipes and replaced with new full cylinders. 2.2 Pneumatic Operation The gas from the left hand (LH) bank of cylinders is connected to the LH inlet of the control panel via the shut-off valve. This is connected to the contents pressure switch, contents pressure gauge and passes through regulator to the outlet pressure gauge. Similarly, the gas from the right hand (RH) bank of cylinders is connected to the RH inlet of the control panel via the shut-off valve. This is connected to the contents pressure switch, contents pressure gauge and passes through regulator to the outlet pressure gauge. The contents pressure gauge indicates the cylinder contents pressure in the LH and RH cylinders respectively, depending on which cylinder bank is open. The outlet pressure gauge indicates the outlet pressure from the regulator. When the pressure in the header falls below the setting of the pressure switch, the other cylinder bank is opened thus allowing the manifold to continue with the gas supply. The empty cylinders should now be replaced with full cylinders. This sequence will continue. - 3 - Manual Gas Manifolds Rev 2

2.3 Manual Control Panel Layout ID Description Qty ID Description Qty 1 Contents Pressure Switch 1 7 Outlet Pressure Gauge 1 2 Header Connector Pipe 2 8 Contents Pressure Gauge 1 3 Manual Backplate 1 9 Pressure Relief Valve 1 4 Shut-off Valve 2 10 Isolation Valve 1 5 Termination Box 1 11 Relief Valve Stub Pipe 1 6 Regulator 1 12 Outlet Stub Pipe 1 2.4 Pressure Settings Component O2 N2O O2/N2O Medical Air (400 kpa) Surgical Air (700 kpa) Regulators (static) 4.0 bar 4.0 bar 4.0 bar 4.0 bar 7.0 bar Pressure Relief Valve 5.5 bar 5.5 bar 5.5 bar 5.5 bar 9.0 bar Cylinder Pressure, Low 68 bar 14 bar 68 bar 68 bar 68 bar - 4 - Manual Gas Manifolds Rev 2

3. Safety 3.1 General This equipment should be installed, operated and maintained by personnel who are suitably trained, are fully conversant with HTM 2022 and are familiar with this product. This equipment should be kept clean and be free from oil and grease at all times. Oil and grease will ignite spontaneously in the presence of oxygen. If you suspect that any equipment is contaminated, do not use it. No attempt should be made to use or modify this equipment for use with a gas other than as identified. This equipment should not be operated at pressures exceeding those stated in HTM 2022 and this manual. Care should be taken when drawing conclusions about the contents of cylinders from their pressures. In the case of compressed gases, such as oxygen and medical air, the cylinder pressure is a good indication of the cylinder contents, for example, if the cylinder is half full, it will show half pressure. Liquefied gases, such as nitrous oxide, show full cylinder pressure as long as there is liquid in the cylinder. This may represent a small percentage of the full cylinder contents. It should not be assumed that a cylinder of nitrous oxide is full simply because the gauge shows full pressure. 4. Installation 4.1 General The manifold is supplied with all internal electrical connections pre-wired. The only external wiring required for the installation is the connection of the contents pressure switch (if required) to a plant alarm system. All regulators and pressure sensors are preset and no further adjustment of these settings should be required. 4.2 Mechanical (a) Secure the manual control panel to the wall using suitable fixings at the height indicated in figure 4.4. Ensure that there is sufficient space either side of the control panel to install the manifold header racks. (b) Secure the header racks to the wall using suitable fixings at the height indicated in figure 4.4. Use the header connecting pipes to assist in the accurate alignment of the racks. (c) Install the outlet stub pipe, pressure relief valve and isolation valve. (d) Install the exhaust pipeline from the pressure relief valve to a safe external location. (e) Install all tailpipes onto the manifold headers. (f) Check and tighten all mechanical joints. (f) Connect cylinders, pressurise the manifold and check for leaks. Make sure that the isolation valve is closed if the connection to the distribution pipeline has been made. 4.3 Electrical (a) Remove the fascia from the termination box mounted on the header rack. (b) Connect the wiring for a plant alarm system into the terminals on the remote PSW board. Refer to figure 4.5. - 5 - Manual Gas Manifolds Rev 2

4.4 Installation Drawings 4.5 Electrical Schematic Diagram - 6 - Manual Gas Manifolds Rev 2

5. Testing 5.1 General Prior to testing the manifold installation, please check the following. (a) All components have been installed are tightened. (b) The isolation valves on the manual manifold are closed. (c) Full cylinders have been fitted to all tailpipes. 5.2 Testing (a) Slowly open one cylinder on each bank and allow the system to pressurise. (b) Check all tailpipes and joints for leaks. (c) Close the cylinders. Create a leak between the regulator and the distribution pipeline and very slowly depressurise the system. (d) At the setting of the contents pressure switch, the RESERVE LOW indicator on the plant alarm system will illuminate. (e) If any pressure settings differ from those as stated in table 2.5, adjust as necessary. 5.3 Commissioning Demonstrate the manifold installation to the client by repeating the abovementioned test procedure. It is advisable to confirm that a central alarm system (if available) is operating correctly at the same time. The manifold must not be used until all testing & commissioning procedures for the pipeline system as detailed in HTM 2022 have been satisfactorily completed and accepted. 6. Maintenance 6.1 General Maintenance of the manifold should be restricted to periodic checking and adjustment and if necessary the replacement of faulty components. It should be noted that although many of the components appear to be standard items, many have been selected and treated to make them suitable for the gases carried and the pressures involved. 6.2 Preventative Maintenance Regular inspections and maintenance of the manifold will prolong it s life and reduce the possibility of sudden, inconvenient component failures. Manifolds should be subjected to regular inspection and testing as detailed below. Monthly; (a) Visually inspect the manifold for signs of damage. (b) Check all mechanical joints for leaks. (c) Check that cylinders connected are full. Annually; (a) Visually inspect the manifold system for signs of damage. (b) Check that one bank of cylinders is open ready for use. (c) Check all mechanical joints for leaks. (d) Test the manifold system as detailed in section 5.2 to confirm correct operation. (e) Observe that all indicators are working correctly. - 7 - Manual Gas Manifolds Rev 2

6.3 All maintenance should be carried out with the knowledge of the hospital engineer and in accordance with the Permit-to-Work system. 7. Spare Parts Lists When ordering spare parts, please quote the batch number of the equipment and a description of the component required to ensure that you receive the component that you require! Item Description Part Number 1 Tailpipe, O2, Pin-index 1-40800-01 2 Tailpipe, O2, Bull-nose 1-40801-01 3 Tailpipe, O2, CGA 540 1-40806-01 4 Tailpipe, N2O, BS 341 No.13 1-40802-02 5 Tailpipe, N2O, CGA 326 1-40807-02 6 Tailpipe, 50% O2/50% N2O, Pin-index 1-40803-03 7 Tailpipe, Medical Air, Pin-index 1-40804-04 8 Tailpipe, Medical Air, Bull-nose 1-40805-04 9 Tailpipe, Medical Air, CGA 346 1-40808-04 10 Pressure Relief Valve, 5.5 Bar 1-46000-09 11 Pressure Relief Valve, 9.0 Bar 1-46001-09 12 Regulator Assembly, 4 Bar 1-46002-09 13 N2O Regulator Assembly, 4 Bar 1-46003-02 14 Regulator Assembly, 7 Bar 1-46004-09 15 N2O Contents Pressure Switch, 14 bar 1-46005-02 16 Contents Pressure Switch, 68 Bar 1-46006-09 17 Contents Pressure Gauge, 0 400 Bar 1-46007-09 18 Contents Pressure Gauge, 0 160 Bar 1-46008-02 19 Outlet Pressure Gauge, 0 6 Bar 1-46009-09 20 Outlet Pressure Gauge, 0 16 Bar 1-46010-09 21 Bodok Seal for pin-index tailpipes 1-46011-09 22 O-ring for bull-nose tailpipes 1-46012-09 23 Manifold Header Link Pipe c/w o-rings 1-46023-09 24 Header O-ring Service Kit (per cylinder) 1-46027-09 25 Remote PSW PCB 1-46026-09 26 HP Manual Shut-off Valve 1-46025-09 27 Tailpipe, Medical Air, CGA540 1-40810-04 Due to our policy of continuous product improvement, Phoenix Pipeline Products Ltd. reserve the right to alter dimensions and/or specification without notice although every effort will be made to advise of any such modifications. Certain aspects of P³ Ltd. designs are protected. - 8 - Manual Gas Manifolds Rev 2