Bruker BioSpin NMR CASE. System Requirements Manual. Version. NMR Spectroscopy. think forward

Similar documents
Bruker Sample Transport

LC-SPE. Organizer 3.0 User Manual. Innovation with Integrity. Version 001 NMR

Animal Bed Rat Cryo. User Manual. Innovation with Integrity. Version MRI

Animal Bed Mouse Cryo

Ascend. User Manual. Innovation with Integrity. Version NMR

Helium Level Measurement Unit

RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats

BSMS/2. PNK Pneumatic Module Technical Manual. Version 001 BRUKER

Ascend DNP. User Manual. Innovation with Integrity. Version NMR

Ascend Aeon. User Manual. Innovation with Integrity. Version NMR

BNL. User Manual. Innovation with Integrity. Version. Accessories

Variable Temperature Storage Dewar Mount Inserts

Fourier Systems. General Safety Considerations User Manual. Innovation with Integrity NMR

Microscopy Cryogenic WORKSTATION. Performance by Design

Updated: Variable Temperature

Pump On Demand. Skimmer System

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Step By Step Instructions for VT Experiments on the Bruker 300 MHz Spectrometer

CRYO INDUSTRIES. Liquid Helium Research Dewars

Bruker BioSpin. SampleXpress Lite. Sample Changer User Manual. Version. NMR Spectroscopy. think forward

EMISSION CHAMBERS. Bulk and Surface Emission Detection for the RAD7 User Manual

A Quick Start Guide for the ER 4123D CW-Resonator

Site Planning for AVANCE Systems

CryoProbe Prodigy. Bruker User Manual. Innovation with Integrity. Version NMR

Varian 400-MR Magnetic Resonance Spectrometer Pre-installation Instructions

Stainless Steel 142 mm Filter Holder YY

ALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD

OptiAir CA09 Pneumaticfluidizing

Manual Solvent Purification System. Innovative Technology, Inc.

Variable Temperature Operation

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

1. Cryogen Tanks. N2 Fill Tutorial for Vanderbilt Magnets. Contents: CSB NMR Facility

Glove Box Installation Manual

Bruker BioSpin. SampleXpress. Sample Changer User Manual. Version. NMR Spectroscopy. think forward

DNP-NMR. DNP Control System Console User Manual. Innovation with Integrity. Version 002 NMR

Current Leads for Cryogenic Systems

Parts Manual. Spray Gun Series Master III. Issue /11/2017 Ref. NR ENG

Operating Instructions Part No

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

SECTION 2 RING MOUNT

Air preparation units

Judson Technologies PB 219. JTC JOULE-THOMSON CRYOSTAT COOLING SYSTEM Operating Instructions. October, 2003

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

Technical Service Bulletin 10001

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

DD2 Site Preparation Checklist Hardware Site Preparation Specification

MoveRoll Conveyor Operating and Maintenance Manual

Operating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100

User Manual. Quantos Automated Dosing Liquid Module

TECHNICAL SPECIFICATION. SPC-1 Stirling Process Cryogenerator

TECHNICAL SPECIFICATION. SPC-4 Stirling Process Cryogenerator

INSTALLATION, OPERATION AND MAINTENANCE MANUAL. Model Sanitary Vent SECTION I

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Hose Reel Unit, OMM. Discipline: TECHNIC Configuration control: X. Rev. Purpose Date Prep. Check. Appr. Changeorder No.

TELEDYNE PB 219. JUDSON TECHNOLOGIES A Teledyne Technologies Company TELEDYNE JUDSON TECHNOLOGIES JTC JOULE-THOMSON CRYOSTAT COOLING SYSTEM

User s Guide. Vacuum Controller for liquid delivery systems

Armatures for analytical probes

ABRASIVE REGULATOR II

INSTALLATION, OPERATION AND SERVICE MANUAL ABS AIR BAG LIFT

cooled, Heated ATH 2800 MT, DN 250 ISO-F, External drive electronics, Water cooled, Heated

MUELLER. Mega-Lite Drilling Machine. Reliable Connections. table of contents PAGE. Equipment 2. Operating Instructions 3-4. Parts Information 5

Performa Cv Twin Alternating and High Flow Systems. Manual Supplement

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

GM Series Dual-Block Multi-Function Gas Control Valves

High Pressure Equipment Taper Seal Valves, Fittings and Tubing

Laboratory Flowmeters

Supplementary instructions. Rod and cable components. for VEGAFLEX series 80. Document ID: 44968

Jupiter Type 4 Tube Owner s Manual

Vacuum. Dry Turbo Pumping Stations. Roughing Pumps for 1 K Helium Pots and Lambda Refrigerators. Hermetically Sealed Pumps for He3 Recirculation

Dual Modular Safety Shutoff Valves with Two-stage operation and Proof of Closure. DMV-ZRD/612 Series DMV-ZRDLE/612 Series

14. Exploded View. Page 68

VACGEN. Vacuum Components. Product Overview

TRADITIONAL HIGH VIBRATION HOSE SYSTEM THE STANDARD FOR MANY AUTOMOTIVE APPLICATIONS

Glove Box with Gas Purification System and Digital Control VGB-6. MTI Corporation

AR740 RPS Redundant Power Supply Quick Install Guide

Agilent 1290 Infinity Pump Head Maintenance

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

TBi RM 50 G. RoboMIG 1.1. Technical data. Robot welding torch TBi RM 50 G

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

MPG-plus. Application example. Pneumatic 2-Finger Parallel Gripper Gripper for small components. Gripping force 38 N 175 N. Sizes

Operating Instructions Part No

BSS 2 Bottle Sampling System. Instruction manual Version 1.8

Sensor for Air Bubble Detection at Liquid Filled Tubes. SONOCHECK Type ABD06.xx. Operating Manual

PFH. Application example. Pneumatic 2-Finger Parallel Gripper Long-stroke Gripper. Weight 2.65 kg 12.6 kg. Gripping force 630 N 2950 N

4/2 way Pneumatic Solenoid Valve

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

Pilot Check Valve: Metal Body Type

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

Paintball Marker. User s Manual. 530 South Springbrook Road Newberg, OR 97132

Constant Pressure Crude Oil Container Model CPCCP

Commercial Softeners. Installation, Operation & Maintenance Guide. Manual 002.1

SANDBLAST HELMET EXTREME

User s Guide. High Pressure NMR Cell. With High Pressure Tether. Bruker & Varian Cells TABLE OF CONTENTS:

Description Advanced Features Typical Applications

Dual Modular Safety Shutoff Valves. DMV-D/602 Series DMV-DLE/602 Series

Budget Range Operators Handbook

INSTRUCTION MANUAL & PARTS

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

Transcription:

Bruker BioSpin NMR CASE System Requirements Manual Version 012 think forward NMR Spectroscopy

Copyright by Bruker BioSpin NMR GmbH All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form, or by any means without the prior consent of the publisher. Product names used are trademarks or registered trademarks of their respective holders. This manual was written by Martin Rindlisbacher February 11, 2004: Bruker Biospin Corporation Billerica, MA, USA This manual was desktop published by Stanely J. Nles January 12, 2009: Bruker Biospin GmbH Rheinstetten, Germany P/N: B2928 DWG-Nr.: 0000012 For further technical assistance on the NMR CASE unit, please do not hesitate to contact your nearest BRUKER dealer or contact us directly at: BRUKER BioSpin GMBH am Silberstreifen D-76287 Rheinstetten Germany Phone: + 49 721 5161 0 FAX: + 49 721 5171 01 Email: service@bruker.de Internet: www.bruker.com

Contents Contents... 3 1 System Requirements... 5 1.1 Warnings... 5 1.2 Non-magnetic Ladder... 5 1.3 Necessary space on top of the magnet cryostat with TWO helium towers... 6 1.4 Necessary space on top of the magnet cryostat with THREE helium towers... 7 1.5 Orientation of the Shim Stack... 8 Precision of the orientation...8 Shimming...8 1.6 Shim Upper Stack Adapter Collars... 9 1.7 The telescopic legs hit a metal eyelet on the cryostat... 10 1.8 The telescopic legs are too long... 10 1.9 Electrical outlet requirement... 10 1.10 Compressed gas requirement... 10 NMR Case motion controller unit...10 NMR Console...10 1.11 Lifting up the helium tower connecting tubes... 11 Mounting... 11 System Requirements Manual Version 012 BRUKER BIOSPIN 3

Contents 4 BRUKER BIOSPIN System Requirements Manual Version 012

System Requirements 1 1 Warnings 1.1 WARNING! The NMR CASE has to be installed on top of a NMR magnet cryostat. DANGER! This magnet creates a very STRONG MAGNETIC FIELD. Objects and tools can be attracted to the magnet with great force and may cause damage. Use only NON-MAGNETIC tools and equipment. A non-magnetic TOOL SET is supplied with the NMR CASE. Read and follow carefully this installation manual.product description, usage, scope of manual, etc. Non-magnetic Ladder 1.2 To install the NMR case on top of the magnet cryostat, a non-magnetic ladder made from wood or aluminum must be available. Bruker does not supply nonmagnetic ladders. WARNING! To avoid damage to the magnet cryostat, use only NON - MAG- NETIC equipment. System Requirements Manual Version 012 BRUKER BIOSPIN 5 (13)

System Requirements Necessary space on top of the magnet cryostat with TWO helium towers 1.3 6 (13) BRUKER BIOSPIN System Requirements Manual Version 012

Necessary space on top of the magnet cryostat with THREE helium towers Necessary space on top of the magnet cryostat with THREE helium towers 1.4 System Requirements Manual Version 012 BRUKER BIOSPIN 7 (13)

System Requirements Orientation of the Shim Stack 1.5 The shim stack orientation must be as shown. The two air inlets must point to the front of the cryostat. The spin rate cable connector must point to the rear. The shim cable at the bottom of the cryostat must point to the rear. Precision of the orientation 1.5.1 The collar for the NMR Case can be rotated by ± 30 degrees; therefore the orientation of the shim stack does not need to be very precise. Shimming 1.5.2 If the shim stack needs to be rotated, complete new shim settings are necessary. Please consult a BRUKER engineer before rotating the shim stack. 8 (13) BRUKER BIOSPIN System Requirements Manual Version 012

Shim Upper Stack Adapter Collars Shim Upper Stack Adapter Collars 1.6 The following adapter collars are necessary and must be specified when ordering an NMR Case: Shim Upper Stack, Shim Upper Part Adapter, Shim Upper Part NMR Case Adapter Collar Type Part Number Serial Number Color Top Diameter Spin rate Sensor (width) Part Number BST Z9523 Z9524 Z9525 Z9526 Z42416 Z9527 Z46308 Z48798 Red Ø 75 mm Built in B2383 Is included with every NMR CASE Standard Bore old type-01 Z5178 Z001103 Z00627 Z5266 Red Ø 50 mm Small (20 mm) B0635 WB > SB adapter insert Z7680 Z3972 Z5297 Z6964 Below 0019 Below 0030 Below 0123 Below 0014 Gray Ø 50 mm Large (30 mm) B0636 WB > SB adapter insert Z7680 Z3972 Z5297 Z6964 Above 0019 Above 0030 Above 0123 Above 0014 Gray Ø 50 mm Small (20 mm) B0635 BST WB/ 99 > SB Z46401. A - L Red Ø 100 mm Built in B3944 System Requirements Manual Version 012 BRUKER BIOSPIN 9 (13)

System Requirements The telescopic legs hit a metal eyelet on the cryostat 1.7 Eyelet covers can be ordered from Bruker USA, the part number is B3156. The telescopic legs will rest on top of the eyelet covers. The telescopic legs are too long 1.8 Short legs can be ordered. The part numbers are B3743 and B3744, two each are necessary. Electrical outlet requirement 1.9 110 Volt AC 60 Hz or 220 Volt AC 50 Hz, 50 Watt. Compressed gas requirement 1.10 NMR Case motion controller unit 1.10.1 The NMR Case motion controller unit must be connected to compressed, oil-free, dry and dust-free air. Oil content < 0.005 ppm (0.005 mg/m3), filter size 1 micron, dew point < 4 degrees C. To ensure the long term operation of the NMR Case, the air filters need to be flushed out and cleaned on a regularly basis. Any dust or dirt will eventually block the valves and air flow reducers of the NMR Case. Any oil mist in the air will eventually weaken the air cylinders inside the NMR Case. A better solution is to use dry nitrogen gas with a minimum purity of 99%. The gas pressure should be between 60 to 90 pounds per square inch (4 bar to 6 bar). The necessary flow rate is 0.5 cubic foot per minute (14 liters per minute). NMR Console 1.10.2 The NMR console has to deliver the sample lift air. Usually, the sample lift airflow needs to be increased above the amount necessary for manual operation. The sample lift airflow is dependent on the weight of the sample spinners and the length and type of shim stack. For the blue 5 mm spinners and a short BST shim stack, an sample lift air flow of 3.5 cfm (100 l/m) is necessary. Heavy spinners and a long shim stack of an older version than BST need 4.5 cfm (140 l/m). Of course this high air flow is necessary only during the duration of the sample eject/insert cycle (ca 30 seconds). Consult the site installation manual and make sure there is enough supply airflow and pressure available to allow for a sample lift airflow set fully to the maximum. Consult the BSMS manual on how to increase the sample lift airflow. 10 (13) BRUKER BIOSPIN System Requirements Manual Version 012

Lifting up the helium tower connecting tubes Lifting up the helium tower connecting tubes 1.11 In some installations the shim stack tube is long and causes the NMR Case sample changer to sit high above the magnet dewar. In this case, there is not enough free space above the NMR Case, because the horizontal helium tower connection tube (T) of the magnet dewar is in the way. Solution: the tower connection tube (T) can be raised by 90 mm with commercial available vacuum components. Example: components from Nor-Cal Products, Inc. 1967 South Oregon Street, P.O. box 518, Yreka, California 96097, phone (916) 842-4457 or (800) 824-4166, fax (916) 842-9130, web page http://www.n-c.com/page.asp?navid=35. Clicking on NW Flanges and Fitting Catalog leads to an Adobe Acrobat file, which shows drawings and prices for the parts as follow: Parts (H): 4 pieces of NW 90 radius elbows part number 2E-NW-25B for $43.00 each (2004). Parts (R): 4 pieces of NW clamps, part number NW-25-CP for $ 7.00 each (2004) 4 pieces of NW aluminum centering rings with Viton o-ring, part number NW-25-CR-AV for $ 3.50 each (2004). Of course, these parts can also be ordered through BRUKER. Mounting 1.11.1 Open the clamps of the two originals quick flange couplings (R) and remove the he tower connection tube from the he towers. Important: to avoid moisture entering the helium towers, immediately block the two openings into the helium towers with Kleenex tissues. Connect the two extension tubes to both sides of the horizontal connection tube. Use two clamps and o-rings. Remove the Kleenex tissues and quickly reconnect the tube assembly to the two helium towers. System Requirements Manual Version 012 BRUKER BIOSPIN 11 (13)

System Requirements 12 (13) BRUKER BIOSPIN System Requirements Manual Version 012

End of Document System Requirements Manual Version 012 BRUKER BIOSPIN 13 (13)

Bruker BioSpin your solution partner Bruker BioSpin provides a world class, market-leading range of analysis solutions for your life and materials science needs. Bruker BioSpin Group Bruker BioSpin B2928 info@bruker-biospin.com www.bruker-biospin.com