Sterisil Ac+ Installation & Operating Manual

Similar documents
UltRo Dual Flow Reverse Osmosis Water System

BASIN AND WALL MIXER INSTALLATION INSTRUCTIONS

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL

ULTIMATE REVERSE OSMOSIS SYSTEM

IMPORTANT PLEASE READ BEFORE COMMENCING INSTALLATION

Operating Instructions Part No

ULTIMATE REVERSE OSMOSIS SYSTEM

Reverse Osmosis System Installation Guide and Operation Manual.

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Installation and Instruction Manual for 6 stage ROCCS Requires Professional Installation

Installation Guide & Owner s Manual. Model V4700. Water Filtration for Dental & Medical Instrument Washing Equipment

ULTRA REVERSE OSMOSIS SYSTEM

Operating Instructions Part No

ULTRA PURE SYSTEMS 625 GPD RODI INSTALLATION AND OPERATION MANUAL

SpectraPure PUMPED RO SYSTEMS (PSP) User s Manual for PSP-1500 Systems

REVERSE OSMOSIS INSTALLATION Watermark 4 Stage

Please Keep this Instruction Manual for future reference TABLE OF CONTENT

INSTALLATION & START-UP INSTRUCTIONS. Clack WS1 & WS1.25 METER WATER SOFTENER SYSTEMS

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

ESSENCE REVERSE OSMOSIS SYSTEM INSTALLATION INSTRUCTION

ROES-PH75 INSTALLATION INSTRUCTION & OWNER S MANUAL

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION GUIDELINES

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

INSTALLATION INSTRUCTION & OWNER S MANUAL

1020 Industrial Drive, Orlinda, TN fax

ULTRA PURE SYSTEMS 400 GPD RODI INSTALLATION AND OPERATION MANUAL

Pressure Dump Valve Service Kit for Series 3000 Units

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

Drinking Water Kit with Non-Air Gap Faucet for RO and RO/DI Systems (Including -MF Systems)

SWIRLFLO Refrigerated fountains with FLEXI-GUARD

Installation, Operation and Maintenance Manual

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

Owner s Manual AMERICAN. LiquaGen 5 Stage (RO/DI) Aquarium Reef System. YEAR Y 1

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

US WATER DENTAL OFFICE PURIFICATION SYSTEM

Flexfence Swaging Unit. Product Manual. Release 02/16.

INSTALLATION GUIDELINES

User's Manual. MixRite TF 10. Edition 05.08

Sanitizers. Sanitizers. Sanitizers for Pools & Spas

AutoChanger Installation & User Guide Issue 2

Pressure Dump Valve Service Kit for Series 2300 Units

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 3035 INSTRUCTION MANUAL

400C & 450C DUAL PERFORMANCE VALUE PACKS

Halsey Taylor Owners Manual 4410 Freeze Resistant Tubular Fountain STOP!

Sanitizers. Sanitizers. Sanitizers for Pools & Spas

Owners Manual. Models 4420BF1U & 4420BF1L Floor Mounted Bi-Level Bottle Filler Tubular Fountain

NGP-250/500 Nitrogen Generator Quick Start Guide

444C DUAL PERFORMANCE VALUE PACK

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

ROUGHNECK II Pumpless Drain

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

Installation Instructions

Installation Instructions

Halsey Taylor Owners Manual STOP!

Halsey Taylor Owners Manual

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS FOR GRT75-PF-RFS PET FOUNTAIN

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

INSTALLATION AND SERVICE MANUAL FOR LANCER DISPENSING TOWERS

Remineralized Reverse Osmosis System

Installation Instructions

200 PSI COMPRESSORS - MODEL NUMBERS

100C Air Compressor Kit

IMPORTANT SAFETY INSTRUCTIONS

WARNING! Equipment for treatment of water intended for human consumption (Directive 98/83 CE).

420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

INSTALLATION, OPERATION AND MAINTENANCE MANUAL ZERO WASTE REVERSE OSMOSIS SYSTEM

INSTALLATION, CARE & USE MANUAL. EMASMB & LMASMB Surface Mount Bottle Filling Stations

PAGE (Rev. A - 06/17)

PRO SINK -R 1 and 2. Dilution system CONTENTS

ACV-10 Automatic Control Valve

3M Water Filtration Products

CART FOR CENTURION 1500+

Installation and commissioning instructions 255 series and 256 series

Model PSI Compressor with 3-Gallon Air Tank 12VDC

Installation & Instruction Manual For Portable Series R.O. units:

RAM 4021-DPX Operation Manual

DRS4-RM Manual. Set Up Instructions for DRS4 Series Single Tank

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Zoi Series. RO System Reverse Osmosis Water Filtration System. DFK-Alpha * MB F. Core System Filter Set. Replacement Membrane

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

TABLE OF CONTENTS. Purchase Records... Introduction... Component List... Diagram... Specifications... Quick Connect Guide...

Installation and Instructions 2. Product Features 3-6. Key Pad Functions 7. Distributor Information Programming Guide 8. Master Programming Guide 9-14

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Pure Water Power Purification System 4-Stage RODI Operations Manual

SANITIZERS SANITIZERS SANITIZERS FOR POOLS & SPAS

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

DK ton, Double-acting, Die-type Crimping Tool

Installation, Operation, and Maintenance Manual

Using the UltraRAE. Firmware 2.35

Budget Range Operators Handbook

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

ALF400 to 900, ALT-5/-9

FAULT CODE TROUBLESHOOTING INDEX

Transcription:

Sterisil Ac+ Installation & Operating Manual 719 622 7200 Sterisil.com

Arbor Dental Associates 2

Table of Contents Introduction 4 Sterisil Ac + Features 5 Installation Instructions 6 Maintenance 10 Troubleshooting and FAQ 12 3

Introduction The ease of use and confidence that Sterisil has brought to dental unit water line treatment now comes to the world of distilled and deionized water for use with your office s table top autoclaves and instrument washers. With a 6 stage treatment process, the Sterisil Ac + produces up to 50 gallons of distilled quality water with 0 parts per million total dissolved solids (ppm TDS). Full system monitoring with audible and visual alarms make knowing when to replace cartridges worry-free. Automatic shut off valves mean that cartridge replacement is also an easy, no-spill process. The Sterisil Ac + requires only minutes of maintenance per year. Its space-saving design allows for easy installation even under sinks. Get ready to reclaim your storage space from all of those plastic jugs and containers with the Sterisil Ac +! This manual will show you how to install and maintain your new Sterisil Ac + unit and answer a few frequently asked questions that you may have. For additional support, please contact Sterisil at (719) 622-7200 or email technical support at support@sterisil.com. 4

Sterisil Ac + Features Simple + Designed for easy sterilization center or under-sink installation. + Optional extended life cartridges for areas with extremely hard water. + Automatic shut off valves for easy, spill-free cartridge replacement. + Optional filler wand for easy tabletop autoclave filling. + Sliding or hinged mounts for flexible installation site choices. + Metal enclosure protects internal components. Effective + Produces distilled quality water with 0 ppm TDS for use with table top autoclaves. + Can produce up to 50 gallons of water per day with up to 14 gallons of reserve water. + Water quality monitors sample the water purity from the system in real-time. + Class B Ultraviolet Irradiation pre-treats water for dental unit water line treatments. + Audible and visual alarms alert you when cartridges need to be replaced. + Dispensing faucet with LED monitoring light. + Backup filters supply your office with water for your table top autoclaves even after audible and visual alarms trigger. + Designed for offices with moderately hard to very hard water. + Sterisil ISO 13485 certification ensures that your new Sterisil Ac + unit has been tested for performance and reliability. 5

Installation Instructions 1. Unpack boxes and take inventory. Sterisil Ac + Back View w/ SlideMount (L) or Hinge Mount (R) & Electrical Connections Reserve Water Bladder Tank w/ Valve 1/4" PEX Tubing Sterisil Ac + Cartridges Stages 1, 2, and 3 (6 Total) Mounting Bracket (L) & Blank Cartridge (R) UV Lamp Transform 1/4 Tee (2) Reducers: 1/2" 3/8 (1, T) 3/8 1/4" (1, B) Saddle Drain Valve Faucet, White (L) & Faucet LED Disk (C) & Data Wire (R) Mounting Screw (7, L) & Dry Wall Anchor (7, R) 6

Ac+ Autoclave System Addendum Instructions 719 622 7200 sterisil.com

April 2015 Addendum to Ac+ Install Directions Update to Box inventory: -Please note that the larger tank will always be considered the reserve tank. -Permeate pumps are included with Sterisil Ac+ boxes, as shown below: -Also included with this Permeate pump is a mounting bracket, which is installed with one red mounting anchor and one screw Addendum to Step 8: -Reserve water should now read Autoclave water (to autoclave faucet) -Drinking water should now read To Permeate Pump (Permeate In) -Autoclave water should now read From Permeate Pump (Permeate Out) Addendum to Step 10 b. The line coming out of the bottom of the Stage 3 cartridge will now be hooked up as the following to accommodate the new permeate pump: -Make a cut in the line that comes from the bottom of the Stage 3 cartridge to the drain saddle or floor drain. -Take the line that is now only connected to the Stage 3 cartridge and insert it into the Brine In port found on the booster pump. -The other end of the line that was cut will now connect to the Brine Out port on the permeate pump with the other end still connecting to the drain saddle or floor drain. Addendum to Step 11 -This port is now yellow and is connected in the same way. Addendum to Step 13 -This port is now green and will read From Permeate Pump (Permeate Out) -At this point, we will be installing the last two lines into the Permeate Pump -Connect the To Permeate Pump (Permeate In) port to the Permeate In port on the Permeate Pump

-Connect the From Permeate Pump (Permeate Out) port to the Permeate Out port on the Permeate Pump -The line going between the From Permeate Pump (Permeate Out) and Permeate Out on the Permeate Pump will be cut to connect to the Reserve Tank in the following manner: -Cut the line between the From Permeate Pump (Permeate Out) and Permeate Out on the Permeate Pump -Insert both ends of that cut into the ¼ Union Tee -The other end of the Union Tee will run directly to the Reserve Tank Addendum to Step 19: -Please note that these drawings are no longer accurate. If you wish to have new drawings, please email Beau at b.sheidenberger@sterisil.com Please contact Beau for any clarification Phone: 719-622-7200 ext 210 Email: b.sheidenberger@sterisil.com

Installation Instructions 2. Note Ac + dimensions and specifications. STERISIL Ac+ SPECIFICATIONS Pressure Min/Max 65 psi / 125 psi Temp. Min/ Max Tubing (PEX, PUR) Elec. Outlets Fittings 45 F / 100 F 1/4" OD 2 110V AC, 3 AMP Quick Connect 3. Prep Installation Site. a. Remove items under sink (for under sink installations) and/or away from wall mount site. b. Ensure (2) 110 V AC electrical outlets are available within reach of the installation site. c. Ensure that the cold water municipal supply and drain are easily accessible at installation site. 7

Installation Instructions 4. Remove Mounting Plate from Sterisil Ac + and test fit at desired location. a. Ensure there is enough room for the completed Ac + with additional room to spare. i. Under Sink (Slide Mount) 1. Standard Cartridges (AC+4 Cartridges) a. Recommended: 20 H x 8 W x 21 D in. b. Minimum: 17 15 / 16 H x 7 W x 18 D in. 2. Extended Life Cartridges (AC+4XL Cartridges) a. Recommended: 22 H x 8 W x 21 D in. b. Minimum: 20 H x 7 W x 18 D in. ii. Wall Mount (Hinged Mount) 1. Standard Cartridges (AC+4 Cartridges) a. Recommended: 20 H x 8 W x 26 D in. b. Minimum: 17 15 / 16 H x 7 W x 24 D in. 2. Extended Life Cartridges (AC+4XL Cartridges) a. Recommended: 24 H x 8 W x 26 D in. b. Minimum: 20 H x 7 W x 24 D in. 5. Install the Mounting Plate. a. Place the Mounting Plate in desired location using a level (mounting hardware supplied). b. Mark the six holes designated below and using a 1/4" drill bit, drill each hole. Mounting Plate (Front View) Mounting Plate with Screws and Anchors (Side View) c. Carefully hammer an anchor in to each of the six holes drilled in step b. d. Line up the six holes in the mounting plate to the six anchors and secure the mounting plate by screwing a mounting screw in to each anchor. 8

Installation Instructions 6. Install the faucet. a. Drill a 5/8 hole into the sink surface at the desired location. b. Insert the threaded end of the faucet in to the LED faucet disk and feed the data wire through the 5/8 hole in the sink surface. Ensure the data wire is not pinched when inserting the threaded end of the faucet through the 5/8 hole in the sink surface. c. Secure the faucet with the black plastic washer, lock washer, and nut included with the faucet. Install the quick connect fitting as seen in the diagram below. 7. Mount the Sterisil Ac + and install cartridges. a. Slide Sterisil Ac + onto Mounting Plate slides, allowing hinges to lock. Test for fist and operation. Leave in extended position and continue to step b. 9

Installation Instructions b. With unit extended and municipal supply turned off, begin installation of cartridges starting with Stage 1 cartridge. 8. Establish Sterisil Ac + Plumbing Connections. a. Establish ¼ PEX tubing connection between cold water municipal supply and Sterisil Ac + SOURCE IN (red port) on side of Ac + with ¼ white PEX tubing supplied with Ac +. NOTE: If separate water source is not provided, shut off water supply and install easy connector and ball valve onto existing cold water supply valve as shown. 10

Installation Instructions b. Establish ¼ PEX tubing connection between white faucet and ¼ DI WATER (green port) on side of Ac + with ¼ white PEX tubing. NOTE: The optional autoclave filler wand (Part No. SS-Wand) can be used in conjunction with the white faucet or without it. If you choose to use the wand without the faucet, connect the blue coiled tubing directly in to the DI WATER (green port) on the side of the Ac +. c. Establish ¼ PEX tubing connection between the reserve bladder tank and the RESERVE TANK (yellow port) on the side of the Ac +. NOTE: The Sterisil Ac + comes standard with a 4 gallon reserve bladder tank. Sterisil also offers 10 gallon or 14 gallon reserve bladder tank options and are available for order. If it becomes apparent that your office uses more than 3 gallons of autoclave water per day, Sterisil recommends that a larger bladder tank be purchased. 11

Installation Instructions d. Install the Drain Saddle Valve for Stage 3 drain. i. Place the saddle valve on the sink drain pipe and mark the desired location for the drain hole on the pipe. ii. Using a 1/4 drill bit, drill a hole only through the first wall of the drain pipe. iii. Place the saddle valve over the hole and tighten in to place. e. Establish ¼ PEX tubing connection between the drain saddle valve and the Stage 3 brine drain with ¼ white PEX tubing. NOTE: Be sure to remove the rubber stopper from the brine drain on Stage 3 before inserting the drain tubing. Push the small, white collar on the Stage 3 brine drain at the bottom of the cartridge toward the cartridge while simultaneously pulling the rubber stopper out. 12

Installation Instructions Sterisil Ac + Operational Mode 9. Connect all electrical connections. a. Plug the standard, black power cable in to a nearby 110 V AC wall outlet. b. Connect the white connector on the black transformer that came with the Ac + to the white connector from the black power conduit. Plug in the other end of the black transformer to the nearby wall outlet. 13

Installation Instructions c. Connect the data cable form the system to one end of the beige modular coupler that came with the Ac + and connect the other end to the data cable from the faucet LED disk. 10. Slide unit in and out to check for snags and to ensure that the tubing is not pinched. 11. Turn the bladder tank valve to the off position and open the municipal water supply valve. Check for leaks and allow the system to charge briefly. NOTE: As the cartridges fill, the Stage 3 R/O drain line will flow as it flushes contaminants in to the drain. 14

Installation Instructions 12. Open the faucet on top of the sink (and/or autoclave filler wand) and run until a slow, constant trickle of water comes out. 13. Turn off the faucet and open the bladder tank valve. Check for leaks and to make sure all fittings are secure. It will take between 4 to 8 hours for the system to fully charge and fill the reserve bladder tank depending on the source pressure and size of the bladder tank(s). You have now completed installation of your new Sterisil Ac +! 15

Maintenance Manual MAINTAINING YOUR STERISIL Ac + Your Sterisil Ac + is equipped with full system monitoring as well as audible and visual alarms that will alert you when a cartridge replacement is necessary. This section will tell you how to maintain your Sterisil Ac +, how to take and understand the different readings, and also when to replace which cartridges based on which alarm. Reading the Sterisil Ac + Monitors Filter Monitor The filter monitor keeps track of the remaining life of stages 1, 2, 3, and the UV light. These cartridges and the UV lamp are designed to last one calendar year. To check on the remaining life of stages 1, 2, 3 and the UV light, press the PWR/STG button to turn the monitor on. Press the PWR/STG button again to select Stage 1. On the left of the screen, there will be a number indicating the remaining number of months before these cartridges must be replaced. Once these cartridges are depleted, the monitor will trigger the Filter Monitor Alarm LED on the faceplate of the Ac +. The LED will flash red and the unit will beep, letting you know that it is time to replace the cartridges. Sterisil recommends always having a backup set of cartridges and purchasing a new set immediately after replacing cartridges. After replacing these cartridges and the UV light, it is necessary to reset the filter monitor. Refer to the monitor reset instructions included in this manual on how to properly reset the Filter Monitor. RO Quality Monitor This monitor will tell you the water quality of your source water and the water coming out of the stage 3 filter by monitoring the parts per million Total Dissolved Solids (ppm TDS) in the water. Total Dissolved Solids are any material in the water that is not water; including dissolved minerals, gases, and oils. This monitor will tell you how the first three filters are performing and are also used during troubleshooting and diagnostics. To take the Source water and RO readings, turn the monitor on by pressing the silver On/Off button in the upper right-hand corner of the monitor. Flip the white slide to the left (SOURCE) and record the reading. This is the water quality of your municipal cold water supply line and is dependent on your geographical location. Next, flip the white slider to the right (RO) and record the reading. This is the water quality after the Stage 3 cartridge. The ideal range of this reading is 10% or below of the SOURCE reading. For example: if you have a source water reading of 200 ppm TDS, the RO reading should be about 20 ppm TDS or less. If this number exceeds 10% of the SOURCE reading, refer to the troubleshooting section of this manual. 16

Maintenance Manual DI Quality Monitor This monitor will tell you the water quality of the water after the first and second stage 4 cartridges. To take the DI 1 reading (the quality of the water after the first Stage 4 cartridge), turn the monitor on by pressing the silver On/Off button in the upper right-hand of the monitor. Slide the white slide to the left (or leave it there if it is already to the left) and record the number. This number should be significantly less than the RO reading from the RO Quality Monitor. To take the DI 2 reading, slide the white slider to the right and record the number displayed. This number should be below 10 ppm TDS. When DI 2 exceeds 10 ppm TDS the audible alarm will sound and the DI Water Quality Alarm LED on the faceplate of the Ac + will turn solid red. This means that the first two Stage 4 cartridges have been depleted and should be replaced. DO NOT replace all three Stage 4 cartridges at the same time as this is wasting one of the cartridges. Sterisil has outfitted the Ac + with three Stage 4 cartridges to ensure that even after the alarms signal the need for a replacement, you will still have the ability to use the DI water from the unit in your table top autoclaves. To replace the depleted Stage 4 cartridges, remove the first two Stage 4 cartridges and dispose of them normally. Move the third Stage 4 cartridge in to the now empty first Stage 4 position. Now install two brand-new Stage 4 cartridges in to the second and third Stage 4 positions. NOTE: It is recommended that you take all four readings from the Ac + on a regular basis (weekly if at all possible) using the Data Log Sheet on the next Page. 17

Sterisil Ac + Data Log Sheet Filter Installation Date: _ Date Months Remaining (Filter Monitor) Source RO DI 1 DI 2 NOTE: Refer to directions on Pg 16 & 17 for how to find the above readings 18

Thank You! Thank you for purchasing your new Sterisil Ac! We know you will be pleased with this Unit! Should you have any installation questions, please contact us at (719)622-7200 or online at Sterisil.com. 19

20

21