NARKOVET ANAESTHETIC UNIT

Similar documents
Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

VAPOMATIC ANESTHETIC VAPORIZER

Selecting and Connecting Breathing Systems

Recommendations on Checking Anaesthesia Delivery Systems

Flow meter. bellow vaporizer. APL valve. Scavenging system

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Breathing Systems. Professor Khalid Bashir

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

M5400 Operators Manual

Checking a Breathing System

Hallowell EMC Anesthesia WorkStation Operating Manual

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

Anesthesia Machine Intro

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Anes Machine Use and Pre-use Check - CA1 focus

SERVICE KITS AND CALIBRATION NOTES

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

SURGITRON USER MANUAL TELEFON Art. Nr

Latvin Luxury Shower Panel. Telephone Product Specification. ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3.

Keeping anaesthetic pollution to minimum in theatre environment

Landmark. Veterinary Anesthesia Systems. Operator s Manual P/N

Breathing Circuits. Product training

1 of 5. Functionn. HTHP mud. measuree. of the filter. when the Power. Structure

Operating Instructions Model and Hydrostatic Test Pump

International Journal of Science, Environment and Technology, Vol. 5, No 5, 2016,

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Product Manual. Description. Specifications. CVS Type 1301F and CVS Type 1301G Regulator. Introduction

Testing anaesthetic machines, ventilators and related equipment

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

MEGR-1627 Instruction Manual

OWNER S TECHNICAL MANUAL

Budget Range Operators Handbook

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

Operating Manual. 3S-R AirElite Full Face Mask Advantage AirElite Full Face Mask Advantage 4000 AirElite Full Face Mask. AirElite - Full Face Masks

64 Series Pressure Reducing Regulators

Medical Instruments in the Developing World

Training. Testor Training Manual

BleaseFocus Anesthesia System VERSATILE ANESTHESIA

FLOWMETER UNITS FLOWMETER DEVICES, BY-PASS SYSTEMS

The centerpiece of a comprehensive anaesthesia workstation

Code AWC20HP Air Compressor

Anesthesiology Research and Scholarship. Follow this and additional works at:

AUTOVENT 4000 VENTILATOR

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

User's Manual. MixRite TF 10. Edition 05.08

Model 405 &407 Small Animal Ventilator. User Manual v3.5

Operating Instructions Models 36200, , 46200, Hydrostatic Test Pumps

ACUC ۰ ROUTINE PROCEDURE

OPERATING INSTRUCTIONS

WW-720. Pressure Reducing Control Valve

IV-ANE. User s Manual. Anesthesia equipment for test animals. Petition

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Propane Conversion Kit Instruction

B.C.D. Buoyancy Compensator Device User Manual

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

Operation and Maintenance of the EPV200 Portable Ventilator

Operating Instructions Models and Hydrostatic Test Pumps

299H Series. Introduction. P.E.D. Categories. Specifications. Installation. Warning. Installation Guide English September 2012

Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable

Techcon Systems TS1254 Pressure Pot

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Service and Repair Manual

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

APUS Anesthesia Machines

Anesthesia. Midmark Animal Health Products

Columbus Instruments

THE UNIVENTOR 400 ANAESTHESIA UNIT

Installation and Maintenance Manual. Compact Medical Gas Outlets

Operating Manual MSA SSR 30/100 / MSA SSR 30/100 B EN K/20/S. Order No /01. MSAsafety.com

ANESTHESIA VAPORIZER SERVICE RECOMMENDATIONS

TEST BENCH SAFETY VALVES ¼ - 5 DN10 DN125

Pulmonex II Training Presentation

VACUUM REGULATORS CONTENTS

Aespire 7100 Essential performance Compact design

WW-720. Pressure Reducing Control Valve

Operating Manual ALU 7 - ALU 10 - ALU 20 - ALU 26 ALU 35 - ALU 60 - ALU 100

The SAV04 Small Animal Ventilator. Contents Page

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Welcome to Maxstar Medical

S/5 Aespire. Quick Reference Guide

Draeger Fabius GS Pre-Use Check

Instruction Manual Updated 7/26/2011 Ver. 2.2

30T A/Manual Hydraulic Shop Press

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

Model VR6 System. Installation, Operation & Maintenance

BioPak 240 Revolution

PRESSURISED PAINT CONTAINER

Purge Star & Ring Purge Systems

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

AR251/AR251-B. Venepuncture and Infusion Arm. Instruction Manual

Operating Procedures for GripTight 15.5 SDR & IPS Test Plugs

Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL

4 ANGLE GRINDER MODEL NO: CAT 52 PART

A200SP Circle Absorber Service Manual

The Univentor 1250 Anaesthesia Unit

Instructions for use Compressed air injector with water trap

Transcription:

NARKOVET ANAESTHETIC UNIT MANUAL Art. No. 213062 TELEPHONE +49 7461 96 580 0 www.eickemeyer.com

Table of Contents 1 Description...4 2 Overview of the Device...4 3 Startup Procedure...5 4 Operation of the Pendulum System (Bain Circuit)...7 5 Preparing the Device before Anesthaesia...8 6 Anaesthetic Procedure...8 7 Cleaning and Maintenance of Anesthaesia Equipment...9 3/12

1 Description The anesthaesia machine NarkoVet is a semi-closed loop system with partial rebreathing. The device is suitable for anesthaesia of small animals up to a body weight of 80 kg. For animals up to 15 kg it is recommended to use the included pendulum system (Bain Circuit). Further advantages of NarkoVet are: - Respiratory valves that are large and easy to see - A soda lime container that is easy to refill and install with a capacity up to 10 hours of anesthesia - Pipes for fresh gas that are easy to adapt and which can also be used for anesthaesia boxes - Flow meter that is easy to adjust and read with a gas flow from 0.2 to 4.2 l/min - Flush button for the oxygen bypass 2 Overview of the Device 1. Precision Flow Meter 2. Oxygen-Bypass 3. Evaporator 4. Safety Filling Device 5. Oxygen Supply Line (out of a Bottle or from the Oxymat) 6. Absorbent Container 7. Direction Valves 8. Pressure Relief Valve 9. Connection for the Breathing Bag 10. Tube Connection for the Patient 11. Absorbent Container Lock Lever 4/12

3 Startup Procedure Connect the oxygen supply line (5) to the oxygen supply. This can be done either through: - Oxygen cylinder and pressure reducer (Part No. 212705 or 21270501) - Oxygen concentrator (Part No. 213175 + 213160) - Air connector (piped oxygen) (Part No. 212503) The input pressure should be about 3 bar. When using an oxygen concentrator the inlet pressure is lower due to the type of construction. Thus, the filling time of the breathing bag is extended when operating the flush button (oxygen bypass (2)). The flow meter, however, always displays the correct oxygen flow. NarkoVet with 10 l oxygen cylinder and pressure reducer (not included in the delivery) Fill the evaporator with Isoflurane. The procedure is shown on the evaporator. 1. Screw the Isoflurane key fill adapter onto bottle (Art. No. 213246) 2. Untighten the screw on the side of the safety filling device, remove the sealing block, introduce the safety filler into the filling tunnel and tighten screw. 3. Turn the black filling screw to the 12 clock position. 4. Fill the evaporator up to the maximum filling mark. 5. Set the black filling screw on 3 clock position 6. Remove the key fill adapter and insert sealing block and turn the screw very tightly. 7. In older evaporators this may be impossible due to the lack of ventilation. In this case loosen the bottle from the filler, so that air bubbles ascend in the bottle so that the anesthetic agent runs into the evaporator. 8. A strong Isoflurane odour deriving during the filling, it indicates a worn out seal plate, which must be changed. 5/12

Filling the soda lime Turn the lock lever of the absorbent container (11) counterclockwise. Remove the soda lime. Fill the container with soda lime. The max. filling line must not be exceeded. Insert the container and lock it by turning the lock lever clockwise. Plug in the patient circuit and the breathing bag. Connect an exhaust pipe to the pressure relief valve. 6/12

4 Operation of the Pendulum System (Bain Circuit) The Bain Circuit supplements the basic unit perfectly for anesthaesia of smaller patients. It contains no breathing valves. The anesthetic gas is transported directly from the evaporator via the fresh gas pipe to the patient. This ensures quicker wash-in and wash-out even with a smaller minute volume. The exhalation takes place into the breathing bag, and in the case of overpressure air is transferred via the open valve in the exhaust pipe. The pressure relief valve should always be fully opened during spontaneous breathing (against clockwise direction). A re-breathing of exhaled carbon dioxide is prevented by the influx of fresh gas into the breathing bag (open system). Recommended quantity of fresh gas for wash-in: 800 ml - 1000 ml/min Concentration of Isoflurane: 2.5% - 3.5% Maintenance values as required, but at least 500 ml/min fresh gas (depending on evaporator). Please pay attention to the correct assembly of the anesthaesia system, as shown in the picture above. 7/12

5 Preparing the Device before Anesthaesia Check the oxygen supply (pressure and access to sufficient oxygen). Connect patient tube and breathing bag to the anesthesia machine and close the pressure relief valve. Keep your thumb on the patient's end of the tube and inflate the breathing bag via the Oxygen bypass until the breathing bag is bulging. The anesthaetic system is sufficiently tight once a fresh gas flow of less than 200 ml is needed to maintain this inner pressure. Now check the function of the overpressure valve by fully opening it very slowly. If the internal pressure is not adequately maintained, there is a leak in the system which has to be found before using the anaesthetic machine. Check the anesthetic level of the Evaporator. In general, due to the built in pressure compensation of the evaporator a loss of anesthetic agent up to 0.5 cm 3 per day can be expected. 6 Anaesthetic Procedure Typically in a semi-closed system a fresh gas flow of 2 l/min at a selected concentration of 3.0 vol% at the evaporator is sufficient for a quick wash-in. The depth of anesthaesia needs to be monitored and the concentration to be changed if necessary. To maintain the anaesthesia, generally it is sufficient to follow the guidelines of the MAC-values + 15% inspiratory add as per the package leaflet at minimum 500 ml/min oxygen inflow (watch status of anaesthesia and breathing continuously). The MAC values are Dog: 1.28% Cat: 1.63% Birds: 1.45% Small mammals: 1.34% Rat: 2.40% Pay attention to the decreased wash-in and wash-out times of an Isoflurane-only anesthaesia in comparison to halothane or mixed anesthaesia. In general, the evaporator should be closed just at the time of the final suture (5 min before the intended end of the OP). 8/12

7 Cleaning and Maintenance of Anesthaesia Equipment Remove breathing bag, the patient tube and other tubes from the device and dry them (if necessary, clean with detergent). Unscrew the lids of the breathing valves and let them dry too. If you remove the valve plates, it will speed up the drying process. But please make sure to insert these plates again before the next anaesthetic, otherwise proper gas exchange is no longer guaranteed. Approximately every 10 hours, please change the absorbent, even if the indicator change did not take place. Please check every 20 hours of anesthaesia all exposed sealing rings for wear and tear and lubricate with silicon vaseline. Once a year, the evaporator should be subjected to a technical inspection to ensure proper functioning. In this case, please contact the EICKEMEYER Customer Service. Precautions for handling Isoflurane: Isoflurane is a highly volatile anaesthetic. Therefore, when used in closed spaces ensure that a suction tube is properly connected to the pressure relief valve or the Bain Circuit. If ventilation of gases is not possible into the open air, we recommend an anesthetic gas filter (Art. No. 213485). 9/12

Notes: 10/12

Notes: 11/12

GERMANY EICKEMEYER KG Eltastraße 8 78532 Tuttlingen T +49 7461 96 580 0 F +49 7461 96 580 90 E info@eickemeyer.de www.eickemeyer.de ITALY EICKEMEYER S.R.L. Via G. Verdi, 8 65015 Montesilvano (PE) T +39 085 935 4078 F +39 085 935 9471 E info@eickemeyer.it www.eickemeyer.it CANADA EICKEMEYER Inc. 250 Briarhill Dr. Stratford, Ont. Canada N5A 7S2 T +1 519 273 5558 F +1 519 271 7114 E info@eickemeyervet.ca www.eickemeyervet.ca DENMARK EICKEMEYER ApS Lysbjergvej 6, Hammelev 6500 Vojens T +45 7020 5019 F +45 7353 5019 E info@eickemeyer.dk www.eickemeyer.dk NETHERLANDS EICKEMEYER B.V. Bedrijventerrein Pavijen-West Bellweg 44 4104 BJ Culemborg T +31 345 58 9400 F +31 345 58 9401 E info@eickemeyer.nl www.eickemeyer.nl POLAND EICKEMEYER Sp. z o.o. Al. Jana Pawła II 27 00-867 Warszawa T +48 22 185 55 76 F +48 22 185 59 40 E info@eickemeyer.pl www.eickemeyer.pl SWITZERLAND EICKEMEYER AG Sandgrube 29 9050 Appenzell T +41 71 788 23 13 F +41 71 788 23 14 E info@eickemeyer.ch www.eickemeyer.ch UNITED KINGDOM EICKEMEYER Ltd. 3 Windmill Business Village Brooklands Close Sunbury-on-Thames Surrey, TW16 7DY T +44 20 8891 2007 F +44 20 8891 2686 E info@eickemeyer.co.uk www.eickemeyer.co.uk