NARKOVET ANAESTHETIC UNIT MANUAL Art. No. 213062 TELEPHONE +49 7461 96 580 0 www.eickemeyer.com
Table of Contents 1 Description...4 2 Overview of the Device...4 3 Startup Procedure...5 4 Operation of the Pendulum System (Bain Circuit)...7 5 Preparing the Device before Anesthaesia...8 6 Anaesthetic Procedure...8 7 Cleaning and Maintenance of Anesthaesia Equipment...9 3/12
1 Description The anesthaesia machine NarkoVet is a semi-closed loop system with partial rebreathing. The device is suitable for anesthaesia of small animals up to a body weight of 80 kg. For animals up to 15 kg it is recommended to use the included pendulum system (Bain Circuit). Further advantages of NarkoVet are: - Respiratory valves that are large and easy to see - A soda lime container that is easy to refill and install with a capacity up to 10 hours of anesthesia - Pipes for fresh gas that are easy to adapt and which can also be used for anesthaesia boxes - Flow meter that is easy to adjust and read with a gas flow from 0.2 to 4.2 l/min - Flush button for the oxygen bypass 2 Overview of the Device 1. Precision Flow Meter 2. Oxygen-Bypass 3. Evaporator 4. Safety Filling Device 5. Oxygen Supply Line (out of a Bottle or from the Oxymat) 6. Absorbent Container 7. Direction Valves 8. Pressure Relief Valve 9. Connection for the Breathing Bag 10. Tube Connection for the Patient 11. Absorbent Container Lock Lever 4/12
3 Startup Procedure Connect the oxygen supply line (5) to the oxygen supply. This can be done either through: - Oxygen cylinder and pressure reducer (Part No. 212705 or 21270501) - Oxygen concentrator (Part No. 213175 + 213160) - Air connector (piped oxygen) (Part No. 212503) The input pressure should be about 3 bar. When using an oxygen concentrator the inlet pressure is lower due to the type of construction. Thus, the filling time of the breathing bag is extended when operating the flush button (oxygen bypass (2)). The flow meter, however, always displays the correct oxygen flow. NarkoVet with 10 l oxygen cylinder and pressure reducer (not included in the delivery) Fill the evaporator with Isoflurane. The procedure is shown on the evaporator. 1. Screw the Isoflurane key fill adapter onto bottle (Art. No. 213246) 2. Untighten the screw on the side of the safety filling device, remove the sealing block, introduce the safety filler into the filling tunnel and tighten screw. 3. Turn the black filling screw to the 12 clock position. 4. Fill the evaporator up to the maximum filling mark. 5. Set the black filling screw on 3 clock position 6. Remove the key fill adapter and insert sealing block and turn the screw very tightly. 7. In older evaporators this may be impossible due to the lack of ventilation. In this case loosen the bottle from the filler, so that air bubbles ascend in the bottle so that the anesthetic agent runs into the evaporator. 8. A strong Isoflurane odour deriving during the filling, it indicates a worn out seal plate, which must be changed. 5/12
Filling the soda lime Turn the lock lever of the absorbent container (11) counterclockwise. Remove the soda lime. Fill the container with soda lime. The max. filling line must not be exceeded. Insert the container and lock it by turning the lock lever clockwise. Plug in the patient circuit and the breathing bag. Connect an exhaust pipe to the pressure relief valve. 6/12
4 Operation of the Pendulum System (Bain Circuit) The Bain Circuit supplements the basic unit perfectly for anesthaesia of smaller patients. It contains no breathing valves. The anesthetic gas is transported directly from the evaporator via the fresh gas pipe to the patient. This ensures quicker wash-in and wash-out even with a smaller minute volume. The exhalation takes place into the breathing bag, and in the case of overpressure air is transferred via the open valve in the exhaust pipe. The pressure relief valve should always be fully opened during spontaneous breathing (against clockwise direction). A re-breathing of exhaled carbon dioxide is prevented by the influx of fresh gas into the breathing bag (open system). Recommended quantity of fresh gas for wash-in: 800 ml - 1000 ml/min Concentration of Isoflurane: 2.5% - 3.5% Maintenance values as required, but at least 500 ml/min fresh gas (depending on evaporator). Please pay attention to the correct assembly of the anesthaesia system, as shown in the picture above. 7/12
5 Preparing the Device before Anesthaesia Check the oxygen supply (pressure and access to sufficient oxygen). Connect patient tube and breathing bag to the anesthesia machine and close the pressure relief valve. Keep your thumb on the patient's end of the tube and inflate the breathing bag via the Oxygen bypass until the breathing bag is bulging. The anesthaetic system is sufficiently tight once a fresh gas flow of less than 200 ml is needed to maintain this inner pressure. Now check the function of the overpressure valve by fully opening it very slowly. If the internal pressure is not adequately maintained, there is a leak in the system which has to be found before using the anaesthetic machine. Check the anesthetic level of the Evaporator. In general, due to the built in pressure compensation of the evaporator a loss of anesthetic agent up to 0.5 cm 3 per day can be expected. 6 Anaesthetic Procedure Typically in a semi-closed system a fresh gas flow of 2 l/min at a selected concentration of 3.0 vol% at the evaporator is sufficient for a quick wash-in. The depth of anesthaesia needs to be monitored and the concentration to be changed if necessary. To maintain the anaesthesia, generally it is sufficient to follow the guidelines of the MAC-values + 15% inspiratory add as per the package leaflet at minimum 500 ml/min oxygen inflow (watch status of anaesthesia and breathing continuously). The MAC values are Dog: 1.28% Cat: 1.63% Birds: 1.45% Small mammals: 1.34% Rat: 2.40% Pay attention to the decreased wash-in and wash-out times of an Isoflurane-only anesthaesia in comparison to halothane or mixed anesthaesia. In general, the evaporator should be closed just at the time of the final suture (5 min before the intended end of the OP). 8/12
7 Cleaning and Maintenance of Anesthaesia Equipment Remove breathing bag, the patient tube and other tubes from the device and dry them (if necessary, clean with detergent). Unscrew the lids of the breathing valves and let them dry too. If you remove the valve plates, it will speed up the drying process. But please make sure to insert these plates again before the next anaesthetic, otherwise proper gas exchange is no longer guaranteed. Approximately every 10 hours, please change the absorbent, even if the indicator change did not take place. Please check every 20 hours of anesthaesia all exposed sealing rings for wear and tear and lubricate with silicon vaseline. Once a year, the evaporator should be subjected to a technical inspection to ensure proper functioning. In this case, please contact the EICKEMEYER Customer Service. Precautions for handling Isoflurane: Isoflurane is a highly volatile anaesthetic. Therefore, when used in closed spaces ensure that a suction tube is properly connected to the pressure relief valve or the Bain Circuit. If ventilation of gases is not possible into the open air, we recommend an anesthetic gas filter (Art. No. 213485). 9/12
Notes: 10/12
Notes: 11/12
GERMANY EICKEMEYER KG Eltastraße 8 78532 Tuttlingen T +49 7461 96 580 0 F +49 7461 96 580 90 E info@eickemeyer.de www.eickemeyer.de ITALY EICKEMEYER S.R.L. Via G. Verdi, 8 65015 Montesilvano (PE) T +39 085 935 4078 F +39 085 935 9471 E info@eickemeyer.it www.eickemeyer.it CANADA EICKEMEYER Inc. 250 Briarhill Dr. Stratford, Ont. Canada N5A 7S2 T +1 519 273 5558 F +1 519 271 7114 E info@eickemeyervet.ca www.eickemeyervet.ca DENMARK EICKEMEYER ApS Lysbjergvej 6, Hammelev 6500 Vojens T +45 7020 5019 F +45 7353 5019 E info@eickemeyer.dk www.eickemeyer.dk NETHERLANDS EICKEMEYER B.V. Bedrijventerrein Pavijen-West Bellweg 44 4104 BJ Culemborg T +31 345 58 9400 F +31 345 58 9401 E info@eickemeyer.nl www.eickemeyer.nl POLAND EICKEMEYER Sp. z o.o. Al. Jana Pawła II 27 00-867 Warszawa T +48 22 185 55 76 F +48 22 185 59 40 E info@eickemeyer.pl www.eickemeyer.pl SWITZERLAND EICKEMEYER AG Sandgrube 29 9050 Appenzell T +41 71 788 23 13 F +41 71 788 23 14 E info@eickemeyer.ch www.eickemeyer.ch UNITED KINGDOM EICKEMEYER Ltd. 3 Windmill Business Village Brooklands Close Sunbury-on-Thames Surrey, TW16 7DY T +44 20 8891 2007 F +44 20 8891 2686 E info@eickemeyer.co.uk www.eickemeyer.co.uk