URN (UNIVERSAL RODENT NOSE CONE) (17mm diameter) Figure 1

Similar documents
INSTRUCTION MANUAL. (Variante 2 / System 2) Surgery table standard (Variante 1 / System 1)

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

GAS ANESTHESIA SYSTEMS. Cat. No / 22100

INSTRUCTION MANUAL. Combi-vet Anesthesia system

IV-ANE. User s Manual. Anesthesia equipment for test animals. Petition

instruction manual Gas Anesthesia System Cat. No / UGO BASILE S.R.L. /

ACUC ۰ ROUTINE PROCEDURE

Anesthesiology Research and Scholarship. Follow this and additional works at:

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

Chapter 9 Accessories

Flow meter. bellow vaporizer. APL valve. Scavenging system

Instructions for using TSI Model mm or 8025-CEN Fit Test Adapter Kit

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Instructions for Use. OptiLife Nasal Lab Mask

Breathing Circuits. Product training

Selecting and Connecting Breathing Systems

M5400 Operators Manual

Checking a Breathing System

G250/G250 HP Compatibility of Upgrade Kit

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

C-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Mirage Vista Nasal Mask

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Anesthesia. Midmark Animal Health Products

We Warm Animals. User Manual For. Darvall Heated Smooth-Wall Circuits. Darvall Heated ZDS Qubes

LISTADO DE CIRCUITOS DE PACIENTE

Phalanx Alpha Phalanx Beta and Ultra Elite

Breathing Systems. Professor Khalid Bashir

UltraFilter. respirator. with Ultravue or Ultra Elite Facepiece. instructions for use and care

Anesthesia Machine Intro

VAPOMATIC ANESTHETIC VAPORIZER

Installation Procedure for your Disco Double Kit, by Airgun Lab, LLC.

Patient Comfort. Oxygen Therapy Compliance Uninterrupted oxygen therapy aids compliance of prescribed oxygen therapy.

MEDICAL BREATHING CIRCUIT SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

Hallowell EMC Vacuum / Scavenging Interface

Tank Blanketing Pressure Regulators RHPS Series

[MSA The Safety Company] TURBO-FLO. Blowers. Replacement Parts. Accessories. Masks. Exploded Diagrams. safetyscope. Product Portfolio for FABA

Instruction Manual. PE-25/32/37 Tapping Machines. POLYTAPP Valves

Chapter 9 Airway Respirations Metabolism Oxygen Requirements Respiratory Anatomy Respiratory Anatomy Respiratory Anatomy Diaphragm

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

Honors Physiology The Respiratory System

STANDARD FLOWMETER USER S MANUAL

Chapter 9 Accessories for the Kirby Morgan BandMasks

AUTOVENT 4000 VENTILATOR

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

anesthesia equipment CDS 9000 Small Animal Anesthesia Machine phone fax

Breathing Process: Inhalation

Respiratory Protection in Agriculture

Invacare Twilight II Nasal Mask

Operational Efficiencies

Respiration Lab Instructions

ATI Skills Modules Checklist for Oxygen Therapy

Airways and Resuscitators. CRC 330 Cardiorespiratory Care University of South Alabama

NITRONOX SCAVENGER SYSTEM USER S MANUAL / INSTRUCTIONS

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Keeping anaesthetic pollution to minimum in theatre environment

The Anesthesia Machine (con t) The primary cause of machine malfunction is failure to check Never start without the American Express items What is the

Installation and Maintenance Manual. Compact Medical Gas Outlets

SCUBAPRO Repair Guide. S600-S550 Second Stages

1. Connect the attached anesthesia mask to your anesthesia system using the included tubing and connectors.

98695A01 - THE CONCEPT OF THE DIVATOR FAMILY

2) an acute situation in which hypoxemia is suspected.

Combination Breathing Apparatus

Hallowell EMC Anesthesia WorkStation Operating Manual

Breathing systems. Non-rebreathing system using a unidirectional valve pair. Non-rebreathing systems

Standard: Knows the general structure and functions of cells in organisms

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

DIGITAL MXR-D FLOWMETER USER S MANUAL

PremAire INSTRUCTIONS FOR VORTEX TUBE MODE OF OPERATION

Welcome to Maxstar Medical

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

NARKOVET ANAESTHETIC UNIT

Operation Manual. Carbon Filtration Industrial Drive, Orlinda, TN fax

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

Balanced SCUBA Second Stage Regulator (P/N ) Maintenance Manual Contents

Model 687 Mouse Ventilator User s Manual. Model 687 Mouse Ventilator

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Operation and Maintenance of the EPV200 Portable Ventilator

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

International Journal of Science, Environment and Technology, Vol. 5, No 5, 2016,

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

L9 Frog Dissection- External Page 1 of 7 STUDENT LABORATORY PACKET. Student s Name Modified from Prentice Hall Lab. Manual Lab Instructor Date Points

Training. Testor Training Manual

SOTR Special Operations Tactical Respirator

PORTER INSTRUMENT COMPANY, INC.

European Technical Standard Order (ETSO)

INSTRUCTION MANUAL EBA-75 CLOSED-CIRCUIT ESCAPE RESPIRATOR (CCER) CAPACITY 3

INSTRUCTION MANUAL FOR EBA MINUTE SELF-CONTAINED SELF RESCUER (SCSR) Manual Number: NH13747 Revision: A, ECN 2384 September 7, 2001

Print PERSONAL PROTECTIVE EQUIPMENT

with O 2 Controller Instruction Manual

SurgiVet Universal CDS 9000 Small Animal Anesthesia Machine Operation Manual

Breathing apparatus. Compulsory Specification for. Published by Government Notice No. R. 370 (Government Gazette 25040) of 20 March 2003

MXR FLOWMETER USER S MANUAL

RG1200 Service and Repair Manual

Ocenco Incorporated. M-20 Self Contained Self Rescuer Instruction Manual

Effective 2/1/17. SC1 Service Call PER Visit $76.00

Part II. Under Construction Station Instructions. Lab Station A - Blue Print: There is O 2 Here!

Transcription:

URN (UNIVERSAL RODENT NOSE CONE) 78914727 (17mm diameter) Figure 1 URN NRB SYSTEM (UNIVERSAL RODENT NOSE CONE NON-REBREATHING SYSTEM) 78909584 (17mm diameter) Figure 2 SMALL RODENT FACE MASK for use with URN NRB System body 78909681 (25mm diameter) Figure 3 RODENT FACE MASKS Small 78909681 Medium 78909682 Large 78909683 Extra Large 78909684 Figure 4 MAPLESON-D NRB With Small Rodent Face Mask 78909636 Figure 5 Page 1 of 6

Face Mask / Nose Cone / URN / Mapleson D NRB User Instructions Application: The URN (designed to be used with a coaxial non-rebreathing system), URN NRB system (a coaxial non-rebreathing system), Small Rodent Face Mask, and Mapleson-D NRB are designed to be used primarily with mice and rats. These devices are usually used in conjunction with an inhalant anesthesia system; however they could be used to administer oxygen and/or other metabolic gases without inhalant anesthesia (for example if the subject were anesthetized with an injectable anesthetic). However, any species can be safely anesthetized with either system as long as: 1. The subject's muzzle and/or breathing apparatus can fit within the nose cone or small face mask. 2. The reservoir of fresh gas is large enough to meet and/or exceed the tidal volume of the subject. 3. The URN and Small Rodent Face Mask is attached to a properly configured coaxial NRB system. (Example: Jackson Rees, Bain, Mapleson D, Ayers T piece, the URN NRB system body, and the Mapleson-D NRB.) 4. The proper oxygen flowrate is maintained to ensure no buildup of CO2 in the fresh gas reservoir. CAUTION: THESE PRODUCTS CONTAINS NATURAL RUBBER LATEX WHICH MAY CAUSE ALLERGIC REACTIONS IN SOME INDIVIDUALS. ACCESSORIES INCLUDED WITH URN AND URN NRB SYSTEM 78914975 O-Ring Diaphragm material 1.25 ft. Figure 6 Figure 7 ACCESSORIES INCLUDED WITH THE RODENT FACE MASKS 78914971 O-Ring for 78909681 Small Face Mask 78914972 O-Ring for 78909682 Medium Face Mask 78914973 O-Ring for 78909683 Large Face Mask 78914974 O-Ring for 78909684 Extra Large Face Mask All face masks include 1 ft. of extra Latex diaphragm material Page 2 of 6

Optional Accessory 78917456 A 6 inch wide X 36 inch long sheet of 12 mil latex diaphragm replacement material. Figure 8 Figure 9 Figure 10 Figure 11 Legend for Figures #9, #10, and #11 (URN NRB and Mapleson-D NRB systems) #1. Silicone "O" ring 78914975 for URN and URN NRB System. Note: Each URN and URN NRB System comes with an extra "O" ring. #2. Silicone "O" ring 78914971 for Small Face Mask (78909681) 78914972 O ring for Medium Face Mask (78909682) 78914973 O ring for Large Face Mask (78909683) 78914974 O ring for Extra Large Face Mask (78909684). Note: Each Rodent Face Mask comes with an extra "O" ring. #3. 12 mil Latex Diaphragm. Note: Each URN, URN NRB System, and Small Rodent Face Mask comes with an extra 6 inch X 12 inch sheet of Latex Diaphragm material to facilitate replacing defective diaphragms. Note: Extra diaphragm material may be purchased separately. 78917456 (6 inch X 36 inch roll). #4. Fresh Gas Feed Tube (note direction of flow of fresh gas marked by white bold arrow). Note: This component cannot be purchased separately. #5. 3.0mm male X 15mm male endotracheal tube adapter 78007275. This adapter plugs into the 15mm Common Outlet (Fresh Gas Outlet) on the inhalant anesthesia system. Note: This item may be purchased separately. #6. Fresh gas reservoir of URN and Small Rodent Face Mask. #7. Body of URN NRB System. The blue 19mm evac tubing fits on this body to collect and manage Waste Anesthetic Gases. Note: This component cannot be purchased separately. #8. Blue 19mm corrugated evac tubing. Note: This component may be purchased separately, (78914311). Page 3 of 6

Principle of Operation: 1. The URN NRB and Mapleson-D NRB Systems are coaxial flow systems in which the fresh gases flow towards the subject via the fresh gas feed tube (see bold white arrows for direction of flow). The exhausted gases (CO2) and other unused waste anesthetic gases flow away from the subject towards the Waste Gas Management System (see bold striped arrow for direction of flow). 2. The flowrate of fresh gas is set relatively high in relation to the subject's tidal volume. This is done to ensure that the exhaled CO2 from the subject is flushed towards the Waste Gas Management System. 3. Suggested minimum flowrate for a mouse: 500cc/minute. 4. Suggested minimum flowrate for a rat: 1 lpm. CAUTION: Using fresh gas flowrates lower than the suggested flowrates may result in CO2 buildup within the fresh gas reservoir. 5. Since the anesthetic gases and the typical carrier gases (Example: 100% Oxygen, Medical Air, Nitrous Oxide, Nitrogen, etc.) are extremely volatile, the fresh gas flowrate is used to continually flush the waste gases into the Waste Gas Management System. It is vital to the proper operation of the NRB System that the diaphragm of the URN and the Small Rodent Face Mask be cut with the proper sized orifice, appropriate for the size and position of the subject. Since rodents are obligatory nose breathers in the sternal recumbency, it is not necessary to insert more than the subject's nose into the fresh gas reservoir. The diaphragm can be cut appropriately (see section entitled "Replacing the Diaphragm") using a pair of delicate sharp/sharp scissors (see Figure 21). a. The circular hole in the diaphragm needs to be small enough such that the diaphragm forms a tight seal around the subject's muzzle. b. The circular hole in the diaphragm needs to be large enough so that it is not too tight and occludes the nares preventing proper spontaneous breathing. CAUTION: Do not cut a cross (+) or an X in the diaphragm for the subject s nose because diaphragm will not seal properly around the subject s muzzle and anesthetic gases will escape into work space. 6. If the rodent is in the dorsal recumbency, it is recommended that the subject's nose and mouth be inside the fresh gas reservoir. When the subject is on its back, there is a risk of the subject breathing through its mouth. If the mouth is outside the fresh gas reservoir, and the subject is allowed to breathe room air through its mouth, the level of anesthetic may be diluted by the room air and the subject may not be sufficiently anesthetized. In the dorsal recumbency, we recommend that the subject s nose and mouth is within the fresh gas reservoir. 7. If the rodent is on its side, there may be the possibility of mouth breathing. Our recommendation is that the subject s nose and mouth be inside the fresh gas reservoir. Page 4 of 6

Replacing the "O" Ring and the Diaphragm: 1. The silicone "O" ring holds the diaphragm material in place on the URN and Rodent Face Mask fresh gas reservoir. A. Silicone is more resistant to photo degradation and oxidation than latex. However, when it becomes cracked or broken and will no longer hold the diaphragm material in place, it is time to change the "O" ring. An extra "O" ring comes with each kit. Additional "O" rings can be purchased separately. 2. The "O" ring can be removed by rolling it from the groove it rests in. Using your thumb, press down and roll the "O" ring to the front to dislodge it from the groove (see figure 12). A. The defective diaphragm material will be discarded and the new diaphragm material installed. 3. The diaphragm is made of 12 mil thickness latex sheet material. This latex is much thicker, more durable, and more resistant to tearing; and more resistant to oxidation than surgery glove material (2-3 mil thickness). We do not recommend that you replace the diaphragms with surgery glove material. For the reasons stated above, and because the surgery glove latex material may be more permeable to the anesthetic gas molecules, there is a possibility of trace anesthetic gases escaping into the work space if surgery glove material is used. Extra diaphragm material comes with each kit. Additional diaphragm material can be purchased separately. 4. The defective diaphragm material can be replaced using the following processes: a. Roll the "O" ring out of the groove of the URN or Rodent Face Mask fresh gas reservoir (see figure 12, 13, and 14) (if not already done). b. Discard the flawed diaphragm material (if not already done). c. Stretch the new diaphragm material over the end of the URN or URN NRB System or Rodent Face Mask fresh gas reservoir (see figure 15). i. Hold the new latex diaphragm material with thumb and forefinger of one hand (see figure 16). d. Reinstall the "O" ring over the new diaphragm material and allow it to rest in the groove in the fresh gas reservoir (see figure 17). i. Make certain that there is no gap around the periphery of the diaphragm where trace anesthetic gases might escape (see figure 18). e. Pull the diaphragm material around the edges to smooth out the diaphragm and create a slight tension on the diaphragm (see figure 18). f. Using a pair of delicate sharp/sharp scissors, cut off the excess diaphragm material from around the "O" ring. Save the rest of the new diaphragm material for subsequent diaphragm replacements (see figures 19 and 20). g. Using delicate sharp/sharp scissors cut a circular hole in the diaphragm the appropriate size of the subject s nose and/or nose and mouth. (see figure 22,23,24). i. Position the circular hole in the diaphragm such that when the subject's nuzzle is placed in the diaphragm, the subject's head is relatively level with its body. The URN, URN NRB, and Mapleson-D NRB fresh gas reservoir can be rotated on its axis to ensure that the orifice in the diaphragm is in the proper position. Page 5 of 6

Figure 12 Figure 13 Figure 14 Figure 15 Figure 16 Figure 17 Figure 18 Figure 19 Figure 20 Figure 21 Figure 22 Figure 23 Figure 24 Page 6 of 6