Biometry tips. Marilyn Watkins-Ramdin, COMT ROUB

Similar documents
Ophthalmology Clinical Pearls A Quick Guide To Crystalens

LENSTAR LS 900 Setting the standard in optical biometry

Scleral Fitting Guide

Getting Started GUIDE LEAVE A LEGACY OF EXCELLENT OUTCOMES FOR PATIENTS WITH ASTIGMATISM. TECNIS TORIC 1-PIECE IOL pg 1

Sofmed breathables XW (comfilcon A) Sofmed breathables XW premium (comfilcon A) Sofmed breathables XW toric (comfilcon A)

Appendix C: Skill Stations Examination - Marking Rubric

TONOMETRY AND PACHYMETRY

LENS EXCHANGE WITH IMPLANTATION OF A BITORIC INTRAOCULAR LENS IN EYES WITH SIGNIFICANT ASTIGMATISM AFTER PREVIOUS CORNEAL SURGERIES

Acknowledgments. Paragon Vision Sciences Adoption of CRT certification slide for this generic talk. Material Considerations

LASIK Wavefront Guided idesign LASIK with IntraLase

Pre- and postoperative refraction after cataract extraction with implantation of standard power IOL

Orthokeratology. What is Orthokeratology? How does it work?

Appendix C: Skill Stations Examination - Marking Rubric

POLYMACON HYDROPHILIC CONTACT LENSES PRACTITIONER FITTING GUIDE

Dr. Thomas R. Reim President, DreimLens, Inc.

An introduction to modern scleral lenses

How images are formed in the eye

EVERY TREATMENT IS A TRUE ORIGINAL

The BE Retainer System Certification Course. Supplement

V E J A R A N O P R O T O C O L FOR T H E M A N A G M E N T O F Q U E R A T O C O N O SURGICAL T I PS

Pardon the Objection (PTO) Contact Lens Clinical Case Debate (2 hours)

FITTING AND TROUBLESHOOTING SCLERAL LENSES

Fitting Guide. Minutes-to-Fit 2.0. by SynergEyes Moving Vision Forward

Basic Concept & Techniques of Phaco Prechop

Financial Disclosures Toric IOL Selec-on and Posi-oning

NAVAL SUBMARINE MEDICAL RESEARCH LABORATORY

NormalEyes Guidelines For Successful Fitting RIGID GAS PERMEABLE SCLERAL CONTACT LENSES. Manufactured in Paragon HDS 1OO (paflufocon D)

Alberta Police Recruit Selection Standards

THE TOP 100 KEYWORDS FOR AN

Injector system for implantation of 1stQ foldable intraocular lenses

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR) ICO-OSCAR Instructor Directions

Learn Connect Succeed. JCAHPO Regional Meetings 2017

DISCLOSURES FITTING AND TROUBLESHOOTING SCLERAL LENSES WHO S A GOOD CANDIDATE SCLERAL LENS ADVANTAGES ANDREW J. BIONDO, OD, FSLS.

In November 1998, the Staar AA4203 toric intraocular. Early rotational stability of the longer Staar toric intraocular lens. Fifty consecutive cases

Procedures for ETDRS 4M Refraction and Vision Testing. Katherine M. Burke, BS, CO, COMT CertifEYED Associates, LLC New York, New York

REFRACTIVE ERRORS in. CORRECTION of OPTICAL DIVERS USING OPTICAL MASKS

New Highly Productive Phased Array Ultrasonic Testing Machine for Aluminium Plates for Aircraft Applications

PACKAGE INSERT RIGID GAS PERMEABLE CONTACT LENSES FOR ORTHOKERATOLOGY. (hexafocon B)

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Phacoemulsification and foldable posterior chamber intraocular lenses in aircrew: Clinical behaviour and aeromedical implications


Training Guide STAR S4 IR Excimer Laser

Management of corneal wounds: some practical tips. Lee SH, MD, FRCS (C ), FAAO

Understanding How the Appearance of Optical Fiber Splices Relates to Splice Quality

Practical method to calculate post-lasik corneal power: the Actual K a+p method

Operating manual Analogue refractometer

Spherical Soft Contact Lens Fitting. J. Perrigin OPTO 6374 Spring 2017

INTRASTROMAL CORRECTION OF PRESBYOPIA AND CLINICAL APPLICATION WITH TECHNOLAS FEMTOSECOND LASER

Andre Litwin BM FRCOphth Consultant Ophthalmologist and Oculoplastic Surgeon

Reusable Ophthalmic Instruments

Sterile, Single-use, Instruments

TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Ocular hazards of playing squash

Ophthalmic Surgical Instruments

International Council of Ophthalmology s Ophthalmology Surgical Competency Assessment Rubric (ICO-OSCAR)

Triple EK/CE/IOL has a cataract needs both DMEK and CEIOL

Level MEASUREMENT 1/2016

CLINICAL SCIENCES. Michael Goggin, FRCSI(Ophth), MS; Sacha Moore, FRCOphth; Adrian Esterman, AStat, PhD

OPESAVER Super Irrigation System

485 Annubar Primary Flow Element Installation Effects

PROFESSIONAL FITTING AND INFORMATION GUIDE. Paragon CRT Manufactured in Paragon HDS (paflufocon B)

Operating manual Analogue Brix Refractometer

RECOMMENDER REQUIRED PRE-AUTH GROUP A PRE-AUTH GROUP B

Figure 1 Schematic of opposing air bearing concept

FL M FLUME FOR HYDRAULIC BENCH

UsER manual for Watersens ph -REDOX

UNITED STATES MARINE CORPS WEAPONS TRAINING BATTALION MARINE CORPS COMBAT DEVELOPMENT COMMAND QUANTICO, VIRGINIA

James T. Schwiegerling CURRICULUM VITAE

ASICO Innovative I/A Tip System (Patented) Maximize your I/A efficiency for Today s MICS and FLACS procedure

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Inspection of CANDU Reactor Pressure Tubes Using Ultrasonics

LaserSnake Underwater Laser Cutting Phase 1 TWI

Ophthalmic Surgical Instruments

Aerodrome Design Manual

^- ; asefehs. a»s.^r,--. - "SSSES. v^yj:-- MEDIGÄL. -^j~ ^ z<_^

Outline Chapter 7 Waves

BIORAPTOR Knotless Suture Anchor Arthroscopic Shoulder Instability Repair. Shoulder Series Technique Guide by: Thomas Youm, MD

Single-use Ophthalmic Instruments

CATALOG CATALOG. Pioneer in Ophthalmic Products. BACK FLUSH INSTRUMENTS VITREORETINAL INSTRUMENTS

25+ instruments answer the most common barrier to 25G instrument flexibility.

Single-use Ophthalmic Instruments

Keeler Digital Applanation Tonometer (D-KAT)

Instrumentation & Data Acquisition Systems

Waves. Kevin Small or

AE3610 Experiments in Fluid and Solid Mechanics OPTICAL/PRESSURE MEASUREMENTS IN SUPERSONIC FLOW

AllPro Foot. Posterior Mount Product Manual

PAPER 2 THEORY QUESTIONS

Insert to. Sponsored by Alcon Laboratories, Inc. February 2014 EXPLORING THE CATARACT REFRACTIVE SUITE FEATURING. The CENTURION Vision System

LESSON 9 THE SWING PLANE AND THE BODY (Part 1)

CONTROL VALVE TESTING

Oxygen tension under a contact lens. Kenneth A. Poise and Marianne Decker

Post-Standardisation

Aeris Activity. Product Manual

Measurement of court speed and bounce. Rod Cross, Physics Department, Sydney University 2nd July 2006

Underwater measurement of photosynthetically active radiation

SMILE for low Myopia

OP CHECKLIST FOR 1D CONSOLIDATION LABORATORY TEST

CHAPTER 7 : SMOKE METERS AND THEIR INSTALLATIONS

Rotator Cuff Solution. Combined Strengths. Precise Control. TWINFIX Ultra FOOTPRINT Ultra Suture Anchors

Transcription:

Biometry tips Marilyn Watkins-Ramdin, COMT ROUB

Types of Biometry Ultrasound (A-Scan) Contact Immersion Laser Interferometry (IOL Master)

Patient History Use the oldest possible MRx to include for calculations since myopic shifts up to 10 diopters can be caused by the changes in the lens. Necessary for Holladay 2 formula Ask about prior refractive surgeries: LASIK, PRK these can change K readings and require special calculations

Patient History Remember Myopic eyes are usually longer 25.16+/- 1.23mm Hyperopic eyes are usually shorter 22.62 +/- 0.6m Emmentropic eyes are usually 23.5mm long

Patient History con t Retinal surgery: Scleral buckles can elongate the eye 0.5-1.00mm Pseudophakic patients find out what IOL material was used: PMMA, Silicone or Acrylic Aphakic: IOL setting must be adjusted PK or corneal opacity Slit lamp exam: Ocular surface disease, Eg. Severe dry eyes

Patient History con t Ocular history: glaucoma, prior surgery, eye meds Medical history: prostate; leads to floppy iris syndrome if patient on Flomax Social history: job type, social activities, drug use

Keratometry Manual Automatic: Autorefractor / IOL Master Pick a method and stick with it Use two methods for K readings to verify Eyes should be within 1D of each other NB: Myopic eyes are long and have flat OA's Hyperopic eyes are short with steep K's Average eye is 23.5mm Average K reading 43.00-44.00D

Keratometry P=(n-1)/r P= power n=standardized keratometric index of refraction usually 1.3375 The index of refraction refers to theoretical single refractive lens with a stable ratio of anterior to posterior corneal curvature

Keratometry con t

Keratometry con t Manual keratometry considered the Gold standard seldom used due to better automated keratometers Always focus your eye piece first as in lensometry since uncorrected refractive errors of the examiner will result in erroneous measurements

Keratometry con t Keep both eyes open when measuring Focus the horizontal meridian first then measure Then refocus mires to measure the vertical axis Much needed in high myopes NB 1D error K s =1D post-op refractive error

Keratometry con t CL s wearers: Soft CL wearers should take lenses out for about one week prior to measurements GP and HL should be out until K s are stable NB. Prior to K reading no procedures ointments or drops are to be used that will distort mires 1 drop of anaesthics GTTs can be used to smooth mires Calibrate weekly

Topography Topography is used to detect any deformities of the corneal surface e.g. Keratoconus, PK, RK The cornea is provides 70% of the eye s refractive power Keratometry measures the central 3mm of cornea VS Topography measuring entire anterior corneal surface More accurate for corneas << 40D or >> 46D & irregular Best when used in conjunction with Keratometry. An average taken between when K s are disparate much needed when measuring for toric IOL s

A-Scans Immersion - You will need scleral shell : Hansen or Prager BSS & anesthetic gtts Position patient as for contact method Have the machine in a position where you can see and easily access screen as for contact Can be done manually or automatically Instill gtts

A-Scans Contact vs Immersion Contact Patient must be comfortable and informed of what you are going to do Patient can be sitting or reclining Instill anesthetic gtts Place probe directly on central cornea Align probe along visual axis come off cornea before making any major movements other than tilting

A-Scans Contact vs Immersion Immersion Insert shell with probe attached for Prager or in the case of Hansen shell just insert into eye Immerse probe in BSS Align probe along visual axis Lateral movements are used to obtain measurements

A-Scans Contd. For either method take 10 readings per eye Immersion is more accurate but difficult at first to learn Messy at first Takes longer at first Contact Corneal compression Corneal abrasion Fluid bridge Dry Eye

Sources of Error You may have some situations that require contact so perfect the skill Causes of errors Short measurements Corneal compression Velocity too low Improper gate placement Gain too high Misalignment of sound beams

Errors Long measurements Fluid bridge Velocity too high Improper gate placement Gain too low Misalignment of sound beam Improper techniques Poor patient fixation: Strabismus, NLP in one eye Internal vs External fixation Pathology: disciform scar or saphyloma

Macula Localizing Technique Find optic nerve Tilt probe to direct sound until scleral spoke shows

Laser Interferometry IOL master uses light to measure axial length Advantages Non-contact Very accurate Disadvantages Cannot penetrate dense cataracts Difficulty measuring Pseudophakic eyes due to high degree of reflection

Axial Length Measurements Good fixation Good corneal tear film Not a fool proof method Nuclear cataracts may require off-axis measuring technique Dense opacity of posterior capsule may produce error message Ametropia of more than 6D may require patient to wear glasses for best results

IOL Master Useful for measuring eye with posterior staphyloma (protrusion of sclera causing an irregular shaped posterior eye) this is because the axial length measurements is required along the axis of patient fixation NB: when using an A-Scan proper alignment is at best estimated with a staphyloma

Corneal Curvature Measurements K Reading's Susceptible to the same error factors as the keratometer Poor tear film Corneal irregularly Image obscured by eyelids or eyelashes Improper focusing

Anterior Chamber Depth Many of the IOL calculation formulas use ACD as part of calculation IOL master uses a projected slit beam to measure ACD Excessive accommodation by patient can cause divinations, if so measure after dilation ACD will NOT be accurate on a Pseudophakic eye

IOL Calculations & Formulas 4 primary pieces of information needed for IOL calculations: A-Scan axial length measurements Average K readings A-constant for IOL to be implanted Desired post-op refraction Formulas: Haigis, HofferQ, Holladay, SRK II, SRK/T Can be linked to Holladay II program and other network system Know which formulas are most appropriate for the measurements you acquired.

IOL Calculations & Formulas Both eyes must be measured to identify any potential sources of error IOL power should be within 1D between both eyes Theoretical formulas 1 st Gen Fyodorov, Binkhorst, Holladay Shammas used A-scan ultrasound and K readings to devise these formulas Are all based on the same equation P=n/L-ACD-n x K/n-K x ACD Regression formulas SRK P=A-2.5L-0.9K Gave better results with normal eyes than the theoretical formulas

Formulas 2 nd Gen formulas were used to improve the accuracy of IOL calculations for short & long eyes Increase in PIOL vs Iris-plane IOLs a factor SRK altered by combining a linear regression analysis for short and long eyes this became SRKII 3 rd Gen formulas arose after doctors express discomfort with a formula based on an artificial linear model. The authors of SRK devised then SRK/T an new theoretical formula. SRK/T is similar to the Holladay & relies on the ACD constant rather than surgeon factor

Formulas 4 th & 5 th Gen formulas are: Holladay,Haigis, Olsen & Barrett Universal II These can use as many as seven variables (keratometry, axial length, anterior chamber depth, lens thickness, horizontal white-to-white measurement, age and pre-op refraction in the case of the Holladay II

Formulas To get the most from many of the latestgeneration formulas, you also need to have an optical biometer such as the IOLMaster 700 or the Lenstar. This is because two of the newer formulas, the Holladay II and the Olsen, require a measurement of the lens thickness that these instruments provide

Clean & Calibrate IOL Master Clean Chin rest and head rest with Alcohol wipes between patients Once per week calibrate with test eye if available or staff member with stable Rx Ultrasound Machine Soak shells in Hydrogen Peroxide prior and after each patient but rinse with water prior to inserting into the eye Probe tip can be soaked in peroxide or wipe with alcohol Calibrate using a staff member

Summary For successful biometry Examiner must have knowledge of proper techniques High quality consistent scans IOL calculations Biometry standards Have a 2nd examiner re-measure patient anytime you are unsure Triple check your work Never stop learning!

NB: take home points For successful post-op outcomes take into consideration : The surgeon s expectations The patient s expectations The increasing use of Premium IOLs The world of biometry is heading to optical vs ultrasound with the increasing demands for optimum visual outcomes

References/ Resources A-Scan -Axial Eye Length Measurements A Handbook For IOL Calculations- Sandra Frazier Bryne Grove Park Publishers https://www.haag-streit.com/haag:streitusa/product/haag-streitdiagnostics/lenstar-biometry/ https://quizlet.com(roubflashcards) www.ophthalmicedge.org

References/ Resources www.jcahpo.org www.atpo.org https://www.reviewofophthalmology.com /article/power-calculation-how-to-upyour-game Ultrasound of the Eye & Orbit 2nd Sandra Frazier Bryne Ronald L. Green Mosby Inc ISBN 0-323-01207-8

THANK YOU!