HPLC troubleshooting guide

Similar documents
Acclaim RSLC. (Rapid Separation Liquid Chromatography)

LC, RSLC, Bio RS and Dual Gradient Systems

no peaks missing peaks too small peaks increasing retention decreasing retention

UltiMate 3000 Viper Capillary Kits for Intelligent LC (LCi) System Solutions

SOP: Buck Scientific BLC-20P HPLC Operation

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

Setting a New Standard in Flash Chromatography Performance

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Figure 1: Plug in ion inlet of mass spec

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

AKTA pure 25 New Owner s Intro

Running HPLC Methods on an ACQUITY UPLC System

GCMSD-Headspace Analysis SOP

DEBUBBLERS & DEGASSERS ACTIVE DEBUBBLERS APPLICATION NOTE WHY DEGAS YOUR MOBILE PHASE? PAGE 177 STAND ALONE DEGASSERS PAGE 175 PAGE 178

Mass Spec will not Autotune

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Dionex IonPac AS19-4µm

Rapid and Reliable Detection of Dissolved Gases in Water

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Autosampler problems!

Standard Operational Procedure for Capillary Electrophoresis

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

Rev. CO Packed Columns for Normal Phase Chromatography. TSKgel NH m INSTRUCTION MANUAL

The instrument will process the samples one by one till all the samples have been cleaned up.

Carrier Gases in Capillary GC

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Ascentis Express HPLC Columns with Fused-Core Technology

MV5: Automated concentration system

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

HPLC. Solvent Delivery Module Systems

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Agilent Dimension Software for ELSD User Manual

Thermo Scientific Chromatography Well Plates

Thermo Scientific Chromatography Well Plates

Connectors. SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses. Top Seller SEE STARRED PRODUCTS. VHP Tees & Crosses for Capillary Tubing

Best Practice for Identifying Leaks in GC and GC/MS Systems

Columbus Instruments

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Extended Application Note

Advances in Capillary Column Backflushing

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Using the Akta Prime plus October 22, 2012

More Speed and Resolution for Current Methods

HPLC systems. Overview

Video Notes. LC Troubleshooting Series Retention Time Shifts

BETTER SAMPLE CARE. Spark Holland introduces. SPH1299 UHPLC Gradient Pump

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Peak Biotech - Company & Product Presentation

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

CORESTA RECOMMENDED METHOD Nº 67

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

DOS in Food Science & Nutrition, Manasagangotri, Mysore

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

This test shall be carried out on all vehicles equipped with open type traction batteries.

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

1999 Microcomponents Technology SLYA017A

The HumiPyc - Model 1 - Gas Pycnometer; Density, Moisture, Permeation Analyzer; RH sensor Calibrator

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Indian Institute of Technology Kanpur Environmental Engineering and Management Department of Civil Engineering

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

FOURIER TRANSFORM INFRARED SPECTROSCOPY

Troubleshooting. Select the Table of Contents here Rheodyne All rights reserved

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

MMS Mercury Monitoring System for Natural Gas

FLUORESCENCE DETERMINATION OF OXYGEN

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Agilent 1220 Infinity II LC Mobile Upgrade Kit

Residual Gas Analysis Systems for Industry

SOR-100 Solvent Racks

Constant Pressure Inlet (CCN) Operator Manual

Customer Responsibilities

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Prep LC Training Manual

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Customer Responsibilities

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

Transcription:

CASE STUDY 70303 HPLC troubleshooting guide A free app that will save time and frustration Introduction During the development of a new HPLC method, there were unexpected results all peaks, except for the injection peak, were negative (Figure 1). Even with advances in chromatography to eliminate problems, something unexpected is bound to happen. How would you go about finding out what s wrong? Although the web is full of HPLC troubleshooting guides, many of these are dozens of pages long. 1,2,3 They can be helpful in teaching systematic approaches to troubleshooting, but when you need an answer fast, digging through these resources to find what you need can be slow and tedious. Thermo Scientific HPLC Troubleshooting Guide an app for iphone, Android, and Windows platforms.

100 0 mau -300 0 5 10 Minutes Figure 1. Negative peaks during method development experiments. An alternative is using a poster, such as the Thermo Scientific HPLC Troubleshooting Guide poster, 4 where all troubleshooting suggestions are available at a glance and a useful suggestion can be found much faster. An alternative for people with smart phones is the Thermo Scientific HPLC Troubleshooting Guide an app for iphone, Android, and Windows platforms. This mobile application incorporates all the information from the HPLC troubleshooting poster, and expands it to include electrochemical and nano LC solutions. In combination with solutions for charged aerosol detection, these additions make the troubleshooting app a comprehensive tool for various system setups. This mobile application is available for free at thermofisher.com/troubleshoothplc System Thermo Scientific UltiMate 3000 x2 Dual System including: SRD-3600 Solvent Rack with six degasser channels DGP-3600RS Dual-Gradient Pump WPS-3000TRS Thermostatted Wellplate Sampler - TCC-3000RS Thermostatted Column Compartment - DAD-3000RS Diode Array UV-Vis Detector with analytical flow cell Conditions Conditions Column: Thermo Scientific Acclaim 120, C18, 5 μm, 4.6 100 mm Standard: Caffeine, anthracene, and phenanthrene (0.05 mg/ml) in 70% ACN Eluents: A: 0.1% Formic acid in water B: 0.1% Formic acid in acetonitrile Gradient: Time (min) %B 0.0 60% 5.0 90% 5.5 90% 6.0 60% 10.0 60% Flow: 1 ml/min Column temp.: 30 C UV wavelength: 285 nm UV bandwidth: 4 nm Thermo Scientific Chromeleon Chromatography Data System (CDS) All modules were connected using 0.005 in. (0.13 mm) i.d. Thermo Scientific Viper Fingertight Fitting System 2

a) b) c) d) Figure 2. Generating probable cause using the Thermo Scientific HPLC Troubleshooting Guide. Troubleshooting Case Study A Few Clicks to Find the Probable Cause Using the problem of negative peaks described above, the HPLC Troubleshooting Guide finds a solution as follows: From the app s home screen, select Start Troubleshooting (Figure 2a). Because the issue is related to peaks, select Peaks (Figure 2b). Choose Negative peaks (Figure 2c). In this example, results were obtained digitally by DAD. Possible causes 2, 4 and 5 in Table 1 relate to analog data acquisition, CAD and FLD, therefore, do not apply. This leaves two other possible causes. Starting with cause 3: Inappropriate reference wavelength setting (DAD), the HPLC Troubleshooting Guide shows that the related solution for an inappropriate reference wavelength is (Figure 3): The sample must not absorb in the range of the reference wavelength. If possible, use a method without reference wavelength. The app lists five Possible cause/solution (Figure 2d), shown in Table 1. Table 1. List of Possible Causes Generated by HPLC Troubleshooting Guide. Possible Causes 1. Absorption/fluorescence of analyte is lower than mobile phase 2. Wrong polarization of analog output interface 3. Inappropriate reference wavelength setting (DAD) 4. Drainage spiking (Charged Aerosol Detection, CAD) 5. The fluorescence of the substance is quenched by matrix or mobile phase (FLD) Discussion of possible causes and solutions Negative peaks can be caused by the detector or by how the detection signal is processed. Consequently, the different causes are specific to the types of detectors. In this example, different causes in combination with UV detection (variable wavelength, VWD) or diode array detection (DAD), analog data acquisition, fluorescence detection (FLD) and charged aerosol detection (CAD) are listed and demonstrate the unique diversity of the Troubleshooting Guide. 3 Figure 3. The app showing probable cause Inappropriate reference wavelength setting (DAD), with the solution. Also note the app has the ability to share the solution with a colleague and links if you want to learn more.

For a UV-Vis DAD, you can select both a reference and a detection wavelength. When used properly, the reference signal corrects for changing conditions such as baseline drift. If the reference wavelength is not chosen correctly and absorbs in the range of your analyte, then you may get a negative response. (See the sidebar Did You Know? for more information.) In this case study, the reference and detection wavelengths for the analysis of butylparabene were selected incorrectly. With a reference wavelength of 240 nm and a bandwidth of 100 nm, the reference absorption was taken from 190 nm to 290 nm. With a detection wavelength of 285 nm and a bandwidth of 4 nm, detection absorption was taken from 283 nm to 287 nm. The range of the reference indeed overlaps in the range of analyte detection, making this the probable cause for negative peaks. To verify that this is cause, the reference wavelength can be switched to off. Reference: off is actually the recommended operation of Thermo Scientific Diode Array Detectors mainly to avoid implications with incorrect reference settings. The resulting chromatogram (Figure 4) shows that this change immediately solves the issue of negative peaks. Did you know? How reference and detection wavelengths work Reference wavelength The reference wavelength is intended to correct for (subtract) absorption that can occur from changes in conditions during analysis. Those changes can sometimes be significant, such as if there is a reduction in the lamp energy (lamp drift). The reference wavelength selected therefore must be in a quiet area of the spectrum where little to no absorption of the sample analytes occurs. Because the signal from the reference wavelength is subtracted from the signal received from the detection wavelength, it has a direct impact on results. Reference bandwidth Reference bandwidth serves to average several photodiode signals above and below the reference wavelength. A reference bandwidth should be selected as wide as possible, for example 30 100 nm, but narrow enough not to interfere with the absorption spectrum of the analytes. Figure 5 below visualizes appropriate reference settings for a butylparabene spectrum. 100 mau 0 0 5 10 Minutes Figure 4. Obtaining the chromatogram without reference (Reference: off) displays the peaks in a normal way and removes side effects such as baseline drift and injection peak. Note that the positive injection peak and the baseline drift are now gone. Both were caused by the absorption of formic acid as part of the mobile phase, affecting wavelengths below 260 nm. The injection peak was positive as the sample diluent lacks formic acid and therefore absorbed less than the mobile phase. With the initial, improper reference wavelength settings, the negative peak was inverted and shown as a positive peak. Figure 5. Butylparabene spectrum with recommended reference wavelength settings. UV-Vis Detection Wavelength In contrast to the reference wavelength, the UV- Vis detection channel wavelength for recording the analyte absorption should be set to the absorbance maxima of the analytes of interest. 5 4

2016 Thermo Fisher Scientific Inc. All rights reserved. PEEK is a trademark of Victrex PLC. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. XX70308-EN 1016S Conclusion A solution is just a few taps away; select your problem and the HPLC Troubleshooting Guide an app that will guide you to the answer. Comprehensive in its inclusion of specific technologies such as fluorescence detection, charged aerosol detection and nano LC, this powerful, free tool effectively diagnoses various potential issues with HPLC analysis. A Share function even lets you send the solution to your team. Simply check the troubleshooting solution of interest and select share, and enter the email address from your contacts list. Download your free HPLC Troubleshooting Guide The app is available for iphone, Android, and Windows phones, free at thermofisher.com/troubleshoothplc or scan this quick response (QR) code to download the app: References 1. Successful HPLC Operation A Troubleshooting Guide Version 1.1, Thermo Electron Corporation, 2004. https://assets.thermofisher.com/tfs-assets/cmd/ Product-Guides/TG-20421-HPLC-Troubleshooting-Guide-TG20421-EN.pdf 2. HPLC Troubleshooting Guide, Bulletin 826E, Sigma-Aldrich, 2009. 3. HPLC Troubleshooting Guide, Phenomenex Inc., 2008. 4. Thermo Scientific HPLC Troubleshooting Guide Poster (XX70308), 2016. 5. Thermo Scientific UltiMate 3000 Series Diode Array Detectors Operating Instructions, Revision 1.4. https://assets.thermofisher.com/tfs-assets/cmd/manuals/man-lc-dad- MWD-Operation-Sep13-DOC4820-8250-1-4.pdf HPLC Troubleshooting Guide quickly resolve application or system issues Peaks No peaks Instrument failure, wrong elution conditions Does the detector show typical baseline noise? If the detector output is just a flat line, the detector or the data transfer failed. Inject known test substance without column and check detector response. No injection, not enough sample volume Ensure sample is drawn into sample loop. Check that injection has taken place (pressure drop at the beginning of run). High background current/noise Check mobile phase quality. See baseline section. (Charged Aerosol Detection) Sample too volatile (Charged Aerosol Detection) Check response by flow injection analysis. Check vapor pressure of analyte(s). Peak tailing Basic compounds interact with silanol groups Use type B (high-purity) silica or shield phases (such as polar-embedded groups). Use competing base such as triethylamine (TEA). Alternatively (with type A silica columns), use polymeric columns. Use buffers of high ionic strength (displacement effects NOT compatible with LC/MS). Insufficient buffer capacity Increase buffer concentration. Chelation with trace metals in stationary phase As above. Too large of an extra-column volume Use short capillary connections. Check inner diameter of connecting capillaries: 0.13 mm (0.005 in.) for UHPLC columns and 0.18 mm (0.007 in.) for conventional HPLC columns. The extra-column volume should not exceed 1/10 of the smallest peak volume. Use Thermo Scientific Viper or Thermo Scientific nanoviper capillaries. Improper capillary connections Check fittings for correct placement of ferrule. Use Viper or nanoviper capillaries. Column degradation Replace column. With high temperatures: Avoid high temp in combination with aggressive buffers (e.g., phosphate) reduce to uncritical temperature (e.g., 40 C), replace column type (e.g., hybrid column). With high ph: Use high ph compatible column type. Column void (particularly at UHPLC pressures) Replace column. Try to flush reversed column (outlet to waste). Preventive actions: Slowly increase the column flow to avoid pressure shocks on the column. Avoid aggressive ph (check column specifications). Routinely operate columns at less than 70 80% of the pressure specification. Peak fronting Blocked frit, particles on column head Replace pre-column frit. If fronting returns quickly, try to locate source of particles (sample, eluents, pump mechanics, injection valve). Channels in column Replace column. Check if application conditions are within the column specifications (e.g., pressure and ph range). Column overload Reduce amount of sample. Increase column volume (use larger i.d.). Sample dissolved in strong eluent Dissolve sample in starting mobile phase. Reduce sample solvent strength or injection volume. Broad peaks Detector cell volume too large Flow cell volume should not exceed 1/10 of the smallest peak volume. Use smaller volume flow cell (i.e., micro or semimicro) with UHPLC or microbore columns. Early peaks broader than later eluting ones Extra-column volume too large. Check capillary i.d. and length, sample loop size, flow cell, etc. Detector response time (time constant) too long Select a response time less than 1/4 of the peak width at half-height of the narrowest peak. Use Thermo Scientific Chromeleon Chromatography Data System program wizard to optimize the response time (time constant) settings. Charged aerosol detection compared to e.g., Charged aerosol detection will broaden peaks more than UV due to nebulization. Use Viper or nanoviper capillary connections for UV detection zero-dead-volume connections by design. High longitudinal dispersion Isocratic separation: Retention time too long. Use gradient elution or stronger isocratic mobile phase; use less retaining stationary phase (e.g., C8 instead of C18). Linear velocity too slow: Check flow rate. Ghost peaks Only some peaks broad: late-eluting peak from Extend run time. Increase elution strength of gradient (higher organic content). At the end of the run, flush column with strong eluent. previous injection Contamination (typically injector or column) Flush sampler, replace parts prone to contamination (e.g., needle or needle seal). Flush column with strong eluent (also check column instructions; flush in reverse flow direction if possible). Contamination of eluents Water quality: Run different amount of water over the column and measure enrichment time as a function of the volume. Replace water with HPLC-grade water. Contaminations may also result from bacterial growth in the degasser, eluent modifiers (e.g., degraded TFA) or improper liquid handling (e.g., amino acid mobile phase prepared without gloves). Interferences in sample Use sample cleanup techniques such as solid-phase extraction (SPE). Contaminated nebulizer (Charged Aerosol Detection) Nebulizer chamber may need to be cleaned. Remove column. Wash your charged aerosol detector with 50/50 water:methanol at 2.0 ml/min for several hours. Adjust flow-rate to 0.2 0.4 ml/min overnight. Re-check noise level of original mobile phase composition. Split or double peaks Contamination on column or guard inlet 1. Replace guard. 2. Flush column with strong mobile phase, backflush to waste (if possible). 3. Replace analytical column. Sample solvent too strong Prepare/dilute sample in mobile phase. Coelution with unknown interference Perform efficient sample cleanup. Adjust selectivity by changing the mobile phase or the column. Worn out rotor seal Replace rotor seal. With extreme ph applications, check compatibility of the seal polymer. Packing integrity loss (UHPLC applications) Replace column. Check pressure rating of the column and work at maximum 70 80% of the pressure spec. Slowly increase pressure on columns to avoid pressure shocks. Temperature mismatch Mainly occurs with column i.d. s > 3 mm in combination with high column temperatures. Use an eluent pre-heater to ensure eluent temperature does not result in temperature gradient over column. Improper capillary connections Check fittings for correct placement of ferrule. Use Viper or nanoviper capillaries. Negative peaks Inappropriate reference wavelength setting (DAD) The sample must not absorb in the range of the reference wavelength. If possible, use a method without a reference wavelength. Use Viper or nanoviper capillaries, which provide zero-dead-volume connections by design. Absorption/fluorescence of analyte is lower than 1. Change UV/fluorescence detection wavelength(s). 2. Use mobile phase with less background absorption/fluorescence. mobile phase 3. Dissolve sample in mobile phase. Wrong polarization of analog output interface Check cable polarity at analog output. Drainage spiking (Charged Aerosol Detection) Ensure that PEEK tubing is present within the drain/vent assembly. learn more at thermofisher.com/liquidchromatography Peak Area Precision Caused by instrument Sample or a sampler problem Perform multiple injections to differentiate between sampler and sample related issues: Increasing retention times Mobile phase composition 1. Check premixed mobile phase, make sure the liquid is homogeneous. 2. Perform an OQ proportioning test. a) High-pressure proportioning: or sample 1. Variations of the sum of the peak areas Injector. Check for leaks or bubbles in aqeous eluent. b) Low-pressure proportioning: Check/adjust proportioning valve. c) UltiMate 3000 NCS/NCP 2. Only some peak areas vary Sample not stable. To verify: Inject known stable mixture peak areas should not vary. systems: Perform pressure sensor and viscosity calibration. 3. If peak areas of some peaks still vary, check system pressure stability and short-term flow stability. Active sites on stationary phase Use organic modifier base (such as TEA), increase buffer strength or use higher coverage column packing. Peak integration settings Positions of integration delimiters vary Check the software integration settings. For instance, with the support of the Chromeleon software help. Best, facilitate the Decreasing flow rate Leaking capillary connections. Check flow rate settings. Chromeleon 7 Cobra algorithm and let it identify the ideal settings for you. Avoid automatic data rate settings and use a fixed data rate. Scattering retention times Insufficient equilibration Increase equilibration times, at least 5 10 times the column volume. Fluctuating baseline Irreproducible integration caused by pump pulsation, Touch this field to be forwarded to Category: Baseline, Symptom: Periodic Baseline Fluctuation. Imprecise eluent proportioning Perform an OQ eluent proportioning test. Clean/replace pump check valves, or call service for adjusting/replacing proportioning valve. mixing ripple, etc. Insufficient buffer capacity Increase buffer concentration > 20 mm. Work at ph within buffering range. Injection volume variation Autosampler draws air from the vial Check the sample filling height and the sampling height of the injector needle. Changing column temperature Place column in column thermostat. Monitor oven temperature. Sample degradation Use appropriate storage conditions, e.g., thermostatted autosampler. No synchronization of injection with pump cycle Activate function in the software, for instance, in the Chromeleon pump driver. Air in the autosampler fluidics Flush out autosampler fluidics following the steps laid out in the respective operating instructions. (low-pressure proportioning) The autosampler, the injection valve, and/or the syringe Check seals and tighten fittings involved in the injection process of the sampler. valve are not tight Baseline Injector needle clogged or needle tip deformed Replace needle. Remove air from autosampler fluidics. Autosampler draw speed too high, sample with high Reduce the draw speed to take at least 2 3 seconds. Program a delay time after the drawing of sample (e.g., 3 seconds). Periodic baseline fluctuation Pressure fluctuations from pump Record a pump pressure channel to identify temporary or permanent pressure ripple. A pump influence is indicated by baseline gas content Degas sample and/or decrease draw speed. fluctuations synchronous with the pump cycle. Purge the pump, check general functionality (for instance, with a diagnostics test), read precompression value from pump panel and compare against values from manual (Thermo Scientific pumps only). Check that an appropriate Leaking injector seal, bubble in syringe Check injector seals, purge syringe. Inappropriate detector Wrong detection wavelength Measuring in a UV/fluorescence spectrum flank can compromise the precision. Choose a detection wavelength or an excitation/emission mixing volume for applied eluents/flow rate is selected. settings wavelength pair near the apex of the spectrum/spectra. If spectra of analytes are very different, a wavelength switch might be required. Air in the pump fluidics 1. Degas eluents, best with on-line vacuum degassing or helium sparging. 2. Purge the pump. 3. Actively draw liquid through the pump, for Response time too short high noise imprecise Make sure to use suitable response time (or time constant) settings. Typically, set a response time that is about 1/4 of the peak width at instance with a syringe at the outlet of the pump. 4. Temporarily replace the water with degassed organic solvent and purge the respective integration at trace level half-height of the narrowest peak. Follow the operating instructions for details. eluent line. Afterwards, switch back to degassed water. Incorrect nebulizer temperature (Charged Aerosol Check nebulizer temperature setting. If using large amounts of THF or halogenated solvents in mobile phase, set temperature to 30 C. Wrong reference wavelength selected (DAD) The sample must not absorb in the range of the reference wavelength. If possible, use a method without reference wavelength. Detection) If analyte is semivolatile, it may be necessary to turn off the nebulizer heater to recover response. Insufficient mixing Periodic, wave-like noise. Intensity of effect is related to the eluents, the type of detector, and the detector layout. Increase the mixing Analyte transfer from Carryover (from needle, sample loop, needle seat) Extend autosampler wash options to reduce carry over. Clean needle, needle seat, and rotor seal. volume depending on the manufacturer s recommendation. previous injection Mobile phase degraded Prepare fresh mobile phase and flush the system with it. If required, backflush the column (increased backpressure). Column memory effect Run blank sample (no injection) after sample. If a peak is detected, then a column memory effect is present. Improper grounding The complete HPLC system should be plugged into the same circuit. Isolate the electrical circuit from strong current consumer devices or Flush column with strong eluting eluent. Reverse column (if allowed) and flush column again. Replace column. use an isolation transformer to filter current fluctuations. Absorbing solvent Check mobile phase for degradation, particularly aqueous buffers. With DADs, switch off the reference wavelength during troubleshooting. Pressure Prepare fresh mobile phase. Improper installation of optical components Check if lamps and flow cell are positioned properly. Pressure too high/increasing Buffer precipitation 1. Check buffer compatibility with max organic content (in beaker). Non-periodic, high noise Mobile phase quality Make sure to use HPLC-grade or better mobile phase quality, particularly for sub-2 µm particle columns and evaporative detectors. Use 2. Flush out system and column with low organic content. mobile phase with lowest particle content whenever possible. 3. Systematically investigate fluidics to locate the blocked part (from detector to pump). Nonvolatile mobile phase (Charged Aerosol Detection) Make sure to use volatile buffers and additives with the Charged Aerosol Detection. Bent capillaries Systematically investigate fluidics to locate the blocked part (from detector to pump). Nebulizer (Charged Aerosol Detection) Nebulizer may need to be cleaned. Remove column. Flush the HPLC-Charged Aerosol Detection system with 80/20 water:methanol at Particles on mixer, frit, guard, or column Use pre-filtered mobile phases suitable for the column particle size (for instance 0.1 0.2 μm for sub-2 μm particles, 0.45 μm for 3 5 μm 2.0 ml/min for several hours. Adjust to 0.2 0.4 ml/min overnight. Re-check noise level with fresh preparation of original mobile phase particles). Filter samples accordingly. Check sealing mechanic parts (piston seal, rotor seal). composition and flow rate. Too high flow or too low column temperature setting Correct setting. Flooding of charged aerosol detector Perform self-test to ensure there are no errors present. Check to be sure all diagnostics are within specification. Column aging/blocking Gradual pressure increase is normal over the lifetime of a column. Unusual pressure increase: Check connection tubings, temperature Nitrogen gas supply (for evaporative detectors) Make sure to use 99% pure nitrogen, free from water vapor and particulates. Replace two-stage filter if necessary. control, replace column. Response time (time constant) setting too short Make sure to use suitable response time. Typically set a response time that is about 1/4 of the peak width at half-height of the Pressure changes with the course of a gradient Normal. The system pressure is dependent on the viscosity of the mobile phase composition in the system fluidics. increased noise narrowest peak. Follow the operating instructions for details. Blocked/clogged liquid inlet or nebulizer (Charged Check backpressure of the charged aerosol detector. Backpressure should be under 10 bar (145 psi). Optical bandwidth setting (UV detectors) too narrow Use a low optical bandwidth setting for best optical resolution. Increase the bandwidth setting for best signal-to-noise. Please refer to the Aerosol Detection) detector s operating instructions for details. Pressure too low Leaking capillary connections 1. Check fittings for tightness, particularly before the column. PEEK capillaries with fingertight fittings may slip out of the ideal position. 2. High baseline drift Causes as discussed in nonperiodic, high noise Solutions as discussed in non-periodic, high noise Charged Aerosol Detection. Use Viper or nanoviper capillary connections for zero-dead-volume connections by design. (Charged Aerosol Detection) Flow too low or column temperature setting too high Correct setting. Gradient operation (Charged Aerosol Detection) Minimize gradient baseline effects by using appropriate mobile phase quality. Use shallow gradient methods whenever possible. Pressure fluctuating Air trapped in pump 1. Degas eluents, best with on-line vacuum degassing or helium sparging. 2. Purge the pump against a gentle backpressure. 3. Actively Use inverse gradient. draw liquid through the pump, for instance, with a syringe at the outlet of the pump. 4. Temporarily replace the water with degassed organic Detector lamp/optics temperature not stable Allow the detector to warm up. Depending on the optical design, this may take 30 minutes to a few hours. See the detector s operating solvent and purge the respective eluent line. Afterwards, switch back to degassed water. instructions for details. Inlet or outlet check valves of the pumps are dirty Take out check valve cartridges, place them in a container with a suitable liquid (e.g., methanol) and clean them in an ultrasonic bath. Mobile phase not homogeneous After a day or more idle time, gently swirl the eluent bottle to homogenize eluents already in their reservoir. Failing pump check valves or piston seals High-pressure proportioning: Identify the failing pump block by relating the pressure drops to the pump cycle. Change the %B and Column contamination Flush the column with strong eluent. If possible, try to reverse the column (check with column specifications). observe how the frequency of pressure drops changes. Use Chromeleon software s pump diagnostic to identify the failing part. Contaminated flow cell Clean the flow cell (refer to the operating instructions for a recommended procedure). If necessary, replace the flow cell. Blocked solvent line frit Replace solvent line frit. New detector lamp For a few hours, allow the lamp to reach a constant light intensity. Non-constant ambient conditions Make sure that the temperature and the humidity in the laboratory are constant. Within Chromeleon software, record temperature Changing Retention Times fluctuations with the help of detector temperature channels. Verify that the lamp and flow cell covers are in their proper position and that the front panel door is closed. Decreasing retention times Degradation of stationary phase Replace column. Check ph requirements of the column, typically ph 2 8 for silica-based RP columns. Slow column equilibration Ion-pairing application Normal. Ion pairing applications always require long initial equilibration times. Shorter alkyl-chain lengths of the ion-pairing agent allow for less equilibration than longer chains. Insufficient equilibration Increase equilibration times, at least 5 10 times the column volume. Spikes Air inside flow cell fluidics Typically, bubbles appear randomly. Identify by: 1. Observing similar spike heights at different wavelengths, or 2. Applying a slight Column overload Reduce amount of sample. Increase column volume (use larger i.d.). backpressure on the flow cell (spikes disappear). Solution: Ensure that connections are tight. Degas mobile phase, use backpressure Mobile phase composition 1. Check premixed mobile phase, make sure the liquid is homogeneous. 2. Perform an OQ proportioning test. 3. a) High-pressure regulator after flow cell (for instance, 100 psi, check flow cell specifications). proportioning: Check for leaks or bubbles in aqeous eluent. a) Low-pressure proportioning: Check adjust proportioning valve. c) Thermo Drainage spiking (Charged Aerosol Detection) Check for leaks at the waste bottle and drain/vent assembly. Check for the presence of internal tubing within drain line. Scientific UltiMate 3000 NCS/NCP systems: Perform pressure sensor and viscosity calibration. Nebulizer (Charged Aerosol Detection) Nebulizer may need to be cleaned. Remove column. Flush HPLC-Charged Aerosol Detection system with 80/20 water:methanol at Increasing flow rate Check flow rate setting, perform OQ on flow rate precision. 2.0 ml/min for several hours. Adjust to 0.2 0.4 ml/min overnight. Re-check noise level with fresh preparation of original mobile phase Stationary phase dewetting at low organic content Use stationary phase compatible with low organic content (e.g., polar embedded phases). Re-wetting the column is facilitated by high composition and flow rate. organic content at increased pressure. Check column care manual for details. Grounding (Charged Aerosol Detection) Ensure that the Charged Aerosol detector is operating on the same circuit as the LC system and is plugged into a surge protector. Column stored without caps Flush column with strong mobile phase and make sure to store column properly, typically (RP phases) in buffer-free, high organic liquid. Electrical interferences Make sure to exclude the previous causes of spikes first. Spikes are typically not random, but for instance, are related to cyclic strong power consuming equipment. In addition, spikes can simply be caused by current fluctuations in the power grid. Solution: Isolate the electrical circuit from strong current consumers or use an UPS (Uninterruptible Power Supply) to filter current fluctuations. Column temperature significantly above boiling point of Use backpressure regulator after the flow cell (for instance 100 psi, check flow cell specifications). Use a post-column cooler (only the mobile phase available with selected column compartments, for instance, the Thermo Scientific UltiMate 3000 TCC-3000RS. Consider switching off the flow cell heating. Thermo Scientific UltiMate 3000 Basic Automated Systems Thermo Scientific UltiMate Standard Systems Thermo Scientific UltiMate 3000 RS and BioRS Systems Thermo Scientific UltiMate 3000 RSLCnano Systems Thermo Scientific EASY nlc-1200 System Charged Aerosol Detector Vanquish Flex UHPLC Systems Vanquish Horizon UHPLC System with Charger Module Chromeleon Chromatography Data System Software Accessories Scan this code to register for a free download of the HPLC Troubleshooting Guide App for iphone, Android, and Windows Phone making UHPLC technology available to all users, all laboratories, and for all analytes For a copy of the troubleshooting poster, please request it from your Thermo Fisher Scientific sales representative. Find out more at thermofisher.com/liquidchromatography 2018 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. CS70303-EN 0418S