Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Similar documents
Hallowell EMC Anesthesia WorkStation Operating Manual

Operation and Maintenance of the EPV200 Portable Ventilator

Unit 15 Manual Resuscitators

Checking a Breathing System

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

Anesthesia Machine Intro

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

NARKOVET ANAESTHETIC UNIT

R409 Plus Veterinary Ventilator Machine

Working version subject to Departmental approval. Drafted by Dr. Olympio and reviewed by clinical members of The Capital Equipment Committee

M5400 Operators Manual

AUTOVENT 4000 VENTILATOR

Recommendations on Checking Anaesthesia Delivery Systems

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

VAPOMATIC ANESTHETIC VAPORIZER

COMPANY PROFILE. Engler manufactures the Sonus V ultrasonic dental unit for the human market

Operation and Maintenance of the MCV100 and MCV100-B Portable Ventilator

Anes Machine Use and Pre-use Check - CA1 focus

Flow meter. bellow vaporizer. APL valve. Scavenging system

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

Manual: Biphasic Positive Airway Pressure (BiPAP) Ventilation

Breathing Systems. Professor Khalid Bashir

S/5 Aespire. Quick Reference Guide

QUICK REFERENCE GUIDE

The SAV04 Small Animal Ventilator. Contents Page

Selecting and Connecting Breathing Systems

Hallowell EMC Vacuum / Scavenging Interface

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

RoVent Jr. Automatic Ventilator Volume & Pressure Controlled. User s Guide

CPR MEDICAL DEVICES INC. OPERATING MANUAL

SERVICE KITS AND CALIBRATION NOTES

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

Aespire 7100 Essential performance Compact design

Capnography in the Veterinary Technician Toolbox. Katie Pinner BS, LVT Bush Advanced Veterinary Imaging Richmond, VA

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Draeger Fabius GS Pre-Use Check

Testing anaesthetic machines, ventilators and related equipment

DPA 02 TM Diamedica Portable Anaesthesia System INSTRUCTIONS AND USER GUIDE

Anesthesia Delivery Systems What Do I Need to Know About My Machine? A Workshop James H. Philip, MEE, MD, CCE

MicroVent 1 Dual Mode Veterinary Ventilator IPPV / HFOV

SurgiVet Universal CDS 9000 Small Animal Anesthesia Machine Operation Manual

Operating & troubleshooting a self inflating bag. Victorian Newborn Resuscitation Project Updated February 2012

Welcome to the Specialized Medical Services Respiratory training webinar series!

Landmark. Veterinary Anesthesia Systems. Operator s Manual P/N

Simplified Automated Ventilator Training Slides

Automatic Transport Ventilator

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

SCUBAPRO. Balanced Power Inflator

Rejuvenation Instructions

Laryngeal Mask Airway (LMA) Indications and Use for the NH EMT-Intermediate and Paramedic

Apollo Safety Features

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

How to use the SAVe II Ventilator

The Merlin Small Animal Ventilator by Vetronic Services Ltd. Contents page. Introduction.2. Features of Merlin. 3

Rev. B. Operating Instructions

Training. Testor Training Manual

VIMA. (Volatile Induction and Maintenance Anesthesia) How and Why. James H. Philip M.E.(E), M.D.

Medical Instruments in the Developing World

MACS Mask CPAP System

OXY-life TM. Oxygen Resuscitator OPERATING MANUAL

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

256 Pneumatic Pressure Indicator

OXY-VIVA 3 RESUSCITATOR OPERATING INSTRUCTIONS

A200SP Circle Absorber Service Manual

Natural Gas Conversion Kit

Keeping anaesthetic pollution to minimum in theatre environment

Mass Casualty Ventilator

Instructions for Use. OptiLife Nasal Lab Mask

Mass Casualty Ventilator Models: MCV100 & MCV100-B

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

Frigitronics CRYO-PLUS

Model 687 Mouse Ventilator User s Manual. Model 687 Mouse Ventilator

Breathing Devices. Chapter 8 KNOWLEDGE OBJECTIVES SKILL OBJECTIVES. 6. List four precautions to take when using oxygen.

Hydraulic Punch Drivers

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

AB800 ABSORBER USER MANUAL

Test your PICU/ NICU/ CHSU Knowledge

Ventilator Training Module

PORTER INSTRUMENT COMPANY, INC.

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

RG1200 Service and Repair Manual

INSTALLATION INSTRUCTIONS MANUAL ON/OFF SAFETY VALVE/PILOT KIT MODEL GA9050A-1 (F0235)

GE Healthcare. Aestiva/5. More than superior ventilation

(Delivery System) Pocket Guide. Series 3 software

Airways and Resuscitators. CRC 330 Cardiorespiratory Care University of South Alabama

Narkomed M Anesthesia System Part No

Your patients, our primary focus

INSTRUCTIONS FOR CONVERTING CONDENSING GAS MOBILE HOME FURNACES MODEL SERIES: CMA3*, CMC1*, VMA3*, VMC1*

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

LISTADO DE CIRCUITOS DE PACIENTE

FLOWMETER UNITS FLOWMETER DEVICES, BY-PASS SYSTEMS

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Transcription:

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator. If you cannot get the ventilator set up and functioning properly, DO NOT USE IT, refer to the Hallowell Manual, or call the Tech Support line. 1. Set Up Front View 1. Bellows 2. Bellows housing 3. Volume Scale 4. On/Off switch 5. Breaths-per-min setting 6. Fine (a, small silver) and coarse (b, large black) tidal volume control Found on 2002 model, the 2000 model has coarse knob (b) only. 7. Inspiratory hold button 8. Maximum working pressure limit setting 2. 1. 3 Black: Bellows Assembly 4 5 6a,b 7 8 Back View 9. Breathing System port 10. Breathing System Exhaust port 11. Driving gas port Bellows side 12. Power supply 13. Driving gas Exhaust port 14. Driving gas port Control side 15. Pressure transducer port 16. Oxygen Driving gas inlet 9 10 11 12 13 14 15 16 1

* The following items should remain set up at all times: The blue exhaust hose connected to the breathing system exhaust port (10), the clear breathing system hose connected to the breathing system port (9), clear small tubing connected to the pressure transducer port, and to the circle system pressure adaptor, the driving gas control (14) and bellows (11) ports connected by black rubber hose. * Before each day of use, the following need to be set up or checked: Oxygen driving gas inlet (16) is connected to house oxygen and to the anesthesia machine, the proper sized bellows (1) is placed on the bellows assembly ring and covered with the bellows housing (2), the power supply (12) is connected to an outlet. Follow steps 1 8 below to completely set up and pressure check the ventilator, but attach a 1L bag to the y-piece of the circle system in place of a patient, and use only oxygen, not inhalant, for a pressure check (read system pressure off the circle system manometer as you do for the anesthesia machine pressure check). Set the breathing rate and volume to the minimum you expect the patient to need (see table pg. 5). Return the circle system to spontaneous breathing set up unless the ventilator will be used immediately, and be sure to pressure check the anesthesia machine itself with the ventilator disconnected. Converting a patient from bag to ventilator (1 8 may be completed in set up): 1. Insert the circle system pressure adaptor between the inspiratory limb of the Y-piece and the anesthesia machine. 2. Attach the blue exhaust hose from the breathing system exhaust port (10) to the scavenge system port (on the anesthesia machine). 3. Remove the breathing bag and attach the clear breathing system hose from the bellows (9) to the anesthesia machine where the breathing bag was. Expel all anesthetic gas from the bag before removing it. 4. Turn the anesthesia machine oxygen flow meter up to 3-4 L/min. 5. Close the pop-off valve on the anesthesia machine. The bellows will begin to fill with anesthetic gas at this point. The patient can breathe spontaneously via the bellows. Do not fill the bellows by pressing the oxygen flush valve; turn up the oxygen flow meter instead so the breathing system is filled with anesthetic gas, not pure oxygen. 6. When the bellows is full, turn the ventilator switch On (4). 7. Adjust the breaths per minute rate control (5) up or down for your patient. You will need to change the volume control after changing the rate: a slower rate delivers a larger volume; a faster rate delivers a smaller volume (see pg. 5). 8. Slowly turn the volume control (6a,b) counterclockwise until the ventilator delivers an adequate breath for the patient and the low pressure alarm no longer sounds at each breath. Use the silver knob for coarse control (with caution), and the black knob for fine control. See page 5 for tidal volume ranges for small animal patients. Set the volume and rate controls so you are giving the proper tidal volume and inspiratory pressure (read directly off the circle system manometer) to your patient at the desired rate. 2

9. Watch the chest rise of the patient, the function of the ventilator, and check the patient and end tidal CO2 monitor for changes as the volume and rate controls are set. 10. Turn down the oxygen flow meter on the anesthesia machine to the proper flow for your patient once the ventilator is functioning properly. 11. Re-check all settings and connections throughout the procedure. Converting a patient from ventilator to bag: 1. It is preferable to wean patients off the ventilator and return them to spontaneous breathing before ending inhalant anesthesia. This is not essential. For very small patients (< 5 kg), you may need to replace the circle system with a non-rebreathing system for this conversion. 2. Three to five minutes before you convert, wean the patient by turning down the breathing rate control to about half the selected operating rate. Lower the volume control to keep the same tidal volume and inspiratory pressure. Watch the patient carefully as rapid awakening may occur during this time. 3. If the patient has not tried to take a breath after 5 minutes at ½ the breathing rate, reduce the tidal volume to ½ your selected volume. This will lower the inspiratory pressure as well. Wait one to two minutes, and watch the patient carefully as rapid awakening may occur during this time. If the patient has not tried to breathe on it s own after 7 minutes of weaning, follow steps 3-7 below and ventilate by hand. 4. As soon as the patient attempts 2 4 breaths on it s own, turn the ventilator Off. 5. Open the pop-off valve on the anesthesia machine. 6. Replace the ventilator breathing system hose on the anesthesia machine with a breathing bag. To fill the bag, turn up the oxygen flow meter; do NOT press the flush valve. 7. Remove the blue ventilator breathing exhaust hose from the anesthesia machine scavenge system port and replace it with the regular anesthesia machine blue exhaust hose. 8. Manually breathe for your patient if needed, or allow the patient to breathe spontaneously. Watch the chest rise of the patient, and check the patient and monitors for changes as the animal breathes on it s own. Changing Bellows sizes: 1. Select the size bellows required for your patient. See tidal volume chart, pg. 5. Make changes BEFORE starting the case. 2. Unscrew the thumb screws holding the bellows housing (2) to the bellows assembly; they need only be loosened, not removed. 3. Twist the bellows housing off the bellows assembly. This is difficult two people make this much easier. 4. GENTLY remove the bellows from the bellows assembly ring. The very last ridge of the bellows should slip over the edge of the ring when the bellows are pulled gently to the side. 3

5. Slip the very last ridge of the new bellows over the appropriate bellows assembly ring. Do not slip more than the last complete ridge of the bellows over the ring or the bellows will leak. 6. Check that the bellows sits flat on the ring and all the ridges are compressed. 7. Replace the appropriately-sized bellows housing on the bellows assembly, twist it firmly in place, and tighten the thumb screws to secure it. 8. It is a good idea to check that the new bellows don t leak by attaching the ventilator to the anesthetic machine and run through steps 1 8 of the bag-toventilator conversion above. Basic Troubleshooting: 1. Bellows won t fill: a. Check the anesthesia machine pop-off valve (Close to ventilate) and breathing hose connection to the anesthesia machine. b. Check for leaks in the anesthesia machine and circle system. c. Check the anesthesia machine oxygen flow rate; it is necessary to increase oxygen flow when first converting to the ventilator from the bag. d. There must be no patient using the anesthesia machine or ventilator for the diagnostics below: i. Check the bellows seating on the bellows assembly ring; only the last ridge should lie over the ring. Check the bellows for an obvious tear or leak. Check that the bellows housing is on properly. ii. Check the internal pop-off valve at the center of the bellows base. (Remove the three red thumb screws, gently remove the valve plate, and inspect the valve base and rim of the outflow tube for debris or hair. Wipe both surfaces with a lint-free cloth moistened with distilled water, dry, and replace the valve.) Note: if the ventilator breathing system exhaust (10) is occluded by hand and the circle system Y-piece is occluded with oxygen flow into the system of > 200 ml/min and the bellows deflates, the problem is not the ventilator pop-off valve, but more likely a bellows tear or system leak. 2. Alarms go off: a. Low pressure: siren sound during breath delivery, or continuous. i. Check for leaks. Check that the volume control is set high enough to deliver an appropriate tidal volume for your patent. A continuous alarm means the breathing circuit has become disconnected somewhere. b. High Pressure: single tone when ventilator delivers a breath. i. Check for obstructions in the breathing circuit and the endotracheal tube of the patient. Check that the maximum operating pressure control (8) is set above the peak airway pressure you require to properly ventilate your patient. 3. Patient wakes up, seems deep, or ventilation is inadequate: 4

i. If in doubt, disconnect the ventilator and manually ventilate your patient. Check fresh gas flow and the vaporizer on the anesthesia machine. Check the breathing circuit for leaks or kinks. Check the ventilation rate and tidal volume both may need to be adjusted during a procedure. 4. Check the manual for more detailed troubleshooting descriptions. For further information: http://www.hallowell.com/index.php?pr=zdoca3667 Ventilation Settings: These are only a guide. Adjust the rate, tidal volume, and peak inspiratory pressure to the exact needs of your patient. Maintain the lowest peak inspiratory pressure which allows an adequate tidal volume at a reasonable rate. *Ventilation of patients below 2 kg using the Hallowell ventilator is not advised. Body Wt. (kg/lb) Bellows Size Breathing Rate (per min) Tidal Volume (ml) *1.0/2.2 small 12-25 10-20 8-12 2.2/5 small 10-25 20-50 8-12 4.5/10 small 10-20 50-100 10-15 6.8/15 small 10-20 60-140 10-15 9.1/20 small 8-15 90-180 10-15 11.3/25 sm/lg 8-15 110-220 10-15 13.6/30 sm/lg 8-15 130-275 10-15 18.1/40 large 8-15 180-360 10-15 22.7/50 large 8-15 220-440 12-20 27.2/60 large 8-12 270-540 12-20 31.8/70 large 6-12 320-640 12-20 36.3/80 large 6-12 360-720 12-25 40.8/90 large 6-12 400-800 12-25 45.4/100 large 6-12 450-900 15-25 49.9/110 large 6-12 500-1000 15-25 54.4/120 large 6-12 540-1100 15-30 Peak Inspiratory Pressure (cm H 2O) When using the Hallowell 2002 ventilator, if you change the breathing rate of the patient, you will change the tidal volume delivered: BPM increase tidal volume decrease BPM decrease tidal volume increase When volume changes, the peak inspiratory pressure will change in the same direction as a result; monitor peak inspiratory pressure using the anesthesia machine manometer. Select your rate first, then adjust volume to meet your patients needs. It is preferable to have a lower rate with a normal volume and pressure than to have a more rapid rate with a lower volume and pressure unless the patient s lungs cannot tolerate a normal PIP. 5