Essential Performance for MED rd ed. PSES San Diego chapter meeting December 11, 2012

Similar documents
Using what we have. Sherman Eagles SoftwareCPR.

Risk Management Series Article 8: Risk Control

IEC :2012 Risk Management Client Completion Form

CHAPTER 2: ELECTRICAL SAFETY

ISO INTERNATIONAL STANDARD

Guidelines on Monitoring in Anaesthesia

Reliability of Safety-Critical Systems Chapter 3. Failures and Failure Analysis

Every things under control High-Integrity Pressure Protection System (HIPPS)

5.1 Introduction. Learning Objectives

Engineering World Health Youth EWH Design Projects That Matter

Guidelines to the standard IEC60601

DETERMINATION OF SAFETY REQUIREMENTS FOR SAFETY- RELATED PROTECTION AND CONTROL SYSTEMS - IEC 61508

Recommendations for Standards. Monitoring. Cardiopulmonary Bypass

Cardiac Arrest Report Guide

Advanced Test Equipment Rentals ATEC (2832) OMS 600

Release: 1. UEPOPL002A Licence to operate a reciprocating steam engine

THE IMPACT ON ENERGY CONSUMPTION CAUSED BY PRESSURE DROP IN A COMPRESSED AIR SYSTEM

10 Best Practices for Electrosurgical Unit Testing

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Resusci Anne Simulator

Entry to the premium class!

Electrical Safety in Hospitals

MECA Ed3.1 Evaluation Package BETA, Through Cl ( ) Page 1 of 65

R S A B C CPR. Basic Life Support Flow Chart Check for danger. Check Response. Send for Help. Check Airway. Check for Breathing.

Safety-critical systems: Basic definitions

Home Oxygen Therapy. Delivering the best in care. UHB is a no smoking Trust

Hazard Operability Analysis

Avalanche Victim Resuscitation Checklist

Morning Time: 1 hour 30 minutes

Spasticity / Dystonia

Ultima. X Series Gas Monitor

SF 6 Product Guide. Gas Analysis Instruments and Accessories

2015 Guidelines Summary HeartSine samaritan PAD Automated External Defibrillators

Simulation for emergency care

McHENRY WESTERN LAKE COUNTY EMS SYSTEM FALL 2014 CONTINUING EDUCATION MANDATORY FOR ALL PRIMARY AND PROBATIONARY ALS SYSTEM PROVIDERS.

Part 2-55: Medical electrical equipment. Particular requirements for the basic safety and essential performance of respiratory gas monitors

ISO INTERNATIONAL STANDARD. Sleep apnoea breathing therapy Part 2: Masks and application accessories

Pediatric Hospice Patients

Upload-pressurized Automatic Sphygmomanometer.

Section1. Knowledge of RPE Sets out the minimum knowledge requirements in respect of respiratory protective equipment. [Mandatory] 1.

CENELEC GUIDE 32. Guidelines for Safety Related Risk Assessment and Risk Reduction for Low Voltage Equipment. Edition 1,

Systems Theoretic Process Analysis (STPA)

eacls TM Skill Performance

A GUIDE TO RISK ASSESSMENT IN SHIP OPERATIONS

5 th year (Control department)

WARNING: EXPLOSION HAZARD

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Fingertip Pulse Oximeter

On-Scene Resuscitation. Brent Myers, MD MPH Director Medical Director Wake County EMS, Raleigh, NC

ISO INTERNATIONAL STANDARD

LIFEPAK. CR2 Defibrillator. Better technology for better outcomes. Essential version

Review and Assessment of Engineering Factors

DEPARTMENT SCHOOLS OF HEALTH

VENTILATORS PURPOSE OBJECTIVES

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

American National Standard

ISO INTERNATIONAL STANDARD

SOP 407: PROTOCOL DEVIATIONS AND UNANTICIPATED PROBLEMS

Understanding safety life cycles

Accidents and accident prevention

Overview. The user must follow the instructions below.

Intelligent Ventilation solution from ICU to MRI

Ultrasound Dose Calculations

Dynamic Automated External Defibrillator (AED) Registry

2017 Edition GUIDANCE NOTE GN 8 INVERTER DRIVES FOR FANS AND PUMPS GN 8 - INVERTER DRIVES FOR FANS AND PUMPS

MAGICFX CONFETTIPISTOL USER AND INSTALLATION MANUAL

New Thinking in Control Reliability

Section 1: Multiple Choice

D-MAS HyperSat DanMedical hyperbaric remote patient monitoring system

System Operating Limit Definition and Exceedance Clarification

Hazard and risk analysis in pharmaceutical products 1

Accident Investigation and Hazard Analysis

M-06 Nitrogen Generator (Nitrogen Making Machine)

TABLE OF CONTENTS SECTION 16

Cardiac Catheterisation Instrumentation. Lancashire & South Cumbria Cardiac Network

How does the paramedics work day look like? Have a look at our field research video.

Pomona College Automated External Defibrillator (AED) Program

Hazard Identification

LQG Based Robust Tracking Control of Blood Gases during Extracorporeal Membrane Oxygenation

RELIABILITY OF ENERGY SYSTEMS

Medical Instruments in the Developing World

A study on the relation between safety analysis process and system engineering process of train control system

Confidence comes with knowing you are Code-Ready.

Decant Filling of Gas Cylinders

Commandant United States Coast Guard

INSPECTION OF MEDICAL DEVICES

What is an AED? Why buy from defibshop?

Effective Workplace Inspections

Distinguishing Features

Powerheart G5 AED Automatic

INFORMATION NOTE No 353: MICROBIOLOGICAL TESTING

የIትዮዽያ የደረጃዎች ኤጀንሲ ETHIOPIAN STANDARDS AGENCY ES DRAFT UPLOAD REQUEST FORM FOR PUBLIC COMMENTS

3. Real-time operation and review of complex circuits, allowing the weighing of alternative design actions.

PERFORMANCE CRITERIA

Your reference Your correspondence of Our reference Date VOU/RLE 13 June 2018

Automated External Defibrillator (AED) Program PROCEDURES

CARDIOVIT AT-104 ergospirometry

Contacts. Quick Start Guide

POLICY: EBBAA (453.11) Approved: September, 2002 Reviewed: January, 2003 Revised: February 23, 2004 Revised: August 18, 2015 DEFIBRILLATOR

Transcription:

Essential Performance for MED 601-1 3 rd ed. Grant Schmidbauer, Nemko USA, Inc. PSES San Diego chapter meeting December 11, 2012 2012 PSES San Diego Chapter http://www.psessymposium.org/ 1

3.27 Essential Performance Old (3 rd ed.): performance necessary to achieve freedom from unacceptable risk New (Amd. 1): performance of a clinical function, other than that related to BASIC SAFETY, where loss or degradation beyond the limits specified by the manufacturer results in an unacceptable risk

3.10 Basic Safety And just as a reminder: BASIC SAFETY freedom from unacceptable risk directly caused by physical hazards when MEE is used under normal condition and single fault condition

3.27 Essential Performance Rationale* It has long been recognized that MEE or an MES that does not perform properly could result in unacceptable risk for patients, operators, or others Hence the concept of "safety" has been broadened from the basic safety considerations in the first and second editions of this standard to include essential performance matters *Rationale from Amd. 1

3.27 Essential Performance Rationale* In order to achieve its intended use, the MEE or MES needs to perform within certain limits These limits are usually specified by the manufacturer but could be specified by this standard, a collateral standard or a particular standard in the IEC 60601 family *Rationale from Amd. 1

3.27 Essential Performance Rationale* Examples of essential performance are: correct administration of a drug by a syringe pump where inaccuracy/incorrect administration would cause an unacceptable risk to the patient; the ability of an electrocardiograph/monitor to recover from the effects of the discharge of a defibrillator where the failure to recover could lead to an incorrect response by the medical staff that would present an unacceptable risk to the patient; *Rationale from Amd. 1

3.27 Essential Performance Rationale* Examples of essential performance are: correct operation of an alarm system in an intensive care or operating room monitoring system where an incorrect/missing alarm signal could lead to an incorrect response by the medical staff that would present an unacceptable risk to the patient; or correct output of diagnostic information from MEE that is likely to be relied upon to determine treatment, where incorrect information could lead to an inappropriate treatment that would present an unacceptable risk to the patient *Rationale from Amd. 1

3.27 Essential Performance Rationale* For purposes of this standard, performance related to basic safety aspects of the MEE, such as the performance of basic insulation, is not considered to be essential performance Particular and collateral standards in the IEC 60601 family are expected to identify specific essential performance *Rationale from Amd. 1

4.3 Essential Performance Rationale* During the initial risk analysis, the manufacturer identifies the performance of the MEE or MES that is necessary to achieving the intended use The manufacturer also identifies other qualitative and quantitative characteristics that could affect the safety of the MEE or MES *Rationale from Amd. 1

4.3 Essential Performance Rationale* The performance limits specified by the manufacturer could be anywhere within the full range of intended performance in NC and SFC, to no performance The manufacturer then determines if the loss or failure to perform within specified limits would result in a risk to the patient *Rationale from Amd. 1

4.3 Essential Performance Rationale* The estimate is often made with the assumption that the performance aspect in question has been lost or degraded beyond the specified limits The manufacturer takes into account the probability of a hazardous situation leading to harm as well as the severity of that harm *Rationale from Amd. 1

4.3 Essential Performance Rationale* The manufacturer then evaluates the risk using their established risk acceptance criteria If the risk is unacceptable, then the identified performance is Essential Performance because risk reduction is required to achieve an acceptable level of residual risk *Rationale from Amd. 1

4.3 Essential Performance Rationale* Once the essential performance is identified, the manufacturer puts in place risk control measures that are appropriate to reduce the risk to an acceptable level When conducting the risk control option analysis, the manufacturer is to follow the priority order listed in ISO 14971, namely: a) Inherent safety by design b) Protective measures c) Information for safety *Rationale from Amd. 1

Life-supporting devices should have adequate performance, perform as instructed by the operator, or signal clearly that a fault exists Even a total failure should alert the operator via internally powered alarms

Similar requirements should apply to diagnostic equipment which might be used in emergency or rescue situations where reliance could be placed solely on the information from the device to determine immediate clinical action, e.g. an ECG monitor used to determine the effect of a cardiac defibrillator discharge

Other non-life-supporting devices may have differing performance specifications and may fail without causing an unacceptable risk In many instances the failure will be obvious to the operator

Examples of Essential Performance requirements include those characteristics where inaccuracy, inadequacy or failure could: directly affect the PATIENT s safety e.g. accuracy of a life-supporting function, incorrect administration of a drug by a syringe pump, etc.

continued.. (are EP) lead to an incorrect response by the medical staff which might be hazardous for the pateint, e.g. inability of an electrocardiograph/monitor to recover from the effects of the discharge of a defibrillator, incorrect operation of an alarm in an intensive care or operating room monitoring system, etc.

continued. (are EP) lead to an incorrect diagnosis where the device is likely to be the only source of evidence, e.g. devices used in accident and emergency situation outside the hospital environment, diagnostic devices prescribed for use by a patient to determine self treatment, results in repetition of a procedure associated with a known risk to the patient.

Examples of requirements that are NOT essential performance requirements but may be suitable for inclusion in other standards or technical reports: accuracy of measuring systems where other evidence is available e.g. pulse rate in an ECG monitor, unless part of an intensive care system with high and low rate alarms,

continued. (not EP) accuracy of presentation of parameters of non vital physiological functions e.g. EEG, EMG, ENG, Audiometry, Kinesiology, Optometry, etc. format of records, displays, communication protocols, etc.,

continued. (not EP) limits of performance e.g. frequency response, electrical noise levels, cross talk, data processing capabilities, etc., output power 2, flow rates, etc.

continued. (not EP) accuracy of a diagnostic advisory function (e.g., ECG rhythm classification or interpretation of ECG morphology) where: medical staff have access to original patient data on which the advice is based, and the advisory function does not automatically initiate patient treatment In this case, the accompanying documentation must explain the limitations of the accuracy and use of the function

It is unnecessary and sometimes even difficult to differentiate between basic safety and other essential performance requirements, for example:.

Specific limits on electrical output might be necessary in order to achieve acceptable risk levels for equipment such as defibrillators, electrosurgical units and therapeutic stimulators The performance necessary to ensure the output does not exceed these limits might be seen as a basic safety requirement if excessive output can result in tissue damage, while the performance necessary to ensure at least adequate output might be seen as essential performance too little output would result in failure to achieve the intended purpose.

Specific thermal requirements may be necessary in equipment such as dialysis machines, blood warmers and heart-lung machines The performance necessary to ensure the extracorporeal circulation is not over-heated might be seen as a basic safety requirement if excessive heating could result in physical damage (haemolysis) while the performance necessary to ensure at least adequate heating might be seen as essential performance.

In some cases one user of the 3rd edition of IEC 60601-1 may even treat a particular requirement as a basic safety issue while another user may treat it as essential performance..

For example, specific accuracy requirements may be necessary for the high voltage generator in diagnostic X-ray equipment. Some might see the performance necessary to achieve this as a basic safety requirement because an inadequate image might result in the need for additional films with an inherent increase in radiation dose and an increased risk of physical damage to the patient (cancer). Others might see this as an essential performance requirement because an inadequate image might result in an incorrect diagnosis and inappropriate treatment.

While ISO 14971 clearly requires the manufacturer of a particular product to differentiate between the performance necessary to achieve freedom from unacceptable risk and other non-essential performance.. it is usually a waste of time differentiating between basic safety and essential performance requirements..

The risk management file required under IEC 60601-1 and ISO 14971 will already identify all the equipment performance necessary to achieve freedom from unacceptable risk and there will be no need to differentiate between basic safety and essential performance

Standards in the IEC 60601 series provide tried-and-true requirements by which some risks can be reduced to acceptable levels in some particular MEE The general standard and the collateral standards clearly do no include an adequate range of tried-and-true requirements for every type of MEE

The role of a particular standard is to modify the general requirements and specify additional requirements where appropriate for a particular type of MEE all of the additional requirements set out in a particular standard address either basic safety or essential performance

as IEC 60601-1 and its associated collateral standards provide adequate basic safety requirements for most MEE, the additional requirements set out in most particular standards are about the essential performance required for the particular type of equipment.

4.3 Essential Performance Frank O Brien Where EP is to be maintained after a particular test, an EP test program is needed.

Contact Information Grant Schmidbauer Nemko USA, Inc. 2210 Faraday Ave, Suite 150 Carlsbad (San Diego), CA 92008 USA Tel: 760-444-3413 Fax: 760-444-3005 Email: grant.schmidbauer@nemko.com Website: www.nemko.com