PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

Similar documents
Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

ali-q TM Gravimetric Verification Procedure

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

Operating Instructions METTLER TOLEDO Pipette Check Application for AX and MX/UMX Balances Version 1.xx

Pipettor. User Manual

Transferpette -8/-12

VITLAB micropipette. Standard Operating Procedure (SOP)

RayPette Plus Autoclavable Pipette. User Manual

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

The pipettes cover a volume range from 0.1µl to 10ml.

THE USE OF Z FACTORS IN GRAVIMETRIC VOLUME MEASUREMENTS

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

MULTICHANNEL. Labnet International, Inc.

VITLAB Volumetric instruments

INTERNATIONAL STANDARD

Ensuring pipette & weighing device performance: Verification & Calibration

Technical Procedure for General Laboratory Equipment

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Transferpette -8/-12

Inquiry Module 1: Checking the calibration of a micropipette

Register your instrument! Eppendorf Research plus. User Adjustment

CALIBRATION AND ADJUSTMENT

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

Calibrating a Micropipette

Eliminating Sources of Pipetting Error in the Forensic Laboratory David M. Epstein Director of Scientific Operations. Abstract.

Register your instrument! Eppendorf Xplorer / Eppendorf Xplorer plus. Adjustment

Transferpette. Lightweight precise rugged reliable!

Calibrating Piston Pipettes

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

MSL Technical Guide 30. Calibrating Piston Pipettes. Introduction. Piston Pipettes. Calibrating Piston Pipettes overview

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

Proline Plus Mechanical Pipette

Transferpette. Lightweight precise rugged reliable!

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Standard Operating Procedure. Air Displacement Pipette Calibration

A N F G J I C D D L A B C D E F

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

New Technological Guidelines Required for Pipette Management (Proposal of a Pipette Management Method)

Installation Procedure Esco Miri Multi-Room Incubator

PIPETMAN F ENGLISH. Other languages available on our website:

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Performance Verification of BioTek s Precision 2000 using Artel s Multichannel Volume Verification (MVV ) System

Density Determination Kit Instruction Manual

Buoyancy Does it Matter?

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

Transferpette. Single and Multichannel Pipettes

Operating Instructions. Rainin Classic Manual Pipette

Gas mixtures. Individual solutions specifically for your application

Real World Metric System

Pipetting Small Volumes

QAM-I-117 Volumetric Equipment Calibration Verification

MET-101 Introduction to Metrology

Measuring Mass and Volume

GMC-I Service GmbH Calibration Center

Improving Pipetting Techniques

SOP No. 14. Recommended Standard Operating Procedure for Gravimetric Calibration of Volumetric Standards Using an Electronic Balance

K V A L I F I K. Our accreditation number: NAH /2015. Valid to:

Adjust the Volume Display

Chapter 2. Pipetting techniques. Guide to Pipetting

A N B P F G J I C D D L A B C D E F

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

Corning Step-R. Repeating Pipettor. Instruction Manual

Tacta Mechanical Pipette User Manual

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

BASIC LABORATORY TECHNIQUES (Revised )

CHM250 Calibration and Measurement Lab. Balance Calibration

INFLUENCE OF MEASURING PROCESS AUTOMATION ON UNCERTAINTY OF MASS STANDARD AND WEIGHTS CALIBRATION.

Varipette Operating manual

Good Pipetting Practice

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

General Accreditation Guidance. User checks and maintenance of laboratory balances

SmartCal. SmartCal Troubleshooting Checklist

Single Channel Pipette Calibration and Operator Competency Assessment Using a Dual-Dye Ratiometric Photometry System

Calibration of Volumetric Glassware

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

Biochemistry Laboratory I CHEM 4401 Pipetting & Scales

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

MICROMAN ENGLISH. Other languages can be consulted and downloaded on our website,

SoftGrip Single Channel Pipettes User Manual New Photo To Come

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

CORESTA RECOMMENDED METHOD N 6

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

INTERNATIONAL STANDARD

Traceable calibration of automatic weighing instruments in dynamic operation

Qi Litre Ultrasonic Cleaner

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

BASIC LABORATORY TECHNIQUES (Revised )

Pipette apparatus. Meet the difference. Manual. T E I

2014 Programme of Training Courses in Measurement & Calibration

Transcription:

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS VER3150999 1

Content: 1.0 Introduction...2 2.0 Existing standards....3 3.0 Biohit gravimetric test process, flow chart 4 4.0 The Calibration Service. 5 5.0 Performance test according ISO 8655......6 5.1 Equipment needed.7 5.2 Simplified test instructions....8 6.0 Performance specifications in field service...9 6.1 The Biohit Field specifications... 10 7.0 Quality Control 13 8.0 The Main Biohit Service Center Approval.....14 9.0 Uncertainty of the measurement..15 9.1 temperature.. 16 9.2 evaporation..17 9.3 humidity..18 9.4 air pressure..19 2

1.0 Introduction: The purpose of this manual is to give the main frame of the Standard Operation Procedure ( SOP ) how to execute the performance and calibration service to all Biohit Proline pipettors. This manual follows the latest features of the future standards and their demands to field technical service. 3

2.0 Existing Standards: Following is short description of all main standards and their special features. 1) DIN 12650, Piston-operated volumetric apparatus Is a German standard series which was launched 1997. Today major part of the pipettor manufacturers are following this standard. DIN 12650 describes in detail how volumetric instruments are defined and should be tested. The specifications in this standard are introduced as a calculation result, which is named as the F-value, German Fehlergrenze. This F-value formula is: accuracy + 2 x CV. DIN 12650 is specifying the maximum error in µl and relative maximum error. When this manual is produced ( September -99 ) the Biohit specifications are calculated in relation to DIN 12650. However the Biohit specifications are better and more strict than what the mentioned standard demands. 2) NCCLS and BS 7532:1991 NCCLS = National Committee for Clinical Laboratory Standards has published the proposed guideline 18-P Determining Performance of Volumetric Equipment. BS 7532 is a British standard Accuracy and precision of mechanical hand pipettes of capacity less than 50µl down to and including 10µl. The NCCLS and BS 7532:1991 are describing accuracy and precision terms while DIN 12650 describes the F-value. Also some difference can be found how to make the gravimetric test. The basics and principles of NCCLS and BS7532 are very close to e.g. DIN 12650 and therefore do not have big significance against what standard a pipettor is tested 3) ISO 8655 Is a new series of standard, which is at the moment ( September -99 ) in Committee Draft (CD ) stage. This new international standard will most obviously be approved to have the draft International standard ( DIS ) level before end 1999. The basis for ISO 8655 is the corresponding DIN 12650 and its parts. However quite many changes have been made to original working drafts ( WD ) and the ISO 8655 will differ from the DIN 12650. The ISO 8655 will describe trueness and precision and also the field technical service is taken in notice. Also the new ISO will refer also to field service performance testing and specifications. 4

3.0 The Biohit Gravimetric test process, flow chart BIOHIT OYJ Name of the quality managerment protocol Logistics/ Production Acitivity defidition Performance test Page 1 Designed by Approved Date Code S. Riikonen O.Suovaniemi 170899 TKK - 3C- 08-01 = File = Document = Task, Responsility, Participate = Decision = Transfer or End OP/ Pipette in to the testing environment VP TH Are the test devices calibrated and in operation No Balance calibration and starting the test equipments OP/ VP TH Yes NTV Record the temperature and humidity, OK? No Adjust the parameters according the given instructions. OP/ VP TH Yes Distilled water according the given parameters? No Change fresh water OP/ VP TH Yes Gravimetric test instructions OP/ Performance test VP ( DIN 12650/ISO 8655 TH ) no Did the unit meet the given No field specifications? Any repair activities Yes needed? NTTK QC-certificate Performance test results Yes OP/ Store the performance data and print QC-certificate VP TH Place a repair tag on the pipette and execute necessary service activities OP= Departmental head VP= Head of manufacturing TH= Production personell NTTK= Performance data file NTV= Environmental parameter control file End LOM-0-01-01 5

4.0 The Calibration service Major part of Biohit technical service departments can already offer to their customers a pipettor performance service or calibration service. However there are several kind of methods in field and misunderstandings how to execute the performance test and what specifications should be used. With the help of following instructions we like to give tools to you for proper calibration/performance service. The original purpose of the calibration service is that a pipettors performance; the present accuracy and precision levels are tested using either gravimetric (or photometric methods). Very rarely it means that pipettor should perform like a new one after several years use. The nature of a calibration or performance service is only to check the pipettor. Therefore our recommendation is: - Pipettor cleaning and regreasing according given instructions - Either gravimetric (or photometric) test. - Calibration / performance certificate. However many times the market situation can demand to include additional features to above content. All these additional features should be decided locally. Also the service price varies depending of the market area. 5.0 Performance test method, according ISO 8655. Pipette manufacturer can decide in which and (in how many) volume(s) the pipettor is performance tested before shipping. This factory test is a QA-test (Quality Assurance), which confirms the performance of the pipettor. This factory QA-test is also called a suppliers declaration. However manufacturers must publish type test or conformity test specifications to following volumes: - nominal volume =upper limit of useful volume range. - 50 % off nominal volume. - the smallest settable volume and/or 10% of the nominal volume. The conformity test is always related to entire system, pipettor and tip used. The original ISO 8655 standard method describes how the type test (or conformity test) should be done. Unfortunately it is not clearly stating how pipettors should be tested in field and normal laboratory environment. 6

The ISO 8655 separates two specification levels: 1) Manufacturers specifications, ( type test specifications, conformity test specifications or type evaluation ) The specks defined by the manufacturer with their internal conformity testing. These specks are achieved in strictly controlled conditions with optimal pipetting features and technique. The type test is done to one or more representative specimens of an instrument type. 2) User specifications ( field specifications ) ISO 8655 allows users to exceed given specifications by the standard not more than 100%, multichannel pipettors. The main responsibility is user to establish the accuracy and precision levels for their own needs and performance demands. Please note that specifications mentioned in ISO 8655 are the maximum error levels defined to pipettor/tip manufacturers to fulfil the demands of this standard. However main part of the pipettor manufacturers introduces specifications which are tighter and better that mentioned in ISO 8655. This is due the result of their own production process quality and /or market demands. When pipettors are tested in field conditions it must be confirmed against what specifications the pipettors are referred to. NOTE! The ISO 8655 allows the technical services to use specification levels given in this standard. 5.1.Equipment needed: Analytical balance: - analytical balance min. 0.01mg readability - weighing vessel with relatively small inner diameter The balance should be calibrated in intervals recommended by the balance manufacturer or when ever relevant. Thermometer: 20-25 C ± 0.5 C The main purpose is to control that pipettor, tips used, balance and water used are at the same testing temperature. Water: grade 3, according ISO 3696 The water should be changed in regular basis and not reused at all. 7

5.2 Simplified field ( user ) test instructions: 1) Temperature : Take care that pipettor, tip, balance and water used are within the same temperature. 2) Pipetting technique: Tip should be immersed into the liquid only 2-3 mm into a vessel containing water. The pipettor should not be tilted. There must be a 1-2 second pause before the tip is removed from water. It is important that using same pipetting technique and smooth plunger operation carry out the complete performance test when testing mechanical pipettors. 3) Disposable tip: The performance of a pipettor is a combination of the pipettor and tip used. Tips of different origin can cause problems to achieve proper performance level. Therefore the original manufacturers specks are valid only when test is executed with the original manufacturers tips. 4) Prerinsing: Prerinsing it 3-5 times prior to test should wet the tip used. 5) Number of repetitions: Since the accuracy (trueness) and precision are statistical values it is recommended to execute 10 subsequent measurements. 6) Microliters vs. milligrams: The milligram reading (mg) is not directly a test result in microliters (µl) and the conversion factor (Zfactor) is dependent on temperature. The detailed test procedure can be found from publications ISO 8655-6. 8

6.0 The Biohit Specifications in field service: The specks defined by the manufacturer are for their internal conformity testing. These specks are achieved in strictly controlled conditions with optimal pipetting features and technique. Unfortunately in field service the control or adjustment of the environment parameters is difficult or many times even impossible. The main purpose of the performance test in field is to find out how good or bad the pipettor is performing in normal environment. Therefore it is not sensible to performance test pipettors against manufacturers original type test specifications or trying to achieve the same performance level as in the strict controlled conditions. The DIN 12650 allows user to exceed the specks mentioned in the standard not more than 100%. However the user should establish their own specification levels related to their needs. NOTE! The specification mentioned in ISO 8655 are not same as the Biohit type specifications. Biohit Oyj recommends that all Biohit technical service centers should take in use field specifications, which are: The basic rule: The original Biohit type specks + 50 % For example; Original accuracy: 1.00% Field: 1.50% Original precision: 0.40% Field: 0.60% Following pages describes The Biohit Field Test Specifications. 9

6.1.The Biohit Field specifications Code Volume Volume Type specifications Service specifications range Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN12650,Fehlerg./µl 710520 0,2-10 10 0,9 0,5 0,2 1,4 0,8 0,3 5 1 0,7 0,2 1,5 1,1 0,3 1 2,5 1,5 0,2 3,8 2,3 0,3 0,2 6 6 0,2 9,0 9,0 0,3 710010 5-100 100 0,4 0,15 0,7 0,6 0,2 1 50 0,7 0,3 0,7 1,1 0,5 1 10 2 1 0,7 3 1,5 1 5 2,5 1,8 0,7 3,8 2,7 1 710030 10-250 250 0,4 0,15 1,2 0,6 0,2 2,5 125 0,6 0,2 1,2 0,9 0,3 2,5 25 1,5 0,8 1,2 2,3 1,2 2,5 10 2 1 1,2 3,0 1,5 2,5 710100 10-500 500 0,4 0,15 3,5 0,6 0,2 5 250 0,7 0,2 3,5 1,1 0,3 5 50 1,5 0,8 3,5 2,3 1,2 5 10 4 1,5 3,5 6 2,3 5 710020 50-1000 1000 0,4 0,15 7 0,6 0,2 10 500 0,6 0,2 7 0,9 0,3 10 100 1,5 0,5 7 2,3 0,8 10 50 2 1 7 3 1,5 10 710040 50-1200 1200 0,4 0,15 8,4 0,6 0,2 12 600 0,7 0,2 8,4 1,1 0,3 12 100 1,5 0,5 8,4 2,3 0,8 12 50 2 1 8,4 3 1,5 12 710500 100-5000 5000 0,5 0,15 40 0,8 0,2 58 2500 0,8 0,2 40 1,2 0,3 58 500 0,8 0,3 40 1,2 0,5 58 710410 5-100/4 100 0,5 0,2 1,5 0,8 0,3 2,3 50 0,8 0,3 1,5 1,2 0,5 2,3 10 1,5 1,5 1,5 2,3 2,3 2,3 5 4 2,5 1,5 6,0 3,8 2,3 10

Code Volume Volume Type specifications Service specifications range Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN12650,Fehlerg./µl 710420 25-250/4 250 0,4 0,15 1,75 0,6 0,2 2,5 125 0,7 0,2 1,75 1,1 0,3 2,5 25 1,5 1 1,75 2,3 1,5 2,5 710200 0,2-10/8 10 0,9 0,5 0,2 1,4 0,8 0,3 5 1,5 0,8 0,2 2,3 1,2 0,3 1 4 4 0,2 6 6 0,3 710210 5-100/8 100 0,5 0,2 0,9 0,8 0,3 1,4 50 0,7 0,3 0,9 1,1 0,5 1,4 10 1,5 1,5 0,9 2,3 2,3 1,4 5 4 2,5 0,9 6,0 3,8 1,4 710220 25-250/8 250 0,4 0,15 1,75 0,6 0,2 2,5 125 0,6 0,2 1,75 0,9 0,3 2,5 25 1,5 1 1,75 2,3 1,5 2,5 710800 50-1200/8 1200 0,5 0,15 9,6 0,8 0,2 13,8 600 1 0,2 9,6 1,5 0,3 13,8 100 4 0,8 9,6 6 1,2 13,8 50 8 1,5 9,6 12 2,3 13,8 710300 0,2-10/12 10 0,9 0,5 0,2 1,4 0,8 0,3 5 1,5 0,8 0,2 2,3 1,2 0,3 1 4 4 0,2 6 6 0,3 710310 5-100/12 100 0,5 0,2 0,9 0,8 0,3 1,4 50 0,8 0,4 0,9 1,2 0,6 1,4 10 1,5 1,5 0,9 2,3 2,3 1,4 5 4 2,5 0,9 6 3,8 1,4 710320 25-250/12 250 0,4 0,15 1,75 0,6 0,2 2,5 125 0,6 0,2 1,75 0,9 0,3 2,5 25 1,5 1 1,75 2,3 1,5 2,5 710810 50-1200/12 1200 0,8 0,15 9,6 1,2 0,2 13,8 600 1 0,2 9,6 1,5 0,3 13,8 100 4 0,8 9,6 6 1,2 13,8 50 8 1,5 9,6 12 2,3 13,8 11

Code Volume Volume Type specifications Service specifications range Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN 12650,Fehlerg./µl 720000/ 0,5-10 10 1 0,8 0,3 1,5 1,2 0,4 720001 5 1,5 1,5 0,3 2,3 2,3 0,4 1 2,5 1,5 0,3 3,8 2,3 0,4 720080 2-20 20 0,9 0,4 0,3 1,4 0,6 0,5 10 1,2 1 0,3 1,8 1,5 0,5 2 3 2 0,3 4,5 3 0,5 720020/ 5-50 50 0,6 0,3 0,6 0,9 0,5 1 720021 25 0,9 0,6 0,6 1,4 0,9 1 Calib.vol. 20 0,9 0,6 0,6 1,4 0,9 1 5 2 2 0,6 3 3 1 720050 10-100 100 0,8 0,15 1,1 1,2 0,2 1,6 50 1 0,4 1,1 1,5 0,6 1,6 Calib.vol. 25 1,3 0,5 1,1 2,0 0,8 1,6 10 3 1,5 1,1 4,5 2,3 1,6 720070 20-200 200 0,6 0,15 1,8 0,9 0,2 2,6 100 0,8 0,3 1,8 1,2 0,5 2,6 Calib.vol. 50 0,7 0,3 1,8 1,1 0,5 2,6 20 3 1 1,8 4,5 1,5 2,6 720030/ 50-200 200 0,6 0,15 1,8 0,9 0,2 2,6 720031 100 0,8 0,3 1,8 1,2 0,5 2,6 50 1 0,4 1,8 1,5 0,6 2,6 Calib.vol. 50 0,7 0,3 1,8 1 0,5 2,6 720060 100-1000 1000 0,6 0,2 10 0,9 0,3 15 500 0,7 0,25 10 1,1 0,4 15 Calib.vol. 200 0,6 0,3 10 0,9 0,5 15 100 2 0,7 10 3 1,1 15 720040/ 200-1000 1000 0,6 0,2 10 0,9 0,3 15 720041 500 0,7 0,25 10 1,1 0,4 15 200 0,9 0,3 10 1,4 0,5 15 Calib.vol. 200 0,6 0,3 10 0,9 0,5 15 12

Code Volume Volume Type spesifications Service specifications range Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN 12650,Fehlerg./µl 720110/ 1-5 ml 5 0,5 0,15 40 0,8 0,2 60 720111 2,5 0,6 0,3 40 0,9 0,5 60 1 0,7 0,3 40 1,1 0,5 60 720120 5-50/4 50 1 0,5 1 1,5 0,8 1,5 25 1,5 1 1 2,3 1,5 1,5 Calib.vol. 20 1,5 1 1 2,3 1,5 1,5 5 3 2 1 4,5 3 1,5 720130 50-250/4 250 0,7 0,25 3 1,1 0,4 4,8 125 1 0,5 3 1,5 0,8 4,8 Calib.vol. 100 1 0,5 3 1,5 0,8 4,8 50 1,5 0,8 3 2,3 1,2 4,8 720210/ 0,5-10/8 10 1,5 1,5 0,45 2,3 2,3 0,7 720211 5 2,5 2,5 0,45 3,8 3,8 0,7 1 4 4 0,45 6 6 0,7 720220/ 5-50/8 50 1 0,5 1 1,5 0,8 1,5 720221 25 1,5 1 1 2,3 1,5 1,5 Calib.vol. 20 1,5 1 1 2,3 1,5 1,5 5 3 2 1 4,5 3 1,5 720230/ 50-250/8 250 0,7 0,25 3 1,1 0,4 4,7 720231 125 1 0,5 3 1,5 0,8 4,7 Calib.vol. 100 1 0,5 3 1,5 0,8 4,7 50 1,5 0,8 3 2,3 1,2 4,7 720240/ 50-300/8 300 0,7 0,25 3,6 1,1 0,4 5,6 720241 150 1 0,5 3,6 1,5 0,8 5,6 Calib.vol. 100 1 0,5 3,6 1,5 0,8 5,6 50 1,5 0,8 3,6 2,3 1,2 5,6 720310 0,5-10/12 10 1,5 1,5 0,45 2,3 2,3 0,7 5 2,5 2,5 0,45 3,8 3,8 0,7 1 4 4 0,45 6 6 0,7 13

Code Volume Volume Type specifications Service specifications range Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN 12650,Fehlerg./µl 720320/ 5-50/12 50 1 0,5 1 1,5 0,8 1,5 720321 25 1,5 1 1 2,3 1,5 1,5 Calib.vol. 20 1,5 1 1 2,3 1,5 1,5 5 3 2 1 4,5 3 1,5 720330/ 50-250/12 250 0,7 0,25 3 1,1 0,4 4,7 720331 125 1 0,5 3 1,5 0,8 4,7 Calib.vol. 100 1 0,5 3 1,5 0,8 4,7 50 1,5 0,8 3 2,3 1,2 4,7 720340/ 50-300/12 300 0,7 0,25 3,6 1,1 0,4 5,6 720341 150 1 0,5 3,6 1,5 0,8 5,6 Calib.vol. 100 1 0,5 3,6 1,5 0,8 5,6 50 1,5 0,8 3,6 2,3 1,2 5,6 Code Volume range Volume Type specifications Service specifications Inacc. % Impr. % DIN 12650,Fehlerg./µl Inacc.% Impr. % DIN 12650,Fehlerg./µl 722001 5 5 1,3 1,2 0,2 2,0 1,8 0,3 722004 10 10 0,8 0,8 0,2 1,2 1,2 0,4 722010 20 20 0,6 0,5 0,3 0,9 0,8 0,5 722015 25 25 0,5 0,3 0,3 0,8 0,5 0,4 722020 50 50 0,5 0,3 0,55 0,8 0,5 0,8 722025 100 100 0,5 0,3 1,1 0,8 0,5 1,7 722030 200 200 0,4 0,2 1,6 0,6 0,3 2,4 722035 250 250 0,4 0,2 2 0,6 0,3 3 722040 500 500 0,3 0,2 3,5 0,5 0,3 5,3 722045 1000 1000 0,3 0,2 7 0,5 0,3 11 722050 2000 2000 0,3 0,15 12 0,5 0,2 17 722055 5000 5000 0,3 0,15 30 0,5 0,2 43 14

7.0 Quality Control More and more demands are directed to technical service and traceability of the gravimetric performance testing. The traceability of the gravimetric test is understood to include: - balance calibration - certified calibration weights - test method and standard used - environmental control Therefore Biohit Oyj has launched a quality control protocol for field service departments which are executing performance or calibration tests to end user. Protocol: 1) Biohit Oyj will send a set of pipettors to be performance tested by the local technical service center. 2) After the pipettors are tested the units and their test results are returned back to Biohit Quality Control who will check the results. The service partner is not allowed to do any cleaning service or calibration adjustment to pipettors in concern. 3) Biohit Quality Control will send The Competence Certificate of the test results. This Quality control test is done once in a year, after 9 months intervals. 8.0 The Main Biohit Service Center Approval: Those service partners which attends the Quality Control program, will get the Biohit Performance Test Certificate where Biohit Oyj approves the partner to execute the performance tests. This certificate also conforms testing is regularly controlled by Biohit Oyj; Quality Control Department. Please note that the traceability of the balance is not included with this protocol. See enclosed certificate copy. 15

9.0 Uncertainty of the measurement The performance test or calibration test of a pipettor suffers from two different sources of uncertaintes: - the delivering process itself i.e. pipettor itself - uncertaintes related to gravimetric process Both of these sources have to be taken care if special quality control management activities has to be fulfilled.the complete description and calculating formulas are mentioned in ISO-document Guide to Expression of Uncertainty in Measurement ( GUM ), 1993. When the performance test or calibration test is executed in field environment, all uncertaintes cannot take in notice and experience shows that the uncertainty budget resulting from gravimetric measuring apparatus is small related to delivering process itself. The gravimetric apparatus in this case are; analytical balance,barometer, humidity meter and thermometer. It is, of course obvious that these apparatus must be handled according the given instructions eg. ISO 8655 or DIN 12650. The uncertainties caused by the incomplete knowledge of air temperature, pressure and humidity are so small,that is most of the cases it is justified to work standard values rather than measured values. The evaporation of the liquid has to be taken care when volumes below 100µl are measured. 9.1 Temperature In field performance test a reliable ( preferably calibrated ) thermometer is needed and it should be placed as close the balance weighing pan as possible. The temperature alters the water density and is therefore needed to control. The water density or z-factor ca be easily to achieved form a separate water density chart. 9.2 Evaporation The water evaporation speed is necessary to know when volumes below 100µl are measured. Both the DIN 12650 and ISO 8655 are describing that a stop watch must be used when the gravimetric test is done and after the evaporation must be measued within same time limit. There are in the market a special evaporation traps combined with the analytical balances. However it is recommended to measure what is the exact evaporation speed in individual test environments. If a dedicated evaporation trap is not used, a set of additional water containers should be used to reduce the evaporation speed during the gravimetric test. Please note that DIN 12650 is stating that the RH% should be 60-90% around ( close ) the weighing area. The evaporation speed can be experimentally measured by controlling the balance reading up to a certain period of time. Next page descibes a one experiment and corresponds the evaporation speed of 0.07µl in 10 seconds. The time is coming from an average stabilizing time of the balance. 16

Time (h) Weight(g) T0 0,0 108,0805 Balance, Ohaus, HN01 0,5 108,0572 Balance calibrated before the test T1 1,0 108,0446 4 pcs additional moisture traps used. Size: 14 x 54 mm 1,5 108,0319 T2 2,0 108,0192 Glass container: 22 x 90 mm 2,5 108,0073 T3 3,0 107,9946 3,5 107,9817 T4 4,0 107,9631 4,5 107,9519 T5 5,0 107,9402 Total: 5 x 60min. 0.1403 (=T0-T5) Evaporation, average: =300 min. 0.07 µl = 10 sec. 0.1403g=140.3 mg 140.3 mg/300min= 0.467mg/min 0.467mg / 6 = 0.078mg/ 10 sec. 1 mg = 1 µl Evaporation2 108,1 108,05 Weight (g) 108 9.3. Humidity 107,95 Weight(g) 107,9 The relative humidity level should be established on the level of RH 60-90%. The main purpose in this case is 107,85 to create the necessary humidity level just above the weighing vessel. This means that either a specially 1 2 3 4 5 6 7 8 9 10 1 1 designed evaporation traps or additional disitilled water vessels should be placed inside the balance weighing chamber. Tim e (h) 17

9.4. Air Pressure The air pressure is depending of the height distance to the sea level. Following chart is describing the differences volumes referred to sea level. 200 ul Manual Pipette; calibrated and adjusted at NTP to provide 200 ul at 200 ul setting Dead Volume Height at 200 ul Air Pressure (NTP) 540 ul 36 mm 10282 mmh2o Altitude Diff Error (ul) Error (%) (km) 0 1,90 0,00 0,00 1 2,09 0,19 0,10 2 2,32 0,42 0,21 3 2,61 0,71 0,36 4 2,99 1,09 0,55 5 3,49 1,59 0,80 18