General Accreditation Guidance. General Equipment - Calibration and Checks, General Equipment Table

Similar documents
K V A L I F I K. Our accreditation number: NAH /2015. Valid to:

2014 Programme of Training Courses in Measurement & Calibration

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

General Accreditation Guidance. User checks and maintenance of laboratory balances

MET-101 Introduction to Metrology

UKAS Accredited Electrical & Mechanical Calibration Services (Laboratory and On-site)

Department of Electronics and Instrumentation Engineering Faculty of Engineering and Technology SRM UNIVERSITY Course Code Course Title

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits

Measurement And Control Appreciation

GMC-I Service GmbH Calibration Center

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Pressure measurement. Absolute pressure D Gauge pressure C Atmospheric pressure

Accreditation programme for calibration laboratories NAR KL Edition 1

WBEA Standard Operating Procedure

Equipment Accuracy / Selection

Australian Standard. Bourdon tube pressure and vacuum gauges AS

Gas Network Craftsperson

Instrumentation & Data Acquisition Systems

Pressure Measurements

SEMEM4-40 Carrying out the testing and calibration of instrumentation control equipment and circuits

IE011: Process Control & Instrumentation Technology

Technical Procedure for General Laboratory Equipment

STARLOG. Capacitive Water Depth Probe

IE068: Process Instrumentation Technology

Instrumentation & Data Acquisition Systems

Unique mounting features for Thermal Mass Flow Meters. By Martin Raab in cooperation with CaTeC b.v.

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

IE039: Practical Process Instrumentation & Automatic Control

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

Overview IMPEM163S. Carry out preventative planned maintenance on instrumentation and control equipment used in food and drink operations

Level MEASUREMENT 1/2016

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18

Fluid Machinery Introduction to the laboratory measurements

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers

KENYA ACCREDITATION SERVICE

Gas Flow Calibration Basics

High Precision Gauge Pressure Indicators

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

PM105 PRESSURE METER USER S MANUAL

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

CO2 incubator with additional process controls

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

P900. P900, Datenblatt Seite 1 DESCRIPTION

PRESSURE TRANSDUCERS WITH UNIFIED (4-20) ma OR (10-50) ma OUTPUT SIGNAL

Engineering: Measurement Technology Pressure/Level (SCQF level 6)

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.

How to specify a product. Process Sensors and Mechanical Instruments

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Schedule of Requirements THERMODYNAMICS LABORATORY- CHEMICAL ENGINEERING DEPARTMENT

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

HYDRAULICS. H89.8D - Hydraulic Bench

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

DTG - LCD Digital Temperature Gauge USER MANUAL

Selecting the right pressure sensor for your application

Elcometer Viscosity Cup. Operating Instructions

FAS-W. Features. Applications. Automatic condensation hygrometer

Improve Process Reliability

Pressure Measurement

CONTENTS 1. INTRODUCTION DESCRIPTION OF TEST SAMPLE TEST RIG GENERAL ARRANGEMENT TEST SEQUENCE...7

AGILENT TURBOMOLECULAR PUMPS MEASUREMENT METHODOLOGY

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

LEAP CO 2 Laboratory CO 2 mixtures test facility

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

APPENDIX 1. PART 10: Leak Detection Inspector, Level 1, 2 and 3 CERTIFICATION SCHEME FOR PERSONNEL. 3 rd Edition, February 2016

Testing of Active Flow Control actuators at harsh environment

Over 20,000 Strain Gage Target flowmeters installed since 1952.

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas

Best Practice Guide, Servomex 2700

Rotary Viscometers Rotary viscosimeters ST-2020

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

MANUAL KPS Pressure Control Valve

ISO 7218 INTERNATIONAL STANDARD

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Gas Measurement Fundamentals Certification. Curriculum

Level transmitter LT 100

SA121, SA122, SA123, SA128 and SA1219 Self-acting Temperature Control Systems (Dial Adjustment)

Pressure transmitters EMP 2

analytical bulk flow distance FLOW level pressure temperature industrial communication

BAPI Pressure Line of Products - FAQs

SEMETS3-17 Checking and calibrating mechanical inspection equipment

CHAPTER 6 MEASUREMENT OF INSPECTION PARAMETERS. Measurement Ports

INSTALLATION AND OPERATING INSTRUCTIONS

Vortex Meters for Liquids, Gas, and Steam

MA PHIT- Food Certificate Program

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

Agilent Flowmeter ADM1000

HYGRODYNAMICS NARROW RANGE HYGROSENSOR CONTENTS

SPECIFICATION FOR REPEATED LOAD TRIAXIAL (RLT) TESTING FOR PAVEMENT MATERIALS

InstrumentationTools.com

DNVGL-CP-0187 Edition March 2016

Paper 2.2. Operation of Ultrasonic Flow Meters at Conditions Different Than Their Calibration

Transcription:

General Accreditation Guidance General Equipment - and Checks, General Equipment Table January 2018

Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright under the Commonwealth of Australia Copyright Act 1968. NATA s accredited facilities or facilities seeking accreditation may use or copy this publication or print or email this publication internally for accreditation purposes. Individuals may store a copy of this publication for private non-commercial use or copy a reasonable portion of this publication in accordance with the fair dealing provisions in Part III Division 3 of the Copyright Act 1968. You must include this copyright notice in its complete form if you make a copy of this publication. Apart from these permitted uses, you must not modify, copy, reproduce, republish, frame, upload to a third party, store in a retrieval system, post, transmit or distribute this content in any way or any form or by any means without express written authority from NATA.

General Equipment- and Checks For NATA s policies on metrological traceability and equipment please refer to the criteria documents Metrological Traceability and Equipment Assurance, Inhouse and Equipment Verification. The recommended calibration and checking s as presented in this document is to be used in conjunction with the guidance on establishing an equipment assurance program as found in the criteria document Equipment Assurance, In-house and Equipment Verification. January 2018 Page 3 of 13

General Equipment Table Accelerometers Piezoelectric types 3 Servo, strain gauge and piezoresistive types (CD or 0Hz response Air flow nozzles 12 Intercomparison. On use Check against vibration calibrator. 2 On use Check by inversion Initial 12 Check throat diameter. Anemometers 1 Angle gauges 2 then 4 Balances 3 The of Weights and Balances EC Morris and KMK Fen 12 Service. 6 Repeatability check. NATA Technical Note 13. 1 One point check. NATA Technical Note 13. Barometers Initial NATA Technical Note 8. Biological safety 1 AS 2252.4 cabinets (BSC)Class I and Class II for personnel and environment protection Callipers 2 AS 1984 Dial gauges 2 AS 2103 Digestion blocks e.g. blocks or mantles used for Kjehldahl Nitrogen, Chemical Oxygen Demand or metal digestions Dimensional Measuring machines Precision scales 10 Geometric tests 5 Micrometer heads 3 Coordinate Measuring 2 Machines (CMMs) Initial, then Temperature variation check across 12 and after working spaces or recovery check repair or with a difficult to digest maintenance standard/sample e.g. nicotinic acid for TKN digestion. January 2018 Page 4 of 13

Displacement transducers (LVDT) Electrical instruments Digital multimeters (DMM), and other types of meters which measure electrical parameters such as volts, resistance, current, capacitance etc. Including: Analog meters, Data loggers, Chart recorders, Watthour and Varhour meters. 2 6 Intermediate volumetric check (eg ball bar). On use Against length standard. 1 Calibrate over all ranges and parameters of use including calibration across frequency (Hz) of use. 6 Compare with meters of similar resolution. Environmentally controlled enclosures including Incubators, Ovens, Furnaces, Conditioning enclosures (ageing), Refrigerators and Freezers, Water baths Temperature 3 Spatial uniformity, IEC 60068-1; 60068-2-38; 60068-2-39; AS 2853 On use Monitor temperature at at least one point Humidity 3 12 Spatial uniformity of temperature. CO 2 On use Monitor level. Infra Red, Ultraviolet 3 and Visible On use Check operation of the lamps. Pressure / Vacuum On use Monitor level. January 2018 Page 5 of 13

Extensometers Contact and Optical 2 AS 1545. Grading requirements apply. Feeler gauges 2 AS 1665 Flowmeters Differential Pressure meters, orifice meters, venturi meters and Anubar Electronic Thermal, Mass Flow 2 6 Flow or dimensional calibration plus inspection for wear and damage. Pressure to be calibrated as appropriate. 1 Where high temperature or corrosive gases are monitored a shorter is recommended. Laminar flow meters 2 6 Inspect for damage or contamination Sonic Nozzle Reference 0.1% 3 6 Inspect and clean. Working 0.5% 6 6 Inspect and clean. Soap Film 2 Positive Displacement 2 Meters Provers 2 6 Thermometer ice points and pressure readout checks for stability Rotary meter 2 6 Inspect for contamination or damage Rotameters Variable 2 area meters 3 Visual inspection for damage to float edges or ball float for pitting Turbine meters 2 6 Inspect for contamination or damage of turbine blades Turbine meters (Pelton 1 Wheel/Miniature) Vortex shedding 2 6 Inspect for contamination of the bluff body Wet test meters 2 January 2018 Page 6 of 13

Before use Set water level before use Force testing machines Dead Weight 5 AS 2193 Elastic Dynamometer 2 AS 2193 Hydraulic, pneumatic 2 AS 2193 6 Cross head speed (for constant rate of extension machines) and pressure Fume cupboards (cabinets) 1 6 Depending on cabinet type either AS/NZS 2243.8 or AS/NZS 2243.9 Gauge blocks 2 then 4 AS 1457 subsequent Hygrometers (Assmann and sling psychrometers) 10 6 Compare thermometers at room temperature with wick dry. AS 2001.1 Appendix C Thermohygrographs 1 (hair) Weekly Check against a calibrated psychrometer. Electronic types (eg. 1 digital psychrometer) Digital psychrometers 3 (not electrical impedance sensors) 6 Check against a calibrated thermometer at ambient temperature. Electrical impedance humidity probes 1 2 yearly if used only under ambient conditions. Dew or frost point 2 hygrometers Levels (precision) 4 12 12- monthly single point check for electronic levels Load cells and Large 2 AS 2193 scale weighing devices On use If amplification is variable, perform shunt calibration check. Luminance meters & Illuminance meters January 2018 Page 7 of 13

Digital 1 Analogue 2 Manometers Liquid 10 Electronic 1 Masses Stainless steel, nickel 3 chromium alloy Other alloy and iron 2 Class III For proof loading purposes 5 Against calibrated load cell (in house) or weighing device, which achieves the specified accuracy. Micrometers 5 AS 2102 1 Zero, one point (against gauge block) and condition of anvils. Optical electronic 2 distance measurement equipment Optical projectors 5 ph meters Daily or on use Check against two buffer solutions as per manufacturer s instructions. Pipettes (POVAs) * * *Equipment calibration and/or checking requirement will be dependent on the criticality of volume dispensed and/or the repeatability of dispensed volume required by the test method. Pressure equipment Test gauges used for 1 AS 1349 for Bourdon tube types calibration of industrial gauges Industrial gauges not 1 AS 1349 for Bourdon tube types subject to shock loading Industrial gauges 6 months AS 1349 for Bourdon tube types subject to shock loading. Digital pressure 1 gauges Pressure transducers 1 Pressure transmitters 1 January 2018 Page 8 of 13

Calibrators 1 Radiation thermometers including Visible and Infrared Pyrometers Disappearing filament pyrometers 2 Initial test of target size dependence should be performed Initial calibration should include sufficient points to confirm linearity 12 Check at one point in range or at ice point 3 Pyrgeometers 3 Sieves Initial Compliance certificate to AS 1152, BS 410. 12 More or less frequent checks may be required against a reference set or a suitable reference material. Sound measuring 2 devices Including Sound level meters & Noise dosimeters On use Check against acoustic calibrator or pistonphone Acoustic calibrators including Pistonphones and sound sources 1 AS/IEC 60942 6 Intercompare Spectrophotometers and Spectroradiometers 6 Wavelength accuracy, bandpass, absorbance, stray light error, linearity of response, repeatability and matching of cells. On use A blank and at least 2 points on the calibration curve must be checked. Tape measures, rules Tape measures and retractable pocket rules Initial AS 1290.4 24 to 60 Check at maximum length, depending on use and accuracy required. Steel rules Initial BS 4372 6 1 point check within operating range. Thermocouples Base metal type, sheathed 2 For use up to 400 C. For use from 400 C to 1300 C the same immersion depth must always be used (or a greater depth of immersion). Homogeneity must be assessed as January 2018 Page 9 of 13

part of their recalibration. Base metal type, wire 2 For use up to 300 o C. Replace if used above 300 o C. Stored reels 10 Reel of wire 4 samples of wire from end points and middle of reel. Rare metal type 3 3 years or after 100 hours above 500 o C whichever is sooner. Thermocyclers * 12 *Equipment calibration and/or checking requirement will be dependent on test method/kit. When performing a verification, measure temperature uniformity across the block for a number of cycles, logging time spent at temperature using a measurement frequency of at least 2 Hz; Check for excessive overshoot and undershoot (recovery rate) of temperature between temperature point. Perform verification of digital display accuracy as required. Thermometers Liquid in glass 5 Resistance -40 C to 250 C 5 <-40 C and >250 C 2 Measuring instrument AC Bridge type, Reference and Working Measuring instrument DC Bridge type 5 2 6 Check at ice point. NATA Technical Note 19 or against reference thermometer at 1 point in range 6 Check at ice point. 6 Check resistance at ice point. 6 Check at ice point. January 2018 Page 10 of 13

Digital indicating systems, with or without a temperature sensor, hand held or bench type, single and multichannel. (Includes temperature loggers) 2 Timing devices Stop watches, clocks (mechanical and electrical devices) Torque wrench and transducers, Screwdrivers Velocity transducers 3 Volumetric glassware Initial (on commissioning and subject to nature of intended use) 1 6 Check against a reference device at the temperature of use. If used at more than one temperature, choose the most critical temperature. Check at ice point if the facility does not have a reference device. (For data loggers the reference device can not be another data logger of the same type). 6 Check using Telephone Speaking Clock or GPS signal 6 In house cross check of overlapping ranges if possible. 24 Check frequency response and sensitivity. AS 2162.1; BS 1797 January 2018 Page 11 of 13

REFERENCES This section lists publications referenced in this document. The year of publication is not included as it is expected that only current versions of the shall be used. Australian Standards AS 1152 AS 1290.1 Specification for test sieves Linear measuring instruments used in construction - General requirements AS 1290.4 Linear measuring instruments used in construction - Retractable steel pocket rules AS 1349 AS 1457 AS 1545 AS/NZS 1665 AS 1984 AS 2001.1 AS 2102- AS 2103 Bourdon tube pressure and vacuum gauges Geometrical Product Specifications (GPS) - Length standards - Gauge blocks Methods for the calibration and grading of extensometers Welding of aluminium structures Vernier callipers (metric series) Methods of test for textiles - Conditioning procedures Micrometer callipers for external measurement Dial gauges and dial test indicators (metric series AS 2162. Verification and use of volumetric apparatus - General - Volumetric glassware AS 2190 AS 2193 Clinical maximum thermometers - Mercury-in-glass and classification of force-measuring systems AS/NZS 2243.8 Safety in laboratories - Fume cupboards AS/NZS 2243.9 Safety in laboratories - Recirculating fume cabinets AS 2252.1 Biological safety cabinets - Biological safety cabinets (Class I) for personnel and environment protection AS 2853 AS IEC 60942 Other Standards BS 410. BS 1797 BS 4372 IEC 60068-1 Enclosures - Temperature-controlled - Performance testing and grading Electroacoustics - Sound calibrators Stainless steel test sieves Schedule for tables for use in the calibration of volumetric glassware Specification for engineers' steel measuring rules Environmental testing. Part 1: General and guidance; IEC 60068-2-38 Environmental testing - Part 2-38: Tests - Test Z/AD: Composite temperature/humidity cyclic test; January 2018 Page 12 of 13

IEC 60068-2-39 Environmental testing - Part 2: Tests. Test Z/AMD: Combined sequential cold, low air pressure, and damp heat test ISO/IEC Guide 99 International vocabulary of metrology -- Basic and general concepts and associated terms (VIM) ISO/IEC 17025 General Requirements for the competence of calibration and testing laboratories NATA Publications General Accreditation Criteria Metrological Traceability General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification General Accreditation Guidance User Checks and Maintenance of Laboratory Balances General Accreditation Guidance Liquid-in-Glass Thermometers Selection, Use and Checks General Accreditation Guidance The in-situ calibration of Barometers Other Publications The of Weights and Balances EC Morris and KMK Fen Amendment Table The table below provides a summary of changes made to the document with this issue. Section Entire document Amendment This document replaces the former General Equipment Table. The document has been reviewed and updated to reflect the new accreditation criteria documentation structure. January 2018 Page 13 of 13