Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

Similar documents
Table of Contents Want full page pictures with most important parts labeled in the beginning.

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

D R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00

DRAFT. Operating Procedures for the NPDGamma Liquid Hydrogen Target in TA-53, Building MPF-35

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

Pump out insert (condenser and still lines) overnight with turbo pump

Dilution Refrigerator manual

Microscopy Cryogenic WORKSTATION. Performance by Design

RC102 & RC110 Sample in Vacuum Continuous Flow Cryogenic Workstation Cryostats

CRYO INDUSTRIES. Liquid Helium Research Dewars

MB-SP Series Solvent Purification System

APPS-160 Gen 2 User Manual

Procedure of Xenon Transfer

256 Pneumatic Pressure Indicator

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Manual Solvent Purification System. Innovative Technology, Inc.

Variable Temperature Storage Dewar Mount Inserts

Inert Air (N2) Systems Manual

Oxford Instruments Austin, Inc.

Review of the Hall B Gas System Hardware. George Jacobs

B. Assemble the Method-25 sample train with a heated probe/filter.

Cryogenics is the science of ultra low temperatures. Low temperatures are achieved by the liquefaction of gases.

Gas Absorption Draft Standard Operating Procedure

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

Glove Box with Gas Purification System and Digital Control VGB-6. MTI Corporation

Liquid Helium Bath Cryostat

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

The Discussion of this exercise covers the following points:

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

NTF-230 OPERATION MANUAL

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Information Sheet. About this information sheet. Floor and wall loadings. Attachment of process module to floor. PlasmaPro Estrelas100

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

NGP-250/500 Nitrogen Generator Quick Start Guide

VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

NTF-515. Operation Manual

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

BIMA cryogenics (Dick Plambeck, 18jan2007)

Trillium US Inc. Cryo-Plex 10 Cryogenic High Vacuum Pump User s Manual Rev A / November 2015

School of Chemistry SOP For Operation Of Glove Boxes

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

STANDARD OPERATING PROCEDURES

Procedure for Filling 2 T Bruker Magnet with Nitrogen

INTRODUCTION. The Quantum Technology system has the following advantages:, as it does not need plastic gas-bags which are volume-consuming,

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

Chem Bio Liquid Nitrogen Facility

Section 3- Repair and Calibration Procedures

OPERATION MANUAL NTF-15

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

Operating Procedures for Metal Evaporator I

Armfield Distillation Column Operation Guidelines

OPERATION MANUAL NTF-60 Plus

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Compressed Gas Properties. Introduction. Compressed Gas Properties. Definition of a Compressed Gas

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

UNLOADING PROCEDURE OF AN AIRFLOW CONTAINER

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Rejuvenation Instructions

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

EHS Laboratory Guidance: Cryogenic Material

CVI Valve Line. Exceeds the industry s highest standards for reliability and performance

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE

Reclaim Basic Set Up

CO 2. (R744) Service Station Presentation. Contents. Goals Basic functions Specification Recovery R477 Vacuum Charge Safety Other features Prototype

Cryogenic Materials Operating Procedure and Safety Requirements

R E D I C O N T R O L S

Low Temperature Superconductivity

Thanks for choosing EJAI SOLAR product

Equipment Operating Procedure Glove Box

KJ4000 Operating Instructions & Parts Manual NSN

XpressFill XF4500C Counter Pressure Can Filler. Operating Instructions

XpressFill XF4500 Counter Pressure Filler. Operating Instructions

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Glove Box Installation Manual

GDS OPERATING PROCEDURE

Liquid Helium Plant. INSTALLATION, OPERATION and ROUTINE MAINTENANCE MANUAL

VACUUM REGULATORS CONTENTS

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

36H SERIES Combination Gas Valve

PRad Target. Chris Keith JLab Target Group

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

COMPRESSOR REPLACEMENT PROCEDURE (R-134A)

Bulletin TCR-104 & 109 Filling and adding to the Glycol pressure system

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Approved by Principal Investigator Date: Approved by Super User: Date:

Norton Group Toepler Pump Operating Instructions. Version 3.1

Liquid Nitrogen Cooling Accessory CE

Liquid Nitrogen for Duct Cleaning

Standard Operating Procedure

PXRF- Tracer Helium Flow Control System

Dean Pump Self-Priming Chemical Process Pumps

Bubble Chamber Detector System

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

Transcription:

Project: Novel Electron Bubble Particle Detector Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat Version 5.1 (Using side filling line) Hand Processed Changes HPC No. Date Page No. Initials Approved: Date: Operating Manuals Page 1 10/16/2007

PROCEDURE FOR OPERATING NEVIS LHE EBC 1. Purpose 1.1 This Operating Procedure Manual provides instructions for operation of the Columbia (Nevis) LHe Electron Bubble Chamber (EBC) System, with explanation of the LHe/LN 2 Cryostat and associated infrastructure involved. 1.2 The Operating Procedure set forth here is required for the safety of the system and of the operators. All participants in the operations must study the Operating Procedure Manual. 1.3 The Log Book for operations must indicate which version of the Operating Procedure Manual is being followed at the time; who is in charge of operations at any given time (Chief); and who else is assisting. All significant conditions in the system and changes must be noted in the Log Book. 2. Responsibilities 2.1 It is the Chief operator s responsibility to ensure that all operators have the appropriate training, and that all procedures in this Manual are followed. 3. Prerequisites 3.1 All operators must have received the required BNL Training prior to working with the LHe EBC system, and have studied this Operating Procedure manual. 4. Precautions 4.1 Be sure that all necessary system components are functioning correctly prior to start-up. 5. Procedure 5.1 BEFORE START-UP 5.1.1 Make sure that the vacuum jacket of the LHe transfer line has been properly evacuated. Periodic re-evacuation of the transfer line jacket is necessary for efficient LHe transfer. 5.1.2 Be sure that all valves (vacuum, ball and globe) except ball valve 16 (Not usually labeled since it comes from a supplier. This vavle refers to the safety relief valve on the LHe Dewar) on the LHe Dewar are closed, and that all relief valves are installed on the LHe Dewar, LN 2 Dewar, Pumping port and Cryostat. 5.1.3 Do not open any valves if there is below 0 (psig) pressure indicated on the pressure gauges P4 and P5 on the LHe Dewar and LN 2 Dewar. This would allow air to rush into the Dewar and form an ice plug. Operating Manuals Page 2 10/16/2007

5.1.4 Be sure to wear safety glasses and insulated gloves when handling liquid helium and nitrogen. 5.2 EVACUATION OF LHE/LN 2 CRYOSTAT VACUUM JACKET 5.2.1 Connect the flexible coupling from the turbo-pump (TP1) to the connector of vacuum valve 103 extending from the Cryostat. 5.2.2 Open vacuum valve 103 and 101 on the top of turbo-pump (TP1), turn on the turbo-pump, and evacuate the cryostat vacuum jacket until the pressure at the pump drops to 1 10-5 torr or less, as measured by the ionization gauge (TC5). It is recommended to evacuate the system overnight before operating. 5.2.3 Do not introduce cryogen into the chamber with the evacuation valve open when the pump is turned off. This may cause cryopumping of vacuum pump oils into the Cryostat vacuum jacket. 5.3 PURIFICATION (OPTIONAL) 5.3.1 Open the globe valve on gn2 cylinder, and set its regulator at 85 psi (100psia), which is needed to provide the high pressure of gn2 to operate the Purifier. 5.3.2 Open the globe valve and additional valve on the ghe (or gne) cylinder, and set its regulator at 30 psi, then open the valve 34 (for ghe) or valve 35 (for gne) 5.3.3 Switch on the power of Purifier (the power button is located on the right side), the controller will perform an auto-test, and three LEDs will illuminate on the front panel HEATER: STANDBY VALVES: BYPASS TEMPERATURE: LOW 5.3.4 The Purifier is now at room temperature. Open valve 32, 30 and 9, then open slightly valve 7. The purge gas (Ne or He) is flowing slowly through the BYPASS of Purifier and is venting out through Bubbler (checking bubble) to atmosphere 5.3.5 After about 10~15 min, push the HEATER ON/OFF button to start heating the Purifier, the HEATER status will change to OPERATING and the VALVE status will change from BYPASS to PURIFY 5.3.6 It will take about 30-45 min for the Purifier to reach working temperature (LOW to SET, ready to purify the gas 5.3.7 Open valve 33 and 31, then close valve 32. Open the valve of the Flow Meter and set its at ~2.5 CFH 5.3.8 After about 2~3 min, close valve 9 and 7, now the gas flowing from the Purifier is pure to fill into e-bubble chamber. 5.3.9 After finish gas filling, close valve 30 and push the VALVE CONTROL button to change its status from PURIFY to BYPASS. 5.3.10 Push the HEATER ON/OFF button to change its status from OPERATING to STANDBY 5.3.11 TEMPERATURE LED will change to LOW after several minutes. 5.4 EVACUATION AND FILLING e-bubble CHAMBER AT DESIRED TEMPERATURE Operating Manuals Page 3 10/16/2007

5.4.1 Before filling e-bubble chamber, make sure to evacuate the central tube by Turbopump (TP2) for about 30 min, then close firmly valve 8. 5.4.2 Open valve 25, 27 and 28, turn on Turbo-pump (TP2) to evacuate the side transfer tube, e-bubble chamber and gas transfer line for about 10 min and open valve 26 5.4.3 After 10 min, turn on the emission of UHV ionization gauge (TC7), which can only start to work at vacuum below 10-3 torr, to check the vacuum until the vacuum drops below 10-5 torr. 5.4.4 Close valves 26 and 27, and turn off Turbo-pump (TP2) 5.4.5 Open valve 30 and allow the pure gas (He or Ne) entering the e-bubble chamber through the side transfer tube until the pressure transducer P3 reaching 4 psi, close valve 28 and 30. 5.4.6 Open valve 27 and turn on the Turbo-pump (TP2) to evacuate the side transfer tube, e-bubble chamber for about 10 min and open valve 26 5.4.7 Repeat the procedures of 5.4.2 and 5.4.3 5.4.8 Open valve 30 and 7, then open slightly valve 9 and allow the pure gas (He or Ne) purge the transfer line for 2 min. 5.4.9 Close valve 7 and 9, then open slightly valve 25 and allow the pure gas (He or Ne) entering the e-bubble chamber through the side transfer tube until the necessary pressure, measuring by Pressure transducer P3 5.4.10 Close firmly valve 25 and 28, the e-bubble chamber are filled with pressurized pure gas. 5.4.11 To reach small temperature gradient of e-bubble chamber, fill small amount of helium gas into central tube by checking pressure transducer P2 to decrease the temperature gradient of T4 and T5 on the top and bottom flange of e-bubble chamber. 5.5 PUMP/PURGE LHe VESSEL AND COOLING CIRCUIT 5.5.1 Never evacuate the LHe vessel unless the Cryostat vacuum jacket has been previously evacuated, collapse of the LHe vessel wall may occur. 5.5.2 Insert the SS filling tube though helium fill port (C1) into LHe vessel until it reaches the bottom, then rise slightly and seal helium fill port C1. 5.5.3 Connect a Teflon OR SS braided transfer hose to the SS filling tube and connect the other end to the quick connector C19 on the LN 2 Dewar. 5.5.4 Open ball valve 1 and 4 on the control panel. 5.5.5 Open ball valves 104 and helium needle valve 105. 5.5.6 Turn on Turbo-pump (TP2), to evacuate LHe vessel and flow passage through the needle valve, capillary tube and cooling circuit loops, until the pressure drops below 0.1 torr (TC2) to remove any air or moisture. 5.5.7 Close ball valves 104 and 1 and turn off the Turbo-pump (TP2). 5.5.8 Set regulator valve 12 (on the ghe bottle) at 0.5~1 psig, and open ball valve 10, allowing helium gas into the LHe vessel. Operating Manuals Page 4 10/16/2007

5.5.9 After 20 minutes, the 0.5psig relief valve R4 should be open and the flow meter at ball valve 3 may indicate flow. Close ball valve 10 and helium needle valve 105. The LHe vessel is now pressurized (0.5~1 psig). 5.5.10 Close needle valve 11. Be sure to close ball valve 22 on helium vent port C2 of the LHe vessel to prevent air and moisture from entering the vessel. 5.6 PRECOOLING LHe VESSEL WITH LN 2 5.6.1 Open globe valve 20 on the LN 2 Dewar, this valve can be adjusted (1/4~1/2 turn) to obtain the proper liquid flow rate. 5.6.2 Open ball valve 22 on the helium vent port (C2). Liquid nitrogen will be transferred into the LHe vessel directly from the pressurized LN 2 Dewar. 5.6.3 During this cool down, open occasionally the helium needle valve 105, by approximately 3~5 turns, to let LN 2 enter the needle valve, capillary tube, and cooling circuit loops. This procedure also prevents the needle valve from freezing shut as it is being cooled along with the rest of the LHe vessel. 5.6.4 After about 15 min, when T1 on the bottom flange of LHe vessel drops below 81K, it indicates that there are some LN2 cumulated inside LHe reservoir. From then on it takes about 10 min to fill ~10 liters of LN 2, which are sufficient to precool LHe vessel. 5.6.5 Close globe valve 20 on the LN 2 Dewar. 5.6.6 Wait several minutes to let LN 2 calm down, disconnect the Teflon transfer hose from the SS filling tube. 5.6.7 Lift slowly the SS filling tube out off helium fill port (C1) on the LHe vessel, and seal helium fill port (C1) by rubber stopper 5.6.8 Close ball valve 22 on the helium vent port (C2) to prevent air and moisture from entering LHe vessel, LN2 vent out through relief valves R4 (0.5 psig). 5.6.9 Keep the Teflon transfer hose on the quick connector 19 of the LN 2 Dewar. 5.7 LIQUID NITROGEN FILLING TO LN 2 VESSEL 5.7.1 Insert the Teflon transfer hose though one of the nitrogen fill/vent ports (A1) into the LN 2 Dewar until it reaches the bottom. 5.7.2 Open globe valve 20 on the LN 2 Dewar, this valve can be adjusted (1/4~1/2 turn) to obtain the proper liquid flow rate. 5.7.3 LN 2 will be transferred into LN 2 vessel directly from the pressurized LN 2 Dewar, while venting though the other two nitrogen vent ports (A2 and A3). 5.7.4 Approximately 40~45 liters (~30 min) of LN 2 is needed for this cool-down and fill. 5.7.5 When the LN 2 starts to be visible from one of the nitrogen vent ports (A2 or A3), close globe valve 20 on the LN 2 Dewar. 5.7.6 Record readings on temperature T10 and T11 periodically during steps 5.6.3 through 5.7.1. 5.7.7 Remove the Teflon transfer hose from nitrogen fill/vent tubes (A1) on the LN 2 vessel and seal it with rubble stopper. Nitrogen vapor will vent through the other two nitrogen vent ports (A2 and A3). Operating Manuals Page 5 10/16/2007

5.8 LOWER TEMPERATURE OF E-BUBBLE CHAMBER 5.8.1 Adjust the needle valve operator 105 at the top flange of the cryostat to change the LN 2 mass flow rate entering the cooling circuit loops from the LHe vessel. 5.8.2 Open helium needle valve 105 3~8 turns, and open ball valves 3 and 4 on the control panel, in this case helium vapor will vent out directly to atmosphere through check valve R12. The venting flow rate can be monitored either by flow meter FM1 or FM2 on control panel. 5.8.3 Monitor temperature drops of T4, T5, T6, T7, T8 and T9 around the e-bubble chamber. 5.9 REMOVE LN 2 FROM LHe VESSEL 5.9.1 Wait and allow LN 2 to completely cool LHe vessel and e-bubble chamber to lower than 80K (~several hours) by checking temperature T4, T5, T6 and T9. 5.9.2 Insert the SS filling tube though the helium fill port (C1) into LHe vessel until it reaches the bottom and connect a Teflon tube to it. 5.9.3 Set regulator valve 12 on the ghe bottle at 2~5 psig. Open needle valve 11 and ball valve 10. Admit warm gas helium into the LHe vessel. 5.9.4 Seal relief valves R4 (0.5 psig) by relief valves R5 (4 psig) to allow pressurization 5.9.5 LN 2 will start flowing out of this tube once the LHe vessel is pressurized. The LN 2 outflow should be transferred to a suitable container (e.g. a bucket). 5.9.6 Continue this process for about 5 minutes until it appears that no LN 2 is flowing out off the Teflon tube. Note that solid nitrogen has a very large heat capacity (an order of magnitude greater than copper), so even an inch left at the bottom of the vessel will require large amounts of liquid helium to cool to 4.2K. 5.9.7 Remove the SS filling tube from LHe vessel and seal the helium fill port (C1) by rubber stopper. 5.9.8 Close ball valve 10. 5.9.9 Check temperature T1 and wait until it is higher than 82 K. 5.10 PUMP/PURGE LHe VESSEL AND COOLING CIRCUIT 5.10.1 Open ball valve 1, 2, 3 and 4 on the control panel. 5.10.2 Open ball valves 104 and helium needle valve 105. 5.10.3 Turn on Turbo-pump (TP2), to evacuate LHe vessel and flow passage through the needle valve, capillary tube and cooling circuit loops, until the pressure drops below 0.1 torr (TC2) to remove any air or moisture. 5.10.4 Close ball valves 104, 1, 2 and 4, and turn off the Turbo-pump (TP2). 5.10.5 Set regulator valve 12 (on the ghe bottle) at 0.5~1 psig, and open needle valve 11, allowing helium gas into the LHe vessel. 5.10.6 After 20 minutes, the LHe vessel is now pressurized (0.5~1 psig). Confirm flow through flow meter after ball valve 3. Operating Manuals Page 6 10/16/2007

5.11 LIQUID HELIUM FILLING TO LHe VESSEL 5.11.1 Before LHe transfer, maintain the pressure inside LHe vessel at 0.5 psig by setting regulator valve 12 at 0.5 psig, and keeping valves 10, 11 and 13 open. 5.11.2 Unplug the rubber stopper on the helium fill port (C1), insert quickly the LHe transfer line to the bottom of LHe vessel through the helium fill port (C1) and tighten its connector. At the same time, 5.11.3 Close ball valve 16 on the LHe Dewar, and remove the plug of quick connector 14, open fill valve 15, insert slowly the LHe transfer line into LHe Dewar, until it reach the bottom of the LHe Dewar, and tighten quick connector 14. 5.11.4 Close ball valve 10 and open ball valve 22. The boil-off caused by the warm LHe transfer line being inserted into the LHe Dewar will cause immediate pressure rise and transfer liquid helium into LHe vessel. 5.11.5 Set the initial transfer rate of liquid helium at a very slow rate, by regulating the pressure in the LHe Dewar (in the range 0.5~1 psig). This guarantees that no liquid helium is accumulated in the LHe vessel until T3 at bottom of 4K-heat shield is cooled below 20K.This slow transfer makes more efficient use of the enthalpy of the liquid helium as it cools the LHe vessel and the attached radiation shield. 5.11.6 During LHe transfer, open and close helium needle valve 105 several times, to clear the path through needle valve and capillary tube and let liquid helium enter the cooling circuit loops for e-bubble chamber cooling. 5.11.7 When T3 at the bottom of the 4K-radiation heat shield has cooled down to approximately 20K, the helium transfer rate can be accelerated a little 5.11.8 Open helium needle valve 105 3~5 turns, and open ball valves 3 and 4 on the control panel, in this case helium vapor will vent out directly to atmosphere through check valve 12 5.11.9 Monitor temperature drops of T4, T5, T6, T7, T8 and T9 around the e-bubble chamber. When the desired lower temperature has been reached, control exhaust by using the computer control to open and close the solenoid valve installed after ball vave 2. 5.11.10 Check the helium level meter occasionally. The LHe vessel is completely filled (21 on the level meter) using ~ 70 liters of liquid helium for cool down and fill. 5.11.11 It is important not to keep the helium level meter on except when you need to record the level to avoid the extra heat load to the LHe vessel. 5.12 AFTER LIQUID HELIUM FILLING 5.12.1 Upon completion of LHe transfer, temperature T3 at the bottom of 4K-radiation heat shield should stabilize at around 16K. 5.12.2 Open ball valve 16 and release the pressure inside LHe Dewar through relief valve R9 (0.5 psig), then lift slowly the LHe transfer line until it just clears fill valve 15, then close it. At the same time, 5.12.3 Lift slowly LHe transfer line from helium fill port (C1), and then seal it. 5.12.4 Close ball valve 22 at the helium vent port (C2), to prevent air from entering the LHe vessel. Helium vapor is only venting out through relief valves R4 (0.5 psig). Operating Manuals Page 7 10/16/2007

5.12.5 The pressure relief valve R4 is set at 0.5. This maintains a positive adjustable pressure inside the LHe vessel which prevents air entering the vessel, while maintaining a constant drive pressure to send LHe through the needle valve, and capillary tube into cooling circuit loops. 5.12.6 After the LHe transfer is complete, refill the LN2 vessel. 5.12.7 Be sure that the LHe transfer line is warmed up and dry before using again, by purging it with warm helium gas. 5.13 LIQUID HELIUM FILLING TO e-bubble CHAMBER AND CENTRAL TUBE 5.13.1 By condensation with filling warm helium gas into EBC very slowly 5.13 DURING NORMAL OPERATION 5.13.1 Monitor temperature sensors for the holding time of the liquid helium and liquid nitrogen vessels. The normal boil-off rate of the combined total liquid helium volume (about 45 liters) is approximately 1 liter per hour, resulting in a hold time of about 2 days. The LN 2 vessel has a net capacity of about 41 liters, and last more than 5 days. At end of LHe run out, the temperature of the 4K-heat shield will start to gradually increase above 20K, indicating that the heat load from the LHe vessel to the 4K-heat shield is slowly increasing. 5.14 LOWER TEMPERATURE OF E-BUBBLE CHAMBER 5.14. 1. Adjust the needle valve operator 105 at the top flange of the cryostat to change the LN 2 mass flow rate entering the cooling circuit loops from the LHe vessel. 5.14. 2. Open helium needle valve 105 1~2 turns, and open ball valves 2 and 3 on the control panel, in this case helium vapor will vent out directly to atmosphere through check valve 12 5.14. 3. Monitor temperature drops of T4, T5, T6, T7, T8 and T9 around the e-bubble chamber. 5.15 SYSTEM WARM-UP 5.15.1 System warm-up consists of allowing the cryogen to boil away and the cryostat to warm to room temperature. There are two modes: one is normal warm-up and the other quick warm-up. If the system is left unattended for any length of time during warm-up, Caution signage and taping of the area should be in place beforehand. 5.15.2 Normal warm-up: 5.15.2.1 Use heaters placed at the top flange of the e-bubble chamber to increase temperature and boll off LHe. 5.15.2.2 Wait for the liquid helium and nitrogen to boil-off and vent out of the pressure relief valves and vent ports. Operating Manuals Page 8 10/16/2007

5.15.3 Quick warm-up: 5.16.3.1 Open globe valve 13 and set regulator valve 12 at 0.5~1 psig. Connect the gas transfer line to the connector of vacuum valve 103 extending from the Cryostat. 5.16.3.2 Open needle valve 11 and purge the gas transfer line. 5.16.3.3 Open vacuum valve 103 and break the vacuum of the Cryostat vacuum jacket with warm helium gas from ghe bottle. 5.15.4 After the liquid helium has boiled away and the Cryostat has warmed up to room temperature, be sure that all ports are closed to prevent air from entering the vessel. 5.16 LIQUID WITHDRAWAL This procedure will normally be used only under exceptional circumstances: 5.16. 1. Purge the LHe transfer line with gaseous helium. 5.16. 2. Insert the LHe transfer line to the bottom of the LHe vessel through helium fill port (C1), at the same time, 5.16. 3. Remove the plug of quick connector 14, open fill valve 15, insert slowly the LHe transfer line into LHe Dewar, until it reach the bottom of LHe Dewar, and tighten quick connector 14. 5.16. 4. Open vent valve 17 on the LHe Dewar. Liquid helium transfer from the LHe vessel will take place as the helium vapor inside the LHe vessel is vented from the vent valve 17. 5.16. 5. When the LHe transfer is complete, the temperature measured by temperature sensor T1 at the bottom of LHe vessel will increase abruptly. 5.16. 6. Wait a few minutes for helium boil-off to settle down, remove LHe transfer line until it just clears the fill valve 15, and then close the fill valve 15. Remove the line completely from the LHe Dewar. 5.16. 7. Close the vent valve 17, open ball valve 16 and let the LHe Dewar vent from ball valve 16. 5.16. 8. Install the plug of quick connector 14 and remove the LHe transfer line from helium fill port (C1). 5.16. 9. Seal helium fill port C1, close ball valve 22 on the helium vent port C2, to prevent air from entering the LHe vessel. 6. Documentation 6.1 EBC Log Book 7. Attachments 7.1 P&ID diagram 7.2 Flow diagram Operating Manuals Page 9 10/16/2007

V9 R5 30 psig V7 VENT TO ATMOS VENT TO ATMOS C19 22 psig R10 V20 V21 P5 LHe FLEXIBLE TRANSFER LINE 250L LN2 DEWAR 0.5 psig 8 psig 10psig R9 R8 R7 V17 V16 P4 250L LHe DEWAR C14 V15 V37 N2 LHe FLEXIBLE TRANSFER LINE LN2 TEFLON TUBE V30 Getter FM3 V31 H2 V32 V33 V36 Ne V28 V29 V35 V27 TP2 He V25 V26 VI TC7 UHV V34 M He VENTS P3 150 psig T10 T14 T4 T6 T7 R2 V22 F2 A1 PR LN2 V10 P2 C2 LHe C1 T1 V105 T15 V8 ghe or LHe R6 150 psig 0.5 psi LHe T13 H1 H2 LT A2 LN2 V104 R4 4K HEAT SHIELD F1 A3 T5 T3 T11 80K HEAT SHIELD He VENT 4 psig R3 T12 R1 4 psig VI TC1 V103 T2 T8 T9 TO ATMOS N2 VENTS N2 VENTS MP3 FM1 R12 V1 V2 V3 V4 FM2 V5 V6 P1 VI TC2 TC6 TC5 Convect VI VI BA Ionization Gauge Gauge V101 TP1 FLOW METER MANUAL VALVE M H MONITOR HEATER PUMNG PORT VI VACUUM GAUGE M RELIEF VALVE LT LIQUID LEVEL METER FILL/VENT PORT MECHANICAL PUMP REGULATING VALVE QUICK CONNECTOR PRESSURE GAUGE TEMP. SENSOR RUBBER TUBE+SLITS CAP TURBO-PUMP LIQUID Figure 7.1 P&ID diagram Operating Manuals Page 10 10/16/2007

SS (1/4" O.D., ~2 Ft.) R5 (30 psi) Flexibel coupling, 1" I.D. ~50" V9 V7 V8 SS tube (1/4" O.D., ~8 Ft.) To TP2 P2 P V10 R6 (150 psi) V104 V15 Pressure gauge (30" Hg ~15 psig) P4 Relief valve P (0.5 psig) R9 V16 Ball valve Quick connector (1/2") C14 Fill valve 1/2" MPT Relief valve (10 psig) R7 V17 R8 Vent valve 1/2" FPT Relief valve (8 psig) C19 R10 Relief valve (22 psig) P5 P Pressure gauge (0-100 psig) Globe valve V203/8" FPT V21 Globe valve 3/8" FPT SS tube (1/4" O.D., ~8 Ft.) H2 V36 V29 V28 V26 V V22 V27 R2 (150 psi) PR P TC1 V25 0.75 O.D. P P3 LHe fill/vent port (C1 and C2) TC7 UHV gauge 16.0 O.D. 14.0 O.D. 6.0 O.D. 5.0 O.D. 0.50 O.D. T10 13.0 O.D. 12.0 O.D. 0.75 O.D. X.028 W 0.37 O.D. Helium Valve R4 (0.5 psi) R3 (4 psi) T12 V103 R1 (4 psi) Nitrogen Vessel 25 M.L.I. Helium Vessel 25 M.L.I. LN2 fill/vent ports (A1, A2 and A3) Flexibel coupling, 1" I.D. ~50" Flexible pumping line 1.5" I.D., ~80" Flow meter 0~1.2 L/s 1/2" MPT To Pump TP2 R12 V2 V1 Digital flow meter/control FM2 0~20 SLM 1/4" MPT FM1 Check valve (~ 0.15 psig) V3 V5 Copper loops 1" O.D. ~60" V4 V37 V30 V35 V34 T1 T14 T13 T2 TC6 Convect Gauge 10-3~760 torr 1/8" FPT male V TC5 Ionization gauge 10-9~10-3 torr V 2.75" CF-F P1 P V6 LHe Dewar (~250L) LN2 Dewar (~250L) N2 V31 Getter FM3 V32 V33 Ne He 9.0 O.D. 11.0 O.D. 14.0 O.D. T4 T8 T15 T6 T9 T7 T5 T3 T11 25 M.L.I. Vacuum valve with NW 40 TP1 V101 Reducer 4-1/2" CF to 2.75" Pfeiffer TSU 071 Turbomolecular pump ~60 L/s, ~10-9 torr With DUO 10 Backing pump ~10 CFM TP2 TC2V Reducer 4-1/2" CF to 2.75" Pfeiffer TSU 071 Turbomolecular pump ~60 L/s, ~10-9 torr With MVP 035 Backing pump ~5 CFM Figure 7.2 Flow diagram Operating Manuals Page 11 10/16/2007