WIDDER TOOLS. HPICAL Calibration Station 1 (203)

Similar documents
HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

HPIC D Hydrostatic Test Pump PRODUCT INFORMATION AND OPERATING INSTRUCTIONS

Operating Instructions Models and Hydrostatic Test Pumps

Operating Instructions Models 36200, , 46200, Hydrostatic Test Pumps

Operating Instructions Models and Hydrostatic Test Pumps

Operating Instructions Models and Hydrostatic Test Pumps

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

Operating Instructions Models and Hydrostatic Test Pumps

Pressure Differential Automatic Manifold For LPG. PDH1000 Series. LPG Installation & Operations Instructions

Operating Instructions Model and Hydrostatic Test Pump

Rejuvenation Instructions

MANUAL BE SERIES Test Benches

ELECTRONIC PLC SERIES Test Benches

Operating Instructions Model and Hydrostatic Test Pump

Installing ASI Umbilical Delivery Systems

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Speed control assembly model A-1

DS05C,D,G Dial Set Pressure Regulating Valves

PNEUMATIC PRESSURE CONTROLLERS

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

Pneumatic Powered - Plunger Pumps

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2"-12" BERMAD. Application Engineering

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

Halsey Taylor Owners Manual 4410 Freeze Resistant Tubular Fountain STOP!

Bermad Pressure Reducing. Model: 42T

CS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information

PXRF- Tracer Helium Flow Control System

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

IMPORTANT SAFETY INSTRUCTIONS READ AND FOLLOW ALL INSTRUCTIONS SAVE THESE INSTRUCTIONS

Glove Box with Gas Purification System and Digital Control VGB-6. MTI Corporation

FloWash. Industrial Filter Cart

D05 Pressure Regulating Valves

PVQ 13 A2 R SE 1 S 20 C-11 D

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

accidents which arise due to non-observance of these instructions and the safety information herein.

Halsey Taylor Owners Manual STOP!

CART FOR CENTURION 1500+

TOOL BOX 033- HOSEREEL SERVICE / FAULT FINDING

SINORIX Valve Reconditioning

INSTALLATION INSTRUCTIONS

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Float Operated Level Controllers

ACCESSORY KIT INSTALLATION INSTRUCTIONS

TECHNICAL DATA. Pressure Regulation 531a. April 24, 2009

HGC-200 Hydro Gas Chlorinator

TECHNICAL DATA. Page 1 of 12

Operation Manual Piston Sensed Gas Pressure Regulators

WW-730. Pressure Sustaining/Relief Control Valve

PAGE (Rev. A - 06/17)

OPERATING AND MAINTENANCE MANUAL

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

LUDLUM MODEL 239-1F FLOOR MONITOR. Revised December 2010

Anderson Greenwood Series 400 Diaphragm Pilot Operated Safety Relief Valves. Installation and Maintenance Instructions for

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Installation/Care/Use Manual Soft Sides Fountains with FLEXI-GUARD

Installation of Your SprayMaster System

Gas Conversion Kit Pilot Ignition Heaters

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

The Discussion of this exercise covers the following points:

Suggested Installation & Operating Instructions for Sidewinder Pumps

2.5 & 5.0 GALLON PRESSURE TANKS

DPC-30 DPC-100. Reference Manual

36C/36D HSI, DSI Proven Pilot Gas Valves INSTALLATION INSTRUCTIONS

Installation Instructions

User s Guide. Pressure Controller for liquid delivery systems

J Air and Water Kit Instructions Part# 02584

Freedom8 ShoeBox Compressor Manual

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

User s Guide. Vacuum Controller for liquid delivery systems

UltRo Dual Flow Reverse Osmosis Water System

Pressure Dump Valve Service Kit for Series 2300 Units

1020 Industrial Drive, Orlinda, TN fax

Installing a Probe. Typical Installation

TECHNICAL DATA CAUTION

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

REPAIR PART NUMBER NUMBER DESCRIPTION 0137 ELE 0136 ELE 12 VOLT SOLENOID

INSTRUCTION MANUAL. January 23, 2003, Revision 0

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL

LINDGREN-PITMAN, INC. LS-3 / LS-4 Line Setter

Interface Devices, Inc. Hydraulic Mini Mule

Operation Manual. Diaphragm Sensed Gas Pressure Regulators

CSA Sample Draw Aspirator Adapter Operator s Manual

User Manual for the Mars Calibration Bench

The X-11 Molding Chamber

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

Parts List. Description. Additional Considerations. Installation Instructions. Dual/Elite Hydraulic Treadmill Kit (supplied by EPI)

LEAK TROUBLESHOOTING

Operator: Save these instructions for future use!

Installation Instructions for NP100 Series Nitrogen Control Panels

Cart-Mounted Airless Package

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

DS06D,G Dial Set Pressure Regulating Valve

Transcription:

WIDDER TOOLS HPICAL-15000 Calibration Station WWW.WIDDERTOOLS.COM 1

1. TEST SET-UP a. Be sure air and water supply lines are installed per basic set-up & installation instructions supplied with calibration station. (available at https://widdertools.com/wp-content/uploads/2017/06/hpical-15000- Install-Manual.pdf) b. Be sure Air Pressure Regulator is backed all the way off, (Counterclockwise) and Air & Water Valves are turned to OFF (counterclockwise). c. Attach system or gauge that is to be tested or calibrated to the Calibration Station using one of the following two options: I. MAIN TEST PORT a. Remove Adapter Cap (PC-1032) from Main Test Port on the right side of the Calibration Station, and store safely. b. Be sure Auxiliary Plug (PC-1031) and Gland Nut (PC-1020-3) are installed in Auxiliary Test Port on top of the unit, and tightened firmly. c. Attach Test Hose (PC-1019) to the Main Test Port and tighten securely. d. Attach appropriate adapter to the high pressure output port of the system to be tested/calibrated. NOTE: Hose end is 9/16 M fitting. One adapter to ½ NPT (PC-1030) is supplied, others are available if required. ***Always use Teflon tape on NPT connections. e. Plug input port on the system to be tested (this is a safety precaution) f. Open Calibration Station Bleed Valve at LEAST ¼ - ½ turn from closed to ensure there is no residual pressure in the system. g. Turn on Calibration Station Digital Gauge. (press POWER button on gauge) h. Zero the Calibration Station Digital Gauge (press ZERO button on gauge and hold for 3 seconds) i. Zero the gauge on the system to be tested (if applicable). j. Start Data Logging at this time (see DATA LOGGING instructions, pages 6 i. Be sure all data on the gage is saved on a computer and deleted from the gage before starting another logging session. ii. Be sure the date and time are correct. iii. Be sure to set logging interval (or gap ) to desired amount. iv. The % sign on the gage should be flashing to indicate the data logging is active. k. Be sure water input and/or Isolation valve on system to be tested are closed or plugged tightly. ***Always use Teflon tape on NPT connections. l. Close the Bleed Valve on the system to be tested tightly. WWW.WIDDERTOOLS.COM 2

m. Close the Bleed Valve on the Calibration Station tightly. n. Turn Calibration Station Water Valve to ON o. Crack open the Bleed Valve on system to be tested slightly, removing as much trapped air as possible; then close tightly. p. Crack open the Bleed Valve on Calibration Station slightly, to remove any residual trapped air, then close tightly. q. Repeat steps n and o as required to be sure as much air as possible is removed form entire system. II. AUXILLARY TEST PORT a. Remove Auxiliary Plug (PC-1031) and Gland Nut (PC-1020-3) from auxiliary test port and store safely. b. Remove Test Hose (PC-1019) from the Main Test Port and store safely, then install Adapter Cap (PC-1032) onto Main Test Port Adapter (PC-1029) and tighten firmly. c. Attach appropriate Test Adapter to Auxiliary Test Port and tighten firmly. NOTE: Calibration Station is supplied with adapter for ¼ NPT. Others are available if necessary. Ref: Test Hose end is 9/16 M fitting. ***Always use Teflon tape on NPT connections. d. Install gauge to be tested/calibrated into Test Adapter & tighten firmly. ***Always use Teflon tape on NPT connections. e. Close the Bleed Valve on the Calibration Station tightly. f. Turn Calibration Station s water valve to ON g. Crack open the Bleed Valve on Calibration Station slightly to remove any trapped air, then close tightly.. h. Repeat step g as required to be sure as much air as possible is removed from the system. Once the system or gauge to be tested or calibrated is connected to the Calibration Station using one of the two above methods, follow these operation instructions, depending on what type of test you are performing: WWW.WIDDERTOOLS.COM 3

2. OPERATION - STANDARD CALIBRATION OR PRESSURE TEST a. Ensure that Calibration Station water valve is ON, and air pressure regulator is backed ALL THE WAY off. (Counter-clockwise) b. Back off Pressure Adjust Valve all the way. (Counter-clockwise) c. Turn Calibration Station air valve to ON. d. Slowly increase air pressure by turning Air Pressure Regulator knob clockwise. Pump will begin building test pressure, noted on the System Pressure and Digital Logging Gauge. e. Slowly increase test pressure until you are approximately 500 psi BELOW desired test pressure. f. Back off the air pressure regulator all the way. (Counter-Clockwise) NOTE: This isolates the pump from the test system, and typically results in a slight test pressure drop as the check valves seat themselves. g. Use the Pressure Adjust Valve to set test pressure to the desired point. (Move clockwise to increase pressure, counter-clockwise to decrease pressure.) This adjustment valve provides very fine control adjust slowly! h. Calibrate gauge according to specific instructions for the gauge being calibrated, using the Calibration Station to build and release pressure as required. Use pressure adjust valve for fine control. Use bleed valve (cracked open slightly, then closed firmly) and air pressure regulator (turned CW to build pressure, then backed off) for gross adjustments. (SEE PAGES FOR SPECIFIC INSTRUCTIONS FOR THE WIDDER HPIC- 10000-D CABINET/ASHCROFT GAUGE) i. When gauge calibration procedure is complete, turn Calibration Station Air and Water Valves to OFF and use Bleed Valve to release system pressure before disconnecting any fittings. j. Stop Data Logging on the Calibration Station gage and output the data to a computer according to the DATA LOGGING Instructions (pg ) 3. OPERATION CONTINUOUS LEAK TEST a. Ensure that Calibration Station Water Valve is ON, and Air Pressure Regulator is backed ALL THE WAY off. (Counter-clockwise) b. Back off Pressure Adjust Valve all the way. (Counter-clockwise) c. Turn Calibration Station Air Valve to ON. d. Slowly increase air pressure by turning Air Pressure Regulator knob clockwise. Pump will begin building test pressure, noted on the System Pressure and Digital Logging Gauge. WWW.WIDDERTOOLS.COM 4

. e. Slowly increase test pressure until desired level is reached. f. At this point the Calibration Station will maintain set pressure so physical leak test can be performed. If system leak is causing drop in pressure, Calibration Station will cycle on as required to try and maintain set pressure. g. When leak test is complete, back off the Air Pressure Regulator all the way (Counter-Clockwise). Turn Calibration Station Air and Water Valves to OFF. Then use Bleed Valve to release the system pressure. WWW.WIDDERTOOLS.COM 5

Additel Digital Gage Data Logging WWW.WIDDERTOOLS.COM 6

Ashcroft Digital Gage Calibration (the Ashcroft Gage is found on all HPIC-10000-D Hydrostatic Test Units) 1. Press the MENU key on the keypad. 2. Press the (up arrow key) or (down arrow key) on the keypad until the word CONFIG appears. 3. Press ENTER. 4. Enter user password if it has been programmed. 5. Press (up arrow key) or (down arrow key) until the word RECAL appears. 6. Press ENTER. 7. The gauge will now flash between INPUT and unit of measure on the lower line and.00 on the top line. Apply zero pressure to the gauge. 8. Press ENTER. Zero pressure is now set. 9. The gauge will display full-scale pressure (10000). Apply full-scale pressure to the gauge. 10. Press ENTER. Full-scale pressure is now set. 11. The gauge will now display mid-scale pressure (5000). Apply mid-scale pressure to the gauge. 12. Press ENTER. Mid-scale pressure is now set. 13. After zero, full-scale and/or mid-scale or factory default calibration have been set, the word SAVE appears on the gauge display. 14. Press ENTER to finalize calibration. WWW.WIDDERTOOLS.COM 7

WIDDER HPICAL-15000 Calibration Station Parts Top View WWW.WIDDERTOOLS.COM 8

WIDDER HPICAL-15000 Calibration Station Parts Inside View WWW.WIDDERTOOLS.COM 9

WIDDER HPICAL-15000 Calibration Station Accessories WWW.WIDDERTOOLS.COM 10

Item Description Part # Top View 1 Cabinet PC-1001 2 On/Off Valves PC-1006 3 Pump Mounting Screw PC-1040 4 Washer (Outer) PC-1043 5 Pump Standoff PC-1045 6 Washer (Inner) PC1043 7 Locknut PI7000-27 8 Air Filter/Regulator PC-1009 9 Gage Trim PI7000-214P1 10 Digital Logger Gage, 15K PC-1003 11 Aux. Port Block Mounting Screw PC-1042 12 Washer PC-1044 13 Auxiliary Plug PC-1031 14 Gland Nut PC-1020-3 15 Pressure Adjust Valve PC-1017PA 16 Valve Mounting Screw PC-1042 17 Bleed Valve PC-1017BV 18 Manifold Mounting Screw PC-1041 19 Washer PC-1043 20 Manifold Standoff PC-1046 21 Cap Screw PI7000-212 22 Washer PI7000-213 23 2 ½ Flange PI7000-11A 24 2 ½ -160 psi Gage PI7000-11 25 Flange Screw, 2 ½ Gage PI7000-20 WWW.WIDDERTOOLS.COM 11

Item Description Part # Inside View 26 Manifold PC-1004 27 Tube, Manifold to Aux. Block PC-1016A 28 Test Port Adapter, 9/16 M PC-1029 29 Adapter Cap PC-1032 30 Manifold Mounting Screw PC-1042 31 Washer PC-1044 32 Tube, Manifold to Bleed Valve PC-1014A 33 Bleed Valve PC-1017BV 34 Adapter PC-1026 35 Flareless Adapter PI7000-78 36 Bleed Tube (St. Steel) PI7000-82 37 Bleed Tube (Plastic) PC-1048 38 Bleed Tube Grommet PC-1047 39 Tee PC-1050 40 Aux. Port Block PC-1011 41 Plug Assy. PC-1021A 42 Digital Logger Gage PC-1003 43 Air Filter/Regulator PC-1009 44 Tube, Filter Drain PC-1049 45 Filter Drain Tube Grommet PC-1047S 46 Air Inlet Bulkhead PC-1008 47 Hose, Air Inlet to Valve PC-1051A 48 Water Inlet Bulkhead PC-1007 49 Reducer Bushing PC-1033 50 Inline Strainer PC-1039 51 Hose, Air Valve to Regulator PC-1052A 52 Air On/Off Valve PC-1006 53 Hose, Air Regulator to Pump PC-1055A WWW.WIDDERTOOLS.COM 12

Item Description Part # 54 Hose, Water Inlet to Valve PC-1053A 55 Water On/Off Valve PC-1006 56 Hose, Water Valve to Pump PC-1054A 57 Pump PC-1002 58 Hose, Air Regulator to Gage PC-1056A 59 Tube, Pump to Check Valve PC-1013A 60 2 ½ -160 psi Gage PI7000-11 61 Check Valve PC-1010 62 Tube, Check Valve to Manifold PC-1015A 63 Digital Gage Bracket, Upper PI7000-210 64 Digital Gage Bracket, Lower PI7000-211 65 Tube, Manifold to Gage PC-1012A 66 Tube, Manifold to Pressure Adj. PC-1014A 67 Pressure Adjust Valve PC-1017PA Accessories 68 Gage Data Cable PC-1003-DC 69 Gage Power Cable PC-1003-PA 70 Test Hose PC-1019 71 Drain Bottle PC-1057 72 Clamp, Drain Bottle PC-1058 73 ¼ NPT Adapter PC-1020-1/4NPT 74 C Cabinet Adapter PC-1030 WWW.WIDDERTOOLS.COM 13