Safety and Longterm Accuracy of Bourdon Tube Pressure Gauges

Similar documents
WIKA INSTRUMENT CORPORATION

Instrument Operating Manual

Notes on equipment protection per EN/IEC and NEMA For Bourdon tube or diaphragm pressure gauges

A hose layline contains important information for specifying the replacement assembly: manufacturer, hose trade name, working pressure and hose ID.

Pressure transmitters EMP 2

Bourdon tube pressure gauge, stainless steel For extremely low ambient temperatures down to -70 C Model PG23LT, with case filling

INSTALLATION AND OPERATING INSTRUCTIONS

Notes on equipment protection to DIN EN and NEMA

TECHNOLOGIES. M Series. Positive return metering pumps [API675]

Expansion Thermometers Stainless Steel Series, Model 70

Sense Element Pump Ripple Fatigue

DS/PI-14 DIAPHRAGM SEAL TEMPERATURE ERROR

Bursting strength of paper. 1. Scope

SOLFRUNT Model /2" (63mm) and 4" (100mm) Stainless Steel Gauges

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating Instructions

P499 Heavy Duty Pressure Transducer

Digital pressure gauge Model CPG500

SEMATECH Provisional Test Method for Pressure Cycle Testing Filter Cartridges Used in UPW Distribution Systems

Hatch cover securing and tightness

TEK-SUB 4800B. Submersible Level Transmitter. Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys LEVEL

Model /2" (63mm), 4" (100mm) and 6" (160mm) Open Front Stainless Steel Gauges

Sizing Pulsation Dampeners Is Critical to Effectiveness

Diaphragm monitoring system with threaded connection For the chemical and petrochemical industries Model DMS34

Bourdon tube pressure gauges with glycerine filling, for chemical applications EN 837-1

Mining scalers are used in. Designing a stable suspension. F l u i d P o w e r. Demanding application requires unique engineering approach.

When does a firefighter realize the value of an engineer? In the middle of a 1,400 F inferno

High-performance submersible pressure transmitter For level measurement Model LH-10

McDaniel. General Service Gauges. (Brass Internals) Stainless Steel Case 1 1 /2 and 2 1 /2 Gauges. All Brass 2 and 2 1 /2 Gauges M C

Pressure measurement. Pressure gauges

Differential pressure gauge For the process industry Model , all-metal media chamber

PRESSURE & TEMPERATURE GAUGES PRESSURE TEMPERATURE. Catalogue H20200a

Certified Accuracy. Ring Force Gauge. Models and Capacities Available. Design and Principle of Operation

Digital pressure gauge Model CPG500

AN31E Application Note

Accu-Drive Pressure Gauges

Mechanical Pressure Measurement

The Criticality of Cooling

Buying more than 11 gauges? Call Lesman for additional discounts.

Bourdon Tube Pressure Gauges Ultra High Purity (UHP) Series Type

Fluid Machinery Introduction to the laboratory measurements

High Precision Gauge Pressure Indicators

FLUID POWER FLUID POWER EQUIPMENT TUTORIAL ACCUMULATORS. This work covers part of outcome 2 of the Edexcel standard module:

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Abstract. 1 Introduction

Product Catalogue. P-SERIES Pressure Gauges

A FINITE ELEMENT PROCEDURE TO MODEL THE EFFECT OF HYDROSTATIC TESTING ON SUBSEQUENT FATIGUE CRACK GROWTH

PULSATION DAMPENER / SUPPRESSOR SERIES PDS PROVIDES MULTIPLE SYSTEM SAFEGUARDS

HCMTCB MATERIALS SAMPLING & TESTING PERFORMANCE CHECKLIST

TECH TIPS: ROPE DEFECTS

Royal Process Gauges, ±0.5% Accuracy 316 SS Tube and Socket Dial Size

Structural Design of Tank Weighing Systems

Spilt body Flange ball valve. TC-205MFF-PN1640 User Manual English Version. Document No: TC-205MFF-PN1640.Ur-manual. Date: 2007/04/2617. Version: 1.

High Purity & Ultra High Purity

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

Guide for Evaluating Your Hose Assembly Supplier

Instrumentation & Data Acquisition Systems

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL

HYDRAULIC INTENSIFIER Max. Pressure: See Intensifier Data Plate Weight: 16 Lbs. (7.26 kg)

REV0709. Parts List. Hydraulic Cylinder Model MODEL C SINGLE-ACTING, SPRING RETURN HYDRAULIC CYLINDER. Max. Capacity: 55.2 Tons at 10,000 PSI

Unit 24: Applications of Pneumatics and Hydraulics

TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES

MB610 Industrial Pressure Transmitter

3. REQUIRED EQUIPMENT

BOSHART INDUSTRIES Telephone: Toll Free: Fax:

Vibration Analysis and Test of Backup Roll in Temper Mill

Type N-10, N-11 Hazardous Area Non-incendive Transmitters

Vibration Related Failures of Small-Bore Attachments

Wave Motion. interference destructive interferecne constructive interference in phase. out of phase standing wave antinodes resonant frequencies

3 1 PRESSURE. This is illustrated in Fig. 3 3.

Section 3- Repair and Calibration Procedures

Redesign of Centrifugal PLACE FOR Compressor TITLE. Impeller by means of Scalloping

Pressure Relief Valve Instruction Manual

Installation and Maintenance Instruction Manual

MANUAL BE SERIES Test Benches

QUALIFICATION ENVIRONMENTS FOR AMP PROGRAM PROPELLANT TANK ALLIANT TECHSYSTEMS, P/N 80520

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

THE PROCESS OF JOINT INTEGRITY

Nominal size dim 63mm. Option with case dim 50 mm. Aluminium, white with black lettering

Upgrade Increased Seaming Spindle Bushing Wear Life & Repair Suggestion

Vibration isolation system 1VIS10W. User manual

Truck-mounted Mass Flow Metering for LPG Delivery

SPECIFYING MOTIONLESS MIXERS

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

PumpAgents.com - Click here for Pricing/Ordering Nominal psi (bar) Cut-In Cut-Out

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

IAPMO GUIDE CRITERIA FOR BALL VALVES IAPMO IGC PURPOSE

ENGINEERING TECHNICAL REPORT

Cost reduction for conveyor systems through self-adjusting chain wheel

Aluminium, white with black lettering. Bar. Option with unit bar/pa, bar/psi, kpa, MPa, psi

Waves. harmonic wave wave equation one dimensional wave equation principle of wave fronts plane waves law of reflection

COMPRESSION TUBE FITTINGS

P-04 Stainless Steel Corrugated Hoses and Metal Bellows Expansion Joints

Simplified Digital Hydraulic Testers Operating Manual. Oil Solutions. DHC DHT HP

Pressure Gauges. Version 2018 Revised 01/2018

Powers Controls D Day-Night Thermostats

VISCOMETER, OFI, 8-SPEED, MODEL 800 PART No

THE TACOMA NARROWS BRIDGE FAILURE Revision A

SPWE Series. Setup Instructions. Version 2.0

Precision Rotary Ball Screw

WHITE PAPER Overview of Diaphragm and Chemical Seals

Transcription:

Technical Brief Safety and Longterm Accuracy of Bourdon Tube Pressure Gauges The demands of quality standards throughout the world such as ISO 9000 and BS5750 have led to the major pressure gauge users demanding instruments that stay in calibration longer. esources are no longer available to recalibrate pressure gauges on a regular basis. Ultimately wear on a gauge movement or plastic deformation of the Bourdon tube can lead to failure of the measuring element. This report outlines some of the development work being researched at WIKA, in conjunction with the major gauge users, to improve the safety and long-term accuracy of Bourdon tube pressure gauges. ecommendations on the best type of pressure gauge for countering the effects of vibration and pulsation in process plants are also included in this report. Fig. 1 Components of a Bourdon tube pressure gauge

1.1 Introduction Bourdon tube pressure gauges represent about 90% of the market for mechanical pressure measurement. But mechanical vibration and pulsating pressure are common causes of failure for this type of gauge. This report looks into the reasons for the failure of Bourdon tube gauges and covers a number of techniques that can prolong the life of the instruments. A close examination is also made of the implications of using damped movements and liquid-filled cases to prolong the life of a gauge. All the research data in this report is based on 4 (100 mm) stainless steel Bourdon tube gauges. 1.2 Common frequency ranges When designing a pressure gauge it is important to take into account some of the common frequencies encountered within the industry, thus preventing these frequencies from matching the resonant frequency of the pressure gauge. Table 1 shows some examples of common frequencies. 2. Initial examination of Bourdon tube measuring elements The combined effect of pressure loading and elastic deformation due to mechanical vibration produce areas of high stress within a Bourdon tube. These areas of high stress initially cause the calibration of the instrument to shift and can ultimately lead to the failure of the tube itself. Table 1 Application Frequency Hz European Alternating Current - Single Pole Motors 50 USA Alternating Current - Single Pole Motors 60 Typical Hydraulic Drive Assembly 120 Five Piston Pump with 50Hz Motor at 1440 PM 24 High Pressure Cleaning Machine at 2000 PSI 300 Common equipment frequencies found in the process industry. When designing pressure gauges it is important that the resonant frequency of the gauges should be different from the common frequencies found in the process industry. The following section looks at these areas of high stress and shows how research into new techniques of Bourdon tube production has led to the development of a much stronger and safer tube. 2.la Location of the maximum tube stress Using FEM (finite element method) calculations, the stresses of a Bourdon tube when under pressure (Fig. 2) and when subjected to mechanical vibration (Fig. 3) were determined. In Fig. 2 the maximum stress caused by steady internal pressure is uniform over the complete length of the tube with the exception of the end sections. The maximum stress where the Bourdon tube is attached to the connection begins at an angle of 30 along the tube from the connection. Fig. 2 Stress of a Bourdon Tube under pressure. The maximum load is over the whole length of the tube with the exception of the ends. 2

Fig. 3 Stress of a Bourdon tube due to mechanical vibration. Maximum load over 30 to 60 from the connection. In Fig. 3 the maximum stress caused by mechanical vibration occurs at an angle of between 30 and 60 along the tube from the connection. It is therefore clear that the area of maximum stress occurs where the effects of the pressure and vibration stresses combine in the area of 30 to 60 along the tube from the connection. The greatest material stress in both cases is at the narrow edge of the Bourdon tube, see Fig. 4, radius. Contrary to common beliefs, the joint between the tube and the gauge connection is in an area of reduced stress. Fig. 4 The greatest tube stress occurs at the narrow edge of the tube -. 2.lb Techniques to strengthen the area of maximum stress A - Cold working and annealing the Bourdon tube By cold working the Bourdon tube during the manufacturing process it is possible to increase the yield point of the tube from 250 N/mm 2 to 400-600 N/mm 2. This stronger tube gives an improved performance and service life. Furthermore it was found that by annealing the stainless steel tube at 950 F for 3 hours, the material stress spikes at the edges of the tube, created from the tube forming process, were significantly reduced. B - The WIKA Bourdon tube design The area of maximum stress on the tube between 30 and 60 was found to be at the narrow edge of the tube (see Fig. 4). The development of a WIKA tube design with a more rounded profile was found to reduce these stresses at the narrow edge (see Fig. 5). The key to this development was Fig. 5 ounder profile of the WIKA tube design reduces the high stresses at. to maintain the rounder profile and still retain sufficient deflection at the end of the tube to drive the gauge movement. C - Damping the tube movement by liquid filling the case Liquid filling the pressure gauge case reduces the effects of mechanical vibration on the Bourdon tube by reducing the amplitude of its vibration. This is covered in greater detail later in the report. 3

2.2a Stresses at the free end of a Bourdon tube The simple and cost effective way of closing the end of a Bourdon tube is to pinch the end and seal it with a weld. FEM stress analysis of this technique shows that this creates an area of high stress at the end of the tube (see Fig. 6). Under pressure cycling tests this method of closing the tube lead to a failure after about 300,000 cycles. Fig. 6 If a tube is closed by compressing the end, a high area of stress is produced. Typically after 300,000 pressure cycles, failure will occur at the tip. 2.2b Techniques to strengthen the free end of a Bourdon tube A - Tube end closing technique By keeping the tube profile unaltered at the end and closing the tube with a weld cap of the same material, the stress was significantly reduced (see Fig. 7). The same pressure cycling tests were repeated using this technique and after 3 million cycles the tube had not failed. B - The WIKA Bourdon tube design When developing the new Bourdon tube design (as covered in section 2.lb), it was found that if both ends of the new tube design were finished with a full circular profile, rather than the elliptical profile of the normal tube, any stress that still remained at the end of the tube was reduced even further (see Fig. 7). Although this made the tubes more costly to manufacture, it was decided that the benefits were significant enough to justify recommending the change. Fig. 7 Closing the tube by end capping on a circular profile removes the high area of stress at the end. This tube design will survive over 3 million pressure cycles without failure. 4

2.3 Calibration stability and safety of WIKA Bourdon tubes WIKA Bourdon Tube - 200 PSI Internal % of Endurance Pressure ange 190 PSI 95% 200,000 pressure cycles before calibration error occurs (0 PSI to 190 PSI) 260 PSI 130% 2 hours continuous overload before calibration error occurs 1,800 PSI 900% Minimum burst pressure of the Bourdon tube 8,600 PSI 4,300% Typical burst pressure of the Bourdon tube Table 2 - The special techniques used to produce the WIKA Bourdon tube lead to a much higher degree of calibration, stability and safety. With the development of the WIKA tube design and the techniques to strengthen this tube, the calibration stability, endurance and safety of the tube have been considerably improved (see Fig. 8) The stronger more consistent tube allows higher standards to be set for its performance. Fig. 8 illustrates the minimum burst pressures for the stronger tube, showing a large safety margin compared to the normal full scale pressure. The actual burst pressures are usually much higher as can be seen from Table 2. PESSUE - PSI MINIMUM BUST PESSUE BOUDON TUBE PESSUE ANGE BOUDON TUBE PESSUE ANGE - PSI Fig. 8 The improved design produces a stronger and safer tube with high burst pressures throughout the range. 5

2.4 esonant frequency of a Bourdon tube Using an oscillation generator, the resonant frequency of different Bourdon tube pressure ranges was determined (see Fig. 9). These tests were conducted without the movement attached to the tube end. For example, the resonant frequenqy for a 60 PSI Bourdon tube is about 100Hz and for a 400 PSI tube just under 200Hz. Frequency Hz Bourdon Tube ange - PSI Deflection mm s Fig. 10 Bourdon Tube ange - PSI The red diamonds and green squares show the deflection at the end of the Bourdon tube at their resonant frequencies. The blue triangles are the design deflections for the Bourdon tube. Fig. 9 The red diamonds show the resonant frequency of different Bourdon Tubes. The red diamonds in figure 10 show the deflection at the end of the tube caused by the resonant frequency oscillating the tube. The blue triangles show the normal design deflection allowed for the tube. Clearly even at quite low accelerations of only 2.5g the tube will quickly fail. For higher pressures the greater tube strength and more rounded geometry reduces the deflection at the end of the tube. As can be seen from the graph at 250 PSI and 2.5g acceleration (red diamond) the deflection falls within the design limits of the tube. The green squares show that by increasing the acceleration to 5g the deflection of the tube once again moves outside its design limits. 2.5 Bourdon Tube endurance at resonant frequency Fig. 11 shows the number of oscillation cycles various Bourdon tube pressure ranges could withstand, before failure, when subjected to their resonant frequency. Failure was defined as when a crack appeared in the tube and the internal pressure dropped. It can clearly be seen that for the tubes which are oscillating well outside their design limits, the failure occurs quickly. But as the oscillation deflections approach the normal design deflection, the life of the tube increases. This shows the importance of preventing the Bourdon tube from oscillating outside its design limits, if the life and safety of the Bourdon tube is to be increased. Tests were performed on four different 250 PSI tube designs to determine which had the best life under its resonant frequency (see Table 3). Oscillation Cycles to Failure x 10,000 Fig. 11 Bourdon Tube ange - PSI The squares and diamonds show the number of oscillations before failure of the Bourdon tube. When the tube deflection is whithin its design limits the life of the tube increases, as at 2.5g on a 250 PSI tube. 6

Design Bourdon Tube Acceleration Number of esonant Variant Assembly (g) Cycles Before upture Design 1 Traditional flat tube H/W = 0.30 2.5g 30,000 Design 2 Traditional flat tube inserted 7 mm into the connection H/W = 0.30 2.5g 30,000 Design 3 WIKA tube design H/W = 0.52 2.5g 100,000 Design 4 WIKA tube design, glycerine filled case - 86.5% concentration 2.5g 10 million - no rupture Design 4 WIKA tube design, glycerine filled case - 86.5% concentration 10g 10 million - no rupture Table 3 The number of cycles before rupture of the different Bourdon tube assemblies subjected to their resonant frequency. The WIKA tube design is 3 times better than a flat tube, but liquid filling adds considerable protection against rupture. Design 1 was a standard Bourdon tube profile, cold worked and annealed, using the cap end tube closing technique. Design 2 was the same as Design 1, but with the tube inserted into the connection by 7 mm. Both these designs had the same endurance to their resonant frequency. This shows that inserting the tube further into the connection does not increase the endurance of the tube on these welded assemblies. Design 3 shows that the WIKA tube design with a rounder profile and circular ends increases the endurance of the tube by a factor of three. Design 4 clearly shows the effect of liquid filling the case of a pressure gauge. The 86.5% glycerine/13.5% water filling prevented the tube from oscillating outside its design limits at its resonant frequency, thus significantly increasing the life of the tube. After 10 million cycles no failure had occurred at 2.5g. The load was then increased to 10g, but still no failure had occurred after 10 million cycles. 3. Examination of different instrument variations The tests examined so far have been with just the gauge connection and Bourdon tube. The following series of tests were conducted to determine the effects of vibration on complete instruments with connection, tube, movement, and pointer all linked. A testing standard developed by one of the leading petrochemical companies for testing Bourdon tube gauges was used as follows: Acceleration : 10g Frequency range : 100-500Hz Frequency change : 1 octave/minute Duration of test : 200 hours 3.1 Instruments with damped movements Damped movements have an extension to the pinion or quadrant running into a small pot filled with a highly viscous liquid which provides the desired damping forces. The pointer on these designs was found to be free of oscillations and also the oscillations of the tube were kept within the design limits. At first this would seem to be a good solution, but the tests showed the opposite. Because the pinion is damped, all of the damping forces are transmitted to the toothing, quadrant and connecting link to the tube end piece. These parts are thus subjected to strong forces and wear quickly. The movement with pinion damping was completely inoperable after 100 hours. The toothing was severely damaged and the bearing necks were ground down, whereupon the quadrant came loose and wedged itself in the movement. The tube which was no longer damped by the movement was then free to oscillate, with the increased risk of a resonant rupture of the tube. The movement with pinion and quadrant damping also suffered similar damage. The two damped components and the PTFE toothing were relatively undamaged, but the problem was shifted further along the movement to the connecting link which became severely worn. After 200 hours the pointer had a zero error of 30% (pointer reading 50 PSI with no pressure on the gauge). This excessive wear on the link is caused by the Bourdon tube trying to oscillate against the damped movement. The relatively rigid linking of the free end of the tube can also lead to harmonics of the tubes natural frequency causing it to oscillate with a full wave or even a double wave form. This puts further additional stresses on the tube. 7

Instrument Type Zero Error After 50 Hrs. 200 Hrs. Instrument Condition After 200 Hours Damped movement - Single pinion, 2.3% 66% Inoperable. Toothing & bearing necks Viscous pot severely damaged. Quadrant broke loose. Tube free to resonate. Damped movement - Double pinion & quadrant, 3.9% 34% Inoperable. Severely worn connecting link. Viscous pots, PTFE toothing Standard gauge - No damping 2.3% 3.0% Still operable. But out of calibration. Glycerine filled case - Standard undamped 0.6% 0.8% Fully operational. Still within calibration limits. movement. 86.5% glycerine / 13.5% water No wear to the movement. Table 4 Four complete gauges subjected to vibration tests. Damping pots put excessive load onto the movement, leading to early failure. Liquid filling is the best solution. 3.2 Undamped pressure gauges With an unfilled and undamped movement the vibration caused the pointer to oscillate significantly at practically all frequencies. Although the gauge had no significant internal damping forces, it was found to be unexpectedly durable in the vibration fatigue test. The only fault found after the 200 hours was a zero deviation of 3%. All of the components were operational. The main concern from these tests was that the tube in the undamped instrument could oscillate with a high amplitude leading to the increased danger of a resonant rupture. 3.3 Liquid-filled instruments In the liquid-filled gauge the oscillations of the Bourdon tube, the movement and the pointer were strongly damped. The damping of the Bourdon tube (as shown earlier in the report) reduced the amplitude of the tube and significantly reduced the loading on the movement. The damage to the instruments as outlined in sections 3.1 and 3.2 did not occur. The liquid filling also acted as a lubricant and coolant to all the moving parts further reducing their wear. The gauge was still fully operational after the test and the reading of the gauge was still within its published accuracy of 1.0%. 4. ecommendations A) A liquid filled case is the best solution to counter the effects of vibration. Liquid filling the case of a gauge dampens the oscillations of the tube at the pressure element itself, preventing the tube from oscillating outside its design limits. Furthermore, the filling acts as a lubricant and coolant to all the moving parts, extending their life (Tables 3 and 4). B) WIKA does not recommend the use of damped movements. The movement provides a good calming effect to the pointer but accelerates the wear on the movement. With severe vibration this can eventually lead to a resonance rupture of the Bourdon tube (Tables 3 and 4). C) The cold working and annealing of the Bourdon tube combined with the WIKA tube design and special tube end closing techniques, increases the calibration stability, endurance and safety of the Bourdon tube (Table 2, 3 & Fig. 8). D) Liquid filling improves the safety of a Bourdon tube by preventing a resonant rupture from occurring should the movement fail (Table 3). E) WIKA recommends the removal of the zero stop on the gauge dial since this disguises calibration errors. The new tube design is strong enough not to need this. Under and over range stops are already on the movement. F) Inserting the tube by up to 7 mm into the connection of the gauge before welding does not increase the strength of the tube. WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 Ph:1-888-WIKA-USA Fax: 770-338-5118 website: wika.com e-mail: info@wika.com 8 5. Implementation Most of these recommendations have already been incorporated into the WIKA standard product design. As well as the research covered in this paper there is considerable effort being put into other areas such as diaphragm gauges, chemical seal systems, longer lasting movements, etc. The benefits of this work have already been seen with gauges that stay in calibration longer, and with a greater degree of safety for their users. Authors: Herman Julien, Klaus Schäfer - July 15, 1993 Copyright 1993, All ights eserved