INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

Similar documents
Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

VACUUM REGULATORS CONTENTS

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

CSA Sample Draw Aspirator Adapter Operator s Manual

256 Pneumatic Pressure Indicator

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

Operation Manual. Diaphragm Sensed Gas Pressure Regulators

BT 800-OX Operating Manual

MANUAL BE SERIES Test Benches

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Rejuvenation Instructions

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Powers Controls RC 195 Single Input Receiver-Controller

Powers Controls RC 195 Multiple Input Receiver- Controller

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Installation Instructions and valve Maintenance Rev. 0 of 21/09/15

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MANUAL KPS Pressure Control Valve

DPC-30 DPC-100. Reference Manual

INSTALLATION GUIDE Gas Rangetops

STAP. Differential pressure controller DN 15-50

36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Operator: Save these instructions for future use!

ACCESSORY KIT INSTALLATION INSTRUCTIONS

36E DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

Powers Controls D Day-Night Thermostats

MyControl Bathing System Signature Series

Instruction Manual Contact Pressure Vacuum Gauge

High-performance submersible pressure transmitter For level measurement Model LH-10

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

PRESSURE COMPARATOR 10000PSI 700bar. Operator Instructions CRYSTAL. engineering corporation

Pneumatic proportional controller Types M Types FM Function tested for use of the float in Ex-zone 0

1805 Series Relief Valves

SPECIFICATIONS ATTENTION

INSTALLATION & MAINTENANCE INSTRUCTION

SUPER PRECISION AIR RELAYS

Hydraulic hand test pump Models CPP700-H and CPP1000-H

SIHI LPH-X - Liquid Ring Compressors

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

PORTABLE 5 GALLON AIR TANK ASSEMBLY & OPERATING INSTRUCTIONS

Training. Testor Training Manual

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

36E03 and 36E38 DSI and HSI Step Opening Combination Gas Valve INSTALLATION INSTRUCTIONS

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

Dialflow Regulator. Instructions for Use

The Discussion of this exercise covers the following points:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Hydraulic comparison test pump Models CPP1000-X, CPP1600-X

MoveRoll Conveyor Operating and Maintenance Manual

36E DSI and HSI Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

STAD-C. Balancing valves DN 15-50

SILENTAIRE TECHNOLOGY

PNEUMATIC PRESSURE CONTROLLERS

Operating Instructions Model and Hydrostatic Test Pump

Operating Instructions Part No

Techcon Systems TS1258 Pressure Pot

Cylinder Tilt Saddle Instruction Manual

Maximum 0.85 MPa pressure setting Long-life, high flow perfect for balancer applications

Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL

STAD. Balancing valves DN (MD60, MD61, MD62)

SILENTAIRE TECHNOLOGY

Pressure Dump Valve Service Kit for Series 2300 Units

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

Oxygen Dialflow Meter. Instructions for Use

WARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.

RK LEL Sample Draw Aspirator Adapter Operator s Manual

STAD-C. Balancing valve ENGINEERING ADVANTAGE

Types 749B and R130 Changeover Manifolds

SUPER PRECISION AIR RELAYS

RC 195 Receiver-Controller

64 Series Pressure Reducing Regulators

Freedom8 ShoeBox Compressor Manual

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2"-12" BERMAD. Application Engineering

Additel 761 Automated Pressure Calibrators

User Instruction Manual

Additel 761 Automated Pressure Calibrators Selection Guide

Pressure Relief Valve Instruction Manual

SIKA Ba_ODWT15_en 05/2018. Operating manual for oil deadweight tester 0.015% accuracy ODWT15

Installation, Operation, and Maintenance Manual

Additel 761 Automated Pressure Calibrators

159 mm 154 mm. Subject to changes 10/2012

Techcon Systems TS1254 Pressure Pot

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

UsER manual for Watersens ph -REDOX

OPERATION MANUAL NTF-15

ALU-CD 12, ALU-CD 25, ALU-CD 35, ALU-CD

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Inverse Prop. Pressure-Relief Cart., Size 2 4

OPERATION MANUAL NTF-60 Plus

Additel 761 Automated Pressure Calibrators Selection Guide

628 Differential Pressure Decay Leak Tester

Instruction Manual Updated 7/26/2011 Ver. 2.2

Transcription:

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

Dear user, We have made every effort to ensure the accuracy of the contents of this manual. Should any errors be detected, we would greatly appreciate to receive suggestions to improve the quality of the contents of this manual. For more detailed technical data about the Instruction manual for Beamex PGL Low Pressure Calibration Pump, please contact the manufacturer. Copyright 2013-2017 BEAMEX OY AB Ristisuonraitti 10 FIN-68600 Pietarsaari FINLAND Tel +358-10 - 5505000 Fax +358-10 - 5505404 E-mail: sales@beamex.com service@beamex.com Internet: http://www.beamex.com 8804200 / PGLuEng / Version 2

CONTENTS DESCRIPTION... 1 Standard Accessories... 1 Optional Accessories... 1 Connections and Parts... 2 Connection Diagram... 3 OPERATION... 3 Positive Pressure Calibration... 3 Negative Pressure (Vacuum) Calibration... 4 Keep in Mind... 5 TROUBLESHOOTING/MAINTENANCE... 6 Seal Replacement... 6 Leak Tests... 7 Leak Test for Positive Pressure... 7 Leak Test for Negative Pressure (Vacuum)... 7 SPECIFICATIONS... 8 WARNINGS... 9

DESCRIPTION The PGL Low Pressure Calibration Pump is designed to manually generate up to 400 mbar (6 psi) for quick and accurate calibration of pressure gauges, transducers and other pressure measurement instruments. Standard Accessories The standard accessories are as follows: Pressure connectors: o G 1/8" male with 60 internal cone to G1/8 including an o-ring (10.1 mm x 1.6 mm NBR 70). Pre-installed in PGL. o G 6/4" nipple to G 1/8 including an o-ring (10.1 mm x 1.6 mm NBR 70). Delivered as a replacement for the pre-installed connector A set of seals (o-rings), type: 10.1 mm x 1.6 mm NBR 70 (Nitrile Butadiene Rubber, Shore A70, black) Guarantee leaflet This Instruction Manual Optional Accessories The optional accessories are as follows: ITEM ORDER CODE Carrying case 8003310 Service Seal kit 8003085 Bag of additional connectors: o G 1/8" male with 60 internal cone to G1/8 incl. an o-ring (10.1 mm x 1.6 mm NBR 70). o G 6/4" nipple to G 1/8 incl. an o-ring (10.1 mm x 1.6 mm NBR 70). 8003090 Tube of grease 0005180 1

Connections and Parts Fine Adjust handwheel (clockwise to increase pressure). Cylinder Coarse Adjust handwheel (clockwise to increase pressure). Pressure connector. Default: G 1/8" male with 60 internal cone. Release valve (clockwise to close/ counterclockwise to open). 2

Connection Diagram Atmosphere Fine Adjust Coarse Adjust Pressure connector Release Valve Calibrator (reference) Instrument to be calibrated (DUT) OPERATION Positive Pressure Calibration 1. Open Release valve. Rotate the the Fine Adjust handwheel to mid position. Rotate the Coarse Adjust handwheel counterclockwise to maximum 2. Connect the calibrator (pressure module, reference) and the instrument to be calibrated (Device Under Test, DUT) to Pressure connector using a Beamex Pressure Measurement Hose or similar. 3. Close Release valve and check that the measurement system is not leaking. See Leak Tests on page 7. 4. Increase the pressure close to the next calibration point by rotating the Coarse Adjust handwheel clockwise. 5. Advance to calibration point by rotating the Fine Adjust handwheel. 6. Continue to next calibration point by repeating stages 4 and 5 until you reach the calibration point with the highest pressure. 7. Decrease the pressure close to the next calibration point by rotating the Coarse Adjust handwheel counterclockwise. 8. Advance to calibration point by rotating the Fine Adjust handwheel. 9. Continue to next calibration point by repeating stages 7 and 8 until all calibration points are done. 10. Start another calibration run from stage 4 onwards or end calibration by opening Release valve. 3

Negative Pressure (Vacuum) Calibration 11. Open Release valve. Rotate the the Fine Adjust handwheel to mid position. Rotate the Coarse Adjust handwheel clockwise to maximum 12. Connect the calibrator (pressure module, reference) and the instrument to be calibrated (Device Under Test, DUT) to Pressure connector using a Beamex Pressure Measurement Hose or similar. 13. Close Release valve and check that the measurement system is not leaking. See Leak Tests on page 7. 14. Decrease the pressure close to the next calibration point by rotating the Coarse Adjust handwheel counterclockwise. 15. Advance to calibration point by rotating the Fine Adjust handwheel. 16. Continue to next calibration point by repeating stages 14 and 15 until you reach the calibration point with the deppest vacuum. 17. Increase the pressure close to the next calibration point by rotating the Coarse Adjust handwheel clockwise. 18. Advance to calibration point by rotating the Fine Adjust handwheel. 19. Continue to next calibration point by repeating stages 17 and 18 until all calibration points are done. 20. Start another calibration run from stage 14 onwards or end calibration by opening Release valve. 4

Keep in Mind The operating principle of the PGL pump is based on an adjustable volume. When changing the volume in a closed pressure system, it causes the pressure to change. The PGL has certain range in the volume adjustment and the effect of that to the pressure change in the system depends directly on the total volume of the system: For a system with very large volume, the PLG can only make small pressure adjustment, and you may not reach the max pressure specified for PGL. For a system with very small volume the PGL can adjust the pressure significantly more, even over the max safety pressure. Please take this into account when using PGL and don t exceed the safety pressure 4bar / 60psi in any circumstances. Immediately after a pressure change, the pressure may change slightly due to the stretching/shrinking of the pressure measurement hose. Thermodynamic effects may also cause pressure variation. In that case, adjust the pressure back to required value using the Fine Adjust handwheel. Do not exceed the max. operating pressure of the pump and the hose. Also observe the pressure limit set by environmental conditions. Ensure that there always is a reliable pressure indicator connected to the pump when the pump is used. If there is a strong counterforce while operating the pump, but no pressure change is indicated, stop using the pump and locate the fault. Always keep a reliable indicator connected to the measurement system. If the pump fails to indicate a pressure change, check to assure that the connections between the pump, the hose end and the attached instrument(s) are tight and retry pumping. Also check to assure that possible unused connectors are properly plugged. If a pressure increase still cannot be obtained, it s possible that one or more of the seals in the Pump is leaking and needs to be replaced. Do not use the pump if the functionality of the pump is not normal. Locate the fault before you continue using the pump. A full list of warnings is at the end of this manual. 5

TROUBLESHOOTING/MAINTENANCE Hard to Adjust Using Coarse Adjust Valve CAUSES SOLUTIONS Release valve is not closed. Close the release valve. The Coarse Adjust Valve's o-ring is broken or loose. Hard to Fine Adjust Using Fine Adjust Valve CAUSES The connections are not tightened. The Fine Adjust Valve's 0-ring is aged or frayed. The thread surface is not smooth. The connector types do not match. Replace the o-ring. SOLUTIONS Tighten all connections. Replace the o-ring. Apply Teflon tape. Use proper adapter. Hard to Turn the Valves CAUSES Release valve closed too tight after the previous calibration. The new pump does not run smooth. Lack of lubrication in threads. SOLUTIONS Do not close release valve too tightly. The new pump needs time to run in. Lubricate the thread. Seal Replacement Depending on the frequency of use, the seals will eventually need replacing. The replacement seals are part of an optional Service kit and the kit includes instructions for replacing/fitting the seals. 6

Leak Tests If you want to test the pump and the connected measurement system, do as follows: If you do not already have a calibration measurement system connected to the pump, connect an accurate enough pressure measurement instrument, e.g. a Beamex 400mC pressure module to the pump. You'll also need a clock to measure time. Leak Test for Positive Pressure Increase pressure using constant change rate up to pump's full scale positive pressure (400 mbar / 6 psi). Start your clock. After five minutes, record the pressure indication, p5. After ten minutes, record the pressure indication, p10. Calculate the leak, Lp, using this equation: p5 p10 Lp p5 If Lp is less than 0.01 (1 %), the system is leak free. Leak Test for Negative Pressure (Vacuum) Decrease pressure using constant change rate down to pump's full scale negative pressure (-400 mbar / 40 kpa / -6.0 psi). Start your clock. After five minutes, record the pressure indication, pv5. After ten minutes, record the pressure indication, pv10. Calculate the leak, Lv, using this equation: Lv = pv5 - pv10 [kpa]. If Lv is less than 1 kpa (approx. 0.14 psi), the system is leak free. 7

SPECIFICATIONS Weight 1.38 kg 3.04 lb Dimensions Height 133 mm approx. 5.2" Width 251 mm approx. 9.9" Depth 117 mm approx. 4.6" Pressure range Safety Pressure Pressure media Temperature 0... 50 C Humidity < 95 %RH Adjusting sensitivity Output connectors -400 mbar to +400 mbar / -6 psi to +6 psi < 4 bar / < 60 psi Clean air 1 Pa / 0.01 mbar / 0.004 inh2o / 15 10-5 psi G 1/8" male with 60 internal cone, for Beamex pressure measurement hose, pre-installed. G 6/4" nipple delivered as an optional connector to replace the one meant for Beamex pressure measurement hose 8

WARNINGS Read the instruction manual carefully prior to setting up and using the pressure pump. Only personnel with good experience and knowledge of pressure media, pressure instruments and connections are allowed to work with the pressure pump. Incorrect use may result in damage to the pump, the instrument connected to the pump and/or personal injury. Use eye shields. The use of a relief valve for over-pressure protection is recommended. If a relief valve is used, remember to occasionally check its functionality. Do not connect the pump to an external pressure source. Vent external systems before connecting to the pump. Ensure that all connections are made correctly and that the hoses and connectors are free of impurities and undamaged. Do not use faulty hoses or connectors. Use only measuring hoses provided by Beamex or other reliable suppliers. Observe the effect of the operational conditions to the maximum pressure allowed in the hose. The environmental conditions (ambient pressure and temperature) may restrict the allowable maximum pressure to a lower level than the pump and the hose enable. Always depressurize PGL when it is left on its own. Avoid possible damage when suddenly apply pressure to low pressure and small chamber gauges. Do not exceed the safety pressure limit (4 bar / < 60 psi). Compressed air may cause problem in explosive or corruptive environment. Do not over tighten connectors to avoid any damage. Store the pump in the dry and non-corruptive environment. If the pump is accidentally dropped, it may be damaged. Do not use the pump before it is inspected at Beamex s service. Do not use PGL in any other way than as described in this manual. Any safety problems or damages caused by incorrect operation, are beyond Beamex's responsibility.

BEAMEX OY AB Ristisuonraitti 10 FIN-68600 PIETARSAARI FINLAND Phone +358-10 5505000 Fax +358-10 5505404 E-mail sales@beamex.com service@beamex.com Internet http://www.beamex.com Beamex Limited Newtown Grange Farm Business Park Desford Road NEWTOWN UNTHANK Leicestershire LE9 9FL United Kingdom Phone 01455 821 920 Fax 01455 821 923 E-mail beamex.ltd@beamex.com Beamex Inc 2152 NW Parkway Suite A Marietta, GA 30067 U.S.A. Phone 800 888-9892, +1-770-951-1927 Fax +1-770-951-1928 E-mail beamex.inc@beamex.com Representative: