Trust Guideline for Neonatal Volume Guarantee Ventilation (VGV)

Similar documents
Neonatal tidal volume targeted ventilation

Volume Targeted Ventilation on the Neonatal Unit

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Introduction to Conventional Ventilation

Mechanical Ventilation

Mechanical Ventilation

Managing Patient-Ventilator Interaction in Pediatrics

excellence in care Procedure Management of patients with difficult oxygenation. For Review Aug 2015

Wales Neonatal Network Guideline

Volume Guarantee New Approaches in Volume Controlled Ventilation for Neonates

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Selecting the Ventilator and the Mode. Chapter 6

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

This is a pre-copyedited, author-produced PDF of an article accepted for publication in Neonatal Network following peer review. The version of record

Physiological based management of hypoxaemic respiratory failure

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

Difficult Oxygenation Distribution: Sydney X Illawarra X Orange X

Wales Neonatal Network Guideline

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

SLE5000 Infant Ventilator with HFO

Mechanical Ventilation. Flow-Triggering. Flow-Triggering. Advanced Concepts. Advanced Concepts in Mechanical Ventilation

VENTILATION STRATEGIES FOR THE CRITICALLY UNWELL

Automatic Transport Ventilator

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

N eonatal respiratory distress syndrome and bronchopulmonary

VENTILATORS PURPOSE OBJECTIVES

Safe and Intuitive Neonatal Ventilation

Initiation and Management of Airway Pressure Release Ventilation (APRV)

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Using Common Ventilator Graphics to Provide Optimal Ventilation

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

The Basics of Ventilator Management. Overview. How we breath 3/23/2019

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Neonatal Ventilators. Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000

Operation of Oxylog 3000 ventilator

? PSV. Lung Protective Ventilation Strategies

Flight Medical presents the F60

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

The aim of this guideline is to describe the indications and procedure for using high flow nasal prong oxygen

Oxylog 3000 plus Emergency & Transport Ventilation

Health Professional Info

Technical Data and Specifications


Newer forms of conventional ventilation for preterm newborns

NOTE: If not used, provider must document reason(s) for deferring mechanical ventilation in a patient with an advanced airway

4/2/2017. Sophisticated Modes of Mechanical Ventilation - When and How to Use Them. Case Study 1. Case Study 1. ph 7.17 PCO 2 55 PO 2 62 HCO 3

QUICK GUIDE: VOLUME GUARANTEE VENTILATION Prematurely Born Infants < Weeks Gestation

Panther 5 Acute Care Ventilator

AUTOVENT 4000 VENTILATOR

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Principles of Mechanical Ventilation: A Graphics-Based Approach

Presentation Overview. Monitoring Strategies for the Mechanically Ventilated Patient. Early Monitoring Strategies. Early Attempts To Monitor WOB

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Invasive mechanical ventilation:

Neonatal Assisted Ventilation. Haresh Modi, M.D. Aspirus Wausau Hospital, Wausau, WI.

ONLINE DATA SUPPLEMENT. First 24 hours: All patients with ARDS criteria were ventilated during 24 hours with low V T (6-8 ml/kg

Guide to Understand Mechanical Ventilation Waveforms

HELPING PROTECT YOUR MOST VULNERABLE NEWBORNS. Puritan Bennett 980 Neonatal Ventilator System

Notes on BIPAP/CPAP. M.Berry Emergency physician St Vincent s Hospital, Sydney

QUICK REFERENCE GUIDE

New Frontiers in Anesthesia Ventilation. Brent Dunworth, MSN, CRNA. Anesthesia Ventilation. New Frontiers in. The amount of gas delivered can be

How does HFOV work? John F Mills MBBS, FRACP, M Med Sc, PhD Neonatologist Royal Children s Hospital. Synopsis

Update to RS-232 commands. Changing patient from NIV to INVASIVE Vent Type. SNDF command

Operating & troubleshooting a self inflating bag. Victorian Newborn Resuscitation Project Updated February 2012

The Puritan Bennett 980 Neonatal Ventilator System. Helping to Protect Our Most Vulnerable NEWBORNS

RESUSCITATION. If baby very floppy and heart rate slow, assist breathing immediately. If baby not breathing adequately by 90 sec, assist breathing

HIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION

Infant Ventilators. Infant Ventilator with Touch-Screen Operation SLE4000

NAVA Neurally Adjusted Ventilatory Assist In Neonates

? PSV. Lung Protective Ventilation Strategies

Complex Care Hub Manual: Pressure Ventilation on the LTV 1200

Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

Selecting and Connecting Breathing Systems

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

Pressure Controlled Modes of Mechanical Ventilation

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

Understanding and comparing modes of ventilation

3100A Competency Exam

Ventilating the Sick Lung Mike Dougherty RRT-NPS

D Beyond essentials DRÄGER SAVINA 300 SELECT

Operating Instructions for Microprocessor Controlled Ventilators

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Inspiration 7i Ventilator

Oxylog 2000 plus. Step up your performance. Emergency Care Perioperative Care Critical Care Perinatal Care Home Care

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

Volume vs Pressure during Neonatal Ventilation

Inspire rpap REVOLUTION FROM THE FIRST BREATH

INTENSIVE CARE VENTILATORS

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

Objectives. Respiratory Failure : Challenging Cases in Mechanical Ventilation. EM Knows Respiratory Failure!

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Infant Ventilators. Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000

Volume Diffusion Respiration (VDR)

Virginia Beach EMS. Oxylator EMX. Debra H. Brennaman, RN, MPA, NREMT-P

Disclosures. The Pediatric Challenge. Topics for Discussion. Traditional Anesthesia Machine. Tidal Volume = mls/kg 2/13/14

Transcription:

A Clinical Guideline For Use in: By: For: Division responsible for document: Key words: Name of document author: Job title of document author: Name of document author s Line Manager: Job title of author s Line Manager: Supported by: Assessed and approved by the: Neonatal Intensive Care Unit Doctors, Nurses, and Advanced Neonatal Nurse Practitioners (ANNPs) Neonates and infants Women and Children Division Guarantee, Ventilation, VGV, Volume, Vent Dr R Roy Consultant Neonatologist Dr David Booth Date of approval: 15/04/2015 Clinical Director Paediatrics Dr Mark Dyke, Consultant Neonatologist Dr David Booth, Consultant Neonatologist Dr Paul Clarke, Consultant Neonatologist Dr Priya Muthukumar, Consultant Neonatologist Dr Helen O Reilly, Consultant Neonatologist Dr Florence Walston, Consultant Neonatologist Clinical Guidelines Assessment Panel (CGAP) If approved by committee or Governance Lead Chair s Action; tick here Ratified by or reported as approved to (if applicable): Clinical Standards Group and Effectiveness Sub-board To be reviewed before: This document remains current after this date but will be under 15/04/2018 review To be reviewed by: Dr Rahul Roy Reference and / or Trust Docs ID No: NICU VGV Version No: 2 Description of changes: Compliance links: Reviewed, no clinical changes None Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 1 of 8

Quick reference guideline (Title of document needed on every page) Initial Settings < 1kg Select AC-PC + add VG TV 5 ml/kg P max 25cm H 2 O (PIP limit) PEEP 5 cm H 2 O Inspiratory time 0.3 seconds Back up respiratory rate (BURR) 40/minutes Initial Settings > 1kg Select AC-PC + add VG TV 4.0 ml/kg P max 30cm H 2 O (PIP limit) PEEP 5 cm H 2 O Inspiratory time 0.35 seconds Back up respiratory rate (BURR) 40/minutes Ongoing Management on VGV Adjust set TV in increments of 0.5m//kg or BURR 5-10/min based on P CO2 and work of breathing Consider checking blood gas within 30min - 1 hour if settings are changed P max limit must be adjusted periodically to keep the limit close to the working delivered PIP pressure Delivered PIP > 26 cmh 2 O consider HFOV Adjustments when PC 02 high Is the set TV and BURR rate adequate? Is the TV low alarm going off? o Exclude airway obstruction/leak o o Is P max limit set adequate? Enough Ti to reach PIP? Does the baby need a chest X-ray? Adjustments when PC O2 low High RR o? agitated/pain -sedation o?autotrigering - Trigger sensitivity & circuit free of water Normal/low RR o Reduce set TV Weaning and Extubation Allow delivered PIP to self-wean Avoid over sedation during the weaning phase Once the set TV is at the low end of the normal range (usually 4ml/kg), P CO2 is allowed to rise to help in self-weaning by improving respiratory drive If the TV is set too high, resulting in a high ph > 7.40, the baby will not have a good respiratory drive and will not self-wean Over distension with normal P CO2 may be best addressed by reducing PEEP If significant oxygen requirement persists FiO2 > 40%, PEEP may need to be increased to keep MAP from falling as delivered PIP is auto weaned, to prevent atelectasis, worsening lung compliance and failure of extubation Reduce BURR to 20 30/min to ensure good respiratory drive Consider Extubation if mean airway pressure is consistently <8 to 10 cm H 2 O or delivered PIP < 12 cm H 2 O in babies < 1kg or delivered PIP < 15 cm H 2 O in babies > 1kg, with set TV 4.0 to 4.5 ml/kg, FiO2 less than 35%, blood gases are satisfactory with good sustained respiratory effort. Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 2 of 8

Glossary AC BPD BURR CLD ETT PIP PLV RDS SIMV SIPPV TV VG VGV Assist-Control Bronchopulmonary Dysplasia Back-Up Respiratory Rate Chronic Lung Disease Endotracheal Tube Peak Inspiratory Pressure Pressure-limited ventilation Respiratory Distress Syndrome Synchronised Intermittent Mandatory Ventilation Synchronised Intermittent Positive Pressure Ventilation Tidal Volume Volume Guarantee Volume Guarantee Ventilation Objective To guide the Neonatal team in using Volume Guarantee Ventilation (VGV) in preterm and term infants Rationale Mechanical ventilation is required to manage neonates with severe respiratory failure. Pressure-limited ventilation (PVL), delivers a fixed Peak Inspiratory Pressure (PIP). This has one major disadvantages of delivering variable tidal volumes as the lung compliance and resistance changes. VGV allows effective control of delivered tidal volumes while offering ventilator support to low birth weight preterm infants. Ventilation allowing large tidal volumes results in volutrauma which is important in the pathogenesis of Bronchopulmonary Dysplasia (BPD). VGV facilitates for the appropriate use of tidal volumes with the aim of reducing lung damage. VGV is possible in small preterm babies due to the development of sensitive and accurate flow sensors which allows for the measurement and tracking of gas flow. Improved software development prevents over/undershoot and rapid obtaining of target tidal volumes. The flow sensors assist in avoiding overexpansion (volutrauma) and under expansion (atelectotrauma). In animal studies ventilation at low and high lung volumes has also been shown to cause lung injury especially in surfactant deficient lungs. This injury is believed to be related to the repeated opening and closing of lung units with each mechanical breath (atelectotrauma). VGV is protective against lung injury as a result of the above changes. Recent systematic review and meta-analysis shows that VGV compared to PLV reduce death and BPD Process to be Followed Initiation Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 3 of 8

VGV is initiated using the Drager VN 500. VG can be used in conjunction with Synchronised Intermittent Mandatory Ventilation (SIMV), Synchronised Intermittent Positive Pressure Ventilation (SIPPV)/Assist-Control (AC) or Pressure Support Ventilation. On the Drager VN 500 the preferred ventilator mode is SIPPV/AC as it is associated with more stable expired Tidal volume (TV), better oxygenation and reduced tachypnea when compared with SIMV+VG. It is a time-cycled pressure limited mode which automatically adjusts the delivered PIP to achieve a target volume set by the operator. An upper limit of P max needs to be set; the machine will adjust the pressure delivered to achieve the target TV selected. If the ventilator cannot deliver the TV selected using the P max set the machine will alarm. Breaths that are unsupported by inflation during continuous mandatory ventilation +VG or SIMV +VG are not volume targeted. Setting the ventilator 1. Select AC-PC on the Drager VN 500 2. Add VG + ensure correct weight has been entered into the ventilator at imitation 3. Setting the desired TV Start at 4mL/kg during the acute phase of Respiratory Distress Syndrome (RDS) in infants larger than 1kg. Babies with birth weight less than 1kg require TV of 5 ml/kg because of the additional 1mL dead space of the flow sensor has a larger impact in the smallest infants. Subsequent adjustment to TV may be needed based on PC O2. The usual increment is 0.5mL/kg. Substantially larger tidal volumes may be needed in infants with chronic lung disease because of increased anatomical dead space due to stretching of the trachea and increased physiologic dead space. TV of 6-8mL/kg is often used to achieve good gas exchange in infants with moderate to severe Chronic Lung Disease. For babies with Congenital Diaphragmatic Hernia after birth start with low TV 4mL/kg and high ventilator rates. 4. Setting the Peak Inspiratory Pressure (PIP) limit (P max ) During VGV the set PIP is not the same as the delivered PIP because the VG mode continuously alters the delivered PIP to achieve the set TV. The set P max needs to be high enough to allow fluctuations around the average delivered PIP. Start VG ventilation with a PIP limit (P max ) of 25 cm H 2 O for babies less than 1 kg and 30cm H 2 O for babies more than 1 kg. This enables the ventilator to choose a PIP lower than this to deliver the set TV. Thereafter, the P max can be adjusted to at least 5 to 10 cm H 2 O above the working delivered PIP, allowing the ventilator flexibility to deliver the set TV during variable Endotracheal Tube (ETT) leaks and untriggered inflations. The Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 4 of 8

patient and ventilator should be assessed if the delivered PIP progressively increases, is persistently high (for example >30 cm H 2 O), or if the ventilator frequently alarms 'low tidal volume'. The main causes are: increasing ETT leak, baby splinting its abdominal muscles against ventilator inflations, untriggered inflations, worsening lung mechanics, air leaks, the ETT slipping down the right main bronchus or ETT kinking. 5. Setting the Trigger Sensitivity The flow-sensor trigger threshold should be set at its greatest sensitivity. If sensitivity is decreased, triggering is delayed and reduces synchrony between the baby's inspiration and ventilator inflation, and increases work of breathing. Movement of water within the circuit (rain out) is often misinterpreted by the flow sensor as the onset of a breath. This inappropriately triggers inflations which are out of synchrony with the respiratory rate of the infant s own ventilation. The circuit must be kept free of condensed water. The flow sensor should be positioned with the plug pointing upwards to reduce condensation in the sensor. It is inappropriate to reduce the humidity in the circuit to prevent rain out. The gas must be delivered to the airways at 37 C and 100% humidity to maintain cilia and mucus function and prevent epithelial dehydration. 6. Setting the Inflation Time (T i ) Sufficient T i is required for the pressure to rise to the PIP selected by the ventilator and form a small plateau. If the T i is too short, the PIP will not rise to a level sufficient to deliver the set TV and a 'low tidal volume' alarm occurs. In an AC mode set the Ti 0.3 s for babies with RDS and < 1kg which is similar to their spontaneous T i. Infants with other lung pathologies may need a longer T i. The appropriateness of the T i can be evaluated by observing the ventilator graphics. If T i is set too long, the pressure plateau is held after the inflation flow has stopped and there is no further increase in TV. 7. Setting the Ventilator Back-Up Respiratory Rate (BURR) When using AC modes select a ventilator back-up BURR. BURR delivers untriggered inflations if the infant's breathing rate falls below this rate. If the BURR is high, the baby has less opportunity to trigger inflations before the ventilator delivers untriggered inflations. In a study comparing BURRs of 30, 40 and 50 per minutes in ventilated spontaneously breathing infants, there were most triggered inflations at a BURR of 30 per minutes. It is important that infants trigger as much inflation as they need because their breathing contributes to the TV, and therefore, the delivered PIP required is lower. However, for infants with apnoea, or a poor respiratory drive, a BURR about 50 to 60 per minutes may be needed to maintain minute volume. 8. Setting the Positive End-Expiratory Pressure (PEEP) Adequate PEEP is vital to maintain functional residual capacity, prevent atelectasis and improve oxygenation. Most intubated infants will require a PEEP Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 5 of 8

5 cm H 2 O due to underlying lung disease and the ETT bypassing the larynx. Insufficient PEEP may contribute to heterogeneous ventilation, increasing the risk of regional lung injury from local volutrauma and shear stress. 9. The Initial Settings for Babies with RDS and < 1 kg are: a. TV 4.5 ml/kg b. P max 25cm H 2 O (PIP limit) c. PEEP 5 cm H 2 O d. Inspiratory time 0.3 seconds e. Back up respiratory rate of 40 breaths/min 10. Weaning and Extubation from Volume Guarantee Ventilation (VGV) VG ventilation automatically weans the PIP as an infant's lung function improves. In AC the infant controls the ventilator rate. The only parameters that should be altered during weaning are the FiO 2 and the set TV. Reducing the ventilator rate during AC has no effect on delivered rate unless respiratory effort is poor or the BURR is greater than the infant's breathing rate. Over-sedation must be avoided during the weaning phase. When the set TV is below the infant's spontaneously generated TV, the PIP will be reduced, Therefore the infant will be breathing on an ETT with continuous positive airway pressure, potentially increasing work of breathing and the risk of subsequent extubation failure due to fatigue. If the above scenario is observed for more than short periods, extubation should be considered. Weaning is not recommended TV below 3.5 ml/kg. Consider extubation: If the mean airway pressure is consistently <8 to 10 cm H 2 O or delivered PIP < 12 cm H 2 O in babies < 1kg or delivered PIP < 15 cm H 2 O in babies > 1kg, with set TV 3.5 to 4.5 ml/kg. FiO2 less than 35%, blood gases are satisfactory with a good sustained respiratory effort. 11. Alarm, Endotracheal Tube (ETT) Leak and Troubleshooting VGV The VG option will generate additional alarms due to low tidal volume (TV low ). A 'low tidal volume' or ETT obstructed alarm occurs if there is Significant fall in lung compliance, Decreased spontaneous respiratory effort, ETT slipping into the right bronchus, Air leaks Forced expiration with tightening of abdominal muscles such that gas entry to the lungs during inflation is prevented, often termed as splinting. Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 6 of 8

It is critical to evaluate the cause of repeated alarms and correct any correctable problems. Large leak results in underestimation of delivered TV and triggers the low tidal volume alarm when the ventilator is unable to reach the target TV at the set P max. In general, VGV can be used with ETT leaks up to 50% because the ventilator adjusts the delivered PIP to deliver the set TV. An increasing ETT leak with a concomitant rise in delivered PIP may be misinterpreted as worsening lung disease. The number of these TV low alarms may be reduced by increasing the P max if the alarms sounds repeatedly in the absence of excessive leak, obstruction and adequate Ti and then investigate the cause. In a ventilated infant who is breathing well and has adequate blood gases despite having a large ETT leak, extubation may be appropriate. Otherwise, if an infant has a large ETT leak it is prudent to place a larger ETT rather than turning off VG and leaving TV delivery uncontrolled. Surfactant administration often causes brief increases in airway resistance and ETT obstruction. To maintain the set TV, the P max should be set 10cm H 2O or more above the pre-surfactant delivered PIP to enable the set TV to be delivered. Within 1 hour, the working delivered PIP falls as lung compliance improves. Clinical Audit Standards To ensure that this document is compliant with the above standards, the following monitoring process will be undertaken: 1. All neonates in need of mechanical ventilation should be started on volume guarantee with assist control mode of ventilation The audit results will be sent to the Women and Children s Clinical Governance Meeting for discussion and to review the results and make recommendations for further action. Summary of development and consultation process undertaken before registration and dissemination This guideline was drafted by Dr Rahul Roy, on behalf of the Neonatal Intensive Care Unit. During its development it has been circulated to consultant neonatologist colleagues for comments. It has been presented and discussed at the neonatal unit clinical guidelines meeting (attended by medical, nursing and ANNP staff of the neonatal unit). This version has been endorsed by the Clinical Guideline Assessment Panel. Distribution List Hospital Intranet References Klingenberg C, Wheeler K, Davis PG, Morley CJ. A practical guide to neonatal volume guarantee ventilation. Journal of Perinatology 2011, 31(9), 575 585. Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 7 of 8

Keszler M. Volume-targeted ventilation. Early Human Dev 2006; 82(12): 811-818. Abubakar K, Keszler M. Effects of volume guarantee with assist / control vs synchronised intermittent mandatory ventilation. J Perinatol 2005,; 25(10), :638-642. Wheeler K, Klingenberg C, McCallion N, Morley CJ, Davis PG. Volume-targeted versus pressure limited ventilation in the neonate. Cochrane Database Syst Rev 2010;11:CD003666. Available via Trust Docs Version: 2 Trust Docs ID: 11166 Page 8 of 8