Neonatal ventilators: how do they differ?

Similar documents
Newer forms of conventional ventilation for preterm newborns

Introduction to Conventional Ventilation

Principles of Mechanical Ventilation: A Graphics-Based Approach

Principles of mechanical ventilation. Anton van Kaam, MD, PhD Emma Children s Hospital AMC Amsterdam, The Netherlands

Neonatal tidal volume targeted ventilation

Mechanical Ventilation. Which of the following is true regarding ventilation? Basics of Ventilation

Mechanical Ventilation

Managing Patient-Ventilator Interaction in Pediatrics

VENTILATORS PURPOSE OBJECTIVES

Mechanical Ventilation. Mechanical Ventilation is a Drug!!! is a drug. MV: Indications for use. MV as a Drug: Outline. MV: Indications for use

Elective high frequency oscillatory ventilation versus conventional ventilation for acute pulmonary dysfunction in preterm infants (Review)

Physiological based management of hypoxaemic respiratory failure

Basics of Mechanical Ventilation. Dr Shrikanth Srinivasan MD,DNB,FNB,EDIC Consultant, Critical Care Medicine Medanta, The Medicity

Mechanical ven3la3on. Neonatal Mechanical Ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on. Mechanical ven3la3on 8/25/11. What we need to do"

Selecting the Ventilator and the Mode. Chapter 6

SLE5000 Infant Ventilator with HFO

RESPIRATORY PHYSIOLOGY, PHYSICS AND PATHOLOGY IN RELATION TO ANAESTHESIA AND INTENSIVE CARE

PART EIGHT HIGH FREQUENCY PERCUSSIVE OSCILLATION (HFPOV )

MEDICAL EQUIPMENT IV MECHANICAL VENTILATORS. Prof. Yasser Mostafa Kadah

Guide to Understand Mechanical Ventilation Waveforms


3100A Competency Exam

The Puritan Bennett 980 Neonatal Ventilator System. Helping to Protect Our Most Vulnerable NEWBORNS

UNDERSTANDING NEONATAL WAVEFORM GRAPHICS. Brandon Kuehne, MBA, RRT-NPS, RPFT Director- Neonatal Respiratory Services

Indications for Mechanical Ventilation. Mechanical Ventilation. Indications for Mechanical Ventilation. Modes. Modes: Volume cycled

Flight Medical presents the F60

Ventilators. Dr Simon Walton Consultant Anaesthetist Eastbourne DGH KSS Basic Science Course

Using Common Ventilator Graphics to Provide Optimal Ventilation

Mechanical Ventilation

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

The Crossvent 2i+ 2. Ventilator Concept (brief theory of operation and features)

Trust Guideline for Neonatal Volume Guarantee Ventilation (VGV)

HAMILTON-C2 HAMILTON-C2. The universal ventilation solution

Wales Neonatal Network Guideline

Technical Data and Specifications

SLE4000. Infant Ventilator with touch-screen operation. When the smallest thing matters

Advanced Ventilator Modes. Shekhar T. Venkataraman M.D. Professor Critical Care Medicine and Pediatrics University of Pittsburgh School of Medicine

Initiation and Management of Airway Pressure Release Ventilation (APRV)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

Neonatal Ventilators. Touch-screen Neonatal Ventilator with High Frequency Oscillation SLE5000

How does HFOV work? John F Mills MBBS, FRACP, M Med Sc, PhD Neonatologist Royal Children s Hospital. Synopsis

Blease900 Ventilator PATIENT CENTERED VENTILATION

Completed downloadable Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 5th Edition by Cairo

Automatic Transport Ventilator

fabian HFO: Modular, Powerful, Versatile.

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives


Automatic Transport Ventilators. ICU Quality Ventilation on the Street.

INTELLiVENT -ASV. The world s first Ventilation Autopilot

Safe and Intuitive Neonatal Ventilation

Potential Conflicts of Interest Received research grants from Hamilton, Covidien, Drager, General lel Electric, Newport, and Cardinal Medical Received

evolution 3e Ventilators

5/9/2017. Definition & Description. HFOV Rationale, Physiology, & Applications. Definition & Description. Rationale. High Frequency Ventilation Types

Volume-Targeted Ventilation and Arterial Carbon Dioxide in Neonates

Medical Instruments in the Developing World

Understanding and comparing modes of ventilation

This is a pre-copyedited, author-produced PDF of an article accepted for publication in Neonatal Network following peer review. The version of record

Ventilating the Sick Lung Mike Dougherty RRT-NPS

PERFORMANCE EVALUATION #34 NAME: 7200 Ventilator Set Up DATE: INSTRUCTOR:

PART SEVEN THE HISTORY AND APPLICATION OF HIGH FREQUENCY OSCILLATORY VENTILATION (HFOV)

INTRODUCTION TO BI-VENT (APRV) INTRODUCTION TO BI-VENT (APRV) PROGRAM OBJECTIVES

Stephanie. Pediatrics. The ventilation system for neonatology

SLE5000 Neonatal Ventilator with High Frequency Oscillation

Minimum size for maximum performance

EMS INTER-FACILITY TRANSPORT WITH MECHANICAL VENTILATOR COURSE OBJECTIVES

HIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION

HAMILTON-C3 HAMILTON-C3. The compact high-end ventilator

V8600 Ventilator. Integrated Invasive & Noninvasive Ventilation

Peter Kremeier, Christian Woll. 2nd. Understanding and comparing modes of ventilation. The Kronberg List of Ventilation Modes

How Ventilators Work. Chapter 3

WILAmed INTENSA. Intensive Care Ventilator System

HELPING PROTECT YOUR MOST VULNERABLE NEWBORNS. Puritan Bennett 980 Neonatal Ventilator System

fabian Therapy evolution Our 2-in-1 device

NAVA Neurally Adjusted Ventilatory Assist In Neonates

O-Two Self-Study Guide. e600 Transport Ventilator Ventilation Modes

ASV. Adaptive Support Ventilation

Supporting you in saving lives!

Mechanical Ventilation. Flow-Triggering. Flow-Triggering. Advanced Concepts. Advanced Concepts in Mechanical Ventilation

Volume Guarantee New Approaches in Volume Controlled Ventilation for Neonates

Inspire rpap REVOLUTION FROM THE FIRST BREATH


Mechanical ventilation: simplifying the. terminology. Rapidly changing technology has unfortunately resulted in ambiguity in the terminology

GE Healthcare. Centiva/5 Critical Care Ventilator. Meet a new level of expectations

VENTILATION SERVO-s EASY AND RELIABLE PATIENT CARE

Test Bank for Pilbeams Mechanical Ventilation Physiological and Clinical Applications 6th Edition by Cairo

OPEN LUNG APPROACH CONCEPT OF MECHANICAL VENTILATION

fabian HFO Our 4-in-1 device

Mechanical Ventilation

HAMILTON-T1 HAMILTON-T1. Intelligent transport ventilation

Infant Ventilators. Infant Ventilator with Touch-Screen Operation SLE4000

V8800 Ventilator System. Your healthcare, we care.

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Expanding versatility. The upgraded Trilogy family of ventilators continues to meet the changing needs of your patients

Operating Instructions for Microprocessor Controlled Ventilators

High Frequency Ventilation. Neil MacIntyre MD Duke University Medical Center Durham NC USA

NEONATAL VOLUME VENTILATION CLINICAL GUIDELINE V2.0

Intelligent Ventilation solution from ICU to MRI

6 th Accredited Advanced Mechanical Ventilation Course for Anesthesiologists. Course Test Results for the accreditation of the acquired knowledge

Meet the changing needs of your patients. Trilogy Series, with advanced ventilator technology

excellence in care Procedure Management of patients with difficult oxygenation. For Review Aug 2015

Transcription:

(2009) 29, S73 S78 r 2009 Nature Publishing Group All rights reserved. 0743-8346/09 $32 www.nature.com/jp REVIEW Neonatal ventilators: how do they differ? Department of Pediatrics, Division of Neonatal-Perinatal Medicine, CS Mott Children s Hospital, University of Michigan Health System, Ann Arbor, MI, USA Remarkable technological advances over the past two decades have brought dramatic changes to the neonatal intensive care unit. Microprocessor-based mechanical ventilation has replaced time-cycled, pressure-limited, intermittent mandatory ventilation with almost limitless options for the management of respiratory failure in the prematurely born infant. Unfortunately, much of the infusion of technology occurred before the establishment of a convincing evidence base. This review focuses on the basic principles of mechanical ventilation, nomenclature and the characteristics of both conventional and high-frequency devices. (2009) 29, S73 S78; doi:10.1038/jp.2009.23 Keywords: prematurity; respiratory failure; mechanical ventilation; neonatal ventilators (HFOV) were widely used in the United States. 2 The decade of the 1990s saw the incorporation of the microprocessor into the neonatal ventilator and this greatly expanded the scope of neonatal ventilation, which became a symphony under the direction of a clinical conductor. The clinician is now able to control target modality, the mode of ventilation, tidal volume delivery, minute ventilation, the cycling mechanism, assist sensitivity and even the rise time on some devices. 3 Sophisticated, lightweight transducers with minimal deadspace now provide real-time breath-to-breath displays of pulmonary waveforms and loops, providing the clinician with valuable information regarding the patient ventilator interaction and allowing for adjustments in strategy as the patient or the disease changes. 4 Successful management may depend on an understanding of how these facets of ventilator management differ and how to best use them in an individual patient. It was the best of times. It was the worst of times. Charles Dickens A Tale of Two Cities (1859) Introduction Neonatal respiratory failure has been treated with mechanical ventilation since the early 1960s. 1 Early attempts used adult ventilators that were modified for neonatal use. The first widespread neonatal devices consisted of continuous flow, timecycled, pressure-limited devices to provide intermittent mandatory ventilation (IMV). For nearly a quarter of a century, this was the preferred method of treating newborns. Options were limited, as clinicians could control only the fraction of inspired oxygen, peak inspiratory pressure, positive end-expiratory pressure (PEEP), inspiratory time, rate and circuit flow. Patient ventilator asynchrony was a common problem, and spontaneous breathing between mechanical breaths was supported only by PEEP. In the 1980s, high-frequency ventilation was introduced as an alternative form of mechanical ventilation, and both high-frequency jet ventilation (HFJV) and high-frequency oscillatory ventilation Correspondence: Dr, Department of Pediatrics, Division of Neonatal-Perinatal Medicine, F5790 CS Mott Children s Hospital, 1500 E. Medical Center Drive, Ann Arbor, MI 48109-0254, USA. E-mail: smdonnmd@med.umich.edu Classification of neonatal ventilators Neonatal ventilators can be classified as either devices that deliver tidal ventilation, usually referred to as conventional mechanical ventilators, or devices that deliver smaller gas volumes at rapid rates, referred to as high-frequency ventilators (Table 1). Conventional (tidal) ventilation Conventional mechanical ventilators may be described by the target or limit modality (pressure or volume) and the manner in which this is achieved (such as pressure-limited or pressure-control, or volume-limited ventilation), and by the cycling mechanism, or mode. This is the manner in which the ventilator initiates and terminates the inspiratory phase of the respiratory cycle, such as by time or changes in airway flow or pressure. 5 There are four modes of ventilation, IMV, synchronized intermittent mandatory ventilation (SIMV), assist/control ventilation (A/C) and pressure support ventilation (PSV), a mode applied only to spontaneous breaths. 6 The latter three are patient-triggered modes, in which inspiration is initiated by the ventilator in response to a signal, which is representative of a spontaneous breath, such as a change in airway flow or pressure. Most current-generation neonatal ventilators are microprocessor based. They offer multiple modes and modalities and sophisticated interfaces, which enable real-time displays of graphic or

S74 Comparison of neonatal ventilator modalities alphanumeric data. This enables the clinician to customize ventilation to the individual needs and pathophysiology of the individual patient. This marks a radical departure from the early days of neonatal ventilation, when babies were essentially managed in a singular manner, largely on the basis of technological limitations. Table 1 Classification of neonatal ventilators Conventional (tidal ventilation) Modes Intermittent mandatory ventilation (IMV) Synchronized intermittent mandatory ventilation (SIMV) Assist/control ventilation (A/C) Pressure support ventilation (PSV) Modalities Pressure targeted Pressure limited Pressure control Volume targeted Hybrid High-frequency ventilation High-frequency jet ventilation (HFJV) High-frequency oscillatory ventilation (HFOV) Hybrids Modes of ventilation The mode of ventilation describes how mechanical breaths are delivered to the patient (Figure 1). IMV describes a mode in which all mechanical breaths are delivered to the patient at regular intervals chosen by the clinician. Breaths are delivered irrespective of patient effort and can thus result in significant asynchrony, wherein the baby may be trying to exhale against positive inspiratory pressure, leading to inefficient ventilation. 7 Asynchrony may result in significant disturbances, including pneumothorax and irregular cerebral blood flow velocity, and it has been associated with intraventricular hemorrhage. 8 In addition, spontaneous breaths are supported only by PEEP and may thus contribute to an increased work of breathing. SIMV uses timing windows to deliver a mandatory mechanical breath, which is synchronized to the onset of a spontaneous breath occurring within the timing window. If the patient fails to breathe, a breath is provided when the window closes. Expiratory asynchrony can still occur, and spontaneous breaths are also unsupported. A/C is a mode in which all spontaneous breaths that exceed the trigger sensitivity result in the delivery of a mechanical breath (assist) synchronous to the patient s inspiratory effort. If the patient fails to breathe or cannot trigger the ventilator, a control breath will be provided at the desired interval. However, the timing mechanism resets after each mechanically delivered breath, and Figure 1 Comparison of IMV, SIMV and A/C.

Comparison of neonatal ventilator modalities S75 hence the control breaths are not truly mandatory as they are in IMV or SIMV. PSV is a means to apply an inspiratory pressure burst to spontaneous breaths. It is usually used in conjunction with SIMV, but may be used alone in patients with reliable respiratory drive. PSV is a triggered mode, initiated and terminated by the patient [true PSV is flow-cycled (see below) and time-limited]. PSV can be adjusted to provide full support (pressure sufficient to deliver a full tidal volume breath) or partial support. 9 Modalities of ventilation The modality refers to the target or limit variable of the mechanical breath. Generally, there are only two modalities of neonatal conventional mechanical ventilation, pressure or volume. In pressure-targeted ventilation, pressure is limited, while volume is variable and is dependent on lung mechanics. In volume-targeted ventilation, volume is limited, while pressure is variable; as compliance improves, pressure is automatically weaned to provide the targeted volume (Figure 2). A mechanical breath may be classified according to three features: trigger, limit and cycle. The trigger refers to the mechanism used to initiate the breath, which may be time, or a change in airway pressure or flow. The limit refers to the targeted variable. Finally, the cycling mechanism is what causes the breath to end, or to cycle from inspiration to expiration. Time cycling is the oldest mechanism. Here, inspiration ends after a pre-set inspiratory time has elapsed, and the exhalation valve opens. The inspiratory time for each mechanical breath will be controlled by the ventilator and will be identical. Flow cycling describes a mechanism whereby inspiration ends when the decelerating inspiratory flow has decayed to a small (that is, 5%) percentage of the peak inspiratory flow. 10 The ventilator reads this as a signal that the patient is about to terminate his own breath, and cycles directly into expiration without reaching a zero-flow state. This enables the patient to control the inspiratory time (and rate) and allows for both inspiratory and expiratory synchrony. It also safeguards against gas trapping and inversion of the inspiratory:expiratory ratio during PTV, where time cycling can result in expiratory times that get progressively shorter the faster the patient breathes. Pressure-targeted modalities Most clinicians are familiar with time-cycled, pressure-limited ventilation, in which mechanical breaths are initiated and terminated by time, and which are limited by a pre-set inspiratory pressure limit that cannot be exceeded. The inspiratory flow is constant and the delivered tidal volume is related to pulmonary compliance. It will also be affected by the degree of asynchrony between the patient and the ventilator. Newer ventilators allow the choice of flow cycling to minimize asynchrony. Time-cycled, pressure-limited ventilation can be delivered using IMV, SIMV or A/C. Pressure control ventilation describes a pressure-targeted modality in which there is variable inspiratory flow, which is proportional to patient effort. Pressure control ventilation produces a rapidly accelerating inspiratory flow waveform, which peaks and then rapidly decelerates. It results in rapid pressurization of the ventilator circuit and delivery of flow (and hence volume) early in inspiration. 11 In other words, the mechanical breath is front end loaded and peak pressure and volume delivery occur early in inspiration. Some devices also have an adjustable inspiratory rise time, which allows a qualitative modification in the slope of the inspiratory pressure waveform. Pressure control ventilation is usually time-cycled, although some ventilators enable it to be flow-cycled. Pressure control ventilation breaths can be applied with IMV, SIMV or A/C. PSV is also pressure-targeted. It is pressure-limited, flow-cycled and has a variable inspiratory flow. It is applied only to spontaneous breaths, either during SIMV or used alone in patients with reliable respiratory drive. Rise time may also be adjusted on some devices. Table 2 summarizes the features of the three pressure-targeted modalities. Volume-targeted modalities Volume-targeted modalities target or limit the delivered gas volume. However, because cuffed endotracheal tubes are not used in neonatal patients, true volume cycling cannot be accomplished. There is always some degree of leak around the endotracheal tube, Table 2 Features of pressure-targeted modalities Figure 2 Comparison of pressure- and volume-targeted modalities. Note that with pressure targeting, the delivered tidal volume depends on compliance; at low compliance, less tidal volume is delivered at the same pressure. With volume targeting, volume delivery is constant; if compliance is low, the ventilator will increase the pressure, and conversely, if compliance improves, pressure will be auto-weaned. Pressure limited Pressure control Pressure support Limit variable Pressure Pressure Pressure Flow Continuous Variable inspiratory Variable inspiratory Cycle mechanisms a Time or flow Time or flow Flow, with time limit a Cycle mechanisms vary according to the device.

Comparison of neonatal ventilator modalities S76 and additionally, a certain amount of the delivered tidal volume will be compressed within the ventilator circuit (especially if compliance is poor). This is referred to as compressible volume loss. 12 Volume-targeted modalities have differing algorithms on various ventilators. Some use a breath-averaging technique or make adjustments on the basis of the previous breath, whereas others control the amount of volume delivered to the circuit. A key element for all systems is the ability to measure the volume of gas that actually reaches the patient by measuring it not at the machine but at the patient wye. 13 Volume-targeted breaths classically produce a square flow waveform, in which inspiratory flow is held constant until inspiration ends. This results in a steady ramping of pressure and volume, such that peak volume delivery and pressure occur at the end of inspiration. Thus, volume target breaths are back end loaded and may have inherent advantages in some clinical situations. Clinical evidence IMV vs PTV Numerous small studies have compared IMV with both SIMV and A/C using mostly physiological (for example, gas exchange, pressure, rapidity of weaning, etc.) or short-term (for example, oxygen dependency at 36 weeks postmenstrual age) outcome measures. 7,14 22 The balance of these studies used first- or secondgeneration devices, and yet virtually all have shown the superiority of PTV, with the exception of the Baumer trial. However, this trial, the largest to date, was an open trial that used a multiplicity of devices and had variable levels of experience among collaborating centers. Because of its large size, it has skewed the meta-analytic conclusion from showing a positive effect to one of no difference. Pressure targeted vs volume targeted Several clinical trials have compared pressure- and volumetargeted ventilation for treating preterm babies with respiratory distress syndrome (RDS). 23 25 All are relatively small and addressed primarily physiological (mainly pneumothorax and duration of ventilation) and short-term outcomes. A recent meta-analysis of these trials indicates that there is a statistically significant reduction in pneumothorax and in the duration of ventilation, as well as a very strong trend toward decreased chronic lung disease (CLD) in babies treated with volume-targeted ventilation. The trial of Singh et al. 25 also looked at longer-term follow-up and showed a medical correlate at 24 months corrected age, with increased hospitalizations, wheezing and cough in infants treated with pressure-targeted ventilation compared with their volume-targeted counterparts. 26 Pressure support ventilation Trials examining the use of PSV in newborns have also been limited to physiological effects and short-term outcome measures, but the results are encouraging and suggest the need for future trials. 6 The studies of Reyes et al. 27 and Gupta et al. 28 show the benefits of PSV in improving spontaneous tidal volumes and minute ventilation. Further work needs to be carried out to address the optimal balance between mechanical and spontaneous support. High-frequency ventilation High-frequency ventilation differs from CMV in providing smaller tidal volumes, often less than the anatomical deadspace, at a much more rapid rate. The major devices used in the United States include HFJV and HFOV. 29 HFJV delivers tidal volumes of 1 to 3 ml kg 1 at rates between 240 and 660 breaths per minute. These are injected directly into the airway or through a special triple lumen endotracheal tube. Exhalation during HFJV is passive, similar to that during CMV. It is used in tandem with CMV, which provides PEEP and may be used to give sigh breaths or low-rate CMV. In clinical practice, management with HFJV is similar to that with CMV. Oxygenation is proportional to mean airway pressure and ventilation is proportional to amplitude (the difference between peak inspiratory pressure and PEEP). Jet pulsations are created by a flow interrupter placed close to the patient wye. This produces a high-velocity laminar flow, which has the ability to bypass airway disruptions, making it suitable to treat significant thoracic air leaks. HFOV differs from HFJV in several ways. First, the delivered tidal volumes are smaller (1 to 2 ml kg 1 ) and are delivered at a faster rate, usually 8 to 15 Hz. Second, the most commonly used device operates by applying a piston to a diaphragm, pushing gas into the lung during inspiration, and actively drawing gas from the airway during expiration. With HFOV, the lung is inflated to a static volume, then oscillated around the mean airway pressure. In this manner, oxygenation and ventilation are uncoupled. Changes in oxygenation are facilitated by adjusting the mean airway pressure, and changes in ventilation are made by altering the amplitude. These can generally be carried out independently of each other. Although the mean airway pressure measured at the proximal airway tends to be higher with HFOV than with CMV, the actual pressure seen at the alveoli is actually lower because of the very short inspiratory time and the damping effect of the airways. Clinical evidence HFJV has been shown to be a more effective means of treating pulmonary interstitial emphysema than rapid-rate CMV, with shorter time to resolution and decreased mortality in the sub-group of smallest infants. 30 Keszler et al. 31 also showed a decreased incidence of CLD and the need for home oxygen in infants enrolled in an RDS treatment trial. Friedlich et al. 32 showed a significant

Comparison of neonatal ventilator modalities S77 reduction in the mean airway pressure in a group of infants with hypoxemic respiratory failure unresponsive to either CMV or HFOV. HFOV is a popular rescue technique but has also been applied as a primary treatment strategy in babies with RDS. Study results have been conflicting. Other than in the Provo trial, 33 mortality does not seem to have been affected. Some studies suggest a modest reduction in CLD in survivors, but with no reduction in short-term neurological outcomes. Two high-frequency rescue trials, one with HFOV and one with HFJV, failed to show a reduction in the need for extracorporeal membrane oxygenation therapy in eligible infants, 33,34 although HFOV seems to be a better rescue strategy than IMV. 35 Concern over an increased incidence in intracranial hemorrhage has been expressed. Two recent studies, published in the same journal and using similar methodologies, found differing results. 36 Courtney et al. 37 compared HFOV with SIMV in infants of 600 to 1200 g birthweight and found a small reduction in the incidence of CLD. However, it is plausible that this difference might be more attributable to the use of SIMV in the control group, and that if A/C had been used, it might have even disappeared. Johnson et al. 38 reported the results of the UKOS study, a comparison of HFOV with time-cycled, pressure-limited ventilation in infants of 23 to 28 weeks gestation, and found no reduction in CLD. The Cochrane review shows no difference in mortality, but there are some trends favoring HFOV, including decreased CLD (oxygen need at 28 to 30 days) in survivors, decreased death or CLD (oxygen need at 28 to 30 days) and decreased CLD (oxygen at 36 to 37 weeks PMA). However, there are trends toward an increase in severe intraventricular hemorrhage and periventricular leukomalacia, and an increased incidence of air leaks. When a high-volume strategy was used in the study design, pulmonary outcomes were more favorable. 39 On balance, HFOV is not recommended as a primary strategy for RDS. 29,39 On the horizon As technological advances continue to occur at rapid rates, new techniques are being explored to improve the outcomes of preterm infants with RDS. Closed loop control of oxygen is under investigation to determine whether more tightly controlled oxygen saturation in ventilated babies can reduce complications. 40 Various forms of adaptive ventilation are also in the preliminary stages of development and investigation, including proportional assist ventilation, 41 mandatory minute ventilation, 42 volume-assured pressure support 43 and adaptive support ventilation. 44 Conclusion Tremendous technological advances have revolutionized the mechanical ventilation of newborns. However, this has been a double-edged sword. Ventilation has never been safer, but it has also never been this complicated. Clinicians need to approach ventilator management in the same manner as that in which they approach pharmaceuticals. They ought to learn the prototype first, master concepts and principles, learn indications and limitations, and then expand the use. In addition, they should pay attention to evidence and allow it to augment experience. Disclosure is a paid consultant for Cardinal Health and has grant support from Dey, LP. This paper was based on a talk presented at the Evidence vs Experience in Neonatal Practices Fifth Annual CME Conference, which was supported by an unrestricted educational grant from Dey, LP. References 1 Delivoria-Papadopoulos M, Levison H, Swyer P. Intermittent positive pressure respiration as a treatment in severe respiratory distress syndrome. Arch Dis Child 1965; 40: 474 478. 2 Bunnell JB. High frequency ventilation: general concepts. In: Donn SM, Sinha SK (eds). Manual of Neonatal Respiratory Care, 2nd edn. Mosby Elsevier: Philadelphia, 2006, pp 222 230. 3 Sinha SK, Donn SM. Newer forms of conventional ventilation for preterm newborns. Acta Paediatr 2008; 97: 1338 1343. 4 Becker MA, Donn SM. Real-time pulmonary graphic monitoring. Clin Perinatol 2007; 34: 1 17. 5 Hagus CK, Donn SM. Pulmonary graphics: basics of clinical application. In: Donn SM (ed). Neonatal and Pediatric Pulmonary Graphics. Principles and Clinical Applications. Futura Publishing Co.: Armonk, NY, 1998, pp 81 127. 6 Sarkar S, Donn SM. In support of pressure support. Clin Perinatol 2007; 34: 117 128. 7 Donn SM, Nicks JJ, Becker MA. Flow-synchronized ventilation of preterm infants with respiratory distress syndrome. J Perinatol 1994; 14: 90 94. 8 Perlman JM, McMenanim JB, Volpe JJ. Fluctuating cerebral blood flow velocity in respiratory distress syndrome: relationship to subsequent development of intraventricular hemorrhage. N Engl J Med 1983; 309: 204 209. 9 Sinha SK, Donn SM. Pressure support ventilation. In: Donn SM, Sinha SK (eds). Manual of Neonatal Respiratory Care, 2nd edn. Mosby Elsevier: Philadelphia, 2006, pp 212 215. 10 Servant GM, Nicks JJ, Donn SM, Bandy KP, Lathrop C. Feasibility of applying flow-synchronized ventilation to very low birthweight infants. Respir Care 1992; 37: 249 253. 11 Donn SM. Pressure control ventilation. In: Donn SM, Sinha SK (eds). Manual of Neonatal Respiratory Care, 2nd edn. Mosby Elsevier: Philadelphia, 2006, pp 210 211. 12 Sinha SK, Donn SM. Volume-controlled ventilation. In: Goldsmith JP, Karotkin EH (eds). Assisted Ventilation of the Neonate, 4th edn. WB Saunders: Philadelphia, 2003, pp 171 183. 13 Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN et al. Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 2000; 162: 2109 2112. 14 Dimitriou G, Greenough A, Cherian S. Comparison of airway pressure and air flow triggering systems using a single type of neonatal ventilator. Acta Paediatr 2001; 90: 445 447. 15 Greenough A, Milner AD, Dimitriou G. Synchronized ventilation. Cochrane Database Syst Rev 2005; Supplement 3, Art No. CD000456. 16 Baumer JH. International randomized controlled trial of patient triggered ventilation in neonatal respiratory distress syndrome. Arch Dis Child Fetal Neonatal Ed 2000; 82: F5 F10.

S78 Comparison of neonatal ventilator modalities 17 Chan V, Greenough A. Randomized controlled trial of weaning by patient triggered ventilation or conventional ventilation. Eur J Pediatr 1993; 152: 51 54. 18 Chan V, Greenough A. Comparison of weaning by patient triggered ventilation or synchronous intermittent mandatory ventilation in preterm infants. Acta Paediatr 1994; 83: 335 337. 19 Dimitriou G, Greenough A, Giffin F. Synchronous intermittent mandatory ventilation modes compared with patient triggered ventilation during weaning. Arch Dis Child Fetal Neonatal Ed 1995; 72: F188 F190. 20 Bernstein G, Mannino FL, Heldt GP, Callahan JD, Bull DH, Sola A et al. Randomized multicenter trial comparing synchronized and conventional intermittent mandatory ventilation in neonates. J Pediatr 1996; 128: 453 463. 21 Roze JC, Liet JM, Gournay V, Debillon T, Gaultier C. Oxygen cost of breathing and weaning process in newborn infants. Eur Respir J 1997; 10: 2583 2585. 22 Dimitriou G, Greenough A, Laubscher B, Yamaguchi N. Comparison of airway pressure-triggered and airflow-triggered ventilation in very immature infants. Acta Paediatr 1998; 87: 1256 1260. 23 Sinha SK, Donn SM, Gavey J, McCarty M. Randomised trial of volume controlled versus time cycled, pressure limited ventilation in preterm infants with respiratory distress syndrome. Arch Dis Child Fetal Neonatal Ed 1998; 79: F202 F205. 24 Lista G, Castoldi F, Bianchi S, Battaglioli M, Cavigioli F, Bosoni MA. Volume guarantee versus high-frequency ventilation: lung inflammation in preterm infants. Arch Dis Child Fetal Neonatal Ed 2008; 93: F252 F256. 25 Singh J, Sinha SK, Clarke P, Byrne S, Donn SM. Mechanical ventilation of very low birth weight infants: is volume or pressure a better target variable? J Pediatr 2006; 149: 308 313. 26 Singh J, Alsop E, Gupta S, Sinha SK, Donn SM. Volume-controlled versus pressurelimited ventilation in very low birth weight infants: long term follow-up of a RCT. Pediatr Acad Soc 2007, Abstract number 752140. 27 Reyes ZC, Claure N, Tauscher MK, D Ugard C, Vanbuskirk S, Bancalari E. Randomized, controlled trial comparing synchronized intermittent mandatory ventilation and synchronized intermittent mandatory ventilation plus pressure support in preterm infants. Pediatrics 2006; 118: 1409 1417. 28 Gupta S, Sinha SK, Donn SM. The effect of two levels of pressure support ventilation on tidal volume delivery and minute ventilation in preterm infants. Arch Dis Child 2009; 94: F80 F83. 29 Lampland AL, Mammel MC. The role of high-frequency ventilation in neonates: evidenced-based recommendations. Clin Perinatol 2007; 34: 129 144. 30 Keszler M, Donn SM, Bucciarelli RL, Alverson DC, Hart M, Lunyong V et al. Multicenter controlled trial comparing high-frequency jet ventilation and conventional mechanical ventilation in newborn infants with pulmonary interstitial emphysema. J Pediatr 1991; 119: 85 93. 31 Keszler M, Modanlou HD, Brudno DS, Clark FI, Cohen RS, Ryan RM et al. Multicenter controlled clinical trial of high-frequency jet ventilation in preterm infants with uncomplicated respiratory distress syndrome. Pediatrics 1997; 100: 593 599. 32 Friedlich P, Subramanian N, Sebald M, Noori S, Seri I. Use of high-frequency jet ventilation in neonates with hypoxemia refractory to high-frequency oscillatory ventilation. J Mat Fetal Neonatal Med 2003; 13: 398 402. 33 Gerstmann DR, Minton SD, Stoddard RA, Meredith KS, Monaco F, Bertrand JM et al. The Provo multicenter early high-frequency oscillatory ventilation trial: improved pulmonary and clinical outcome in respiratory distress syndrome. Pediatrics 1996; 98: 1044 1057. 34 Clark RH, Yoder BA, Sell MS. Prospective, randomized comparison of high-frequency oscillation and conventional ventilation in candidates for extracorporeal membrane oxygenation. J Pediatr 1994; 124: 447 454. 35 Engle WA, Yoder MC, Andreoli SP, Darragh RK, Langefeld CD, Hui SL. Controlled prospective randomized comparison of high-frequency jet ventilation and conventional ventilation in neonates with respiratory failure and persistent pulmonary hypertension. J Perinatol 1997; 17: 3 9. 36 Moriette G, Paris-Llado J, Walti H, Escande B, Magny JF, Cambone G et al. Prospective randomized multicenter comparison of high-frequency oscillatory ventilation and conventional ventilation in preterm infants of less than 30 weeks with respiratory distress syndrome. Pediatrics 2001; 107: 363 372. 37 Courtney SE, Durand DJ, Asselin JM, Hudak ML, Aschner JL, Shoemaker CT et al. High-frequency oscillatory ventilation versus conventional mechanical ventilation for very-low-birth-weight infants. N Engl J Med 2002; 347: 643 652. 38 Johnson AH, Peacock JL, Greenough A, Marlow N, Limb ES, Marston L et al. High-frequency oscillatory ventilation for the prevention of respiratory distress syndrome of prematurity. N Engl J Med 2002; 347: 633 642. 39 Bhuta D, Henderson-Smart DJ. Rescue high-frequency oscillatory ventilation versus conventional ventilation for pulmonary dysfunction in preterm infants. Cochrane Database Syst Rev 1998; Supplement 2, Art No. CD000438. 40 Claure N, Gerhardt T, Everett R, Musante G, Herrera C, Bancalari E. Closed-loop controlled inspired oxygen concentration for mechanically ventilated very low birth weight infants with frequent episodes of hypoxemia. Pediatrics 2001; 107: 1120 1124. 41 Schulze A, Rieger-Fackeldey E, Gerhardt T, Claure N, Everett R, Bancalari E. Randomized crossover comparison of proportional assist ventilation and patient-triggered ventilation in extremely low birth weight infants with evolving chronic lung disease. Neonatology 2007; 92: 1 7. 42 Guthrie SO, Lynn C, Lafleur BJ, Donn SM, Walsh WF. A crossover analysis of mandatory minute ventilation compared to synchronized intermittent mandatory ventilation in neonates. J Perinatol 2005; 25: 643 646. 43 Amato MB, Barbas CS, Bonassa J, Saldiva PH, Zin WA, de Carvalho CR. Volume-assured pressure support ventilation (VAPSV). A new approach for reducing muscle workload during acute respiratory failure. Chest 1992; 102: 1225 1234. 44 Arnal JM, Wysocki M, Nafati C, Donati S, Granier I, Corno G et al. Automatic selection of breathing pattern using adaptive support ventilation. Intensive Care Med 2008; 34: 75 81.