STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

Similar documents
STERILIZABLE DISSOLVED OXYGEN SENSORS

OxyProbe II Instruction Manual

LogIT Dissolved Oxygen Probe

PRODUCT SHEET. Order probe only as RXPROBE02

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

DISSOLVED OXYGEN SENSOR BT34i

Oxygen Sensor with Electrode

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Vinmetrica Dissolved Oxygen System User Manual

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

BI-680 Online Dissolved Oxygen Controller Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Heavy Duty Dissolved Oxygen Meter

Dissolved Oxygen Guide

SS32L Dissolved O 2 Probe

Dextens Thermo Conductivity Sensors. Type 52101, 54101, 55101

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Dissolved Oxygen Probe (Order Code DO-BTA)

Dissolved Oxygen/Temp Waterproof Tester

Oxymax H COS21D. Technical Information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

NEULOG OXYGEN LOGGER SENSOR GUIDE

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

Best Practice for Calibrating LTH Conductivity Instruments

12mm Dissolved Oxygen Sensor Care and Use Instructions

Instruction Manual PN SO/rev.A April Model 1181SO. Dissolved Oxygen Two-Wire Transmitter

& Dissolved Oxygen Sensor. Instruction Manual

Dissolved Oxygen Probe (Order Code DO-BTA)

MODEL GT820 OXYGEN SENSOR

A LAZAR LABORATORIES HANDBOOK DISSOLVED OXYGEN PRIMER

Columbus Instruments

Powers Controls RC 195 Multiple Input Receiver- Controller

WELLS JOHNSON HIGH VOLUME CANISTERS

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

T02.0.3X.6C-06 Page 1 of 6. Operating Manual for Dissolved Oxygen-Measuring transducer OXY 3610 MP

Powers Controls RC 195 Single Input Receiver-Controller

HI 9143, HI 9143M HI 9145

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

Unisense In Situ UniAmp Connector System

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

QUICK START GUIDE TO DRINKING WATER MONITORING

Oxygen Dipping Probe PSt3 / PSt6 / PSt9

Instruction Manual HI 9146 Dissolved Oxygen and Temperature Meter

Aerobic Respiration. Evaluation copy

PRACTICAL TIPS FOR TRACE-LEVEL OXYGEN MONITORING WITH ELECTRO-CHEMICAL SENSORS

Dissolved Oxygen Sensor

PERME- OX2/230 Oxygen Transmission Rate test system

Free Chlorine Sensor For additional information, please visit our website at

Radnoti Liver Perfusion System

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Oxygen Measurement Systems High Reliability and Wide Application Coverage

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

Generating Calibration Gas Standards

1. ABOUT THE PRODUCT:

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Instruction Manual MicroChem 2 Chlorine Probes Series CL4000

Installation, operating and maintenance Instructions for Seemag bypass level indicator

INSTRUCTIONS FOR USING THE CREATECH PERMEOMETER MODEL 210T FOR MEASURING OXYGEN PERMEABILITY OF CONTACT LENS MATERIALS BY I. FATT

Best Practice Guide, Servomex 2700

A N F G J I C D D L A B C D E F

Gas Sampling Fitting with MK2.1 and L-Probe

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

MARTINI. instruments. THANK YOU FOR CHOOSING. Sales and Technical Service Contacts:

Digital Model 10 and Model 20 Controller

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

O3 3E 1 F Gas Sensor Module

ph measuring system for hygienic process applications

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

OxiQuant S. Operating Instructions

with O 2 Controller Instruction Manual

ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL

OPERATION. Estimated kerf width compensation. HPR260 Manual Gas Instruction Manual 4-9

Quick Start Guide LIQ-QSG-499ACL-03, Rev I July Rosemount 499ACL-03. Monochloramine Sensor

Oxymax W COS51D. Technical Information. Digital sensor for measuring dissolved oxygen Membrane-covered amperometric sensor

Technical Information Oxymax COS22D/COS22

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

PRO20 USER MANUAL. English. Français. Español. Deutsch

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

User Manual DOCUMENT # Pro20 Pro20i USER MANUAL. English

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal

CalCheck. Instrument User Manual V1.8

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

OWNER S MANUAL NMB SERIES

Operation Manual. O2 Quickstick. Oxygen Analyzer

HUNTER Beeper Collar Manual

Transcription:

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS oxygen amplifiers and other manufacturer s amplifiers designed to the same specifications. The probe is designed to withstand severe conditions, like high temperature, high pressure, and moisture during sterilization. SPECIFICATIONS Electrode body Membrane Cathode Anode : Polished 316L stainless steel. : PTFE/Silicone (stainless steel reinforced). : Platinum, diameter 0.10mm. : Tubular Ag. Output in air-saturated DI water : 50 na. Residual Signal : less than 1 na. Response time : less than 60 seconds to 98% of full response. THEORY OF OPERATION The polarographic dissolved oxygen electrode was first described by Clark in 1956. It consists of polarized platinum and silver electrodes, with the electrolyte separated from the sample by a PTFE/silicone gas permeable membrane. Oxygen diffuses across the electrode membrane and is reduced to hydroxyl ions at the platinum cathode according to the reaction; (1) O 2 + 2H 2 O + 4e - = 4OH - The electrons necessary for this process are produced by a reaction at the silver anode. Because the electrolyte contains chloride ions, this reaction occurs as; (2) 4Ag + 4Cl - = 4AgCl + 4e - At any given temperature, the current flow between cathode andanode is directly proportional to the level of oxygen outside the membrane. ZERO POINT OF AMPLIFIER The electrical zero point of the amplifier must be first set. (Follow the manufacturer s operating instructions).

TEMPERATURE COMPENSATION Temperature compensation is not required when calibration is made at the same temperature as the controlled fermentation temperature. The fermentation must be temperature controlled, with no temperature variation. POLARIZATION When the amplifier is switched on, polarization voltage is applied to the anode and the cathode. Initial current is very strong, but falls off exponentially and settles down to a steady state after a few hours. Since the equilibrium period of the polarization current is relatively long, it is advisable to keep the electrode connected to the amplifier when not in use. The small current flowing through the electrode will not shorten the life of the electrode. If, for any reason, the electrode must be disconnected, or the amplifier switched off, the electrode must be repolarized before it can be used. During the polarization period, the electrode current will decrease even in oxygen free solutions. An excessive zero current may indicate incomplete polarization. ELECTRODE CONSTRUCTION The external portion of the electrode is constructed of a polished stainless steel cylinder. The functional portion of the electrode is constructed of a ryton chamber enclosing an ultra-pure tubular Ag anode and a 0.10 mm diameter cathode. A thin PTFE/silicone membrane, reinforced by a stainless steel mesh at the tip, provides for efficient sealing of the system and is in contact with the platinum cathode and electrolyte. The membrane is permeable to oxygen but impermeable to water and electrolyte. PREPARATION The electrode is shipped filled with electrolyte and with PTFE/silicone membrane in place. The protective vinyl cap on the sensing tip must be removed before and during sterilization. Upon receiving the electrode, inspect for damage. If any damage is noticed, notify phoenix Electrode Company Service Department immediately. ELECTRODE CHECK Remove protective vinyl cap. Connect the detachable electrode cable to the probe and amplifier. Follow polarization procedures (see POLARIZATION SECTION) Leave electrode in air saturated DI water. Set the range selector on the amplifier to zero. Adjust the zero current to read zero on the indicator. Then calibrate the meter to 100% saturation degree. Pass nitrogen gas over the electrode membrane. The electrode will react immediately increasing the reading, which will decrease after 45 seconds. The reading should be under 10%. (Refer to the amplifier manufacturer s procedure on checking the electrode).

ZERO POINT OF THE ELECTRODE The zero current of the dissolved oxygen electrode (electrode current in pure N 2 ) is usually negligibly small and almost identical with the amplifier zero point. Nonetheless, the electrode zero point should be periodically checked as some electrode faults result in excessive zero current. Moreover, checking the zero point calibration is necessary before measurement of low oxygen concentration. The zero current(residual current) of the electrode may be compensated by adjusting the zero button on the meter. After zeroing the electrode, recalibrate it in air-saturated water. ELECTRODE SLOPE Zero point adjustment must precede slope calibration. Unlike zero point calibration, the aqueous phase is preferred for slope adjustment. The following problem arise in calibration in air: 1) Membrane permeability slightly differs in air and water. 2) Relative air humidity rarely attains 100%. 3) The temperature is poorly defined. When calibrating in a fermentor, slope calibration should be performed only after sterilization or the electrode slope may be altered. This alteration may be comparatively large, particularly when a new membrane is used for the first time. After cooling, allow the vessel to reach operating temperature. This is very important when a temperature compensator is not used. Saturate the aqueous phase with air in the vessel by purging air at the maximum flow rate for approximately 20 minutes, allowing the output signal to stabilize. Adjust the display module to 100%. Calibration is usually most effective at an air saturation of 100%. Since calibration depends on pressure, the 100% air saturation should be done under operating pressure. (Refer to amplifier manufacturer s operating procedures). STERILIZATION Electrodes should be installed in the headplate of the fermentor or in the side port, depending on the type of fermentor. During sterilization, particular care should be exercised during autoclaving when localized boiling tends to occur. Either maintain a slow cooling rate or pressurize the autoclave during sterilization. Consult your autoclave manufacturer for information regarding a pressure balance feature and proper autoclaving procedures to eliminate localized boiling during the cooling cycle. After sterilization and cool down, electrolyte may be replenished by following the filling procedure. DO NOT remove the electrode from the fermentor.

CALIBRATION Connect the electrode cable to the DO probe and to the DO amplifier. Follow the polarization procedure as described, along with the amplifier manufacturer s operating manual. Dissolved oxygen electrodes should be recalibrated prior to every measurement. If work is performed under sterile conditions, the system must be calibrated after sterilization, but prior to inoculation. MEMBRANE REPLACEMENT The membrane should be examined routinely after each fermentation cycle and replaced if any deterioration is evident. To remove the membrane, unscrew the bottom stainless steel cap with the membrane in place. (Figure 1). Inspect the glass part and platinum cathode at the very tip of the glass with tissue paper. Remove a new membrane from it s package. Use a screwdriver to replace membrane as shown in Figure 1. Fill the membrane cap with electrolyte and screw the cap onto the tip of the electrode. After the membrane cap is securely in place, the PTFE/silicone membrane should be stretched across the glass cathode. The electrode is to be filled with electrolyte, following the above procedures (see PREPARATION SECTION). After re-assembling the membrane and filling the electrode with electrolyte, it must be checked. The polarization and sterilization procedures should be followed. TROUBLESHOOTING The probe when in air-saturated water generates no potential (zero output). Check that all signal cables connections are correct. Check that the probe is filled with electrolyte. The output signal is reversed. Reverse the signal wires connected to the display module or recorder. The calibration potentiometer is on maximum but the signal does not reach 100%. Change the membrane. Change the electrolyte.

The output signal is unstable and shows random drift. Check the grounding of the fermentor and of the instrumentation. NOTE: The output signal may not reach 100% if the ambient airtemperature is less than 15%. TROUBLESHOOTING DURING FERMENTATION During the sterilization process, 0.5-1.0 ml of electrolyte will be lost due to expansion. Check that the signal is not being affected by external switching gear, power relays, etc. Check that the slow random drift is not due to poor mixing of culture medium. If the fault persists, it may be due to a badly fitted or incorrectly tightened membrane. The output signal is stable, but shows a regular sinusoidal drift. Check the proportional band setting of the DO control. Check the DO control on the fermentor. During a long fermentation, the signal drifts to a value that appears incorrect. Check to see if the membrane is split or has been fouled by the culture medium. MAINTENANCE After the electrode has been used for a period of time, (at least five (5) cycles), the residual current may rise. To minimize the residual current, the following cleaning method is recommended: Gently unscrew the membrane cover from the body, rinse the inside of the membrane with DI water. Flush the electrolyte chamber with DI water. Soak the silver anode and the glass tip in silver cleaning solution (commercially availableas jewelry cleaning solution) for about 10 minutes. Thoroughly rinse the soaked portion with DI water, blot dry with tissue paper. Fill the electrode with electrolyte as described above. PACKING LIST phoenix Electrode Company polarographic DO probe Electrolyte 1 Spare membrane 1 Membrane Mounting tool 1 O-rings 2 Plastic syringe 1

Polarographic Oxygen Electrode Membrane Replacement Procedure A E B F D J G D C H I G F E D A. Detachable Lead B. O-Ring Seal C. 316 Stainless Steel Body Figure 1 H D. O-Ring Seal E. Platinum Cathode F. Ag/AgCl anode G. Replaceable PTFE/silicone & stainless steel mesh membane H. Membrane Cover I. 16mm OD with M18 x 1.5 thread J. Membrane replacement tool