SS32L Dissolved O 2 Probe

Similar documents
PRODUCT SHEET. Order probe only as RXPROBE02

Evaluation copy. Interdependence of Plants and Animals. computer OBJECTIVES MATERIALS

12mm Dissolved Oxygen Sensor Care and Use Instructions

Experiment C-6 Gas Solubility

Dissolved Oxygen in Water. Evaluation copy. Table 1. Temperature Range ( C) Trout Smallmouth bass Caddisfly larvae

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Aerobic Respiration. Evaluation copy

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

The Oxygen Sensor can be connected to all types of einstein Tablets, einstein LabMate, and einstein LabMate +.

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Heavy Duty Dissolved Oxygen Meter

DISSOLVED OXYGEN SENSOR BT34i

A180 DO/BOD/OUR/SOUR Meter Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Oxygen Sensor with Electrode

Recording established using the VO2 and RER graph template file (h19.gtl)

Dissolved Oxygen Meter. ino2. Instruction Manual. Serial No..

Introduction The Smart Q Oxygen Adaptor enables an Oxygen electrode to be connected to an EasySense unit.

Dissolved Oxygen/Temp Waterproof Tester

Fish Respiration and Q 10

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

LogIT Dissolved Oxygen Probe

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

QUICK START GUIDE TO DRINKING WATER MONITORING

STERILIZABLE DISSOLVED OXYGEN SENSORS

Vinmetrica Dissolved Oxygen System User Manual

Optical Dissolved Oxygen Meter

Bante904 Benchtop Conductivity/Dissolved Oxygen Meter Instruction Manual

Instruction Manual DO 6. Economy Hand-held Dissolved Oxygen Meter. 68X Rev 2 Apr 08. Technology Made Easy... Part of Thermo Fisher Scientific

Oxymax H COS21. Technical Information. Sensor with long-term stability for frequent sterilization and autoclaving

Instruction Manual HI Portable Dissolved Oxygen Meter. DO meter HI 8043 STB C OFF O 2 ZERO SLOPE C. ins rum n s

SOLUBILITY OF A SOLID IN WATER

Operating manual. TA 197 Oxi. Dissolved Oxygen Depth Armature

Water Quality Monitoring Using a Handheld Sonde for in situ Measurements Date Prepared: 11/17/2017

Three Lakes Association Hydrolab for Dummies. Table of Contents:

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Dissolved Oxygen Sensor

Refillable ORP/Redox Probe: Model MTC30101 or MTC30103

FLUORESCENCE DETERMINATION OF OXYGEN

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

HI 9143, HI 9143M HI 9145

NEULOG OXYGEN LOGGER SENSOR GUIDE

Biology Unit 2, Structure of Life, Lab Activity 2-3

Optical Dissolved Oxygen Meter

SOLUBILITY OF A SOLID IN WATER

Lesson 12 New Procedure PULMONARY FUNCTION I

Dissolved Oxygen Probe (Order Code DO-BTA)

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Instruction Manual. DO 700 Dissolved Oxygen/ºC/ºF Bench Meter. Technology Made Easy...

Vapor Pressure of Liquids

Aqua-DP. Aqua-DP Waterproof Dissolved Oxygen - ph -mv Temperature Meter. TPS Australia - Quality hand-made instruments since 1968.

Vapor Pressure of Liquids

YSI Model 550a Dissolved Oxygen (DO) Meter

Instruction Manual DO 6+ Dissolved Oxygen/Temp

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

SensoDirect Oxi200 Dissolved Oxygen Meter Operating instructions

OPTIMA 7 - BIOGAS Calibrating the gas sensors SENSOR ADJUSTMENT. Diamond Scientific 625 Peachtree St., Cocoa, Florida /Tel:

D-Opto Dissolved Oxygen Sensor Operation Manual for the D-Opto 4-20mA

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

Micro600DO_ZI INST /1/17 13:16 Page 1. Palintest. Micro 600 DO Meter. Palintest ZI INST 1240

MARTINI. instruments. THANK YOU FOR CHOOSING. Sales and Technical Service Contacts:

Vapor Pressure of Liquids

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

What does the % represent on the beakers?

Dissolved Oxygen Probe (Order Code DO-BTA)

A LAZAR LABORATORIES HANDBOOK DISSOLVED OXYGEN PRIMER

Vernier Optical DO Probe (Order Code ODO-BTA)

Instruction Manual MicroChem 2 Chlorine Probes Series CL4000

Using the Akta Prime plus October 22, 2012

Radnoti Liver Perfusion System

Evaluation copy. Wind Chill. computer OBJECTIVES MATERIALS

What is photosynthesis? Photosynthesis II: Dissolved O2

Experiment P18: Buoyant Force (Force Sensor)

Measuring the Conductivity of Pure Water Using USP<645>

ACQUIDATA FAMILY CALIBRATION PROCEDURE CALIBRATION CHECKING

6-SERIES ENVIRONMENTAL MONITORING SYSTEMS. Service Manual

Yeo-Kal Electronics Pty Ltd 18/26 Wattle Road, Brookvale NSW Australia Telephone Fax USER MANUAL

Instruction Manual. Multiparameter Benchtop Meter

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

CBL Lab GAS PRESSURE & VOLUME MATHEMATICS CURRICULUM. High School. Florida Sunshine State Mathematics Standards

AKTA pure 25 New Owner s Intro

UsER manual for Watersens ph -REDOX

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

OxyProbe II Instruction Manual

Vapor Pressure of Liquids

OPERATION MANUAL DIVE CALIPER NOT SCALE / IF IN DOUBT ASK* *ALL UNITS IN MM/DEGREES UNLESS SPECIF. Dive Caliper General DC-051

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

Free Chlorine Sensor For additional information, please visit our website at

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES

Quick Start Guide LIQ-QSG-499AOZ, Rev M July Rosemount 499AOZ. Dissolved Ozone Sensor

ph measuring system for hygienic process applications

Visit us at Code

Physics Experiment 17 Ideal Gas Law Qualitative Study

Transcription:

BIOPAC WWW.biopac.com Systems, Inc. Application Note PH-185 SS32L Dissolved O 2 Probe SS32L Dissolved O 2 Probe BIOPAC Software SS32L Specifications BSL PRO v. 3.6.6 BIOPAC Hardware SS32L Dissolved O 2 Probe (includes electrolyte solution) Other Hardware Components: Dissolved O 2 Probe Electrolyte Solution Replacement Membrane Cap Replacement O-ring Tool Solution Syringe Compatible fluids: Primarily, the sensor is intended for file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (1 of 9)3/24/2006 3:20:19 PM

Culture tube (such as Pyrex rimless 16x100mm or larger) Zero O 2 solution measuring water, blood, or urine. Note: Calibration procedures remain the same for each substance. Type: galvanic; platinum and silver electrodes Membrane: 2 Mil Output: mv per mg/l min. sample velocity: 20 cm/sec Min. submersion depth? Pressure: Temperature Compensation: Not required if calibration is made at the same temperature as testing. Salinity Compensation: manual Pressure Compensation: manual Oxygen range: 0-40 ppm Accuracy:??? Temp range:??? Response time: Sample size: Minimum 2 ml Sterilized: steam sterilizable Lead content: yes Probe body: cm x cm Weight: grams Cable: 3 meters file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (2 of 9)3/24/2006 3:20:19 PM

Objective: This Application Note will explain the functionality, calibration, recording and storage/care instructions for the BIOPAC Dissolved O 2 Probe (SS32L) The probe can be used to determine the Biological Oxygen Demand (BOD) of organic matter. Abstract: This SS32L Dissolved O 2 Probe is designed to measure the amount of oxygen dissolved in a unit volume of a given liquid. The amount of dissolved oxygen that a liquid contains is influenced by pressure, temperature, and purity. As pressure decreases, DO 2 decreases. As temperature decreases, DO 2 increases. As purity decreases (i.e., salinity increases), DO 2 decreases. A liquid is "saturated" when the pressure of the oxygen in the liquid equals the pressure of the oxygen in the air at a given temperature; water saturation usually occurs at 5-10ppm (particles per million). DO 2 is commonly used as a measurement of water quality and hydro-carbon co-monomer quality. Notes For Taking Good Measurements: A. Allow a sufficient "warm-up" period before use (generally, 5-10 minutes should suffice). B. Keep the liquid moving across the surface of the membrane--stir the probe or stir the liquid--to prevent bubbles from collecting on the membrane tip, which will distort the O 2 reading. C. Allow the probe to "recharge" for 1-2 hours after you change the electrolyte solution or the membrane cap. D. Keep the membrane moist while the probe is connected to the MP30. E. Do not drop, shake or excessively impact the probe. file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (3 of 9)3/24/2006 3:20:19 PM

Setup: 1. First use: Unscrew the probe tip, fill it with electrolyte solution, screw the tip back on. 2. Remove the protective cap and immerse the probe in water for 2-3 hours before first use. 3. Plug the SS32L cable into a channel on the MP unit, and fasten the screws on either side of the attachment. 4. Immerse the probe in Zero O 2 Solution. 5. Turn on the MP unit. 6. Launch the BSL PRO software. 7. Wait for about 10 minutes as a probe "warm up" period. Procedure: You must repeat the warm up period any time the probe is disconnected from the MP30 unit. Calibration: 1. In BSL PRO click MP30>Setup Channels. 2. In the Setup Channels dialog, activate the three check boxes next to the "CH1" label by clicking inside them file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (4 of 9)3/24/2006 3:20:19 PM

3. In that same row, click the "View/Change Parameters" wrench icon to generate the Input Channel Parameters dialog. 4. Click the "Scaling" button. 5. In the Change Scaling Parameters dialog, label the units "mg/l" for milligrams per liter or "ppm" for particles per million (equal measures). To use units of % saturation, click here. Cal 1: 6. Stir the probe and Click Cal 1. Calibration values should fall in the range of 0.2-0.5 V. If your values vary, move the probe to release a possible bubble on the tip. file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (5 of 9)3/24/2006 3:20:19 PM

Keep the liquid moving across the membrane surface to prevent air bubbles. 7. Set Cal 1 Scale value to 0. 8. Rinse the probe tip and blot it dry. 9. Immerse the probe tip in air-saturated water. 10. Wait one minute, then stir the probe and Click Cal 2. 11. Click "OK" and close out of any remaining dialog boxes. 12. Rinse the probe tip and blot it dry. Cal 2 13. Immerse the probe in a calibration standard. 14. Click Cal 2. 15. Set Cal 2 Scale Value to known value --> see chart for lab temp/pressure adjustment. 16. Label the units %. 17. Click "OK" and close out of any remaining dialog boxes. 18. Rinse the probe tip and blot it dry. Experiment: Dissolved O 2 : 1. Immerse the probe in the liquid sample. 2. Click on the Start button in the PRO software. 3. Stir the probe for one minute or more. file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (6 of 9)3/24/2006 3:20:19 PM

4. Click on the Stop button in the PRO software. 5. Rinse the probe drip and blot it dry. BOD Option: 1. Collect two water samples. 2. Measure the dissolved O 2 (ppm) of one sample. 3. Store the other sample at 20 C/68 F for 5 days in complete darkness. Use an incubator or wrap the sample bottle in aluminum foil or black electrical tape and store it in a dark place at room temperature (20 C/ 68 F). 4. After 5 days, measure the dissolved O 2 (ppm) of the stored sample. 5. Subtract the Day 5 reading from the Day 1 reading to determine the BOD level (ppm). Analysis: Total volume of dissolved O 2 file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (7 of 9)3/24/2006 3:20:19 PM

1. Open your data file. 2. Set a measurement for "value." 3. Click the cursor on a point and review the result. If the signal value varies widely, measure the peak two or three times and average your results. Care of the SS32L Cleaning & Storage Rinse the dissolved O 2 probe with tap water after use and before storage. Use a bristle brush to remove deposits as necessary. Periodically replace the membrane cap. Periodically replace the electrolyte solution. For short-term storage (<24 hours), leave the probe immersed in water. For long-term storage (>24 hours) unscrew the probe tip and empty the electrolyte solution, then rinse, dry and replace the tip. Unscrew the tip to fill/empty the electrolyte solution Maintenance Replacing the Membrane Cap and/or O-ring 1. Unscrew the tip. 2. Align the TOOL against the membrane cap and push it through the tip to release the cap. file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (8 of 9)3/24/2006 3:20:19 PM

3. As necessary, remove and reseat the O-ring from the tip. O-ring at end of tip 4. Insert the new membrane cap (membrane down) into the tip. 5. Hold the probe and use the TOOL to push the cap securely into the tip. Hold the probe to prevent pressure against the membrane surface, such as might occur if you pushed down on the cap while it was resting on a surface. 6. Use the syringe to fill the tip about 2/3 full (??ml) of electrolyte solution, tap it to release any bubbles, and then screw it back on. 7. Wait 2-24 hours before using the probe. Return To Application Note Menu file:////ntserver4/doc%20pdf/appnotes/app185/ph185dissolvedo2.htm (9 of 9)3/24/2006 3:20:19 PM