Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Similar documents
Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

6890 GC Site Preparation

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Please refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

7820A GC Site Preparation Checklist

Customer Responsibilities. Important Customer Information Series Q-TOF LC/MS Systems Site Preparation Checklist

490 Micro GC Site Preparation Checklist

Customer Responsibilities

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Agilent GC, GC/MS, and ALS

Customer Responsibilities

Customer Responsibilities. Important Customer Information. Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

be present throughout these services; otherwise, they will miss important operational, maintenance and safety information.

28 Site Preparation. Supplying valve actuator air

Agilent 5975T LTM GC/MSD

Customer Responsibilities. Important Customer Information. Agilent Infinity II Preparative HPLC System Site Preparation Checklist

490 Micro GC Site Preparation Checklist

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

Varian 240-MS GC/MS Ion Trap Mass Spectrometer Pre-installation Instructions

Agilent 7697A Headspace Sampler

Customer Responsibilities. Important Customer Information. Cary 4000/5000/6000i UV-Vis spectrophotometer Site Preparation Checklist

Agilent 7667A. Mini Thermal Desorber. Site Preparation

Mass Spec will not Autotune

Micro GC Fusion Pre-Installation Checlist

Pre-installation Checklist

UltraLo 1800 Alpha Particle Counter

Customer Responsibilities

Best Practice for Identifying Leaks in GC and GC/MS Systems

Agilent G3180B Two-Way Splitter Kit

Varian 325-MS LC/MS Quadrupole Mass Spectrometer Pre-installation Instructions

Customer Responsibilities

Agilent G Rebuilt Electron Capture Detector

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

6890 GC Not Ready Messages (A16010)

3 Flow and Pressure Control

Customer Responsibilities

Agilent 7697A Headspace Sampler

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Your local gas generation partner. Precision series Modular gas generation solution for GC.

CP-3800 GC Getting Started Manual

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

The Open-Split Capillary Interface. Part No

STANDARD OPERATING PROCEDURES

Agilent Intuvo 9000 Gas Chromatograph

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Preparing Your Laboratory for the Optima 7300 V and 8300 ICP-OES Spectrometers

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Discovery HP-TGA 75/750. Site Preparation Guide

Avio 500 ICP-OES. ICP-Optical Emission Spectroscopy PREPARING YOUR LAB. Preparation Checklist. Suitable Working Area

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

User Manual. Agilent 1260 Infinity Analytical- and Preparative-scale Fraction Collectors

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Your local gas generation partner. Precision series Modular gas generation solution for GC.

no peaks missing peaks too small peaks increasing retention decreasing retention

INSTALLATION GUIDE Gas Rangetops

DD2 Site Preparation Checklist Hardware Site Preparation Specification

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Workhorse-15 Oxygen Generator

Superior Result from Analytical Instruments The constant purity of the nitrogen improves system stability and ensures reproducible results

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

CSA Sample Draw Aspirator Adapter Operator s Manual

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Super-Clean Gas Filters from Restek

Discovery SDT 650. Site Preparation Guide. Revision G Issued December Page 1

Oxford Instruments Plasma Technology. PlasmaPro NGP80. Installation Data Sheet. Original Instructions.

If you want to offer a more professional look to your

Mass Spec will not pump down

Columbus Instruments

5890II GC Standard Operating Procedure 9/2/2005

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

Operating Instructions

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

Workhorse 8 Oxygen System Model No 5665, 5673, 5679

Agilent 1220 Infinity II LC Mobile Upgrade Kit

INSTRUMENT INSTRUMENT YL INSTRUMENT

Carrier Gases in Capillary GC

Agilent G3185B QuickSwap Accessory

Agilent Auxiliary Gas Module AGM 2. User s Guide

Tank Blanketing Pressure Regulators RHPS Series

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

Multiple Gas#5 GC configuration Jan 2016

Instant Connect Helium Saver Injector Module

Zebron Gas Management

ATIS OPERATION MANUAL

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Preinstallation Checklist and Instructions for the TriStar These Preinstallation Checklist and Instructions were reviewed and approved by:

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Thermo Scientific. TRACE GC Ultra. Gas Chromatograph Site Preparation and Installation Manual. PN , Revision May 2010

NexION 1000/2000 ICP-MS

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Transcription:

Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 5975C Series Mass Selective Detector (MSD). This checklist is designed to prevent delays during installation, familiarization, and the initial use of the MSD system in your application. This checklist outlines the space and utility requirements for a 5975C MSD. It also recommends tools and consumables that may help you get started. Use it along with the 5975C Site Preparation Manual. Note: The 5975C VL MSD Series refers to the 5975C MSD (G3170A); 5975C inert MSD (G3171A); 5975C inert XL (G3172A); and 5975C inert XL EI/CI MSD (G3174A). Customer Responsibilities Make sure your site meets this specification, including: the necessary space, electric outlets, gases, tubing, operating supplies, consumables and other usage dependent items such as columns, vials, syringes and solvents required for the successful installation of instruments and systems. If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Important Information If you need assistance, please contact your local Agilent Technologies office. Assistance with this checklist and with user specific applications is available and will be contracted separately. Dimensions and Weight Select the laboratory bench space before your system arrives. Pay special attention to the total height requirements. Avoid bench space with overhanging shelves. Allow at least 20 cm clearance between back of GC and wall to dissipate air. 5975C Series MSDs Diffusion Pump Standard Turbo Pump Performance Turbo Pump Performance CI/EI Turbo Pump Height Width Depth Weight 40.8 cm 40.8 cm 40.6 cm 40.6 cm 39 kg 85 lb 39 kg 85 lb 41 kg 90 lb 46 kg 100 lb GCs Series Height Width Depth Weight 6890/7890 21.3 in 54.5 cm 21.4 in 21.3 in 50 kg 112 lb 6850 50.0 cm 19.7 in 29 cm 11.4 in 21.3 in 23 kg 51 lb Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 1 of 7

A simple system that includes a GC, an automatic liquid sampler, and a computer would require about 153 cm or 5 feet of bench space, add an addition 41cm for a LaserJet printer (195cm or about 6.5 feet); access to the MSD could require moving the instrument so an additional (30cm or 1 foot) should be available. A total of 244 cm or 8 feet of bench space should be available for a full GCMS system. Some repairs to the MSD or to the GC will require access to the back of the instrument. Please note: the length of the vacuum hose is 130cm or about 4.24 feet from the high vacuum pump to the foreline pump, while the length of the foreline pump power cord is 2M or about 6.6 feet. The use of a Dry foreline pump requires the exhaust to be plumbed to an exhaust hood or exhaust line. During the installation and familiarization it is okay to use the silencer, but while running samples other than the installation checkout sample it is recommended the exhaust of the pump goes to an exhaust hood or exhaust line. Power Consumption The number and type of electrical outlets depends on the size and complexity of your system. The MSD power consumption and requirements depends on the country the unit is shipping to. The electrical outlet for the unit should have a dedicated ground. Product Line voltage Frequency Current Rating (amps) Maximum continuous power consumption Outlets Requried 5975C Series 120VAC 50/60 Hz ± 5% 15A 1100VA (400VA for 1 MSD foreline pump only) 220-240VAC 50/60 Hz ± 5% 15A 1100VA (400VA for 1 foreline pump only) 200VAC 50/60 Hz ± 5% 15A 1100VA (400VA for 1 foreline pump only) ChemStation PC 120VAC 50/60 Hz ± 5% 15A 1000VA 3 to 5 system (monitor, CPU, printer) 200-240VAC 50/60 Hz ± 5% 15A 1000VA 3 to 5 6890/ Standard Americas:120V AC (1) 48-66 Hz 18.8 2250 VA 1 7890 single phase 220/230/240V single/split 48-66 Hz 10.2 / 9.8 / 9.4 2250 VA 1 phase Fast Japan: 200V split phase 48-66 Hz 14.8 2950 VA 1 220/230/240V (2)(3) single/split phase 48-66 Hz 13.4 / 12.8 / 12.3 2950 VA 1 6850 Standard Americas:120V AC (1) 48-66 Hz 1 single phase 220/230/240V single/split 48-66 Hz 1 phase Fast Japan: 200V split phase 48-66 Hz 1 Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 2 of 7

220/230/240V (2)(3) single/split phase 48-66 Hz 1 Notes 1. Americas 120V requires 20 amp dedicated line. Americas 240V requires 15 amp dedicated line. 2. Option 003, 208V fast oven, uses a 220V unit with operating range of 193 to 231V. Most labs have 4-wire service resulting in 208V at the wall receptacle. It is important to measure the line voltage at the receptacle for the GC. 3. Power line conditioners should not be used with 6890/7890 GCs. Heat Dissipation The following table may help you calculate the additional BTU s of heat dissipation from this new equipment. Maximums represent the heat given off when heated zones are set for maximum temperatures. Oven type 6890 series 5975C series Standard oven ramp 7681 BTU / hour maximum 3000 BTU / hr including GC/MSD interface Fast oven ramp (options 002 and 003) 10,071 BTU / hour maximum 3000 BTU / hr including GC/MSD interface Oven type 6850 series 5973 series Standard oven ramp 3000 BTU / hr including GC/MSD interface Fast oven ramp 3000 BTU / hr including GC/MSD interface Environmental Conditions Operating the GCMS within the recommended ranges insures optimum instrument performance and lifetime. Instrument needs space for proper convection of heat and ventilation. Performance can be affected by sources of heat and cold from heating, air conditioning systems, or drafts. Product Conditions Operating temp range Operating humidity range Maximum altitude 6890/7890 Series Standard oven ramp 20 27 C 50 60% 4,615.38 m Fast oven ramp (options 002 and 003) 20 27 C 50 60% 4,615.38 m Storage 5 to 40 C 5% - 95% 6850 Series Standard oven ramp 20 27 C 50 60% 4,615.38 m Fast oven ramp (options 002 and 003) 20 27 C 50 60% 4,615.38 m Storage 5 to 40 C 5% - 95% Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 3 of 7

5975C Series Operation 15 to 35 C (59 to 95 F ) Storage -20 to 70 C (-4 to 158 F ) 20% - 80% 4,615.38 m * 0% - 95% * 7,500 feet (2.3 Km) for the 5975C VL MSD Gas Supply Gases are supplied by tanks, internal distribution system, or gas generators. Tank supplies require two staged, pressure regulation. To connect tubing to the supply, it must have one 1/8-inch Swagelok female connector for each gas. Make sure that your regulator has the appropriate sized adapter to end with a1/8-inch Swagelok female connector. (The URL of SwagLok s web site is http://www.swagelock.com to help assist is finding connectors.) The following table lists minimum and maximum pressures in psi for inlets and detectors measured at the bulkhead fitting at the back of the 7890, 6890 and 6850 Series GCs. 6890 Series Inlets and Detectors FID NPD TCD ECD FPD S/splitless 150 psi Hydrogen 35-100 35-100 45-100 Air 55-100 55-100 100-120 Make up 55-100 55-100 55-100 55-100 55-100 Reference 55-100 S/splitless 100 psi Oncolumn Purged packed Carrier max 170 120 120 120 120 Carrier min 20 psi above pressure used in method Conversions: 1 psi = 6.8947 kpa = 0.068947 Bar = 0.068 ATM Notes 1. If you have not requested option 305, you must supply pre-cleaned, 1/8-inch copper tubing and a variety of 1/8-inch Swagelok fittings to connect the GC to inlet and detector gas supplies. 2. Cyrogenic cooling with Liquid N2 requires 1/4-inch insulated copper tubing. 3. Cyrogenic cooling with Liquid CO2 requires 1/8-inch heavy-walled, stainless steel tubing. 4. Valve actuation requires a separate pressurized, dry air at 55 psi. 5. Never use liquid thread sealer to connect fittings. Never use chlorinated solvents to clean tubing or fittings. PTV 5975C Series Gas Flow Limitations Feature G3170A G3171A G3172A G3174A High vac pump Diffusion Pump Standard turbo Performance turbo Performance turbo, EI/PCI/NCI Optimal gas flow 1.0 1.0 1.0 to 2.0 1.0 to 2.0 ml/min (a) Maximum 2.0 2.0 4.0 4.0 recommended gas flow, ml/min Maximum gas flow, ml/min (b) 2.4 2.4 6.5 4.0 Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 4 of 7

Max column id 0.32mm (30m) 0.32mm (30m) 0.53mm (30m) 0.53mm (30m) a Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) b Expect degradation of spectral performance and sensitivity 5975C Series Carrier and Reagent Gases Carrier and reagent gas requirements Typical pressure range (psi) Typical flow (ml/min) (required) 50 to 80 20 to 50 (column and split flow) Hydrogen (optional) (a) 50 to 80 20 to 50 (column and split flow) Methane reagent gas 15 to 25 1 to 2 (required for CI operation) Isobutane reagent gas (optional) 15 to 25 1 to 2 Ammonia reagent gas (optional) 5 to 8 1 to 2 Carbon dioxide reagent gas (optional) 15 to 20 1 to 2 a Hydrogen gas can be used for the carrier gas but specifications are based on as the carrier gas and please observe all hydrogen gas safety cautions. Gas Selection Agilent recommends that carrier and detector gases be 99.9995% pure. Air needs to be zero grade or better. Agilent also recommends using traps to remove hydrocarbons, water, and oxygen. The following table lists gases for capillary columns. ECD FID FPD NPD TCD Carrier Preferred makeup 2 nd choice Detector, anode purge, or reference Hydrogen Argon/methane Anode purge must be same as makeup Argon/methane Argon/methane Argon/methane Argon/methane Hydrogen Hydrogen Argon Hydrogen Must be same as carrier and reference Must be same as carrier and reference Hydrogen and air for detector Hydrogen and air for detector Hydrogen and air for detector Reference must be same as carrier and makeup 5975C Series Carrier and Reagent Gases Purity Carrier and reagent gas requirements Purity (Carrier) 99.9995% hydrocarbon free Hydrogen (Carrier) 99.9995% SFC Grade Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 5 of 7

Methane reagent gas 99.999% Research or SFC grade (required for CI operation) Isobutane reagent gas (optional) 99.99% Instrument grade Ammonia reagent gas (optional) 99.9995% Research or SFC grade Carbon dioxide reagent gas (optional) 99.995% SFC Grade For both the GC and MSD it is recommend two (2) step regulators be used with 1/8 size outlets. Other considerations Exhaust Venting Requirements for the GCMS For the MS vent external to building via ambient-pressure vent system, within 460 cm (15 ft) of both GC split vent and MSD foreline pump or vent to fume hood. Exhaust vent system is not part of environmental control system of building that recirculates air. Exhaust venting need to comply with all local environmental and safety codes. For the GC with the deflector (outlet diameter 10cm 4in) installed the exhaust is about 65 CuFt/min (1.840CuM/min), without deflector 99 CuFeet/min (2.8M3) Basic Tools Your GCMS comes with a few basic tools and consumables depending on the specific inlet and detector that you ordered. Here is a general list which one will get with the instruments or should have on-hand. Tool or consumable Used for Inlet wrench Replacing inlet septa and liners. T10 and T20 Torx wrenches Remove tray. Remove covers to access EPC modules, traps, and possible leaks. ¼-inch nut driver FID jet replacement. FID flow measuring insert FID troubleshooting. Column cutter Column installation. 1/8-inch Tee, Swagelok, brass Connect gas supplies 1/8-inch nuts & ferrules, Swagelok, brass Connect gas supplies Inlet septa appropriate for type Injection port seal Inlet insert or liner Injection port 1.5 mm and 2.0 mm hex driver Source maintenance (disassembly) Tool bag Used to hold GC and MS tools Q-Tips Used to clean source parts Cloths Used to keep surfaces clean and parts clean Gloves Used to reduce contamination on parts GC and MS MSD Maintenance supplies Description Part number Abrasive paper, 30 µm 5061-5896 Alumina powder 1kg sample 8660-0791 Cloths, clean (package of 300) 05980-60051 Cloths, cleaning (package of 300 9310-4828 Cotton swabs (package of 100) 5080-5400 Foreline pump oil, P3 6040-0621 Gloves, clean, large 8650-0030 Gloves, clean, small 8650-0029 Grease, Apiezon L, high vacuum 6040-0289 Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 6 of 7

Ferrules Description Part number Blank, graphite-vespel 5181-3308 GC/MSD interface 0.3-mm id, 85% Vespel 15% graphite, for 0.10-mm id columns 5062-3507 0.4-mm id, 85% Vespel 15% graphite, for 0.20-mm id and 0.25-mm id columns 5062-3508 0.5-mm id, 85% Vespel 15% graphite, for 0.32-mm id columns 5062-3506 0.8-mm id, 85% Vespel 15% graphite, for 0.53-mm id columns 5062-3538 Injection port 0.27-mm id, 90% Vespel 10% graphite, for 0.10-mm id columns 5062-3518 0.37-mm id, 90% Vespel 10% graphite, for 0.20-mm id columns 5062-3516 0.40-mm id, 90% Vespel 10% graphite, for 0.25-mm id columns 5181-3323 0.47-mm id, 90% Vespel 10% graphite, for 0.32-mm id columns 5062-3514 0.74-mm id, 90% Vespel 10% graphite, for 0.53-mm id columns 5062-3512 Miscellaneous parts and samples Electron multiplier horn 05971-80103 Electron multiplier horn for the Triple Axis Detector G3170-80103 Filament assembly (EI) G2590-60053 Filament assembly (CI) G1099-80053 Foreline pump oil (1 liter) 6040-0621 Foreline exhaust oil mist trap G1099-80039 Octafluoronapthalene (OFN), 1 pg/ul 5188-5348 Octafluoronapthalene (OFN), 100 fg/ul 5188-5347 Perfluorotributylamine (PFTBA), certified (10 gram) 8500-0656 Perfluorotributylamine (PFTBA) sample kit 05971-60571 PFDTD, CI Calibrant 8500-8510 PFHT 5188-5357 Sample, evaluation A, hydrocarbons 05970-60045 Updated May 2010 PN# G3170-90183 LSCA GCMS Technical Marketing Page 7 of 7