5890II GC Standard Operating Procedure 9/2/2005

Similar documents
System Overview. TCD Detector temperature setpoint Regulator

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Operation of the Perkin Elmer TGA-GC/MS

This GC has two columns (I think new columns can be ordered from Restek)

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

AKTA pure 25 New Owner s Intro

Armfield Distillation Column Operation Guidelines

6890 GC Site Preparation

GCMSD-Headspace Analysis SOP

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

SOP: Buck Scientific BLC-20P HPLC Operation

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

6850 Series Control Module User Information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Using the Akta Prime plus October 22, 2012

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

6890 GC Not Ready Messages (A16010)

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

3 Flow and Pressure Control

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

SOP for use of the SX19 Stopped-Flow Instrument

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

CP-3800 GC Operator s Manual

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Figure 1: Plug in ion inlet of mass spec

FROG-4000 TM Chemical Analysis System

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

FROG-5000TM Chemical Analysis System User s Manual

FROG-4000 TM Chemical Analysis System

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

Multiple Gas#5 GC configuration Jan 2016

Cover Page for Lab Report Group Portion. Head Losses in Pipes

AX5000 Operational Manual

Experiment GC : Analysis of BTEX by GC-FID

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Quick Reference Guide. OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

Prep LC Training Manual S.V.

A New GC/MS System Designed for Helium Carrier Gas Conservation

A SCENT OF FUTURE. Master DHS. Dynamic Head Space Sampler. Operating Manual

SHIMADZU LC-10/20 PUMP

University of Minnesota Nano Center Standard Operating Procedure

Cover Page for Lab Report Group Portion. Flow Visualization in a Water Channel

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Rapid and Reliable Detection of Dissolved Gases in Water

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Operating Instructions

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

no peaks missing peaks too small peaks increasing retention decreasing retention

Manual for continuous distillation

Agilent 7820A GC. Site Preparation Guide. Agilent Technologies

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

FIG: 27.1 Tool String

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

CE 428 CATALYTIC ISOMERIZATION OF BUTENE

Vapor Pressure of Liquids

Agilent G3185B QuickSwap Accessory

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

Technology. Using Bluetooth

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

The MRL Furnaces USED FOR THIS MANUAL COVERS

MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL

Discovery HP-TGA 75/750. Site Preparation Guide

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Cover Page for Lab Report Group Portion. Pump Performance

PC Configuration software for Discovery MkVI v 1.03 User guide

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

SOP: Cation Exchange Chromatography of t-pa using AKTA Pure system

CLEANLINESS STANDARD NCPRE tool contamination classification

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Gas Absorption Draft Standard Operating Procedure

Your local gas generation partner. Precision series Modular gas generation solution for GC.

ELE INTERNATIONAL COMPRESSION TESTER Standard Operating Procedure

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Evaluation copy. Vapor Pressure of Liquids. computer OBJECTIVES MATERIALS

IVIS Spectrum Whole Animal Imager

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

CP-3800 GC Getting Started Manual

Experiment HE-9: Resting, Active, and Exercising Metabolic Rates

The. «Modular Alliance» Gas Generators. ALLIANCE, the partnership that keeps the gas flowing. One step ahead!

The Gas Laws: Boyle's Law and Charles Law

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Transcription:

5890II GC Standard Operating Procedure 9/2/2005 This procedure is for the analysis of natural gas using the HP 5890II GC with the Restek ShinCarbon column. Detector Type: TCD Column Type: Restek Micropacked, ShinCarbon ST 100/120 1 meter, 0.75mm ID, 0.95mmOD, Silco Max Temp 330 degrees C Injector type: Packed purged inlet. Injector used: B - Injector A is broken Detector used: A TCD Gasses used: Helium Helium is the carrier gas; this is the gas that carries the injected material through the column. Helium is also used to provide Makeup gas, also known as Aux gas. Makeup gas is added at the end of the column to increase the gas flow entering the detector. Reference gas is helium. Reference gas is compared with the sample at the detector. 1. Turn on the power The power switch for the GC is on the lower right side towards the back of the GC. 2. Open Gas Valve: Do not adjust the cylinder regulator valves! Open the Helium main valve on the tank labeled 5890 one turn counterclockwise. Some gasses have a small valve on the outlet side of the regulator this valve should be open. Helium Tank Pressure U> 500U Outlet Pressure U45 60 PSIU If the pressure readings are out of range consult the lab manager. Do Not Adjust! Make sure the (He) toggle switch located to the left of the GC on the HP 5890 II GC gas switch panel is set to the horizontal position.

3. Set gas flow. Gas flow rates should be preset. Consult lab personnel if you have questions about gas flow. DO NOT ADJUST GAS FLOW. Column flow = 8 10 ml/min. Aux gas + column flow = 9 15 ml/min. Measure as combined flow. Set Reference Gas: Reference Gas should be 27 45 ml/min. (Column flow + Aux Gas + Reference Gas) = 36 60 ml/min. 4. Measure column gas flow: Verify Helium Flow. Do not adjust flow rates. Verify Helium flow through the column by measuring the flow at detector A using the gas flow meter or a glass bubble type flow meter. Caution! Detector is HOT. Column flow which is helium flow rate with REF AND AUX GAS OFF should be between 8 and 10 ml/min. The recommend starting value is 10ml/min. 5. Open AUX gas valve for TCD detector (A) 2 to 3 turns. Valve is located on upper left hand area of GC. Verify reference gas valve located below AUX valve is off. Measure combined flow of column flow and aux gas. Flow should be within 9 15 ml/min. 6. SET REFERENCE GAS: Open reference (REF) gas valve. Reference gas should be 20 to 50 % higher than column flow, approximately 27 45 ml/min. Measure this as a percentage of total flow. Total flow with Column Gas + Aux Gas + Reference Gas = 36 60 ml/min. 7. Once gas flow is set turn on TCD detector. DET, ON. 8. Start Software: a. Log on to computer (This Computer), NOT CHE - using the user log on, GC analysis: Password = GCuser b. Open Chrom Perfect software from the desktop. c. Select Acquisition to acquire data. d. Under the (Control) section claim the instrument you wish to use. Note the detector type for each instrument is indicated under the (instrument name) heading. e. Locate the (Status) and (Selection) tabs on the right hand side of the screen. Select (Status) and click on the (Control) icon to bring up the instrument control panel.

f. From the instrument control panel select (Download). This opens the Acquisition Parameters window. You can select the Method file in this window. Edit the base name as desired. Use the NatGasB.met for the 5890 II, TCD detector running natural gas, propane or other flammable gases. Select the (Tutorial.MET) for injecting hydrocarbons using the 5890A, FID detector. g. The 5890II (TCD) will automatically start the Chrom Perfect software when the start button on the GC is depressed. The 5890A requires the user to start the software using the (Start Run) icon on the instrument control window. When running the 5890A start the software immediately injecting your sample. h. Once the analysis is running the plot display will appear. You can edit the scale and run time or stop the run by right clicking within the plot and selecting the desired parameter. i. When the run is complete go to the desktop and select (Analysis) from the main Chrom Perfect menu. The data is stored is C:\CPData\GC Student Data\ (description of GC & detector). All files have the.raw extension. j. Select the icon to the right of the drop down icon. This will display a list of files. The most recent file is at the top of the list. Open your file and edit as needed. Select (Integration), click on (Report) to generate a report of your chromatograph. Temperature: Temperature will affect gas flow. Bring injector, oven and detector to set point temperatures before measuring gas flows. Temperature set points will depend on the type of analysis you are doing. See supplemental information for further guidelines on temperatures. Recommended starting temperatures for volatile gases are list below. Detector = < Max column temperature (Recommended temp = 100degrees C) Injector = 50 120 degrees C Oven = 200-300 degrees C IF NOT READY LIGHT IS ON: SET DETECTOR B TO OFF, SET BOTH SIGNALS TO DETECTOR A. The not ready light will go off when the heaters have reached their set point and the TCD detector is on. Once flow of carrier gas through the column is confirmed turn on the 3 heaters for the GC. The controls for all the heaters are located on the upper two rows of the right front control panel. Press INJ A TEMP The digital display shows the temperature of injector A, it is normally about room temperature. To the right of the temperature reading is the on/off indication. Be sure the injector and detector you are using is set to on. First press the ON button on the keypad then if the set point is not the one you want you may press the numbers for the desired set point then ENTER to change it.

The second heater is for the oven. Press the OVEN TEMP button then press the ON button. If you need to adjust the set point do it in the same manner as the injector heater. The third heater is for the detector. You will be using detector B on this GC. Press the DET B TEMP button then press the on button. You may adjust this in the same manner as the other two heaters. If the GC has not been used for a few days it is good practice to set the oven temperature at 300 degrees C and let the column cook for 10 15 minutes. Set Temperature Ramp: For this lab the recommended range for initial temp is 40 to 120 P PC. To set up a program that is not isothermal press the INIT TIME button and enter the time you would like the oven to remain at the initial temperature. For this experiment start with a value of 2 minutes. To set the rate of the ramp press the RATE button and enter the rate. The recommended starting value is 12 degrees per minute. Set the FINAL VALUE button to set the desired final temperature. The recommended final temperature is 300 degrees C. Set the amount of time you want to oven to remain at the final temperature. FINAL TIME, enter value for final time. Recommended final time = 10mins. o Injection Guidelines Wear safety glasses. Use a gas syringe not a liquid syringe. Hold the needle above injection port A. Don t touch the port since it may be hot. Keep your thumb above the plunger in case the pressure pushes the plunger up. Then push the entire needle straight down through the plunger until you can t go any further. Then push the plunger down and pull the whole syringe out. Immediately press the start button on the GC and the program will begin. SHUTDOWN Turn off the oven heater. Turn off the injector heater. Turn off the detector heater.. Press the DET button and turn the detector off. Allow the GC oven to cool to room temperature, then turn the Helium gas off at the tank. If the GC is not cool Oxygen will get in the column and destroy the inner coating of the column. If the GC is not cool leave the helium on. Turn OFF GC.

Chrome Perfect Tips: Data for the HP 5890 II is stored in C:\CPData\GC Student Data\5890II TCD Data Data for the HP 5890A is stored in C:\CPData\GC Student Data\ 5890A FID Data Data for the GOW-MAC is stored in C:\CPData\GC Student Data\GOW MAC Data To zoom in on selected peaks right click and select peaks of interest then left click to view. To return to normal view left click and select undo expansion. To remove baseline noise select integration Auto then enter a number greater than zero in the threshold window. Other options available by selecting integration manual.