SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Similar documents
SRI Multi Gas Analyzer 2016

Multiple Gas#5 GC configuration Jan 2016

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

System Overview. TCD Detector temperature setpoint Regulator

Retention Time Locking: Concepts and Applications. Application

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Agilent 490 Micro GC Natural Gas Analyzer

This GC has two columns (I think new columns can be ordered from Restek)

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

New improved robot for gas kinetics

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

ISO INTERNATIONAL STANDARD

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Buck Scientific, Inc.

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Carrier Gases in Capillary GC

Rapid and Reliable Detection of Dissolved Gases in Water

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Carrier Gases in Capillary GC

Applied Technology and Best Practices in CEE. Conference

5890II GC Standard Operating Procedure 9/2/2005

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

VERIFYING GAS CHROMATOGRAPH OPERATION AT CUSTODY TRANSFER LOCATIONS Murray Fraser Daniel Measurement & Control

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Your local gas generation partner. Precision series Modular gas generation solution for GC.

6890 GC Site Preparation

REFINERY GAS ANALYSIS BY GAS CHROMATOGRAPHY

Quantitative Analysis of Hydrocarbons by Gas Chromatography

Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Your local gas generation partner. Precision series Modular gas generation solution for GC.

AN-03 Rotary valves applications

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Experiment GC : Analysis of BTEX by GC-FID

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

During today's webinar we're going to look at the following topics:

ISO 7941 First edition

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

FUNDAMENTALS OF GAS CHROMATOGRAPHY, GAS QUALITY & TROUBLESHOOTING Larry Ewing Chandler Engineering Company LLC

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Improved Reliability in Natural Gas Energy Measurement with the Revised ISO Standard. Gerard Nieuwenkamp - VSL

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

7820A GC Site Preparation Checklist

GAS CHROMATOGRAPHY MAINTENANCE USING UNCERTAINTY BASED CBM

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Zebron Gas Management

Operating Manual. AC OxyTracer Analyzer on 7890 GC Manual part number version 2011/1.0

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

Gas Leak Measurement of TRT Barrel Module

MODEL 131/132 GAS CHROMATOGRAPH

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

By: Eng. Ahmed Deyab Fares - Mobile:

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

no peaks missing peaks too small peaks increasing retention decreasing retention

Technical Manual. Sensepoint XCD Gas Detector

Automated Determination of Dissolved Gases in Water Anne Jurek

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

GAS CHROMATOGRAPHY AND SAMPLE LET DOWN SYSTEM CONDITIONAL BASED MONITORING AND LIVE UNCERTAINTY CALCULATION

Disclosure to Promote the Right To Information

GCMSD-Headspace Analysis SOP

Best Practice for Identifying Leaks in GC and GC/MS Systems

Mr Roger Wood Measurement & Process Group (M&P) Network Strategy National Grid Gas plc Brick Kiln Street Hinckley Leicestershire LE10 0NA

Agilent G3180B Two-Way Splitter Kit

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

Simple and Worry Free Gas Filters

The. «Modular Alliance» Gas Generators. ALLIANCE, the partnership that keeps the gas flowing. One step ahead!

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

CP-3800 GC Operator s Manual

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

North Sea Flow Measurement Workshop th October 2017

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

28 Site Preparation. Supplying valve actuator air

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

IMPACT OF THE HELIUM SHORTAGE ON PROCESS GAS CHROMATOGRAPHS

PRECISION GAS MIXTURES

Svensk läkemedelsstandard

IRwin Methane Leak Detector. Innovation makes the difference

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

Agilent 7697A Headspace Sampler

Space product assurance

CONNECTIONS. The Thermal Conductivity Detector

Transcription:

The SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration is a versatile low cost way of analyzing many different kinds of gas samples. The GC pictured at right has two Multiple Gas #3 ( MG#3 ) configurations implemented in a single GC chassis so there are two gas sampling valves and four columns as well as four detectors. This is why the column oven looks so crowded. The MG#3 GC configuration is almost identical to the MG#1 GC configuration except there is an additional valve which when activated by the PeakSimple data system stops the flow of carrier gas in column 1. When the valve is actuated ( typically while the gas sampling valve is in the INJECT position ), column 1 has the same pressure applied to both its inlet and outlet. This stops the flow of carrier gas in column 1. The peaks which were in column 1 simply stop moving without broadening or distortion. The flow of carrier gas in column 2 actually increases while the is actuated since the full carrier gas head pressure is now applied across a shorter restriction ( one column instead of two in series ). Column1 Column 2 The MG#3 GC configuration is slightly more flexible than the MG#1 because the stop flow capability allows a wider selection of columns to be used, where the MG#1 only works with silica gel as Column 1 and Mole- Sieve 13X as Column 2. Instructions Page 1

The chromatograms shown on this page are a mix of natural gas and sulfur compounds. The top chromatogram shows the sulfur selective FPD response. The middle chromatogram shows the FID response. The two lower chromatograms show the FPD response ( black ) overlaid with the FID response ( red ). The PeakSimple event table shown at right rotates the valve from Load to Inject at.1 and then back to Load at 1.00. Because even the first peak ( methane ) has not migrated from Column 1 though to Column 2 at this time, the equivalent effect is that the FPD Sulfur only Column 1 60 meter MXT1.53mm 5 mic. Column 2 15 meter RTX Q-Plot.53mm FID Hydrocarbons peaks are injected into and are separated by Column 1 only, as if Column 2 was not even connected. You can see by the overlaid chromatograms that COS co-elutes with Propane quenching its FPD response. FPD and FID overlaid Just this section enlarged Instructions Page 2

Instead, the MG#3 allows the Stop Flow to actuate at 3.5 just after the Propane and COS migrate into Column 2 ( 15meter RTX QPlot.53mm ). FPD Sulfur only This traps the peaks after Propane in Column 1 while the peaks in Column 2 ( Methane, Ethane, Propane, H2S and COS ) separate and elute. Unlike column 1 which does not separate COS and Propane, the peaks are well separated on Column 2 so quenching does not occur. Once Propane elutes from Column 2 ( about 10 ) the valve rotates back to the Load position and the Stop Flow is de-energized. The peaks which were trapped on column 1 now elute to the detectors ( Butanes, Pentanes, Mercaptans etc. ) Oven temperature 40C for 10 then 20C/min to 200C turns off and valve rotates to Load at 10.00, just after propane Instructions Page 3 turns on at 3.5, stopping flow in

The MG#3 GC configuration is also useful with other column combinations. In this example, Column 1 is a 3 HaysepD and Column 2 is a 6 MS13X. The sample is first run on the 3 HaysepD using the event table shown at right. Because the valve is rotated back to the Load position almost immediately after injection Oven temperature 40C for 10 then 20C/min to 200C on at.9 This would be a good time (.9 ) to activate the stop-flow. Just after the Methane migrates onto Column2 but before the CO2 and heavier peaks off (.1 ) the separation occurs as if Column2 was not even connected. ( no hardware changes are required to produce this effect ). There is a convenient gap between Methane and CO2 where it would make sense to activate the stopflow valve to immobilize the CO2 and heavier peaks in Column1 while the H2, O2, N2, Methane and CO peaks elute from Column1. Instructions Page 4

The same sample is injected again using the Event table shown at right. The valve stays in the Load position until 4.00. The Stop-Flow is actuated at.9 ( determined from the chromatogram on the previous page ) and de-activated at 4.00 mintes. This results in H2, O2, N2, CH4 and on at.9 Stop-Flow on at.9 and off at 4.00. Valve back to Load position at 4.00 off CO migrating onto Column2 ( Mole- Sieve13X ) where they separate and elute into the. Once CO elutes ( about 4.00 ), the valve is rotated back to the Load position and the Stop-Flow is de-energized. The concept of immobilizing peaks by stopping the flow is applicable to many situations and many column combinations, not just the two examples presented here. Instructions Page 5

Another example is Natural Gas. Set the Event table up to inject and then immediately rotate the valve back to Load after.1 in the Inject position. This has the effect of performing the analysis as if Column2 was not in the system. on at.9 Zoom in off Column 1 is a 3 Haysep D and Column 2 is a 6 MS13X. The Haysep D does not separate Oxygen and Nitrogen or CO. Set the Stop-Flow time by finding the gap between Methane and CO2, in this case about.9. The gap between Methane and CO2 is a good time to actuate the Stop-Flow Instructions Page 6

With the Event table modified, the Oxygen, Nitrogen and Methane separate on the MS13X. Then the Stop-Flow is de-energized and valve rotated back to Load position ( both at 4.00 ) and the remaining peaks ( Ethane, Propane, Water, Butanes, and Pentanes ) which were immobilized on the on at.9 Zoom in Off at 4.00 Haysep D ( column 2 ) elute normally. The Stop-Flow is actuated at.9 and de-energized at 4.00. The valve is rotated back to Load also at 4.00 Instructions Page 7