Blood Pressure Monitoring: Arterial Line

Similar documents
Blood Sampling: Arterial Lines

Chest Drains. All Covenant Health Intermediate Care Nursery staff. Needle Aspiration

Suctioning: Esophageal/Gastric

HEALTH SERVICES CODE A.6 CATEGORY: RN - SNP

Endotracheal Suctioning: In Line ETT

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

AQUARIUS Continuous Renal Replacement Therapy Competency

CENTRAL IOWA HEALTHCARE Marshalltown, Iowa

NEXUS TKO. Specialty Neonatal/Pediatric. Safety Infusion Sets. Blood Reflux Protection 24/7! Anti-Reflux Technology Provides

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Assisting with Insertion. Care of Intraspinal Catheters

ARTERIAL MONITORING. Final Approval: VP Operations, Chief Nursing Officer Distribution: ICU, PACU POLICY:

Clinician Guide. Baxter Elastomeric Devices. Promoting patient-centric care

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Troubleshooting Guide Aquarius

Contacts. Quick Start Guide

Standards and guidelines for care and management of patients requiring oxygen therapy.

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

MANITOBA RENAL PROGRAM

HAEMOFILTRATION GUIDELINES

The NAMIC Dream Kit. Single Operator Control. Reduce Contrast Waste. Less Force to Inject. Compensator * Manifold. How the Compensator Manifold Works:

Making Life Easier Easypump II

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

Nita Newborn Model 1800 User Manual

TRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS. Hereditary Angioedema

Xcela Hybrid PICC. Clinician Reference Tool

New. The high security connection

Quiz for Module 2: Aquarius Training (Rev. 1) Lesson 1: Introduction to Aquarius

Skills & Scenarios 81

Summary of Cardiac Anesthesia Technician Responsibilities

Introduction. Medley TM Medication Safety System. Alaris Medical Systems

Skills & Scenarios 81

C O L O G Y. New. g u O N. l o t a C a. The high security connection

Digital Display large color coded LED display is easy to read precise measuring shows actual cuff pressure

PRE-TRANSFUSION GUIDELINES

Shared Haemodialysis Care Handbook. Name Hospital number: Shared HD Care Nurse: Date: Machine type: Dialysis Unit:

Patient Guide. Baxter Elastomeric Devices. Promoting patient-centric care

Chapter 40. Fluid, Electrolyte, and Acid Base Balance. Procedures Checklist INTRAVENOUS THERAPY. Procedure 40.1: Starting an Intravenous Infusion

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

Alaris Pump module FAQs

PreludeSYNC. Radial Compression Device INSTRUCTIONS FOR USE

Gastrostomy Tubes. Information for Parents and Carers

Bleeding: Chapter 22 page 650

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

Indications for Use: Caution: Note:

Compatible Baxter IV Sets

Patients Guide for Self Administering Intravenous Antibiotics General points

Instructions For Use. Elastomeric Infusion Device

PUMP WITH FIXED FLOW RATE. Instructions for Use

MyoSure Setup Simplified

Checking a Breathing System

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

Module 2: Aquarius Training

CLINICAL PROCEDURE Policy # Date Introduced Supercedes Policy# Review Data due November 2005 July 2007 Author/s:

ECMO Levitronix CentriMag Protocols Neonatal & Infant

Cardiac Catheterisation Instrumentation. Lancashire & South Cumbria Cardiac Network

DIAGNOSTIC CARDIOLOGY

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

PRODUCT CATALOGUE OF

User s Guide. CO2 & O2 Controller for miniature incubators

HIGH FREQUENCY JET VENTILATION (HFJV): EQUIPMENT PREPRATION

User s Guide. CO2 Controller for miniature incubators

Port - Body in a Box User s Manual

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

Varithena Delivery System Instructions for Use

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

ACQUIDATA FAMILY CALIBRATION PROCEDURE CALIBRATION CHECKING

SEP-1 Additional Notes for Abstraction for Version 5.0b

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

A complete family of Pediatric Perfusion Systems

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Lucid System In-Service Training

TSE Blood Pressure Monitor Invasive series

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

Section 2: Using the Invasive Blood Pressure Module - An Introduction

Recommendations on Checking Anaesthesia Delivery Systems

OPERATOR S MANUAL. Pediatric/Infant cap-one Mask YG-232T/YG-242T A. General. Safety Information WARNING WARNING WARNING. Components WARNING

Tourniquet Systems Pressure Infusion

UMBILICAL CORD MODEL FOR UMBILICAL VEIN CATHETERIZATION TRAINING

Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.

INTERVENTIONAL CARDIOLOGY

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

Alfred ECMO Circuit Priming and Storage Guide

PREPARATION AND SELF-ADMINISTRATION INSTRUCTIONS FOR PATIENTS

County of Santa Clara Emergency Medical Services System

HAEMODYNAMIC MONITORING SETUP IN THE CATHETERISATION LABORATORY MAHAT SAPION

PAIN RELIEF SYSTEM PATIENT GUIDELINES

HEALTHCARE PROFESSIONAL GUIDE. Elastomeric Products

CADD-Micro. Model 5900 Operator s Manual. Note: Colors are Black and Cyan (as second color) Deltec. Ambulatory Infusion Pump

Infant IV Leg LF03636U Instruction Manual

MIC *, MIC-KEY * GASTRIC-JEJUNAL (GJ) FEEDING TUBE PATIENT USE & CARE GUIDE

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Newborn Nursing Skills and ALS Simulator LF01400U Instruction Manual

6900 Maintenance Instruction System Flush

TOURNIQUET SYSTEMS & PRESSURE INFUSION

Transcription:

Approved by: Blood Pressure Monitoring: Arterial Line Gail Cameron Senior Director, Operations, Maternal, Neonatal & Child Health Programs Dr. Ensenat Medical Director, Neonatology Neonatal Nursery Policy & Procedures Manual : Date Effective Next Review May, 2018 Dr. Sharif Shaik Medical Director, Neonatology Purpose Policy Statement Equipment To ensure safe management of arterial lines and allow frequent arterial blood sampling. To ensure safe management of arterial lines, continuous blood pressure monitoring and allow for frequent arterial blood sampling Accurate blood pressure monitoring is essential for the optimal management of the neonatal patient. Blood pressure is a physiologic parameter that aids in evaluation of clinical stability. Recognition and prompt treatment of blood pressure instability is vital in the Neonatal Nursery. Blood pressure may be measured using noninvasive or invasive methods. Noninvasive measurements are obtained using oscillatory arterial blood pressure measurement, whereas invasive blood pressure monitoring may be obtained via peripheral or central line arterial access A monitoring device also provides a closed blood sampling system, allowing continuous blood pressure monitoring, minimizing blood loss with blood sampling and decreasing the risk for nosocomial infections. Arterial lines may present a risk for acute blood loss, therefore a monitoring device will be utilized to detect and prevent potential concerns relating to damaged lines, misconnections, accidental dislodgement and occlusions Heparinized intravenous solution as ordered Anti-reflux tubing 10 cc syringe Line clamp Disposable Vamp Jr. transducer set Edwards blood pressure cable Procedure ACTION RATIONALE 1. Perform hand hygiene, gather equipment, and don gloves. Maintain no touch technique for equipment in contact with IV solution. 2. Prepare anti-reflux tubing. Attach prepared IV bag to anti-reflux tubing. Fill set with solution removing all visible air. Standard precautions. Systemic infection may occur as a result of poor aseptic technique. IV solution is ordered. Refer to Heparin Use Policy for recommended guidelines for arterial line infusions. 3. Attach anti-reflux set to IV line port of Vamp Jr. set. Keep anti-reflux set open to allow for fluid movement during transducer set up.

Page 2 of 7 4. Open reservoir plunger to approximately ½ ml. Facilitates the flow of priming solution. 5. Open shut-off valves (parallel to tubing), and hold sampling site above the reservoir at 45 angle. Allows air to rise and to be removed during the priming procedure. 6. Prime transducer set. Depress syringe plunger and provide flow by pulling Snap- Tab of the pressure transducer. Slowly deliver priming solution to remove air. Once the transducer head is primed, change the initial white open cap to the alternate white closed cap provided in the package. The initial white cap has an open end to help with priming. Replacing it with the alternate closed white cap provides a closed system and decreases the risk for introduction of infection.

Page 3 of 7 7. Close plunger and remove all air from the system. Ensure that Vamp Jr. and IV line is filled with fluid. Air emboli may occur as a result of air in the tube. Symptoms include a decrease in BP; weak, rapid pulse; cyanosis; and loss of consciousness. Turn patient on left side with head of bed down if air emboli are suspected. Air in the pressure system compromises the dynamic response of the system and interferes with accurate blood pressure measurement. 8. Load the IV administration set into the IV pump and start the pump. 9. Clamp the extension set on a peripheral arterial line or the UAC itself. Attach the Vamp Jr. set up to the end of the extension set or the UAC. IV pump rate should not exceed 2 ml/hr to minimize fluid intake and should not be less than 0.5 ml/hr to ensure line patency. No connector is used on luer lock UAC s to reduce the amount of blood withdrawal and flush requirements. 10. Connect blood pressure cable to monitor and to disposable transducer. 11. Calibrate the monitoring system now and every shift using the procedure outlined below. 12. Restrain the limb when a peripheral artery catheter is in the arterial access Calibration helps to ensure accurate blood pressure values. Hemorrhage may occur if the arterial line is accidently dislodged due to limb

Page 4 of 7 site. movement. 13. For peripheral arterial line use, secure the Vamp Jr. syringe reservoir with tape to prevent use. Blood sampling for PAL s will not use the syringe reservoir. 14. Change the IV administration system and the transducer every 96 hours or if blood backs up into the pressure dome. This does not include the extension set immediately connected to the intracath on a PAL. 15. Observe the system hourly for blood clots, a dampened or flat waveform, the limb s color and perfusion, and perfusion to the legs with an UAC. Discard the transducer when it is changed or discontinued into Biohazard waste bins. Clean the monitoring cable and return it to its proper storage space. Thrombosis may occur as a result of arterial damage or inadequate heparin infusion. If occlusion is suspected or waveform dampens or flattens, notify the physician/designate immediately. Calibration Procedure Once an infant is connected to arterial pressure monitoring, the monitor is calibrated immediately, at the beginning of each nurse s shift, and whenever the line has been disconnected. ACTION RATIONALE 1. Position the transducer head horizontally so that the stopcock port is at heart level.

Page 5 of 7 2. Clamp the arterial line closest to the patient by either clamping the extension set or by turning the valves on the Vamp Jr. off (turning the valve perpendicular to the line). Ensure the line is off to the patient to prevent accidental hemorrhage. 3. Turn the stopcock off to the patient side. Open the transducer head. Alternate white closed cap does not need to be completely removed, just loosened. 4.. Check to ensure that there is no backflow of blood. In the event a line is not clamped off to the patient, blood will back up quickly. 5.. Depress the soft key labeled zero on the blood pressure module until a tone is emitted. Alternately, touch the blood pressure reading on the screen to bring up a list of options, scroll down to and press Zero ABP. 6.. The zero and calibration process is complete when the monitor states that the line has been calibrated. 7. Close the transducer head. 8. Open the stopcock. 9. Unclamp the arterial line to the patient by either unclamping the extension set or by

Page 6 of 7 turning the valves on the Vamp Jr. open (turning the valve parallel to the line). 10. Document in the nursing flowsheet that calibration has been done. Possible causes of a dampened waveform: Catheter too small Clot forming on catheter Loose connections Catheter or blood pressure tubing kinked Air between transducer diaphragm and head Occlusion forming Related Documents Adapted with permission from Stollery Children s Policy and Procedure Manual: http://www.intranet2.capitalhealth.ca/nicu/pages/policiesprocedures/policiesprocedures_new.htm Blood Pressure Arterial Lines, December 2009 RELATED POLICIES AND PROCEDURES Assessment of the Newborn Blood Sampling, Arterial Lines Heparin Use References Inservice Bulletin http://ht.edwards.com/resourcegallery/products/pressuremonitoring/pdfs/1158vampjrinsvc.pdf Revisions July 2005 December 2009 May 2015

Page 7 of 7 Signing Original Signed GAIL CAMERON SENIOR DIRECTOR OPERATIONS MATERNAL, NEONATAL & CHILD HEALTH PROGRAMS COVENANT HEALTH GREY NUNS & MISERCORDIA HOSPITALS 7 May 2015 DATE DR. SANTIAGO ENSENAT MEDICAL DIRECTOR NEONATAL PROGRAM COVENANT HEALTH GREY NUNS HOSPITAL DATE Original Signed 14 May 2015 DR. SHARIF SHAIK DATE MEDICAL DIRECTOR NEONATAL PROGRAM COVENANT HEALTH MISERCORDIA HOSPITAL