Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Similar documents
In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Pump out insert (condenser and still lines) overnight with turbo pump

Oxygen Contamination

INSTRUCTION MANUAL. January 23, 2003, Revision 0

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

Glove Box Installation Manual

Rejuvenation Instructions

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

1)! DO NOT PROCEED BEYOND THIS MARK

INFLATOR TDUX-IT-16 AND GAS CYLINDER E

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

Microfluidic Demonstration

Air Ball! Evaluation copy

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

ACV-10 Automatic Control Valve

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

Scanning Electron Microscope JEOL JSM F

Chromat Calibration Updated October 27th, 2017

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

Product Information News

NITROGEN INFLATION SYSTEM NG-4, NG-6, NG-12, NG-18

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Agilent Auxiliary Gas Module AGM 2. User s Guide

University of MN, Minnesota Nano Center Standard Operating Procedure

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

PXRF- Tracer Helium Flow Control System

with O 2 Controller Instruction Manual

User s Guide. Vacuum Controller for liquid delivery systems

OPERATION OF THE DIMPLER

Pressure Dump Valve Service Kit for Series 2300 Units

User s Guide. Pressure Controller for liquid delivery systems

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

256 Pneumatic Pressure Indicator

Ammonia Treatment Instructions Page 1 of 5

Operation Manual Piston Sensed Gas Pressure Regulators

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

Standard Operating Manual

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Installing ASI Umbilical Delivery Systems

6900 Maintenance Instruction System Flush

Table of Contents Want full page pictures with most important parts labeled in the beginning.

Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System)

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

INSTALLATION INSTRUCTIONS

Model PSI Compressor with 3-Gallon Air Tank 12VDC

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

FAULT CODE TROUBLESHOOTING INDEX

Maintenance Manual. Note: This document is broken up into the following 6 sections of prescribed maintenance.

Met One E-BAM Particulate Monitor

Armfield Distillation Column Operation Guidelines

General remarks regarding ALS robot system:

The X-11 Molding Chamber

MAINTENANCE PROCEDURE FOR X 650

DCIF Training Guide Variable Temperature Bruker B400 (last update )

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

OWNER'S MANUAL LOCK-N-LOAD BULLET FEEDER (PISTOL)

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

Float Operated Level Controllers

8. Carefully layout all the parts of the reservoir setup and clean any dirt, grime or dust from the parts. ( image 8 ) 8

PRESSURE-TEMPERATURE RELATIONSHIP IN GASES

Installation Instructions

DBML-60/80 Squeeze Tool

5890II GC Standard Operating Procedure 9/2/2005

PANACEA P200 STAINLESS STEEL USER MANUAL

#LZ400 LEAKALYZER. Water Loss Sensor INSTRUCTION MANUAL Country Dr. #190 St. Paul, MN

Section 3- Repair and Calibration Procedures

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

User Instruction Manual

SPUTTER STATION STANDARD OPERATING PROCEDURE

INSTALLATION & START-UP INSTRUCTIONS. Clack WS1 & WS1.25 METER WATER SOFTENER SYSTEMS

Revised: June 7, 2017

OPERATING INSTRUCTIONS

MANUAL BE SERIES Test Benches

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Columbia Taping Tools. Hydra Reach Handle. Manual

WIDDER TOOLS. HPICAL Calibration Station 1 (203)

Halsey Taylor Owners Manual

RG1200 Service and Repair Manual

Operation Manual. Diaphragm Sensed Gas Pressure Regulators

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

INSTALLATION INSTRUCTIONS

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

[USING THE VITROBOT April 9, Using the Vitrobot

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

Installation Instructions

GM Series Dual-Block Multi-Function Gas Control Valves

Shimano Di2 Installation on S5

Oerlikon Sputtering Evaporator SOP

Installation Instructions

RM-80 respiration monitor

User's Manual. MixRite TF 10. Edition 05.08

INSTALLATION & MAINTENANCE INSTRUCTION

Transcription:

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures range: 2 K to 300 K See IRM Quarterly, Spring 1992, Vol. 2, No. 1; Fall 1993, Vol. 3, No. 3; Spring 1994, Vol. 4, No. 1 (magnetic quantities and units); Winter 1998-1999, Vol. 8, No. 4; Winter 2002-2003, Vol. 11, No. 4 Contents: I) Sample Preparation II) Sample Loading/Unloading III) Sample Centering IV) Software Setup and Running Experiments V) Helium Transfer

The MPMS consists of an instrument station, a control/electronics rack/station, a computer, a sample holder rod, various small tools, and supplies. There are two possible configurations for the instrument: 1) Standard. This configuration incorporates an air lock, allowing samples to be loaded or removed at low temperature. 2) RSO (reciprocating-sample option). This configuration increases sensitivity for measurement of very weak samples, but lacks an air-lock mechanism, and thus requires that the system be at room temperature for sample loading or removal. I. SAMPLE PREPARATION (For either RSO or standard mode) 1) Prepare the sample for the drinking straw holder: a) Gel-cap method: Insert some fiberfrax or bits of kimwipe in the ends of a gel-cap (to keep the sample snug). After placing the sample in the gel-cap and closing it until it snaps shut, seal the side of the gel-cap with enough Capton tape to make a snug fit in the drinking straw. b) Second straw method: First, place the sample chip in a short piece of drinking straw. Then, insert this sideways into the main straw. Recommended only for single chips. 2) Determine instrument configuration before proceeding to load your sample into the MPMS (see Fig. 1below). 3) Cut the straw to the appropriate length and position the sample in the straw according to the tem- Standard-mode transport head (oblique view) RSO-mode transport head (posterior view) Figure 1. Two possible head configurations for the MPMS. Standard (left), and reciprocating sample option (RSO; right).

plates. 4) Place a small piece of Capton tape over the bottom of the straw to prevent sample from falling out (standard mode only). 5) Puncture straw near bottom and top to allow air to escape during airlock evacuation (standard mode only). II. SAMPLE REMOVAL/LOADING 1) Standard Configuration. Steps a-e are for removing a sample from the instrument. If there is no sample installed, proceed to Step f to load your sample. a) b) c) Figure 2. Sample airlock. a) Airlock viewport. You should be able to see entire sample/straw assembly in window before closing off airlock. b) Top of airlock. Note difference between thumb screws and sample locks. c) Lower airlock controls. a) Loosen thumb screws (NOT the sample locks!! see Fig. 2b) and rotate the sliding-clamp bracket out from under the thumb screws. b) Pull the sample rod assembly up until the sample is visible in the window (Fig. 2a). Make sure that you can see the bottom of the straw! c) Close the sample chamber valve (Fig. 2c). The venting light will go on for a few seconds. d) When the light goes out, loosen sample locks (Fig.2b) and remove sample rod assembly. e) Replace blue plug and tighten sample locks. Press Purge Airlock button (Fig. 2c). f) Exchange sample on rod: i) Gently lay the standard rod on the long counter top with the bracket ( sliding clamp in Fig. 3) hanging just over the edge. ii) Slide the glass sleeve (Fig. 3) to the left, exposing the straw/sample assembly. iii) Remove old sample. iv) Attach new sample securely to the rod with Capton tape. v) Position sample in straw according to template. vi) Puncture straw near bottom and top several times to allow air to escape. vii) Slide glass sleeve back as far to the right as it will go.

Figure 3. Sample rod assembly (standard configuration). Glass sleeve shown in left drawing has been moved to left out of drawing above. Sample/straw is attached to top (left-most) centering plug with Capton tape. Bottom (right-most) centering plug may not always be used. g) Take the rod/sample back to the MPMS. The blue plug should be in place, the sample locks closed, and the green Ready light on. h) Open sample chamber valve slightly and close it again. This should vent the airlock and the green light will go out. i) Loosen sample locks and remove blue plug. j) Insert prepared sample rod assembly with white dot facing forward, and tighten sample locks. k) Push purge air lock button. The venting light will go on and off for about 30 seconds. Watch the sample through the window to make sure it does not move i or fall out of the straw ii. i) If the sample moves only a little (<few mm), it is still OK; proceed to next step. If it moves up or down too much, it will not be possible to center it. In this case, wait for the green ready light, open the sample chamber valve, then close it again, and return to step d to remove and reposition your sample. ii) If the sample falls out of the straw, find an IRM staff member for assistance. DO NOT ATTEMPT TO RETRIEVE IT YOURSELF! k) When the green ready light comes on, open the sample chamber valve. If the ready light fails to show after a few minutes, there are two possible causes: i) The system is cooling to a set-point temperature. Check the current T and set-point T. If the system is cooling, you can interrupt it: click on the temperature display, select Set Temperature, and enter a value near (slightly above) the current temperature. Once the new set-point is registered, the green light should come on immediately. You can then re-enter the original set-point T and continue with sample installation as the system cools. ii) If the temperature is stable and the green light still fails to illuminate, there is probably a vacuum leak. Find an IRM staff member ASAP! l) Push the sample rod assembly down. At least once per day, apply a thin coat of vacuum grease to the rod as you insert it. m) Fit the bracket under the thumb screws, and tighten the thumb screws. 2) Sample Loading for RSO Configuration. a) Check system temperature. Look at the system status panel on the bottom of the computer screen and find the current temperature (both the set-point and actual temperature are shown). If the temperature is below 280K do not continue with sample loading until T is >= 280K. To change the temperature set-point, click on the temperature indicator and an edit box will open,

where you can enter a value of 300. Wait until T is at least 280K before continuing. b) Click on the Sample menu on the top of the screen and select Install/Remove Sample. You will hear the sample space being flushed with He gas, and the top vacuum seal will break open. c) Remove the black cap (which the software will refer to as the servo) and set it on the counter upside-down (so the rubber o-ring does not pick up dirt from the counter top). d) Remove the clear plug from the top of the RSO transport head. Invert it and use the threaded tip to grab the matching threaded piece on the blue plug and gently remove the RSO sample rod. Replace the clear plug. e) Exchange sample on rod: i) Gently lay the RSO rod on the long counter top with the blue plug hanging just over the edge. ii) Remove the previous straw/sample assemblage from the rod and remove the black end-plug from the straw. iii) Press your straw/sample onto the RSO rod. iv) Adjust the position of your sample in the straw, according to the RSO 4-cm position on the template. v) Insert the end-plug. f) Take the rod and sample back to the MPMS. Remove the clear plug, lower the sample rod (twisting gently if you encounter any resistance), press the blue plug into place (using a fingertip or the threaded plug tip), and replace the clear plug. g) Click on the Purge button on the computer display. You will hear the sample space being evacuated and flushed with He gas. This will be repeated several times, and then the green Ready light on the RSO head will come on. h) An editing box will open on the screen, allowing you to enter the new sample name, mass, etc. Then proceed to section III. III. SAMPLE CENTERING (For either RSO or standard mode) 1.0E-04 1) Make sure temperature is stable before centering for best results. 2) Click on the Center menu on the top of the screen. 3) Select DC Centering. 4) Click on Initialize Transport. 5) Run the Center procedure. A SQUID response curve will be displayed in about 30 seconds. The response curve should be symmetric with the large central bump (up for positive fields or moments) centered on the plot at ~2 cm, as in the example (Fig. 4). Check the magnetic moment; if > 0.5 emu, remove your sample, decrease the amount of material and re-start. If there SQUID voltage 8.0E-05 6.0E-05 4.0E-05 2.0E-05 0.0E+00-2.0E-05-4.0E-05 0 1 2 3 4 position [cm] Figure 4. Output from Center procedure. measured best fit is not a recognizable central peak (positive or negative), there are two things to try: a) Apply a small field (e.g., 10 G or 100 G; click on the field display panel on the bottom of the

screen and enter the desired value). For weak samples that have not been previously magnetized, this may help to generate a measurable signal. Repeat Centering proceedure with field on. b) Try a Full DC Scan. If the sample is far from the correct location, this full-length scan is a more effective way to detect its position. 5) Adjust sample position. a) Adjust automatically. If sample location is only slightly (a few 10s of mm) offset from the 2 cm mark, click on Adjust Position and run the Adjust Automatically procedure. Check the box indicating you want to re-run the Center procedure when finished. b) Adjust manually. If the sample is too far off-center, you will not be able to adjust the position automatically. i) Standard configuration: you can run the Adjust Manually procedure, which involves loosening and tightening the knurled nut on the sample rod, as instructed by the software. ii) RSO configuration: you will have to remove the sample, re-position manually, and start over. IV. SOFTWARE SET-UP AND RUNNING EXPERIMENTS The panel on the left side of the screen contains controls for selecting and running sequences (experiments) and for specifying data files for output. 1) Set sample name. The top box indicates the sample name. You can click on the Change button below to change the sample name, mass, etc. for the header of the data file. (Note that mass is not used.) 2) Select experiment sequence. The second box shows the currently selected Sequence File. Experiments are run by executing Sequence Files, which contain all the necessary instructions for temperature and field control, data acquisition, etc. To view the contents of the current sequence file, click on the Edit button. An edit window will open, and will display the contents of the selected file. To close the edit window, click on File and then Close. To view a list of sequence files, click on the Change button below the sequence file name. You can scroll through the list displayed, and select a file by doubleclicking on its name. If you wish to modify any of the existing files, consult an IRM staff member. 3) Set data file name. The third box on the left panel is labeled Sequence Base Data File Name. All data collected by the sequence as it executes will be stored in the file designated in this box. To specify a new file, click on the Change button below the sequence file name. You can use the control buttons to change directories, or to create a new directory for your files. 4) Start sequence. The fourth and final box on the left-side panel is the Sequence Control box. Click on the Run button to begin the experiment. 5) Display plot. Once the sequence begins collecting data, you can display a graph of measurements as they are acquired. Click on the View button below the Sequence Base Data File Name box. A list of relevant files will appear. The dat file is most probably what you want to display: moment vs. temp, field, etc. The lastscan file allows you to look at the SQUID response curve for the previous measurement. VI. HELIUM TRANSFER (IRM Personnel ONLY) 1) This is a two-person job. 2) Collect the following near the MPMS: liquid helium dewar

helium gas tank (in the corner behind the door) helium transfer tube (on the back wall) dewar exit valve adapter (on the transfer tube) two pair of gloves (in the cabinets) heat gun (in the cabinets). 3) Purge both the hose from the helium gas tank and the connector to the dewar with helium gas, then connect the hose to the dewar. Set the regulator for about 1 psi, but leave the inlet valve on the dewar closed. 4) Bleed the 0.5 psi relief valve on the dewar, then open the valve at the top of the dewar. Insert the long end of the transfer tube, making sure to seat the outer sleeve of the tube in the dewar adapter. Push in slowly. 5) Close the 0.5 psi relief valve and open the inlet valve to start gas flow into dewar. 6) Bleed off the pressure in the MPMS relief valve, then remove the cap when ready to insert the short end of the transfer tube [See next step]. 7) When the sound of the gas plume coming out of the transfer tube starts to change (indicating liquid is nearing), insert the short end of the transfer tube into the MPMS. Do this quickly to avoid freeze-up, but with finesse to negotiate around all the internal baffles. Again make sure that the outer sleeve of the transfer tube seats in the connector to the MPMS. 8) Adjust the regulator to get about 1 psi for transfer pressure. 9) In the Utilities menu, select Helium Fill, and start the count-down timer by clicking on the up arrow once or twice. The whole transfer should take about 10 minutes. 10) Close the regulator when the MPMS is about 95% full. Pull the transfer tube out of the MPMS when the helium-level jumps up to 98% full. CAUTION! The tube is dangerously cold! USE GLOVES! Pull the other end of the transfer tube out of the dewar. USE GLOVES! 11) Replace the cap on the MPMS. Close the inlet valve on the dewar and open the 0.5 psi relief valve. Close the dewar exit valve, and remove the adapter 12) Stop and close the graph. The gas flow on the MPMS meters should drop back to below 100 cc/min in a few minutes. 13) Put everything away.