TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

Similar documents
ANSI N323A and ANSI N323B

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Accreditation for In-Service Inspection of Transportable Pressure Receptacles (TPRs)

Gas mixtures. Individual solutions specifically for your application

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

GUIDANCE FOR THE APPLICATION OF EQUIPMENT CALIBRATION AND TRACEABILITY OF MEASUREMENTS ON INSPECTION BODIES

General Accreditation Guidance. User checks and maintenance of laboratory balances

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

SEMETS3-17 Checking and calibrating mechanical inspection equipment

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

Service Calibration are you doing it properly?

Supersedes: The copy of this document located on Measurement Canada s website is considered to be the controlled copy.

CALIBRATION SYSTEM REQUIREMENTS. ESCC Basic Specification No

INSPECTION FOCUS ON:

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z CALIBRATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017 ANSI/NCSL Z

Standard Operating Procedure Measuring & Testing Equipment

Certificate of Accreditation

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Recommended Practices for Calibration Laboratories

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z TROXLER ELECTRONICS LABORATORIES, INC. James Pratt Phone:

Requirements for the certification of movements and mechanical watches resistant to magnetic fields of 1.5 T ( G)

ANNEX A. (Normative) Tank Calibration Frequency and Re-computation of Calibration Tables

ATT-66/96, DENSITY TEST, ASBC CONTROL STRIP METHOD

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Light Lab. Reliable Measurements through Regular Calibration. DAkkS

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

FPG8601 Force Balanced Piston Gauge

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

2014 Programme of Training Courses in Measurement & Calibration

Traceable calibration of automatic weighing instruments in dynamic operation

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

1. Scope: This document describes the procedure for verifying the performance of vaccine vial monitors.

Peak Pressure Indicator MSI-3

PTB Ex PT Scheme. Procedure Instruction of program Explosion Pressure - Test Round 2017

I N N O V A T I O N C O N T I N U E S... International Patent Filed in 45 Leading Countries

EN ISO/IEC Technical Requirements

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

SIKA Ba_ODWT15_en 05/2018. Operating manual for oil deadweight tester 0.015% accuracy ODWT15

General Accreditation Criteria Reference Material Producers. ISO Standard Application Document

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

Denise L Seman City of Youngstown

Gerald D. Anderson. Education Technical Specialist

Moor, Tend Mooring And Unmoor Ship - Supervisor Level -

Any use of this material except in accordance with a written agreement with Western Power is prohibited.

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

Dead-weight tester High-pressure version Model CPB3800HP

Guide No. 2/2009. Design, training and equipment Martin Firlus. Physikalisch Technische Bundesanstalt Braunschweig und Berlin

UKAS Accredited Electrical & Mechanical Calibration Services (Laboratory and On-site)

INSPECTION, TESTING AND CERTIFICATION OF GASFITTING WORK DONE UNDER SUPERVISION

Certificate of Accreditation

A Code of Practice for the Calibration of Industrial Process Weighing Systems

technical guide Laboratory Balances Calibration Requirements

Part 1: General principles

SAFE EGRESS FROM DEEP STATIONS Flawed Criteria in NFPA 130

Stability and traceability of calibration gases for EN 14181

EUROPEAN pr I-ETS TELECOMMUNICATION September 1995 STANDARD

CORESTA Guide N 6. February A User Guideline for the use of balances for cigarettes and cigarette related products

AN31E Application Note

Best Practice for Calibrating LTH Conductivity Instruments

GMC-I Service GmbH Calibration Center

Transferpette -8/-12

GAS FILLING MACHINES PROCESS MONITORING PRESSURE BALANCE

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

OD plug gauge. for bores and internal profiles from dia. 2,5 mm. OD plug gauge in open design. Special designs Accessories. OD plug gauge for various

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF MASS METROLOGY

INTERNATIONAL STANDARD

Siting classification for Surface. Michel Leroy, Météo-France

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

INTERIM ADVICE NOTE 171/12. Risk Based Principal Inspection Intervals

Columbus Instruments

Instrument Operating Manual

SAUTER FK Version /2016 GB

Procedure Instruction of program Pressurized Enclosure - Test Round 2017

Pressure Equipment Directive PED 2014/68/EU Commission's Working Group "Pressure"

Lab # 03: Visualization of Shock Waves by using Schlieren Technique

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Certified Reference Materials FOR REFRACTOMETERS AND POLARIMETERS

No. 2 TEST WEIGHT ASSESSMENT

Verification Of Calibration for Direct-Reading Portable Gas Monitors

Certified Accuracy. Ring Force Gauge. Models and Capacities Available. Design and Principle of Operation

New Thinking in Control Reliability

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE CO 2 PRESSURE. (+34)

LONE WORKING HEALTH AND SAFETY GUIDANCE

MET-101 Introduction to Metrology

Systems of Accounting for and Control of Nuclear Material

Service ISO17025 explained

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Safe management of industrial steam and hot water boilers A guide for owners, managers and supervisors of boilers, boiler houses and boiler plant

Instrumentation & Data Acquisition Systems

BRINGING YOU THE SINCE 1980

Characterizers for control loops

Independent Ethics Committees

Temperature Controllers

Transcription:

GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION Approved By: Senior Manager: Mpho Phaloane Revised By: Field Manager: Neville Tayler Date of Approval: 2015-08-26 Date of Implementation: 2015-08-26 SANAS Page 1 of 9

CONTENTS: 1. Purpose... 3 2. Scope... 3 3. Definitions... 3 4. References... 3 5. Calibration Intervals... 3 6. Recalibration... 4 7. Methods for Determining Calibration Intervals... 4 8. Suggested Initial Calibration Intervals for Certain Dimensional Equipment... 8 ADDENDUM 1: Amendment Record... 9 SANAS Page 2 of 9

1. Purpose The purpose of this document is to provide guidelines concerning calibration intervals and recalibration. 2. Scope This document is applicable to accredited bodies as defined in the Accreditation Act, Act No. 19 of 2006. 3. Definitions Calibration Interval: Accredited body: The period over which there is an estimated high probability that the performance of the instrument will remain within the desired specification. means an organisation or facility that has been accredited by SANAS or by a member of the recognition arrangements of the International Laboratory Accreditation Co-operation (ILAC) or the International Accreditation Forum (IAF); 4. References ILAC G24:2007 Act 19 of 2006 Guidelines for the determination of calibration intervals of measuring instruments. Accreditation Act No. 19 of 2006: Accreditation for Conformity Assessment, Calibration and Good Laboratory Practice Act, 2006 5. Calibration Intervals 5.1 Gauges and instruments should be recalibrated before drift or wear cause them to be unfit for their intended purpose. 5.2 The rate of drift or wear in a given application is estimated by projection from past calibration history. [This is why SANAS calibration certificates should contain actual calibration measurement results (before and after any necessary adjustment), not just pass of fail. Only the former can provide the data needed to determine the calibration interval]. 5.3 In the case of standard lamps, consideration must be given to burning time. 5.4 Calibration intervals should be chosen to give the desired confidence in measurements for an acceptable cost of calibration. The desired confidence depends on the risk and consequences of an inaccurate measurement. These factors are determined by the user and are therefore subject to discussion between the laboratory and the user. 5.5 As per ISO/IEC 17025:2005 Calibration Laboratories are not permitted to include any recommendation on the calibration interval unless this is agreed to by the customer, or is a legal requirement. SANAS Page 3 of 9

6. Recalibration Calibration is necessary in order to establish confidence in the performance of a gauge or instrument. A calibration history supported by certificates contributes to this confidence. It is therefore good practice to compare new calibration results with old ones. In certain circumstances it may be possible to reduce the amount of new calibration work on the basis of consistency with previous results. For example, the full procedure for calibrating gauge blocks by interferometry requires that a measurement be made with one face of the gauge block wrung to the platen, followed by a second measurement with the other face wrung to the platen. The average of the two measurements is taken as the final result. However, if the first measurement is found to agree well with a previous measurement as recorded on a certificate, it may be decided not to proceed with the second measurement, thereby reducing costs. Again, for example, the mandatory parameters listed for the calibration of screw gauges and plain rings and plugs include roundness. It is unnecessary to perform roundness measurements when recalibrating such gauges. 7. Methods for Determining Calibration Intervals 7.1 Introduction The purpose of this document is to provide guidance on the determination of calibration intervals. An important aspect of the efficient operation of a calibration system is the determination of the maximum period between successive calibrations of measurement standards and measuring equipment. A large number of factors influence the frequency of calibration. The most important of these factors are: Uncertainty of measurement; type of equipment; manufacturer s recommendations; trend data obtained from previous calibration recommendations; recorded history of maintenance and servicing; extent and severity of use; tendency to wear and drift; frequency of cross-checking against other measuring equipment, particularly measuring standards; frequency and formality of inhouse check calibrations; environmental conditions (temperature, humidity, vibration, etc); accuracy of measurement sought; the penalty of an incorrect measured value being accepted as correct because the measuring equipment has become faulty. The cost of calibration cannot normally be ignored in determining calibration intervals and this may therefore be a limiting factor. It is obvious from all these stated factors that a list of calibration intervals which can be universally applied cannot be constructed. It is more useful to present guidelines on how calibration intervals may be established and then reviewed once calibration on a routine basis is underway. There are two basic and opposing criteria which are required to be balanced when deciding on the calibration intervals for each item of measuring equipment. These are: the risk of measuring equipment failing to conform to specification when in use should be as small as possible; keeping the calibration costs to a minimum. SANAS Page 4 of 9

Therefore, methods are presented in this document for the initial selection of calibration intervals and for the readjustment of these intervals on the basis of experience. 7.2 Initial Choice of Calibration Intervals The basis of the initial decision in determining the calibration interval is invariably the so-called engineering intuition. Someone with experience of measurements in general, or of the measuring equipment to be calibrated in particular, and preferably with knowledge of the intervals used by other laboratories, makes an estimate for each item of equipment or group of items as to the length of time it is likely to remain within tolerance after calibration. Factors to be taken into account are: the equipment manufacturer s recommendation; the expected extent and severity of use; the influence of the environment; the required uncertainty of measurement; published data about the same or similar items; any legal requirement; the accuracy of measurement sought. 7.3 Methods of Reviewing Calibration Intervals A system which maintains calibration intervals without review, determined only by so-called engineering intuition, is not considered to be sufficiently reliable. Once calibration on a routine basis has been established, adjustment of the calibration intervals should be possible in order to optimize the balance of risks and costs as stated in the introduction. It will probably be found that the intervals initially selected are not giving the desired optimum results; items of equipment may be less reliable than expected; their usage may not be as expected; it may be sufficient to carry out a limited calibration of certain items instead of a full calibration; the drift determined by the regular calibration of the equipment may show that longer calibration intervals are possible without increasing the risks, and so on. If shortage of money or shortage of staff means that extended calibration intervals are necessary, it should not be forgotten that the costs of using inaccurate measuring equipment may be significant. If an estimate of these costs is made, it may well be found to be more economic to spend more money on calibration and to reduce the calibration intervals. A range of methods is available for reviewing the calibration intervals. These differ according to whether: items of equipment are treated individually or as groups (for example, by maker or by type); items fail to comply with their specifications due to drift with the lapse of time, or by usage; data are available and importance is attached to the history of calibration of the equipment. No one method is ideally suited for the full range of equipment encountered. 7.3.1 Method 1: Automatic or Staircase Adjustment Each time an item of equipment is calibrated on a routine basis; the subsequent interval is extended if it is found to be within tolerance, or reduced if it is found to be outside tolerance. This 'staircase' response may produce a rapid adjustment of intervals and is easily carried out without clerical effort. When records are maintained and used, possible trouble with a group of items, indicating the desirability of a technical modification or preventive maintenance, will become apparent. SANAS Page 5 of 9

A disadvantage of systems which treat items individually may be that it is difficult to keep the calibration work-load smooth and balanced and that it requires detailed advanced planning. 7.3.2 Method 2: Control Charts The same calibration points are chosen from every calibration and the results are plotted against time. From these plots, both scatter and drift are calculated, the drift being either the mean drift over one calibration interval or, in the case of very stable equipment, the drift over several intervals. From these figures the effective drift may be calculated. The method is difficult to apply, in fact very difficult in the case of complicated equipment, and can virtually only be used with automatic data processing. Before calculations can commence, considerable knowledge of the law of variability of the equipment, or of similar equipment, is required. Again, it is difficult to achieve a balanced work-load. However, considerable variation of the calibration intervals from those prescribed is permissible without invalidating the calculations; reliability can be calculated and, in theory at least, it gives the efficient calibration interval. Furthermore, the calculation of the scatter will indicate if the manufacturer's specification limits are reasonable, and the analysis of the drift which is found may help in indicating the cause of the main drift. 7.3.3 Method 3: Calendar Time Items of measuring equipment are initially arranged into groups on the basis of their similarity of construction and of their expected similar reliability and stability. A calibration interval is assigned to the group, initially on the basis of engineering intuition. In each group, the quantity of items which return at their assigned calibration interval and are found to have excessive errors or to be otherwise nonconforming is determined and is expressed as a proportion of the total quantity of items in that group which are calibrated during a given period. In determining the 'nonconforming' items, those which are obviously damaged or which are returned by the user as suspect or faulty, are not included as they are not likely to cause measurement errors. If the proportion of 'nonconforming' items of equipment is excessively high, the calibration interval should be reduced. If it appears that a particular sub-group of items (such as a particular make or type) does not behave like the other members of the group, this sub-group should be removed to a different group with a different calibration interval. The period during which the performance is assessed should be as short as possible, compatible with obtaining a statistically meaningful quantity of calibrated items for a given group. If the proportion of nonconforming items of equipment in a given group proves to be very low, it may be economically justifiable to increase the validation interval. Other statistical methods may be used. 7.3.4 Method 4: 'In-Use' Time This is a variation on the foregoing methods. The basic method remains unchanged but the calibration interval is expressed in terms of hours of use rather than in calendar months of elapsed time. An item of equipment may be fitted with an elapsed time indicator, and is returned for calibration when this indicator reaches a specified value. The important theoretical advantage of this method is that the number of calibrations performed, and therefore the cost of calibration, varies directly with the length of time for which the equipment is used. Furthermore, there is an automatic check on equipment utilization. SANAS Page 6 of 9

However, the practical disadvantages are many, and include: it cannot be used with passive measuring instruments (for example, attenuators) or with passive measurement standards (resistors, capacitors, etc.); it should not be used when equipment is known to drift or deteriorate when on the shelf, or when handled, or when subjected to a number of short on-off cycles: it should in any case have a calendar-time backup; the initial cost of the provision and installation of suitable timers is high and, since users may interfere with them, supervision may be required which will again increase costs; it is even more difficult to achieve a smooth flow of work than with the other methods mentioned, since the calibration laboratory has no knowledge of the date when the calibration interval will terminate. 7.3.5 Method 5: In-Service or Black-Box Testing This method is complementary to a full calibration. It can provide useful interim information on characteristics of measuring equipment between full calibrations and can give guidance on the appropriateness of the calibration programme. This method is a variation on methods 1 and 2 and is particularly suitable for complex instruments and test consoles. Critical parameters are checked frequently (once per day or even more often) by portable calibration gear or, preferably, by a 'black-box' made up specifically to check the selected parameters. If the equipment is found to be nonconforming by using the 'black-box', it is returned for a full calibration. The great advantage of this method is that it provides maximum availability for the equipment user. It is very suitable for equipment which is geographically separated from the calibration laboratory, since a complete calibration is only done when it is known to be necessary or at extended calibration intervals. The main difficulty is in deciding on the critical parameters and in designing the 'black-box'. Although the method theoretically gives a very high reliability, this is slightly ambiguous, since the equipment may be failing on some parameter which is not measured by the 'black-box'. In addition, the characteristics of the 'black-box' itself may not be constant and it also needs to be regularly calibrated. 7.4 SANAS Requirements Concerning the Determination of Calibration Intervals 7.4.1 The manager of a SANAS accredited laboratory should discuss with his/her customer all calibration requirements. The conditions under which gauges or instruments will be used should be included in the discussion. The calibration interval or re-calibration date may only appear on a SANAS certificate of calibration if requested in writing by the customer. 7.4.2 If a gauge or instrument is abused, damaged, or subjected to conditions which cause wear or drift, it is the responsibility of the user of the gauge or instrument to request recalibration at appropriate intervals. SANAS Page 7 of 9

8. SUGGESTED INITIAL CALIBRATION INTERVALS FOR CERTAIN DIMENSIONAL EQUIPMENT As noted in Addendum 1 of this document, the initial discussion in determining calibration intervals is invariably based on "engineering intuition". A list of suggested initial calibration intervals for dimensional equipment has been drawn up on the above basis and is given below. Instrument or Gauge Gauge Blocks Secondary Standards Plain Ring Gauges Secondary Standards Plain Plug Gauges Secondary Standards Screw Ring Gauges Screw Plug Gauges Lab Standards Taper Gauges Angle Plates: Steel & Granite Box Parallels: Suggested Calibration Interval 3 years 1year, thereafter 2 years 6 months - 3 years 6 months - 6 months - 3 years 6 months - 6 months -2 years Remarks Maximum 4 years for steel, 4 years for tungsten carbide Ring gauges should be calibrated at 6 months, and 2 years Calibrate at 6 months and thereafter every 2 years Calibrate at 6 months and thereafter every 2 years Steel & Granite 6 months - 2 years Cylindrical Square 6 months - 2 years Calibrate at 6 months and thereafter every 2 years Granite Surface Table 6 months - 2 years Calibrate at 6 months and at 12 months Length Bars 6 months - 3 years Optical Flats Secondary Standards 3 years Visual inspection for scratches before every use Probe Tips Reference Standards for Surface Finish Visual inspection for scratches before every use Reference Standards for Visual inspection before every use Roundness 2 years Sine Bars 6 months - Visual inspection before every use Steel Square 6 months - Visual inspection before every use Straight Edge 6 months - Visual inspection before every use Thread Measuring Cylinders Thread Measuring Vee-pieces Visual inspection before every use Visual inspection before use SANAS Page 8 of 9

ADDENDUM 1: Amendment Record Proposed By: Section Change FM 2. Scope changed from SANAS Accredited laboratories, to accredited bodies as defined in the accreditation Act, Act 19 of 2006. FM 3 Added definition of Accredited Body FM 4 Reference changed from ISO 10012 to ILAC G24:2007 Added reference to Act 19 FM 5 Added clause 5.5 FM 7 Added additional detail from ILAC G24:2007 SANAS Page 9 of 9