Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS

Similar documents
Rapid and Reliable Detection of Dissolved Gases in Water

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

May 1, By John Heim,Doug Staples

Retention Time Locking: Concepts and Applications. Application

M/s. Thermo fisher Scientific

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

Impact of Air Leaks on the Productivity of GC and GC/MS Systems

Experiment GC : Analysis of BTEX by GC-FID

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Operation of the Perkin Elmer TGA-GC/MS

STANDARD OPERATING PROCEDURES

Determination of capsaicinoid profile of some peppers sold in Nigerian markets

Carrier Gases in Capillary GC

GCMSD-Headspace Analysis SOP

Hiden QIC Series Gas Analysers Real-time gas analysis systems

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Instant Connect Helium Saver Injector Module

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

SUPPORTING INFORMATION

Zebron Gas Management

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

A New GC/MS System Designed for Helium Carrier Gas Conservation

APPLICATION NOTE. GC Integrated Permeation Device

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Total Pressure Measurement Using Residual Gas Analyzers

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Thermo Scientific TRACE 1300 Series Gas Chromatograph. Evolve into GC 2.0. powerful breakthroughs

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

Longterm Detection of Anabolic Steroid Metabolites in Urine

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

no peaks missing peaks too small peaks increasing retention decreasing retention

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Sorbent Pen Heavy & Labile SVOC s Taints, Off-Flavors, Odors

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Automated Determination of Dissolved Gases in Water Anne Jurek

During today's webinar we're going to look at the following topics:

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Agilent G3185B QuickSwap Accessory

Quadruple mass spectrometers (transducer type)

This test shall be carried out on all vehicles equipped with open type traction batteries.

Space product assurance

Choosing a Quadrupole Gas Analyzer Application Note #9

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Mass Spec will not Autotune

3. Coupled Techniques in Thermal Analysis

Best Practice for Identifying Leaks in GC and GC/MS Systems

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Residual Gas Analysis Systems for Industry

Gas chromatographic determination of 1,4-dioxane at the

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Quantitative Analysis of Hydrocarbons by Gas Chromatography

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Gun Vacuum Performance Measurements February 2003

3 Flow and Pressure Control

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

CHEMTRACE Fremont Blvd. Fremont, CA 94538, Tel: (510) Fax: (510)

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

Practical approach and problems in in-situ RGA calibration

Agilent 7800 ICP-MS. Specifications and Typical Performance. Fast track metals analysis with Solution-Ready ICP-MS

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Quick Reference Guide. OPTIC 3 - PTV on Shimadzu GC-2010 Gas Chromatograph

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Agilent G Rebuilt Electron Capture Detector

Real-time PCB Gas Monitoring System (MOHMS-21GP)

The analysis of complex multicomponent

Agilent Inlet EPC Leak Test

ISO INTERNATIONAL STANDARD

Versatility. Accuracy. Compact. Versa. Automated Headspace Vial Sampler

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

We release Mascot Server 2.6 at the end of last year. There have been a number of changes and improvements in the search engine and reports.

Ultimate Performance in Gas Chromatography

Transcription:

TECHNICAL NOTE 1598 Comparative analysis of mass spectral matching for confident compound identification using the Advanced Electron Ionization source for GC-MS Authors Amit Gujar, 1 Tim Anderson, 1 Daniela Cavagnino, 2 and Amy Patel 1 1 Thermo Fisher Scientific, Austin, TX, USA 2 Thermo Fisher Scientific, Milan, Italy Keywords Gas chromatography-mass spectrometry, GC-MS, ISQ 7, advanced electron ionization, AEI, library match, NIST, Wiley, spectral quality, compound identification Goal To assess the spectral match quality of the ion spectra obtained using the Thermo Scientific Advanced Electron Ionization (AEI) source against the reference mass spectra of commercial libraries. Introduction Single quadrupole GC-MS operated in electron ionization (EI) mode has long been employed as a convenient technique for a variety of applications spanning environmental monitoring, food safety, forensic and toxicological, pharmaceutical, and industrial analysis, among others. GC-MS serves as both a qualitative and quantitative analytical tool for known compounds as well as for unknowns analysis. The unknowns analysis by GC-MS often relies on compound identification using spectral matches against mass spectral libraries (ex: NIST, Wiley, etc.) commercially available in addition to mass spectral interpretation by the user. 1 The fragmentation pattern of the EI mass spectrum is compared via spectrum matching with the reference mass spectra in the library, and a list of most probable identities is produced based on the best (forward or reverse-search) mass spectrum matching score, with 999 being a perfect match. 2

Counts The Thermo Scientific ISQ 7 single quadrupole mass spectrometer can be configured with a Thermo Scientific Advanced Electron Ionization (AEI) source. The AEI source is specifically designed to increase the ionization efficiency of the compounds eluting from the GC capillary column, which ultimately leads to increased instrument sensitivity. The AEI source utilizes a filament that is in line with the cyclotroning magnetic field and the ionized beam path and together with electrostatic mirroring of electrons leads to significant increase in the yield of ions generated within the source. 3 5, 4, 3, 2, 1, AEI source 45 ev Area=8859 >1 increase (absolute peak area) Traditional source 7 ev Area=651 Traditionally, to obtain the best sensitivity possible, the electron energy used in EI experiments is 7 ev. Thus, most commercially available libraries have spectra that have been acquired at this energy. The efficiency of ionization of most molecules increases exponentially between 1 and 2 ev, followed by plateauing of the fragment ions produced at around 35 ev. 4 It is well known that electron ionization above 4 ev generates highly reproducible spectra, providing stable molecular fingerprints. 5 With the AEI source, the optimum ionization efficiency occurs when the electron energy is set to 45 ev. Nevertheless, according to the minimum acceptable criteria for contributing to the NIST MS database, as stated in the NIST documentation, it is required to acquire at an electron ionization of not less than 2 ev under conditions such that isotope ratios are not grossly distorted. 6 Therefore, it is viable to leverage electron energy other than 7 ev (but not less than 2 ev) for NIST library searching, assuming that the required matching is achieved. An example of increased sensitivity of the AEI source is shown in Figure 1. An increase in area counts greater than a factor of 1 was observed for 1 pg on-column of terbuthylazine acquired in full-scan mode using a mass range of 35 to 55 Da when compared against the signal obtained on a traditional source, like the Thermo Scientific ExtractaBrite EI source. 12.73 13. 13.5 14. 14.5 15.19 Minutes Figure 1. Example of increased sensitivity with the AEI source versus a traditional ion source for 1 pg terbuthylazine on-column, 35 55 Da full-scan In practice, the best identification of a compound is often simply the one with the highest similarity score that represents the goodness of fit between a measured spectrum and a library reference spectrum. For this, the NIST library guidelines in Table 1 were used for interpreting the mass spectral match quality. 2 Table 1. NIST library guidelines Match Factor (SI) or Reverse Match Factor (RSI) Thresholds Mass Spectral Match 9 and above Excellent 8 9 Good 7 8 Fair <6 Poor Three distinct classes of compounds that are often analyzed by GC-MS were tested: semi-volatile organic compounds, pesticides, and terpenes. The aim of this technical note is to demonstrate that the EI mass spectra obtained using the AEI source operated at its optimum electron voltage gives excellent spectral matching when compared with the commercially available reference mass spectra. 2

Experimental Samples Semi-volatile organic compounds Restek 827 MegaMix standard was diluted to 1 ppm with dichloromethane before being analyzed. Pesticides Two multi-residue pesticide standards (organochlorine compounds #2, Restek Cat.# 364 and organophosphorus compounds #5, Restek Cat.# 367), were diluted to 2 ppm with toluene before analysis. Terpenes Individual terpenes standards were obtained from Santa Cruz Biotechnology, Inc. (Dallas, TX, USA) and diluted to 1 ppm in hexane before analysis. The Thermo Scientific ISQ 7 single quadrupole GC-MS equipped with the AEI source and with the ExtractaBrite EI source has been used for the analysis of the above mentioned standard mixtures. The acquired full scan spectra have been compared and matched with commercial MS libraries. Mass spectral libraries For the semi-volatiles and the multi-residue pesticide studies, the NIST 17 mass spectral library (mainlib) was used. For the terpenes study, the Wiley s Flavors and Fragrances of Natural and Synthetic Compounds (FFNSC), 2nd Ed. was used in conjunction with the NIST 17 library. GC and MS conditions Typical conditions used for these experiments are detailed in Table 2. Data was acquired in full-scan mode and processed with Thermo Scientific Chromeleon 7.2 Chromatography Data System (CDS) software. Table 2. GC and MS parameters Thermo Scientific TRACE GC system Column: Liner: SSL mode: Inlet temperature: 3 C Thermo Scientific TraceGOLD TG-5ms GC column 3 m. mm. µm (P/N 2698-142) Thermo Scientific LinerGOLD Splitless liner single taper with wool, 4 mm ID, 78.5 mm length (P/N 453A924) Split (split ratio varied) Septum purge flow: 5. ml/min Carrier flow: Oven program: Constant He flow of 1.2 ml/min 35 C (2. min), 18 C/min to 1 C (.1 min), 23 C/min to 28 C (.1 min), 1 C/min to 32 C (5. min) for SVOC and pesticides, 35 C (2 min), 2 C/min to C (1. min) for terpenes ISQ 7 single quadrupole GC-MS system MS acquisition: Full-scan MS transfer line temperature: 3 C Ion source temperature: 3 C Ionization mode: EI Electron energies: AEI at 45eV and ExtractaBrite at 7 ev Emission current: 5 µa Mass range: 35 5 Da for SVOC and pesticides 35 3 Da for terpenes Dwell time:.2 s 3

Results and discussion Semi-volatile organic compounds (SVOCs) SVOCs are an important class of compounds that are routinely analyzed by GC-MS especially in the field of environmental monitoring. Good spectral matching is key for confident compound identification of environmental contaminants in soil, water, and other matrices. Figure 2 shows the NIST library match factors (SI scores) for the SVOCs analyzed using both the AEI source and the ExtractaBrite ion source for the chromatographically resolved compounds. When searching NIST the match factors for both sources gave very similar results with the majority of compounds having scores of above 9 (excellent match), while all the compounds have match factors above 87 indicating good to excellent spectral match with the NIST library spectra. No significant difference in the average match factor scores was observed for the thirty-nine compounds between spectra acquired using the AEI source (average match factor 929) and the spectra obtained using the ExtarctaBrite ion source (average match factor 921). Figure 3 shows an example of spectra for hexachlorobutadiene acquired with both the ion sources under study, in comparison to the NIST reference spectra (bottom pane). In both cases, the library match factors were excellent and very similar (6 with the ExtractaBrite ion source and 8 with the AEI source). Match Factor (SI) Figure 2. NIST 17 mass spectral library match factors of selected semi-volatile compounds analyzed on the conventional ExtractaBrite ion source and the AEI source 4

12 1 A 2 12 1 B 2 % 5 12 143 155 19 192 229 26 8 % 5 94 16 12 19 188 141 155 192 26 229 8-2 12 1 Library Hit: 1,3-Butadiene, 1,1,2,3,4,4-hexachloro- 6 2-2 12 1 Library Hit: 1,3-Butadiene, 1,1,2,3,4,4-hexachloro- 8 2 % 5 12 141 153 19 188 192-2 12 5 1 15 2 3 229 26 8-2 12 5 1 15 2 Figure 3. Mass spectra of hexachlorobutadiene obtained with (A) the ExtractaBrite ion source and (B) the AEI source % 5 12 141 153 19 188 192 229 26 8 3 Pesticides Pesticides are compounds that are routinely monitored in food and environmental matrices. Figure 4 shows the NIST library match factors for various organochlorine and organophosphorus pesticides. The results from the NIST search show that the mass spectra obtained from the AEI source are comparable to those from the ExtractaBrite ion source, and in both cases the match factors resulted >9 for the majority of the compounds. An example is shown in Figure 5 for the methacrifos mass spectra, which results in a match factor of >95 for the AEI source. Match Factor (SI) Figure 4. NIST 17 mass spectral library match factors of selected pesticides for the ExtractaBrite ion source and the AEI source 5

12 1 A 1 12 1 B 1 % 5-2 12 1 55 8 11 99 19 Library Hit: Methacrifos 918 1 18 28 29 24 % 5-2 12 1 56 8 99 11 1 Library Hit: Methacrifos 952 165 18 28 29 24 % 5 62 99 11 127-2 35 5 1 15 2 3 18 28 29 24 % 5-2 35 5 1 15 2 3 Figure 5. Mass spectra of methacrifos obtained with (A) the ExtractaBrite ion source and (B) the AEI source 62 99 11 127 18 28 29 24 Terpenes Identification of terpenes is an important GC-MS application especially in the food and flavors industry. Many terpenes are structurally similar and therefore have very similar fragmentation patterns and common fragments; hence it is of paramount importance that GC-MS gives a high library matching score to distinguish the correct terpene and provide a reliable compound identification. Figure 6 shows the NIST match factors (SI) and reverse match factors (RSI) for various terpenes that were tested by individual injections on the ISQ 7 system with the AEI source. All the terpenes injected onto the system returned as the first hit following the library search with forward and reverse scores above 9. Figure 7 shows an example of AEI mass spectra for linalool compared against the Wiley s FFNSC MS spectral library. Figure 8 shows the match factor scores for 16 consecutive injections for linalool showing excellent repeatability from injection to injection with all of them showing greater than 95 for the SI scores. Match Scores Figure 6. NIST 17 mass spectral library match factors for terpenes tested with the AEI source 6

12 1 % 5-2 12 Library Hit: Linalool 959 1 41 55 % 5 69 8 67 42 53 92 121 59 72 77 82 17 136-2 35 6 8 1 12 14 16 18 2 Figure 7. Mass spectral fragmentation pattern for linalool and its comparison with Wiley s FFNSC, 2ed. Library Match factor (SI) score for Linalool 43 55 42 1 95 9 85 8 53 57 69 67 68 8 82 92 SI average value (n=16) = 955 7 1 2 3 4 5 6 7 8 9 1 11 12 13 14 15 16 Injection number Figure 8. Match factor (SI) repeatability for linalool using the AEI ion source Conclusions The results of these studies demonstrate that: 17 The AEI source is designed to deliver higher sensitivity compared to the conventional EI source. It shows the best performances at an optimum electron voltage setting of 45 ev without compromising on spectral results. 121 136 Mass spectra obtained using the AEI source showed a match that was excellent for the majority of the compounds analyzed. The minimum value of the forward match score was >87 for all compounds analyzed. NIST match scores obtained with the AEI source were found to be equivalent to the ones obtained with a traditional EI source. The spectral match scores obtained for the 16 terpenes analyzed using the AEI source were in excellent agreement with those contained within the Wiley and NIST 17 libraries, in all instances returning the correct identification as the first hit. Taken together, these results indicate that the ISQ 7 GC-MS system configured with the AEI or with the Extractabrite ion sources delivers in both cases very high quality mass spectra. In both configurations, the ISQ 7 GC-MS can be confidently used for identification of untargeted compounds in screening experiments using the available commercial mass spectral libraries. The added advantage of the AEI source to increase ionization efficiency and extend the robustness towards difficult matrices, offers the possibility to expand the analytical capability of the laboratory. The ISQ 7 GC-MS system configured with AEI source is therefore a valuable solution any time ultra sensitivity and robustness is required to respond to challenging analytical needs. References 1. Sparkman, O. D.; Penton, Z. E.; Kitson, F. G. Gas Chromatography and Mass Spectrometry A Practical Guide; Academic Press, 211. 2. Basis for Interpretation of the Library Search Results, NIST/EPA/NIH Mass Spectral Library (NIST 17) and NIST Mass Spectral Search Program (Version 2.3) user s guide, page 41. 3. Thermo Fisher Scientific Technical Note TN 1597: Practical determination and validation of instrument detection limit for the Thermo Scientific ISQ 7 Single Quadrupole GC-MS system with Advanced Electron Ionization source, 218. 4. Watson, J. T.; Sparkman, O. D. Introduction to Mass Spectrometry; Wiley, 28, page 318 319. 5. Stein, S. Mass spectral reference libraries: An ever-expanding resource for chemical identification, Anal. Chem. 212, 84, 7274 7282. 6. APPENDIX 7: Contributing EI Mass Spectra to the NIST/EPA/NIH Mass Spectral Library in NIST Standard Reference Database 1A, NIST/EPA/NIH Mass Spectral Library (NIST 17) and NIST Mass Spectral Search Program (Version 2.3) user s guide. Find out more at thermofisher.com/isq7 218 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries unless otherwise specified. NIST is a registered trademark of The National Institute of Standards and Technology, Agency of the Government of the United States. Wiley is a registered trademark of John Wiley & Sons, Inc. Restek and MegaMix are registered trademarks of Restek Corp. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. TN1598-EN 218S