OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Similar documents
Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

METROTEC GSM. Type GSM. Oxygen measurement and control unit. *** Version 1.6 ***

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Columbus Instruments

Technical Data Sheet MF010-O-LC

Instruction manual. for. high purity gas pigtails 200 bar / 300 bar

DPC-30 DPC-100. Reference Manual

testo Leakage detector for refrigerants Instruction manual

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

The New PHD 4 Portable Helium Detector

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

NTF-515. Operation Manual

GasLab SAN-102 Micro Oxygen Monitor User Manual

EC214 EC215 - EC215R Bench Conductivity Meters

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

OPERATION MANUAL NTF-15

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

Ventam 85 Installation & Commissioning Instructions

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

OPERATION MANUAL NTF-60 Plus

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

GAS ANALYSER OXYBABY 6.0 for O 2. or O 2 /CO 2

SAUTER FK Version /2016 GB

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

SPECIFICATIONS APCEPH1

Mass Flow Controller (MFC) for Gases

NTF-230 OPERATION MANUAL

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

HI 2314 HI 2315 HI 23151

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Discovery HP-TGA 75/750. Site Preparation Guide

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

GasLab SAN-102 Micro Oxygen Monitor User Manual

ACV-10 Automatic Control Valve

METROTEC GSM-V6. Typ GSM. Oxygen measurement and control unit. *** Version 1.5 ***

Operation Manual. Pro Trimix Alarm TM. Helium & Oxygen Gas Analyzer. Rev

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

CENTRAL GAS SUPPLY SYSTEMS : CYLINDER BATTERIES AND HIGH PRESSURE HOSES

Instructions for Use

Oxygen Flue Gas Analyzer BA 2000

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

NEULOG OXYGEN LOGGER SENSOR GUIDE

NIV EAU MATIC ST User Manual Made in Canada

NGN Series nitrogen generator

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Operating Instructions Part No

Hand-held pressure calibrator with integrated pump Model CPH6600

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Tank Blanketing Pressure Regulators RHPS Series

Heavy Duty Dissolved Oxygen Meter

PRESSURE REGULATORS AND FLOWMETERS, ALL MEDICAL GASES

The affordable consumption counter for compressed air and gases R 3 /4 (DN 20)

Quick Installation Guide Doc nr.: C Date:

Mass Flow Meter (MFM) for Gases

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

The Univentor 1250 Anaesthesia Unit

Instruction Manual Dräger MSI P7 and MSI P7 plus

Model Description

COM280. Technical Information. Oxygen measurement Battery-driven pocket meter for dissolved oxygen measurement

450P- RV AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

This test shall be carried out on all vehicles equipped with open type traction batteries.

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

450P AUTOMATIC PORTABLE COMPRESSOR EXTREME SERIES

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Flow VA 520. incl. temperature measurement. Intelligent solutions for ac- for compressed air and gases R 1/4 (DN 8) DN 15 R 3/4 (DN 20) DN 25

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

combustion efficiency monitoring General Purpose Applications Flue gas oxygen analyzers Combustion & Environmental Monitoring An Company

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

PM105 PRESSURE METER USER S MANUAL

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

combustion efficiency monitoring Flue gas oxygen analyzers General Purpose Applications Combustion & Environmental Monitoring

MODEL CALIBRATION GAS DELIVERY SYSTEM

Purge Star & Ring Purge Systems

SDX Submersible Depth Transmitter User Manual

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Gas Sampling Fitting with MK2.1 and L-Probe

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Nitrogen Advantage. Better Tire Pressure Retention. Improved Fuel Economy. Extended Tread Life. Elimination of Oxidation. Reduction in Tire Failures

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Installation and Operation Manual

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

TECHNICAL DATA CAUTION

Mass Flowmeters. Mass Flowmeters for Gases

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

Operating Instructions Part No

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

PURA. Pure Gas Dewpoint Transmitter. Users Guide

User Manual. GPL 3000 e. gas detector

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Transcription:

OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1

INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle of Measurement 5 Measurement Electronics 6 Sensor 6 GENERAL DESIGN 7 Top View 7 Side View 8 Front View 8 Display 9 OPERATION 10 Turning on the Instrument 10 Program Functions 11 Programming 11 Measuring 12 RECHARGING OF THE BATTERY PACK 13 Recharging Instructions 13 Docking Station for Recharging 14 Status of Battery 15 Energy Saving 15 Replacing Battery Pack 15 Manual Integral 2006 2

Switching Unit Off 16 Measuring Different Gases 17 Filter Life 18 TECHNICAL DATA 18 Technical Data Sensor 18 Technical Data Battery Charger 18 IMPORTANT INFORMATION 19 NOTES 19 Manual Integral 2006 3

Preface This instrument determines the partial pressure of oxygen in gas atmospheres by using an oxygen sensor. Such sensors operate at very high temperatures. Precautions must be taken to ensure that no flammable gas mixtures reach either the sensor or the instrument. Should the ceramic portion of the sensor be damaged, the gas measured may escape or may be diluted by air. Precautions must be taken to avoid damaging either components or the environment. Damage to the equipment and false readings can arise due to incorrect operation, leaks, corrosion, condensation, etc. Regular servicing of all components is essential. The oxygen measuring instruments and their accessories are built and tested under quality control standards in accordance with DIN EN-ISO 9001. The instruments must be installed and used in compliance with all applicable local and specific regulations. The equipment must be periodically checked for accuracy and function depending on its degree of use, and should be checked in accordance with the calibration and test instructions with which it was supplied. This measuring instrument is not suitable for use with explosive gas mixtures, especially with pure hydrogen. There is a risk of explosion. Manual Integral 2006 4

Introduction Operating Manual OXY Integral Principle of measurement Oxygen measuring instruments are designed to process the output from an oxygen sensor made of stabilised zirconium oxide. This material is a ceramic, also known as a solid electrolyte, and is ideally suited at high temperatures as an oxygen ion conductor. Such ion conductors within a temperature range, depending on the composition of the material, possess the capability to fill empty spaces in their crystal structure with oxygen ions. The oxygen ions attach themselves to a conductive contact surface most commonly platinum. The degree of oxygen concentration in a gas when measuring determines the degree of oxygen activity and thus the quantity of oxygen ions. A sensor consists in principle of a solid electrolyte, which is exposed to gas on both sides. One side is in contact with a reference gas, air for example, the other with the gas to be measured. The physical construction of the sensor keeps the two gases apart, so that any mixing is prevented. In applications as described above both heated and unheated sensors are used. Unheated sensors are normally used with ovens. Where gases are under 600 C, a heated sensor is used. Heated sensors require a minimum temperature of between 500 C- 650 C as required by the measurement principle. Heated sensors are held at a set temperature by a controller built into the electronic processor. The temperature of both heated and unheated sensors is measured by the processor and is an essential component in the calculation of the oxygen content (oxygen partial pressure). Manual Integral 2006 5

Electronics The electronics of the "Oxy - Integral " are capable of the following functions: Measuring accurately the rest oxygen component in inert gases such as argon and helium as well as nitrogen. The setting for the type of gas used is chosen using the P button in combination with the Plus+ button or the Minus button (see page 17) The maximum hydrogen content in the purge gas cannot exceed 10%. Forming gases are gas mixtures of nitrogen and hydrogen. This type of sensor cannot accurately measure rest oxygen levels in these types of gases. What can be shown is a tendency and an accurate value. Sensor The built in sensor includes the measuring element of platinized zirconium oxide, the heating element and the heating controller. The heating element is required to keep the sensor at approximately 550 C. Manual Integral 2006 6

General Design Top View Filter Gas Inlet Hook Ring Gas Outlet Housing Display Operational & Programming Buttons Manual Integral 2006 7

Side View Cable Battery Charger (115 V / 230 V) Front View Digital Interface (Data Transfer, Extra Charging Possibilities through an Orbital Welding Unit) Gas Outlet Gas Inlet Manual Integral 2006 8

Display 3 Digit Display (Max. Measuring Range: 999 PPM) Plus+ (Increases Upper Limit Value after Pressing the P Button) Minus- (Decreases Upper Limit Value after pressing the P Button) P (Button to Program Upper Limit Value or Select Gas Types) Pump (On and Off Button for the Integrated Pump) ON/OFF Manual Integral 2006 9

Operation Turning on the Unit Operating Manual OXY Integral Pressing and holding down the ON/OFF button for approx. 1 second will turn on the instrument. The element heats up the sensor, which needs to be brought up to a specific temperature for accurate operation. During this warm-up phase no measurement is possible. The digital display will rise slowly until it reaches 999. If the Pump button is pressed during the warm up phase, the display will show 3 small letters ccc, indicating the sensor is not ready to use. After the warm up phase the display will indicate the maximum value 999. This will take between 45 seconds to 1 minute. The unit is now fully operational. Manual Integral 2006 10

Program Functions When using the OXY Integral it is possible to set a desired upper limit within the entire measuring range, which is 0 999 PPM. The operator will be alerted through the flashing of the display if the value displayed is higher than the upper limit. When the setting has been reached the display will no longer flash and welding can proceed. Important: During the welding procedure the probe must be removed! The sensor will not function during the welding procedure and the displayed value shown during that period does not correspond to the actual oxygen content. Programming 1. Press the P button (3 dots will flash and after that an upper limit setting, for example 20 PPM will be displayed). While in the programming function there will be 3 dots behind the numbers as in picture 1. Picture 1 Picture 2 2. To change the upper limit, press the Plus+ or the Minus- buttons until the desired limit is selected, for example push the Plus+ once and the value goes to 21 PPM as in picture 2. 3. Press the P button to confirm programmed upper limit setting. This setting is now retained in the unit s processor. This upper limit remains even if the unit is switched off or if the battery is replaced. Manual Integral 2006 11

Measuring After the zirconium oxide sensor has reached operating temperature the unit is fully functional. An upper limit setting for QC purposes can be programmed through the P button, which is recommended but not necessary for measuring purposes. The probe at the end of the sensor s hose can be inserted into the purge chamber or root gap and the Pump button is pressed. Now the sample gas is drawn through the oxygen sensor hose (approx. 50l/h) into the unit for analysis. Important: During the welding procedure the probe must be removed! The sensor will not function during the welding procedure and the displayed value shown during that period does not correspond to the actual oxygen content. Manual Integral 2006 12

Recharging of the Battery Pack Recharging This oxygen analyser is powered by internal battery pack (2300 ma) with a life between charges of approx. 3 4 hours. When replacing a complete battery pack, use batteries from the same manufacturer with the same charge capacity. All five batteries must be replaced at the same time. Important: Only use rechargeable batteries! When the battery pack is low and recharging is necessary, three dots in a row flash from right to left. This feature displays when the battery depletion is approaching 45% of capacity. This suggests that the operator return the unit to the supplied docking station for recharging. The unit is fully functional even while in the docking station. Important: The sensor must be turned on during the charging phase! If the sensor is not turned on, the charging station only provides a small maintenance charge of approx. 50 ma in order to prevent further battery depletion. The sensor also has an energy saving feature to extend battery life and internal components. After 10 minutes of inactivity the unit will shut itself down unless the unit is charging or the pump is on. Recharging Necessary (Flashing three dots from right to left) Charging on the Docking Station (Flashing three dots from left to right) Manual Integral 2006 13

Docking Station for Recharging This oxygen analyser is a portable handheld battery operated unit, which makes it excellent for job site work. After use the unit is placed back in the docking station so that the battery pack can be recharged. The unit is placed into the middle of the docking station where a receptacle with spring loaded clamps attaches itself to the metallic contacts on the back of the sensor. Important: Ensure that the sensor is seated correctly into the docking station; a misaligned sensor will not charge! Spring loaded clamps in the middle of the docking station. Switching to the local power requirements either 230 V / 115 V Important: Before opening docking station, disconnect power supply! Unit in picture is using 230 V Power; This is visible by the number on the switch (see arrow). Manual Integral 2006 14

Status of Battery Pack Energy Saving By pressing the Plus+ and the Minus button at the same time the status of the battery pack can be viewed. At approx. 45% capacity the unit will indicate that it needs recharging by flashing three dots in a row from right to left. In order to extend battery life, especially if the unit is accidentally left on, the unit is programmed with an energy saving function. If within 10 minutes of non-use or not being placed back in the docking station, the unit will shut itself down. The sensor is fully functional again after it has been turned on for approx. 1 minute. This feature extends battery life and internal components. Changing Battery Pack Remove the black end cap by pressing the side tabs and pulling. Then by pressing up on the grey tab on the battery pack and pulling, the battery housing is removed and batteries can be inserted. Important: When replacing battery pack please note polarity! Manual Integral 2006 15

Turning Unit OFF By pressing and holding down the ON/OFF button continuously for 3 seconds the unit shuts off. This delay prevents turning the unit off accidentally. When pressing the ON/OFF button continuously for the 3 seconds the display indicates 8.8.8. and turns off. Manual Integral 2006 16

Measuring Different Gases Press and hold down the P button and the Minus- button at the same time to program for the measurement of rest oxygen in Argon. Display will read 0 as shown above. Press and hold down the P button and the Plus+ button at the same time to program for the measurement of rest oxygen in Forming Gases. Display will read 1 as shown above. Note: Factory default for this unit is always argon! In most cases the unit is exposed to situations where argon is used. To prevent errors the unit always resets to the default, which is argon, when the unit is shut off. If hydrogen based forming gases are used the selection must be made again (see above). In principle the sensor is not suited to measure oxygen content in forming gases such as hydrogen mixtures and atmosphere. When using the oxygen analyser to measure rest oxygen in forming gases the following is to be noted and followed: Maximum concentration of H2 is 10 %. The reaction time of the sensor on the H2 elements is much longer than with inert gas such as argon, nitrogen and helium. When switching from H2 mixtures back to inert gases, the sensor is contaminated for a while. To purge the unit, turn pump on and let regular air be sucked through the unit. Unlike inert gases, which can be measured accurately, rest oxygen cannot be accurately measured with forming gases. This is why a small test weld to visually compare rest oxygen exposure enables what the indicator displays to be used as a reference point. Important: The value displayed when measuring forming gases indicates a tendency and not a fixed value. Important: During the welding procedure the probe must be removed! The sensor will not function properly during the welding procedure and the value shown during that period does not correspond to the actual accurate oxygen content. Manual Integral 2006 17

Filter Life To measure rest oxygen in the sample gas the unit requires a minimum flow of approx. 50 l/h. If this amount is not delivered to the sensor the unit is not able to measure accurately. The most common cause for this is if the filter is plugged. A flashing ccc indicates the sensor is measuring below the minimum flow required. The value of the sampling measurement as well as the flashing of the ccc will alternate approx. every 100 ms. This situation will not stop the measurement from taking place but make the operator aware that the filter needs changing. The value displayed may not be accurate and could possibly be higher than indicated. Important: Check hose and connections for leaks when changing filter. Technical Data Measurement Instrument (Dimensions and Weight) Length 290 mm 11 3/8 Width 130 mm 5 1/8 Height 60 mm 2 3/8 Weight 0, 8 kg 1.8 lbs. Power Supply: 6 V Digital Interface for documentation Interface is compatible with Orbital Welding System TIGTRONIC Compact Docking Station (Dimensions and Weight) Length 140 mm 5 ½ Width 125 mm 5 Height 80 mm 3 3/16 Weight 0, 5 kg 1.1 lbs. Power Supply: 115 V / 230 V Manual Integral 2006 18

Important Information Do not insert probe into high-pressure lines or attempt to use in strong vacuum environments. Do not open docking station when power cord is still connected to power supply. Battery packs should be replaced as a component (battery packs can be ordered from INTERCON). Do not measure without the filter (no warranty). Use only original filter from INTERCON. Avoid exposure to organic gases such as melting plastic. These gases can destroy the sensor. Do not measure gases with hydrogen content higher than 10%. RISK OF EXPLOSION! Notes: Manual Integral 2006 19