DEVELOPMENT AND VALIDATION OF RP HPLC METHOD TO DETERMINE NANDROLONE PHENYLPROPIONATE IN DIFFERENT PHARMACEUTICAL FORMULATIONS

Similar documents
A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

IJPAR Vol.6 Issue 1 Jan - Mar Journal Home page:

SIMPLE, EFFICIENT AND SIMULTANEOUS DETERMINATION OF ANABOLIC STEROIDS IN HUMAN URINE

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

FRANZ CELL TEST OF NANO GLUTATHIONE WITH HPLC ANALYSIS

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Extended Application Note

CHAPTER 3 Development and Validation of a RP-HPLC method for Dutasteride and its impurities in bulk drug

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

Chemical Analysis for Methyltestosterone

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

A VALIDATED STABILITY-INDICATING RP-HPLC ASSAY METHOD FOR BOLDENONE UNDECYLENATE AND ITS RELATED SUBSTANCES

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Sensitive Analysis of Genotoxins by HPLC-ECD Using Boron-Doped Diamond Electrochemical Cell

The Application of QuEChERS in the Extraction of Anabolic Steroids in Whole Blood

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

CORESTA RECOMMENDED METHOD Nº 67

Agilent Dimension Software for ELSD User Manual

Improved Performance in Capillary Electrophoresis using Internal Standards

May 1, By John Heim,Doug Staples

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

SOP: Buck Scientific BLC-20P HPLC Operation

Rapid and Reliable Detection of Dissolved Gases in Water

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

Syed Ghulam Musharraf 1,2* and Umair Gulzar 2

Gas chromatographic determination of 1,4-dioxane at the

Extend the time window for the detection of low level anabolic androgenic steroids and their metabolites

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

APPLICATION NOTE. GC Integrated Permeation Device

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Analysis of Melamine and Cyanuric Acid in Food Matrices by LC-MS/MS

Carrier Gases in Capillary GC

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

The instrument should have a mass range from 2 to 2000 m/z or better Interface

Development of a Clinical Research Method for the Measurement of Testosterone. Dihydrotestosterone

Setting a New Standard in Flash Chromatography Performance

International Journal of Research and Development in Pharmacy and Life Sciences. Research Article

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

TESTOSTERONE AND ESTERS* Latest Revision: June 23, 2005

RP-HPLC Method Development and Validation for Nitroxynil in Active Pharmaceutical Ingredient Manufacturing.

Analysis of Deuterium Enrichment by Fourier Transform Infrared Spectrometry (FTIR): Practice. Christine Slater PhD

ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Dissolution test results showing biorelevant media made from SIF Powder Original is the same as media prepared using methylene chloride

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

WADA Technical Document TD2014EAAS. Endogenous Anabolic Androgenic Steroids Measurement and Reporting

Svensk läkemedelsstandard

Retention Time Locking: Concepts and Applications. Application

Now it is. easy to switch. CHS TM Steroid Profiling kit and software. Brochure not for distribution in the USA and Canada

Chapter Pipette service & maintenance. Pipette specifications according to ISO Repair in the lab or return for service?

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

SOP: Derivatized testosterone_01 Revision: 01 Date Effective: 05/17/15

International Journal of Theoretical & Applied Sciences, 4(2): (2012)

A Depletion Compensated Wet Bath Simulator For Calibrating Evidential Breath Alcohol Analyzers

Effects of Dissolved Gases in Surfactant Dissolution Media

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Integration of amino acid, acylcarnitine and steroids analysis in single FIA/LC-MS/MS platform

Automated Determination of Dissolved Gases in Water Anne Jurek

Determination of Selected ASTM Perfluorinated Compounds (PFCs) Using the Shimadzu LCMS No. LCMS-083. No. SSI-LCMS-083

Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)

LEVANT WORMSEED FOR HOMOEOPATHIC PREPARATIONS CINA FOR HOMOEOPATHIC PREPARATIONS

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

GCMSD-Headspace Analysis SOP

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

Novel empirical correlations for estimation of bubble point pressure, saturated viscosity and gas solubility of crude oils

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

Applications of a Magnetic Sector Process Mass Spectrometer to the Analysis of Variable Vacuum Samples

Carrier Gases in Capillary GC

HUMAN IL6 KITS PROTOCOL

AKTA pure 25 New Owner s Intro

Linezolid 600 mg Tablets WHOPAR part 6 February 2017 (Hetero Labs Limited), TB299

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

LCMSMS Solutions For Steroid Analysis

Experiment GC : Analysis of BTEX by GC-FID

CHM250 Calibration and Measurement Lab. Balance Calibration

Increasing the productivity of ICP-OES with the Thermo Scientific Qtegra ISDS Software

SAM510: SAM Methyltransferase Assay A Non Radioactive Colorimetric Continuous Enzyme Assay

Introduction. Experimental

Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

INTRODUCTION TO THE SPECTROPHOTOMETER AND PIPETTING SKILLS

Analysis of Casein and Whey Protein in Whole, 2%, and Skim Milk by Capillary Gel Electrophoresis

The instrument will process the samples one by one till all the samples have been cleaned up.

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

Int. Res J Pharm. App Sci., 2012; 2(6): ISSN:

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

More Speed and Resolution for Current Methods

Applied Technology and Best Practices in CEE. Conference

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

Transcription:

Acta Poloniae Pharmaceutica ñ Drug Research, Vol. 68 No. 2 pp. 155ñ160, 2011 ISSN 0001-6837 Polish Pharmaceutical Society ANALYSIS DEVELOPMENT AND VALIDATION OF RP HPLC METHOD TO DETERMINE NANDROLONE PHENYLPROPIONATE IN DIFFERENT PHARMACEUTICAL FORMULATIONS JAYANTI MUKHERJEE, AYAN DAS, UDAY SANKAR CHAKRABARTY, BIJAY KUMAR SAHOO, GOUTAM DEY, HIRA CHOUDHURY and TAPAN KUMAR PAL* Bioequivalence Study Centre, Department of Pharmaceutical Technology, Jadavpur University, Kolkata 700 032, India Abstract: This study describes development and subsequent validation of a reversed phase high performance liquid chromatographic (RP-HPLC) method for the estimation of nandrolone phenylpropionate, an anabolic steroid, in bulk drug, in conventional parenteral dosage formulation and in prepared nanoparticle dosage form. The chromatographic system consisted of a Luna Phenomenex, CN (250 mm 4.6 mm, 5 µm) column, an isocratic mobile phase comprising 10 mm phosphate buffer and acetonitrile (50:50, v/v) and UV detection at 240 nm. Nandrolone phenylpropionate was eluted about 6.3 min with no interfering peaks of excipients used for the preparation of dosage forms. The method was linear over the range from 0.050 to 25 µg/ml in raw drug (r 2 = 0.9994). The intra-day and inter-day precision values were in the range of 0.219ñ0.609% and 0.441ñ0.875%, respectively. Limits of detection and quantitation were 0.010 µg/ml and 0.050 µg/ml, respectively. The results were validated according to International Conference on Harmonization (ICH) guidelines in parenteral and prepared nanoparticle formulation. The validated HPLC method is simple, sensitive, precise, accurate and reproducible. Keywords: nandrolone phenylpropionate, RP-HPLC, anabolic-androgenic steroid, nanoparticles Figure 1. Nandrolone phenylpropionate (CAS number 62-90-8) Nandrolone phenylpropionate (NPP) (Fig. 1), 17β-hydroxyestr-4-en-3-one 3-phenylpropionate (1), is an anabolic-androgenic steroid. The low androgenicity and enhanced anabolic activity of NPP has shown to positively influence calcium metabolism and to increase bone mass in osteoporosis. In women with disemminated mammary carcinoma, this has been reported to produce objective regressions for many months. Furthermore, it has a nitrogen-saving action. This effect on protein metabolism has been established by metabolic studies and is utilized therapeutically in conditions where a protein deficiency exists such as during chronic debilitating diseases and after major surgery and severe trauma (2, 3). Nandrolone has been reported to improve respiratory muscle function in COPD (Chronic Obstructive Pulmonary Disease) (4). The administration of anabolic androgenic steroids increases skeletal muscle mass (hypertrophy) and protein synthesis (5). NPP is commercially available as injections, a conventional pharmaceutical formulation. The UV spectrophotometric method of assay has been reported in British Pharmacopoeia and United States Pharmacopoeia (6, 7). Analysis of nandrolone esters in oily injections to discriminate them from nandrolone alcohol, a potential impurity, by reversedphase high performance liquid chromatography has * Corresponding author: e-mail: proftkpal@gmail.com; phone: +91-33-24146967 (office); +91-9830036297 (mobile); fax: 91-33-24146186 155

156 JAYANTI MUKHERJEE et al. also been reported (8). The work on the development and validation of a multi-residue method for the determination of a wide range of hormonal anabolic compounds in hair using gas chromatographytandem mass spectrometry had been carried out (9). Findings on the effect of side chain, the injection site and the volume of the injection on the pharmacokinetics and pharmacodynamics of nandrolone esters in an oil vehicle in men is also available (10). However, HPLC assay for determination of NPP in any novel drug delivery system has not yet been reported and therefore, this study claims to be a novel one. The objective of the present study was to develop a simple, precise, accurate and validated RP-HPLC method for the estimation and routine analysis of NPP in nanoparticulate drug delivery system and also in commercially available injectables. The results of the analysis were validated according to ICH guidelines (11). EXPERIMENTAL Chemicals and reagents NPP was obtained from Govind Laboratories Pvt. Ltd., Ahmedabad, Gujarat. Injection of brand Durabolin (25 mg/ml) (Batch No. 013895), manufactured by Organon, Kolkata, West Bengal, India were procured from a local pharmacy. Dextran sulfate (MW 500,000), mannitol, and zinc sulfate were all purchased from Sigma Chemical (St. Louis, MO, USA). Chitosan (MW 30,000 with 95% deacetylation) was purchased from Aquapremier, Bangkok, Thailand. Acetonitrile and water of HPLC grade were purchased from Merck, India. All other chemicals were of analytical grade and purchased from Merck, India. Milli-Q water was used for the mobile phase preparation. Cellulose ester dialysis membrane tubing with a molecular weight cutoff (MWCO) of 1,000,000 (Spectra/Por CE) was purchased from Fischer Scientific (Chicago, IL, USA). Preparation and characterization of nanoparticles Nanoparticles preparation The preparation of nanoparticles was done by polyelectrolyte complexation at room temperature (12). The required amount of aqueous solutions of polymers and drug in ethyl alcohol was stirred at 600 rpm. Then, the aqueous chitosan solution was added dropwise and the resulting nanoparticles were stirred for 5 min. A few milliliters of zinc sulfate solution (1 M) were added as a stabilizer. The resulting stabilized NPP nanoparticles were dialyzed for 24 h in the dark (as the drug is light sensitive) against DI water using Spectra/Por CE MWCO 1,000,000 dialysis membrane. Mannitol was added as lyoprotectant to the purified, loaded nanoparticles during lyophilization (13, 14). Particle size determination The particle size was determined by laser diffractometry using Mastersizer 2000 (Malvern laser diffraction particle size analyzer, Malvern Instruments, U.K.). The Mastersizer comprises of helium-neon laser as a light source. This is then focused by a Fourier lens to a detector, which consists of a large number of photosensitive elements radiating outward from the center. The intensity of the scattered light is measured. The volumetric particle size distribution and specific surface area is calculated using an optical model and mathematical deconvolution procedure. For particle size analysis, the samples were dispersed in Milli-Q water. Measurements were carried out at 30 O C using a helium-neon laser at an angle of 90 O. Instrumentation and chromatography The HPLC system was of Knauer (Berlin, Germany), consisting of a solvent delivery pump (LC-1000 Knauer), a Rheodyne injector and 2500 variable wavelength UV/VIS detector with EZChrom (version 3.1.6) software for integration. Separation was achieved using a Luna Phenomenex CN column (250 mm 4.6 mm, 5 µm). The isocratic mobile phase pumped at a flow rate of 1 ml/min consisted of 10 mm phosphate buffer and acetonitrile (50:50, v/v) freshly prepared, filtered through 0.45 µm filter (Millipore, Milford, MA, USA) and degassed by sonication for 15 min. The injection volume was 50 µl and the wavelength used was at 240 nm. All separations were performed at room temperature. Standard and sample preparation Standard preparation Stock solution (1 mg/ml) of NPP was prepared in ethyl alcohol. From this stock, a solution of concentration 10 µg/ml was prepared with the mobile phase. Standard solutions of NPP (0.05, 0.1, 0.25, 0.5, 1.0, 2.5, 5, 10, 15, 20 and 25 µg/ml) were prepared by subsequent dilution using the mobile phase. Sample preparation for assay and recovery study For the analysis of the dosage form, Durabolin injection manufactured by Organon Ltd. (India) was selected. The content of the ampoule was poured in a 250 ml volumetric flask and ethyl alcohol (ana-

Development and validation of RP-HPLC method to determine nandrolone... 157 lytical grade) was added and shaken thoroughly to extract the drug from the excipients. Dried powdered sample equivalent to 10 mg NPP was dissolved in about 40 ml of ethanol in a separate 100 ml volumetric flask and shaken for 2 h. Both the solutions were filtered through Whatman filter paper (No. 41). A volume of 1 ml of each sample was pipetted out and the final volume was made up to 10 ml with mobile phase to prepare a concentration of 10 µg/ml. After suitable dilution, the drug concentration with six replicates for the parenteral solutions, as well as solutions of prepared dosage form was determined by HPLC using the calibration curve. The recoveries were determined by adding known amounts of NPP reference substance (2.5, 5.0, 7.5 and 10.0 µg) to the formulated samples. These samples were compared with the standard solutions (i.e., 2.5, 5.0, 7.5 and 10.0 µg). Method validation The method was validated according to the ICH guidelines for the validation of analytical procedures. The parameters which were used to validate the method of analysis were linearity range, limit of detection (LOD), limit of quantification (LOQ), specificity, accuracy, precision and recovery. RESULTS AND DISCUSSION Nanoparticles were successfully prepared by polyelectrolyte complexation technique. The speed was kept constant at 600 rpm. to obtain the desired results concerning average particle size and narrow particle size distribution profile. The model formulation used in the proposed RP-HPLC analysis showed an average particle diameter of 530.97 nm. The chromatographic conditions were optimized and separation was performed on a Luna Phenomenex CN column using a mobile phase consisting of phosphate buffer and acetonitrile. The proposed mobile phase composition allowed suitable retention time of NPP and good selectivity towards interference from the excipients of the formulation. Calibration curve was constructed using standard NPP solutions in the range of 0.05ñ25 µg/ml. The linearity of the calibration curve (Fig. 2) was validated by high value of correlation coefficient (r 2 = 0.9994). Under the chromatographic conditions described, NPP was eluted after about 6.3 min. Good baseline resolutions and peak shape can be observed (Fig. 3). The limit of detection (LOD) and the limit of quantitation (LOQ) were 0.010 and 0.05 µg/ml, respectively. Low values of LOD and LOQ indicate that the method is sensitive. The specificity of the proposed method demonstrated that the excipients present in market formulation and also in the prepared formulation do not interfere with the drug peak. Furthermore, wellresolved peaks indicate the specificity of the method (Fig. 3). Thus the proposed method is useful to Figure 2. Linearity curve of nandrolone phenylpropionate

158 JAYANTI MUKHERJEE et al. Figure 3. (A) Chromatogram of nandrolone phenylpropionate standard solution, (B) chromatogram of NPP in parenteral dosage form, (C) chromatogram of NPP in nanoparticle solution

Development and validation of RP-HPLC method to determine nandrolone... 159 Table 1. Intraday precision and accuracy. Concentration Mean Accuracy Precision Relative ± SD* (µg/ml) (n = 6) (%) (% RSD)** error 8.0 8.03 0.03 101.40 0.31 1.40 8.5 8.52 0.05 100.28 0.61 0.28 9.0 9.14 0.04 101.56 0.47 1.56 9.5 9.60 0.04 101.04 0.43 1.04 10.0 10.03 0.02 100.34 0.22 0.34 10.5 10.51 0.02 100.12 0.26 0.12 11.0 11.10 0.03 100.99 0.40 0.99 11.5 11.58 0.05 100.69 0.44 0.70 12.0 12.13 0.05 101.04 0.44 1.05 * Standard deviation ; ** relative standard deviation Table 2. Interday precision and accuracy. Concentration Mean Accuracy Precision Relative ± SD* (µg/ml) (n = 6) (%) (% RSD)** error 8.0 8.05 0.04 100.64 0.54 0.64 8.5 8.56 0.06 100.69 0.73 0.69 9.0 9.09 0.08 101.01 0.88 1.01 9.5 9.56 0.05 100.62 0.54 0.62 10.0 10.09 0.05 100.93 0.53 0.93 10.5 10.56 0.08 100.59 0.78 0.59 11.0 11.12 0.05 101.05 0.44 1.05 11.5 11.58 0.06 100.68 0.53 0.67 12.0 12.08 0.06 100.66 0.50 0.66 * Standard deviation ; ** relative standard deviation Table 3. Assay of pharmaceutical formulations. Sample Assay (n = 6) (%) ± SD* RSD** Durabolin (Injection) 02.48 0.65 0.63 Formulation 100.87 0.71 0.71 * Standard deviation ; ** relative standard deviation Table 4. Recovery studies in different formulations. Dosage Mean recovery form (n = 6) ± SD* RSD** Durabolin Injection 98.91 % 0.80 0.81 Formulation 97.84 % 0.93 0.95 * Standard deviation; ** relative standard deviation quantify NPP in different pharmaceutical formulations. The precision was determined by analyzing nine different concentrations of the bulk drug on three different days. The intra-day and inter-day precisions were calculated as 0.218ñ0.875% and 0.441ñ0.875%, respectively, whereas the accuracies were calculated as 100.284ñ101.564% and 100.587ñ101.053% as shown in Table 1 and 2, respectively. The recovery results are very close to 100%, which prove the suitability and accuracy of the proposed method for both the formulations. The accuracy of both the formulations was tested by assaying the marketed formulation and the

160 JAYANTI MUKHERJEE et al. prepared nanoparticles on the same day and three different days. The results of formulation assay are shown in Table 3. CONCLUSION The described HPLC method provides simple, sensitive, precise, accurate and reproducible quantitative method for routine analysis of NPP in conventional dosage form such as injections as well as in novel drug delivery system like nanoparticles. Simple sample preparation procedure and a retention time less than 10 min make this method unique and suitable for analyzing a number of samples in a short period of time. The formulations assays were within the acceptable limit. High percentage of recovery (Table 4) ensures the method to be free from interference of excipients and additives, used in the formulations. ACKNOWLEDGMENT: One of the authors, Jayanti Mukherjee, is grateful to All India Council of Technical Education (AICTE), Govt. of India, for financial assistance and to Bioequivalence Study Centre, Jadavpur University, Kolkata, India to carry out this entire research work. The authors are also grateful to Govind Laboratories, Ahmedabad, Gujarat, India for sending nandrolone phenylpropionate as a gift sample. REFERENCES 1. The Merck Index, An encyclopedia of chemicals, drugs and biologicals, 14 th ed. OíNeil M.J. Ed., 6364 (2006). 2. Griggs R., Kingston W., Jozefowicz R., Her B., Forbes G., Halliday D.: J. Appl. Physiol. 66, 498 (1989). 3. Haupt H., Rovere G.: Am. J. Sports Med. 12, 469 (1984). 4. Wijnhoven H., Ennen L., Rodenburg R., Dekhuijzen PN.: Int. J. COPD. 1, 83 (2006). 5. Bagchus W., Smeets J., Verheul H., Jager-Van Der Veen S., Port A., Geurts T. B.: J. Clin. Endocrinol. Metab. 5, 2624 (2005). 6. British Pharmacopoeia: Volume II, 1443 (2007). 7. United States Pharmacopoeia, Asian Edition, United Pharmacopoeial Convention, Inc., Rockville, 1051 (1995). 8. Cavrini V., Di Pietra A.M., Raggi M.A., Sarti R.: J Pharm. Biomed. Anal. 1, 21 (1987). 9. Rambaud L., Monteau F., Deceuninck Y., Bichon E., Andre F., Bizec B.: Anal. Chim. Acta 586, 93 (2007). 10. Minto C., Howe C., Wishart S., Conway A., Handelsman D.: J. Pharmacol. Exp. Ther. 1, 93 (1997). 11. ICH Steering Committee, ICH Q2B Validation of Analytical Procedures: methodology, European Agency for the Evaluation of Medicinal Products, International Commission on Harmonisation, London (CPMP/ICH/281/95), (1996). 12. Burgess D.J.: J. Colloid Interf. Sci. 140, 227 (1990). 13. Tiyaboonchai W., Woiszwillo J., Middaugh C.R.: J. Pharm. Sci. 90, 902 (2001). 14. Lim S.J., Kim C.K.: Int. J. Pharm. 243, 135 (2002). Received: 12. 02. 2010