Capillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions

Similar documents
The Open-Split Capillary Interface. Part No

ms NoVent II Instruction Booklet

Connecting a SSL/PTV Backflush System

6890 GC Site Preparation

Markes Thermal Desorption Systems Site Preparation Page 1 of 8

MULTIDIMENSIONAL CAPILLARY GC SYSTEM FOR AGILENT TECHNOLOGIES 6890 GC

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division

Running HPLC Methods on an ACQUITY UPLC System

Agilent G3180B Two-Way Splitter Kit

Super-Clean Gas Filters from Restek

Connectors. SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses. Top Seller SEE STARRED PRODUCTS. VHP Tees & Crosses for Capillary Tubing

Laboratory Flowmeters

Advances in Capillary Column Backflushing

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Best Practice for Identifying Leaks in GC and GC/MS Systems

SIMPLAIR PIPING TOOL. JXT Company

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

no peaks missing peaks too small peaks increasing retention decreasing retention

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Instruction Manual Updated 7/26/2011 Ver. 2.2

MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Olfactory Detector Port OP275 Operations Manual

Inflatable Packer Single & Double. Single & Double Packer Dimension. Wireline Packer. Water Testing Packer (WTP) Packer

CP-3800 GC Getting Started Manual

3 Flow and Pressure Control

Crimpers and Decappers

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

Installation, Operation, and Maintenance Manual

Your local gas generation partner. Precision series Modular gas generation solution for GC.

System Overview. TCD Detector temperature setpoint Regulator

High Pressure Heatable Liquid Cells GS05910, GS05915, GS05920, GS05925 Series User Manual

CI-SfB 732 (5-1) x. - laboratory fittings

Agilent 7667A. Mini Thermal Desorber. Site Preparation

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Mass Spec will not Autotune

Installation, Operation, and Maintenance Manual

CORESTA RECOMMENDED METHOD Nº 67

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A

Valves FLUIDICS VALVE OVERVIEW & FUNCTIONS ROTARY SHEAR VALVES STAND ALONE VALVE PRODUCTS MANUAL VALVES SPARE PARTS VALVE ACCESSORIES

HICHROM. HPLC Accessories

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Precision and Instrument Regulators 1/4" port size

Instant Connect Helium Saver Injector Module

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

LC, RSLC, Bio RS and Dual Gradient Systems

Pressure Dump Valve Service Kit for Series 3000 Units

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Quantitative Analysis of Hydrocarbons by Gas Chromatography

ALE-40. R06 Miniature Brass Body Regulator. Water and Compressed Air Service 1/8" and 1/4" PTF Port Sizes. Air line equipment.

GILMONT ACCUCAL FLOWMETERS

5890II GC Standard Operating Procedure 9/2/2005

MV-Series High-Pressure Valves USER GUIDE

Description Supply pressure Pressure range Connection Device Page max. bar bar thread

VIC offers a variety of Calibrated Gas Leaks

Customer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist

Simple to install connection fails. The Talbot Pushfit fitting is pressure rated PN 16.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

AREA VALVE SERVICE UNITS

Air Operated Hydraulic Pumping Systems to 50,000 psi

Halsey Taylor Owners Manual

ATIS OPERATION MANUAL

Installation Instructions

Model: AT-1000 L.P.G. FREEZE VALVE

Model: AT-1000 L.P.G. FREEZE VALVE

SAILTEC, INC CONGER COURT, OSHKOSH, WI USA

Multiple Gas#5 GC configuration Jan 2016

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

RG1200 Service and Repair Manual

About. providing superior performance and reliability. Bromic regulators are AGA or UL144 approved.

7820A GC Site Preparation Checklist

Simple and Worry Free Gas Filters

AirCheck Kit Model K901

UNITY 2 TM. Installation Manual. Version 1.9. July QUI-1056 UNITY 2 Installation Manual

Instruments AQDC. Aluminium Quick Disconnects. Light-weight, quick and reliable method a making a number of pneumatic connections.

Agilent G Rebuilt Electron Capture Detector

Manual Solvent Purification System. Innovative Technology, Inc.

JR Series. Gas Pressure Reducing Valve GAS PRESSURE REDUCING VALVE

Avenger Model 33M Installation & Operation Instructions

HYDRAULIC PUNCH DRIVER 38456, 38520, 7306 / 7306SB, 7310 / 7310SB, 7506, 7606SB, 7610SB, 7625 / 7625Pg / 7625PgSB, 7646 / 7646Pg / 7646PgSB

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Chlorinator. MODEL C-731 (Floor Mounted)

Description Advanced Features Typical Applications

SB-E ISS.01. Operation Manual: KB-522 & KB-522-SS Pressure Feed Cup E P ITW Finishing Systems and Products

High Pressure Bleed Valves, Gauge Valves and Block & Bleed Valves

Warner Instruments, Inc Dixwell Avenue, Hamden, CT (800) / (203) (203) fax

GCMSD-Headspace Analysis SOP

SP-400-M1/SP-403K-M1 INSTRUCTIONS

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

For additional information about our solutions, please visit our web site at US/Pages/HomePage.

500 deg C Max in air unlimited in an inert gas Soft and Malleable. & with No phase

Transcription:

Operating Instructions Check the detector for the correct operating conditions. If make-up gas is being used a flow of 30ml/min. of carrier plus make-up is normally recommended. This can most effectively be measured at the exit of the detector after first turning off other gases to the detector. To ensure correct functioning of the system, the test chromatogram provided with an SGE capillary column should be reproduced. Set up the conditions as given in the computerised test report and inject the test sample provided with the column. A similar result to the test chromatogram should be obtained. Capillary Conversion System CCS-4/NC & CCS-05/NC Installation and Operating Instructions SGE International 7 Argent Place, Ringwood, 3134 Publication No.MN-0068-A Australia 11/92

Introduction This kit is designed to be a simple and inexpensive system for converting a gas chromatograph configured for packed column to operate with capillary columns. When correctly installed, true high resolution capillary chromatography can be achieved. Useful Tips 1. The most accurate way of setting the carrier gas flowrate in the column is to inject a volatile solvent while the column is heated to a temperature at which the solvent will not be retained. Measure the time it takes for the solvent peak to first leave the base line and calculate from the column length, the linear gas velocity through the column. 2. The split flow can be used to adjust the sample size onto the column. If the split flow is turned down too low, however, broadening of the peaks will be seen. When the gas flow through the vaporiser tube is too slow, the sample is delayed and thus is unable to enter the column as a sharp band. It is recommended NOT to use a split flow below 20ml/min. measured at the outlet of the split valve. 3. When a new system or column is installed, we recommend that it is conditioned with carrier gas flow for a few hours near the maximum continuous temperature of the capillary column. NB. Make sure that the carrier gas is clean, dry and free from oxygen. Fit adsorption tubes and oxygen filters to the carrier supply line if not already installed. See SGE catalogue for: P/N 103486 GFO Oxygen Scrubber P/N 103486 GFMS Molecular Sieve Absorber.

Installation Packing List Figure 1. CCS-4 N/C Capillary Conversion System (1/4 Injection Ports). Qty Description 1 R8286AMBS60 Pressure Regulator See Figure 5. 1 Splitter tee See Figure 3. See Figure 4. 1 1/4 x 4mm ID x 135mm GLT 1 1/4 x 4mm ID x 90mm GLT 1 Molecular Sieve Trap (AT-P) connected to a BMCV-1 Brass Mini Control Valve See Figure 2. 1 3SST-16/0.8 Stainless Steel Tubing 1 GVF/004 Vespel Ferrules (Pkt 5) 1 GVF/005 Vespel Ferrules (Pkt 5) 1 GVF/008 Vespel Ferrules (Pkt 5) 1 OGF/4 Graphite Ferrules (Pkt 5) 1 VSR/16 (Pkt 10) 1. Remove the existing packed column from the instrument. If 1/8 O.D. packed columns are being used, the 1/8 adaptors should be removed to accommodate the larger CCS capillary injector and detector fittings. 2. (Refer to Figure 2.) CCS-4 systems are supplied with two 1/4 O.D. GLT vaporiser tubes. Insert the shorter vaporiser tube into the gas chromatograph s existing packed column injector (5). The tube should be inserted the same distance into the injector as the packed column which has been replaced. At least 20mm of the tube should extend into the oven from the injector to leave sufficient room for the splitter tee and other components to be fitted. If there is insufficient room, use the longer (135 mm) GLT tube provided in the kit. 1 GFF/4 Graphite Ferrules (Pkt 10) 1 DC-4 Detector Connector Kit 1 1/4-5/16 AF Spanner 1 Piece Carborundum 1 Instructions CCS-05 systems are supplied with a vaporiser tube which is 5mm O.D. GLT.

Figure 2. Vapouriser Tube Packing List CCS-05 N/C Capillary Conversion System (5mm Injection Ports). Qty Description (22) VSR/16 Vespel Sealing Ring (25) Packed ColumnRetaining Nut (24) OGF/4 Graphite Ferrule 1/4" OD End of GLT Vaporiser Tube (23) GFF/4 Graphite Ferrule (1) Splitter Tee (22) VSR/16 Vespel Sealing Ring (21) GVF/004, 005 or 008 (see Installation Notes) 1 R8286AMBS60 Pressure Regulator 1 Splitter tee 1 5mm O.D., 135mm x 3mm ID GLT. Bushed to 1/4. 1 Molecular Sieve Trap (AT-P) connected to a BMCV- 1 Brass Mini Control Valve 1 3SST-16/0.8 Stainless Steel Tubing 1 GVF/004 Vespel Ferrules (Pkt 5) 1 GVF/005 Vespel Ferrules (Pkt 5) 1 GVF/008 Vespel Ferrules (Pkt 5) One end of this tube is bushed up to 1/4 O.D. The split tee (1) is connected to the 1/4 section. For both the CCS-4 and CCS-05, connect the splitter tee (1) using the 1/4 ferrules provided. First fingertighten and then tighten a further 11/4 turns with a spanner. Place the packed column retaining nut (25) which is normally used with the instrument (not provided in kit) over the vaporiser tube (2). Then slide a ferrule of the type normally provided to hold a packed GC column in the injection port over the vaporiser tube. Do not use metal ferrules as they will damage the glass layer inside the tube. 1 GFF/05 Graphite Ferrules (Pkt 5) 1 GFF/4 Graphite Ferrules (Pkt 10) 1 VSR/16 Vespel Sealing Rings (Pkt 10) 1 DC-05 Detector Connector Kit 1 1/4-5/16 AF Spanner 1 Piece Carborundum 1 Instructions Finger tighten the nut (25) and then tighten a further 11/4 turns with a spanner. 3. (Refer to Figure 3.) For capillary split injection technique it is essential to use a pressure regulated carrier gas system rather than flow controlled. If your instrument is fitted with a pressure regulator, this device can be used. If it is fitted with a mass flow

controller, the pressure regulator (12) supplied in the kit must be installed as detailed below: Figure 3. (5) Existing Packed Column Injector (11) Carrier Gas Inlet (12) R8286AMS60 Pressure Regulator a: Find a suitable position on the chromatograph to mount the pressure regulator (12). A 1/2 diameter hole in a panel is required to fix the regulator in place using the locking nuts. b: Connect the carrier gas source to the IN fitting on the regulator. Use VSR/16 Sealing Ring for making this seal. c: There will be a carrier gas inlet line (11) to the packed column injector from the original gas control device. Connect this line to the OUT position on the pressure regulator and seal with a VSR/16 vespel sealing ring. d: If the pressure gauge in the instrument is already incorporated in the carrier gas line, nothing further is required except to ensure the GA position on the pressure regulator is blocked off. If it is not, the line from the pressure gauge (13) should be connected to the GA position on the pressure regulator. 4. (Refer to Figure 4.) Select a suitable position on the chromatograph, outside the oven, to place the absorption trap (8) and splitter control valve (9) assembly. Cut a sufficient length of the 1/16 O.D. x 0.8mm I.D. stainless steel tubing (6) provided and connect from the split outlet (3) on the injector to the end of the absorption trap (7). Use VSR/16 sealing rings for making these connections. OUT GA IN Reordering Part Numbers This listing details individual parts. The part numbers represent the SGE order numbers for replacement items. FOR CCS-4/NC KIT Item Part No. Order Code Description 1 0930042 GISS Tee 1/4 x 1/16 Splitter tee 2A 0827701 1/4 x 4mm ID x 135mm GLT 2B 0827702 1/4 x 4mm ID x 90mm GLT 3 103403 SSN/16 Stainless Steel 1/16 nut 4 103405 SSNE 16/012 1/16 S/S extended nut 1.2mm bore 5 0624161 3SST 16/0.8 1/16 x 0.8mm S/Steel tubing 6 103403 SSN/16 1/16 S/S nut 7 0932184 AT-P Molecular sieve absorption trap 8 1236282 BMCV-1 Micro needle valve 9 103404 BN/16 Brass 1/16 nut 10 0624161 3SST 16/0.8 1/16 x 0.8mm S/Steel tubing 11 175788 PR8286 AMBS60 Pressure regulator 12 0624161 3SST 16/0.8 1/16 x 0.8mm S/Steel tubing (to gauge) 13 0624161 3SST 16/0.8 1/16 x 0.8mm S/Steel tubing (to carrier source) 14 103462 DC-4 1/4 Detector Connector Kit 15 (Part supplies with DC-4-Item 15) 1/16 x 0.8mm ID GLT 16 1034632 SSDUT Make-up tee (Part of DC-4 Kit) 17 103403 SSN/16 1/16 s/s nut (Part of DC-4) 18 0624161 3SST 16/0.8 1/16 x 0.8mm S/Steel tubing (to make-up source) 19 103405 SSNE 16/012 1/16 S/S extended nut, 1.2mm bore 20 072663 GVF/004 Graphitised Vespel Ferrule (Pkt 10) OR 072654 GVF/005 Graphitised Vespel Ferrule (Pkt 10) OR 072655 GVF/008 Graphitised Vespel Ferrule (Pkt 10) 21 072653 VSR/16 Graphitised Vespel Ferrule (Pkt 10) 24 072601 GFF/4 Graphite Ferrule (Pkt 10) 25 072621 OGF/4 Graphite Ferrule (Pkt 10)

5. (Refer to Figure 5.) The GC detector connections should be modified to accept capillary columns. 1/4 (CCS-4) or 5mm (CCS-05) detector fittings are provided. Detector connector kits are supplied with a length of 1/16 O.D. x 0.55mm I.D. GLT (fitted) and a spare interchangeable piece of 1/16 O.D. x 0.8mm I.D. GLT. The 0.55mm I.D. GLT should be used for 0.10, 0.15, 0.22 and 0.32mm I.D. capillary columns. The 0.8mm I.D. GLT is suitable for use with 0.53mm I.D. capillary columns. Figure 4. Splitter Gas Outlet (9) BMCV-1 Brass Mini Control Valve (8) AT-P Molecular Seive Trap FOR CCS-05/NC KIT Reordering Part Numbers Item Part No. Order Code Description 1 0930042 GISS Tee 1/4 x 1/16 Splitter tee 2 0827704 5mm x 3mm ID x 135mm GLT Vaporiser tube bushed to 1/4 at one end. 3 103403 SSN/16 Stainless Steel 1/16 nut 4 103405 SSNE 16/012 1/16 S/S extended nut 1.2mm bore 5 0624161 3SST 16/0.8 1/16 x 0.8mm s/steel tubing 6 103403 SSN/16 1/16 S.S. nut 7 0932184 AT-P Molecular sieve absorption trap 8 1236282 BMCV-1 Micro needle valve 9 103404 BN/16 Brass 1/16 nut 10 0624161 3SST 16/0.8 1/16 x 0.8mm s/steel tubing 11 175788 PR8286 AMBS60 Pressure regulator 12 0624161 3SST 16/0.8 1/16 x 0.8mm s/steel tubing (to gauge) 13 0624161 3SST 16/0.8 1/16 x 0.8mm s/steel tubing (to carrier source) 14 103462 DC-05, 5mm detector connector kit 15 (Part supplied with DC-05 Item 15) 1/16 x 0.8mm ID GLT 16 1034632 SSDUT Make-up tee (Part of DC-4 Kit) 17 103403 SSN/16 1/16 s/s nut (Part of DC-4) 18 0624161 3SST 16/0.8 1/16 x 0.8mm s/steel tubing (to make-up source) 19 103405 SSNE 16/012 1/16 s/s extended nut, 1.2mm bore 20 072663 GVF/004 Graphitised Vespel Ferrule (Pkt 10) OR 072654 GVF/005 Graphitised Vespel Ferrule (Pkt 10) OR 072655 GVF/008 Graphitised Vespel Ferrule (Pkt 10) 21 072653 VSR/16 Graphitised Vespel Ferrule (Pkt 10) 22 072601 GFF/4 Graphite Ferrule (Pkt 10) 23 072621 OGF/4 Graphite Ferrule (Pkt 10) Make-up gas may be required to reduce the effective dead volume and ensure maximum detector sensitivity. A tee-piece (17) is already connected to the stainless steel stem when supplied. This is used when make-up gas is required. The tee-piece (17) should be removed and replaced with the straight through union (supplied) if make-up gas is not required. (6) SST-16/0.8 Stainless Steel Tubing Figure 5. (22) VSR/16" Vespel Sealing Ring A length of 1/16 O.D. glass lined tubing (GLT) bevelled at one end, passes through the stainless steel stem to the base of the union and is locked into the connector with a high temperature graphite ferrule (GFF/16). The GLT is supplied longer than may be necessary in order that it can be positioned as close to the base of the detector as possible. Makeup Gas Inlet (15) Detector Connector (21) GVF/004, 005 or 008 (see installation notes) If necessary, both the stainless steel stem and the GLT liner may be cut to a more convenient length. The GLT can be cut using a wet silicon carbide cut-off wheel or the SGE Terry Tool (P/N 082780). A simple, although less effective method is to

score the GLT in a deep line all around using a fine file then snap the tubing. It should then be filed square on the end, deburred, washed, dried and then flame polished by heating to a medium red colour in a bunsen flame. Use the standard packed column detector retaining nut and ferrule supplied with the gas chromatograph to fix the detector connector (15) into the GC detector inlet. If make-up gas is being used, connect a line from the make-up tee (17) on the detector connector to a controlled supply of make-up gas. The existing flow controller in the gas chromatograph which has been replaced by the pressure regulator, can be used for this purpose. 6. At this stage the capillary column can be installed. It is recommended the column and cage is supported in some way in the centre of the GC oven. Unwind approximately 20cm of fused silica from each end of the column. Insert the column through the nut (4) and then through a suitable ferrule (refer to table 1. for the selection of the correct ferrule). The tapered end of the vespel ferrule is reversed into the nut as shown. Be very careful not to get pieces of ferrule or any other material inside the column. It is good practise at this stage to break a small piece from the column end by scratching the column approximately 2cm from its ends. Bend the column away from the scratch and it will break cleanly. TABLE 1. Column I.D. (mm) Ferrule Type 0.10 GVF/004 0.15 GVF/004 0.22 GVF/004 0.32 GVF/005 0.53 GVF/008 The column is inserted into the injector so that the end of the column is approximately 1cm past the nut (25) holding the capillary injector into the packed injector. Do not insert the column too far into the injector. The procedure for connecting the column into the detector connector is the same as for the injector except the end of the column should be inserted as close to the flame jet as possible. UNITED STATES OF AMERICA SGE, Incorporated Telephone : +1-512-837 7190 Toll Free : (800) 945 6154 Fax : +1-512-836 9159 E-mail : usa@sge.com AUSTRALIA AND PACIFIC REGION SGE International Pty Ltd Phone: +61-3-9837 4200 Fax: +61-3-9874 5672 E-mail: support@sge.com UNITED KINGDOM SGE Europe Ltd Phone: +44-1908-568 844 Fax: +44-1908-566 790 E-mail: uk@sge.com FRANCE SGE Europe Ltd. (France) Phone: +33-1-69 29 80 90 Fax: +33-1-69 29 09 25 E-mail: france@sge.com GERMANY SGE GmbH (Germany) Phone: +49-6151-860 486 Fax: +49-6151-860 489 E-mail: germany@sge.com JAPAN SGE Japan Inc Phone: +81-45-222 2885 Fax: +81-45-222 2887 E-mail: japan@sge.com INDIA SGE Laboratory Accessories Pvt Ltd Phone: +91-22-24715896 Fax: +91-22-24716592 E-mail: sgeindia@vsnl.com CHINA SGE China Service Centre Phone: +86-10-6588 8666 Fax: +86-10-6588 6577 E-mail: sales@jjindustries.com.cn MIDDLE EAST Tel: 00 971-6 - 557 3341 Fax: 00 971-6 - 557 3541 email: gulfsupport@sge.com Publication No. MN-0068-A Rev:01 12/95