ASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist

Similar documents
Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Automated Determination of Dissolved Gases in Water Anne Jurek

Automated Determination of Dissolved Gases in Water By Headspace Calibration of Mixed Gases Anne Jurek

Rapid and Reliable Detection of Dissolved Gases in Water

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

GCMSD-Headspace Analysis SOP

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Experiment GC : Analysis of BTEX by GC-FID

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Retention Time Locking: Concepts and Applications. Application

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

CORESTA RECOMMENDED METHOD Nº 67

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Determination of Dissolved Gases in Ground Waters. By Ray Martrano Laboratory Director Seewald Laboratories Inc.

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

A New GC/MS System Designed for Helium Carrier Gas Conservation

Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen

Gas Chromatographic Applications of the Dielectric Barrier Discharge Detector

Standard Test Method for Residual Acrylonitrile Monomer Styrene-Acrylonitrile Copolymers and Nitrile Rubber by Headspace Gas Chromatography 1

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

NCASI METHOD IM/CAN/WP IMPINGER/CANISTER SOURCE SAMPLING METHOD FOR SELECTED HAPS AND OTHER COMPOUNDS AT WOOD PRODUCTS FACILITIES

Multiple Gas#5 GC configuration Jan 2016

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Flare Gas Composition Analysis and QA/QC Lessons Learned and Lessons Lost SPECTRUM ENVIRONMENTAL SOLUTIONS, LLC 1

no peaks missing peaks too small peaks increasing retention decreasing retention

Quantitative Analysis of Hydrocarbons by Gas Chromatography

DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Determination of Oxygenates in Gasoline by Gas Chromatography and Oxygen Selective Flame Ionization Detector Method

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Development of DMS and Acetonitrile, and Formaldehyde gas standards. Gwi Suk Heo, Yong Doo Kim, Mi-Eon Kim, and Hyunjin Jin

6890 GC Site Preparation

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Low-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI

CHM250 Calibration and Measurement Lab. Balance Calibration

Carrier Gases in Capillary GC

APPLICATION NOTE. GC Integrated Permeation Device

This standard is due for Five Year Review. This process is required by the SEMI Regulations to ensure that the standard is still valid.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Carrier Gases in Capillary GC

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

Buck Scientific, Inc.

3 Flow and Pressure Control

Background Statement for SEMI Draft document 4657B NEW STANDARD: SPECIFICATIONS FOR TUNGSTEN HEXAFLUORIDE (WF 6 )

Gas chromatographic determination of 1,4-dioxane at the

5890II GC Standard Operating Procedure 9/2/2005

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port

Agilent 7697A Headspace Sampler

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

METHOD: 1024, Issue 2 EVALUATION: FULL Issue 1: 15 August 1987 Issue 2: 15 August 1994

Zebron Gas Management

System Overview. TCD Detector temperature setpoint Regulator

NOTE: Performance of this method should not be. attempted by persons unfamiliar with the operation of a gas

New improved robot for gas kinetics

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module

Gas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation

PREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director

Addressing the World Shortage of Helium

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

DRAFT. GC2 PFT Analysis: User Guide. Introduction. Apparatus, Reagents, & Materials. Laboratory Apparatus. Reagents. Draft 6/7/2011; Rev.

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

This GC has two columns (I think new columns can be ordered from Restek)

Best Practice for Identifying Leaks in GC and GC/MS Systems

May 1, By John Heim,Doug Staples

Quality Assurance Charting for QC Data

INSTRUCTOR RESOURCES

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

acrolein, acetaldehyde and acetone( cm -1 ); methanol (1306 cm -1 ); ethylene (949 cm -1 ); and isoprene (893 cm -1 ).

Applied Technology and Best Practices in CEE. Conference

Traceability of greenhouse gases and volatile organic compounds for air monitoring in South Africa

SOP: Buck Scientific BLC-20P HPLC Operation

Agilent G Rebuilt Electron Capture Detector

A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF ICATIBANT IN FORMULATIONS

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

MSD 96-Well MULTI-ARRAY and MULTI-SPOT Human Granulocyte Colony Stimulating Factor (hg-csf) Ultrasensitive Assay

ISO INTERNATIONAL STANDARD

7820A GC Site Preparation Checklist

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

Chemical Analysis for Methyltestosterone

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

Agilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist

A Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction

IMPACT OF THE HELIUM SHORTAGE ON PROCESS GAS CHROMATOGRAPHS

Advancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013

Simultaneous Determination of a Panel of 22 Steroids in Urine and Serum by SPE and LC-MS/MS

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

instrumentation Pulsed Discharge Detectors Non-Radioactive, Multiple Mode Electron Capture / Helium Photoionization

Comparison of Different DGA (Dissolved Gas Analysis) Methods for the Condition Assessment of Power Transformers

Transcription:

ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist

Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking process. There is no standardized method for the determination of dissolved gases in water.

Rationale The most common method for testing for dissolved gases is RSK-175. The RSK-175 procedure is a standard operating procedure, not a formal method. There have been assorted interpretations and modifications of RSK-175 in order to determine the amount of gas dissolved in the water.

RSK-175 Summary Displace 10% of the sample volume with high purity helium Shake the sample for 5 minutes Inject the headspace of the sample onto the GC Detection performed by a Thermal Conductivity Detector (TCD) or Flame Ionization Detector (FID). Concentration calculated using the Henry s Constant, the headspace volume versus the sample vial volume, and the temperature of the sample

WK 43267 Standardizes the method for the determination of dissolved gases in water. Gases: Methane Ethane Ethylene Propane

Calibration Approach Saturate water with the gas to be analyzed Perform a serial dilution of the saturated water in order to establish a calibration curve. Note: By saturating the water with the gas, there is no need to back calculate the gas concentration using the Henry s constant.

Dissolved Gas Solution Preparation Order high purity gases from gas supplier Fill 500ml volumetric flask with DI water Prepare 20 C recirculating water bath Place volumetric flask in water bath and let the DI water come to the 20 C temperature. Purge water at 200 to 250ml/min gas flow in volumetric flask with the high purity gas for 30 minutes for full saturation

Saturation Time Study Using the procedure outlined previously Deionized water was purged with the designated gas for two hours at 20 C with samples taken from the solution every 10 minutes for the first hour and every 30 minutes for the second hour. The samples were then sampled and analyzed for the respective compound area counts to determine when the gas had saturated the water solution.

Propane Saturation Time Results 140000000 120000000 100000000 80000000 60000000 Propane Area Count Run 1 Propane Area Count Run 2 Propane Area Count Run 3 40000000 20000000 0 10 20 30 40 50 60 90 120 Ave. Std. Dev. %RSD

Dissolved Gas Solution Concentrations Saturated Gas Concentrations: Methane 23.2 mg/l at 20 C Ethane 62.0 mg/l at 20 C Ethylene 149.0 mg/l at 20 C Propane 76.7 mg/l at 20 C Concentrations Referenced from Lange s Handbook of Chemistry, 14 th edition and encyclopedia.airliquide.com

Dissolved Methane Curve Preparation Saturated Methane Gas Solution Curve Preparation at 20 C Standard Amount Final Concentration Saturated Solution 50ml 11600µg/L Saturated Solution 25ml 5800µg/L Saturated Solution 5ml 1160µg/L Saturated Solution 1ml 232µg/L Saturated Solution 500ul 116µg/L Saturated Solution 100ul 23µg/L Saturated Solution 25ul 6µg/L *Samples Prepared in a 100ml Volumetric Flask

Curve Ranges Methane 6 to 11600 µg/l Ethane 6 to 5000 µg/l Ethylene 6 to 5960 µg/l Propane 8 to 5500 µg/l

GC/FID Experimental Parameters GC/FID Agilent 5890 Inlet Temperature 250 C Inlet Pressure 9psi Gas Helium Inlet Split/Splitless Split Ratio 20:1 Column Flow 14.33ml/min Column Restek RT Q-bond 30m x 0.53mm x 20µm Oven Program 45 C hold for 1 minute, ramp 16 C/min to 180 C hold for 1.06 min, 10.5 min total runtime FID Temperature 250 C

Sampling Process Saturated gas calibration standards or dissolved gas samples Displace a volume of saturated gas standard Heat and mix the standard Inject an aliquot of the headspace on GC/FID for separation and analysis.

Step 1 Instrument Vial Preparation Place a stir bar into an empty sampling vial. Sweep vial with nitrogen gas. Using a gas tight syringe spike the vial with surrogate and cap vial. Pull a volume of air from the sampling vial with 25ml syringe fitted with a luer lock needle leaving the luer lock needle in the vial septa and removing only the syringe thus insuring the integrity of the sample vial. Note: Volume of air removed to be the equivalent of volume of sample to be added

Step 2 Sampling from Vial Bring sample or standard vial to room temperature. Insert gas tight syringe with luer lock needle into the sample vial. Purge vial with an inert gas in order to pull a volume of sample into the syringe.

Step 3 Transferring Sample for Headspace Analysis Transfer sample into the prepared vial for sample agitation and headspace sampling. Take a 1ml aliquot of the headspace and inject it into the GC/FID for separation and analysis.

Sample/Standard Storage Samples collected in clean 40ml vials with butyl rubber septa Filled to volume with no headspace Stored at approximately 6 C Expire after 7 days with no preservative or 14 days if preserved with sulfuric acid Samples collected in duplicate Careful to minimize sample agitation during sampling and transport

Preservation Study Methane Standards Prepared at 5800ug/L in replicates of five Each set of replicates was preserved differently Preserved samples were run for initial results and then stored for 4 weeks with a standard run once a week to account for analyte loss

Preservation Study Results Preservation Scheme Ave. % Lost in One Week Std. Dev. One Week Ave. % Lost in Two Weeks Std. Dev. Two Weeks Ave. % Lost in Three Weeks Std. Dev. Three Weeks None -2.59 5.68-21.59 5.77-35.94 10.93 HCl -8.01 3.06-28.88 7.09-32.59 7.81 H2SO4-3.32 5.56-18.02 8.53-15.16 2.19 Thick Septa -1.61 4.47-16.07 1.72-9.14 10.42 Butyl Rubber Septa -3.55 3.75-7.60 3.53-1.10 10.36 Butyl Rubber Septa + H2SO4 0.55 2.86-6.10 3.71-4.01 7.19

Preservation Study Results

Curve Requirements Curves to be 5 points minimum Low point on the curve needs to be at or below laboratory quantitation limit Curve should bracket the linear range of the instrument Curve should be linear with better than 0.995 linear regression

Quality Control Initial demonstration of low system background must be done before any samples are analyzed Background subtraction may be done, if it is proven that a consistent amount of background contamination exists in the method blanks

Quality Control Method Blank for compound analyzed Calibration verification every 12 hours, alternating a mid-range standard and a low range standard, calibration standard to pass, +/-30% for mid range and +/-50% for low range, verification needs to be done for every compound to be analyzed Surrogate needs to pass +/-30% of expected for standards, blanks and samples

Quality Control LCS/LCSD used in lieu of MS/MSD LCS/LCSD should be at or around the mid-point of the curve with recoveries to be +/-30% of expected LCS/LCSD needs to be demonstrated for each analyte tested LCS/LCSD needs to be done for every 20 samples tested or one batch

Precision and Bias A single lab participated in the study Tested six levels of each compound Ten replicates of each level All samples spiked with surrogate

Precision and Bias At the time of the study, there was no accepted reference material, for determining the bias for the dissolved gases in the method. Bias for the surrogate was less than 10% for the single lab study. The single lab study produced an approximate 95% probability of correct results.

Repeatability Limit Methane Analyte Average Repeatability Standard Deviation Repeatability Limit Percent Relative Standard Deviation Percent Recovery Methane-46 36.53 3.40 9.52 8.83 79.41 Methane-116 100.68 5.97 16.72 5.63 86.79 Methane-232 204.88 15.88 44.47 18.25 93.63 Methane-116 0 Methane-580 0 Methane-928 0 Surrogate-50 0 1026.22 65.77 184.15 6.08 88.47 4676.24 298.39 835.48 6.05 80.62 7563.47 521.05 1458.93 6.54 81.50 489.99 25.19 70.53 4.88 98.00

Repeatability Limit Ethane Analyte Average Repeatability Standard Deviation Repeatability Limit Percent Relative Standard Deviation Percent Recovery Ethane-32 32.30 1.16 3.25 3.41 100.94 Ethane-64 75.20 7.32 20.48 9.23 121.29 Ethane-310 284.00 23.34 65.36 7.80 91.61 Ethane-1250 1063.40 98.96 277.09 8.83 85.07 Ethane-3100 3224.50 342.81 959.87 10.09 104.02 Ethane-4380 4198.40 201.00 562.81 4.54 95.85 Surrogate-500 462.70 32.97 91.49 6.70 92.54

Repeatability Limit Ethylene Analyte Average Repeatability Standard Deviation Repeatability Limit Percent Relative Standard Deviation Percent Recovery Ethylene-30 25.80 4.50 12.70 16.54 85.93 Ethylene-60 59.20 4.60 13.00 7.43 98.67 Ethylene-238 227.80 25.90 72.50 10.78 95.71 Ethylene-1190 1036.80 138.80 388.80 12.70 87.13 Ethylene-2980 2337.60 155.00 434.00 15.67 88.39 Ethylene-4470 3857.90 494.00 1383.20 12.15 86.31 Surrogate-500 537.50 57.30 160.50 10.12 107.50

Repeatability Limit Propane Analyte Average Repeatability Standard Deviation Repeatability Limit Percent Relative Standard Deviation Percent Recovery Propane-40 32.70 4.81 13.46 13.95 81.75 Propane-196 188.90 32.58 91.21 16.36 96.38 Propane-790 753.60 108.31 303.28 13.64 95.39 Propane-1200 1180.40 123.51 345.82 9.93 98.37 Propane-2400 2211.00 216.01 604.83 9.27 92.13 Propane-4700 5007.90 285.11 798.30 5.40 106.55 Surrogate-500 525.90 42.04 117.71 7.58 105.18

Current Status Work Item has been submitted to Subcommittee Ballot for approval New revision to be re-balloted Working with Commercial Company to try and develop Certified Reference Materials

References 1. Dean, John A., Lange s Handbook of Chemistry. 14 th Edition, R.R. McGraw-Hill, Inc., 1992. 2. Gas Encyclopaedia. (2009). Retrieved November 15, 2011 from http://encyclopedia.airliquide.com/encyclopedia.asp 3. ConocoPhillips Company, Drilling and Completion, Retrieved January 20, 2012, from http://www.powerincooperation.com/en/pages/drilling-andcompletion.html 4. Hudson, Felisa, RSKSOP-175, Revision No. 2, May 2004. 5. EPA New England, Technical Guidance for the Natural Attenuation Indicators: Methane, Ethane, and Ethene. Revision 1, July, 2001. 6. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/ FID), PADEP 3686, Rev. 0, April 2012.

THANK YOU... QUESTIONS EST Analytical Phone: 513-642-0100 Email: ajurek@estanalytical.com Web: www.estanalytical.com