Conversion factors for Thermal Mass flow meters for different operating mediums: An Experimental study

Similar documents
Influence of Reynolds number on Cd of Critical flow venturi nozzle and validation with ISO standards

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

800 Series Mass Flow Meters and Controllers. High Performance Mass Flow Meters and Controllers

Mass Flow Controller (MFC) for Gases

Compiled by: B Beard. Approved by: SH Carstens. Description of requirements and procedures for compact provers to be used as verification standards.

TEK-THERMAL 1700B. Thermal Mass Flowmeter. FLOW. Technology Solutions

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

ABSTRACT. flotek.g Innovative Solutions in Flow Measurement and Control - Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India

Mass Flow Meter (MFM) for Gases

Bubble Tube Installations

TH-1800-T TRX-700-CNG Converter

Thermal Mass Flowmeters. 5 Mass Flowmeters

Truck-mounted Mass Flow Metering for LPG Delivery

Gas flow calibrations performed at the National Metrology Institute of South Africa (NMISA)

analytical bulk flow distance FLOW level pressure temperature industrial communication

Precision level sensing with low-pressure module MS

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Design Features. General Description

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Performance Overview calibration Laboratory EP Instruments Messtechnik und Kalibrierung GmbH

Turbine Flowmeters for Gas Applications (Metric)

Cover Page for Lab Report Group Portion. Pump Performance

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

NMISA NEW GAS FLOW CALIBRATION FACILITY. Deona Jonker 7 October 2013

Gas Flow Calibration Basics

The University of Hong Kong Department of Physics Experimental Physics Laboratory

Establishing Primary Gas Pressure Standard facility (0.140 to 70) bar Gas Piston Gauge at FCRI and its applications

Model 601CV. Mass Flow Controller. Product Features:

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

INSTRUMENTS A THERMAL MASS FLOW SENSOR USING A CONSTANT DIFFERENTIAL TEMPERATURE ABOVE THE AMBIENT GAS TEMPERATURE

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book.

Torque Tube TB300 Digital Transmitters

Ron Gibson, Senior Engineer Gary McCargar, Senior Engineer ONEOK Partners

Level MEASUREMENT 1/2016

Pneumatic dead-weight tester Model CPB3500

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

How to specify a product. Process Sensors and Mechanical Instruments

KMT. Thermal Mass Flow Meter. for Compressed Air and Gases. Measuring range: m3n/h m3 N

Fluid Machinery Introduction to the laboratory measurements

Turbine Flowmeters for Gas Applications

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Mass Flow Controllers

Radar, Ultrasonic and RF Level Transmitters

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

CORESTA RECOMMENDED METHOD N 6

780S Ultra High Purity Series

Mass Flow Controller (MFC) for Gases

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

The Discussion of this exercise covers the following points:

Aera Industrial MFC Product Guide. Innovative Control at Work

CHEM 355 EXPERIMENT 7. Viscosity of gases: Estimation of molecular diameter

Flow meter. bellow vaporizer. APL valve. Scavenging system

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

Vehicle- or rack-mounted liquefied gas meters, pump supplied

Gerald D. Anderson. Education Technical Specialist

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Pressure Measurement. Introduction. Engr325 Instrumentation. Dr Curtis Nelson 3/12/18

Vehicle-mounted meters, pump supplied

Thermal Mass Flow Controllers for High Accuracy Process Gas Applications

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

PHYS 101 Previous Exam Problems

SideTrak 830/840. Process Gas Mass Flow Meters and Controllers. FEATURES DESCRIPTION

STANDARD FOR CONTROL VALVE SEAT LEAKAGE

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

TEK-SUB 4800B. Submersible Level Transmitter. Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys LEVEL

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

NMISA New Gas Flow Calibration Facility

CALCULATING THE SPEED OF SOUND IN NATURAL GAS USING AGA REPORT NO Walnut Lake Rd th Street Houston TX Garner, IA 50438

Application Note AN-107

Turbine Flowmeters for Gas Applications

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Over 20,000 Strain Gage Target flowmeters installed since 1952.

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Flow and Energy Measurement F-5400 THERMAL MASS FLOW METER

mass of container full of air = g mass of container with extra air = g volume of air released = cm 3

Pressure transmitters EMP 2

Orifice Plate Flow Meter

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

Figure 1 Schematic of opposing air bearing concept

Vortex Flow Meter Wafer or Flange Connection. - Steam - Liquid - Gas

Kinetic-Molecular Theory

MASFLO-OVAL Ⅱ THERMAL MASS FLOWMETER/CONTROL VALVE for GAS MEASUREMENT MODEL: FHC, FHD series

Products Solutions Services. Gas Flow Measurement

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

THERMAL MASS FLOW MEASUREMENT FOR GASES

White Paper August , Rev AA

3 1 PRESSURE. This is illustrated in Fig. 3 3.

Technical Data Sheet High Performance Digital Gas Mass Flow Meters and Controllers Features Description

AUTOMATIC HOSE TEST UNIT, TYPE SPU

ONSITE PROVING OF GAS TURBINE METERS Daniel J. Rudroff Invensys Metering Systems

T EK-COR 1100A. Coriolis Mass Flowmeter. FLOW. Technology Solutions

Measurement of Low GAS & Liquid Flows

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Detector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

Transcription:

flotek.g 2017- Innovative Solutions in Flow Measurement and Control -Oil, Water and Gas August 28-30, 2017, FCRI, Palakkad, Kerala, India Conversion factors for Thermal Mass flow meters for different operating mediums: An Experimental study Unnikrishnan.P.N (*), Sivaprakash.K.G (**),Dr.Jacob Chandapillai (***) (*)Junior Research Engineer, (**) Deputy Director, (***) Director Fluid Control Research Institute (Under Ministry of Heavy Industries & Public Enterprises) Kanjikode West, Palakkad-678 623, Kerala, India Website: www.fcriindia.com E-mail ID: fcri@fcriindia.com Phone :91-491-2566206,2566120, Fax :91-491- 2566326 ABSTRACT Thermal Mass flow meters use the thermal properties of the fluid to measure the flow of a fluid flowing in a pipe or duct. In a typical thermal mass flow meter, a measured amount of heat is applied to the heater of the sensor. Some of this heat is lost to the flowing fluid. As flow increases, more heat is lost. The amount of heat lost is sensed using temperature measurement(s) in the sensor. The transmitter uses the heat input and temperature measurements to determine fluid flow. Therefore, it is important for the thermal mass flow meter supplier to know the properties of the fluid so that the proper calibration factor can be used to determine the flow rate accurately for operating mediums different from actual calibration medium. Due to this constraint, thermal mass flow meters are commonly applied to measure the flow of pure gases. Suppliers can provide appropriate calibration information for other gases, however the accuracy of the thermal mass flow meter is dependent on the actual gas being the same as the gas used for calibration purpose. In other words, the accuracy of a thermal flow meter calibrated for a given gas will be degraded if the actual flowing gas is different from the Calibrated gas. Theoretical conversion from one gas to another is very difficult and can result in large errors, especially if using the manufacturer's conversion factors. A series of flow performance tests were carried out on a 22 lpm Thermal Mass flow meter of inlet size less than 1inch with different mediums. Gases used for the studies are Air, Nitrogen, Argon, Helium and Carbon dioxide. The correction factors in Thermal mass flow meter for its application in different mediums estimated using experimental methods were compared with specified factors for different mediums recommended by the manufacturer of the flow meter. A Primary proving system was used for measurement of actual flow in the test run. Keywords: Thermal Mass flow meters, Gas Correction factor, Calibration 1. ABOUT FCRI The Fluid control Research Institute (FCRI) an autonomous R&D Institute was established in 1987 with active assistance and participation from UNDP and UNIDO, under the Ministry of Industry (Govt. of 1

India). The Institute is accredited by different National/International bodies such as National Accreditation Board for Testing and Calibration Laboratories (NABL, India); Underwriters Laboratory, USA; Chief Controller of Explosives, Nagpur-India; Bureau of Indian Standards (BIS); NMI, Netherlands; Department of weights and measures, India; Central Pollution Control Board(CPCB, India); Department of Science and Technology, India; etc. FCRI regularly participates in International round robin proficiency test programs in association with NIST, USA; CEESI, USA; NEL, UK; DELFT, Holland; DTI, Denmark; and KRISS, Korea etc. Other than flow laboratories with Air, Water and Oil as media, FCRI has got supporting laboratories like metrology section, Noise and Vibration section, and Electronics and Instrumentation section etc. FCRI regularly conducts National/International level Conferences/Training programs. FCRI also develops software in the field of flow meter design and selection valves/pumps, natural gas metering etc. 2. PRIMARY CALIBRATION FACILITIES AT FCRI Primary Air flow Calibration facility at FCRI is equipped with internationally accepted primary flow standards like Bell Provers and Piston Provers of various capacities for precise flow measurement. Calibration of flow meters at low pressure is carried out here for flow ranges up to 40 m 3 /h. As per the norms stipulated by the ISO, Primary air flow lab is always maintained at controlled ambient conditions (20±0.5 0 C, 55±5 % RH) for ensuring metrological qualities of the master flow meters and thereby providing highest quality and precision in calibration of flow meters. Blower Limit switch Support column S.S. pipe Tank Digiruler sensor head Solenoid valve Ball valve Pressure connector Solenoid valve Resistance temperature detector Digiruler Scale Fig 1. Schematic of 500ltr Bell Prover at Air Flow Laboratory 3. DESCRIPTION OF 50 LTR BELL PROVER SYSTEM A primary standard Bell prover calibration system is used to accurately measure a displaced volume of gas at a constant pressure. Interface electronics are provided to measure displaced volume and the time over which the volume is displaced so that a precise flow rate can be determined. To calibrate a flow meter using the Bell prover, the flow meter must be installed upstream of the bell so that all gas flowing through it is captured by the calibrator. The System consists of a vertical inner tank surrounded by an outer shell. The annular space between the inner tank and outer shell is filled with sealing liquid (Oil). An inverted tank, called the bell, is placed over the inner tank with the wall of the bell riding in the annular space filled with sealing liquid. The liquid provides a bubble-tight, friction-less seal which allows for smooth linear movement of the bell at essentially zero pressure drop.the bell is precisionbalanced so that pressure remains constant throughout the stroke. A large counterweight is used to balance the weight of the bell while a small counterweight hanging from an involute cam is used to compensate for buoyancy and other linear effects experienced by the bell as it moves along its stroke.the Bell prover has a feature that a large counterweight is Pulley Cross bar Strips Bell RTD indicator Digiruler display Data Acquisition System Medium : Air Range : 0 to 40 m 3 /h Uncertainty : ±0.2% Make : SIM Brunt, Italy 2

suspended in small tank which is hydraulically connected to the annular space in which the bell rides. The counterweight is machined so that its crosssectional area exactly matches the crosssectional area of the bell. As the bell moves out of the sealing liquid, the counterweight moves into the liquid so that the level of the sealing liquid remains constant. This allows for a more accurate/repeatable calibration since the varying liquid level is no longer an influence. The bell is a precision cylinder so that linear movement of the bell is proportional to displaced volume. A rotary encoder is directly coupled to the pulley shaft and produces pulses proportional to the linear displacement of the bell. Interface electronics provide a means of counting the encoder pulses. An exact number of pulses and corresponding time are accumulated over a calibration run. The following are the salient features of the facility. Method : Bell Prover Primary parameter : Volume Medium : Compressed Air/Nitrogen/ Temperature : 20 ± 0.5 o C. Flow rate : 0.03 to 16m3/h Uncertainty : ± 0.3% 4. THERMAL MASS FLOW METER OPERATING PRINCIPLES 4.1 Capillary Thermal Mass flow meter (CTMF meter) A typical CTMF meter consists of a meter body and flow sensor. The flow sensor is mounted into the meter body. A defined portion of the gas flow from the meter body is diverted through the (By pass) flow sensor, through which the gas flow rate is measured. Fig 3 shows a simplified CTMF meter with a flow sensor consisting of a thin tube and two temperature sensors. Depending upon the meter manufacturer the heater can either be combined with each temperature sensor or be located separately in the middle of the flow sensor ie between the temperature sensor upstream (T1) and the one downstream (T2) of the gas flow. A precision power supply delivers constant heat to the flow sensor. Under stopped flow conditions, both sensors measure the same temperature. As the flow rate increases, heat is carried away from the upstream sensor (T1) towards the downstream (T2). A bridge circuitry interprets the temperature difference and an amplifier provides the flow rate output signal. The measured temperature difference between the two sensors is proportional to the mass flow rate. Fig 2-50ltr Bell prover system at FCRI Fig 3 -Simplified CTMF meter 3

Where 1. Upstream temperature sensor (T1) 2. Downstream temperature sensor (T2) 3. Upstream temperature sensor (T1) with heater 4. Downstream temperature sensor (T2) with heater 5. Constant power supply P 6. Heater 7. Bridge circuitry 8. Amplifier 9. Flow signal output (0V to 5V dc or 4 ma to 20mA) The flow sensor measures the mass flow rate as a function of temperature difference. This can be expressed according to first law of thermodynamics (heat in= heat out, for no losses) for which the following equations apply. P=q m C p (T 2-T 1) Where q m is the mass flow rate, expressed in kg/s Cp is the specific heat, expressed in joules per kilogram per Kelvin of the gas at constant pressure T 2-T 1 is the temperature difference, expressed in Kelvin P is the constant input power, expressed in watts 5. CALIBRATION OF THERMAL MASS FLOW METER USING BELL PROVER WITH DIFFERENT MEDIUMS A 22 slpm Thermal mass flow meter was used for the study and a series of flow performance tests were conducted with different mediums. Gases used for the studies are Air, Nitrogen, Argon, Helium and Carbon dioxide. A Primary proving system (Bell Prover) was used for the measurement of actual flow in the test run. The Thermal Mass flow meter was mounted at the upstream of the reference Bell Prover. After ensuring the leak tightness, the calibration was done for specified flow rates covering the entire range of the Thermal Mass flow meter. The required flow rate was maintained using a Mass flow controller and it was allowed to pass through the Thermal mass flow meter and then to the reference Bell prover. Bell prover is provided with an optical encoder, generating a pulse representing an incremental change in the displaced volume. Provision was made for the measurement of pressure and temperature at reference Bell Prover location. The reference flow rate was measured from the data obtained from the Bell prover and it was compared with the flow rate indicated by the Thermal mass flow meter. The Calibration was repeated using different gases. The specification of the Thermal Mass flow meter was given below. Type: Thermal Mass Flow Meter Range: 0-22Slpm (For Air medium) 6. CONVERSION FACTORS FOR VARIOUS MEDIUMS (THEORETICAL PUBLISHED DATA) All the Calibrations cannot be performed using the desired process gas due to various reasons. It this cases it is a common practice to use a reference gas like air or nitrogen for calibration. Gas conversion factors K have been used since the first commercialization of capillary tube thermal MFMs and MFCs to convert the flow calibration for a reference gas to any other gas. Most manufacturers provide a list of gas conversion factors relative to a single primary reference gas, usually air or nitrogen. 4

The Conversion factor K is defined as = Cp.ref ρn.ref Cp.proc ρn.proc Where Cp.ref is the specific heat, at constant pressure, of the reference gas; Cp.proc is the specific heat, at constant pressure, of the process gas; ρn.ref is the density at normal conditions, of the reference gas ρn.proc is the density at normal conditions, of the process gas The below table shows the K-factor of commonly used gases provided by the manufacturer for conversion from one gas to another. Density Heat capacity Thermal conductivity,k Conversion factor,k ρn (kg/m3) Cp-ca l[cal/g.k] (W/m.K) 20 C, 1 atm Name of the gas Symbol 0 C, 1 atm 20 C, 1 atm 20 C Air Air 1.293 0.241 0.025596 1 Nitrogen N2 1.25 0.249 0.025468 1 Argon Ar 1.784 0.125 0.017391 1.4 Helium He 0.1785 1.24 0.14786 1.41 Carbon dioxide CO2 1.977 0.213 0.016225 0.74 Table 1- Gas properties and Conversion factor for different gases Where Q ref is the reference flow rate from the Bell Prover for a particular gas. Q ind is the flow rate indicated by the Thermal Mass flow meter for air medium. 8. RESULTS Indicated flow rate, flow rate after applying manufacturer specified conversion factor, actual flow rate, estimated conversion factor and error in flow rate after applying manufacturer conversion factor are given in Tables 2 to 8 for different gases. Meter under Calibration Actual Error Sl. Indicated flow rate (q i -q a )/q a *100 No. flow, q i q a (slpm) (slpm) (% rdg) 1 2 3 4 1 22.06 22.05 0.06 2 17.57 17.55 0.12 3 15.18 15.19-0.06 4 12.55 12.55-0.02 5 10.00 10.01-0.05 6 7.45 7.44 0.08 7 5.18 5.18 0.10 8 3.17 3.17-0.13 Table 2- Results of Calibration of Thermal Mass flow meter using Air medium 7. CONVERSION OF FLOW FROM ONE GAS TO ANOTHER (EXPERIMENTAL METHOD) The conversion factors in Thermal mass flow meter for its application in different mediums were estimated using the experimental methods relative to air medium. Estimated Conversion factor K, = Qref. Qind. Meter under Calibration Actual Estimated Error after Sl. Indicated Flow rate after flow rate Conversion factor using manufactuer No. flow rate (air medium) applying manufacturer Kest conversion factor qi conversion factor, qt qa (qa/qi) (qt-qa)/qa*100 (slpm) (slpm) (slpm) - (% rdg) 1 2 3 4 5 6 1 21.99 21.99 22.05 1.0026-0.26 2 17.49 17.49 17.54 1.0029-0.29 3 15.27 15.27 15.32 1.0031-0.31 4 12.44 12.44 12.48 1.0035-0.34 5 10.20 10.20 10.24 1.0043-0.43 6 7.40 7.40 7.43 1.0041-0.40 7 5.10 5.10 5.12 1.0039-0.39 8 3.15 3.15 3.18 1.0108-1.07 Table 3- Results of Calibration of Thermal Mass flow meter using Nitrogen medium 5

Meter under Calibration Actual Estimated Error after Sl. Indicated Flow rate after flow rate Conversion factor using manufactuer No. flow rate (air medium) applying manufacturer Kest conversion factor qi conversion factor, qt qa (qa/qi) (qt-qa)/qa*100 (slpm) (slpm) (slpm) - (% rdg) 1 2 3 4 5 6 1 15.96 22.34 23.72 1.4859-5.78 2 14.19 19.87 20.91 1.4733-4.97 3 12.36 17.30 18.05 1.4603-4.13 4 10.88 15.23 15.78 1.4499-3.44 5 8.98 12.57 12.93 1.4397-2.76 6 7.57 10.60 10.86 1.4351-2.45 7 5.92 8.29 8.46 1.4290-2.03 8 4.55 6.37 6.49 1.4274-1.92 9 3.45 4.83 4.92 1.4252-1.77 Table 4- Results of Calibration of Thermal Mass flow meter using Argon medium Sl no. Average Deviation between Experimental Theoretical & Theoretical Experimental Conversion Conversion factor, K-factors Name of the gas Symbol factor, K t K e 1 Air Air 1-2 Nitrogen N2 1 1.0044 0.44% 3 Argon Ar 1.4 1.4473 3.37% 4 Helium He 1.41 1.3893-1.47% 5 Carbon dioxide CO2 0.74 0.7868 6.32% Table 7- Theoretical and Experimental conversion factors Meter under Calibration Actual Estimated Error after Sl. Indicated Flow rate after flow rate Conversion factor using manufactuer No. flow rate (air medium) applying manufacturer Kest conversion factor qi conversion factor, qt qa (qa/qi) (qt-qa)/qa*100 (slpm) (slpm) (slpm) - (% rdg) 1 2 3 4 5 6 1 15.79 22.26 21.26 1.3462 4.74 2 15.61 22.01 21.10 1.3515 4.33 3 12.50 17.63 17.12 1.3692 2.98 4 10.70 15.09 14.80 1.3832 1.94 5 8.57 12.08 11.97 1.3970 0.93 6 6.51 9.18 9.18 1.4094 0.04 7 4.49 6.33 6.39 1.4241-0.99 8 3.09 4.36 4.43 1.4340-1.67 Table 5 - Results of Calibration of Thermal Mass flow meter using Helium medium Sl no. Heat capacity Thermal conductivity,k Deviation between Cp-ca l[cal/g.k] (W/m.K) Theoretical & Experimental Name of the gas Symbol 20 C, 1 atm 20 C K-factors 1 Air Air 0.241 0.025596-4 Helium He 1.24 0.14786-1.47% 2 Nitrogen N2 0.249 0.025468 0.44% 3 Argon Ar 0.125 0.017391 3.37% 5 Carbon dioxide CO 2 0.213 0.016225 6.32% Table 8- Thermal Conductivity of gases and Deviation between Theoretical & Experimental conversion factors 7.00% Meter under Calibration Actual Estimated Error after Sl. Indicated Flow rate after flow rate Conversion factor using manufactuer No. flow rate (air medium) applying manufacturer Kest conversion factor qi conversion factor, qt qa (qa/qi) (qt-qa)/qa*100 (slpm) (slpm) (slpm) - (% rdg) 1 2 3 4 5 6 1 23.37 17.29 19.43 0.8312-10.97 2 22.03 16.30 18.09 0.8213-9.90 3 17.89 13.24 14.26 0.7968-7.13 4 14.13 10.46 11.02 0.7798-5.11 5 12.20 9.03 9.44 0.7738-4.36 6 10.11 7.48 7.77 0.7681-3.65 7 7.35 5.44 5.60 0.7623-2.93 8 5.15 3.81 3.92 0.7610-2.76 Table 6- Results of Calibration of Thermal Mass flow meter using CO2 medium Deviation betweentheoretical & Experimental K- 6.00% 5.00% 4.00% 3.00% 2.00% 1.00% 0.00% 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16-1.00% -2.00% Thermal conductivity, k (W/m.K) Helium (0.14786), Nitrogen (0.025468), Argon (0.017391) Carbon dioxide (0.016225) Fig 4- Thermal Conductivity of gases and Deviation between Theoretical & Experimental conversion factors 6

Errorin % Error Percentage -10-12 6 4 2 0 0 0.02 0.04 0.06 0.08 0.1 0.12 0.14 0.16-2 -4-6 -8-10 -12 6 4 2-4 -6-8 Thermal conductivity, k (W/m.K) Helium (0.14786),Air (0.025596) Nitrogen (0.025468), Argon (0.017391) Carbon dioxide (0.016225) Helium Air Nitrogen Argon Carbon Dioxide 100% flow 80% Flow 70% Flow 50% Flow 40% Flow 30% Flow 20% Flow Fig 5- Thermal Conductivity of gases Vs Percentage Error for different flow rates 0 0 10 20 30 40 50 60 70 80 90 100 110-2 Percentage Flow Fig 6- Percentage flow Vs Percentage error for different mediums 9. OBSERVATIONS BASED ON EXPERIMENTAL DATA Theoretical conversion for Thermal mass flow meters can usually be expected to be within 11% or better when using manufacturer's factors based on the experimental data for different mediums. Maximum deviation of 6.3% is observed in the average experimental Conversion factor from the theoretical conversion factor for the CO 2 medium from the air medium. Minimum deviation of 0.44% is observed in the average experimental Conversion factor from the theoretical conversion factor for the Nitrogen medium from the air medium and this may be due to the similar properties of air and Nitrogen. For higher thermal conductivity gases the error is observed in the positive side and vice versa in the case of low thermal conductivity gases. As the thermal conductivity of the fluid is higher which in turn results in higher heat transfer and hence the higher temperature difference between the sensors. Higher temperature difference causes more error on the positive side and vice versa. Helium is having the largest thermal conductivity(k=0.14786 W/m.K) in the Calibrated gases, which is having a higher positive error of approximately 5% and the Carbon dioxide is having the lowest thermal conductivity(k= 0.016225W/m.K) which resulted in a higher negative error in the order of approximately - 11%. 10. CONCLUSION Thermal Mass flow meters are capable of good accuracy (1% or better) if they are calibrated on the actual process gas. Theoretical conversion from one gas to another is very difficult and can result in large errors, especially if using the manufacturer's conversion factors. The manufacturer conversion factor is derived based on the Specific heat and density of the reference and process gases. But the above experimental result shows that the error of the Thermal mass flow meter is also depending on the Thermal conductivity of gases which is not addressed. 7

REFERENCES [1] ISO 14511: 2001 Measurement of fluid flow in closed conduits-thermal mass flowmeters [2] Flow Measurement Engineering Handbook, Richard W. Miller [3] Industrial Flow Measurement, David W Spitzer 8