Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Similar documents
Unblocking the MicroSprayer Aerosolizer Model IA-1B With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Instructions for use of a Penn-Century MicroSprayer Aerosolizer - Model IA-1C and FMJ-250 High Pressure Syringe

Flexi-pro Aerosol Delivery System (FADS) Manual

MODEL 1329 Tank Gauge

Passing a Nasogastric Tube. Information for Parents and Carers

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Installation of Your SprayMaster System

Patients Guide for Self Administering Intravenous Antibiotics General points

Final Assembly Instructions Bikes with Quill Stems

8. Carefully layout all the parts of the reservoir setup and clean any dirt, grime or dust from the parts. ( image 8 ) 8

INSTRUCTION MANUAL FOR

Regulators repair and maintenance. XS Compact 2nd stage. January 2014 Rev XSC /3 Ed. C /14 1

Installation Procedure for your Disco Double Kit, by Airgun Lab, LLC.

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

Basic Flows in a Microfluidic Device

USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Technical Service Bulletin June 2015 TSB122.06

Hydraulic Punch Drivers

BOYLE S / CHARLES LAW APPARATUS - 1m long

Ultrasound Guided Thoracentesis Simulator Strap-on Type

DL 8 Dome Port. Product Registration. Product Number 75340

Drive Belt Instructions

PEDIATRIC INJECTABLE TRAINING ARM SIMULATOR LF00958U INSTRUCTION MANUAL

Solvent System Walkthrough

!!!! SERVICE MANUAL PRESSURE POT 2 GALLON. Service Manual: LT Washington St 931 Progress Ave., #7

RG1200 Service and Repair Manual

Using the Akta Prime plus October 22, 2012

Articulating Legs with Intraosseous Infusion Simulator LF03840U Instruction Manual

MAINTENANCE PROCEDURE FOR X 650

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

Final Assembly Instructions Bikes with Threaded Headsets

2: How to Fill Veloci-T Cartridges

Unisense In Situ UniAmp Connector System

Final Assembly Instructions Bikes with Threaded Headsets

Installation Instructions

VAPOMATIC ANESTHETIC VAPORIZER

Technical Service Bulletin Oct 2017 TSB122.07

QUIK-jet Micro-Injection System

Instruction Manual Updated 7/26/2011 Ver. 2.2

3 Scotch-Weld Polyurethane Reactive Adhesive Applicator Replacement Part Installation Guide

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

Instruction Manual - War Department Baldwin Live Steam Instruction Manual

Ultrasound Guided Thoracentesis Simulator Strap-on Type

Instruction Manual. B&O WITH SLOPE BACK TENDER Live Steam

Final Assembly Instructions Bikes with 16 Wheel Size

Product Information News September 26, 2003

1. A proper preparation of the cone and other material is inevitable prior to a CPTU test.

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

PREPARING GAS SAMPLES IN LARGE PLASTIC SYRINGES

User manual STARLOCK. Careways International - Innovation Quality Performance

4 ANGLE GRINDER MODEL NO: CAT 52 PART

Powder Measure. Care & Usage Instructions. Instructions # Product # Revision: A

LABORATORY TECHNIQUES. Pouring Liquids

Lipodrop 2.0 device coating

TECH SHEET ORANGE PIVOT TOOL INSTRUCTIONS

model - CYPRESS DX W

RB70 Automatic Diluent Valve Maintenance Manual. Version 1.1 November 2006 Written by Tino de Rijk. Page 1 of 23

APP pumps APP and APP Disassembling and assembling

Multi-Probe Sampling System User Guide & Manual

Maintenance and Repair. AirHawk II Air Mask. Second Stage Regulator. Order No.: /02 Prnt. Spec (I) MSAsafety.

Before you start sampling, be sure to read

Flexfence Swaging Unit. Product Manual. Release 02/16.

MODEL 840 AIR HYDRAULIC PUMP INSTRUCTION

INSTALLATION and OPERATION INSTRUCTIONS

LIQUID LEVEL GAUGE FOR WELDING ZGAU 716 O&M, 717 O&M, 718 O&M, 719 O&M

PANACEA P100 LOW PRESSURE USER MANUAL

OXYTEMP Instructions 1

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

Model 23H Hand Crank Seamer

15ME-014 INSTRUCTION MANUAL OF STERN TUBE SEALING TYPE EVK2RV. URL

6900 Maintenance Instruction System Flush

ACCUCRAFT UK LTD Unit 4, Long Meadow Industrial Estate, Pontrilas, Herefordshire. HR2 0UA Tel:

PANACEA P200 STAINLESS STEEL USER MANUAL

OWNER S MANUAL AirTop Large and AirTop Small

Assembly Drawing: W-311B-A01, or as applicable Parts List: W-311B-A01-1, or as applicable Special Tools: , , &

Trike-Bike Assembly Manual

AirPort Extreme A1521 Logic Board Replacement

FOR SAFE AND CORRECT USE

BELT DRIVE INDOOR CYCLING BIKE SF-B1712

UltRo Dual Flow Reverse Osmosis Water System

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

DM-MARD (English) Dealer's Manual. ROAD MTB Trekking. City Touring/ Comfort Bike REAR DERAILLEUR XTR RD-M9100 RD-M9120

USER MANUAL. All rights reserved

500SE SECOND STAGE SERVICE PROCEDURE

Installation Instructions

TJF-Q180V Cleaning and Disinfection Checklist

Assembly, Fitting, Care & Maintenance

Intro: by firechim on October 3, 2013

SILENT-SR SOUND SUPPRESSOR

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

Type BBS-03, BBS-05, BBS-06, BBS-25

Operating Instructions

Installation and Service Manual

AR-15 Rifle Operator Maintenance

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL

Transcription:

Penn-Century, Inc. 7670 Queen Street Suite 200 Wyndmoor, PA 19038-8031 USA 215-753-6540 Tel 215-220-3040 Fax www.penncentury.com Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter Instructions for Use The High Pressure Reverse Cleaning Adapter permits the FMJ-250 High Pressure Syringe to be used to force hot water or solvent backwards that is, in reverse direction from the tip to the hub of the MicroSprayer Aerosolizer Model IA-1C. If successful, this process will loosen and force out whatever is blocking (clogging) the MicroSprayer tip. Before you begin: CAUTIONS! The Adapter will not work in all cases. In some cases, material may be stuck in position or may have hardened over time, and may not be possible to clear using this Adapter. This Adapter has been successfully used by Penn-Century s technical staff for several years, but it requires patience and skill to operate successfully. The right amount of manual force/pressure must be applied to the plunger at the right rate - otherwise the MicroSprayer may be damaged beyond repair. WARNINGS! BEFORE USING THIS ADAPTER, USERS MUST READ ALL OF THE FOLLOWING INSTRUCTIONS TO AVOID CAUSING DAMAGE TO YOUR DEVICES. ONLY USE THIS ADAPTER KIT IF THE CLEANING METHODS RECOMMENDED IN THE INSTRUCTIONS FOR CLEARING OF THE MICROSPRAYER AEROSOLIZER - MODEL IA-1C DO NOT WORK AFTER REPEATED ATTEMPTS. DISCLAIMER PENN-CENTURY ACCEPTS NO RESPONSIBILITY FOR DAMAGE THAT MAY OCCUR AS A RESULT OF MISUSE OR INCORRECT USE OF THIS ADAPTER. Penn-Century, Inc. Revised 20 December 2013 1

High Pressure Reverse Cleaning Adapter The Reverse Cleaning Adapter consists of three parts: A - Syringe Adapter Fitting B Sleeve Fitting C Sprayer Tip Adapter A B C In addition you will need your FMJ-250 High Pressure Syringe and the MicroSprayer Aerosolizer Model IA-1C that is blocked. 1. Soak MicroSprayer Aerosolizer in boiling hot water. To begin, pour a small amount of boiling water into a clean bowl and immerse the very tip of the MicroSprayer Aerosolizer in the boiling water for 1-2 minutes. This will help to soften any material that may have hardened inside. 2. Ultrasound may also be helpful for this purpose. You may safely immerse and sonicate the MicroSprayer for several minutes in hot water or any organic solvent in the ultrasound device. If you prefer, you can place the MicroSprayer inside of a small plastic dish, cover it with boiling water or organic solvent and place the dish inside the ultrasound device. Take note if any material appears to emerge from either the tip or the hub. 3. Remove the protective black cap from the FMJ-250 High Pressure Syringe. Fill the FMJ-250 Syringe with warm filtered water or de-ionized water. Note: Avoid subjecting the syringe to extreme changes of temperature, as this may cause the glass liner inside of it to crack, causing damage that cannot be repaired. 4. Firmly attach the beige Syringe Adapter Fitting (A) to the screw threading of the FMJ-250 Syringe, as shown below. A B C 5. Next, attach the wide end of the long red Sleeve Fitting (B) to the Syringe Adapter (A) as shown above. Tighten both parts firmly. Do not over-tighten. (Note that there will be a gaps between the parts, and the red Sleeve Fitting (B) will NOT cover the screw threads entirely. Penn-Century, Inc. Revised 20 December 2013 2

6. Hold the Sprayer Adapter (C) in your non-dominant hand with the hub end facing you. Note the small hole at the center. 7. In your dominant hand, hold the MicroSprayer Aerosolizer by the thin tubing near the distal tip. 8. SLOWLY, CAREFULLY AND GENTLY, push the MicroSprayer tip through the hole in the hub of the Sprayer Adapter (C). Try to maintain the tubing of the sprayer in a straight line and not at an angle to the adapter. This hole will feel tight and you will feel some resistance at first. 9. Keep your finger tips near to the tip of the MicroSprayer to support the fragile tubing. Apply moderate pressure as you guide the tip through the hole until the MicroSprayer tip is protruding about 2 cm beyond the bottom of the threaded screw portion of the Sprayer Adapter C. WARNING! TAKE CARE! DO NOT HOLD THE MICROSPRAYER BY THE HUB WHEN INSERTING IT INTO THE HOLE IN THE ADAPTER. YOU MAY ACCIDENTALLY PUSH TOO HARD AND CAUSE A SHARP BEND IN THE DELICATE TUBING. THIS WILL DAMAGE IT BEYOND REPAIR! 10. After the tip of the MicroSprayer is inserted in the Sprayer Adapter ( C), remove the protective white cap from the hub of the MicroSprayer Aerosolizer. Penn-Century, Inc. Revised 20 December 2013 3

11. Next, move the tip of MicroSprayer toward the opening of the red Sleeve Fitting (B). Hold only the rim of the Sprayer Adapter (C). Allow the MicroSprayer to move freely. Insert the Sprayer Adapter (C) inside the Sleeve Fitting (B) and screw the two parts together. Tighten both parts firmly. Do not over-tighten. The MicroSprayer will now be held tightly in one position. 12. Hold the syringe and rest the MicroSprayer Aerosolizer hub on a paper towel or clean surface. Slowly begin to push the plunger of the syringe. Push firmly and continuously but do not use excessive force. Notice if the plunger moves forward. Observe if a drop of water passes through the MicroSprayer and leaks out from the hub. 13. At first, you may feel no movement - only resistance. This may give way suddenly or gradually and you may observe drop of water leak out of the hub of the MicroSprayer. 14. If this does not occur and nothing is coming out, do not use excessive force. 15. Be patient. It may take 2 3 attempts to free the blockage. If the first attempt has not worked: a. Detach the Reverse Cleaning Adapter parts from the syringe. Unscrew the Sprayer Adapter (C ) and pull out the MicroSprayer Aerosolizer. Soak the MicroSprayer again in boiling hot water for a few more minutes. b. Refill the FMJ-250 Syringe. If you used warm water the first time, you can now safely fill it with very hot water the second time. c. Or you may try soaking the MicroSprayer in an organic solvent instead of hot water as suggested in our Instructions, and also filling the syringe with organic solvent. d. Repeat steps 4-13. Penn-Century, Inc. Revised 20 December 2013 4

16. If you are able to feel the resistance that is blocking the tip has given way and the plunger can be pushed in with minimal force, remove the MicroSprayer from the Adapter and remove the Adapter parts from the syringe. 17. TEST THE MICROSPRAYER. Fill the FMJ-250 Syringe with filter or de-ionized water, attach the MicroSprayer Aerosolizer to the FMJ-250 Syringe as normal, and test everything. Is the aerosol uniform? Does the device feel as if it is working as it normally does - without undue resistance? If you feel the device is still partially blocked, repeat the Reverse Cleaning steps. If the device feels and looks normal when tested, you may wish to repeatedly flush it with water and/or solvent until you are assured that everything that was blocking it is now fully removed. 18. Note that if air bubbles are trapped in the syringe, you may observe streaming rather than a uniform cloud of aerosol. If so, use the Air Purge Kit provided with your syringe to eliminate any air bubbles, reload and try again. If the resistance in the plunger of your FMJ-250 Syringe will not give way and the MicroSprayer Aerosolizer still appears to be blocked, do not force it. Contact Penn-Century to inquire about our repair services at 215-753-6540 or email us at info@penncentury.com. Penn-Century, Inc. Revised 20 December 2013 5