Liquidator 96 Manual benchtop pipetting system

Similar documents
Operating Instructions. Liquidator 96 Manual benchtop pipetting system

RayPette Plus Autoclavable Pipette. User Manual

The pipettes cover a volume range from 0.1µl to 10ml.

Pipettor. User Manual

User s Manual. Autoclavable & UV r e s i s t a n t. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

User s Manual. Autoclavable & UV resistant. Multi-channel digital micro pipette. In Vitro Medical Diagnostic Devices (98/79/EC)

Operating Instructions. Rainin Classic Manual Pipette

Corning Step-R. Repeating Pipettor. Instruction Manual

MULTICHANNEL. Labnet International, Inc.

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Improving Pipetting Techniques

Varipette Operating manual

Chapter 2. Pipetting techniques. Guide to Pipetting

602 STRINGING MACHINE OWNER'S MANUAL

600 / 600FC OWNER'S MANUAL

Fisherbrand. Finnpipette INSTRUCTIONS FOR USE PARK LANE Pittsburgh PA

Adjust the Volume Display

High-Throughput. Faster, Better 96 & 384-Well Pipetting Ultra Efficient and Extremely Versatile

User s Manual. Digital micro pipette for liquid handling CERTIFIED ISO9001

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

Nichipet EX Plus II. User s Manual. Autoclavable & UV resistant. Digital micro pipette for liquid handling

User s Manual. Autoclavable & UV resistant Nichipet EX II. Digital micro pipette for liquid handling. In Vitro Medical Diagnostic Devices (98/79/EC)

NOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

CP-D/CP-M Circular Polarizer for Phoroptor. User s Guide

Improve pipetting results in pharmaceutical formulation by using MICROMAN E

Portable Massage Table User Manual

453 Series Steam Heated Vaporizing Regulator

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor

PIPETMAN F ENGLISH. Other languages available on our website:

Transferpette. Lightweight precise rugged reliable!

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

OWNER'S MANUAL. Copyright 2003 GAMMA - All Rights Reserved

ROPV R40 E Series User Manual

X-6 STRINGING MACHINE OWNER'S MANUAL. Issue 1 - May Copyright 2004 GAMMA Sports - All Rights Reserved

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

Drawn up by: / RK Revision: 01 Reviewed: Page: 1 of 23 HI-SR 540 USA. Operating instruction SR 540 Face Shield

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

92831 TEL: (714) FAX:

TMP. PepperBall User Manual

This document to be used with Hurley traditional Davits

ali-q TM Gravimetric Verification Procedure

A N B P F G J I C D D L A B C D E F

Standard Collection Setup Manual

AND OPERATION INSTRUCTIONS

A N F G J I C D D L A B C D E F

200 STRINGING MACHINE

User's Manual. Autoclavable Nichipet EX Plus Digital micro pipette for liquid handling CERTIFIED ISO9001

WARNING NOTES TABLE OF CONTENTS

VIAFLO 96. Electronic 96 Channel Hand Held Pipettor

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

STYRIGGER PRODUCT MANUAL

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

OWNER'S MANUAL. Copyright 1999 ATS - All Rights Reserved

Transferpette. Lightweight precise rugged reliable!

Model ASSEMBLY and OPERATING INSTRUCTIONS

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

Other languages available on our website:

HMO500. Headrest Hardware. Stealth s Multi-Axis Headrest Hardware. Installation Manual INSTALLATION INSTRUCTIONS FOR. Stealth Products HMO500 Hardware

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Pipetting Small Volumes

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Transferpette -8/-12

User Manual. ECC-Opto-Gas. Release Electrochemical test cell EL-CELL GmbH

Prepared By: Lisa McAvoy 30/Nov/17 dd/mmm/yy. Approved By: Veronica Harris-McAllister 30/Nov/17 dd/mmm/yy

Tacta Mechanical Pipette User Manual

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

WITOPET PREMIUM PIPETTE USER S GUIDE

Laboratory #2 Pipetting Technique and Micropipette Calibration Skills=40 pts

*******WARNING******

Instructions for use Compressed air injector with water trap

3M Liqui-Cel EXF-4x13 and 4x28 Series Membrane Contactors

DISCOVERY PRO - features

X-6FC STRINGING MACHINE OWNER'S MANUAL. Issue 1 - May Copyright 2004 GAMMA Sports - All Rights Reserved

Remington 700. Disclaimer Limited Warranty Satisfaction Guarantee Installation Instructions

User Manual. Quantos Automated Dosing Liquid Module

Wallmax Professional Aneroid Sphygmomanometer Model 222 User Manual

E-Man Hybrid. Manual pipette with real-time display. Continuously- Adjustable Microliter Pipette with LCD. volume ranges from 0.

INSTRUCTION MANUAL. January 23, 2003, Revision 0

VWR HIGH PERFORMANCE PIPETTOR

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

PCS 340 Core Saturator : 115 Volt : 230 Volt. Instruction Manual

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps

Not a toy. Adult supervision required. Misuse or careless use may cause serious injury or death. May be dangerous up to 500 yards (457 meters).

NEO. User s Guide. Do not lubricate this pipette. The use of lubricant cancels the warranty of this pipette. NOTICE

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

CALIBRATION AND ADJUSTMENT

Air Displacement Pipetting Modules. Total Integrated Liquid Handling

Pipette Tips. Purity and Performance Rainin BioClean Ultra Tips Deliver

Good Pipetting Practice

25/04/56. LTC Carmen A. Bell Chief, Laboratory Operations USAMC-AFRIMS

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

Eppendorf Research fix variable multi Instruction Manual Bedienungsanleitung Mode d emploi Instruzioni d impiego Manual de Instrucciones

Inflatable Stand Up Paddle Board Manual

The pipette of Choice for Problem Liquids!

Transcription:

Liquidator 96 Manual benchtop pipetting system

Liquidator 96 Controls 8 3 4 2 1 5 7 9 10 6 1. Transport lock 2. Pipetting grip & lever 3. Volume selector 4. Volume lock 5. Tip mounting lever 6. Tip rack position 7. Adjustable height post 8. Tip ejector levers 9. Moveable plate holder 10. Adjustable feet (6) Warnings, Safety Notices Liquidator 96 is for research purposes only. If used for other purposes RAININ will not be liable for resulting damage. Always use safe laboratory practices when working with Liquidator 96. Do not work directly under the pipetting head. Move it over to have free access to the working place required. Never move Liquidator 96 without first locking the pipetting head see manual for details. Pipetting strong acids, bases or other aggressive liquids may damage seals, nozzles, or other parts of Liquidator 96. Avoid allowing any such liquids to contact Liquidator 96 or enter the nozzles. Clean up any contamination or accidental splashes of aggressive liquids immediately. Follow the instructions in this manual for safe long-term operation. Keep the manual close to the instrument for reference purposes. It is the responsibility of the user to follow this instruction manual and to work in accordance with the standard operating procedures and general safety guidelines set up for the particular workplace.

Contents Introduction,... 1 Unpacking,Unlocking and Locking Head... 2 Leveling Liquidator 96... 3 Environment... 3 Height Adjustment Posts... 3 Operation... 4 Clean Operation... 9 Recommendations for Pipetting... 9 Maintenance and Safety... 10 Dimensions, Specifications... 10 Ordering Information... 11 Warranty... 11 Appendix - Using the 384-Well Adapter... 12 Introduction Ideal for high throughput applications, Liquidator 96 is a manual benchtop pipetting system with 96 separate channel nozzles in an 8 x 12 microtiter plate standard pattern. Liquidator 96 operates following the same operational principles as a handheld air-displacement pipette except the pipetting head can be moved along a vertical (z) and a horizontal (x) axis for whole-plate pipetting. The pipetting head will only move in the x axis when it is in the fully up position. Two plate holders, each with two working positions, move in a horizontal (y) axis and allow a maximum of three different plates with SBS footprint to be used simultaneously. Adjustable feet at each corner of the base and under each tip loading position allow you to operate Liquidator 96 in a perfectly level position, and provide support to prevent the unit flexing when mounting tips. The rear positions of the right and left plate holders (right side (6) shown opposite) each retain the specially-designed Liquidator 96 Tip Rack. Liquidator 96 works only with Liquidator LTS tips. See page 10 for tip information. Manual, Liquidator 96 1

Unpacking It is recommended that you follow exactly the steps outlined in the separate publication 9920-363, Unpacking/Repacking Instructions, Liquidator 96. It is critically important to follow the steps for repacking whenever the unit is to be shipped or transported to a new location. It is highly recommended to keep all the packaging material. Unlocking the Liquid Head The Liquidator 96 pipetting head incorporates a locking device to prevent uncontrolled movement of the head during movement, say between labs, or from a bench to a laminar flow cabinet. Release a locked head as follows: Pull the black knob of the transport lock (1) on the right side of the device and turn it about 15 so it stays in the out position, sticking out about 1 cm. This allows the pipetting head to move; check free movement by moving the head up and down. Locking the Liquid Head The pipetting head must be secured each time Liquidator 96 is moved, even from bench to bench: Move the pipetting head fully to the right or left and down to the end position and hold it in this position. Pull the black transport lock knob on the right side and turn it about 15 to the locked position. When you release it, you will hear the transport lock click into place, indicating that the pipetting head is safely locked the knob will be in the in position. See below. Never move Liquidator 96 without locking the pipetting head. Any damage to the Liquidator 96 caused by movement with the head unlocked is not covered by the warranty. PRESS LIQUID HEAD DOWN, TURN AND RELEASE LOCKING KNOB 2 Manual, Liquidator 96

Levelling Liquidator 96 Six adjustment feet (10) are provided on the Liquidator 96 base to allow you to level the instrument, and to provide support for tip loading. You can determine whether the instrument is level by placing a plate of water on the base, and adjusting the feet so that the surface of the water in the plate is level. Or check the level by moving the liquid head to the center of the horizontal rail. If the head moves toward one side, twist the adjustment feet on that side clockwise to raise and counterclockwise to lower. Repeat as necessary until the head does not move when released. Environment Liquidator 96 should be placed level in a dry and clean place avoiding direct sunlight or air-conditioner fans. Liquidator 96 can be operated on a bench or in a laminar flow cabinet; the small footprint means you can use it almost anywhere. The recommended ambient temperature is 20-25 C. Any sudden temperature or humidity changes (such as a draft from an air-conditioner fan, or sunlight from a window) may lead to a loss of precision. Before operation allow the instrument to adjust to ambient conditions (temperature and humidity) for at least 24 hours. Equilibration for Liquidator 96 will require a longer time than usual for conventional pipettes. Height Adjustment Posts LIQ-AP The vertical position of the tips can be limited by individual adjustable posts. Using these posts, the tip end height is limited to prevent tip ends reaching the bottom of the plate, e.g. to prevent self-sealing or possible destruction of coatings or cell layers. Screw the bottom of the post into the thread in the plate holders. Then use the adjustment screw at the top end of the post to set the desired height. ADJUSTMENT SCREW AT END OF POST Manual, Liquidator 96 3

Operation Liquidator 96 is designed for simple ergonomic operation similar to conventional manual pipettes from RAININ. It is simple to understand routine operation and to work efficiently with the device. 1. Tip Mounting (photos show right plate use. Left plate similar) a) With the transport lock (1) unlocked, move the pipetting head smoothly up and to its full left or right position. Slide the right (or left) plate holder fully forward and place a new Liquidator 96 LTS tip rack onto the rear working position (6). Make sure that the tip rack is correctly positioned, using the four alignment stubs in the base plate which lock into the channels in the base of the tip rack. See below. ALIGNMENT STUBS IN TIP TRAY b) Now move the pipetting head smoothly to its right or left position. Ensure that the tip rack is correctly aligned and the tray as well as the pipetting head are fully in their end positions. Move the pipetting head down so the nozzles enter the tips. The two levers (8) on the tip ejection mechanism move outwards, indicating that the nozzles of the pipetting head are inserted correctly into the tips. LEVERS Nozzles not seated in tips Nozzles fully seated in tips c) Firmly press down the tip mounting lever (5) to fully mount the tips on the nozzles. 4 Manual, Liquidator 96

This requires a specific amount of pressure on the mounting lever you are mounting 96 tips at the same time. You should feel a distinct lock as the tips are mounted firmly to the nozzles. A quick, sharp insertion is recommended for best results. (If too much force is required, check for proper alignment of nozzles and tips). Always watch the tip ejecting grips (8) when mounting tips! These grips are an indication of proper tip mounting they move outward when you move the pipetting head downwards onto the tip rack to pick up fresh tips, showing that the nozzles are aligned properly into the tips. Do not press the tip mounting lever down (5) until the grips (8) have moved slightly outwards as you move the pipetting head onto the tip rack. As soon as the grips (8) have moved outwards completely, tips are fully mounted and you should release pressure on the tip mounting lever (5). With the tips fully mounted, move the pipetting head up and the plate holder fully back, and place a vessel containing your liquid sample in the working position in front of the tip rack. After doing so the tips should be directly above the sample vessel. 2. Adjustment of pipetting volume The volume is set and indicated by the volume selector (3) which must only be used while the pipetting lever (2) is squeezed. VOLUME SETTING VOLUME LOCK VOLUME DISPLAY Do not adjust the volume selector (3) without squeezing the pipetting lever (2). Manual, Liquidator 96 5

Selecting volumes is done in a similar way as for RAININ manual pipettes. First unlock the volume lock by twisting it counterclockwise a couple of turns. Adjust the volume a few µl higher than the target, then adjust down to the exact volume, to minimize the effects of mechanical backlash. Then twist the volume lock to secure the set volume. Never force the volume selector beyond its limits as this may damage the mechanism. 3. Liquid Aspiration Gently squeeze the pipetting lever (2) until the first stop. Note: Like a conventional manual pipette, the first stop is an indicator that the set volume is ready to be aspirated; the second stop is the blowout position when dispensing. Move the pipetting head down slowly until the tip ends are immersed into the liquid to a depth of about 2-4 mm. Allow the lever (2) to slowly move to the open position while the liquid is aspirated into the tips. Do not release the lever suddenly, which would aspirate sample too quickly and possibly draw air into the tip. Pre-rinsing filling and emptying the tips 2 or 3 times is highly recommended and often leads to better results. If using chilled samples it is recommended to chill the tips to the same temperature before use, for best results. Note: the vertical position of the tips can be limited with the Height Adjustment Post shown on page 3. When the set volume of the sample is fully aspirated, allow the counterweighted pipetting head to rise clear of the sample, as shown below. 6 Manual, Liquidator 96

4. Dispensing Place a 96-well tray on the front tray on the plate holder and move the plate holder fully back for proper alignment. Move the pipetting head to position it over the target plate. Carefully and slowly lower the head so that the tip ends are in or close to the wells but not touching the well bottoms. The height adjustment post can be used to limit the travel of the pipetting head. Dispense the liquid by slowly squeezing the lever (2) to the first stop. Dispensing too quickly may cause splashing. Mixing in the tip can be performed by repeatedly pulling the lever to the first stop and then releasing the lever. The blowout stroke removes the residual liquid remaining in the tips. To blowout, squeeze the lever (2) to the second stop. 5. Tip ejection a) Move the pipetting head over an empty tip rack and downwards until the tip ends slide into the holes of the rack. b) Squeeze together the two tip ejection levers (8) on top of the pipetting head to eject the tips into the used rack. To reuse empty racks do not eject used tips into the tip rack but into another suitable container, e.g. a used deep well plate. Manual, Liquidator 96 7

384-Well Adapter LIQ-384PA This adapter holds a 384 well plate and allows the plate to move into four discrete positions so that all 384 wells can be filled with four passes of a 96-place tip array. To access all the wells the 384-well plate is moved fully into each of the four corners of the adapter. 384-Well Adapter 384-Well Plate in Adapter 1. Place the 384-well plate into the adapter and move it fully into one corner. 2. Carefully align the tip ends into 96 of the wells and dispense 96 samples into the wells. 3. Move the 384-well plate to the next corner. 4. Move the tips to the sample vessels and aspirate sample. 5. Dispense into the next 96 wells. 6. Repeat until all wells are filled. See Appendix (p.12) for more information on 384-well adapter. 8 Manual, Liquidator 96

Clean Operation Liquidator 96 may be operated in a laminar flow cabinet. To prepare Liquidator 96 wipe the surface of the device with alcohol or a suitable non-toxic and non-corrosive laboratory disinfectant. Do not expose Liquidator 96 to UV radiation more than necessary, and never overnight. The radiation and ozone generated may damage the finish and compromise sealing. Do not autoclave any parts of Liquidator 96. Never use autoclaved tips or racks. The tips as well as the racks will shrink slightly during autoclaving, which may lead to nozzle damage. Pre-sterilized Liquidator LTS tips are available from RAININ. Recommendations for Pipetting Before operation allow the instrument to adjust to ambient conditions (temperature and humidity). Pre-rinse the tips - i.e. fill and empty 2-3 times with the same liquid - for better accuracy and precision. Proper pipetting speed depends on many factors, including the nature of the liquids being pipetted. The optimum operation speed for your sample will be discovered by usage. When pipetting small volumes (<10 µl) it is recommended to touch-off the tips against the receiving vessel or liquid surface whilst gently dispensing (by moving the plate holder very slightly forward). Alternatively, dispense the liquid above the destination plate such that it forms a droplet at the end of the tip. By slowly approaching the surface of a plate or the liquid in a plate these drops will be drawn from the tips when approaching to a close distance. Experienced users may also use the so-called tip touch technique from robotic systems: Carefully move the tray with the destination plate under the pipetting head so that the ends of the tips do not touch the center of the wells but instead the side walls. The liquid is then drawn from the tip by capillary action. The tip touch procedure has to be performed very carefully to avoid nozzle damage. Manual, Liquidator 96 9

Maintenance and Safety When used according to the instructions in the manual, and with a little routine maintenance and regular cleaning, Liquidator 96 should provide years of trouble-free operation. Liquidator 96 surfaces can be cleaned with a soft cloth slightly moistened with water or a mild detergent. It is important that the user does not try to dismantle the unit, as specified performance can only be guaranteed if the pipetting head is undisturbed from its factory setting. Dimensions 16 41 cm 14-1/2 37 cm 14-1/2 37 cm 13 33 cm 18 46 cm Specifications Inaccuracy (average over multiple full plates): 200 µl: ± 1.0% 100 µl: ± 1.0% 20 µl: ± 2.0% 5 µl: ± 5.0% Precision: 200 µl: 0.5% 100 µl: 0.8% 20 µl: 1.5 % 5 µl: 3.5% 10 Manual, Liquidator 96

Ordering Information Cat. No. Description Volume Range LIQ-96-200 96-channel Manual Benchtop Pipetting System 1-200 µl Service Plans LIQ-SVC LIQ-SVC4 LIQ-C-A LIQ-C-B Accessories LIQ-AP LIQ-TRAY LIQ-384PA LIQ-384PA-B LIQ-20PA LIQ-20PA-B Annual Service Plan 4-Year Service Plan LIQ 96 CalPM Plan A LIQ 96 CalPM Plan B Height adjustment posts Deep-well 8-section (longitudinal) reservoir 384-well adapter plate stage - white 384-well adapter plate stage -black 20 µl Tip adapter plate stage white* 20 µl Tip adapter plate stage black* BioClean Liquidator Tips Racked tips LQR-200 Liquidator 96 Tips, Racked 5-200 µl LQR-200S Liquidator 96 Tips, Racked, Sterile 5-200 µl LQR-200F Liquidator 96 Tips, Filter, Racked 5-200 µl LQR-20 Liquidator 96 Tips, Racked* 1-20 µl LQR-20S Liquidator 96 Tips, Racked, Sterile* 1-20 µl LQR-20F Liquidator 96 Tips, Filter, Racked* 1-20 µl Stacked tips LQS-200 Liquidator 96 Tips, Stacked 5-200 µl LQS-200S Liquidator 96 Tips, Stacked, Sterile 5-200 µl LQS-20 Liquidator 96 Tips, Stacked* 1-20 µl LQS-20S Liquidator 96 Tips, Stacked, Sterile* 1-20 µl * The maximum volume that can be used with LQR-20x and LQS-20x tips is 20 µl. If volumes greater than 20 µl are used, sample may be aspirated into the liquid head, resulting in damage to the liquid head, expensive repair, and voiding the warranty. As these 20 µl tips are shorter than standard 200 µl tips, you are highly recommended to use a plate height adapter to raise the working plate up to the correct operational height for the tips. Limited Warranty and Limitation of Liability a.) Seller warrants the merchandise to conform to specifications. Under no circumstances shall Seller be responsible for alleged nonconformities with respect to any merchandise which has been used for purposes or in any manner for which it was not intended, or any merchandise which has been customized or modified without Seller s prior written consent, or damaged or misused. As Buyer s exclusive remedy in the event of breach of warranty, Seller shall repair or replace, as its option, any nonconforming merchandise or parts thereof for a period of one (1) year after delivery. All claims must be made in writing to the Seller. Any claims not made within the period specified above shall be deemed waived and released. b.) THE PROVISIONS OF THE FOREGOING WARRANTIES ARE IN LIEU OF ANY OTHER WARRANTY, WHETHER EXPRESS OR IM- PLIED, WRITTEN OR ORAL (INCLUDING ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, TITLE OR NON-INFRINGEMENT). SELLER S LIABILITY ARISING OUT OF THE MANUFACTURE, SALE OR SUPPLYING OF A PRODUCT OR ITS USE OR ITS DISPOSITION, WHETHER BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE, SHALL NOT FOR ANY REASONS EXCEED THE AGGREGATE PURCHASE PRICE PAID BY BUYER FOR SUCH PRODUCT. IN NO EVENT SHALL SELLER BE LIABLE TO BUYER OR ANY OTHER PERSON OR ENTITY FOR SPECIAL, INCIDENTAL, CONSEQUENTIAL OR EXEMPLARY DAMAGES (INCLUDING, BUT NOT LIMITED TO, DAMAGES FOR LOSS OF PROFITS, LOSS OF DATA OR LOSS OF USE) ARISING OUT OF THE MANUFACTURE, SALE, SUPPLY, USE, MARKETING, RESALE OR OPERATION OF THE MERCHANDISE, EVEN IF SELLER HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES OR LOSSES. Manual, Liquidator 96 11

Appendix Using the 384-Well Adapter This adapter holds a 384 well plate and allows the plate to move into four discrete positions so that all 384 wells can be filled with four passes of a 96-place tip array. To access all the wells the 384-well plate is moved fully into each of the four corners of the adapter. Operation: 1. Place the Adapter into the working position, and place the 384-well plate fully right and to the bottom as shown. 2. Aspirate sample, move the pipetting head over the 384-well plate, and dispense into 96 of the wells as shown here. Well number A1 (top left) is the first well filled 12 Manual, Liquidator 96

3. Move the 384-well plate fully left and to the bottom and aspirate the next samples. Move the pipetting head over the 384-well plate and dispense into the next 96 wells as shown. Well number A2 is filled 4. Move the 384-well plate fully left and to the top and aspirate the next samples. Move the pipetting head over the 384-well plate and dispense into the next 96 wells as shown. Well number B2 is filled 5. Move the 384-well plate fully right and to the top and aspirate the next samples. Move the pipetting head over the 384-well plate and dispense into the remaining 96 wells as shown. Well number B1 is filled Manual, Liquidator 96 13

Contacting RAININ Technical Information: T: 800-543-4030 F: 510-564-1617 tech.support@rainin.com Pipette Service T: 800-662-7027 F: 781-935-7631 service@rainin.com Direct Order Line: T: 800-472-4646 F: 510-564-1617 pipets@rainin.com www.rainin.com From outside North America: T: +1-510-564-1600 global@rainin.com (from outside the U.S.) www.mt.com/rainin Rainin Instrument, LLC 7500 Edgewater Drive, Box 2160, Oakland, CA 94621-0060 ISO 9001:2008 Certified a METTLER TOLEDO Company Prices and specifications are subject to change without notice. RAININ and LTS are registered trademarks, and Liquidator and BioClean are trademarks, of Rainin Instrument, LLC. Copyright 2009, Rainin Instrument, LLC. 9920-360 RevC Printed in USA.