Paper Session II-C - Verification of International Space Station Component Leak Rates by Helium Accumulation Method

Similar documents
The latest automotive systems require innovative leak test methods and fixturing.

Japanese Experiment Module ( JEM)

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE

Leak Checking Large Vacuum Chambers

Chapter 6 Leak detection

The Helium Leak Detector

FUGITIVE EMISSIONS EXPERIMENTAL MEASUREMENTS AND EQUIVALENCY

LESER Deutschland Standard Functional Tightness Test (Cryo) Content

GSE CAGING AND SPINUP BUFFER LINE INSTALLATION

GDS OPERATING PROCEDURE

Application Note AN-107

PHOENIX L300i Best-in-class Helium Leak Detectors

CLEANLINESS TEST OF PROBE SPINUP EXHAUST LINES

Commercial Practice Test Method Internal Vapor Analysis of Hermetic Devices

Oxygen Contamination

Finding Leak Testing Success Through Calibration and Validation. Paul Chamberlain President, CEO

Implication of Multiple Leak Tests and Impact of Rest Time on Avionic Hybrids

DESIGN VERIFICATION AND QUALIFICATION OF NEW POPPET AND BUSHING ON GMA FLIGHT SOLENOID VALVES

This test shall be carried out on all vehicles equipped with open type traction batteries.

ACCURATE PRESSURE MEASUREMENT FOR STEAM TURBINE PERFORMANCE TESTING

Memorandum Background: Results and Discussion:

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

GMA TO SPACE VEHICLE PLUMBING HOOKUP

A portable tool to detect leaks in large diameter pipes AGBAR

AC : MEASUREMENT OF HYDROGEN IN HELIUM FLOW

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES

Current Trends in Leak Testing Technology. A. P. Fonseca, H. P. Marques, A. M. C. Moutinho and O. M. N. D. Teodoro

MIL-STD-883K w/change 2 METHOD SEAL

DURRIDGE STAINLESS STEEL SOIL GAS PROBE

CONTAINMENT OF CO2 REFRIGERATION INSTALLATIONS

Procedure of Xenon Transfer

SCRUB LAUNCH RECOVERY PROCEDURE FOR THE GMA

A Conceptual Understanding of Leakage During Soil-Gas Sampling

Memorandum. 1. Introduction. 2. Procedure. To: File ALMA Number: FEND A-TDR. Neil Horner Ralph Groves John Effland.

INTERNAL LEAKAGE CHECK OF THE FLIGHT GMA

GA / 3.02 TL 4 TL 6. Cat. No Calibrated leaks GA / 3.02 TL 4 TL 6. Cat. No

Residual Gas Analysis Systems for Industry

Standard Test Methods for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode 1,2

Protec P3000 (XL) Helium Sniffer Leak Detector

(PTP) PROBE PRESSURE MEASUREMENT SYSTEM INSTALLATION AND LEAK CHECK PROCEDURE

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Test Plans & Test Results

CALIBRATING THE GMA PRESSURE SENSORS AND THE ECU ENGINEERING UNIT

HOW WISE TECHNOLOGY WORKS INFICON Protec P3000 sensor powered by Wise Technology, consists of an evacuated gas cell covered by a quartz membrane.

Proto-qual Vibration Test Procedure 2.5 and 6 Vacuum Valves (3179 and 3223) Retest of Flight Spare Valves. GP-B P0735 Rev.-

Validation of Custody Transfer Metering Skid at Site After Laboratory Proving

SEMASPEC Test Method for Determination of Cycle Life of Automatic Valves for Gas Distribution System Components

EFFECT OF PACKING MATERIAL AND CONFIGURATION ON ISOLATION VALVE STUFFING BOX LEAKAGE AT CRITICAL OPERATING CONDITION

Helium Leak Detection by Detector Probe Technique : Is it a Leak Detection or Concentration Detection Method?

OMNISTAR / THERMOSTAR. The efficient solution for gas analysis. Intelligent software. Wide range of applications.

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Rejuvenation Instructions

Development of Xenon feed system for a 300-W Hall-Thruster

VIC offers a variety of Calibrated Gas Leaks

Results of mathematical modelling the kinetics of gaseous exchange through small channels in micro dischargers

HELIUM PRESSURE REGULATOR

IEP SENSOR BRACKET CHECKOUT AND VERIFICATION. This is a non-hazardous test.

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Succeeding with Production Air Leak Testing Methods. Paul Chamberlain President, CEO

Analysis of Instrumentation Failure Data

MIL-STD-202G METHOD 112E SEAL

Addressing the World Shortage of Helium

GMA VENT EXTENSION SERVICE PROCEDURE

GSE CAGING BALLAST SYSTEM INSTALLATION

Statesville, NC. Results of AM ASHRAE 110 Laboratory Fume Hood Performance Evaluations Venturi Hood

Transfeu WP2 Subtask Development of small-scale test method for fire effluents

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Canadian Light Source Inc. Vacuum Component Leak Test Technical Procedure

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

Inert Air (N2) Systems Manual

GMA PREPARATION FOR E28 TESTING

CLASS D - SENSITIVE LEAK TEST GAS AND BUBBLE METHOD. 1.1 To provide definitive requirements for PNEUMATIC pressure testing of piping systems.

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Model Description

STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA

Industrial Compressed Air. Ultrasonic Leak Detection Survey Guide

Setting a New Standard in Flash Chromatography Performance

Research on the Sealing Detection Technology for No Leak Detection Interface Specimen Yingjun Huanga, Xudong Liaob, Guoyun Baic, Tao Chend, Miao Loue

Hazards associated with the gas system and how to mitigate them

Evaluation of Leak Check Procedures. Suzie Nawikas, Dave Balkenbush H&P Mobile Geochemistry, Inc. AEHS San Diego - March 20, 2013

Advantages of Carrier Gas Leak Detection using Novel Helium or Hydrogen Leak Detectors with Specific Sensor Types

300 Held Drive Northampton Pa

SPIRIT III Radiometer Saturation Effect

3M Steri-Vac Sterilizer/Aerator GSX Series. For Life Science Applications. Product. Profile

Cryogenic system for SCC horizontal test

Practical approach and problems in in-situ RGA calibration

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

METHOD FOR GLOBAL AND LOCAL LEAK DETECTION APPLICATION FOR HYDRAZINE PROPULSION SYSTEMS

Agilent Auxiliary Gas Module AGM 2. User s Guide

Level MEASUREMENT 1/2016

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

Acoustic Emission Testing of The Shell 0f Electromagnetic Valve

Proof of concept of a Bio-Containment System for Mars Sample Return Mission

P-5215 Differential Pressure Transmitter

LUDLUM MODEL 239-1F FLOOR MONITOR. Revised December 2010

FLUID CONTROL. Highly engineered valves Smart valves. Safety. Automotive. Calibration. Pressure. Reliable. Innovative. PU 178-V3

Measuring mass flows in hermetically sealed MEMs & MOEMs to ensure device reliability

Testing of Active Flow Control actuators at harsh environment

On-Stream Tightness Testing of Vacuum Process Installations

Transcription:

The Space Congress Proceedings 2003 (40th) Linking the Past to the Future - A Celebration of Space May 1st, 1:30 PM - 4:30 PM Paper Session II-C - Verification of International Space Station Component Leak Rates by Helium Accumulation Method Steve D. Underwood Sherry L. Smith Follow this and additional works at: http://commons.erau.edu/space-congress-proceedings Scholarly Commons Citation Steve D. Underwood and Sherry L. Smith, "Paper Session II-C - Verification of International Space Station Component Leak Rates by Helium Accumulation Method" (May 1, 2003). The Space Congress Proceedings. Paper 19. http://commons.erau.edu/space-congress-proceedings/proceedings-2003-40th/may-1-2003/19 This Event is brought to you for free and open access by the Conferences at ERAU Scholarly Commons. It has been accepted for inclusion in The Space Congress Proceedings by an authorized administrator of ERAU Scholarly Commons. For more information, please contact commons@erau.edu.

40 th Space Congress 2003 International Space Station Advances, Mission Milestones, Mission Operations Verification of International Space Station Component Leak Rates by Helium Accumulation Method Steve D. Underwood Sherry L. Smith Contact Author: Steve Underwood The Boeing Company Missiles and Space Division International Space Station Program 499 Boeing Blvd. P.O. Box 240002, M/S JT-03 Huntsville, AL 35824-640 Work: (256) 961-1030 steve.d.underwood@hsv.boeing.com ABSTRACT Discovery of leakage on several International Space Station U.S. LAB ammonia system quick disconnects (QDs) led to the need for a quick process to quantify total leakage without removing the QDs from the system. An innovative solution was proposed allowing quantitative leak rate measurement at ambient external pressure without QD removal. The method utilizes a helium mass spectrometer configured in the detector probe mode to determine helium leak rates inside a containment hood installed on the test component. The method was validated through extensive developmental testing. Test results showed the method was viable, accurate and repeatable for a wide range of leak rates. The accumulation method has been accepted by NASA and is currently being used by Boeing Huntsville, Boeing Kennedy Space Center and Boeing Johnson Space Center to test welds and valves and will be used by Alenia to test the Cupola. The method has been used in place of more expensive vacuum chamber testing which requires removing the test component from the system.

TEST METHOD DEVELOPMENT The Space Station Modules have many components, which are leak tested at the bench level before module integration. For various reasons it is sometimes necessary retest the components in place, after integration. The Lab Module Ammonia QDs were originally tested at the manufacturer in a vacuum chamber utilizing a helium mass spectrometer. The vacuum chamber is an accurate method for quantifying leak rates and verifying the QDs met their leakage requirement of 1E-4 sccs helium. However after installation, QD leakage due to seal contamination was detected during Lab Module ammonia system weld testing. The detector probe method used to test the welds is a qualitative method and did not provide the accuracy needed to quantify the QD leakage. It was determined through a series of developmental tests that QD leak rates could be measure in place, by concentrating QD leakage in a containment hood and comparing peak detector probe responses with those of a known leak in the same enclosure. The developmental test setup is shown in Figure 1. A containment enclosure made of plastic and tape is installed over a QD volume simulator. The helium mass spectrometer is configured for detector probe operation and the probe is inserted into the bag for initial background measurement. Calibrated sources of various leak rates are inserted into the bag for a predetermined time period as shown in Figure 2. At the end of the time period the detector probe is inserted into the bag and the maximum mass spectrometer response is recorded. Enclosure sensitivity is determined from this data. Helium Containment Hood Detector Probe QD Volume Simulator Mass Spectrometer Leak Detector Figure 1: Accumulation Developmental Leak Test Setup

Helium Containment Enclosure Standard Leak Source QD Volume Simulator Figure 2: Enclosure Calibration DEVELOPMENTAL TESTING RESULTS Developmental tests were performed to collect accumulation data for various enclosure volumes, leak sizes and accumulation times. Initial development tests were performed with the following standard leaks: 1E-6 sccs He, 4.2E-6 sccs He, 1.0E-4 sccs He, 1.3E-3 sccs He. Sensitivities were determined for the enclosure with each standard leak. Ambient background measurements were taken and each standard leak was inserted into the enclosure for an accumulation time of 15 minutes. After 15 minutes the detector probe was inserted into the enclosure and the maximum mass spectrometer reading was recorded. The enclosure was purged with air before each calibration to ensure a consistent helium background throughout the test. Fig 3 shows the mass spectrometer output resulting from each standard leak. The standard leak size and the resulting enclosure sensitivity are denoted on the figure. More developmental tests were conducted to gather similar accumulation data on different types and sizes of enclosures, and longer and shorter accumulation times, prior to applying the method to flight hardware acceptance testing.

Accumulation Data 1.00E+06 Mass Spectrometer Response (Divisions) 1.00E+05 1.00E+04 1.00E+03 Standard Leak 1E-6 sccs He Standard Leak 4.2E-6 sccs He Standard leak 1E-4 sccs He Standard Leak 1.3E-3 sccs He Sensitivity = 4.5E-9 sccs He/div Sensitivity = 1.0E-8 sccshe/div Sensitivity = 9.5E-9sccs He/div Sensitivity = 1.1E-8 sccs He/div 1.00E+02 0 10 20 30 40 50 60 70 80 Minutes Figure 3: Accumulation Developmental Test Typical Results FLIGHT HARDWARE ACCEPTANCE TESTING Leak testing of the flight QDs on the Lab Element ammonia system QDs began in May of 1999. The method has since been utilized for leak testing welds, heat exchangers, racks, hoses, and will be used to test the Space Station Cupola Module. Several QD problems were identified and corrected through use of the of the accumulation method. Figure 4 shows actual data from Lab Element ammonia quick disconnect (QD-1), with leakage exceeding the 3E-4 sccs He requirement. The first response on the graph is from the 1E-4 sccs He standard leak used for the enclosure calibration. The resulting enclosure sensitivity is denoted on the graph. Calibration accumulation time was fifteen minutes. The ammonia system was pressurized to 500 psig He and allowed to stabilize for three hours. At the end of three hours the enclosure was purged with air, which began the fifteen minute accumulation period. The second response on the graph is from the QD leakage and the leak rate was determined to be 7.3E-3 sccs. Figure 5 shows data from Lab Element ammonia quick disconnect (QD-2), with leakage below the 3E-4sccs He requirement. The test performed on QD-2 was identical to QD-1 and the leak rate was determined to be 5.3E-5 sccs He.

Flight QD-1 1000000 Mass Spectrometer Output (Divisions) 100000 10000 1000 Calibrated Leak 1E-4 sccs He QD-1 Response QD-1 Leak Rate = 7.3E-3 sccs He Enclosure Sensitivity = 4.3E-8 sccs He/div 100 0 5 10 15 20 25 30 35 40 45 50 Time (Minutes) Figure 4: QD-1 Test Results 6000 Flight QD-2 Mass Specrometer Resoponse (divisions) 5000 4000 3000 2000 1000 Standard Leak 1E-4 sccs He Enclosure Sensitivity = 2.2E-8 sccs He/div QD-2 Response QD-2 Leak Rate = 5.3E-5 sccs He 0 0 5 10 15 20 25 30 35 40 45 50 Minutes Figure 5: QD-2 Test Results

CONCLUSIONS AND SUMMARY ISS leak test engineers verified that it is possible to accurately and consistently quantify the leakage rate of pressurized components at ambient test conditions, utilizing a helium mass spectrometer configured in the detector probe mode. Verification was performed on volumes ranging from a few cubic centimeters to 30 cubic feet, and leak rates ranging from 1E-6 sccs helium up to 1E-2 sccs. It is not necessary to know the enclosure volume or the test article volume and it is not critical that the enclosure have a specific leakage as this is taken into account when the sensitivity is determined.