TROUBLESHOOTING. Dwell Volume Revisited. Gradient transfer does not have to be problematic.

Similar documents
Video Notes. LC Troubleshooting Series Retention Time Shifts

AKTA pure 25 New Owner s Intro

GUIDE TO ULTRA-FAST HPLC. Quick Tips for Converting Conventional Reversed-Phase HPLC Separations to Ultra-Fast Separations

Evaluation of linear and step gradient performance and retention time precision of the Agilent 1200 Series Rapid Resolution LC

SHIMADZU LC-10/20 PUMP

Ghost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager

GCMSD-Headspace Analysis SOP

extraction of EG and DEG from the matrix. However, the addition of all diluent at once resulted in poor recoveries.

Running HPLC Methods on an ACQUITY UPLC System

Acclaim RSLC. (Rapid Separation Liquid Chromatography)

Agilent Dimension Software for ELSD User Manual

Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note

Application Note. Rapid performance verification of AZURA systems with refractive index detector. Summary. Introduction

Retention Time Locking: Concepts and Applications. Application

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Gas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems

Agilent 1290 Infinity LC System Applications requiring the Agilent Ultra-Low Dispersion Kit

(2 pts) Draw the line of best fit through the data and estimate the concentration of Fe in your sample solution.

User s Booklet for the Wyatt minidawn Light Scattering Instrumentation

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

Rapid and Reliable Detection of Dissolved Gases in Water

Setting a New Standard in Flash Chromatography Performance

Using the Akta Prime plus October 22, 2012

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

The instrument will process the samples one by one till all the samples have been cleaned up.

Quantitative Analysis of Hydrocarbons by Gas Chromatography

[CROSS COUNTRY SCORING]

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer

International Journal of Generic Drugs ANALYTICAL METHOD PROCEDURES THIS SOP IS 'SWITCHED : OFF : ON'

SOP: Buck Scientific BLC-20P HPLC Operation

Three Columns Gas Chromatograph Analysis Using Correlation between Component's Molecular Weight and Its Response Factor

Carrier Gases in Capillary GC

Experiment GC : Analysis of BTEX by GC-FID

Ensuring Best Possible UHPLC Performance Through an Innovative Capillary and Fitting Design

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

CHM Gas Chromatography: Our Instrument Charles Taylor (r10)

no peaks missing peaks too small peaks increasing retention decreasing retention

Improving the Measurement Accuracy of Water Partial Pressure Using the Major Constituent Analyzer

High speed UHPLC-MS/MS determination of multiple steroids in human serum using the Nexera MX system for multiplex analysis

CONNECTIONS. The Thermal Conductivity Detector

Prep LC Training Manual S.V.

Gases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Trace-Level Analysis of Metanephrines in Plasma by HILIC LC-MS/MS

FROG-4000 TM Chemical Analysis System

[CROSS COUNTRY SCORING]

Prep LC Training Manual

ANALYTICAL SCALE HPLC AGILENT SYSTEMS OPERATIONAL QUALIFICATION

HPLC systems. Overview

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

Organic Elemental Analysis Series II CHNS/O Elemental Analyzer. proven performance. reliable results

LCMSMS Solutions For Steroid Analysis

SRI 8610C Gas Chromatograph Multiple Gas #3 GC configuration

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.

Air Operated Hydraulic Pumping Systems to 50,000 psi

Optimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler

Chromat Calibration Updated October 27th, 2017

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Environmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!

Precisely HPLC. Hints, Tips, Guidelines and Important Considerations. Presenters: Wilhad Reuter / Tom Breslin

Supporting Information for Micro-Collection of. Gases in a Capillary Tube: Preservation of Spatial

Best Practice for Identifying Leaks in GC and GC/MS Systems

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Operation of the Perkin Elmer TGA-GC/MS

Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.

Extended Application Note

Introduction. Experimental

Automated Determination of Dissolved Gases in Water Anne Jurek

More Speed and Resolution for Current Methods

Your local gas generation partner. Precision series Modular gas generation solution for GC.

Multiple Gas#5 GC configuration Jan 2016

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

FIG: 27.1 Tool String

General Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!

Weiyi Zheng, Xiaoxia Dong, Evan Rogers, Jimmie C. Oxley, and James L. Smith*

FROG-4000 TM Chemical Analysis System

Figure 1: Plug in ion inlet of mass spec

APPLICATION NOTE. Fast Analysis of Coal Mine Gas Using the INFICON 3000 Micro GC ABSTRACT

Automated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:

Analysis of Benzenesulfonic Acid and P-Toluenesufonic Acid Esters in Genotox Monitoring using UPLC/UV-MS

Jaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation

Peak Biotech - Company & Product Presentation

DOS in Food Science & Nutrition, Manasagangotri, Mysore

Gas Leak Measurement of TRT Barrel Module

Svensk läkemedelsstandard

CHEMICAL ENGINEERING LABORATORY CHEG 239W. Control of a Steam-Heated Mixing Tank with a Pneumatic Process Controller

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Agilent 1100 Series Analytical and Preparative Scale Fraction Collectors. User s Guide

ASTM 3612/TOGA/Dissolved Gas GC Revised October 2013

Organic Elemental Analysis Series II Nitrogen Analyzer. proven performance. reliable results

Dionex IonPac AS19-4µm

HPLC. Solvent Delivery Module Systems

Lab 3 Introduction to Quantitative Analysis: Pumps and Measurements of Flow

Fast method for Ginseng Analyses using Agilent Poroshell 120 Columns Scaled from a Traditional Method

Module 9. Advanced Data Processing

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

CHE 4115 Chemical Processes Laboratory 2 Experiment 1. Batch Distillation

Fail Operational Controls for an Independent Metering Valve

Air Quality 2b - PFGC/FID Analysis of Benzene and Toluene in Automobile Emissions

Transcription:

458 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com LC TROUBLESHOOTING Dwell Volume Revisited Gradient transfer does not have to be problematic. John W. Dolan LC Troubleshooting Editor This month s installment of LC Troubleshooting is prompted by an error that I made in January s installment (1). A reader had inquired about the source of an observed change in peak width and retention time when moving a gradient liquid chromatography (LC) method from one LC system to another. I misconstructed the example chromatograms of Figure 1 in (1), which confused the associated discussion. What concerns me more than the error is the fact that only two readers brought the error to my attention usually if I make a mistake, I get dozens of e-mails pointing out my error. This made me decide to return to the topic of dwell volume for this month s column. Dwell Volume and Dwell Time When we refer to dwell volume, we mean the system volume from the point at which the mobile phase solvents are mixed until they reach the head of the column. For high-pressure-mixing LC systems, this comprises the mixer, connecting tubing, and autosampler loop as the primary components (Figure 1). Low-pressure mixing systems combine the solvents upstream from the pump, so additional tubing plus the volume of the pump head (or heads) is added to the components of the high-pressure mixing system (Figure 2). Typical dwell volumes for today s LC systems are 1 3 ml for high-pressure mixing and 2 4 ml for low-pressure mixing systems. These volumes can be reduced to 0.5 ml for systems modified for use with mass spectrometry detection (LC MS) or can be 5 8 ml, or even larger, for some of the older equipment still in use. Dwell volume (V D ) is measured easily, as described in the sidebar. Dwell volume is of practical importance only for gradient applications. When mobile phase components are mixed on-line for isocratic separations, the dwell volume still exists, but because the mobile phase concentration is constant, there is no observed difference between chromatograms run on different dwell-volume systems. Gradient methods, on the other hand, rely on a change in the concentration of mobile phase over time to facilitate the separation. The delay created by the dwell volume can make a difference in the appearance of the chromatogram for different gradient systems. We generally report chromatographic retention in units of time, not volume, so it often is convenient to express the dwell volume instead as dwell time. Dwell time (t D ) is obtained by dividing the dwell volume by the flow rate. So a system with a dwell volume of 3.0 ml run at a flow rate of 1.5 ml/min would have a dwell time of 2.0 min. The Chromatogram Example chromatograms for LC systems with 1.5- and 3.5-mL dwell are shown in Figures 3a and 3b, respectively. Note that this is identical to the erroneous Figure 1 of reference 1 except the chromatograms are interchanged. If you keep LC Troubleshooting for future reference, I encourage you to photocopy Figure 3 and paste it over Figure 1 of the January 2006 installment. At a flow rate of 2 ml/min, the gradient reaches the head of the column at (1.5 ml/2 ml/min) 0.75 min in the run of Figure 3a, as

460 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com Controller Injector Pumps Mixer Dwell volume To column Figure 1: Schematic of a high-pressure-mixing LC system showing the components contributing to the dwell volume (inside dashed lines). noted by the arrow in the gradient overlay. In a similar manner, the dwell time for Figure 3b is 1.75 min. As an oversimplified model of gradient elution, we can think of a compound being frozen at the column inlet until a strong enough solvent arrives to wash it through the column. If this were the case, all the peaks in two runs on different dwell-volume systems would be shifted by the difference in dwell time. In the present example, the difference in dwell time is 1.00 min, so one would expect the peaks of Figure 3b to be delayed by 1 min relative to those of Figure 3a. As a first approximation, this is true, especially for more strongly retained peaks. Table I compares the retention times between the two runs, and the last few peaks differ by 0.9 min. However, the first peaks differ by much less than 1 min. This is because some peak migration nearly always takes place during the isocratic hold created by the dwell volume. The longer the hold, the more migration takes place before the gradient reaches the column, so the dif- Circle 23

462 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com Table I: Retention data for chromatograms of Figure 3 Controller Pump Solvent proportioning valve Mixer Injector To column Dwell volume Figure 2: Schematic of a low-pressure-mixing LC system showing the components contributing to the dwell volume (inside dashed lines). ference in retention will be less than that expected just from the dwell volume. This is illustrated in Table I for the first two peaks, which differ in retention by 0.5 min. Because the peaks in the two runs of Figure 3 spend a different proportion of their time under isocratic and gradient conditions, it is not surprising that there are differences in relative retention and, thus, changes in resolution (R s ). The change is a minor change in the depth of Peak Number Retention Time (min) Difference (min) t d 0.75 min t d 1.75 min 1.00 1 3.859 4.300 0.441 2 3.982 4.455 0.473 3 4.448 4.991 0.543 4 5.036 5.645 0.609 5 5.640 6.311 0.671 6 6.086 6.782 0.696 7 7.861 8.670 0.809 8 8.188 9.004 0.816 9 9.659 10.525 0.866 10 10.431 11.314 0.883 11 10.948 11.835 0.887 12 12.287 13.209 0.922 13 12.503 13.415 0.912 14 13.176 14.100 0.924 the valley between the first two peaks in Figure 3; R s 1.05 in Figure 3a and R s 1.18 in Figure 3b. In other cases, the difference can be dramatic. Two Concerns Two concerns arise from the changes in the appearance of the chromatogram, as in Figure 3, when a method is moved between LC systems of different dwell volume. These account for the perceived difficulty of transferring a gradient method from one laboratory to another. First, a shift in retention can mean that the data system settings for one system probably will not work with the other. Peaks are identified based upon their retention times, so the data system is set to recognize a peak if it appears in a narrow retention time window. If the peak retention time has shifted with the system, it is unlikely that it will be eluted Circle 25

464 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com (a) 0 2 4 6 8 10 12 (b) 0 2 4 6 8 10 12 14 Figure 3: Simulated chromatograms for LC systems with (a) 1.5-mL and (b) 3.5-mL dwell volumes. Column: 150 mm 4.6 mm, 5- m d p ; gradient: 50 90% B in 14 min; flow rate: 2.0 ml/min. within the same retention window. This is not particularly significant if the method documentation is written to align the peak detection window with Time (min) the retention time of an injected standard in such cases the adjustment from one system to the other is almost automatic. On the other hand, many methods are written such that the retention time must fall within a specific time window. This makes it much more difficult to transfer a gradient method between two systems. If you develop and document methods, be sure to specify retention relative to a standard, so you do not unnecessarily restrict future use of the method. The second major concern when moving a gradient method between LC systems of different dwell volume is the change in resolution often observed for early eluted peaks. In terms of practical use of the results, resolution generally is more critical than retention. If a change in dwell volume reduces resolution sufficiently that the data are no longer usable, you could be in trouble. One way to avoid this problem is to check the method, before validation, on systems that cover the expected dwell volume range so that it (and the method documentation) can be modified if necessary to work satisfactorily on all LC systems. The simplest way to deal with dwell volume differences between equipment is to develop the method with sufficient Circle 28

466 LCGC NORTH AMERICA VOLUME 24 NUMBER 5 MAY 2006 www.chromatographyonline.com t D Gradient trace t 1/2 Time 50% Figure 4: Dwell volume measurement from a blank gradient. See text for details. Dwell Volume Measurement Determination of dwell volume is a simple measurement that I recommend checking at least once a year on every LC system in the laboratory. The system setup is simple. Use water for the A-solvent and water spiked with 0.1% acetone for the B-solvent. Replace the column with 1 m of 0.005-in. i.d. tubing, set the detector to 265 nm, and set the flow rate so that there is sufficient backpressure for reliable check valve operation (for example, 2 ml/min). Run a 0 100% B gradient in 20 min. The data system output should be a curve similar to the one below. You can measure the dwell time (t D ) by drawing a tangent to the main part of the gradient curve (dashed line in Figure 4) and extend the baseline to intersect this tangent. The time it takes from the start of the program to this intersection is the dwell time. Multiply by the flow rate to get the dwell volume. A simpler technique of measurement can be done from the computer console: measure the offset between the initial and final baselines and determine the midpoint (t 1/2 ) on the gradient curve. The retention time of this point on the gradient equals half the gradient time (20 min/2 10 min in the previous example) plus the dwell time. resolution that it will tolerate the changes encountered on systems of different dwell volume. There are two other approaches that compensate for differences in dwell volume, such that the chromatogram is unchanged when a different system is used. We ll look at these next. Maximum Dwell Volume Methods If you know in advance the largest dwell volume system on which the method will be run, you can build the method accordingly. In the example of Figure 3, let s assume the development system is the 1.5-mL dwell system, and the maximum dwell that will be encountered is 3.5 ml. In this case, just increase the dwell volume of the development system to match the maximum dwell volume, then develop the method in the normal manner. So we would add 2 ml of volume to the 1.5-mL system. This isn t very practical, but there is a simpler, and equally effective, alternative just add the difference in delay time as an isocratic hold at the beginning of each run. So each gradient in our 1.5-mL dwell system would start with a 1.0-min isocratic hold (assuming a flow rate of 2 ml/min). Now, as far as the column is concerned, the two systems of Figure 3 would be identical. The method could be written to allow adjustment of the initial hold so that the true dwell time plus the hold equals 1.75 min. Several years ago, I developed a series of methods for a multinational pharmaceutical company. The challenge was that the largest dwell volume of any system in their laboratories was 4.5 ml, so they made a policy that all gradient methods would be developed with the equivalent of a 4.5-mL dwell. Our LC systems had a 2.3-mL dwell, so we added the equivalent of 2.2 ml of hold for each run. The methods were written so that part of the method setup procedure included adjustment of the hold so that the total equivalent dwell volume was 4.5 ml. The methods transferred easily all over the world. Zero Dwell Volume Methods Another approach to addressing dwell volume differences is to set the dwell volume to zero for all methods. Of course this cannot be done from a plumbing standpoint, but many, if not most, new LC systems have a feature that allows you to inject the sample after the gradient has started. For example, with a 2-min dwell time, you would start the gradient and wait 2 min before injection. This will only work for LC systems that have this capability, but as older systems are replaced, the feature should be more and more common. If this approach is taken, be sure to allow sufficient time for column equilibration before the next injection, because the timing will be a bit different than normal. The main danger of developing methods using this technique is that someone will try to transfer the method to an older system without injection delay capabilities. Summary The practical impact of the system dwell volume on retention and resolution is not something to take lightly. It is unfortunate that many chromatographers ignore dwell volume considerations when developing and transferring gradient LC methods. As a result, they have unnecessary problems with the methods, and gradient elution as a technique gets a bad name as being unreliable. This month, we ve seen that dwell volume is not such a big mystery measure the dwell volume for each LC system and plan ahead for the use of gradient methods on different equipment and you should have few problems. References (1) J.W. Dolan, LCGC 24(1), 30 36 (2006). John W. Dolan LC Troubleshooting Editor John W. Dolan is Vice- President of LC Resources, Walnut Creek, California; and a member of LCGC s editorial advisory board. Direct correspondence about this column to LC Troubleshooting, LCGC, Woodbridge Corporate Plaza, 485 Route 1 South, Building F, First Floor, Iselin, NJ 08830, e-mail John.Dolan@LCResources.com. For an ongoing discussion of LC troubleshooting with John Dolan and other chromatographers, visit the Chromatography Forum discussion group at http://www. chromforum.com.